thermo scientific orion 2111ll sodium monitor · thermo scientific orion 2111ll sodium monitor user...
TRANSCRIPT
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc.
AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher .
1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTO-STIR, Auto-Test, BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, pHuture MMS, pHuture Pentrode, pHuture Quatrode, pHuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable pH, Universal Access are trademarks of Thermo Fisher .
Guaranteed Success and The Technical Edge are service marks of Thermo Fisher .
PerpHecT meters are protected by U.S. patent 6,168,707.
PerpHecT ROSS are protected by U.S. patent 6,168,707.
ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753.
ionplus electrodes and Optimum Results solutions are protected by US Patent 5,830,338.
ROSS Ultra electrodes are protected by US patents 6,793,787.
Orion ORP Standard is protected by US Patent 6,350,367.
Orion NoCal electrodes with stabilized potential patent pending.
© 2007 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
The specifications, descriptions, drawings, ordering information and part numbers within this document are subject to change without notice.
This publication supersedes all previous publications on this subject.
�Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Table of Contents
Chapter I General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-�Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-1Features .and .Benefits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-2Principles .of .Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-3Principles .of .Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-5Fluidics .Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-7Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-8
Chapter II Instrument Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . II-�Unpacking .the .Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .II-1Mounting .and .Plumbing .Instructions . . . . . . . . . . . . . . . . . . . . . . . . . .II-2Electrical .Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .II-4 . . Safety .Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .II-4 . Warning .Labels .and .Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .II-5Wiring .the .Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .II-6Terminal .Assignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .II-8Installation .of .Reagent .& .Diffusion .Tubing . . . . . . . . . . . . . . . . . . . . . .II-9Conditioning .& .Installation .of .New .Sodium .Electrode . . . . . . . . . . . .II-13Installation .of .ATC .probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .II-13Installation .of .New .Reference .Electrode . . . . . . . . . . . . . . . . . . . . . . . .II-14
Chapter III Instrument Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . III-�Description .of .Basic .Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . III-1Description .of .Keypad .Icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . III-2Use .of .Setup .Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . III-3Shutdown .& .Start-Up .Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . III-30
Chapter IV Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IV-�Positioning .of .Valves .for .Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . IV-1Flow .Cell .Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IV-2 .Rinsing .the .Flow .Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IV-3Air .Regulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IV-4Before .Performing .DKA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IV-5Performing .DKA .Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IV-6Calibration .At .Custom .Concentrations .Using .DKA . . . . . . . . . . . . . . IV-13Span .Check .Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IV-14Offline .Calibration .Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IV-15
�Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Chapter V Instrument Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . V-�Weekly .Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .V-1Monthly .Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .V-2Yearly .Preventative .Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .V-5
Chapter VI Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VI-�Diagnostics .Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VI-1Slope .Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VI-9Troubleshooting .Matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VI-11Error/Event .Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VI-14Resetting .the .Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VI-17Serial .Number .and .Software .Revision . . . . . . . . . . . . . . . . . . . . . . . . . VI-18Service .and .Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VI-19
Chapter VII Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .VII-�Notice .of .Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VII-1Declaration .of .Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VII-2Terms .and .Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VII-3
Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-�Mounting .Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1ISE .Default .Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-2Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-3Ordering .Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-7Recommended .Consumables .for .Annual .Operation . . . . . . . . . . . . . . .A-10Recommended .Field .Replaceable .Space .Parts . . . . . . . . . . . . . . . . . . . .A-10Pipette .Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-11Pipette .Techniques . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-12
I-�Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Chapter I General Information
This .manual .covers .operation, .maintenance .and .troubleshooting .for .the .Thermo .Scientific .Orion .2111LL .sodium .monitor, .which .incorporates .state .of .the .art .technology .designed .for .ease .of .use .while .offering .the .lowest .limit .of .detection .available . .
Monitoring .the .sodium .ion .content .of .steam .and .water .circuits .to .produce .accurate .and .reproducible .results .requires .a .very .well .designed .and .maintained .system . . .The .system .must .optimize .the .fluidic .design .with .the .sensing .technology .to .enable .very .low-level .(ppb) .measurement .of .the .contaminants .as .well .as .measuring .across .the .linear .range .of .the .analyzer . .
The .Orion .2111LL .sodium .monitor .meets .all .of .the .criteria .for .accurate .and .dependable .sodium .monitoring .and .more . . .The .2111LL .incorporates .patented .Orion .technologies .including:
• . ROSS® .and .ROSS .Ultra® .electrodes . .• . Newly .developed .flow .cell .design .• . Marquee .help .screen .• . Orion .pump-less .reagent .addition .and .DKA .calibration .system
Orion 2111LL ionplus® Low Level Sodium Monitor
• . . Power• . . Semiconductor• . . Chemical .& .Petrochemical• . . Pulp .& .Paper
• . . Ultra .pure .water• . . Boiler .feed .water• . . Drum .boiler .water• . . Ion .exchange .breakthrough• . . Steam• . . R/O .system
Introduction
Markets:
Applications:
I-�
General Information
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
The .Thermo .Scientific .Orion .2111LL .sodium .monitor .meets .all .of .the .criteria .for .accurate .and .dependable .sodium .monitoring .and .more . . .Our .30 .years .of .sensor .expertise .in .sodium .measurements .combined .with .patented .Orion .technologies .are .skillfully .incorporated .in .the .2111LL .system .
• . Accurate .and .precise .measurements .even .at .the .lowest .levels .of .detection .(range .0 .001 .ppb .to .10 .ppm):
. • . .Reliable, .low .level .measurements .and .wide .range .with .selectable .resolution .
• . Patented .ROSS .Ultra® .reference .and .ROSS® .sensing .electrodes: . . • . .Superior .accuracy .and .stability .over .a .wider .temperature .range .
• . Patent-pending .flow .cell .with .air .stirring .and .sample .air .transport: . • . .Automatic .sample .handling .and .contamination .control .with .no .
moving .parts .
• . Patented .scrolling .marquee: . • . Intuitive .menu-driven .digital .user .interface .
• . Data .log .of .previous .measurements .and .calibration: . • . Measurement, .calibration .and .error .history .
• . Self .diagnostics: . • . Ease .of .maintainability .
• . Password .protection: . • . Security .and .peace .of .mind .for .your .operation .
• . Auto-ranging .electronics .with .large, .backlit .and .easy .to .read . .LCD .display:
. • . Monitor .determines .best .range .
Features and Benefits
I-3
General Information
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
The .sample .enters .the .Thermo .Scientific .Orion .2111LL .sodium .monitor .and .passes .through .inlet .valve, .bypass .valve .assembly, .inlet .filter, .pressure .regulator, .flow .meter, .calibration .shut-off .valve .and .into .the .restrictor .tubing . .The .sample .then .passes .through .the .reagent .manifold .into .a .reagent .bottle .through .a .diffusion .tubing .assembly .where .pH .adjustment .takes .place . .The .pH-adjusted .sample .then .flows .into .the .flow .cell .via .the .diverter .valve, .where .air .is .introduced .from .the .air .pump .to .ensure .proper .mixing .and .fast .response . .The .sample .then .flows .into .an .atmospheric .drain .
Principles of Operation
Monitor
1 2 3 4 5
Power
4-20mA Output
Air Pu
mp
Referen
ce Solu
tion
Main Feed
Cal
Sample
Inlet Valve
Needle Valve& Filter
Regulator Valve
Safety Drain
for DIPA overflow
Cal Shutoff Valve
3 way Right Valve
Restrictor Tubing
3 way Left Valve
Diverter Valve
Recirculation Block
Flowcell Block
Measure Level
Calibration Level Air Filter
Pressurizes flowcell
to initiate siphoning
Calib
rate
Calib
rate
Flow
meter
Mix (m
easure)
Calibration Port
DIPA
Reagent Bottle
Reagent Manifold
Diffusion Tubing
Drain
Sense
REF
TEMP
Fluid Path
Fluid Restrictor Tubing
Electric
Air Path
Reference Solution Path
DIPA Overflow Path
Diffusion Tubing
K E Y
Diverter Valve Positions
Calibration Mode
Sample/Measure Mode
Calibrate
Valve Position
Siphon/ Measure
Calibrate
Valve Position
Mix/Measure
Figure I-� Orion 2111LL Schematic
I-4
General Information
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
The .sensing .electrode .responds .logarithmically .to .changes .in .the .sodium .ion .concentration . . .This .response .is .described .by .the .Nernst .equation: .
E = Eo + 2.3 (RT/nF) log (C/Ciso)
Where: .
E . = .measured .electrode .potential, .mV .
Eo . . = .potential, .when .C .equals .Ciso, .mV .
R . = .ideal .gas .constant .
T . . = .temperature .of .sample, .degrees .K .
n . . = .valence .of .ionic .species .(+1 .for .sodium .ion) .
F . . = .Faraday’s .constant .
C . = .effective .sodium .ion .concentration .(activity) .
Ciso .= . .concentration .(activity) .of .sodium .ion .where .potential .E .is .temperature .independent .(isopotential .point) .
The .above .equation .indicates .that .the .measured .potential .varies .with .both .temperature .and .the .concentration .of .the .ion .of .the .interest . . .In .order .to .eliminate .error .caused .by .fluctuations .in .sample .temperature, .the .2111LL .microprocessor .constantly .updates .temperature .corrections .from .data .supplied .by .the .ATC .probe . .
From .the .Nernst .equation, .the .theoretical .response .of .a .sodium .ion-selective .electrode .to .a .ten-fold .change .in .concentration .at .25 .˚C .is . .59 .16 .mV . . .This .is .referred .to .as .the .electrode .slope .(S) . . .Most .electrodes, .however, .do .not .exhibit .a .theoretical .slope . . .Therefore, .the .instrument .is .calibrated .to .determine .its .actual .value . . .Two .standards .are .used .to .provide .information .necessary .for .the .microprocessor .to .compute .the .actual .slope .and .E0 .for .use .during .sample .analysis . .
In .order .to .eliminate .interference .from .hydrogen .ions, .which .can .become .significant .when .measuring .low .levels .of .sodium, .the .2111LL .adjusts .sample .pH .to .approximately .11 . . .This .pH .adjustment .is .accomplished .by .the .patented .passive-diffusion .process .wherein .the .sample .passes .through .a .length .of .tubing .contained .in .the .reagent .bottle . . .The .reagent .diffuses .through .the .tube .wall .and .mixes .with .the .sample, .raising .sample .pH .to .approximately .11 . .
I-5
General Information
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Principles of Calibration
Calibration .procedures .for .an .analytical .instrument .are .important .and .must .be .performed .carefully . . .The .calibration .procedure .used .in .the .Thermo .Scientific .Orion .2111LL .is .a .variation .of .Double .Known .Addition .(DKA) .using .ROSS .Ultra® .electrode .technology .and .patent-pending .flow .cell .technology .in .combination .with .the .passive .diffusion .system . . .This .method .has .the .distinct .advantages .of .being .fast, .easy, .and .accurate .
Before .calibration .begins, .the .diverter .valve .is .turned .to .divert .flow .from .the .measure .drain .to .the .re-circ .tube, .allowing .the .flow .cell .to .fill . .
The .two .air .valves .serve .dual .purposes .during .the .calibration .sequence .
The .first .function .is .to .stop .the .airflow .through .the .sample, .and .redirect .the .airflow .to .the .headspace .of .the .flow .cell . . .Thus, .the .flow .cell .is .filled .without .causing .air .bubble .disturbance .in .the .sensing .tube . . .The .second .function .ensures .proper .siphoning .operation . . .The .combination .of .these .functions .provides .an .accurate .volume .for .calibration .
At .the .first .step .of .the .DKA .calibration .the .actual .concentration .in .the .sample .is .unknown . . .The .instrument .measures .the .potential .(Es) .and .stores .this .value .in .the .microprocessor . . .A .known .amount .of .standard .1 .solution .is .added .to .the .flow .cell, .which .increases .the .concentration .(Cs) .with .a .corresponding .known .amount .(dC1) . . .During .this .process, .air .is .pumped .into .the .re-circ .tube, .thoroughly .mixing .sample .and .standard .in .a .closed-loop .system . . .The .new .potential .(E1) .is .measured .and .stored .automatically .when .stability .is .reached . . .Adding .standard .2, .preferably .10 .times .more .
Siphon Tube
Siphon Drain
Sensing Tube
Double-Known Addition (DKA)
Figure I-� Flow Cell Volume for DKA
I-6
General Information
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
concentrated .than .standard .1, .increases .the .concentration .(dC2) .in .the .sample .reservoir . . .Again, .the .new .potential .(E2) .is .measured .and .stored .when .stable . . .Now, .we .have .the .following .three .unknowns: .
Es =Eo + S(Ts/298.15) log (Cs/Ciso)
E1 =Eo + S(T1/298.15) log [(Cs + dC1)/Ciso]
E2 =Eo + S(T2/298.15) log [(Cs+ dC1+ dC2)/Ciso]
S .is .the .Slope .at .25 .˚C .(298 .15 .K)
T .is .the .temperature .in .Kelvin, .measured .when .the .potential .E .is .measured .
Es, .E1, .E2 .have .been .determined .during .the .calibration .procedure . . .The .microprocessor .solves .these .three .equations, .to .obtain .the .values .of .S .and .Eo . . .The .calibration .result .is .stored .for .use .during .online .monitoring .to .convert .the .measured .potential .and .temperature .in .the .sample .into .concentration .values .in .either .ppm .or .ppb . .
When .the .calibration .is .complete .the .flow .cell .drains .as .the .sample .flow .returns . . .The .flow .cell .volume .returns .to .the .measurement .level . . .After .allowing .approximately .30 .minutes .for .concentrated .calibration .solution .to .be .flushed .from .the .system, .the .Orion .2111LL .monitor .can .begin .sample .measurement .again . .
In .addition .to .Double-Known .Addition .(DKA), .the .2111LL .also .allows .the .user .the .ability .to .perform .an .offline .calibration .
The .offline .calibration .feature .of .the .Orion .2111LL .allows .the .user .to .adjust .the .monitor .to .values .determined .by .alternate .methods .used .in .their .laboratory .such .as .elemental .spectroscopy .and .ion .chromatography .
It .is .essentially .a .one .point .calibration . .To .perform .offline .calibration, .a .sample .is .taken .from .the .bypass .of .the .instrument; .the .sample .concentration .value .is .stored .in .memory; .the .sample .is .analyzed .by .an .alternate .method .of .choice; .the .previously .stored .reading .is .adjusted .to .the .lab .method .result; .and .the .instrument .is .then .returned .to .the .analysis .mode . . .The .term .“offline .calibration” .refers .only .to .the .fact .that .a .sample .from .2111LL .bypass .is .taken .“offline” .for .laboratory .analysis; .in .fact, .no .downtime .is .experienced .during .the .procedure .and .the .instrument .remains .online .throughout . .
Offline Calibration
I-7
General Information
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Fluidics Diagram
50
40
30
20
10
Scrolling Marquee
LCD Display
Warning Icon
Mode Indicator
Calibration Shut Off Valve
Mounting Hole
Keypad
Left Air Valve
Flow Cell Reservoir
Siphon Tube
Sensing Tube
Flow Cell Block
Siphon Drain
Re-Circulation Tube
Measurement Drain
Diverter Valve
Inlet Valve
Reagent Bottle
Diffusion TubingAssembly
Reagent Bottle Adapter Assembly
Pressure Regulator
Air Pump
Status Indicators
Bypass Valve
Inlet Filter
Flow Meter
Ref. Electrode Filling Solution
Right Air Valve
Restrictor Tube Assembly
BottleBracket
Reagent Manifold
Figure I-3 Fluidics Diagram
I-8
General Information
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Refer .to .Figure I-3 .
Inlet Valve – Accepts .the .sample .stream .via .1/4 .inch .NPTF .connector . . .The .user .must .supply .the .sample .with .a .pressure .between .14 .and .100 .psig . .
Inlet Filter – 60-micron .stainless .steel .filter .traps .particulate .matter .in .sample .stream . .
Bypass Valve – Use .to .redirect .flow .in .bypass .system .
Pressure Regulator – Adjusts .flow .on .incoming .sample .stream .
Flow Meter – Measures .sample .flow .rate .(40 .mL/min .nominal .flow .is .required) .
Calibration Shut Off Valve – Use .to .stop .flow .of .sample .from .flow .meter .to .restrictor .tubing .
Restrictor Tube Assembly – Use .in .conjunction .with .pressure .regulator .to .lower .downstream .pressure .
Reagent Manifold – Directs .sample .flow .in .and .out .of .reagent .bottle .assembly .
Reagent Bottle Adapter Assembly – Connects .reagent .bottle .assembly .to .manifold .
Diffusion Tubing Assembly – Semi-permeable .tubing .through .which .reagent .diffuses .into .the .sample .
Reagent Bottle – Contains .water-soluble .amine, .which .raises .the .sample .pH .to .approximately .11 . .
Reagent Bottle Bracket – Secures .reagent .bottle .
Flow Cell – Contains .reference .electrode, .sample .electrode, .and .ATC .probe . .
Diverter Valve – Allows .the .flow .cell .reservoir .to .fill .during .calibration .by .forming .a .closed-loop .system . .
Re-circulation Tube – Tube .through .which .sample .(assisted .by .air) .is .pumped .into .the .flow .cell .– .used .to .mix .solution .in .a .closed-loop .during .calibration . .
Glossary
I-9
General Information
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Sodium Electrode – Senses .sodium .ions .in .sample .stream .and .produce .an .electrical .potential .dependent .on .sample .concentration .
Reference Electrode Filling Solution – Provides .a .constant .reference .potential .and .completes .the .measurement .circuit .
Reference Solution Bottle – Provides .constant .flow .of .electrolyte .solution .through .reference .electrode .for .maximum .stability .
ATC probe – Measures .sample .temperature .and .inputs .data .to .microprocessor .for .automatic .temperature .compensation .(ATC) .
Calibration Port – Allows .introduction .of .standards .to .sample .reservoir .during .calibration .
Air Pump – Use .to .mix .sample .during .both .measurement .and .calibration .
Right Air Valve – Adjust .airflow .to .flow .cell . . .Controls .airflow .for .mixing .sample .
Left Air Valve – Adjust .airflow .to .flow .cell . . .Controls .airflow .for .siphon .of .sample .
LCD Display – Provides .digital .readouts .of .concentration, .temperature, .millivolts .and .error .codes . .
Keypad – Consists .of .five .mode .keys, .four .prompt .indicator .lights, .two .scroll .keys .and .one .key .for .entering .data . . .Mode .and .error .indicators .are .also .incorporated .on .keypad . .
Status Indicator – 2 .LED .lights .that .illuminate .according .to .current .status .of .the .instrument .
Green Light: indicates .system .is .in .correct .working .condition .
Yellow Light: indicates .a .warning, .system .in .hold, .or .maintenance .is .required .
Red Light: indicates .something .is .seriously .wrong .
Note: When .either .the .yellow .or .red .LED .is .lit, .there .may .be .an .entry .in .the .diagnostic .menus .indicating .the .error . . .Logging .feature .must .be .initiated .in .the .setup .menu . . .Refer .to .Chapter III, Use of Setup Mode .for .instructions . . .s
II-�Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Chapter II Instrument Preparation
The .Thermo .Scientific .Orion .2111LL .sodium .monitor .has .been .assembled, .tested, .and .packaged .with .great .care . .
Report .any .obvious .damage .of .shipping .container .to .carrier .and .hold .for .inspection . . .The .carrier .(not .Thermo .Fisher .Scientific) .is .responsible .for .any .damage .incurred .during .shipment .
1 . . Open .outer .box . . .Remove .top .two .foam .corner .support .pieces . .
2 . . Open .inner .box . . .This .box .should .contain .(1) .each .of .the .following .items:
. • . .2111LL .low .level .sodium .monitor .
. • . .ROSS® .sodium .electrode .box .
. • . .ROSS .Ultra® .reference .electrode .box .
. • . . .Standards .kit: .- . .(2) .2 .oz . .bottles .of .standard .1: . .concentration: .19 .1 .ppm .Na+ .- . .(2) .2 .oz . .bottles .of .standard .2: . .concentration: .192 .ppm .Na+ .- . .(1) .2 .oz . .bottle .of .etching .solution .- . . .(1) .Options .kit: . .cable .glands, .conduit .fitting, .fuse .kit, .green . .
screw .terminal .
. • . Instruction .manual .CD . .
3 . . Remove .the .cardboard .retaining .shell .by .sliding .over .the .entire .mounting .board .and .the .monitor . .
4 . . Carefully .remove .entire .mounting .board .with .monitor .from .inner .box . .
Note: Do .not .lift .or .pull .instrument .by .the .fluidics .or .electronic .components . . .s
5 . . Unbolt .the .monitor .from .mounting .board .by .removing .the .4 .mounting .bolts .with .a .9/16 .wrench . .These .bolts .may .be .discarded .
6 . . Carefully .place .the .monitor .at .a .convenient .location .until .proper .installation .can .be .completed .
Unpacking the Instrument
Figure II-� Unpacking the Instrument
II-�
Instrument Preparation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Refer .to .Appendix A1, Mounting Dimensions.
Warning: Do .not .connect .power .prior .to .mounting .and .plumbing .of .the .instrument . . .s .
• . Select .a .site .for .the .instrument .that .allows .it .to .be .permanently .bolted .with .ample .height .for .atmospheric .drain .operation . . .Be .sure .there .is .ready .access .to .both .electronic .controls .and .flow .cell .calibration .port . .
• . A .clearance .of .15 .inches .(about .40 .cm) .must .be .allowed .above .the .calibration .port . . .Insert .the .pipette .vertically .(not .angled) .during .the .calibration . .
• . The .analyzer .location .must .permit .connection .to .a .plumbed .in .sample .line, .a .drain, .and .AC .power .supply .and .any .connections .for .output .devices . .
• . The .analyzer .should .be .mounted .as .close .to .the .sampling .point .as .possible, .this .ensures .the .fastest .possible .response .to .a .changing .sample .condition . . .Refer .to .Appendix A-4, Sample Conditions .
• . For .proper .flow .cell .operations, .instrument .must .be .installed .straight .and .level .upon .its .mounting .location . . .Failure .to .level .may .cause .poor .siphoning .in .flow .cell .
1 . . Prepare .mounting .holes . . .Carefully .lift .the .analyzer .and .bolt .into .place . .Do .not .lift .the .instrument .by .holding .on .to .any .of .the .plumbing .or .fluid .handling .components .
2 . . Connect .a .waste .line .to .the .outlet .of .the .instrument, .which .is .3⁄4 .NPT .male . . .The .waste .line .should .be .connected .to .a .drain .of .sufficient .capacity, .0 .5 .inch .(1 .27 .cm) .OD .is .recommended .
3 . . Connect .a .sample .line .to .the .inlet .of .the .instrument, .which .is .1⁄4 .NPT .female . . .It .is .recommended .that .a .shut-off .valve .be .installed .at .the .sampling .point .
4 . . The .system .must .be .mounted .and .leveled, .vertically .for .proper .operation .
Mounting and Plumbing Instructions
Recommendations
Instructions
II-3
Instrument Preparation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Additionally .listed .in .Appendix A-3, Specifications.
Sample inlet connection – 1/4” .NPTF . . .If .particulate .matter .is .present .in .sample, .pre-filtration .is .necessary . . .The .60-micron .stainless .steel .filter .located .after .inlet .valve .will .remove .moderate .amounts .of .particulates . .
Flow rate – 40 .mL/min .(nominal) . .
Pressure – 8-100 .psig . . .Consult .Customer .Support .for .Orion .products* .for .details .on .sample .handling .if .pressure .is .outside .range . .
Temperature – Temperature .must .be .between .5-40 .˚C . .
Sodium level – Sodium .levels .are .read .directly .in .ppb .or .ppm, .when .calibrated .with .Orion .standards .1 .and .2 .(Cat . .No . .181140) . .
Sample acidity – Sample .acidity .should .not .be .more .than .50 .ppm .CaCO3 .equivalent . . .For .higher .sample .acidity, .contact .Customer .Support . .for .Orion .products* . .
* Customer Support for Orion Products
. .US .& .Canada: . .800-225-1480 .Dom . .Fax: .978-232-6015 .Int’l . .Fax: .978-232-6031
. www .thermo .com/water .
Sample Requirements
II-4
Instrument Preparation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Warning: Be .sure .to .read .and .observe .the .following .safety .recommendations! . .s
Warning .icon .provides .important .information .that .should .be .strictly .followed .when .using .the .unit .for .your .own .safety . . .Failure .to .follow .these .instructions .may .result .in .injuries .
• . Prior .to .wiring, .a .switch .or .circuit .breaker .for .disconnecting .the .instrument .from .power .supply .should .be .installed .• . .It .should .be .in .close .proximity .to .the .monitor .and .with .easy .reach .of .
the .operator .• . . . It .should .be .marked .as .the .disconnecting .device .for .the .monitor .
• . To .reduce .the .risk .of .shock .hazard, .disconnect .power .prior .to .opening .instrument .
• . Before .connecting .the .instrument .to .the .main, .make .sure .that .the .voltage .lies .within .either .range: .85-132V; .200 .mA; .or .170-264V; .100 .mA; .50-60 .Hz .AC .
• . Cutting .off .the .power .by .disconnecting .power .source .will .not .reset .the .2111LL . . .This .unit .incorporates .a .non-volatile .memory .and .will .maintain .calibration .and .settings .after .power .failure . . .Battery .power .is .supplied .to .the .display .for .the .date .and .time .functions .
• . If .a .repair .should .be .required, .or .to .arrange .a .Return .Material .Authorization, .call .Customer .Support .for .Orion .products .at . .1-800-225-1480 .(Domestic .toll .free) .or .contact .your .authorized .dealer . .
• . Installation .and .wiring .of .the .2111LL .may .only .be .carried .out .in .accordance .with .applicable .local .and .national .codes .per .this .instruction .manual . .
• . Be .sure .to .observe .the .technical .specifications .and .input .ratings .
Electrical WiringSafety Requirements
Recommendations
II-5
Instrument Preparation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Warning Labels and Locations
Warning: Provides .important .information .that .should .be .strictly .followed .when .using .the .unit .for .your .own .safety . .Failure .to .follow .these .instructions .may .result .in .injuries . . .s
Safety .warning .icons .are .used .in .two .locations .on .the .2111LL . . .
• . Faceplate . . .Refer .to .Figure II-2 .
• . Power .supply . . .Refer .to .Figure II-3 .
Replace .fuse .only .with .fuse .of .same .rating .
Figure II-� Faceplate
Fuse HolderFuse
Figure II-3 Power Supply
II-6
Instrument Preparation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Wiring the Instrument
Warning: Be .sure .to .read .and .observe .the .following .requirements! . .If .you .install .the .wrong .fuse .for .your .system, .you .could .damage .the .instrument . . .Make .sure .you .select .the .correct .fuse .rating .and .discard .the .additional .fuses .supplied .in .the .fuse .kit . . .s
• . Options .kit- .fuses, .cable .glands, .conduit .fitting, .and .green .screw .terminal .• . Phillips .head .screwdriver .• . 2 .mm .Blade .flat-head .screwdriver .
1 . . Open .the .faceplate: .Loosen .the .four .screws .using .a .Phillips .head .screwdriver . . .Electronics .faceplate .will .open .via .the .hinge .pin .connection .
2 . . Remove .one .or .two .of .the .two .unused .cable .glands .as .required .for .wiring .power .cable .or .auxiliary .connections . . .Power .cable .optional .hole .locations .in .Figure II-4 .
3 . . Select .and .install .appropriate .size .cable .gland .or .conduit .fitting .as .required .
4 . . Feed .power .cable .through .conduit .or .cable .glands .as .appropriate .
Figure II-4 Electronics Enclosure
with Cable Glands
4 Compression Screws
Hinge Pin
Power Cable Hole Locations
Cable Glands
Electrode Cables
Required Tools
II-7
Instrument Preparation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
5 . . Wire .the .power .cable .to .the .green .screw .terminal .connector .from .the .options .kit . . .Select .correct .terminal .for .hot .conductor .depending .on .line .voltage . . .Refer .to .Figure II-5 .for .terminal .connector .location .and .Figure II-3 .for .terminal .wiring .position.
6 . . Plug .to .terminal .connector .into .power .supply . . .Refer .to .Figure II-3 .
7 . . Select .the .correct .fuse .from .the .fuse .kit . . .Install .by .inserting .fuse . .in .fuse .holder .and .secure .by .twist .and .lock .method . . .Refer .to . .Figure II-3 . . .The .fuses .are .clearly .labeled .with .the .appropriate .voltages .for .your .system . . .Refer .to .Figure II-3 . . .Refer .to .the .table .below .for . .fuse .selection .
AC Voltage Fuse Rating
115V 200mA, 250V, Fast Acting
230V 100mA, 250V Fast Acting
Figure II-5 Terminal Connector Location
Terminal Connector
II-8
Instrument Preparation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Alarms / Relays Analog Outputs Temperature ISE
� Relay 1 (normally open)
9 Sout (mA) Sodium Signal
�8 ATC ��- ��
Not Used
� Relay 1 (normally open)
�0 GND Common Ground*
�9 ATC �3 Electrode Shield **
3 Relay 2 (normally open)
�� Tout (mA) Temp. Signal
�0 Solution Ground
�4 Reference Electrode
4 Relay 2 (normally open)
��-�7
Not Used �5 Not Used
5 Relay 3 (normally closed)
�6 Sodium Electrode
6 Relay 3 (normally closed)
7 Air pump
8 Air pump
Key Electrical Notes for the Orion 2111LL:
• . Fluid .ground .is .provided .through .the .Orion .ATC .Probe .(30 .kΩ), . .Cat . .No . .2100TP . . .Only .an .Orion .ATC .probe .can .be .used .to .provide .the .fluid .ground .
• . Do .not .connect .to .any .of .the .unused .terminals: .12-17, .21-22 .and .25 . .
• . Leave .factory .installed .jumpers .in .place .on .terminals: . .13-15-16 .
• . . Be .sure .to .hook .shield .to .underside .of .PCB .to .prevent .interferences . . .Refer .to .Figure II-7 .
* Terminal 9 & 11 are not isolated from each other. They share terminal 10 as a common ground.
** Do not attach external ground shield connections to terminal 23
*** Units shipped after October 2006 do not require the shield in Figure II-7
Terminal Assignments
Figure II-7 Shield***
Figure II-6 Terminal Assignments
II-9
Instrument Preparation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Warning: Be .sure .to .read .and .observe .the .following .requirements! . .Diisopropylamine .is .hazardous . . .Refer .to .MSDS .for .DIPA, .available .at .www .thermo .com/water . . .Use .protective .glasses .and .gloves . . .Refer .to .bottle .label .for .precautions . . .Installation .of .DIPA .reagent .requires .a .fume .hood .or .well-ventilated .area . . .s
If .monitor .is .offline .or .sample .is .shut .off .from .the .instrument .for .a .prolonged .period .of .time .be .sure .to .follow .the .Chapter III, Shutdown & Start-Up .Procedure .
1 . . Unscrew .thumbnut .and .remove .reagent .bottle .from .the .reagent .manifold . .• . .New .instruments .will .ship .with .a .rubber .coated .glass .reagent .bottle .
and .complete .diffusion .tubing .assembly .installed . . .Only .installation .of .reagent .is .necessary .for .first .time .start .up .
2 . . Twist .open .reagent .bottle, .separating .glass .bottle, .reagent .bottle .adapter .assembly .and .plastic .cap .to .reagent .bottle .
3 . . Unscrew .the .fittings .from .base .of .the .reagent .bottle .adapter .assembly . .dispose .of .used .diffusion .tubing .assembly .accordingly .
The .new .diffusion .tubing .assembly .for .use .with .the .Orion .2111LL .low .level .sodium .monitor .provides .superior .locking .ability .to .maximize .uptime .and .prevent .against .costly .blowouts . . .The .newly .designed .HDPE .locking .mechanism .is .precisely .engineered .to .remain .connected .under .variable .temperature .conditions .and .pressure .fluctuations . . .The .new .style .diffusion .tubing .assembly .is .now .suspended .in .the .headspace .of .the .reagent .bottle- .decreasing .the .rate .of .consumption .of .reagent-saving .you .time .and .money .
The .diffusion .tubing .ships .fully .assembled .for .quick .and .easy .installation . .Occasionally .the .assembly .will .loosen .during .shipment . .
• . Be .sure .to .check .the .diffusion .assembly .for .completeness .of .components .prior .to .installation .upon .the .reagent .bottle .cap . .
• . Do .not .use .Teflon .tape .during .installation
• . Twisting .of .the .clear .silicone .tubing .will .compromise .tubing .integrity .and .may .cause .ruptures . . .Be .careful .to .avoid .kinks .and .twisting .during .installation .
Installation of Reagent & Diffusion Tubing
Recommendation
Diffusion Tubing Assembly
Tips Before Installation
Part Description Quantity
O-ring .seal .fitting 2Twist .and .lock .Cap 2Pipe .Fitting 2O-ring .EPDM 6Ferrule 4Silicone .Tubing 1
II-�0
Instrument Preparation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Figure II-8 Diffusion Tubing Assembly
1 . . Make .sure .that .ferrule .is .in .correct .orientation- .large .end .of .ferrule .is .at .end .of .PVC .barb . . .The .ferrule .should .also .be .connected .to .diffusion .tubing .PVC .barb .completely . . .See .Figure II-8 .
2 . . Carefully .thread .the .O-ring .seal .fitting .and .ferrule .tightly .into .the .reagent .bottle .adapter .to .create .a .suitable .seal . . .Fitting .can .be .tightened .with .a .wrench .
3 . . Slide .down .and .seat .O-ring .in .top .of .seal .fitting . . .See .Figure II-9a .
4 . . Slide .twist .lock .cap .down .over .O-ring .fitting .compressing .and .rotate .90 .degrees .to .lock .in .place . . .See .Figure II-9b .
5 . . Slide .second .O-ring .down .and .seat .in .top .of .the .twist .lock .cap .recess .
Installation
Ferrule (2)
Pipe Fitting (2)
O-ring EPDM (4)
O-ring Seal Fitting (2)
Twist & Lock Cap (2)
Silicone Tubing
a. b.Figure II-9a & II-9b
II-��
Instrument Preparation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
6 . . Repeat .steps .1-5 .for .the .other .side .of .bottle .adapter . . .Final .assembly .should .appear .as .shown .in .Figure II-10a . .
. Avoid .twisting .or .kinks .in .Silicone .tubing .prior .to .installation .in .diisopropylamine .reagent .
7 . . Carefully .bend .tubing .to .secure .over .hook .on .reagent .bottle .adapter . .See .Figure II-10b . .
• . Check .to .make .sure .that .O-ring .seal .fitting .is .tight .and .there .is .not .a .large .gap .between .the .reagent .bottle .adapter .
• . Check .to .make .sure .that .the .twist .lock .cap .is .in .correct .orientation- .rotated .90 .degrees .from .flat .surface .of .O-ring .seal .fitting . .
• . Check .to .make .sure .that .ferrule .is .oriented .in .the .proper .direction- .large .end .of .ferrule .is .at .end .of .PVC .pin . . .See Figure II-8 . .• . This .ferrule .may .fall .off .in .bag .during .shipment- .2 .additional .
ferrules .have .been .added .to .the .packaged .assembly . . .Be .careful .to .remove .the .additional .ferrules .from .packaging .before .discarding .
1 . . Secure .the .diffusion .tubing .assembly .to .the .base .of .the .reagent .bottle .adapter .assembly .by .tightening .the .fittings . . .The .fittings .must .be .perpendicular .and .straight .when .tightened .into .adapter .assembly . . . .Refer .to .Figure II-8 .• . Be .sure .to .prevent .twisting .of .the .diffusion .tubing .while . .
tightening .fittings . .
2 . . Place .a .new .reagent .bottle .on .a .flat .surface, .and .remove .the .cap . . .Ventilation .or .a .hood .is .recommended .for .this .step .
3 . . Holding .the .bottle .steady .with .one .hand, .install .the .reagent .bottle .adapter .assembly .by .tightening .its .cap .onto .the .bottle . . . .• . Ensure .that .the .bottle .label .is .oriented .so .that .it .will .be .visible .when .
the .entire .assembly .is .mounted .to .the .instrument .
b.a.Figure II-�0a & II-�0b
Troubleshooting
Reagent Bottle Installation
II-��
Instrument Preparation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
4 . . Replace .large .single .O-ring .in .reagent .bottle .adapter .assembly . . .Use .needle .nose .pliers .if .necessary .
5 . . Replace .all .3 .O-rings .on .face .of .reagent .manifold . . .Refer .to . .Figure II-11 .
6 . . Slide .the .reagent .bottle .adapter .assembly .over .the .threaded .screw .of .the .reagent .manifold . . .Refer .to .Figure II-11 .
7 . . Tighten .the .thumbnut .to .secure .
Figure II-�� Reagent Manifold & Reagent
Bottle Adapter Assembly
(3) O-rings for Manifold
Manifold Screw
O-ring for Bottle Adapter
Diffusion Tubing
Reagent Bottle
Reagent Bottle Adapter
Reagent Manifold
II-�3
Instrument Preparation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
The .Orion .ROSS® .sodium .electrode .(Cat . .No . .210048) .must .be . .used .in .conjunction .with .the .Orion .Ross .Ultra® .reference .electrode . .(Cat . .No . .210058) .
Warning: Be .sure .to .read .and .observe .the .following .requirements! . .Only .sodium .sensing .electrode .is .etched . . .Do .not .etch .reference .electrode . . .Safety .glasses .must .be .worn .during .entire .conditioning .procedure . . .Gloves .must .be .worn .while .etching .the .electrode . . .s
1 . . Unpack .the .sodium .electrode .(Cat . .No . .210048) .carefully .remove .the .protective .cap . . .Save .the .cap .for .future .storage .of .the .electrode . .
2 . . Immerse .the .sodium .electrode .in .a .bottle .of .etch .solution . .(Cat . .No . .181113) .for .one .minute . .
3 . . Remove .the .electrode .from .the .bottle, .and .rinse .it .with .de-ionized .water . . .Wipe .the .bulb .with .clean .lab .tissue .paper . . .Rinse .again . .
4 . . Insert .the .sodium .electrode .into .its .port .in .the .flow .cell .cap . . .Refer .to .Figure II-12 .for .location . .
5 . . Plug .the .electrode .cable .marked .“Sensing” .into .the .top .of .the .electrode . .Be .sure .to .push .back .black .cap .to .verify .a .secure .connection .between .male .and .female .pin .connection .prior .to .tightening .
6 . . Tighten .screw .cap .connection .to .cable . .
Note: Do .not .twist .cables .while .tightening .connection . . .Twisting .may .cause .damage .requiring .premature .replacement .of .the .cables . . .s
7 . . Wait .at .least .one .hour .before .calibrating .analyzer . .
Refer .to .Figure II-12 .for .location .
The .automatic .temperature .compensation .(ATC) .probe .(Cat . .No . .2100TP) .is .already .connected .to .the .correct .terminal .for .temperature .upon .delivery . .Insert .ATC .into .its .port .in .the .flow .cell .cap . .
Conditioning & Installation of New
Sodium Electrode
Air Flow Barbs
Flow Cell CapCalibration Port
Sensing Electrode Port
Reference Electrode PortATC Port
Figure II-�� Flow Cell Cap
Installation of ATC Probe
II-�4
Instrument Preparation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
1 . . Unpack .the .reference .electrode .(Cat . .No . .210058) .and .its .instruction .sheet .from .the .shipping .box . . .
2 . . Carefully .remove .protective .caps .from .bottom .and .sidearm .of .electrode . .Save .caps .for .future .storage .
3 . . Shake .out .as .much .of .the .fill .solution .through .the .fill .hole .as .possible . .Drain .through .the .sidearm .or .use .a .pipette .or .syringe .if .needed .
4 . . While .passing .1/8 .inch .tubing .into .electrode .sidearm, .slide .1⁄4 .inch .tubing .over .sidearm . . .The .outside .tubing .should .extend .3/8- .1⁄2 .inches .over .sidearm . . .Refer .to .Figure II-13 .
5 . . Remove .cap .and .fluid .seal .from .reference .electrode .filling .solution .bottle .(Cat . .No . .150072) . . .Hold .bottle .in .upright .position . . .Check .that .rubber .gasket .is .properly .aligned, .then .connect .cap .end .of .tubing .assembly .to .bottle . . .The .1/8-inch .tubing .should .extend .into .the .bottle .
6 . . Hold .reservoir .bottle .above .electrode .with .bottle .cap .end .down . .Electrode .should .be .horizontal .with .sidearm .pointing .up . . .Gently .shake .electrode .to .allow .any .trapped .air .bubbles .to .rise .into .bottle .as .electrode .fills .with .solution .
7 . . Dry .off .ceramic .frit .on .base .of .the .electrode .with .tissue .paper . . .Squeeze .bottle .for .a .few .seconds . . .A .small .amount .of .filling .solution .should .bead .up .on .frit .surface, .indicating .good .fill .solution .flow . . .If .no .moisture .is .visible .electrode .is .clogged .and .should .be .replaced .
Installation of New Reference Electrode
Figure II-�3 Reference Electrode
with Filling Solution
Reference ElectrodeFilling Solution
Gasket
Tubing Bottle Adaptor
Small 1/8” Tubing Inside
Bottle Cap
Large 1/4” Tubing
Electrode Sidearm
ROSS™ Coil
Ceramic Frit
II-�5
Instrument Preparation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
8 . . Invert .electrolyte .bottle .and .snap .into .clip . . .Refer .to .Figure II-14 . . .Use .pushpin .supplied .to .puncture .three .air .vents .on .bottom .of .bottle . .
Caution: Failure .to .vent .fill .solution .bottle .will .lead .to .noisy .and .drifting .output .signals . . s
9 . . Plug .the .electrode .cable .marked .“Reference” .into .the .top .of .the .electrode, .and .tighten .the .screw .cap . . .Be .sure .to .push .back .black .cap .to .verify .a .secure .connection .between .male .and .female .pin .connection .prior .to .tightening .
10 . .Tighten .screw .cap .connection .to .cable . .
Note: Do .not .twist .cables .while .tightening .connection . . .Twisting .may .cause .damage .requiring .premature .replacement .of .the .cables .
11 . .Insert .the .reference .electrode .into .its .port .in .the .flow .cell .cap . . .Refer .Figure II-12 .for .location . .
Figure II-�4 Reference Mounting Clip
Air Vents
Bottle
Clip (on Panel)
III-�Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Chapter III Instrument Operation
Description of Basic Unit Controls
Figure III-� Faceplate
B
A
CD
FE
Parameter Location on Display Options Default
A Mode Indicator Top right corner Hold, Cal, Setup, Measure, Diagnostic Measure
BMarquee Display Top left corner Customer prompts through scrolling
messagingNo ATC connected 25 ˚CATC connected: measured at actual
BTemperature Display Top left corner Celsius Measured at actual
CMain Data Display
Middle Line:
Lower Line:
Concentration
mV
conc in ppb9999: At initial installation if no electrodes are connected: Connect electrodes and calibrateLower line will be disabled at installation. Refer to use of setup mode to initiate lower line visibility.
D Measurement Units Left & right side of display ISE concentration ppb/ppm Auto ranging: ppb - ppm
ESystem Status Yellow /Green Indicator Light:
Left side of display above View icon
Green: System OKYellow: System warning
Yellow: Needs calibration
FSystem Error Red Indicator Light:
Right side of display above Setup icon
Red: System failure Red: At initial installation, needs electrodes installed and calibration.
III-�
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Icon Parameter/Mode Action Operational Selections
Enters calibration mode Calibration mode with user prompts DKA, Offline
Scroll up numbersScroll up in cal & setup mode
Used to edit valuesUse to select available modes by using up arrow.
0-9 selectable by digit, first digit selectable 0-19
Enters setup mode System setup mode at the last parameter used by the customer.
DKA, Password, Date, Time, Log, Hold, Reset, Display, Measure, Temp, Alarm, mAmp, Test, Calibration
Move cursor to the right digit Used to edit values When move to final digit, system will wrap around to first digit
Enters test mode Use enter key to advance through sequence of displaysDiagnostics will appear in top right of screen
Last calibration slope & Eo, error list, log points by time & date, error log by time & date, software rev., serial #, displays measured mV and noise, mAmp output values (sensor & temp), and display test
Scroll down numbersScroll down in cal & setup mode
Used to edit values that can be changed by the operatorUse to select available modes by using down arrow
0-9 selectable by digit, first digit selectable 0-19
view Log-changes between date and data that was logged at that specific time interval
Use to view data in calibration, measurement & status logs
+/- Toggles negative/positive sign when editing numbers
enter function Accepts value or selection displayed on LCD, stores in memory
enter (in test mode Only) Used in test mode to display additional information for selected menus
exit to measure Exits from setup or cal modes and puts the monitor into sample mode.
Instrument automatically enters sample mode when first turned on and after calibration.
last Screen Used in test and setup mode User may move back to previous menu. Can not be used in calibration mode: If exit calibration, user must re-do calibration
decimal point Sets the decimal point position in some menus with decimal point numbers between toggles ppm & ppb.
Used in set decimal position
Description of Keypad Icons
Figure III-� Keypad Icons
III-3
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Before .first .sample .measurements .on .the .Thermo .Scientific .Orion .2111LL .sodium .monitor .can .be .performed, .a .successful .calibration .must .be .performed .and .accepted .by .the .user .and .stored .in .the .memory .
• . Press . .to .enter .setup .mode .• . Setup: .appears .in .mode .indicator .• . Hold: .displayed .while .in .setup .mode .• . System .returns .to .the .last .parameter .used .by .the .customer .
• . Press . ./ . .to .loop .through .14 .menu .selections .
• . Press . .to .select .the .desired .menu .selection .and .enter .submenu .
• . Press . ./ . .to:• . Toggle .between .On/ .Off .in .selected .sub-menus . .• . Scroll .and .set .first .digit .value .from .0-19• . Scroll .and .set .remaining .digit .values .from .0-9
• . Press . .to .scroll .right .to .set .each .digit .(4 .digits .total) .
• . Press . .to .exit .current .screen .and .return .to .previous .screen . .
• . Press . .to .exit .setup .menu .and .return .to .measurement .mode . . .If . .is .pressed, .pressing . .again .will .not .return .the .user .to .the .setup .
mode . . .User .must .enter .into .setup .mode .by .pressing . .
Use of Setup Mode
Navigating tips for the setup mode
Once in a submenu of the selected setup menu
III-4
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default password: .0000 .— .indicates .password .protection .has .not .been .activated .
System password: .Management .secured .password .protection .of .setup .and .calibration .process .
Calibration password: .Operator .secured .password .for .protection .of .calibration .process .only .
If Password(s) are activated:
• . System .prompts .user .to .enter .system .password .• . Marquee: .ENTER .PASSWORD• . Main .display .top: .PASS• . Main .display .bottom: .0000 .– .flashing
• . Correct .password: .allows .user .to .enter .setup .mode .for .custom .programming .options .
• . Incorrect .password: .Password .incorrect .or .not .entered .correctly .• . System .password
• . Marquee: .SYSTEM .PASS .INCORRECT• . Main .display: .E035 .
• . Calibration .password• . Marquee: .CAL .PASS .INCORRECT• . Main .display: .E034 . .
• . Verify .password .and .re-enter .password .
If Password(s) are de-activated:
• . System .enters .the .last .set .up .menu .screen .used .by .the .user . .• . Marquee: .Flashes .current .menu .selection .• . Main .display: .SEL .SCrn
Password protection
III-5
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Features 14 programmable menus: Order .of .menu .selection .is .dependent . .upon .the .direction .the .customer .scrolls .when .pressing . .or . . . .Listed .below .using . .
Set .either .of .2 .password .options .• . .System .password .– .Setup .settings .protected . .Accessed .by .authorized .
users .only• . .Calibration .password .– .Calibration .menu .data .is .protected . .Accessed .
by .authorized .users .only• . .Default .password: .0000 .– .Disables .both .passwords• . Forgot .your .password? .Contact .Customer .Support .for .Orion .
products .at .800-225-1480
Set .date .in .US .or .Europe .format .• . .Enter .month, .day .and .year• . .Default .date: .System .will .continue .to .keep .date .and .time .due .to .
battery .back .up . .Customer .must .set .in .accordance .to .local .time .zone .• . If .the .battery .is .removed .the .system .will .show .01/01/2000
Set .instrument .time .• . .Enter .hour .and .minutes• . .Default .time: .System .will .continue .to .keep .date .and .time .due .to .
battery .back .up . .Customer .must .set .in .accordance .to .local .time .zone• . If .the .battery .is .removed .the .system .will .show .00:01
Set .data .logging .interval .for .Measure .Log . .Calibration .and .Error .logs .are .accessed .in .Test Mode .section .
• . .Set .log .interval .– .Hour: .Minutes• . .Measure .log .(Sens)• . .System .log .(Stat)• . .Default .log .interval: . .00:00 .– .logging .disabled• . Minimum .log .interval .is .5 .minutes
Set .time .system .remains .in .hold .before .system .will .display .actual .measurements .after .calibration .
• . .Once .hold .time .expires, .system .implements .any .custom .programmable .changes .to .the .setup .parameters .from .the .defaults
• . .After .calibration, .hold .allows .rinse .down .of .flow .cell .prior .to .recording .actual .sodium .values
• . .Default .hold .time: .30 .minutes• . In .setup .mode . .will .release .hold
Setup Mode
Password
Date
Time
Log
Hold
III-6
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Reset .the .instrument .to .factory .defaults .for .setup .parameters .• . .Used .to .troubleshoot .system . .(A .hard .reset .can .be .used .if .the .keypad .and .software .is .not .
responding .to .touch . . .Refer .to .Chapter VI, Resetting the Computer) .
Warning: . .Resetting .the .computer .will .lose .all .stored .information .including .relay, .logs .and .calibration .settings . . .s
Controls .the .automatic .lighting .options .for .backlit .display .• . .Auto .– .system .will .toggle .brightness .in .response .to .ambient .light .
source• . .On .– .backlit .display .always .on• . .Off .– .backlit .display .always .off• . .Default: .auto
Sets .significant .digits, .mV, .and .concentration .units .displayed .when .in .measure .mode . .
Set number of significant digits displayed.• . .Toggle .between .2, .3, .and .4 .significant .digits• . .Default .significant .digits: .3Enables 2nd LCD line (mV) values to display. • . .Toggle .between .mV: .On .or .Off• . .Default .mV: .OffSelect concentration unit for sodium values.• . .Auto .– .auto-ranges .from .ppb .to .ppm• . .ppb .– .part .per .billion .of .sodium .ions• . .pp .– .parts .per .million .of .sodium .ions .(pp .= .ppm)• . .Default: .auto .ranging
Allows .user .to .adjust .the .temperature .reading .from .ATC .probe .by .± .1˚C• . .Default: .0 .0 .degrees .C
Reset
Display
Measure
Temp
III-7
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Sets .up .to .3 .alarms: .high, .low, .and .an .error .signaling .contact .• . .Relays .1 .and .2 .(RLY1, .RLY2) .are .normally .open .dry .contacts .used .
to .set .high .and .low .alarms .for .measurement .values• . .RLY .1 .options: .Off, .Hi, .Low• . .RLY .2 .options: .Off, .Hi, .Low• . .Relay .3 .(RLY3) .is .a .normally .closed .contact .that .can .be .dedicated .
to .errors . .This .will .close .if .power .to .monitor .is .lost . .This .alarm .is .effected .by .calibration, .errors .and .off-line .or .hold .status
• . .RLY .3 .options: .Off, .Cal, .Hold, .Err• . .Default: .all .alarms .(1, .2, .& .3) .are .set .to .off
Set .either .of .2 .analog .current .outputs .(Sout .& .Tout), .set .scale .values .as .linear .or .logarithmic .curves .for .concentration .
Analog outputs: • . .Toggle .between .4-20 .mA .or .0-20 .mA .current .signals .• . .The .outputs .share .a .common .return, .but .are .isolated .from .the .main .
circuitry .of .the .monitor .• . .Default .output .current: .4-20 .mA• . .Sout: .dedicated .sodium .output .• . .Tout: .dedicated .temperature .output .• . .Toggle .between .logarithmic .(Log) .and .linear .(Lin) .scale .for .each .• . .Set .Low .(LO) .and .high .(HI) .limits .for .Sout .and .Tout .• . .Default .scale .values: .for .both .Log .and .Lin .
Default Concentration:• . .Low .limit: .1 .0 .ppb• . .High .limit: .100 .ppb
Default Temperature:• . .Low .limit: .5 .0 .degrees .C• . .High .limit: .45 .0 .degrees .C
mAmp
Alarm
III-8
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Allows .user .to .test .relays .and .analog .output .lines .
• . .User .selectable .method .to .activate/deactivate .relays .and .outputs .to . .be .tested
• . .Verify .the .accuracy .of .the .analog .outputs .lines .if .used .with .an .external .loop .calibrator
Diagnostics: now .indicated .in .mode .window . .Will .provide .the .following .test .values/settings .for: .mAmp .output, .relays, .and .the .air .pump . . .s
mAmp output:• . 4-20 .or .0-20• . Sout .= .LO .and .HI .values• . Tout .= .LO .and .HI .valuesRelays: • . RLY .1 .= .Toggle .On/Off• . RLY .2 .= .Toggle .On/Off• . RLY .3 .= .Toggle .On/OffAir pump: • . AIR .= .Toggle .On/Off
Set .calibration .frequency .in .hours .(h) .
• . High .limit: .19999 .h .• . Low .limit: . .00000 .h• . Default: .720 .hours
User .selectable .values .for .customized .Double .Known .Addition .(DKA)
Programmable .for .volume .(mL) .of .flow .cell, .concentration .(ppm) .of .standard .1 .and .standard .2, .and .volume .and .volume .(mL) .of .additions .
• . Default .flow .cell .volume: .190 .0 .mL• . Default .concentration .(std .1): .19 .1 .ppm• . Default .volume .addition .(std .1): .0 .10 .mL• . Default .concentration .(std .2): .192 .0 .ppm• . Default .volume .addition .(std .2): .0 .10 .mL
Test
Cal
DKA
III-9
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
SETUP
Press to enter setup mode• Setup: appears as mode indicator
mode window• Hold: displayed while in Setup mode• System will enter setup mode at last
used menu
PASSDATETIMELOGHOLDRSETDISPMEASTEMPALRMmAMPTESTCALDKA
SELSCrn
Press to loop through 14 menu selections.
Press to select the desired menu selection and enter submenu to customize setup parameters
III-�0
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
PASSPASS(flashing)
SELSCrn
Press to set new system
and calibration passwords
0 0 0 0 SET-UP NEW SYSTEM PASSWORD
PASS# # # #(All digits flashing)
Press / to set first digit SET-UP NEW SYSTEM PASSWORD
PASS# # # #(Flashing digit will change)
Do not scroll value for first digit above 9
Press to move to set each digit SET-UP NEW SYSTEM PASSWORD
PASS
# # # #
(Flashing digit will change)
Press / to set the values of remaining digits for system password
SET-UP NEW SYSTEM PASSWORD
PASS
# # # #
(Flashing digit will change)
Press to accept system password and advance to next screen to set calibration password
0 0 0 0 SET-UP NEW CALIBRATION PASSWORD
PASS# # # #(All digits flashing)
Press / to set first digit SET-UP NEW CALIBRATION PASSWORD
PASS# # # # (Flashing digit will change)
Press to move to set each digit SET-UP NEW CALIBRATION PASSWORD
PASS# # # # (Flashing digit will change)
Press / to set the values of remaining digits for calibration password
SET-UP NEW CALIBRATION PASSWORD
PASS# # # # (Flashing digit will change)
Press to return user to main setup menu for PASS submenu
PASS (Flashing)
SELSCrn
Press to scroll to next menu
III-��
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
DATEDATE (flashing)
SELSCrn
Press to set date
US SET USA OR EUROPEAN US(Flashing)
Press / to toggle between US and EUrO
SET USA OR EUROPEAN US or EUrO(Flashing)
Press to select and advance to next screen to set date
0�/0�/�000 ENTER DATE MM/DD/YYYY (US) ENTER DATEDD/MM/YYYY (EUrO)
# # . # # (Month. Day) US / (Day.Month) EUrO
2 0 # # (Year) (All digits flashing)
Press / to set first digit ENTER DATE MM/DD/YYYY (US) ENTER DATEDD/MM/YYYY (EUrO)
# # . # # (Month. Day) US / (Day.Month) EUrO
2 0 # # (Year) (Flashing digit will change)
Press to move to set each digit ENTER DATE MM/DD/YYYY (US) ENTER DATEDD/MM/YYYY (EUrO)
# # . # # (Month. Day) US / (Day.Month) EUrO
2 0 # # (Year) (Flashing digit will change)
Press / to set values of remaining digits. Set values of last two digits in the year 2 0 # #
ENTER DATE MM/DD/YYYY (US) ENTER DATEDD/MM/YYYY (EUrO)
# # . # # (Month. Day) US / (Day.Month) EUrO
2 0 # # (Year)(Flashing digit will change)
Press to return user to main setup menu for DATE submenu
DATE (Flashing)
SEL SCrn
Press to scroll to next menu
III-��
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
TIMETIME (flashing)
SELSCrn
Press to set time
00:0� ENTER TIME HOUR MINUTE # # : # # (Hour: Minute)(All digits flashing)
Press / to set first digit for hour ENTER TIME HOUR MINUTE # # : # # (Hour: Minute)(Flashing digit will change)
Set in �4 hour time format
Press to move to set each digit ENTER TIME HOUR MINUTE # # : # # (Hour: Minute)(Flashing digit will change)
Press / to set values of remaining digits for hour and minutes
ENTER TIME HOUR MINUTE # # : # # (Hour: Minute)(Flashing digit will change)
Press to return user to main setup menu for TIME submenu
TIME (Flashing)
SELSCrn
Press to scroll to next menu
III-�3
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
LOGLOG (flashing)
SELSCrn
Press to set log interval
0 0 : 0 0 SET LOG TIME IN HOUR MINUTES
# #: ## LOg(All digits flashing)
Press / to set first digit SET LOG TIME IN HOUR MINUTES
# # : # # LOg(Flashing digit will change)
To disable the log enter 0000 for log interval The minimum log interval is 5 minutesThe maximum log interval is 99 hours and 59 minutes
Press to move to set each digit SET LOG TIME IN HOUR MINUTES
# # : # # LOg(Flashing digit will change)
Press / to set values
of remaining digits for log interval
SET LOG TIME IN HOUR MINUTES
# # : # # LOg(Flashing digit will change)
Press to return user to main setup menu for LOG submenu
LOG(Flashing)
SELSCrn
Press to scroll to next menu
III-�4
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
HOLDHOLD (flashing)
SELSCrn
Press to set hold time
0 0 : 3 0 ENTER HOLD TIME HOUR MINUTE
# # : # #HOLd(All digits flashing)
Press / to set first digit ENTER HOLD TIME HOUR MINUTE
# # : # # HOLd(Flashing digit will change)
Press to move to set each digit ENTER HOLD TIME HOUR MINUTE
# # : # # HOLd(Flashing digit will change)
Press / to set values of remaining digits for hold time
ENTER HOLD TIME HOUR MINUTE
# # : # # HOLd(Flashing digit will change)
Press to return user to main setup menu for HOLD submenu
HOLD (Flashing)
SELSCrn
Press to scroll to next menu
III-�5
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Warning: The .Reset .command .will .erase .all .user .settings, .logs, .and .calibration .data! .The .Instrument .needs .to .be .set-up .and .calibrated .before .it .can .resume .operation . . .s .
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
RSETRSET (flashing)
SELSCrn
Press to reset
PRESS TEST VIEW ENTERTO RESET
rSEt?
To Reset:
Press , then
Press , then
Press
When reset is complete system returns to measure mode
User will need to re-enter the setup menu to continue to program setup parameters-
Press to return to the setup menu.
This command resets all previously set parameters to factory default values. Use this command only to set the device to original factory setup values.
PRESS TEST VIEW ENTERTO RESET
rSEt?
To Abort Reset:
Press to return user to main setup menu for RSET submenu
PRESS TEST VIEW ENTERTO RESET
rSEt?
RSET (flashing)
SELSCrn
Press to scroll to next menu
Warning: Resetting .the .computer .will .lose .all .stored .information .including .relay, .logs .and .calibration .settings . . s
III-�6
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
DISPDISP (flashing)
SELSCrn
Press to control the automatic lighting options for backlit display
AUTO BACKLIGHT LItEAUTO
Press / to scroll through AUtO, On, or OFF settings
BACKLIGHT LItEAUtO, On or OFF(flashing)
Press to return user to main setup menu for DISP submenu
DISP (Flashing)
SELSCrn
Press to scroll to next menu
III-�7
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
MEASMEAS (flashing)
SELSCrn
Press to set significant digits
3 SET NUMBERS OF SIGNIFICANT DIGITS
SIg(2, 3, or 4)
Press / to scroll through 2, 3, and 4 significant digit options
SET NUMBERS OF SIGNIFICANT DIGITS
SIg(2, 3, or 4)
Press to select and advance to next screen to enable second LCD line (mV) value on measurement screen
OFF SHOW mV ON LINE 2 OFF (Flashing)
Press / to toggle between OFF and On options
SHOW mV ON LINE 2 On or OFF (Flashing)
Press to select and return to next screen to set concentration units for sodium (ISE) measurements
AUTO SELECT ISE UNIT UnItAUtO(Flashing)
Press / to scroll through AUtO, PPb, and PP (ppm)* options
SELECT ISE UNIT UnItAUtO, PPb or PP (ppm)*(Flashing)
*PP indicates ppm
Press to return user to main setup menu for MEAS submenu
MEAS(Flashing)
SELSCrn
Press to scroll to next menu
III-�8
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
TEMPTEMP (flashing)
SELSCrn
Press to set
temperature adjustment
0 . 0 C ADJUST TEMPERATURE AdJ# .# C(Flashing)
Press / to set first digit to adjust temperature
ADJUST TEMPERATURE AdJ# .# C(Flashing digit will change)
The maximum temperature adjustment is ± �˚C
Press to move to set next digit
Press / to set next digit
ADJUST TEMPERATURE AdJ# .# C(Flashing digit will change)
Press to set a positive or negative temperature adjustment
ADJUST TEMPERATURE AdJ- # .# C
Press to return user to main setup menu for TEMP submenu
TEMP(Flashing)
SELSCrn
Press to scroll to next menu
III-�9
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
ALARMALRM (flashing)
SELSCrn
Press to set up to 3 alarms: high, low, and an error signaling contact (rLY1, rLY2, & rLY3)
OFF SELECT ALARM 1 HIGH LOW OR OFF
rLY1OFF(Flashing)
Press / to select HI, LO or OFF for rLY1
SELECT ALARM 1 HIGH LOW OR OFF
rLY1HI, LO or OFF(Flashing)
Press to set rLY1 alarm value If OFF option is selected the system will proceed to rLY�
�00 ppb To set HI or LO values for rLY1:
Press to move decimal and ppm/ppb
Press / to set digit values
Press to move to set next digit
ENTER VALUE rLY1### [concentration]
Press to accept value and advance to next screen for rLY2
OFF SELECT ALARM 2 HIGH LOW OR OFF
rLY2OFF(Flashing)
Press / to set HI, LO or OFF for rLY2 SELECT ALARM 2 HIGH LOW OR OFF
rLY2HI, LO or OFF(Flashing)
Press to set rLY2 alarm value If OFF option is selected the system will proceed to rLY3
� ppm To set HI or LO values for rLY2:
Press to move decimal and ppm/ppb
Press / to set digit values
Press to move to set next digit
ENTER VALUE rLY2
### [concentration]
Press to accept value and advance to next screen for rLY3
III-�0
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
ALARM (cont’d)
OFF SELECT CALIBRATION HOLD ERROR OR OFF
rLY3OFF(Flashing)
Press / to set CAL, HOLD, Err or
OFF for rLY3
SELECT CALIBRATION, HOLD, ERROR OR OFF
rLY3CAL, HOLD, Err or OFF(Flashing)
Press to return user to main setup menu for ALRM submenu
ALRM(Flashing)
SELSCrn
Press to scroll to next menu
III-��
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
mAMPmAMP (flashing)
SELSCrn
Press to set analog output parameters
4-�0 Select 0-20 or 4-20 4-20 (Flashing)
Press / to toggle between 4-20 or 0-20 output values
Select 0-20 or 4-20 4-20 or 0-20(Flashing)
Press to set analog output and advance to next screen for sensing output (SOUt) for sodium
LOG Press / to toggle
between LOg and LIn output
SELECT LOG OR LINEAR OUTPUT
SOUtLOg or LIn(Flashing)
The default scale values the same for both LOg and LIn
Press to set SOUt low (LO) limit
� .0 ppb mA S ENSOR OUTPUT LOW VALUE
LO#.# ppb(Flashing)
Press to move decimal and ppm / ppb
Press / to set digit values
mA SENSOR OUTPUT LOW VALUE
LO# # #. #(Flashing digit will change)
Press to move to set each digit
Press / to set values of remaining digits for SOUt low value
mA SENSOR OUTPUT LOW VALUE
LO # # #. #(Flashing digit will change)
Press to accept value and advance to next screen to set SOUt high (HI) limit
�00 ppb mA SENSOR OUTPUT HIGH VALUE
HI#.# ppb(Flashing)
Press to move decimal and ppm / ppb
Press / to set digit values
mA SENSOR OUTPUT HIGH VALUE
HI# # #. #(Flashing digit will change)
Press to move to set each digit
Press / to set values of remaining digits for SOUt high value
mA SENSOR OUTPUT HIGH VALUE
HI###. # (Flashing digits will change)
III-��
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
mAMP (cont’d)
Press to accept value and advance to next screen for temperature output (tOUt)
SET TEMP OUTPUT VALUE tOUt (Flashing)
Press to set tOUt low (LO) limit
5 .0 C mA TEMP OUTPUT LOW VALUE
LO#.# C(Flashing)
To set tOUt low (LO) limit value:
Press / to set digit values
Press arrow to move to set next digit
Press to set a positive or negative temperature value
mA TEMP OUTPUT LOW VALUE
LO# #.# C(Flashing digit will change)
Press to accept values and advance to next screen to set tOUt high value
45 C mA TEMP OUTPUT HIGH VALUE
HI# #.# C(Flashing)
To set tOUt high (HI) limit value:
Press / to set digit values
Press arrow to move to set next digit
mA TEMP OUTPUT HIGH VALUE
HI# ##. # C(Flashing digit will change)
Press to return user to main setup menu for mAMP submenu
mAMP(Flashing)
SELSCrn
Press to scroll to next menu
III-�3
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
TEST TEST (flashing)
SELSCrn
Press to test mAmp outputs (4-20), relays (rLY), and the air pump (AIr)
DIAGNOSTICS appears above SETUP in mode window
To Select: tOUt
4-�0 Press / to scroll through 4-20, rLY, and AIr setting. Select 4-20
SELECT mA OR RELAY tESt4-20(Flashing)
Press to test 4-20 outputs
SELECT TEMP OR SENSOR OUTPUT tOUt or SOUt(Flashing)
Press / to toggle between SOUt and tOUt. Select tOUt
SELECT TEMP OR SENSOR OUTPUT tOUt(Flashing)
Press for tOUt; low value (LO)
Actual LO Value Displayed
mA TEMP OUTPUT LOW VALUE LO#. # C
4 .0 mA are sourced at output terminal ��
Press for tOUt; high value (HI)
Actual HI Value Displayed
mA TEMP OUTPUT HIGH VALUE HI#. # C
�0 mA are sourced at output terminal ��
Press to return user to main setup menu for TEST submenu
TEST(Flashing)
SELSCrn
Press to test mAmp outputs (4-20), relays (rLY), and the air pump (AIr)
III-�4
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
TEST (cont’d)
To Select: SOUt
4-�0 Press / to select 4-20 SELECT mA OR RELAY tESt4-20(Flashing)
Press to test 4-20 outputs
SELECT TEMP OR SENSOR OUTPUT tOUt or SOUt(Flashing)
Press / to toggle between SOUt and tOUt. Select SOUt
SELECT TEMP OR SENSOR OUTPUT SOUt(Flashing)
Press for SOUt; low value (LO)
Actual LO Value Displayed
mA SENSOR OUTPUT LOW VALUE LO#. # C
4 .0 mA are sourced at output terminal 9
Press for SOUt; high value (HI)
Actual HI Value Displayed
mA SENSOR OUTPUT HIGH VALUE HI#. # C
�0 .0 mA are sourced at output terminal 9
Press to return user to main setup menu for TEST submenu
TEST(Flashing)
SELSCrn
Press to test mAmp outputs (4-20), relays (rLY), and the air pump (AIr)
To Select: rLY
RLY Press / to select rLY SELECT mA OR RELAY tEStrLY(Flashing)
Press to test rLY outputs
III-�5
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
TEST (cont’d)
Actual rLY � Status:
Press / to toggle rLY1 setting On or Off
USE ARROWS TO TOGGLE RELAY 1 rLY 1On or OFF
Relay contact according to screen indication
Press to accept rLY1 and move to rLY2 status
Actual rLY � Status:
Press / to toggle rLY2 setting On or OFF
USE ARROWS TO TOGGLE RELAY 2 rLY 2On or OFF
Relay contact according to screen indication
Press to accept rLY2 and move to rLY3 status
Actual rLY 3 Status:
Press / to toggle rLY3 setting On or OFF
USE ARROWS TO TOGGLE RELAY 3 rLY 3On or Off
Relay contact according to screen indication
Press to accept rLY3 and advance to air pump setting (AIr)
Actual AIr Status
Press / to toggle air pump On or OFF
USE ARROWS TO TOGGLE AIR PUMP STATE
AIrOn or OFF
Relay contact according to screen indication
Press to return user to main setup menu for TEST submenu
TEST(Flashing)
SELSCrn
Pres to test mAMP outputs (4-20), relays (rLY), and the air pump (AIr)
To select AIr:
Press / to scroll through 4-20, rLY, and AIr setting. Select AIr
SELECT mA OR RELAY tEStAIr(Flashing)
III-�6
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
TEST (cont’d)
Press to test air pump
Actual AIr Status
USE ARROWS TO TOGGLE AIR PUMP STATE
AIrActual Status
Press / to toggle AIr setting On or OFF
USE ARROWS TO TOGGLE AIR PUMP STATE
AIrOn or OFF
Relay contact according to screen indicator
Press to return user to main setup menu for TEST submenu
TEST(Flashing)
SELSCrn
Press to scroll to next menu
III-�7
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
CALCAL(flashing)
SELSCrn
Press to set calibration frequency in hours (h)
7�0 ENTER CALIBRATION FREQUENCY IN HOURS
CAL###(Flashing)
To disable alarm set value to 000
Press / to set calibration frequency value
ENTER CALIBRATION FREQUENCY IN HOURS
CAL# # #(Flashing digit will change)
Press to move to set each digit
Press / to set values of remaining digits for calibration frequency value
ENTER CALIBRATION FREQUENCY IN HOURS
CAL# # # (Flashing digit will change)
Press to return user to main setup menu for CAL submenu
CAL(Flashing)
SELSCrn
Press to scroll to next menu
III-�8
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
DKADKA(flashing)
SELSCrn
Press to program user customized Double Known Calibration values
�90 .0 mL ENTER FLOW CELL VOLUME IN mL FCEL# # #. #(Flashing)
This section may not appear, depending on the system software revision . If this section does appear, leave the flow cell volume set at �90 .0 mL
Press / to set flow cell volume (mL)
ENTER FLOW CELL VOLUME IN mL FCEL# # #. #(Flashing digit will change)
Press to move to set each digit
Press / to set values of remaining digits for flow cell volume (mL)
ENTER FLOW CELL VOLUME IN mL FCEL# # #. # (Flashing digit will change)
Press to accept values and advance to next screen to set concentration for custom standards
�9 .� ppm SET STD 1 CONCENTRATION CONC# #. #(Flashing)
To set value for Standard 1 concentration (ppm or ppb):
Press to set decimal and ppm / ppb
Press / to set value of digits
Press to move to set each digit
SET STD 1 CONCENTRATION CONC# # . #(Flashing digit will change)
Press to accept values and advance to next screen to set volume for Standard 1
0 .�0 mL SET STD 1 VOLUME IN mL Std 1#. # #(Flashing)
To set value for Standard 1 volume (mL):
Press / to set value of digits
Press to move to set each digit
SET STD 1 VOLUME IN mL Std 1# #. #(Flashing digit will change)
III-�9
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action Marquee Scrolling
Main Displaytop linebottom line
Notes
DKA (cont’d)
Press to accept value and advance to next screen to set concentration for Standard 2
�9� .0 ppm SET STD 2 CONCENTRATION CONC# # # (Flashing)
To set value for Standard 2 concentration (ppm or ppb):
Press to move decimal and ppm/ppb Press / to set values of digits
Press to move to set each digit
SET STD 2 CONCENTRATION CONC# #. #(Flashing digit will change)
Press to accept values and advance to next screen to set volume for Standard 2
0 .�0 mL SET STD 2 VOLUME IN mL Std 2#. # #(Flashing)
To set value for Standard 2 volume (mL):
Press / to set values of digits
Press to move to set each digit
SET STD 2 VOLUME IN mL Std 2# #. #(Flashing digit will change)
Press to return user to main setup menu for DKA submenu
DKA(Flashing)
SELSCrn
Press to scroll to next menu
III-30
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
The .following .steps .should .be .taken .if .a .loss .of .sample .flow .is .expected .for .more .than .one .day . . .These .procedures .will .prevent .possible .build-up .of .caustic .reagent .vapors .in .the .monitor . .
Warning: Be .sure .to .read .and .observe .the .following .requirements! . s
Danger: Turn .off .air .pump .prior .to .removing .reagent .bottle . . .The .air .pump .will .spatter .caustic .reagent .outward .as .bottle .is .removed . . .Wear .rubber .gloves .and .safety .goggles .to .avoid .possible .injury .from .reagent .residues .in .the .system . . s
To .access .air .pump, .press . . . .Press . ./ . .to .scroll .though .parameters .until .TEST .appears .in .marquis .window .
Press . .to .access .test .menu . . .Press . ./ . .to .scroll .through .submenu .until .AIr .appears .in .main .display .
Press . . . .Press . ./ . .to .toggle .between .On/OFF .setting .for .air .pump .power .
Note: Changing .air .pump .On .or .OFF .will .be .indicated .by .a .click .sound . .
Press . .to .accept .the .setting .and .returns .operator .to .test .sub .menu .
Press . .to .return .to .measure .mode .
1 . . Shut .off .sample .flow .prior .to .the .analyzer .inlet . . .Refer .to .Chapter IV, .Flow Regulation .
2 . . Disconnect .pPower . . .Refer .to .Chapter II, Wiring the Instrument .
3 . . Drain .flow .cell .
4 . . Remove .reagent .bottle .and .store .in .a .well-ventilated .area .such .as .a .laboratory .fume .hood .
5 . . Carefully .pull .sodium .and .reference .electrodes .out .of .the .top .of .the .flow .cell .and .let .them .hang .by .their .connectors . .
6 . . Locate .protective .end .cap .from .reference .electrode .kit .and .place .on .base .of .reference .electrode . . .This .will .prevent .the .reference .electrode .from .drying .out . .
Shutdown & Start-Up Procedure
Shutdown
III-3�
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Start-up 1 . . Replace .the .diffusion .tubing .if .sample .flow .has .been .off .for .more .than .a .few .days . . .Tubing .becomes .brittle .with .long-term .exposure .to .reagent . .If .age .of .reagent .is .not .known, .replace .it .and .note .in .maintenance .records . . .Re-install .bottle .on .monitor . .
2 . . Etch .sodium .electrode . . .Rinse .with .de-ionized .water, .and .carefully . .re-install .on .monitor . . .Refer .to .Chapter II, Conditioning & Installation of New Sodium Electrode . . . .
3 . . Restore .sample .flow .to .analyzer . . .If .necessary, .adjust .pressure .and .flow .rate .through .monitor .to .normal .range .of .40 .mL/min .
4 . . Remove .protective .end .cap .from .reference .electrode . . .Reinstall .electrode .on .monitor . . .Be .careful .not .to .disconnect .reservoir .tubing .from .sidearm . .
5 . . Allow .monitor .to .stabilize .for .approximately .1 .hour, .and .recalibrate . .Refer .to .Chapter IV, Calibration .
IV-�Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Chapter IV Calibration
Valve Reference Guide:
Figure IV-1 .represents .the .valve .positioning .for .the .Thermo .Scientific .Orion .2111LL .sodium .monitor .during .measurement .and .calibration .modes . . .This .quick .reference .tool .indicates .the .proper .position .of .the .4 .main .valves .used .when .starting .calibrating .the .instrument . .
• . Calibration .shut .off .valve
• . Left .air .valve
• . Right .air .valve
• . Diverter .valve
Each .2111LL .is .shipped .with .a .Valve .Reference .Guide .as .a .self-adhesive .tool .to .mount .either .to .the .back .PVC .panel .or .on .the .inside .of .the .2111LL .enclosure .(Cat . .No . .2111EN) . . .
Using the Valve Reference Guide:
Measure Mode: Turn valves to black arrow position.
Calibrate Mode: Turn valves to red arrow position.
Flow Regulation: Turn valves indicated in blue to either Open or Close position for sample flow adjustment.
Positioning of Valves for Calibration
Figure IV-�
IV-�
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
The .sample .reservoir .in .the .flow .cell .as .shown .in .Figure IV-2, .has .two .sample .volumes: .a .normal-operation .volume .of .20 .mL, .and .a .highly .accurate .and .individualy .calibrated .flowcell .volume .of .190 .mL . . .The .lower .volume .results .in .fast .system .response .while .on-line, .and .the .higher .volume .ensures .accuracy .in .calibration .
When .the .Orion .2111LL .is .first .commissioned, .it .is .advisable .to .flush .out .fluidics .system .overnight .prior .to .initial .calibration .and .use . . .The .electronics .need .not .be .turned .on .at .this .time . . .Refer .to .Chapter IV, Rinsing the Flow Cell .
• . For .accurate .volume .displacement, .install .ATC, .sensing, .and .reference .electrodes .in .respective .flow .cell .cap .ports . . .Refer .to Chapter II, Installation of New Sodium Electrode, Reference Electrode, and .ATC Probe .
• . Close .calibration .port
• . When .rinsing .before .a .low-level .standard .calibration, .be .sure .to .rinse .off .inside .of .reservoir .lid .with .de-ionized .water .to .prevent .build .up .of .sodium .contamination .a .long .the .lid .
• . To .ensure .the .instrument .achieves .stability .and .minimizes .interference .during .calibration, .the .operator .should .minimize .activity .on .system .during .calibration .
Measure Level
Flowcell Block
Calibration Level
Calibration Port
Sense
REF
TEMP
Flow Cell Operation
Figure IV-�
Recommendation
IV-3
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Warning: Be .sure .to .follow .the .air .regulation .procedure .for .rinsing .the .flow .cell . . .Refer .to .Chapter IV, Air Regulation . . .Failure .to .initiate .proper .air .settings .will .result .in .flow .cell .overfill . . .Sample .will .leak .out .the .top .of .flow .cell .cap . . .s
1 . . Open .calibration .shut .off .valve, .see .detail . A . .
. • . Open .position .shows .valve .step .in .vertical .position .facing .up .
2 . . Open .the .sample .inlet .valve, .see .detail . B . .
. • . Positioning .valve .with .pointed .end .facing .up .for .open .position .
3 . . Open .the .bypass .valve .by .turning .counterclockwise, .see .detail . C .
. • . Check .for .appropriate .sample .flow . .
. • . Adjust .pressure .regulator .if .required . . .See .step .4 .
4 . . Adjust .the .knob .on .the .regulator, . .see .detail . E , .so .that .ball .of .flow .meter .reads .40 .mL/min, .see .detail . D . . .
. • . Pulling .the .knob .will .unlock .the .setting, .allowing .adjustment . . .
. • . . Turning .the .knob .clockwise .will .increase .flow .
. • . Pushing .the .know .back .in .will .lock .into .position . .
5 . . Turn .arrow .on .diverter .valve .to .Cal .position, .with .the .arrow .pointing .up . . .Refer .to .Chapter IV, Position of Valves for Calibration . .
. • . Flow .cell .will .begin .to .fill .
6 . . Continue .to .flush .flow .cell .up .to .60 .minutes . . .Sample .will .drain .from .siphon .tube .into .siphon .drain .to .waste .
7 . . . Proceed .with .DKA .Calibration . . .Refer .to .Chapter IV, Before Performing DKA Calibration .
Rinsing the Flow Cell
50
40
30
20
10
A
D
E
C
B
Figure IV-3
IV-4
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
The .2111LL .utilizes .pressure, .gravity .and .fluid .dynamics .as .the .principal .force .of .operation .for .the .fluidic .system .
• . Air .pump, .see .detail . A .
• . Left .air .valve, .see .detail . B
• . Right .air .valve, .see .detail . C
• . Airflow .regulation .is .critical .to .proper .siphoning, .mixing .and .operation .of .the .flow .cell .during .measurement .and .calibration .modes .
• . Successful .air .regulation .requires .proper .air .pump .operation .
1 . . Right .air .valve .is .in .Mix position, .arrow .pointing .down . . .Refer .to .Figure IV-4, . C
2 . . Left .air .valve .is .in .Siphon .position, .arrow .pointing .down . . .Refer .to .Figure IV-4, . B .
. • . .System .now .maintains .a .consistent .pressure .in .flow .cell .to .initiate .siphon .during .flushing .sequence . .
3 . . At .this .point .refer .to .Chapter IV, Rinsing the Flow Cell .for .proper .fluid .flow .to .rinse .flow .cell .
Air Regulation
50
40
30
20
10
Cal
Rinse Position
SiphonMeasure
Cal
MixMeasure
A
CB
Major Components of Air Regulation System
Recommendation
Rinsing the Flow Cell
Figure IV-4
IV-5
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Maximum .system .accuracy .is .ensured .through .a .fast, .easy, .and .accurate .calibration .performed .in .the .expected .sample .range . . .This .calibration .procedure .uses .equipment .supplied .in .calibration .kit .(Cat . .No . .181148) . .
• . Perform .monthly .maintenance .procedure .including .conditioning .of .electrode . .
• . Wait .at .least .one .hour .after .changing .reagent .or .etching .electrode . .Calibration .must .be .initiated .within .2-3 .hours .of .electrode .etching .to .ensure .faster .response . .
• . Check .that .present .sample .concentration .is .equal .or .less .than .standard .1 .(10 .ppb); .if .not, .calibration .can .still .be .performed .with .a .loss .in .accuracy .as .long .as .background .concentration .is .not .significantly .larger . . .Accuracy .can .be .attained .if .alternate .standards .are .employed .but .standard .1 .should .not .be .less .than .10 .ppb .
• . Proper .pipette .technique .is .crucial .to .calibration . . .If .operator .is .not .familiar .with .pipette .equipment .or .technique, .refer .to .Appendix A-11, Pipette Operation .
• . Do .not .touch .pipette .tip .with .bare .fingers . . .Risk .of .sodium .contamination .is .possible . . .Use .powder .free .gloves .to .prevent .contamination . .
• . When .using .pipette .to .inject .standards .during .calibration, .use .fresh .pipette .tip .for .each .standard .concentration .and .each .injection .
• . With .a .fresh .tip .installed, .rinse .the .tip .with .the .same .concentration .as .the .intended .injection . .
• . Fill .the .tip .and .dispense .the .tip .with .the .standard .1-3 .times . . .Be sure to discard the standard in the drain, between the rinses.
• . Wipe .any .excess .liquid .from .the .outside .of .the .tip .with .out .touching .the .opening .
• . Insert .pipette .tip .into .calibration .port .on .flow .cell .cap . . .
• . Be .sure .to .hold .pipette .completely .vertical .while .dispensing . . .Do .not .inject .standard .on .side .of .flow .cell, .as .a .film .may .be .retained .on .the .inside .of .the .wall .resulting .in .errors .in .precision .of .calibration . . .Refer .to .Figure IV-5 .
Before Performing DKA Calibration
Figure IV-5 Pipette Dispensing
Recommendations
IV-6
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
• . To .begin .DKA, .system .starts .from .measurement .mode . . .The .air .regulation .must .be .in .the .following .positions .to .perform .Step .1 .
. • . Left .air .valve .is .in .Siphon position, .arrow .pointing .down . . • . Right .air .valve .is .in .Mix .position, .arrow .pointing .down . . • . Calibration .port .is .closed .for .proper .flow .cell .pressure .
Note: You .may .also .refer .to .Chapter IV, Valve Reference Guide .for .quick .reference .of .position .
Performing DKA Calibration
Stage User Action Marquee Main Displaytop linebottom line
Valve position / Notes
� Actual temperature reading Actual readings in concentration and mV
�* Press to start calibration.* If calibration password is
activated - user must scroll in digits 0-9 using or to set digits and to move selected digit. Press to accept
password.
ENTER PASSWORD PASS0000
CAL appears as mode
3 DKA or OFFL (Flashing)
CALtYPE
4 Press / to select DKA. DKA(Flashing)
CAL tYPE
5 Press .
6 Follow sensor etching procedure Chapter II, Conditioning & Installation of New Sodium Electrode.
ETCH SENSOR PRESS ENTER
EtCH SEnS HOLD CAL appears as mode
7 Press .
8** Check that the cal port on the flow cell cap is closedTurn Diverter Valve to cal positionArrow pointing upWait 60 minutes** Some system software
revisions will display the combined steps shown in stage 8 . For all other systems refer to stages 8a, 8b, and 8c .
CHECK CAL PORT CLOSEDTURN DIVERTER VALVE SO ARROW POINTS UPWAIT 60 MINUTES
Actual readings in ppb & mV flashing
Minutes will be counted down on Marquee. Marquee will display PRESS ENTER after 60 minutes.
IV-7
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Stage User Action Marquee Main Displaytop linebottom line
Valve position / Notes
8a Turn Diverter Valve to cal positionArrow pointing up
TURN DIVERTER VALVE SO ARROW POINTS UPPRESS ENTER
FILL FCEL
8b Press .
8c Wait 60 minutes WAIT 60 MINUTES Actual readings in ppb & mV flashing
Minutes will be counted down on Marquee. Marquee will display PRESS ENTER after 60 minutes.
9 Press .
�0 Turn Left Air Valve to calibrate position. Arrow pointing upTurn Right Air Valve to calibrate position. Arrow pointing up
TURN LEFT AND RIGHT VALVES UPPRESS ENTER
Alr
�� Press .
�� When liquid level in the flow cell has risen above the siphon tube and begins to drop, turn the Calibration Shut Off Valve (above flow meter) to calibrate position.
WHEN LIQUID LEVEL BEGINS TO DROP TURN SHUTOFF VALVE RIGHT PRESS ENTER
FLO OFF
�3 Press .
�4 Turn Left Air Valve to siphon position. Arrow pointing downFluid passing through siphon to drain will be observed.
TURN LEFT VALVE DOWN PRESS ENTER
SIPH
�5 Press .
�6 When siphoning stops:Turn Left Air Valve to calibrate position. Arrow pointing up Open cal port on flow cell cap. Turn Right Air Valve to mix position. Arrow pointing down
WHEN SIPHONING STOPS TURN LEFT VALVE UP
OPEN CAL PORTTURN RIGHT VALVE DOWN PRESS ENTER
Alr
IV-8
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Stage User Action Marquee Main Displaytop linebottom line
Valve position / Notes
�7 Press .
�8 Wait 7 minutes STARTING CALIBRATION MEASURING BACKGROUNDWAIT 7 MINUTES
Actual readings in ppb & mV flashing
System circulates sample. Minutes will be counted down on Marquee.
�9 Wait 0 - 5 minutes AWAITING DATA STABILITY Actual readings in ppb & mV flashing
System continuously evaluates data by applying stability criteria until a “stable/pass” condition is met. If system stabilizes follow stage �0.If system stability is not reached a timeout occurs. Error E041 is issued. Jump to step Error E04�.
�0 Add Std 1 through cal port ADD STD1 PRESS ENTER
Add Std1
�� Press .
�� Wait 5 minutes MEASURING STD1CIRCULATINGWAIT 5 MINUTES
Actual readings in ppb & mV flashing
System circulates sample. Minutes will be counted down on Marquee.
�3 Wait 0 - 5 minutes AWAITING DATA STABILITY Actual readings in ppb & mV flashing
System continuously evaluates data by applying stability criteria until a “stable/pass” condition is met. If system stabilizes follow stage �4.If system stability is not reached a timeout occurs. Error E041 is issued. Jump to step Error E04�.
�4 Add Std 2 through cal port ADD STD2 PRESS ENTER
Add Std2
�5 Press .
�6 Wait 4 minutes MEASURING STD2CIRCULATINGWAIT 4 MINUTES
Actual readings in ppb & mV flashing
System circulates sample. Minutes will be counted down on Marquee.
IV-9
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Stage User Action Marquee Main Displaytop linebottom line
Valve position
�7 Wait 0 - 5 minutes AWAITING DATA STABILITY
Actual readings in ppb & mV flashing
System continuously evaluates data by applying stability criteria until a “stable/pass” condition is met. If system stabilizes follow stage �8.If system stability is not reached a timeout occurs. Error E041 is issued. Jump to step Error E04�.
�8 No action necessary System calculates new Eo and slope and checks for errors. If no errors are found follow stage �9.If there is a slope/E0 or other cal data related error, jump to appropriate error code.
�9 Press to accept the new slope and Eo valuesorPress to abort calibration. The values will not be stored. The last saved values for calibration prior to the aborted values will be used in measurement mode.
NEW CALIBRATION SLOPE AND E0
PRESS ENTER TO ACCEPT OR PRESS EXIT TO ABORT WITHOUT SAVING.
Calculated Slope and E0 System presents the new slope and E0 values.If is pressed, continue to follow stage 30.If is pressed, refer to step Exit � or Exit 4, depending on the system software revision.
Some system software revisions will not display the following sections that are highlighted in gray
NEW CALIBRATION SLOPE AND E0 SAVED. PRESS ENTER
CAL StOr Calibration values for Slope and Eo are stored.Successful calibration completed.
Press
IV-�0
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Stage User Action Marquee Main Displaytop linebottom line
Valve position / Notes
30*** Turn Left Air Valve to siphon position. Arrow pointing down.Turn Right Air Valve to mix position.Arrow pointing down.For correct position in measurement mode.Turn Diverter Valve to sample position. Arrow pointing left.Turn Calibration Shut Off Valve to measure position. Arrow pointing up. Close cal port on flow cell cap.
TURN LEFT VALVE DOWNTURN DIVERTER VALVE SO ARROW POINTS LEFTTURN SHUFOFF VALVE UPCLOSE CAL PORT PRESS ENTER
FLO On
*** Some system software revisions will display the combined steps shown in Stage 30. For all other systems refer to Stages 30a, 30b, 30c, 30d, and 30e.
30a Turn Left Air Valve to siphon position. Arrow pointing downTurn Right Air Valve to mix position.Arrow pointing downFor Correct position in measurement mode
TURN LEFT AND RIGHT VALVE TO DOWNPRESS ENTER
PrEP rd1
30b Press .
30c Turn Diverter Valve to sample position. Arrow pointing left
TURN DIVERTER VALVE SO ARROW POINTS LEFTPRESS ENTER
PrEP rd2
30d Press .
30e Turn Calibration Shut Off Valve to measure position.Close cal port on flow cell cap.
TURN SHUTOFF VALVE UP CLOSE CAL PORTPRESS ENTER
FLO On
IV-��
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Stage User Action Marquee Main Displaytop linebottom line
Valve position
3� Press .
3� System will proceed to measurement mode
Actual temperature reading
Actual readings in ppb & mV
System enters measurement mode, but will remain on hold until hold timer expires. (Default 30 minutes)
Calibration Abort- Exit Steps
Exit � ARE YOU SURE? PRESS ENTER TO ABORT OR PRESS LAST SCREEN TO BACKUP.
SUrE? Abort calibration vs. Last Screen option:The system allows the user two options when is pressed at any step during the calibration. Go to Exit �
Exit � Decide whether or not you really want to exit the calibration process and lose the new slope and E0:To Abort calibration: Follow step Exit 3:User enters into the measure mode.User must refer to stage 30 for proper air and fluid flow for measurement mode.To Continue calibration: If is pressed, system will go back to last screen used in calibration sequence prior to pressing .Continue to follow calibration according to Marquee and main display directions.
Exit 3 Press .
Exit 4 CALIBRATION ABORTEDPRESS ENTER
CAL Abrt
Exit 5 Press . Refer to stage 30 for proper air and fluid flow for measurement mode. System return will return to measure mode. (Begin again with stage � to retry calibration sequence)
IV-��
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Stage User Action Marquee Main Displaytop linebottom line
Valve position / Notes
Stability Timeout
Error 4� Decide to continue calibration despite instability or abort
READINGS WERE UNSTABLE. PRESS ENTER TO CONTINUE CALIBRATION OR PRESS EXIT TO ABORT
E041 Stability was not reached before timeoutIf is pressed system will continue from step where stability timed out.If is pressed refer to Exit � or Exit 4, depending on the system software revision
Calibration Error
Error 4� Review Marquee text to understand the failure that occurred
INVALID (NEGATIVE) SLOPEPRESS ENTER TO ABORT CALIBRATION
E042 Invalid slope continue at Exit 3
Error �09 Review Marquee text to understand the failure that occurred
BAD SLOPE PRESS ENTER
E109 Bad SlopePress to continue at stage �9
Error ��5 Review Marquee text to understand the failure that occurred
BAD E0
PRESS ENTERE125 Bad E0
Press to continue at stage �9
IV-�3
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
The .Thermo .Scientific .Orion .2111LL .sodium .monitor .software .has .default .values .to .accommodate .0 .1 .mL .additions .of .Orion .standard .1 .and .standard .2 .and .flow .cell .volume .of .190 .mL . . .
The .operator .has .the .option .to .use .alternate .standards .by .changing, .in .Setup .mode .the . .flow .cell .volume, .the .standard .1 .and .standard .2 .values .and .the .volumes .to .be .used .
Let . VC . . = . .Flow .cell .volume . . Fixed: .190 .0 .mL .
. C1 . . = . .Standard .1 .concentration . . Default: .19 .1 .ppm .
. C2 . . = . .Standard .2 .concentration . . Default: .192 .ppm .
. V1 . . = . .Volume .of .standard .1 .added . . Default: .0 .1 .mL .
. V2 . . = . .Volume .of .standard .2 .added . . Default: .0 .1 .mL .
. dC1 . .= . .Concentration .change .due .to .the .first .addition .
. dC2 . .= . .Concentration .change .due .to .the .second .addition .
. . . . dC1 . .= . .C1 .x .V1 ./ .(V1 .+ .VC)
. . . . dC2 . .= . .C2 .x .V2 ./ .(V1 .+ .V2 .+ .VC)
• . dC1 .should .be .at .least .twice .the .normal .concentration .in .the .sample
• . dC2 .should .be .about .ten .times .of .dC1 . .dC1 .should .not .be .less .than . .10 .ppb . .
Calibration At Custom Concentrations Using DKA
For maximum accuracy
IV-�4
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
The .2111LL .sodium .monitor .remains .in .sample .mode .during .the .entire .procedure .
1 . . . .Electrode .must .be .etched .within .2-3 .hours .of .this .procedure .
2 . . . .Fill .flow .cell .to .calibration .level .(V0, .190 .mL) . .
3 . . . .Note .concentration .displayed .when .reading .is .reasonably .stable .(C0) . .
4 . . . .Add .standard .1, .volume .V1, .with .concentration .C1 . .
5 . . . .When .reading .is .reasonably .stable, .note .concentration .(CS) . .
6 . . . .The .noted .value .should .be .within .± .10% .of .the .calculated .value .according .to: .
. CS .= .(C0 .x .V0+ .C1 .x .V1) ./ .(V0 .+ .V1) . . • . C1 .should .be .sodium .concentration .actually .in .the .standard .bottle . . . . • . For .example, .standard .1 .of .(Cat . .No . .181140) .is .equal .to .19 .1 .ppm . .
Span Check Procedure
IV-�5
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
The .offline .calibration .feature .is .simply .a .one-point .calibration .without .slope .change . .The .term .“offline .calibration” .refers .only .to .the .fact .that .a .sample .from .2111LL .bypass .is .taken .“offline” .for .laboratory .analysis; .in .fact, .no .downtime .is .experienced .and .the .instrument .remains .online .throughout .the .procedure .
Offline Calibration Procedure
Stage User Action Marquee Main Displaytop linebottom line
Notes
� Open bypass valve to redirect flow. Take QC sample from bypass of 2111LL.
Actual temperature reading
Actual readings in concentration &mV
�* Press to start calibration* If calibration password is
activated - user must scroll in digits 0-9 using . ./ .
.to set digits and to move selected digit.
Press to accept password.
ENTER PASSWORD PASS0000
CAL appears as Mode.
3 DKA or OFFL (Flashing)
CALtYPE
4 Press . ./ . .arrows to select OFFL
OFFL(Flashing)
CAL tYPE
5 Press . HOLD CAL appears as mode.
6 Wait 0 - 5 minutes AWAITING DATA STABILITY
Actual readings in ppb & mV flashing
System continuously evaluates data by applying stability criteria until a “stable/pass” condition is met. When stability is reached the values are stored and the system automatically returns to the measurement mode.
7 Perform laboratory analysis on QC sample and return with results
System remains in measurement mode and operates as normal. The CAL and MEASURE mode icons will both be displayed.
IV-�6
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Stage User Action Marquee Main Displaytop linebottom line
Notes
8* Press . .* If calibration password is
activated - user must scroll in digits 0-9 using . ./ .
.to set digits and to move selected digit.
Press to accept password.
ENTER PASSWORD PASS 0000
CAL appears as Mode
9 DKA or OFFL (Flashing)
CALtYPE
�0 Press . ./ . to select OFFL.
OFFL(Flashing)
CAL tYPE
�� Press . HOLD CAL appears as mode
�� OFFLINE CAL CONC# #.#(Flashing)
�3 Adjust the displayed value to the QC value obtained from laboratory analysis.Use to set decimal and ppm / ppb.
Use . ./ . to set the first digit. Use to move to set each digit.
Use . ./ . to set the values of remaining digits.
OFFLINE CAL CONC# #.#(Flashing digit will change)
�4 Press . The system recalculates the E0 and returns to the measurement mode. Only the MEASURE mode icon should be displayed.
V-�Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Chapter V Instrument Maintenance
The .Thermo .Scientific .Orion .2111LL .sodium .monitor .is .designed .for .simple .maintenance . . .Follow .the .instructions .in .this .section .to .ensure .proper .operation .of .your .instrument .
To .ensure .proper .maintenance .and .good .monitor .performance, .a .service .logbook .should .be .kept .
• . Record .maintenance .date .and .type .service .work .completed . .• . Mark .date .and .fluid .levels .of .reagent .and .date .when .replaced . .• . Tag .each .electrode .cable .with .installation .date .of .electrode .
1 . . . Check .that .sample .flow .rate .is .between .35 .and .45 .mL/min . . .To .alter .flow .rate, .pull .out .locking .knob .of .pressure .regulator, .then .rotate .black .knob .to .increase .(clockwise) .or .decrease .(counterclockwise) .sample .flow . . .Push .in .to .lock .rate, .if .desired . .
2 . . . Inspect .unit .for .leakage . . .Diffusion .tubing .leakage .is .indicated .by .a .rise .in .level .of .reagent . .
3 . . . Check .that .there .are .no .error .indications .and .current .concentration .level .is .reasonable . .
4 . . . Check .that .reference .fill .solution .is .adequate . .
5 . . . If .required .wipe .instrument .with .damp .cloth .to .remove .dirt .and . .dust .particles . .
Maintenance Schedule
Weekly Maintenance
Recommendation
V-�
Instrument Maintenance
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Warning: Safety .glasses .and .gloves .must .be .worn .during .entire .conditioning .procedure . . .Read .and .follow .MSDS .precautions . . .Only .sodium .sensing .electrode .is .etched . . .Do .not .etch .reference .electrode . . .s
1 . . . Immerse .sodium .electrode .in .bottle .of .etch .solution .for .one .minute . .
2 . . . Remove .and .rinse .with .de-ionized .water . .
3 . . . Replace .electrode .into .sensing .port .in .flow .cell .cap .
Calibration .frequency .is .user .dependent .for .the .most .accurate .and .precise .measurements .at .low .levels . . .
• . Etching .of .the .sodium .electrode .is .required .for .effective .calibration . .Calibration .must .be .completed .within .2-3 .hours .of .etching .electrode .for .optimal .results .
• . Replacement .of .DIPA .reagent .requires .calibration . .Refer .to . .Chapter IV, Calibration .
• . Calibration .intervals .can .be .programmed .using .the .setup .mode . . .This .will .prompt .the .user .to .recalibrate .after .a .specific .amount .of .time .has .passed . . .Refer .to .Chapter III, Use of Setup Mode .
Refer .to .Chapter II, Installation of New Reference Electrode .
Reference .electrode .filling .solution .is .meant .to .be .a .several .month .supply . .Therefore, .if .reservoir .is .less .than .one .quarter .full, .replace .solution . . .This .should .be .checked .each .month .to .ensure .the .electrode .does .not .run .dry .
1 . . . Remove .spent .electrolyte .bottle .from .clip . . .Unscrew .cap .and .discard .bottle . .
2 . . . Remove .cap .and .seal .from .new .electrolyte .fill .solution . . .Check .that .rubber .gasket .is .properly .aligned, .then .connect .cap/tubing .assembly .to .bottle . . .The .1/8” .tubing .should .extend .into .the .bottle . .
3 . . . Invert .electrolyte .bottle .and .snap .into .clip . .
4 . . . Use .pushpin .to .puncture .three .air .vents .in .bottom .of .bottle . .
Calibration
Monthly Maintenance Conditioning
Sodium Electrode
Replacement of Reference Electrode Filling Solution
V-3
Instrument Maintenance
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
1 . . . If .sample .flow .can .be .adjusted .with .the .pressure .regulator .to . .50 .mL/minute .or .more, .skip .this .procedure . .
2 . . . Turn .off .sample .flow .by .turning .the .inlet .valve .to .the .right . .
3 . . . Remove .filter .hex .cap .using .1-inch .open-ended .wrench . .
4 . . . Separate .filter .cap .with .bypass .valve .assembly .from .filter .body, . .retaining .ring .
5 . . . Remove .filter .element .and .soft .gasket . .(if .necessary) .
6 . . . Replace .gasket .and .press .new .sintered .element .into .filter .body . .
7 . . . Replace .hex .cap .and .bypass .valve .assembly .and .reposition .bypass .drain .line .into .drain .assembly .
8 . . . Retighten .hex .cap .to .approximately .75 .pound-inch .in .torque . .
9 . . . Turn .on .sample .flow .by .1⁄4 .turn .of .inlet .valve .to .the .left . .
10 . .Check .for .leaks . .
Warning: Replacement .of .DIPA .reagent .is .dependent .on .operating .conditions .and .usage . . .Replacement .may .occur .as .long .as .60 .days .from .installation . . .Be .sure .to .monitor .the .DIPA .levels .weekly .to .assess .proper .replacement .schedule . . .DIPA .is .hazardous . . .Use .protective .glasses .and . .gloves . . .Refer .to .bottle .label .and .MSDS .for .precautions . . .Work .in .a . .well-ventilated .area . .
Danger: Turn .off .the .air .pump .before .beginning .procedure; .otherwise, .the .air .pump .will .spatter .caustic .reagent .outward .as .bottle .is .removed . . .s
Refer .to .Chapter III, Use of Setup Mode for .instructions .to .turn .air .pump .On/Off . .
1 . . . Remove .thumbnut .and .slide .reagent .bottle/adaptor .from .manifold .assembly . .
2 . . . Unscrew .bottle .cap, .and .lift .adaptor .from .bottle . .
Replacement Sample Inlet Filter
Replacement of DIPA Reagent and Diffusion
Tubing Assembly
V-4
Instrument Maintenance
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
3 . . Unscrew .the .fitting .to .disengage .the .diffusion .tubing .assembly . . .Properly .dispose .of .spent .reagent .and .tubing . . .Refer .to .Figure V-1 .
4 . . Replace .reagent .bottle .cap .adapter .O-ring . .
5 . . Follow .instructions .in .Chapter II, Installation of DIPA Reagent .and .Diffusion Tubing . .
DIPA .reagent .will .last .2 .months .if .the .temperature .is .less .than .25 .˚C . . .Higher .temperatures .will .require .bottle .change .out .sooner, .about .6 .weeks . . .We .generally .change .reagents .when .level .is .less .than .1 .inch .remaining . . .The .reagent .bottle .can .be .used .almost .to .dryness, .less .than .50 .mL .remaining .if .readings .are .not .extremely .low, .< .0 .5 .ppb . . .It .is .imperative .that .bottle .doesn’t .go .completely .dry .or .the .system .will .be .compromised .
Replacement .of .reagent .manifold .face .O-rings . . .Refer .to .Figure V-2 .
• . Remove .reagent .bottle .adapter .assembly . .• . Remove .all .three .O-rings .from .face .of .manifold . .• . Insert .new .O-rings .
Notes on DIPA
Figure V-� Reagent Manifold
O-ring Location Reagent Manifold O-rings (3)
Manifold Screw
Bottle Adapter O-ring
Diffusion Tubing Assembly
O-ring Seal Fitting
Reagent Bottle Adapter
Figure V-� Exploded View
V-5
Instrument Maintenance
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Yearly Preventive Maintenance
Electrodes
Flow Cell
1 . . Replace .Orion .ROSS .Ultra® .reference .electrode .(Cat . .No . .210058) . .Refer .to .Chapter II, Instrument Preparation .for .step-by-step .instructions .
2 . . Replace .Orion .ROSS® .sensing .electrode .(Cat . .No . .210048) . .Refer .to .Chapter II, Instrument Preparation .for .step-by-step .instructions .
3 . . Replace .calibration .and .etching .solution .(Cat . .No . .181140)
Dispose .of .all .old .solutions .according .to .MSDS .instruction . . .MSDS .can .be .downloaded .at .www .thermo .com/water .
Using .the .flow .cell .O-ring .kit .from .the .Orion .annual .maintenance .kit . .(Cat . .No . .2111MK) . . .Refer .to .Figure V-3 for .O-ring .location . .
1 . . Follow .shut .down .procedure .to .stop .sample .flow .to .monitor . . .Refer .to .Chapter III, Shutdown Procedure .
2 . . Remove .electrodes .and .ATC .from .flow .cell .cap . .
3 . . Drain .flow .cell .completely .
4 . . Remove .flow .cell .cap . . .Replace .large .O-ring .in .flow .cell .cap . .
5 . . Pull .flow .cell .chamber .straight .up .to .remove .from .flow .cell .base . . .Be .careful .not .to .twist . . .The .flow .cell .base .has .a .pin .to .hold .the .flow .cell .chamber .in .the .correct .position .
6 . . Rinse .flow .cell .chamber .with .de-ionized .water .
7 . . Remove .sensing .tube .by .pulling .vertical .from .flow .cell .base .
8 . . Replace .sensing .tube .O-ring .in .base .and .install .new .sensing .tube .to .original .location .
9 . . Replace .large .O-ring .in .flow .cell .base .
10 . .Install .flow .cell .chamber . . .Line .up .pinhole .in .chamber .to .pin .in .flow .cell .block . . . .Push .straight .down . . .Do .not .twist .
11 . .Replace .flow .cell .cap .on .flow .cell .chamber .
12 . .Re-install .electrodes .and .ATC .according .to .their .port .locations . .
13 . .Follow .start .up .procedure .to .return .sample .flow .to .instrument . .Refer .to .Chapter III, Shutdown Procedure .Figure V-3
Flow Cell Assembly
V-6
Instrument Maintenance
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Using .the .restrictor .tube .assembly .(Cat . .No . .2111RT) .from .the .Orion .annual .maintenance .kit .(Cat . .No . .2111MK) . .Refer .to .Figure V-4 .for .restrictor .tubing .connections .
1 . . Close .sample .inlet .valve .to .stop .sample .flow .through .monitor .by .turning .valve .to .the .right .
2 . . Loosen .restrictor .tube .fitting .connection .from .top .of .calibration .shut .off .valve .
3 . . .Loosen .restrictor .tube .fitting .from .reagent .manifold .
4 . . Pull .out .the .restrictor .tube .assembly .at .each .end . .
5 . . Unscrew .clip-holding .restrictor .to .panel .
6 . . Replace .with .new .restrictor .tubing .assembly . .
7 . . Be .sure .all .fittings .are .secure, .to .prevent .leaks .
8 . . Open .sample .inlet .valve .to .start .sample .flow .through .monitor .by .turning .valve .to .the .left .
Replacement of Restrictor Tubing
Figure V-4 Restrictor Tubing
50
40
30
20
10
Reagent Bottle
Air Pump
Right Air Valve
Restrictor Tube Assembly
VI-�Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Chapter VI Troubleshooting
• . Press . .to .enter .the .diagnostics .mode .
• . The . .key .allows .access .to .menus .used .for .system .diagnostics . .
• . Each .menu .is .sequential . . .Press . .to .scroll .through .the .menu .selections . . .
• . .can .be .pressed .at .any .time .to .return .to .the .measurement .mode .
Diagnostics Mode
Menu User Action Marquee Scrolling Main Displaytop linebottom line
Notes
Diagnostics
Press . to enter diagnostics mode.System will enter diagnostics mode at calibration log menu.
Press . to step through the menu items.
CALIBRATION SLOPE AND E0
ERROR LIST PRESS VIEWSELECT TIME/DATE TO VIEW MEASUREMENT LOGSELECT TIME/ DATE TO VIEW STATUS LOGSOFTWARE REV NUMBERSERIAL NUMBERMODEL NUMBERMEASURE mV AND NOISEmA OUTPUT VALUES SENSOR/TEMPKEYPAD TEST DISPLAY TEST
The display will change with each menu
DIAGNOSTICS appears in mode window Keypad test menu shown immediately after display test
VI-�Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
• . Displays .up .to .the .last .five .calibrations .using .three .information .screens:1 . . Slope .and .E0
2 . . Date .and .Time3 . . Temperature .and .Electrode .Response .Time
• . Use . .to .scroll .between .the .three .information .screens .for .one .calibration
• . In .the .Date .and .Time .screen .use . ./ . .to .view .up .to .the .last .five .calibrations
Calibration Log
Menu User Action Marquee Scrolling Main Displaytop linebottom line
Notes
Press . .
Calibration Log CALIBRATION SLOPE AND E0 ##.# (Slope) ##.# (E0)
DIAGNOSTICS appears as mode
Press . to display the calibration time and date.
LAST CALIBRATION TIME/DATE HH:MMMM/DD or DD/MM
In the calibration time and date display press
./ . to scroll through the last five calibrations
Press . .to display the calibration temperature and electrode response.
CALIBRATION TEMP + ELECTRODE RESPONSE TIME
##.# C (Temp)### (Response)
Press . .
VI-3Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
• . Displays .a .list .of .system .events, .warnings, .and .error .codes .that .cause .the .yellow .or .red .LED .to .light . .
• . The .top .line .of .the .main .display .shows .the .error .code .and .the .bottom .line .shows .the .number .of .error .codes .(1-5 .indicates .that .one .of .five .error .codes .is .being .displayed) .
• . Press . ./ . .to .scroll .through .the .error .list .
• . Press . .to .display .a .description .of .each .error .code .on .the .Marquee .
Error List
Menu User Action Marquee Scrolling Main Displaytop linebottom line
Notes
Error List ERROR LIST PRESS VIEW E#### - #
Press . to display description of error code on Marquee.
ACTUAL DESCRIPTION OF ERROR CODE E#### - #
Press . ./ . to scroll through the error list.
ACTUAL DESCRIPTION OF ERROR CODE E#### - #
Press . .
VI-4Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
• . Displays .logged .measurements .according .to .log .interval .entered .in .setup .menu .
• . Press . .to .toggle .between .the .measurement .value .and .the .data .and .time .
• . Press . ./ . .to .scroll .through .the .log .entries . .
Measurement Log
Menu User Action Marquee Scrolling Main Displaytop linebottom line
Notes
Measurement Log SELECT TIME/DATE TO VIEW MEASUREMENT LOG
HH:MMMM/DD or DD/MM
Press . ./ . to scroll through the log entries
SELECT TIME/DATE TO VIEW MEASUREMENT LOG
HH:MM MM/DD or DD/MM
Press . to toggle between the measurement log and the data and time display
Temperature for logged measurement
Concentration and mV values for logged measurement
Press . .
VI-5Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
• . Displays .a .list .of .system .events .
• . Press . ./ . .to .scroll .through .the .log .entries .
• . Press . .to .toggle .between .the .log .code .and .the .data .and .time .
• . Displays .the .software .revision .numbers .of .the .two .processors .
Status Log
Software Revision
Menu User Action Marquee Scrolling Main Displaytop linebottom line
Notes
Status Log SELECT TIME/DATE TO VIEW STATUS LOG
HH:MM MM/DD or DD/MM
Press . ./ . to scroll through the log entries
SELECT TIME/DATE TO VIEW STATUS LOG
HH:MM MM/DD or DD/MM
Press . to toggle between the log code and the data and time
PRESS UP/DOWN KEY TO VIEW STATUS LOG
Event code
Press . .
Software Revision
SOFTWARE REV NUMBER r # . # #r # . # #
Press . .
VI-6Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
• . Displays .the .serial .number .of .the .electronics .assembly .
• . Displays .the .model .number .of .the .system .
• . Displays .the .current .mV .measurement .value .and .the .noise .associated .with .the .measurement .
• . Displays .current .mA .values .for .sensor .and .temperature .outputs .
Electronics Serial Number
Model Number
mV and Noise Measurements
mA Output Values
Menu User Action Marquee Scrolling Main Displaytop linebottom line
Notes
Electronics Serial Number SERIAL NUMBER # # ## # #
Press . .
Model Number MODEL NUMBER # # # ## # # #
Press . .
mV and Noise Measurements MEASURE mV AND NOISE Current mV measurement and noise
Press . .
mA Output Values mA OUTPUT VALUES SENSOR/ TEMP
Current mA output values for sensor and temperature
Press . .
VI-7Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
• . Press . .to .start .display .test .1 . . All .the .display .segments .will .light .up .and .the .system .will .cycle .
through .the .LED .colors . .2 . . Verify .that .all .the .display .segments .are .on .and .that .the .LED .colors .
are .functional .and .press . .3 . . All .the .display .segments .will .turn .off .except .for .the .Marquee .display, .
which .should .show .PRESS ENTER TO CONTINUE .4 . . Verify .that .all .the .display .segments .are .off .and .press . .5 . . The .system .will .show .a .counting .number .pattern .on .both .lines .of .
the .main .display .6 . . Verify .that .the .display .segments .are .functional .and .press . .to .
proceed .to .the .keypad .test .
Display Test
Menu User Action Marquee Scrolling Main Displaytop linebottom line
Notes
Display Test PRESS ENTER FOR DISPLAY TEST
dISPtESt
Press . .
Verify that all the display segments are on and that the LED colors are functional.
All segments on All segments on System will cycle through the LED colors
Press . .
Verify that all the display segments are off.
PRESS ENTER TO CONTINUE
All segments off
Press . .
Verify that the display segments are functional.
PRESS ENTER TO CONTINUE
Count patternCount pattern
Press . .
VI-8Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
• . The .main .display .will .show .0 .and .the .Marquee .will .display .KEYPAD TEST PRESS ALL KEYS ONE AT A TIME .
• . Press .all .eight .keys .one .at .a .time .in .any .order .
* . . After .the .first .key .is .pressed .Error .E033 .will .be .displayed .if .a .new .key .is .not .pressed .within .10 .seconds .
Keypad Test
Menu User Action Marquee Scrolling Main Displaytop linebottom line
Notes
Keypad Test KEYPAD TEST PRESS ALL KEYS ONE AT A TIME
0
Press all keys one at a time in any order
KEYPAD TEST PRESS ALL KEYS ONE AT A TIME
Display will show numbers 1 through 8 as keys are pressed
After the first key is pressed Error E033 will be displayed if a new key is not pressed within 10 seconds
VI-9
Troubleshooting
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Less .than .45 .mV
• . Check .electrode .function, .etch .electrode, .recalibrate, .if .continues .to .read .low, .replace .electrode . .
• . Check .electronics .to .read .mV .input .correctly . . .Use .electronic .test .kit .(Cat . .No . .180029) .procedure .for .checking .electronics .function . .
• . Background .concentration .too .high .for .standards .used .
. Ion-selective .electrodes .(ISEs) .can .be .shown .as .DC-voltage .sources .with .50 .to .1000 .Mohm; .a .very .high .output .impedance . . .ISEs .respond .to .ion .concentration .changes .with .a .voltage .change .of .approximately .60 .mV/decade .for .monovalent .ions .as .sodium . . .Knowing .electrode .and .instrument .parameters .like .offset .potential, .slope, .etc ., .allows .testing .of .various .instruments .by .applying .a .known .mV-voltage .
Input Signal Expected Reading: ± (0 .5 mV + 0 .�%)
0 mV - 0.5 mV to + 0.5 mV
- 1000 mV -1001.5 mV to – 998.5 mV
+ 1000 mV + 998.5 mV to + 1001.5 mV
• . Standards .contaminated .or .incorrectly .made . . .Verify .fresh .standards . .
• . Standard .1 .and .standard .2 .interchanged .– .Check .values .on .standards . .• . Standard .1 .value: .19 .1 .ppm• . Standard .2 .value: .192 .ppm• . If .using .custom .standard, .verify .values .for .each .addition .
• . pH .adjustment .is .not .correct . . .Check .level .and .flow .for .DIPA .reagent . .Replace .DIPA .if .contamination .or .level .is .less .than .50 .mL . .Refer .to .Chapter II, Installation of Reagent & Diffusion Tubing .
• . Standard .1 .and .standard .2 .added .in .wrong .sequence . .Refer .to . .Chapter IV, Calibration .
Slope ProblemLow Slope
VI-�0
Troubleshooting
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Slope .greater .than .63 .mV
• . Improper .calibration .technique .– .Refer .to .Chapter IV, Calibration .
• . Calibration .standards .contaminated .– .Replace .with .new .standards . .
• . Electronics .failure .– .Try .to .reset .monitor . .Refer .to .Chapter VI, Resetting the Monitor .
• . Improper .pipette .technique .– .Refer .to .section: .Appendix A-11, Pipette Operation .and .Technique .
• . Background .concentration .too .high .for .standards .used .
• . Reference .and .sensing .cables .interchanged .– .verify .wiring .of .cable .connections .to .terminal .strip . .Refer .to .Chapter II, Wiring the Instrument .
• . Check .the .connection .of .the .cable .markers .for .the .corresponding .electrode . .(Cable .markers .are .the .white .tape .at .the .cap .connector .on .the .cable):• . “Ref” .for .reference .electrode . .• . “Sensing” .for .the .sodium .electrode .
• . .Sensing .electrode .coated .– .Re-etch .or .replace .sensing .electrode .
High Slope
VI-��
Troubleshooting
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Malfunction Possible Cause Remedy
Noisy Blocked reference junction
Make sure electrolyte bottle is flowing properly. Check that bottle has been vented. Refer to Chapter II
Sodium electrode failure Check electrode cable connection and location of wiring to terminal strip. Refer to Chapter II
Temperature probe Disconnect ATC. Display should read default 25 ˚C (E101). Replace if necessary
Air pump failure Monitor airflow for steady stream of bubbles. Replace if necessary
Faulty electrode cables Replace electrode cables
Calibration standard carryover
Disconnect air pump intake tube. Rinse flow cell and electrodes with de-ionized water by lifting flow cell cap. Reconnect air pump
Excessive Drift Sample concentration varying
Perform span check. If span check passes, monitor is okay. Refer to Chapter IV
Blocked reference junction
Make sure electrolyte bottle is vented. Check tubing to electrode for obstruction. Refer to Chapter II
Reference electrode Missing or not using correct filling solution (Cat. No. 150072)
Sodium electrode response
Etch sodium electrodeInstall new sodium electrode. Refer to Chapter II
Temperature probe Disconnect ATC. Display should read default 25 ˚C (E101). Replace if necessary
Burst diffusion tubing Observe liquid level in reagent bottle. If liquid level rises, replace reagent and diffusion tubing. Tubing connector may need replacement. Refer to Chapter II
Low Flow Rate Sample pressure below 8 psi
Check sample pressure. If less than 8 psi, increase. Refer to Chapter II
Pressure regulator set too low
Increase pressure by pulling on red locking ring and turning black knob clockwise. Refer to Chapter II
Inlet filter clogged Replace or clean filter
Restrictor tubing crimped or clogged
Replace restrictor tubing. Refer to Chapter II
Troubleshooting Matrix
VI-��
Troubleshooting
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Malfunction Possible Cause Remedy
No Air Bubbles Air line crimped or disconnected
Check air line. Repair as required
Pump failure Replace if necessary
Does Not Calibrate Properly
Contaminated standards, pipette or pipette tips
Use new standard solutions. Replace tips
Standards intercharged Repeat standards in correct sequence using standard 1 and then standard 2. Refer to Chapter IV
Background concentration too high for standards used.
Repeat standards in correct sequence using standard 1 and then standard 2. Refer to Chapter IV
Pipette technique Use correct pipette technique. Refer to Appendix
Reagent diluted or contaminated
Replace DIPA reagent. Refer to Chapter II
Flow cell contaminated Rinse flow cell with de-ionized water. Flush flow cell overnight with flow at calibration liquid level. Refer to Chapter IV
Temperature probe failure Disconnect ATC. Display should read default 25 ˚C (E101). Replace if necessary
Electronics failure / Bad setup
Try resetting the monitor. Refer to Chapter VI
High Readings Monitor out of calibration Recalibrate. Refer to Chapter IV
Monitor flow cell contaminated
Rinse flow cell with de-ionized water. Flush flow cell overnight with flow at calibration liquid level. Refer to Chapter IV
Inlet filter just replaced Flush one hour. Refer to Chapter IV . If offline calibration performed, verify accuracy of alternate method value. Refer to Chapter IV
Reagent spent or diluted Replace reagent. Refer to Chapter II
VI-�3
Troubleshooting
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Malfunction Possible Cause Remedy
Over-Range Readings
Electrode connections Verify electrode connections to electrode cables. Silde electrode cable back to expose male connector. Be sure to secure female connector on electrode to male connector before twisting to tighten. Refer to Chapter II
Electrode wiring Electrodes not properly wired to terminal strip. Refer to Chapter II
Reference electrode Missing or blocked filling solution. Refer to Chapter II
Electrode failure Electrodes are shorted, cracked, damaged. Install new probes. Refer to Chapter II
Electronics failure / Bad setup / Needs calibration
Try resetting the monitor. Refer to Chapter VI
Default Power loss while back-up battery was not fully charged
Allow battery to charge overnight. Recalibrate and reprogram. Refer to Chapter IV
Battery failure Contact Customer Support for Orion products at 1-800-225-1480
Electronics failure Try resetting the monitor. Refer to Chapter VI or contact Customer Support for Orion products at 1-800-225-1480
Low Readings Instrument okay. Sample very pure
Perform span check procedure. If span check passes, instrument is okay and sample is very pure. Refer to Chapter IV
Electronics failure Try resetting the monitor. Refer to Chapter VI or contact Customer Support for Orion products at 1-800-225-1480
If offline calibration performed, verify accuracy of alternate method value. Refer to Chapter IV
VI-�4
Troubleshooting
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Error/Event .codes .will .be .in .the .format .“E###” . . .Some .of .these .are .errors, .some .are .warnings, .and .some .are .purely .informational . . .The .first .digits .identify .the .type .of .error .or .event .as .identified .in .the .table .below .
Error/Event Codes
Displayed Event Code
Description Cause Troubleshooting
System Status Codes
A7�� ALARM1 CH1 ON Alarm 1 Channel 1 engaged (closed)
A7�� ALARM2 CH1 ON Alarm 2 Channel 1 engaged (closed)
A7�3 ALARM3 CH1 ON Alarm 3 Channel 1 engaged (open)
A7�4 AIR PUMP ON Air pump engaged
A8�� ALARM1 CH1 OFF Alarm 1 Channel 1 disengaged (open)
A8�� ALARM2 CH1 OFF Alarm 2 Channel 1 disengaged (open)
A8�3 ALARM3 CH1 OFF Alarm 3 Channel 1 disengaged (closed)
A8�4 AIR PUMP OFF Air pump disengaged
C500 CAL MODE Calibration mode entered
C509 CAL ISE DKA DKA calibration complete
C50B CAL ISE OFFLINE Offline calibration complete
R400 MEAS MODE Measurement mode entered
S600 SETUP MODE Setup mode entered
E0�8 POWER FAIL Brown out or power failure and system has stopped processing.
Verify custom settings
E033 KEYPAD FAILURE User did not press the keys during a diagnostic mode keypad test, or keypad broken
Repeat diagnostic mode keypad test
E034 CAL PASS INCORRECT User entered incorrect calibration password Verify password
E035 SYSTEM PASS INCORRECT User entered incorrect system password Verify password
E036 USER VALUE INCORRECT User entered a value that is out of range Verify value and re-enter
E037 POWER RETURN System started processing after brown out or power failure
Verify custom settings
E04� CALIBRATION TIMEOUT The electrode has not reach stability within the specified time
The user can press enter to force stability. The system code may be due to out of range measurements or an invalid previous calibration.
E04� NEGATIVE SLOPE (INVALID SLOPE)
Invalid slope obtained during calibration Check electrodes, electrode cables, and connectors for defects or shorts. Replace if necessary. Recalibrate in the correct sequence using standard 1 then standard 2.
VI-�5
Troubleshooting
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Displayed Event Code
Description Cause Troubleshooting
System Event/Error Codes - Yellow LED
E004 DEFAULT VALUES Instrument has been reset or is new out of box
Calibrate the instrument
E0�� HARDWARE FAULT System Error Contact Customer Support for Orion products 1-800-225-1480
E�0� TEMP OUT OF RANGE Temperature sensor on channel 1 out of range (5.1 ˚C to 45 ˚C)
Check ATC cable and probe connections. Replace if needed.
E��� CAL DUE User calibration required. No calibration has been performed in user specified time interval.
Recalibrate the instrument.
E��7 SOUT mA OVER RANGE Measured sodium value above Sout high limit value
Verify customer selectable Sout high limit value in setup menu
E��8 SOUT mA UNDER RANGE Measured sodium value below Sout low limit value
Verify customer selectable Sout low limit value in setup menu
E��9 TOUT mA OVER RANGE Measured temperature value above Tout high limit value
Verify customer selectable Tout high limit value in setup menu
E�30 TOUT mA UNDER RANGE Measured temperature value below Tout low limit value
Verify customer selectable Tout low limit value in setup menu
System Error Codes - Red LED
E001 HARDWARE FAULT System Error Contact Customer Support for Orion products 1-800-225-1480
E002 HARDWARE FAULT System Error Contact Customer Support for Orion products 1-800-225-1480
E005 HARDWARE FAULT System Error Contact Customer Support for Orion products 1-800-225-1480
E006 HARDWARE FAULT System Error Contact Customer Support for Orion products 1-800-225-1480
E018 HARDWARE FAULT System Error Contact Customer Support for Orion products 1-800-225-1480
E019 HARDWARE FAULT System Error Contact Customer Support for Orion products 1-800-225-1480
E020 HARDWARE FAULT System Error Contact Customer Support for Orion products 1-800-225-1480
E022 HARDWARE FAULT System Error Contact Customer Support for Orion products 1-800-225-1480
E024 HARDWARE FAULT System Error Contact Customer Support for Orion products 1-800-225-1480
E026 HARDWARE FAULT System Error Contact Customer Support for Orion products 1-800-225-1480
E027 HARDWARE FAULT System Error Contact Customer Support for Orion products 1-800-225-1480
E029 HARDWARE FAULT System Error Contact Customer Support for Orion products 1-800-225-1480
E030 HARDWARE FAULT System Error Contact Customer Support for Orion products 1-800-225-1480
E038 HARDWARE FAULT System Error Contact Customer Support for Orion products 1-800-225-1480
VI-�6
Troubleshooting
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Displayed Event Code
Description Cause Troubleshooting
System Error Codes - Red LED (cond’t)
E�03 MEAS OUT OF RANGE Sodium measurement out of range, the display should also be flashing ‘9999’ or ‘0000’
Check if water fills flow cell, electrode is not connected, electrode cables are faulty, or reference junction is fouled.
E�09 BAD SLOPE Last calibration produced a bad slope. Slope is more than 63 mV/decade or less than 45 mV/decade, or the calibration standards may have been interchanged.
Recalibrate the instrument using fresh standards. Recalibrate in the correct sequence using standard 1 then standard 2.
E��3 CAL OVERDUE Calibration is overdue by more than 50% of the user specified time interval.
Recalibrate the instrument
E��� TEMP BROKEN Temperature sensor on channel 1 broken or out of range.
Check ATC cable connections and probe for damage. Replace if needed.
E��5 BAD OFFSET Last calibration produced a bad Eo offset.
Recalibrate the instrument using fresh standards. Check reference electrode and sensing electrode connections. Re-etch the sensing electrode. Replace electrodes if needed.
VI-�7
Troubleshooting
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
The .Thermo .Scientific .Orion .2111LL .sodium .monitor .allows .the .user .to .reset .the .monitor .through .the .setup .mode .or .by .a .hard .reset . . .Resetting .the .monitor .will .restore .all .settings .in .the .setup .menu .to .factory .default .values . .(See .ISE .Default .Values)
Warning: .Resetting .the .monitor .will .erase .all .data .logs .including .calibration, .and .password .protection .settings .for .setup .and .calibration . . . .The .instrument .will .reset .setup .parameters .to .factory .default .settings .
User Action MarqueeScrolling
Main Displaytop linebottom line
Notes
Press . to enter setup mode.
Last menu used in setup mode
SELSCrn
SETUP appears as mode
Press . ./ . until RSET appears in the Marquee.
RSET(Flashing)
SELSCrn
Press . .
PRESS TEST VIEW ENTER TO RESET
rSEt?
Press in the following sequence:
1. .
2. . .
3. . .System resets to default values and returns to the measurement mode. Actual values for concentration and temperature will be displayed if the electrodes are operational.
PRESS TEST VIEW ENTER TO RESET
rSEt?
Resetting the Monitor
VI-�8
Troubleshooting
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
A .hard .reset .should .be .performed .if .the .unit .becomes .unresponsive .or .the .status .indicators .are .flashing . . .Perform .the .following .sequence: .
• . Carefully .open .front .cover .of .electronics .case .(loosen .4 .corner .screws) .
• . Press .small .button .located .on .raised .portion .of .PCB .to .reset .electronics . .
• . Close .front .cover .of .electronics .case .and .tighten .the .screws .
System .settings .and .calibrations .are .preserved .and .instrument .returns .to .the .measure .mode . . .Actual .values .for .concentration .and .temperature .will .be .displayed .if .electrodes .are .properly .installed . . .If .system .does .not .reset, .contact .Call .Customer .Support .for .Orion .products .at .1-800-225-1480 . .for .assistance .
In .the .event .you .require .technical .assistance, .please .have .the .serial .number .available .when .calling .for .assistance .along .with .the .software .version .
• . The .monitor .serial .number .is .located .on .the .underside .of .the .electronics .enclosure .or .if .panel .mounted .then .on .the .backside .of .the .electronics .enclosure . .
• . The .software .revision .is .accessed .through .the .test .mode .
Hard Reset
Serial Number and Software Revision
VI-�9
Troubleshooting
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Service and RepairExtended Warranty
The .Thermo .Scientific .Orion .2111LL .sodium .monitor .provides .the .lowest .limits .of .detection .for .sodium .through .a .combination .of .our .superior .ROSS .Ultra® .electrode .technology, .patent-pending .flow .cell .design .and .user .friendly .scrolling .Marquee .to .set .a .new .industry .standard .for .reliability .and .modularity .in .electrochemistry . .Similarly, .we .now .offer .multiple .levels .of .service .programs .to .meet .your .needs .beyond .the .One-Year .Product .Warranty .period .
A .cost .effective .way .to .blend .your .in-house .expertise .with .our .service .and .support .experts, .to .get .the .support .you .need .at .a .specified .annual .contracted .price . . .This .warranty .plan .must .be .purchased .within .ninety .days .of .the .initial .product .purchase .date .
• . Extends .the .features .of .the .standard .One-Year .Product .Warranty .for .a .second .or .third .year . .
• . Provides .access .to .the .manufacturer’s .technical .experts .at .Thermo .Scientific .Orion .Support .Call .Center, .800-225-1480, .Monday .– .Friday, .8:00 .AM .– .5:00 .PM .EST . .
• . After .telephone .consultation, .we .will .send .you .the .required .parts .for .installation .by .your .on-site .technician .
With .in-house .resources .at .a .premium, .many .organizations .require .immediate .access .to .manufacturer .expertise . . .Purchase .this .service .kit .within .ninety .days .of .the .initial .product .purchase .date .to .eliminate .the .costly .effects .of .down .time . .
• . Provides .a .contracted, .cost-effective .means .to .enhance .level .of .support .offered .during .the .product .warranty .period . .
• . Provides .priority .access .to .the .manufacturer’s .technical .experts .at .Thermo .Scientific .Orion .Support .Call .Center .
• . Provides .a .service .kit .to .expedite .on-site .repairs .in .conjunction .with .telephone .consultation .with .our .technical .experts . .
. The .service .kit .includes:• . . Power .supply• . . Front .panel .display• . . Air .pump• . . Sensing .and .reference .electrode .cables
Service Kit to Expand Standard Warranty
VI-�0
Troubleshooting
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
When .down-time .is .a .cost .- .factor .to .be .strictly .controlled, .the .advanced .replacement .service .plan .is .often .times .the .best .choice . . .
• . Enhances .level .of .support .offered .during .the .standard .One-Year .Product .Warranty .
• . Provides .access .to .the .manufacturer’s .technical .experts .at .Thermo .Scientific .Support .Call .Center .and .repair .facilities .
• . If, .after .reasonable .remote .diagnostics .and .trouble-shooting .attempts, .we .determine .the .instrument .to .be .non-functioning, .we .will .make .all .reasonable .efforts .to .get .required .parts .or .equipment .out .to .your .facility .by .the .following .day .for .installation .by .your .on-site .technician . .
• . Offers .predictable .expenditure .through .fixed .annual .contract .price .
Ask .your .Thermo .Scientific .Orion .products .Sales .Representative .to .put .you .in .touch .with .our .service .plan .manager . . .We .have .additional .service .options .that .are .sure .to .address .your .concerns . . .
To .help .you .achieve .optimum .performance, .rely .on .Thermo .Scientific .Orion .products .to .provide .you .with .an .efficient .and .quality .installation .and .the .start-up .support .you .need .to .be .up .and .running .quickly . . .Our .service .representatives .are .highly .trained, .experienced, .and .certified .for .your .product .and .will .be .there .to .make .sure .that .your .instrumentation .delivers .to .specifications .
This .is .a .one-day .service .engineer .on-site .visit .to .your .facility .for .remedial .service .events .at .standard .response .time . . .Consider .this .plan .an .excellent .uplift .to .one .of .the .aforementioned .programs .
All .precision-made .scientific .equipment .requires .preventive .maintenance .and .calibration .checks .to .perform .at .their .maximum .effectiveness . . .Consider .a .fixed-fee .preventive .maintenance .contract .to .have .a .Thermo .Scientific .Orion .products .expert .perform .the .following .tasks:• . Sampling .check• . Full .validation .test• . Visual .check .for .worn, .loose, .or .damaged .parts• . Replacement .of .consumables .(additional .charge .for .consumables)• . Diagnostic .check .to .verify .all .operating .parameters .are .within .the .
factory .specifications• . Check .instrument .electrodes, .flows .and .pressures• . Clean .and .replace .any .wearable .items .such .as .tubing, .O-rings, .reagents .
Advanced Replacement
Not sure which plan is best for you?
Installation and Start-Up
Remedial Service
Preventive Maintenance Programs
VII-�Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Chapter VII Customer Service
This .equipment .generates, .uses, .and .can .radiate .radio .frequency .energy .and .if .not .installed .and .used .in .accordance .with .the .instruction .manual, .may .cause .interference .to .radio .communications . . .It .has .been .tested .and .found .to .comply .with .the .limits .for .a .Class .A .computing .device .pursuant .to .Subpart .J .of .Part .15 .of .FCC .Rules, .which .are .designed .to .provide .reasonable .protection .against .such .interference .when .operated .in .a .commercial .environment . . .Operation .of .this .equipment .in .a .residential .area .is .likely .to .cause .interference .in .which .case .the .user, .at .his .own .expense, .will .be .required .to .take .whatever .measure .may .be .required .to .correct .the .interference .
“This .digital .apparatus .does .not .exceed .the .(Class .A) .limits .for .radio .noise .emissions .from .digital .apparatus .set .out .in .the .Radio .Interference .Regulations .of .the .Canadian .Department .of .Communications .”
“Le .présent .appareil .numérique .n’émet .pas .de .bruits .radioélectriques .dépassant .les .limites .applicables .aux .appareils .numériques .(de .la .class .A) .prescrites .dans .le .Régiement .sur .le .brouillage .radioélectrique .édicté .par .le .ministère .des .Communications .du .Canada .”
Notice of Compliance
VII-�
Customer Service
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Manufacturer: . Thermo .Fisher .Scientific .Inc .
Address: . .166 .Cummings .Center .Beverly, .Ma .01915 .USA
We .declare .that .the .following .products
Model . Electrical .Rating: .2111LL . 85- .132 .VAC; .0 .1 .A .or .170-250 .VAC; .0 .05A, . .50 .- .60 .Hz
Equipment Class: . Measurement, .Control .and .Laboratory .Equipment
Conform .to .the .following .Directives .and .Standards
Safety: . Low .Voltage .Directive .(LVD), .73/23/EEC
. .EN61010-1:2001, .Safety .requirements .for .electrical .equipment .for .measurement, .control .and .laboratory .use .– .general .requirements
EMC: . Electromagnetic .Compatibility .(EMC), .89/336/EEC
. .EN .61326:1997 .+ .A1:1998 .+ .A2:2001 .– .Electrical .equipment .for .measurement, .control, .and .laboratory .use . .EMC .requirements .
Environmental: . IP66
These .products .have .been .manufactured .in .compliance .with .the .provisions .of .the .relevant .Thermo .Fisher .Scientific .Inc . .manufacturing .and .test .documents .and .processes . . .Further, .these .documents .and .processes .are .recognized .as .complying .with .ISO .9001 .by .QMI, .listed .as .File .#001911 .
Place .and .date .of .issues:
Beverly, .MA
July .8, .2005
. . . .Glenn .Benvie .Manger .of .Quality .& .Manufacturing .Engineering
Declaration of Conformity
VII-3
Customer Service
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
For .products .not .listed .in .this .warranty .statement, .please .visit .our .website .at .www .thermo .com/water .
For .updated .contact .information, .visit .www .thermo .com/water .
Thermo .Fisher .Scientific .Inc . .Environmental .Instruments .Division .Water .Analysis .Instruments .166 .Cummings .Center . .Beverly, .MA .01915 .Toll .Free: . .800-225-1480 .Tel: . .978-232-6000 .Dom . .Fax: . .978-232-6015 .Int’l .Fax: . .978-232-6031
The .minimum .order .requirement .is .$100 .for .Orion .Online .products . . .An .order .is .considered .to .be .a .purchase .order .for .products .to .be .shipped .to .a .single .location . . .International .minimum .order .requirements .may .vary . .Contact .your .international .coordinator .for .details . .
For .customers .in .the .U .S ., .rush .orders .received .before .12 .pm .Eastern .Time .will .be .shipped .the .same .day . . .Rush .orders .received .after .12 .Noon .Eastern .Time .will .be .shipped .the .next .business .day .
For .customers .and .dealers .in .Canada, .rush .orders .will .be .shipped .within .2 .business .days . . .For .customers .and .dealers .outside .the .U .S . .and .Canada, .contact .your .international .coordinator .for .rush .order .scheduling . . .All .rush .orders .carry .a .$50 .incremental .charge .per .order . . .FOB: .Beverly . . .Freight .charges .are .prepaid .and .added .or .freight .collect . . .All .rush .order .processing .is .subject .to .stock .availability .
Permission .to .return .merchandise .purchased .from .manufacturer .must .be .obtained .prior .to .return . . .Contact .within .30 .days .of .receipt .of .goods .for .a .return .authorization .number .
Terms and Conditions
Minimum Order
Rush Orders
Returning Goods
Contact Information
VII-4
Customer Service
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Some .materials .are .designated .corrosive/oxidizer .by .DOT .and .IATA . .Some .materials .may .require .special .labeling .and .handling . . .Carriers .may .add .additional .freight .charges .for .handling/transporting .these .materials . .Consolidating .such .material .with .other .products .may .be .prohibited . .Additional .freight .charges .are .billed .to .you .per .FOB .terms . . .Advise .manufacturer .of .shipping .instructions .for .these .hazardous .materials .to .reduce .your .freight .costs .
Permission .to .return .new, .excess .inventory .must .be .obtained .prior .to .return . . .If .any .item .is .authorized .to .be .returned .for .credit .as .a .result .of .an .incorrect .purchase .without .a .reorder, .a .25% .restocking .charge .of .the .price .paid .for .the .product .will .be .made . . .International .customer’s .restocking .fee .of .25% .will .be .off .the .international .price .
Only .new .(in-the-box) .goods .may .be .returned .within .30 .days .of .shipment .from .manufacturer . . .Older .items, .9-digit .parts, .and .discontinued .items .cannot .be .returned .for .credit .
Manufacturer .must .be .notified .within .30 .days .of .receipt .of .invoice .of .any .item .or .billing .discrepancies . . .All .substantiated .claims .will .be .remedied .by .a .credit .memo .and .a .new .order .placed .for .short .shipment . . .Any .shipment .discrepancy .claimed .after .30 .days .of .invoice .date .will .not .be .honored .and .credit .will .not .be .issued .by .manufacturer .
Manufacturer .shall .not .be .liable .for .failure .to .perform .or .for .delay .in .performance .due .to .fire, .flood, .strike, .or .other .labor .difficulty, .act .of .God, .act .of .any .governmental .authority .or .of .the .purchaser, .riot, .embargo, .fuel .or .energy .shortage, .wrecks .or .delays .in .transportation, .inability .to .obtain .necessary .labor, .materials, .or .manufacturing .facilities .from .usual .sources, .or .due .to .any .cause .beyond .its .reasonable .control . . .In .the .event .of .a .delay .in .performance .due .to .any .such .cause, .the .date .of .delivery .or .time .for .completion .of .performance .will .be .extended .by .a .period .of .time .reasonably .necessary .to .overcome .the .effect .of .such .delay .
Restocking Charge
Short Shipments
Force Majeure
Hazardous Materials
VII-5
Customer Service
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Orion .Online .monitoring .products .are .warranted .to .be .free .from .defects .in .material .and .workmanship .for .a .period .of .12 .months .from .date .of .installation .or .18 .months .from .date .of .shipment, .whichever .is .earlier, .when .used .under .normal .operating .conditions .and .in .accordance .with .the .operating .limitations .and .maintenance .procedures .given .in .the .instruction .manual, .and .when .not .having .been .subjected .to .accident, .alteration, .misuse, .or .abuse . . .This .warranty .is .also .conditioned .upon .expendable .and .consumable .items .(diffusion .tubing, .electrodes, .and .all .solutions) .being .stored .at .temperatures .between .5 .°C .and .45 .°C .(40 .°F .and .110 .°F) .in .a .non-corrosive .atmosphere .and .within .the .shelf .life .printed .on .the .product .
In .the .event .of .failure .within .the .warranty .period, .manufacturer, .or .its .authorized .dealer, .will, .at .the .option .of .manufacturer, .repair .or .replace .the .product .nonconforming .to .the .above .warranty, .or .will .refund .the .purchase .price .of .the .unit .
The .warranty .described .is .exclusive .and .in .lieu .of .all .other .warranties .whether .statutory, .express, .or .implied .including, .but .not .limited .to, .any .implied .warranty .of .merchantability .or .fitness .for .a .particular .purpose .and .all .warranties .arising .from .the .course .of .dealing .or .usage .of .trade . . .The .buyer’s .sole .and .exclusive .remedy .is .for .repair .or .replacement .of .the .non-conforming .part .thereof, .or .refund .of .the .purchase .price, .but .in .no .event .shall .manufacturer .(its .contractors .and .suppliers .of .any .tier) .be .liable .to .the .buyer .or .any .person .for .any .special .indirect, .incidental, .or .consequential .damages .whether .the .claims .are .used .in .contract, .in .tort .(including .negligence), .or .otherwise .with .respect .to .or .arising .out .of .the .product .furnished .hereunder .
Online .products .used .at .overseas .nuclear .facilities .are .also .subject .to .manufacturer’s .nuclear .terms .and .conditions . .Contact .manufacturer .if .you .do .not .have .a .copy .
Representations .and .warranties .made .by .any .person, .including .its .authorized .dealers, .distributors, .representatives, .and .employees .of .manufacturer, .which .are .inconsistent .or .in .addition .to .the .terms .of .this .warranty .shall .not .be .binding .upon .manufacturer .unless .in .writing .and .signed .by .one .of .its .officers .
Warranty
A-1Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Appendix
Mounting Dimensions
A-2
Appendix
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
E0 default: - 70.0 mV
Low limit of E0: - 130 mV
High limit of E0: - 10 mV
Slope default: 59.15 mV/dec
Slope range: 45.0 mV/dec to 62.0 mV/dec
CISO: 115 ppm = 115000 ppb
Volume of flow cell (fixed): 190 mL
Concentration of standard 1 added during cal = 19.1 ppm
Volume of standard 1 added during cal = 0.1 mL
Concentration of standard 2 added during cal = 192 ppm
Volume of standard 2 added during cal = 0.1 mL
ISE Default Values
A-3
Appendix
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Sodium ISE
Range 0.001 ppb – 10 ppm
Resolution 2, 3, or 4 significant digits
Accuracy (w/DKA cal) ± 5% or 0.01 ppb, whichever is greater. Accuracy not guaranteed above 10 ppm.
Accuracy (w/DKA cal & offline cal) ± 2.5% or 0.01 ppb, whichever is greater (assuming “true” value from the reference)
Precision Std. Dev. ± 2.5% or 0.01 ppb, whichever is greater
Response time 95% with in 2 minutes: with freshly etched electrode
• Reach 95% of final reading within 2 minutes of injecting a standard solution
or
• 2 minutes, max. for time from 5% to 95% of measured change following dynamic addition of a standard solution
Units displayed ppb, ppm
Temperature compensation Automatic
Range ± 1999.9 mV
Resolution 0.1 mV
Relative accuracy ± (0.5 mV + 0.1%)
Temperature Measurement
Range -10 to 120 ˚C
Resolution 0.1 ˚C
Relative accuracy ± 0.5 ˚C
Failure detection Manual compensation
ATC probe 30 K thermistor
LED Status Indicator
Green OK
Yellow Warning
Red Failure
Specifications
A-4
Appendix
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Sample Conditions
Temperature 5 to 45 ˚C
Total acidity Less than 50 ppm CaCO3
Inlet pressure 8 to 100 psig
Flow rate 40 mL/min nominal set by pressure regulator 1/4” NPTF tube fitting
Sample drain 3/4” NPT male
Grab sampler Optional
Reagents Diisopropylamine
Display Features
Type Backlit custom LCD
Size 54 x 76 mm (2 1/8 in x 3 in)
Marquee Temperature, user prompts
Middle line Concentration, error codes
Lower line mV (selectable)
Text display Scrolling marquee for English language instruction for setup, calibration and diagnostics
Calibration
DKA calibration Yes
DKA calibration points 3 points
Off-line calibration Yes
Off-line calibration points 1 point
Pre-programmed standard values Yes
Custom calibration User programmable for concentration and volume
User input DKA, customer programmable, off-line
Keypad 8 electro-mechanical keys with graphic overlay
Inputs (potentiometric)
Sodium ion input ROSS® sodium
Sodium ion reference ROSS Ultra® reference
A-5
Appendix
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Outputs
Number of analog outputs 2 (shared ground)
Galvanically isolated Yes
Output selections 0 to 20 mA, 4 to 20 mA
Programmable range YES
Channel assignments Sodium & temperature
Relative accuracy ± (0.05 mA + 0.5%)
Max. load 500 ohm or 10 V
Log & linear output options Yes, user selectable
Alarm Outputs
Number of relay outputs 3
Max relay load 250 VAC/5A, 30 VDC/5A, resistive load only
Min. value alarm Yes
Max. value alarm Yes
Error alarm Yes
Offline/cal alarm Yes
Programmable min. & max. values Yes
Meter features
Non-volatile memory Yes
Battery backup (replaceable) Yes
Data logging Yes
Software Features
Self-test & diagnostics Yes
Real time clock Yes
Programmable alarms High, low, error, cal/offline
Reset function Yes
Measurement, active Signal noise, temperature
Calibration Sensor response, drift, E0, slope, temperature
A-6
Appendix
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Miscellaneous
Graphics Custom chemical resistant
Cleaning instructions Wipe with damp cloth
Case color Grey
Case material Valox 364
Warranty 12 months from date of installation or 18 months from date of shipment, whichever is earlier
Service policy Field service and factory
Manual CD/printed/web
Regulatory and safety CE, CSA, FCC class A limits
Physical size of monitor 65 x 45 x 27 cm (26 x 17 x 11 in)Fits on 1811EL panel mounting hardware footprint Weight: 22.7 kg (50 lbs)
Physical size of electronics module Size (mm): 144 x 144 x 187 (in): 5.67 x 5.67 x 7.36
Weight (kg): 1.7 (lbs): 3.75
Power input 85 - 132 V; 200 mA or 170-264 V; 100 mA 50 - 60 Hz AC
Over voltage category II
Environmental Operating Conditions
Waterproof enclosure IP66 & NEMA 4X
Overvoltage category II
Ambient temperature 5 to 45 ˚C
Relative humidity Up to 95% non-condensing
Pollution degree 2
Storage temperature -20 to 60 ˚C
Storage humidity Up to 95%, non-condensing
Altitude Sea level to 2000 M
Shock and Vibration
Vibration, shipping/handling 0 - 60 Hz @ 1 G Load
Shock, drop test in packaging 36” on all sides and corners
A-7
Appendix
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Cat. No. Description
2111LL Low level sodium monitor with ROSS® sodium electrode (210048), ROSS Ultra® reference electrode (210058), ATC probe (2100TP) and standard solutions kit (181140)
2111LLEN Low level sodium monitor with enclosure, ROSS® sodium electrode (210048), ROSS Ultra® reference electrode (210058), ATC probe (2100TP) and standard solutions kit (181140)
2111FP Low level sodium monitor, fluidics only
211100 Low level sodium monitor, electronics only
181148 Calibration kit with carrying case, 0.1 mL pipette, box of pipette tips, wash bottle, (1) x 2 oz bottle standard 1 (19.1 ppm Na+ standard), (1) x 2 oz bottle standard 2 (192 ppm Na+ standard) and (1) x 2 oz bottle of etch solution
181140 Sodium standard solution kit with (2) x 2 oz bottles of standard 1 (19.1 ppm Na+ standard), (2) x 2 oz bottles of standard 2 (192 ppm Na+ standard) and (1) x 2 oz bottle of etch solution (181113)
181141 High level sodium standard solution kit with (2) x 2 oz bottles of standard 1 (1910 ppm Na+ standard), (2) x 2 oz bottles of standard 2 (19200 ppm Na+ standard) and (1) x 2 oz bottle of etch solution
181113 Sodium electrode etch solution, (1) x 2 oz. bottle
210048 ROSS® sodium electrode with screw cap
210058 ROSS Ultra® reference electrode with screw cap
150072 (5) x 2 oz bottles of reference electrode filling solution
2111EK Electrode kit with (1) sodium electrode (210048) and (1) reference electrode (210058)
180029 Electronic test kit for troubleshooting 1500/1700/1800/2100 series analyzers
21003M (2) x 3 meter extension cables for 210048 and 210058 electrodes
2100TP 2100 series automatic temperature compensation probe with ground (30 kΩ)
2001XT 10 meter extension cable and 2100 series automatic temperature compensation probe
211192 Sodium consumables kit for one year - includes: (6) Diisopropylamine (DIPA) reagent (6 x 0.8 L bottle) with (6) diffusion tubing kit (211194), reagent bottle adapter (2111RBA), (1) standard solutions kit (19.1 ppm and 192 ppm Na+ standards and etch solution), (1) sodium electrode, (1) reference electrode, (5) x 2 oz bottles reference filling solution, (1) O-ring kit, (1) check valve, (1) set of 100 pipette tips, (1) restriction tube assembly, (2) 60 micron inlet filters and (1) filter gasket
211190 (1) Diisopropylamine (DIPA) reagent (1 x 0.8 L bottle) with diffusion tubing kit
211191 (6) Diisopropylamine (DIPA) reagent (6 x 0.8 L bottle) with (6) diffusion tubing kits
211194 Diffusion tubing kit (1 complete assembly and O-rings)
Ordering Information
A-8
Appendix
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Cat. No. 2111LL Accessory Options
2111EN 2100 series enclosure for use with 2111LL low level sodium monitor
18GRAB 1800/2100 series QC sampler includes (1) x 1 L polypropylene sample bottle, tubing, fittings, power supplies and instruction sheet.
18GBTL 4 x 1 L polypropylene bottles with covers, for use with 18GRAB
18GTBK QC sampler replacement tubing kit for use with 18GRAB
224329-A01 Wall adapter, 115 V for use with 18GRAB
224330-A01 Wall adapter, 230 V for use with 18GRAB
180300 Steam sample conditioner for high-pressure steam samples for use with 1800/2100 series monitors
15DC15 Dynamic calibrator, complete with syringe pump and syringe kits (150096 and 180096) and mounting bracket, 115 V
15DC20 Dynamic calibrator, complete with syringe pump and syringe kits (150096 and 180096) and mounting bracket, 230 V
216970-A01 Low-level verification kit, includes T-fittings and tubing for use with dynamic calibrator systems (15DC15) and (15DC20)
A-9
Appendix
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Cat. No. 2111LL Field Replaceable Parts
2111FP Fluidics panel assembly
2111NV Inlet valve assembly
2111BV Needle valve assembly
2111RG Regulator assembly
2111FM Flow meter assembly
2111SV Calibration shut off valve assembly
2111RT Restrictor tube assembly
2111RMA Reagent manifold assembly
2111RBA Reagent bottle adapter assembly
2111DBB Reagent bottle bracket for DIPA reagent bottle
2111NT Inlet tube assembly
2111EC (2) x 1 meter electrode cables for 210048 and 210058
2111PA Air pump assembly
2111RVA Right air valve
2111LVA Left air valve
2111AF Air filter
2111VC Check valve
2111TG Tygon tubing (4 feet)
2111BP Steel back panel
2111EP 2111LL electronics faceplate
2111PS Power supply assembly
2111FK115 Fuse kit assembly, 115V
2111FK230 Fuse kit assembly, 230V
2111FC Flow cell assembly (complete)
2111CA Flow cell cap assembly
2111LLOK O-ring kit for 2111LL
2111ST Sensing tube (flow cell)
2111MK Maintenance kit includes: (2) electrode cables, (1) diffusion tubing kit, (1) sample filter, (1) O-ring kit, (1) check valve, (1) sensing tube, (1) shut off valve, (1) restrictor tubing assembly, (1) air filter, (1) inlet tubing assembly, and (1) reagent bottle cap and adapter
181170 (2) Inlet sample filters
A-10
Appendix
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Cat. No. Description Quantity
211192 Sodium consumables kit for one year - includes: (6) Diisopropylamine (DIPA) reagent (6 x 0.8 L bottle) with (6) diffusion tubing kit (211194), reagent bottle adapter (2111RBA), (1) standard solutions kit (19.1 ppm and 192 ppm Na+ standards and etch solution), (1) sodium electrode, (1) reference electrode, (5) x 2 oz bottles reference filling solution, (1) O-ring kit, (1) check valve, (1) set of 100 pipette tips, (1) restriction tube assembly, (2) 60 micron inlet filters and (1) filter gasket
1
181148 Calibration kit with carrying case, 0.1 mL pipette, box of pipette tips, wash bottle, (1) x 2 oz bottle standard 1 (19.1 ppm Na+ standard), (1) x 2 oz bottle standard 2 (192 ppm Na+ standard) and (1) x 2 oz bottle of etch solution
1
Cat. No. Description Quantity
2111RG Regulator assembly 1
2111PA Air pump assembly 1
2111FM Flow meter assembly 1
2111EK Electrode kit with (1) 210048 sodium electrode and (1) 210058 reference electrode
1
2111MK* Maintenance kit includes: (2) electrode cables, (1) diffusion tubing kit, (1) sample filter, (1) O-ring kit, (1) check valve, (1) sensing tube, (1) shut off valve, (1) restrictor tubing assembly, (1) air filter, (1) inlet tubing assembly and (1) reagent bottle cap and adapter
1 per unit
2111LLSK* Service kit includes: (1) electronics faceplate (2111EP), (1) air pump assembly (2100PA), (1) regulator assembly (2100RG), (2) electrode cable with gland assemblies for sensing and reference electrodes, (1) power supply assembly (2100PS) and (1) fuse kit assembly (2100FK115 and 2100FK230)
1 per unit
* For customers with multiple units to support
Recommended Consumables for
Annual Operation
Recommended Field Replaceable Spare Parts
A-11
Appendix
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
To help eliminate the risk of contamination, each pipette is fitted with a tip ejector system. The tip ejector system consists of a soft-touch tip ejector and specially designed gearing mechanism. To release the tip, point the pipette at suitable waste receptacle and press the tip ejector with your thumb.
You can attach the pipette shelf hanger on a counter, pipette stand or anywhere where you want to hang your pipette.
Clean the area where you plan to attach the shelf hanger. Apply two stickers to the underside of the shelf hanger. Press the shelf hanger firmly into place – on a shelf, counter top or pipette stand. To use, hang the grippy finger rest on the shelf hanger.
Pipette OperationTip ejection
A-12
Appendix
Thermo Scientific Orion 2111LL Sodium Monitor User GuideROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Special care shall be taken to avoid liquid being aspirated into the interior working of the pipette. Push and release the push button slowly at all times, particularly when working with high viscosity liquids. Never allow the push button to snap back. Make sure that the tip is firmly attached to the tip cone. Check for foreign particles in the tip.
• Do not touch pipette tip with bare fingers. Risk of sodium contamination is possible. Use powder free gloves for protection.
Before you begin your actual pipetting work, fill and empty the tip 3-5 times with the solution that you will be pipetting. Hold the pipette in an upright position while aspirating liquid. The grippy should rest on your index finger. Make sure that the tips, pipette and solution are at the same temperature.
Fill a clean reagent reservoir with the liquid to be dispensed.
1. Depress the push button to the first stop.
2. Dip the tip under the surface of the liquid in the reservoir to a depth of about 1 cm and slowly release the push button. Withdraw the tip from the liquid touching it against the edge of the reservoir to remove excess liquid.
3. Deliver the liquid by gently depressing the push button to the first stop. After a delay of about one second, continue to depress the push button all the way to the second stop. This action will empty the tip.
4. Release the push button to the ready position. If necessary, change the tip and continue pipetting.
Pipette TechniquesGeneral
Forward Technique
Environmental Instruments Water Analysis Instruments
North America 166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 Dom. Fax: 1-978-232-6015 Int’l Fax: 978-232-6031 www.thermo.com/water
Europe12-16 Sedgeway Business Park Witchford, Cambridgeshire England, CB6 2HY Tel: 44-1353-666111 Fax: 44-1353-666001
AsiaBlk 55, Ayer Rajah Crescent #04-16/24, Singapore 139949 Tel: 65-6778-6876 Fax: 65-6773-0836© 2007 Thermo Fisher Scientific Inc. All rights reserved.
250596-003 RevE 0807