tlc-accutofdart2008/10/23 · lc/ms reverse-phase normal-phase 9possible to confirm the mass...
TRANSCRIPT
![Page 1: TLC-AccuTOFDART2008/10/23 · LC/MS Reverse-phase Normal-phase 9Possible to confirm the mass information of products and by-products. 9Impossible to know the condition for isolating](https://reader031.vdocuments.net/reader031/viewer/2022011910/5f8086bcd738b629fa6018b1/html5/thumbnails/1.jpg)
TLCTLC--AccuTOFAccuTOF DARTDART
10ème anniversaire du Club de CCM23/10/2008
Akihiko KUSAI, JEOL (Europe) SAS
![Page 2: TLC-AccuTOFDART2008/10/23 · LC/MS Reverse-phase Normal-phase 9Possible to confirm the mass information of products and by-products. 9Impossible to know the condition for isolating](https://reader031.vdocuments.net/reader031/viewer/2022011910/5f8086bcd738b629fa6018b1/html5/thumbnails/2.jpg)
AccuTOF DART
Unless marked otherwise, DART is interfaced to a JEOL AccuTOFtm
time-of-flight mass spectrometerHigh resolution and exact mass measurements provide selectivityHigh dynamic range makes it easyto measure mixturesExact masses and accurate isotopic abundances permit identification of unknownsFast data acquisition for high throughputMass-independent abundance response (small molecules)
![Page 3: TLC-AccuTOFDART2008/10/23 · LC/MS Reverse-phase Normal-phase 9Possible to confirm the mass information of products and by-products. 9Impossible to know the condition for isolating](https://reader031.vdocuments.net/reader031/viewer/2022011910/5f8086bcd738b629fa6018b1/html5/thumbnails/3.jpg)
What is DART?
DART is a rapid and non-contact surface sampling technique for mass spectrometry at atmospheric pressure under ambient conditions. (Ambient ionization technique)DART can be used to analyze any compounds; gases, liquids, solids and materials on surfaces.DART is suitable to analyze small-moleculecompounds (with some exceptions).
Analyte mast be vaporized in the gas phase.
![Page 4: TLC-AccuTOFDART2008/10/23 · LC/MS Reverse-phase Normal-phase 9Possible to confirm the mass information of products and by-products. 9Impossible to know the condition for isolating](https://reader031.vdocuments.net/reader031/viewer/2022011910/5f8086bcd738b629fa6018b1/html5/thumbnails/4.jpg)
DART Review
Compounds in gas phase are Ionized by metastables
He, N2, …
Electron, charged molecules, and energetic atoms or gas molecule are formed by glow discharge.
Electrodes block the charged particles and electrons.
![Page 5: TLC-AccuTOFDART2008/10/23 · LC/MS Reverse-phase Normal-phase 9Possible to confirm the mass information of products and by-products. 9Impossible to know the condition for isolating](https://reader031.vdocuments.net/reader031/viewer/2022011910/5f8086bcd738b629fa6018b1/html5/thumbnails/5.jpg)
Features of DART
Analysis time is very quick: analysis completes in seconds.
TOFMS is suitable for DART due to fast data samplingDART operation is very simple.
Open-air operationNo vacuum, No solvent, No carryover
DART is not susceptible to high levels of saltsDART is applicable to any compounds in gas phase
Gases, liquids, solids and ,materials on surfacesDART spectrum is very simple like a APCI spectrum
No alkali-metal adducts or multiple charges
![Page 6: TLC-AccuTOFDART2008/10/23 · LC/MS Reverse-phase Normal-phase 9Possible to confirm the mass information of products and by-products. 9Impossible to know the condition for isolating](https://reader031.vdocuments.net/reader031/viewer/2022011910/5f8086bcd738b629fa6018b1/html5/thumbnails/6.jpg)
With a glass rod for liquid sample
With tweezers for solid sample
Sample introduction
With a weighting paper for powder sample, not solved sample, etc.
![Page 7: TLC-AccuTOFDART2008/10/23 · LC/MS Reverse-phase Normal-phase 9Possible to confirm the mass information of products and by-products. 9Impossible to know the condition for isolating](https://reader031.vdocuments.net/reader031/viewer/2022011910/5f8086bcd738b629fa6018b1/html5/thumbnails/7.jpg)
DART Applications
Forensics
Drugs (medicinal, herbal, illicit, counterfeit)
ExplosivesChemical warfare agents
Metabolites
Synthetic organics
Foods & beveragesPesticides
Toxic industrial materials
Polymer & additives
![Page 8: TLC-AccuTOFDART2008/10/23 · LC/MS Reverse-phase Normal-phase 9Possible to confirm the mass information of products and by-products. 9Impossible to know the condition for isolating](https://reader031.vdocuments.net/reader031/viewer/2022011910/5f8086bcd738b629fa6018b1/html5/thumbnails/8.jpg)
A BA + B → AB
PurificationPurification
CABD
Silica-gel
A Bsample
TLC
AB?
Reaction Reaction monitoringmonitoring
LC/MS Reverse-phase
Normal-phase
Possible to confirm the mass information of products and by-products.
Impossible to know the condition for isolating the products
EasySimple
Normal-phase
How DART can be used in the synthetic organics
BAB C
DA
Direct reaction monitoring by DART
Check product Check product by DARTby DART
To the next step
![Page 9: TLC-AccuTOFDART2008/10/23 · LC/MS Reverse-phase Normal-phase 9Possible to confirm the mass information of products and by-products. 9Impossible to know the condition for isolating](https://reader031.vdocuments.net/reader031/viewer/2022011910/5f8086bcd738b629fa6018b1/html5/thumbnails/9.jpg)
1,5 hexanediol acetylation (add excess HOAc at 90 minutes)
-20
0
20
40
60
80
100
120
0 20 40 60 80 100 120 140
Time
Rel
. abu
ndan
ce (%
)
1,5-HexanediolMonoacetateDiacetate
Direct reaction monitoring
1,5-Hexanediol acetylation (add excess HOAc at 90 min.)
1,5-Hexanediol
Diacetate
Monoacetate
![Page 10: TLC-AccuTOFDART2008/10/23 · LC/MS Reverse-phase Normal-phase 9Possible to confirm the mass information of products and by-products. 9Impossible to know the condition for isolating](https://reader031.vdocuments.net/reader031/viewer/2022011910/5f8086bcd738b629fa6018b1/html5/thumbnails/10.jpg)
Published application by TLC-DART
Molrock, Gertrud and Ueda, Yoshihisa. New coupling of planar chromatography with direct analysis in real time mass spectrometry. Jounal of chromatography A, 1143, 2007, 243-251Natalle J. Smith, Marek A Domin, and Lawrence T. Scott, HRMS Directly From TLC Slides. A Powerful Tool for Rapid Analysis of Organic Mixtures, Organic Letters, 2008, Vol.10, No.16, 3493-3496.
![Page 11: TLC-AccuTOFDART2008/10/23 · LC/MS Reverse-phase Normal-phase 9Possible to confirm the mass information of products and by-products. 9Impossible to know the condition for isolating](https://reader031.vdocuments.net/reader031/viewer/2022011910/5f8086bcd738b629fa6018b1/html5/thumbnails/11.jpg)
TLC plate sampler
Maximum size of TLC plate: 10mm × 100mm
TLC sampler
![Page 12: TLC-AccuTOFDART2008/10/23 · LC/MS Reverse-phase Normal-phase 9Possible to confirm the mass information of products and by-products. 9Impossible to know the condition for isolating](https://reader031.vdocuments.net/reader031/viewer/2022011910/5f8086bcd738b629fa6018b1/html5/thumbnails/12.jpg)
Overlapped spots
MS[1];0.311..0.329;- 1.0*MS[1];0.346..0.391; / ESI+ / azo test 001
100 150 200 250 300Ž ¿—Ê“ d‰×” ä(m/ z)
0
10
20
30
40
‹ -“ x (41066)x103
198.10
199.11
197.10
m/z 245
MS[1];0.390..0.412;- 1.0*MS[1];0.329..0.374; / ESI+ / azo test 001
100 150 200 250 300Ž ¿—Ê“ d‰×” ä(m/ z)
0
1000
2000
‹ -“ x (2955)
245.13
227.16
244.12
226.15120.08
246.14228.17177.12135.10
119.07 298.20163.14
m/z 198
m/z 227 m/z 245
4-aminoazobenzene
3,3’-dimethoxybenzidine4,4’-diamino-3,3’-dimethyldiphenylmethane
Analysis of azo compounds by TLC/DART
![Page 13: TLC-AccuTOFDART2008/10/23 · LC/MS Reverse-phase Normal-phase 9Possible to confirm the mass information of products and by-products. 9Impossible to know the condition for isolating](https://reader031.vdocuments.net/reader031/viewer/2022011910/5f8086bcd738b629fa6018b1/html5/thumbnails/13.jpg)
Feature of TLC Auto-Slider
Constant speed drivingControllable of driving speed (0.5, 1.0, 2.0 and 4.0 mm/sec)
Good reproducibilityImprovement of making a chromatogramAvailable to obtain Rfvalue
![Page 14: TLC-AccuTOFDART2008/10/23 · LC/MS Reverse-phase Normal-phase 9Possible to confirm the mass information of products and by-products. 9Impossible to know the condition for isolating](https://reader031.vdocuments.net/reader031/viewer/2022011910/5f8086bcd738b629fa6018b1/html5/thumbnails/14.jpg)
ƒ }ƒ Xƒ Nƒ �[1];748.50700..748.62900; / ESI+ / clarith TLC 004
0 0.2 0.4 0.6 0.8 1.0Œo‰ßŽ žŠ Ô[min]
0
10
‹ -“ x (14352)x103
0.113
Rf = 0.113(min) / 0.5(min) = 0.23Spot size = ca. 5(mm)
ca. 5 sec
Clarithromycin (C38H69NO13)
Sliding speed : 1.0 mm/sec
MS[1];0.104..0.128;- 1.0*MS[1];0.187..0.222; / ESI+ / clarith TLC 004
200 400 600 800 1000Ž ¿—Ê“ d‰×” ä(m/ z)
0
5
10
‹ -“ x (13241)x103
748.56
149.10
749.57
590.46750.58
[M+H]+748.56
Analysis of macrolide antibiotics extracted by CHCl3
Measurement end
CHCl3/CH3OH=9/1
30mm
<JMS<JMS--T100TD>T100TD>
![Page 15: TLC-AccuTOFDART2008/10/23 · LC/MS Reverse-phase Normal-phase 9Possible to confirm the mass information of products and by-products. 9Impossible to know the condition for isolating](https://reader031.vdocuments.net/reader031/viewer/2022011910/5f8086bcd738b629fa6018b1/html5/thumbnails/15.jpg)
Conclusion
DART analysis is fast, easy and “no sample prep” (most of the time…)Simple mass spectra is obtained by DART. It is easy to make a interpretation.TLC/MS is the useful tools for high-throughput identification in the synthetic organics.