top-down analysis of immunoglobulin g isotypes 1 and 2
TRANSCRIPT
Top-down analysis of immunoglobulin G isotypes 1 and
2 with electron transfer dissociation on a high-field
Orbitrap mass spectrometer.
Luca Fornelli1¶, Daniel Ayoub1, Konstantin Aizikov2, Xiaowen Liu3,4, Eugen
Damoc2, Pavel A. Pevzner5, Alexander Makarov2, Alain Beck6, and Yury O.
Tsybin1,7*
1 Biomolecular Mass Spectrometry Laboratory, Ecole Polytechnique Fédérale de
Lausanne, 1015 Lausanne, Switzerland
2 Thermo Fisher Scientific GmbH, 28199 Bremen, Germany
3 Department of BioHealth Informatics, Indiana University-Purdue University
Indianapolis, 46202 Indianapolis, IN, USA
4 Center for Computational Biology and Bioinformatics, Indiana University
School of Medicine, 46202 Indianapolis, IN, USA
5 Department of Computer Science and Engineering, University of California in
San Diego, 92093 San Diego, CA, USA
6 Centre d’Immunologie Pierre Fabre, 74160 St Julien-en-Genevois, France
7 Spectroswiss Sàrl, EPFL Innovation Park, 1015 Lausanne, Switzerland
_________________________________________________________________________________ This is the author's manuscript of the article published in final edited form as:Fornelli, L., Ayoub, D., Aizikov, K., Liu, X., Damoc, E., Pevzner, P. A., … Tsybin, Y. O. (2017). Top-down analysis of immunoglobulin G isotypes 1 and 2 with electron transfer dissociation on a high-field Orbitrap mass spectrometer. Journal of Proteomics, 159, 67–76. https://doi.org/10.1016/j.jprot.2017.02.013
2
*Correspondence should be addressed to Dr. Yury O. Tsybin, Spectroswiss Sàrl,
EPFL Innovation Park, Building I, 1015 Lausanne, Switzerland. Email:
[email protected], phone +41 21 693 98 06
¶ Present address: Northwestern University, 2170 Campus Dr, 60208 Evanston,
IL, USA
Present address: Luxembourg Institute of Health, 1445 Strassen, Luxembourg
Running title: Top-down ETD of IgGs on Orbitrap Elite
List of abbreviation:
Automatic gain control, AGC
Electron transfer dissociation, ETD
Electron transfer dissociation – higher energy collisional dissociation, EThcD
Fourier transform mass spectrometry, FTMS
Keywords: electron transfer dissociation, ETD; immunoglobulin G, IgG;
Orbitrap, top-down
Manuscript date: June 29, 2017
3
(Abstract)
The increasing importance of immunoglobulins G (IgGs) as biotherapeutics calls
for improved structural characterization methods designed for these large (~150
kDa) macromolecules. Analysis workflows have to be rapid, robust, and require
minimal sample preparation. In a previous work we showed the potential of
Orbitrap Fourier transform mass spectrometry (FTMS) combined with electron
transfer dissociation (ETD) for the top-down investigation of an intact IgG1,
resulting in ~30% sequence coverage. Here, we describe a top-down analysis of
two IgGs1 (adalimumab and trastuzumab) and one IgG2 (panitumumab)
performed with ETD on a mass spectrometer equipped with a high-field Orbitrap
mass analyzer. For the IgGs1, sequence coverage comparable to the previous
results was achieved in a two-fold reduced number of summed transients, which
corresponds, taken together with the significantly increased spectra acquisition
rate, to ~six-fold improvement in analysis time. Furthermore, we studied the
influence of ion-ion interaction times on ETD product ions for IgGs1, and the
differences in fragmentation behavior between IgGs1 and IgG2, which present
structural differences. Overall, these results reinforce the hypothesis that gas
phase dissociation using both energy threshold-based and radical-driven ion
activations is directed to specific regions of the polypeptide chains mostly by the
location of disulfide bonds.
4
1. Introduction
Monoclonal antibodies (mAbs) are an effective therapeutic tool in a variety of
medical conditions such as cancer and inflammatory disease.[1] These large
glycoproteins constitute the fastest growing class of biotherapeutics.[2, 3] The
main class of antibodies employed as biopharmaceuticals is immunoglobulin G
(IgG). IgGs are large (~150 kDa) biomolecules with a quaternary structure made
of two ~25 kDa light and two ~50 kDa heavy chains, identical among themselves,
with the latter being N-glycosylated, Supporting Information Figure S1. The four
chains are interconnected and their tertiary/quaternary structure is reinforced
by inter- and intra-molecular disulfide bridges. Heavy chains can be translated
from different genes, and their sequence determines the so-called IgG isotype (or
subclass). In the pharmaceutical industry, three of the four IgG isotypes are of
interest: IgG1, IgG2, and IgG4. Although all isotypes share ~90% of homology in
their amino acid sequences, the structure of their hinge regions and the disulfide
bond connectivity differ significantly. IgG1 and IgG4 contain 16 disulfide bridges,
two in the hinge region linking the heavy chains, one disulfide bridge linking
each light chain to the corresponding heavy chain and 12 intramolecular bridges
(2 and 4 for each light and heavy chain, respectively). IgG2 contains 18 disulfide
bonds with four in the hinge region linking the two heavy chains together.[4, 5]
Figure S1 compares the structures of an IgG1 and an IgG2. Note that the C-
terminal cysteine of the light chain is linked to the third cysteine (from the N-
terminus) of the heavy chain in IgG2 (and also IgG4), whereas in IgG1 it is linked
to the fifth cysteine.
5
Due to their relevance as biotherapeutics, and with the raise of the so-
called biosimilars in the pharmaceutical market,[3, 6] high-throughput and
extensive characterization of the sequence and structure of IgGs is needed.[2]
Mass spectrometry (MS) plays a pivotal role in analysis of IgGs. MS-based
methods have been developed and implemented across all stages of mAb
production.[4] The basic workflow may include intact IgG mass measurement,
IgG fragment (reduced light and heavy chains or papain-produced fragments)
mass measurement (referred to as middle-up MS approach),[7] middle-down,[8,
9] extended bottom-up[10] and bottom-up MS[2, 11] approaches. Although
bottom-up approaches provide the most structural information today, they suffer
a number of drawbacks such as artifact introduction and lengthy sample
preparation.[12-14] Top-down mass spectrometry (TD MS)[15] refers to the gas-
phase dissociation of intact biomolecules and subsequent mass measurement of
the fragmentation products. Applied to IgGs, TD MS is a convenient way to obtain
primary structure information without the need of reducing the inter-chain
disulfide bonds in solution. Furthermore, this approach generally allows access
to sequence confirmation of terminal regions as well as variable domains. It also
allows the identification and localization of major modifications and the
characterization of major glycoforms.[16] It has the advantage of requiring a
highly simplified sample preparation and therefore minimal artifact introduction
induced by sample processing. Despite these advantages, TD MS is not yet
routinely used for the characterization of mAbs, primarily due to the large
number of highly charged products yielded by IgG fragmentation, which requires
6
high-resolution MS and the applications of techniques aimed at improving
sensitivity, such as spectral summation (averaging).[17]
Electron capture dissociation (ECD)[18, 19] and electron transfer
dissociation (ETD)[20] of whole proteins usually generate larger sequence
coverage than slow-heating activation methods (such as collision-induced
dissociation, CID).[21] ECD and ETD MS/MS are also known as powerful
methods for characterization of labile post-translational modifications, e.g.,
glycosylation.[22, 23] Fourier transform ion cyclotron resonance (FT-ICR) MS
provides the highest resolution and can be hyphenated with ECD, making it the
technique of choice for earlier TD MS experiments.[24-26] Notably, Mao et al.
achieved 25% sequence coverage of a purified IgG1 using qFT-ICR MS and ECD
when isolating a single charge state in a direct infusion ionization mode. They
increased the sequence coverage to ~33% with a broadband tandem mass
spectrometry (MS/MS), e.g., using all the IgG1 ions present in the charge state
envelope as precursors for ECD.[27] However, the general use of FT-ICR MS
instruments for IgG characterization is rare,[4] presumably due to their limited
presence in biopharmaceutical laboratories owing to the tedious maintenance
and operational costs, as well as oftentimes low efficiency of heavy precursor ion
transfer and capture in the ICR cells for subsequent ECD reaction.
With the advent of ETD efficiently performed in an external ion trap with
product ion detection in a high-resolution mass analyzer, e.g., Orbitrap
FTMS[28], FT-ICR MS,[29] or time-of-flight (TOF) MS,[30, 31] the performance of
TD MS has improved. Particularly, TD MS of proteins on-line separated with
7
reversed phase liquid chromatography (LC) prior to electrospray ionization (ESI)
became possible routinely.[32, 33] We have previously described the use of ETD
with a high-resolution qTOF MS reporting the overall sequence coverage of 21%
for mouse IgG1 and 15% for human IgG1 which provided structural information
on variable domains of both the heavy and the light chain.[34] As a result of a
more extensive investigation, we achieved 33% sequence coverage of a human
IgG1 using ETD with a hybrid linear ion trap-Orbitrap FTMS (considering c-, z-
and y-type product ions) equipped with a standard (moderate electric field, D30
type) Orbitrap mass analyzer (Orbitrap Velos Pro). The results obtained with
narrow (i.e., 100 m/z) and large (i.e., 600 m/z) isolation windows for precursor
ions were combined to increase the final number of identified product ions.[17]
In all the aforementioned TD MS studies, over one thousand (micro)scans (or
transients in a time-domain domain, in the case of FTMS) were summed
(averaged) to boost the sequence coverage.[17, 27, 34] Typically, several LC
MS/MS runs were acquired followed by data summation (averaging) to provide
the increased number of scans. The current state-of-the-art in TD MS, however,
does not enable the full sequencing of the disulfide bond protected domains of
light or heavy chain, neither it allows the direct localization of the glycosylation
site in the CH2 domain of the heavy chain. This is due to the compact protein
structure and its disulfide bond network (see Supporting Information, Figure
S1), which is believed to constitute the major limitation in IgG TD MS
characterization. Overall, the performance of all three approaches, using ETD on
8
a qTOF MS or on a linear trap-Orbitrap FTMS, as well as ECD on FT-ICR MS,
provide comparable results.
Here we investigated the potential of ETD implemented on a high-field
compact Orbitrap FTMS, namely the Orbitrap Elite,[35] for faster top-down
characterization of three therapeutic mAbs: adalimumab and trastuzumab (both
of the IgG1 subclass), and the IgG2 panitumumab. The high-field Orbitrap mass
analyzer is characterized by a higher frequency of ion oscillations along the
central electrode compared to the prior generation of Orbitrap mass
analyzers.[36, 37] Increased frequency of ion oscillations leads to the
corresponding enhancement of the resolving power over a fixed acquisition time.
Moreover, the implementation of a mixed absorption and magnitude mode FT of
spectral representation (known as enhanced FT, or eFT algorithm)[38] allows a
further gain (of up to two-fold) in acquisition rate at a constant resolution.[39] In
addition to the advantages of the employed mass analyzer, we optimized
precursor ion isolation and product ion transfer between the linear trap and the
Orbitrap to increase the efficiency of ETD-based TD MS. Finally, IgG TD MS was
performed under the optimized conditions with a two-stage ion activation
method, a combination of ETD and higher-energy collision dissociation
(HCD)[40] termed EThcD,[41] which has been recently applied to TD MS
experiments, demonstrating sequence complementarity with ETD.[42, 43]
9
2. Materials and methods
2.1 Chemicals. Water, acetonitrile (ACN), methanol (MeOH), formic acid (FA)
and trifluoroethanol (TFE) were obtained in LC-MS purity grade. Water, ACN and
MeOH were purchased from Fluka Analytical (Buchs, Switzerland). FA was
obtained from Merck (Zug, Switzerland), and TFE from Acros Organics (Geel,
Belgium).
2.2 Sample preparation. Therapeutic monoclonal antibodies of the IgG1 class,
namely adalimumab (Humira, Abbott Laboratories) and trastuzumab (Herceptin,
Genentech), as well as the IgG2 panitumumab (Vectibix, Amgen) are the
European Medicines Agency approved versions and formulations, available
commercially to the general public. Prior to direct infusion analysis, all
antibodies were buffer exchanged against a 150 mM ammonium acetate solution
pH 6.7 using Micro-Spin desalting columns (Zeba 75 µL, Thermo Fisher
Scientific, Zug, Switzerland).
2.3 Sample delivery and ionization. The antibodies were injected into the
mass spectrometer either by on-line coupled LC or by direct infusion in a positive
ion mode of ESI. In the former case, we employed an UltiMate 3000 LC system
(Thermo Fisher Scientific) equipped with a reversed phase C4 column (Hypersil
Gold, 1x100 mm, 5 µm particle size, Thermo Fisher Scientific, Runcorn, UK),
applying a gradient of ACN from 5 to 80% in 20 min, in presence of 0.1% FA, and
10
a flow rate of 100 µl/min generated by the loading pump. 1.5 µg of antibody (~10
pmol), diluted in 20% ACN and 0.1% FA, were typically loaded in each injection,
resulting in ~2 min elution of the IgG. The IonMax ion source (Thermo Fisher
Scientific) was operated at 3.7 kV, with sheath gas set to 20 and auxiliary gas to
10 arbitrary units. In the latter case, the antibody, buffer exchanged as described
above, was diluted in 50% ACN and 0.1% FA to a final concentration of ~10 µM;
6 µl of this IgG solution were loaded in a 20 µl loop connected to the injection
valve of the LC system and infused through a silica capillary to the nanoESI
source (Nanospray Flex ion source, Thermo Fisher Scientific) equipped with a
metallic emitter to which a 2.4 kV voltage was applied. The composition of the
solution used to push the IgG from the loop to the source was 39.9% water, 30%
ACN, 20% MeOH, 10% TFE, and 0.1% FA. The flow rate generated by the nano-
pump of the LC system was 0.6 µl/min.
2.4 Mass spectrometry. All experiments were performed on a hybrid dual linear
ion trap high-field Orbitrap FT mass spectrometer (Orbitrap Elite, Thermo
Scientific, Bremen, Germany) equipped with ETD and HCD. Similarly to our
previous report,[17] the S-lens RF level was set to 70% and the temperature of
heated transfer capillary was 350° C. Broadband mass spectra of intact IgGs and
tandem mass spectra of the same molecules were recorded in separate
experiments using ion detection in the Orbitrap FTMS (m/z 200-4000). All
Orbitrap FTMS scans were recorded averaging 10 microscans to improve the
SNR. Orbitrap FTMS was calibrated in the high mass range (m/z 2000-4000)
11
using ion at 2021.93954 m/z of commercial LTQ Velos ESI Positive Ion
Calibration Solution (Thermo Fisher Scientific). Intact mass measurements of
IgGs were performed at 15’000 resolution at 400 m/z with a target value of
charges (automatic gain control, AGC) of 1 million. For ETD experiments,
precursor ions were isolated between m/z 2400 and 2900 in the high pressure
LTQ and subsequently subjected to ETD reaction. The AGC value for
fluoranthene radical anions was set to 2 million charges, anion maximum
injection time was set to 50 ms and ETD duration (i.e., ion-ion interaction time)
was set either to 10 or 25 ms. Product ion detection in the Orbitrap mass analyzer
was performed with 120’000 resolution (at 400 m/z, eFT enabled) and an AGC
for precursor ions of 1 million charges.
For both intact mass measurements and MS/MS, the gas (N2) “delta
pressure” was of 0.1E-10 torr. The delta pressure corresponds to the difference
in the pressure measured in the so-called ultra-high vacuum region (i.e., UHV
region, or Orbitrap chamber) when nitrogen is injected into the HCD cell and
when the gas is completely shut down. A delta value of 0.1E-10 torr likely
corresponds to a pressure of ~1-2 mtorr of N2 in the HCD cell. Typically the delta
pressure for bottom-up proteomics (i.e., smaller size ions) is ~0.35E-10 torr.
Therefore, for TD MS the N2 gas pressure in the HCD cell is reduced compared to
the bottom-up proteomics. To further improve ion transmission to the high-
resolution mass analyzer from the linear trap, ions ejected from the linear trap
were first collisionally cooled in the HCD cell (a technique known as “HCD
12
trapping”), then transferred backward to the C-trap, and finally injected into the
Orbitrap as previously described.[44]
EThcD experiments were performed as follows. After ETD with 10 ms
duration, all ETD products, including both charge-reduced species and product
ions (precursors are generally completely consumed in ETD under the applied
conditions), were transferred from the linear trap to the HCD cell. Here an axial
potential was set to 25 or 40 V, so that instead of ion capture and relaxation a
second ion activation event could occur. Finally, ions were transferred back to
the C-trap and further into the Orbitrap for detection.
2.5 FTMS data processing and analysis. The data processing routine for
obtaining tandem mass spectra with an improved SNR was performed as
previously described.[17] Briefly, in both LC-MS/MS and direct infusion MS/MS
experiments, Orbitrap FTMS time-domain transients were recorded in MIDAS
.dat format[45] and underwent summation (averaging) before being processed for
time-to-frequency conversion. In this way it was possible to sum 384 ms
transients derived from separate LC runs or distinct direct infusion experiments.
Enhanced Fourier transform, eFT, or a standard magnitude mode Fourier
transform, mFT, was then applied to process transients and yield frequency-
domain spectra (internal processing at Thermo Fisher Scientific).[38] The thus
obtained frequency spectra were then converted into .RAW mass spectra (internal
processing at Thermo Fisher Scientific) for further viewing and processing with
XCalibur software (Thermo Fisher Scientific), as well as data analysis (vide infra).
13
Final ETD mass spectra were derived from summed transients recorded for 10
and 25 ms ETD duration experiments. For an in-depth evaluation of the effect of
ETD duration on IgG fragmentation, 10 and 25 ms ETD datasets were also
analyzed separately. The resolution gain of eFT versus mFT is known to be two-
fold.[38] However, to demonstrate the practically useful and presumably
sufficient resolution requirement for analysis of top-down ETD mass spectra of
IgGs, only the mFT-processed data are reported here. Results obtained with eFT
signal processing were found to be comparable to the mFT processing in terms
of protein sequence coverage (data not shown).
Previously, the combination of Xtract deconvolution software and ProSight
PC 3.0 product ion assignment software (both from Thermo Fisher Scientific) was
successfully employed for IgG TD MS data analysis.[30] Here, we extended this
approach by adding a workflow combining the MS-Deconv deconvolution
algorithm and MS-Align+ product ion assignment software (both developed at
University of California San Diego, CA, USA).[46] A comparison between the
results produced by MS-Deconv/MS-Align+ and by Xtract/ProSight PC is
presented in Figure S2, Supporting Information. Both the data analysis packages
led to comparable results. In greater detail, the MS-Deconv/MS-Align+ analyses
were realized according to the following steps: (i) ReAdW software was used to
convert .RAW files of ETD mass spectra into centroided mzXML files; (ii) a
customized version of MS Deconv was used to deconvolute the mzXML files using
a SNR cutoff of 3. The custom version of MS-Deconv was written to process the
data sets containing MS/MS information only, whereas the standard version of
14
MS-Deconv requires both MS and the directly corresponding to it MS/MS data
sets; (iii) MS-Align+ was used for product ion identification, considering c- and
z·- type ions for ETD experiments and c-, z·-, b- and y-type ions for EThcD data.
Cysteine residues were considered as not modified and all the mass tolerances
were set to 20 ppm. Manual validation was performed for cleavage sites identified
from product ions present only in a single charge state. The graphical
fragmentation maps reported herein were realized using either ProSightPC or
ProSight Lite software.[47]
2.6 Product ion abundance analysis. Abundances of identified product ions
were derived from MS-Align+ analysis. For each cleavage site, all the related
product ions were considered and grouped. Final abundances were obtained by
dividing the abundance values reported for each product ion over the ion charge
state, and summing all the resulting charge-normalized abundance values of
product ions generated from a single cleavage site. For plotting together product
ion abundance (PIA) analysis results of different sets of data, the abundances
were expressed as relative percentages of the most abundant product ion cluster
(i.e., the group of all the z-type ions, with different charge state, derived from a
single cleavage site) in each data set.
3. Results and discussion
3.1 Optimization of ETD MS/MS of IgG. To maximize ETD efficiency, we sought
to first optimize precursor cation selection and product ion transfer. In the
15
former work based on an Orbitrap Velos Pro mass spectrometer we isolated
adalimumab precursor ions around m/z 2750 (isolation window width: 100 m/z,
precursor ions from 53+ to 55+) or 2900 (isolation window width: 600 m/z,
precursor ions from 47+ to 57+). Considering the charge state envelope obtained
with the current LC-MS setting on the Orbitrap Elite for intact adalimumab, we
centered the isolation window at m/z 2700, with a width of 600 m/z, thus
including precursors from 50+ to 61+ (Figure S3, Supporting Information).
Higher charge state precursor ions are supposed to increase the fragmentation
efficiency in ETD and facilitate the separation of obtained product ions bound by
non-covalent bonds. Furthermore, cleavage preferences within protein structure
shift as a function of charge location. In regard to product ion transfer, we
applied a combination of reduction of the gas pressure in the HCD cell and the
“HCD trapping” technique, vide supra. The gas pressure reduction improves SNR
of large product ions by improving their transmission from LTQ into the HCD
cell and further within the Orbitrap analyzer. Trapping of ions in the HCD cell
(and not directly and only in the C-trap) allows the improved transmission also
of lighter ions, presumably counterbalancing the reduced damping effect in the
C-trap resulting from the lower gas pressure (Figure S4, Supporting
Information).
Overall, the implementation of these settings allowed for SNR increases in
the MS/MS mode for high m/z ions, which resulted in a reduction of the number
of transients that needed to be summed to obtain high-quality mass spectra in
comparison to our previous work, vide infra. Each ETD mass spectrum derived
16
from a combination of transients was obtained applying two different ion-ion
interaction times, 10 and 25 ms, Figure 1. These two ETD durations were chosen
to allow a direct comparison with our previous work, where we combined ETD
MS/MS data resulting from shorter and longer ion-ion interaction times, pre-
selected for fragmentation of intact denatured IgG precursor ions. Increasing
ETD duration enhanced the formation of product ions with a reduced charge
state, likely because of the occurrence of multiple electron transfer and
consequent charge reduction events. Nevertheless, the separate analysis of mass
spectra obtained with 10 or 25 ms ETD duration revealed that most of the
identified cleavage sites are shared between the two experiments (see product
ion abundance analysis results below), suggesting that ETD is primarily limited
by the retention of a compact conformation of the IgG in the gas phase. Summing
transients from these two experimental sets, though, increased the confidence
in the assignment of cleavage sites, as product ions obtained by the same
backbone cleavage are identified in multiple charge states.
3.2 ETD and EThcD MS/MS of intact human IgG1. The analysis of product
ions obtained by ETD MS/MS was performed for two IgG1 mAbs, adalimumab
and trastuzumab. Their sequences are 92% homologous, the differences being
almost entirely localized to the complementarity determining regions (CDRs)
involved in the antigen binding. Figure 2 A displays the ETD fragmentation map
of adalimumab, with the typical fragmentation pattern expected for this IgG
isotype. Extensive fragmentation is achieved on the heavy chain for the loop
17
interconnecting VH and CH1, the loop between CH2 and CH3 and part of the CH3
domain; on the light chain, the characterized region is represented by the loop
between VL and CL. For both chains, therefore, the sequenced areas are mainly
disulfide bond-free loops exposed to the solvent. The results of ETD MS/MS
application to trastuzumab are nearly identical to those for adalimumab (Figure
S5, Supporting Information). For both the IgGs only the CDR3 domains of light
and heavy chains are fully sequenced, and only a few product ions positioned on
CDR2 of heavy and light chain have been assigned for trastuzumab (which shows
slightly higher sequence coverage). As summarized in Table 1, 310 ETD MS/MS
scans (which corresponds to 3100 microscans or transients) were summed to
result in 23.8% and 26.5% sequence coverage for adalimumab and trastuzumab,
respectively. Notably, from the comparison of these results and the previously
reported ones it is apparent how, although the total sequence coverage (i.e., sum
of cleavages in light and heavy chains) is similar, the distribution of identified
cleavage sites between the two chains is changed, with the heavy chain being
more sequenced in the present report than in the previous one. For trastuzumab
we reached 29% sequence coverage which represents the highest value obtained
so far for the heavy chain of an intact IgG1 by ETD TD MS. On the other hand,
the light chain has been less sequenced with the new settings and data analysis
routine than in the past, but also for the drastically reduced number of scans
considered. Manual inspection of MS/MS spectra reveals that the missing
product ions are mainly those arising from cleavage sites located at the two
termini of the light chain: these are light, lowly charged ions that presumably
18
were not favorably transmitted with the newly employed settings (Figure S4,
Supporting Information). The fully sequenced region in the light chain,
corresponding to the loop between VL and CL, as mentioned above, remained the
same between the two studies, and it is equal between adalimumab and
trastuzumab (both of which often show complementary c- and z-type ions for the
same cleavage site). Cleavages outside this area do not follow any clear pattern
and seem randomly distributed.
To improve the sequence coverage of the IgG1 and, specifically, of the light
chain, we applied EThcD on adalimumab. We summed transients obtained with
EThcD (i.e., both product ions and charge-reduced species) performed at both
25 and 40 V. As shown in Figure 2B, EThcD produced fragmentation mainly in
the same areas previously covered by ETD, but also induced new cleavages at
the N-terminus of both light and heavy chain, and provided b- and y-type ions
from the N-terminal side of proline that could not be previously identified from
ETD data. The sequence coverage for the heavy chain of adalimumab is lower
than for ETD, but it passed from ~19% to ~28% for the light chain, as reported
in Table 1. Overall, the combination of 310 ETD scans and 260 EThcD scans,
for a total of 570 scans, produced 30.5% sequence coverage, which is the same
result obtained by averaging 1360 scans in our previous work based on the
Orbitrap Velos Pro. In addition to the reduced number of scans, a two-fold scan
speed improvement due to the implementation of the high-field Orbitrap mass
analyzer (higher frequency of ion axial oscillations) decreased the required
experimental time for mFT signal processing. Further two-fold improvement can
19
be achieved by the application of eFT signal processing. For instance, at 120’000
resolution (at 400 m/z, eFT signal processing) and under LC-MS/MS conditions,
120 transients (microscans) could be recorded per minute of IgG elution on the
Orbitrap Elite FTMS versus the previous 40 transients (microscans) on the
Orbitrap Velos Pro (mFT signal processing, 100’000 resolution at 400 m/z). The
ETD MS/MS experimental sequence, including cation accumulation, anion
injection and subsequent cation/anion interaction largely contributes to the
duty cycle, partially counterbalancing the time gain in FTMS data acquisition
due to combination of high-field Orbitrap and eFT.
Finally, from the comparison of the ETD TD MS results obtained with
different ion transfer parameters (especially the pressure conditions in the HCD
cell), those reported here and in the previous work, it seems beneficial to acquire
ETD TD MS data with both conditions and pool the data together. Indeed, results
reported in Figure S6 (Supporting Information) show that the sequence coverage
for ETD MS/MS of adalimumab with only c and z-ions is increased to 41.3% and
34.7% for light and heavy chain, respectively, with a total sequence coverage of
36.8%.
3.3 Product ion abundance analysis of ETD MS/MS of IgGs1. A PIA analysis
was conducted on ETD MS/MS data for the C-terminal portions of the almost
completely sequenced heavy chains of the investigated IgGs1. Figure 3 shows a
comparison between the relative abundances of z-ions from 10 ms ETD of
adalimumab, as well as 10 ms and 25 ms ETD of trastuzumab. Note that the
20
two IgGs share identical sequence in the analyzed area. For all three data sets,
the average abundance of product ions generated by the disulfide-free loop
connecting CH2 and CH3 is ~40% of the maximum, and global maxima are all
located there. Conversely, ions derived from the disulfide-protected area of CH3,
as well as those from the disulfide-free C-terminal end (which is, structurally
speaking, a part of the CH3 domain), generally have an abundance lower than
10% of the corresponding maximum. Furthermore, the global maxima for the
two 10 ms ETD data sets are both located at the cleavage site between K105 and
G106 (which results in z106 ions). Other local maxima correspond to the formation
of z96, z93, and z86 ions. The 25 ms data set shows a similar trend, with z106 still
abundant but surpassed by z93, z91, and z86. Therefore, the ETD results are
conditioned by the secondary and tertiary structures of the gaseous protein ions.
Moreover, similarly to ECD MS/MS,[48] the ETD radical process seems to be
affected by the electron transfer site, as the presence of basic residues at one
side of the cleavage maxima would suggest. Finally, the shift in maxima
occurring when passing from 10 to 25 ms ETD might be explained by the
increased probability of secondary electron transfer events when the ETD
duration is prolonged. The occurrence of secondary electron transfer and
consequent charge neutralization events might be confirmed by the reduced
average charge state of the product ions identified for 25 ms ETD in comparison
to 10 ms ETD experiments (Table S1, Supporting Information). The data suggest
that these events are most prominent at the flexible loop and with a CH3 domain
in a compact conformation. Therefore, despite the second fragmentation event,
21
the abundance of product ions originated within the sequence of the CH3 domain
is not higher than for the 10 ms experiments. Instead, we observe the shift
between z106 and z86 as maxima in the disulfide-free region. Visually, this is
represented in Figure 4 through a color-coded map based on crystal structure
of trastuzumab (image elaborated with The PyMol Molecular Graphics System,
version 1.3, Schrodinger, LLC). The passage from 10 ms to 25 ms ETD duration
induces the translation of highly frequent cleavage sites (represented in red)
towards the C-terminus, but still in solvent accessible positions. On the contrary,
the compact “immunoglobulin-like domain” corresponding to CH3, which
resembles the structure of a beta barrel, and contains an intra-molecular
disulfide bridge, remains poorly sequenced (and thus represented in black-blue).
These results reinforce the hypothesis that IgGs retain a relatively compact high-
order structure in the gas phase, even under denaturing ionization conditions.
3.4 ETD MS/MS of intact human IgG2. ETD MS/MS was performed on
panitumumab, an IgG2, to provide a confirmation to our hypothesis that
limitations in sequence coverage achieved for IgGs1 are mainly attributable to
gas-phase retention of a structured, compact conformation by the protein. As
displayed in Figure 5, identified product ions for this IgG are localized in the
areas sequenced in the experiments involving adalimumab and trastuzumab,
e.g., the solvent accessible loops interconnecting consecutive disulfide-protected
domains and the CH3 domain. In-depth analysis reveals that the C-terminal
portion of the heavy chain is highly sequenced, with 83 assigned z-type ions
22
versus 82 z-ions identified for trastuzumab, which has overall the highest
sequence coverage for its heavy chain (Table 1). Nevertheless, given that, in
accordance to what is observed for IgGs1, the hinge region is not sequenced also
in the case of panitumumab, differences between the two IgG isotypes are
present in the loop interconnecting VH and CH1. In the case of IgGs1, this loop
was fully sequenced, whereas for panitumumab only partial sequencing was
achieved. This can be explained by the presence in the loop of the IgG2 of Cys133,
involved in an inter-molecular disulfide bridge connecting the heavy with the
light chain. Notably, most of the identified c-type product ions are located to the
N-terminus of C133. Further studies are needed to determine whether ETD
cleavages were primarily impeded in the portion of the loop at the C-terminal
side of C133 or, conversely, cleavages occurred but primarily produced product
ions with portions of the light chain still present and were therefore not
identifiable by the present version of the data analysis approach. Finally,
panitumumab’s light chain has been sequenced in its central area (i.e., disulfide-
free loop between VL and CL), similarly to IgGs1, but its incomplete sequencing
and, importantly, the limited number of identified z-ions (nine, corresponding to
only ~50-30% of those assigned in the other experiments), could indicate that
the inter-molecular disulfide bridge with the heavy chain is less efficiently
cleaved in IgG2 than in IgG1.
23
4. Concluding remarks
Developments in Orbitrap FTMS technology mark a substantial improvement in
top-down MS performance. Experimentally, around three-fold faster acquisition
of IgG ETD data under optimized conditions in the Orbitrap Elite (high-field
compact Orbitrap analyzer) combined with a more than two-fold reduction in the
number of transients to be summed as required for a comprehensive top-down
analysis of IgGs resulted in about six-fold reduction of acquisition time in the
present study relative to the analysis with the Orbitrap Velos Pro. For the latter,
a standard trap with mFT yielded resolution of 100’000 at 400 m/z. Here,
resolution of 60’000 (at 400 m/z) was provided with mFT signal processing and
120’000 (at 400 m/z) when eFT algorithm was employed. The (moderate) level of
product ion spectral complexity generated in this top-down experiment, perhaps
due to a symmetric subunit structure of IgGs, demonstrated similar performance
in sequence coverage for mFT and eFT-generated mass spectra. The following
factors contributed to the reduction of number of scans needed for high quality
TD MS analysis: (i) selection of higher charged precursors, (ii) optimization of
transfer of large ETD product ions from the linear trap to the Orbitrap by
lowering the HCD cell gas pressure, (iii) combination of experiments conducted
with different ion-ion reaction times, and (iv) use of EThcD with two different
collisional energy values. Nevertheless, the previously determined limit in
sequence coverage, ~30%, could not be overcome. The primary cause for the
limited sequence coverage is the gas-phase retention of highly structured areas
in the IgG, mainly in correspondence of the immunoglobulin-like domains.
24
To facilitate user access to the described technology, algorithms and
methods for spectral summation across different experiments or/and for
transient acquisition and subsequent signal processing (absorption mode FT) of
summed transients where needed by spectral complexity have been made
commercially available (Spectroswiss, Lausanne, Switzerland). Further
extending the sequence coverage in TD MS of IgGs may be provided by an
improved ETD reaction cell,[49] chemical ionization source for front-end ETD on
Orbitrap FTMS,[50] and the use of ion activation methods alternative to ETD,
e.g., ultraviolet photodissociation.[51] Furthermore, improved sequence coverage
is achievable by combining ETD experimental data acquired with high HCD gas
pressure (i.e., maximizing light product ion transmission), and low gas pressure
(i.e., maximizing heavy product ion transmission), as it results from the
combination of present data with those acquired in standard operational
conditions with the Orbitrap Velos Pro. Finally, the probable presence of internal
products and product ions retaining parts of both light and heavy chains bonded
by inter-molecular disulfide bridges will require developments in the data
analysis tailoring this specific class of proteins.
5. Acknowledgements
We thank Kristina Srzentić and Anton Kozhinov for discussions and technical
support. We express our gratitude to Thermo Fisher Scientific for providing
access under license to Orbitrap Elite FTMS transient signals and related
functionality enabled by the developers’ kit. The work was supported by the
25
Swiss National Science Foundation (Projects 200021-125147 and 200021-
147006) and the European Research Council (ERC Starting grant 280271 to
YOT). Xiaowen Liu was supported by the National Institute of General Medical
Sciences, National Institutes of Health (NIH) through Grant R01GM118470.
26
6. References
[1] A. Beck, S. Sanglier-Cianferani, A. Van Dorsselaer, Biosimilar, biobetter, and
next generation antibody characterization by mass spectrometry, Anal Chem
84(11) (2012) 4637-46.
[2] D. Ayoub, W. Jabs, A. Resemann, W. Evers, C. Evans, L. Main, C. Baessmann,
E. Wagner, D. Suckau, A. Beck, Correct primary structure assessment and
extensive glyco-profiling of cetuximab by a combination of intact, middle-up,
middle-down and bottom-up ESI and MALDI mass spectrometry techniques,
MAbs 5(5) (2013) 0--1.
[3] G. Walsh, Biopharmaceutical benchmarks 2014, Nat Biotechnol 32(10) (2014)
992-1000.
[4] A. Beck, E. Wagner-Rousset, D. Ayoub, A. Van Dorsselaer, S. Sanglier-
Cianferani, Characterization of therapeutic antibodies and related products, Anal
Chem 85(2) (2013) 715-36.
[5] G. Vidarsson, G. Dekkers, T. Rispens, IgG subclasses and allotypes: from
structure to effector functions, Front Immunol 5 (2014) 520.
[6] S.A. Berkowitz, J.R. Engen, J.R. Mazzeo, G.B. Jones, Analytical tools for
characterizing biopharmaceuticals and the implications for biosimilars, Nat Rev
Drug Discov 11(7) (2012) 527-40.
[7] B. Wang, A.C. Gucinski, D.A. Keire, L.F. Buhse, M.T. Boyne, Structural
comparison of two anti-CD20 monoclonal antibody drug products using middle-
down mass spectrometry, Analyst 138(10) (2013) 3058-3065.
27
[8] L. Fornelli, D. Ayoub, K. Aizikov, A. Beck, Y.O. Tsybin, Middle-down analysis
of monoclonal antibodies with electron transfer dissociation orbitrap fourier
transform mass spectrometry, Anal Chem 86(6) (2014) 3005-12.
[9] V.C. Cotham, J.S. Brodbelt, Characterization of Therapeutic Monoclonal
Antibodies at the Subunit-Level using Middle-Down 193 nm Ultraviolet
Photodissociation, Analytical chemistry 88(7) (2016) 4004-4013.
[10] K. Srzentic, L. Fornelli, U.A. Laskay, M. Monod, A. Beck, D. Ayoub, Y.O.
Tsybin, Advantages of Extended Bottom-Up Proteomics Using Sap9 for Analysis
of Monoclonal Antibodies, Analytical chemistry 86(19) (2014) 9945-9953.
[11] Z. Zhang, H. Pan, X. Chen, Mass spectrometry for structural characterization
of therapeutic antibodies, Mass Spectrom Rev 28(1) (2009) 147-76.
[12] O.V. Krokhin, M. Antonovici, W. Ens, J.A. Wilkins, K.G. Standing,
Deamidation of -Asn-Gly- sequences during sample preparation for proteomics:
Consequences for MALDI and HPLC-MALDI analysis, Anal Chem 78(18) (2006)
6645-50.
[13] L.W. Dick, C. Kim, D.F. Qiu, K.C. Cheng, Determination of the origin of the
N-terminal pyro-glutamate variation in monoclonal antibodies using model
peptides, Biotechnology and Bioengineering 97(3) (2007) 544-553.
[14] K. Diepold, K. Bomans, M. Wiedmann, B. Zimmermann, A. Petzold, T.
Schlothauer, R. Mueller, B. Moritz, J.O. Stracke, M. Molhoj, D. Reusch, P. Bulau,
Simultaneous assessment of Asp isomerization and Asn deamidation in
recombinant antibodies by LC-MS following incubation at elevated temperatures,
PLoS One 7(1) (2012) e30295.
28
[15] T.K. Toby, L. Fornelli, N.L. Kelleher, Progress in Top-Down Proteomics and
the Analysis of Proteoforms, Annu Rev Anal Chem (Palo Alto Calif) 9(1) (2016)
499-519.
[16] B.Q. Tran, C. Barton, J. Feng, A. Sandjong, S.H. Yoon, S. Awasthi, T. Liang,
M.M. Khan, D.P. Kilgour, D.R. Goodlett, Y.A. Goo, Comprehensive glycosylation
profiling of IgG and IgG-fusion proteins by top-down MS with multiple
fragmentation techniques, J Proteomics 134 (2016) 93-101.
[17] L. Fornelli, E. Damoc, P.M. Thomas, N.L. Kelleher, K. Aizikov, E. Denisov, A.
Makarov, Y.O. Tsybin, Analysis of intact monoclonal antibody IgG1 by electron
transfer dissociation Orbitrap FTMS, Mol Cell Proteomics 11(12) (2012) 1758-67.
[18] R.A. Zubarev, N.L. Kelleher, F.W. McLafferty, Electron capture dissociation
of multiply charged protein cations. A nonergodic process, Journal of the
American Chemical Society 120(13) (1998) 3265-3266.
[19] R.A. Zubarev, D.M. Horn, E.K. Fridriksson, N.L. Kelleher, N.A. Kruger, M.A.
Lewis, B.K. Carpenter, F.W. McLafferty, Electron capture dissociation for
structural characterization of multiply charged protein cations, Analytical
chemistry 72(3) (2000) 563-573.
[20] J.E.P. Syka, J.J. Coon, M.J. Schroeder, J. Shabanowitz, D.F. Hunt, Peptide
and protein sequence analysis by electron transfer dissociation mass
spectrometry, Proceedings of the National Academy of Sciences of the United
States of America 101(26) (2004) 9528-9533.
[21] J.M. Wells, S.A. McLuckey, Collision-induced dissociation (CID) of peptides
and proteins, Methods Enzymol 402 (2005) 148-85.
29
[22] S. Meier, Y. Tsybin, P. Dyson, B. Keppler, C. Hartinger, Fragmentation
methods on the balance: unambiguous top–down mass spectrometric
characterization of oxaliplatin–ubiquitin binding sites, Analytical and
Bioanalytical Chemistry 402(8) (2012) 2655-2662.
[23] H. Molina, R. Matthiesen, K. Kandasamy, A. Pandey, Comprehensive
Comparison of Collision Induced Dissociation and Electron Transfer
Dissociation, Analytical chemistry 80(13) (2008) 4825-4835.
[24] Y. Ge, I.N. Rybakova, Q. Xu, R.L. Moss, Top-down high-resolution mass
spectrometry of cardiac myosin binding protein C revealed that truncation alters
protein phosphorylation state, Proc Natl Acad Sci U S A 106(31) (2009) 12658-
63.
[25] S.G. Valeja, N.K. Kaiser, F. Xian, C.L. Hendrickson, J.C. Rouse, A.G.
Marshall, Unit mass baseline resolution for an intact 148 kDa therapeutic
monoclonal antibody by Fourier transform ion cyclotron resonance mass
spectrometry, Anal Chem 83(22) (2011) 8391-5.
[26] J.D. Tipton, J.C. Tran, A.D. Catherman, D.R. Ahlf, K.R. Durbin, J.E. Lee,
J.F. Kellie, N.L. Kelleher, C.L. Hendrickson, A.G. Marshall, Nano-LC FTICR
tandem mass spectrometry for top-down proteomics: routine baseline unit mass
resolution of whole cell lysate proteins up to 72 kDa, Anal Chem 84(5) (2012)
2111-7.
[27] Y. Mao, S.G. Valeja, J.C. Rouse, C.L. Hendrickson, A.G. Marshall, Top-down
structural analysis of an intact monoclonal antibody by electron capture
30
dissociation-Fourier transform ion cyclotron resonance-mass spectrometry, Anal
Chem 85(9) (2013) 4239-46.
[28] G.C. McAlister, W.T. Berggren, J. Griep-Raming, S. Horning, A. Makarov, D.
Phanstiel, G. Stafford, D.L. Swaney, J.E. Syka, V. Zabrouskov, J.J. Coon, A
proteomics grade electron transfer dissociation-enabled hybrid linear ion trap-
orbitrap mass spectrometer, J Proteome Res 7(8) (2008) 3127-36.
[29] S. Bourgoin-Voillard, N. Leymarie, C.E. Costello, Top-down tandem mass
spectrometry on RNase A and B using a Qh/FT-ICR hybrid mass spectrometer,
Proteomics 14(10) (2014) 1174-1184.
[30] R.G. Hartmer, D.A. Kaplan, C. Stoermer, M. Lubeck, M.A. Park, Data-
dependent electron transfer dissociation of large peptides and medium size
proteins in a QTOF instrument on a liquid chromatography timescale, Rapid
Commun Mass Spectrom 23(15) (2009) 2273-82.
[31] L. Fornelli, J. Parra, R. Hartmer, C. Stoermer, M. Lubeck, Y.O. Tsybin, Top-
down analysis of 30-80 kDa proteins by electron transfer dissociation time-of-
flight mass spectrometry, Anal Bioanal Chem (2013).
[32] Z. Tian, N. Tolic, R. Zhao, R.J. Moore, S.M. Hengel, E.W. Robinson, D.L.
Stenoien, S. Wu, R.D. Smith, L. Pasa-Tolic, Enhanced top-down characterization
of histone post-translational modifications, Genome Biol 13(10) (2012) R86.
[33] D.R. Ahlf, P.D. Compton, J.C. Tran, B.P. Early, P.M. Thomas, N.L. Kelleher,
Evaluation of the compact high-field orbitrap for top-down proteomics of human
cells, J Proteome Res 11(8) (2012) 4308-14.
31
[34] Y.O. Tsybin, L. Fornelli, C. Stoermer, M. Luebeck, J. Parra, S. Nallet, F.M.
Wurm, R. Hartmer, Structural Analysis of Intact Monoclonal Antibodies by
Electron Transfer Dissociation Mass Spectrometry, Analytical Chemistry 83(23)
(2011) 8919-8927.
[35] A. Michalski, E. Damoc, O. Lange, E. Denisov, D. Nolting, M. Muller, R.
Viner, J. Schwartz, P. Remes, M. Belford, J.J. Dunyach, J. Cox, S. Horning, M.
Mann, A. Makarov, Ultra high resolution linear ion trap Orbitrap mass
spectrometer (Orbitrap Elite) facilitates top down LC MS/MS and versatile
peptide fragmentation modes, Mol Cell Proteomics 11(3) (2012) O111 013698.
[36] A. Makarov, E. Denisov, O. Lange, Performance evaluation of a high-field
Orbitrap mass analyzer, J Am Soc Mass Spectrom 20(8) (2009) 1391-6.
[37] Y.O. Tsybin, Structural Analysis of Complex Molecular Systems by High-
Resolution and Tandem Mass Spectrometry, in: B. Pignataro (Ed.), Discovering
the Future of Molecular Sciences, Wiley-VCH Verlag GmbH & Co. KGaA2014.
[38] O. Lange, E. Damoc, A. Wieghaus, A. Makarov, Enhanced Fourier transform
for Orbitrap mass spectrometry, International Journal of Mass Spectrometry 369
(2014) 16-22.
[39] E. Denisov, E. Damoc, O. Lange, A. Makarov, Orbitrap mass spectrometry
with resolving powers above 1,000,000, International Journal of Mass
Spectrometry 325 (2012) 80-85.
[40] J.V. Olsen, B. Macek, O. Lange, A. Makarov, S. Horning, M. Mann, Higher-
energy C-trap dissociation for peptide modification analysis, Nat Methods 4(9)
(2007) 709-12.
32
[41] C.K. Frese, A.F. Altelaar, H. van den Toorn, D. Nolting, J. Griep-Raming, A.J.
Heck, S. Mohammed, Toward full peptide sequence coverage by dual
fragmentation combining electron-transfer and higher-energy collision
dissociation tandem mass spectrometry, Anal Chem 84(22) (2012) 9668-73.
[42] A.M. Brunner, P. Lossl, F. Liu, R. Huguet, C. Mullen, M. Yamashita, V.
Zabrouskov, A. Makarov, A.F. Altelaar, A.J. Heck, Benchmarking multiple
fragmentation methods on an orbitrap fusion for top-down phospho-proteoform
characterization, Anal Chem 87(8) (2015) 4152-8.
[43] Y. Zheng, L. Fornelli, P.D. Compton, S. Sharma, J. Canterbury, C. Mullen,
V. Zabrouskov, R.T. Fellers, P.M. Thomas, J.D. Licht, M.W. Senko, N.L. Kelleher,
Unabridged Analysis of Human Histone H3 by Differential Top-Down Mass
Spectrometry Reveals Hypermethylated Proteoforms from MMSET/NSD2
Overexpression, Mol Cell Proteomics 15(3) (2016) 776-90.
[44] S. Rosati, R.J. Rose, N.J. Thompson, E. van Duijn, E. Damoc, E. Denisov, A.
Makarov, A.J. Heck, Exploring an orbitrap analyzer for the characterization of
intact antibodies by native mass spectrometry, Angew Chem Int Ed Engl 51(52)
(2012) 12992-6.
[45] M.W. Senko, J.D. Canterbury, S. Guan, A.G. Marshall, A high-performance
modular data system for Fourier transform ion cyclotron resonance mass
spectrometry, Rapid Commun Mass Spectrom 10(14) (1996) 1839-44.
[46] X. Liu, Y. Inbar, P.C. Dorrestein, C. Wynne, N. Edwards, P. Souda, J.P.
Whitelegge, V. Bafna, P.A. Pevzner, Deconvolution and database search of
33
complex tandem mass spectra of intact proteins: a combinatorial approach, Mol
Cell Proteomics 9(12) (2010) 2772-82.
[47] R.T. Fellers, J.B. Greer, B.P. Early, X. Yu, R.D. LeDuc, N.L. Kelleher, P.M.
Thomas, ProSight Lite: graphical software to analyze top-down mass
spectrometry data, Proteomics 15(7) (2015) 1235-8.
[48] K. Breuker, H. Oh, C. Lin, B.K. Carpenter, F.W. McLafferty, Nonergodic and
conformational control of the electron capture dissociation of protein cations,
Proc Natl Acad Sci U S A 101(39) (2004) 14011-6.
[49] C.M. Rose, J.D. Russell, A.R. Ledvina, G.C. McAlister, M.S. Westphall, J.
Griep-Raming, J.C. Schwartz, J.J. Coon, J.E. Syka, Multipurpose dissociation
cell for enhanced ETD of intact protein species, J Am Soc Mass Spectrom 24(6)
(2013) 816-27.
[50] L. Earley, L.C. Anderson, D.L. Bai, C. Mullen, J.E. Syka, A.M. English, J.J.
Dunyach, G.C. Stafford, Jr., J. Shabanowitz, D.F. Hunt, P.D. Compton, Front-
end electron transfer dissociation: a new ionization source, Anal Chem 85(17)
(2013) 8385-90.
[51] J.B. Shaw, W.Z. Li, D.D. Holden, Y. Zhang, J. Griep-Raming, R.T. Fellers,
B.P. Early, P.M. Thomas, N.L. Kelleher, J.S. Brodbelt, Complete Protein
Characterization Using Top-Down Mass Spectrometry and Ultraviolet
Photodissociation, Journal of the American Chemical Society 135(34) (2013)
12646-12651.
34
Figure captions
Figure 1. Top-down ETD Orbitrap FTMS of IgG1 (adalimumab) with ion-ion
interaction time of 10 and 25 ms. (A) Combined mass spectrum including
transients from both 10 and 25 ms measurements, with product ion population
arising from the former experiment highlighted in blue, and from the latter
indicated in red. (B) and (C) Expanded views of the two product ion populations,
with assigned product ions derived from the heavy chain. Note the difference in
the average charge state of the product ions of the two populations.
Figure 2. Fragmentation map of adalimumab by top-down Orbitrap FTMS. (A)
ETD MS/MS. (B) EThcD MS/MS. The heavy chain information is on top, the light
chain information is at the bottom.
Figure 3. Product ion abundance analysis of ETD top-down MS of adalimumab
with 10 ms ETD duration and trastuzumab with 10 and 25 ms duration. In the
case of trastuzumab, the color-coded histogram demonstrates the improvement
in sequence coverage obtained through the combination of 10 ms (magenta) and
25 ms (cyan) ETD MS/MS data. Whereas medium and large product ions (e.g.,
z117 and z63) are generally produced using 10 ms duration, smaller ions such as
z16, z17 and z18 are detected only in the case of a longer ETD duration, presumably
as the result of secondary fragmentation (i.e., re-fragmentation of larger product
ions).
35
Figure 4. Color-coded representation of ETD PIA results for the C-terminus of
the heavy chain of trastuzumab. (A) 10 ms ETD. (B) 25 ms ETD. The displayed
image is based on X-ray crystallography data (PDB entry: 3D6G).
Figure 5. Fragmentation map of panitumumab (IgG2) obtained with top-down
ETD Orbitrap FTMS.
Table 1. Sequence coverage obtained for different IgGs1 and IgG2 by top-down
ETD Orbitrap FTMS. Indicated scans consist of 10 microscans each. Note, only
c- and z-type ions were considered for ETD MS (for both “high-field” and
“standard” Orbitrap data), whereas also b- and y-type ions were included in
EThcD MS analysis.
Table 1.
Number of
scans
N‐terminal
product ions
C‐terminal
product ions
Unique
cleavage sites
Sequence
coverage
N‐terminal
product ions
C‐terminal
product ions
Unique
cleavage sites
Sequence
coverage
high‐field Orbitrap
Adalimumab ETD 310 34 19 41 19.2% 39 78 117 26.0% 23.8%
Adalimumab EThcD 260 45 29 59 27.7% 22 58 80 17.8% 21.0%
Adalimumab combined 570 50 32 67 31.5% 53 82 135 30.0% 30.5%
Trastuzumab ETD 310 37 19 45 21.1% 55 75 130 29.0% 26.5%
Panitumumab ETD 300 34 9 34 15.9% 23 83 106 23.9% 21.3%
standard Orbitrap
Adalimumab combined 1360 54 59 84 39.4% 60 79 119 26.4% 30.6%
Light chain Heavy chainTotal sequence
coverage
36
Figure 1.
37
Figure 2.
38
Figure 3.
39
Figure 4.
40
Figure 5.