troubleshooting of gc systems
TRANSCRIPT
-
8/20/2019 Troubleshooting of GC Systems
1/16
Troubleshooting of GC Systems
Agilent RestrictedPage 1
Troubleshooting ofGC Systems
Techniques, Tips, and Tricks
Daron Decker
Chromatography Technical Specialist
Troubleshooting of GC Systems
Page 2
Troubleshooting Starts with Isolating the problem
Is it injection, flow, column, detection, electronic or some combination?
Know what can and can’t cause the symptom and what to do about it
Broad Solvent Peak – Injector Problems, Technique, Sample Problems
Split Peaks – Injector Problems, Mixed Solvent
No Peaks – Wasn’t Introduced, Wasn’t Detected
Bonus Peaks – In Sample or Back Flash
Response Changes – Activity, Inlet Discrimination, Detector Problem
Peak Tailing – Flow Path or Activity
Peak Fronting – Overload or Solubility Mismatch
Shifting Retention – Leaks, Column Aging, Contamination or Damage
Loss of Resolution – Peak Broadening, Separation DecreasingBaseline Disturbances – Column Bleed, Contamination, Temperature
Noisy or Spiking Baseline – Electrical or Contaminated Detector
Negative or Flat-topped Peaks – Detector Problems
Quantitation Problems – Activity, Inlet or Detector Problems
Logical Troubleshooting
-
8/20/2019 Troubleshooting of GC Systems
2/16
Troubleshooting of GC Systems
Page 3
Broad Solvent Front (or Major Component)
Troubleshooting of GC Systems
Page 4
Broad Solvent FrontTechnique
Installation problem
Large injection volume
Very slow injection rate
Leak in the injector
-
8/20/2019 Troubleshooting of GC Systems
3/16
Troubleshooting of GC Systems
Page 5
Broad Solvent FrontInjector Parameters
Split ratio too low
Purge off for split injection
No or long purge activation time (splitless)
Injector temperature too low/high
Troubleshooting of GC Systems
Page 6
Broad Solvent FrontSample Related
Dirty sample
Mixed sample solvent
Impure solvent
Column temperature < solvent BP
-
8/20/2019 Troubleshooting of GC Systems
4/16
Troubleshooting of GC Systems
Page 7
Split Peaks
Injector problem (poor sample intro)
Mixed sample solvent (polarity difference)
Sample in syringe needle
Sample degradation in injector
Injection technique - erratic
Troubleshooting of GC Systems
Page 8
Injection technique - erratic
Which peaks? Early or late? Trends?
-
8/20/2019 Troubleshooting of GC Systems
5/16
Troubleshooting of GC Systems
Page 9
No Peaks
Detector/recorder not operational
Plugged syringe/plunger not moving
Wrong injector or detector
Broken column
Huge leak
No carrier gas flow
No column
Troubleshooting of GC Systems
Page 10
A Real Troubleshooting Example
No Peaks
10 20 30 40 50
14
16
18
20
22
24
26
28
30
-
8/20/2019 Troubleshooting of GC Systems
6/16
Troubleshooting of GC Systems
Page 11
Is the flame Lit?
put glass piece over FID outlet---- Answer in this case, Water condenses
look at output in instrument guage-- is the digital value greater than 0.0?
Answer in this case is approximately 16.2 pico amps
Is there flow through the column?
disconnect column from detector and measure flow with bubble solution or meter
Answer in this case was YES THERE IS FLOW
Assess the observations
Flame is lit and we have flow from end of column
Hypothesis: Sample not getting on column-syringe plugged?
Take syringe out and make injection manually on a dry paper towel
Answer – towel stayes dry (Syringe was clogged with septum)
Pull plunger out top, add solvent and replace plunger will usually dislodge septum particle
(should hear a little pop) If you can’t dislodge plug, Replace syringeReassemble the Injector & Re-inject
Logical Steps Taken to Find Peaks(most of our problems are leaks and plugs)
Troubleshooting of GC Systems
Page 12
Peaks !!
0 10 20 30 40 50
14
16
18
20
22
24
26
28
30
-
8/20/2019 Troubleshooting of GC Systems
7/16
Troubleshooting of GC Systems
Page 13
Bonus Peaks or Ghost Peaks
Contamination in injector, column or carrier gas
Carry-over from a backflash or previous sample
Septum bleed
Troubleshooting of GC Systems
Page 14
Peak Response All Change in Size
DETECTOR response problem
Attenuation value
High background signal
Detector zeroed at a high signal level
Split ratio set incorrectly
Wrong purge activation time
Septum purge flow too high
Injector temperature too low
INJECTOR
Leaky syringe
-
8/20/2019 Troubleshooting of GC Systems
8/16
Troubleshooting of GC Systems
Page 15
Peak ResponseSome Change in Size
Adsorption of active compounds (-OH, -NH, -SH) by liner,column or contaminants
Decomposition or evaporation from sample
Inlet Temperature Change – Discrimination
Troubleshooting of GC Systems
Page 16
Peak Tailing
Reversible adsorption of active compounds
(-OH, -NH, -SH) by liner, column or contaminants
Flow path problem - dead volume, obstruction, poorinstallation, or severe column contamination
Miscellaneous - overloading of PLOT columns, co-elution,polarity mismatch between phase, solute or solvent, and somecompounds always tail
-
8/20/2019 Troubleshooting of GC Systems
9/16
Troubleshooting of GC Systems
Page 17
Peak FrontingShark Fin Shaped
Overload
– More pronounced with large solute andphase polarity differences
Troubleshooting of GC Systems
Page 18
Peak FrontingOther
Compound very soluble ininjection solvent
Column installation
Injection technique
Co-elution
Mixed sample solvent
-
8/20/2019 Troubleshooting of GC Systems
10/16
Troubleshooting of GC Systems
Page 19
Retention Time Shift
2.00
3.25
4.75
2.75
4.00
5.50
Leak in the septum
Change in gas velocity
Change in injection solvent
Large change in sampleconcentration
Change in temperature
Loss of stationary phase
Contamination
Damaged stationary phase
Troubleshooting of GC Systems
Page 20
min8 9 10 11
1400 ng
7 ng
Effect of Sample Overload on
Retention Time and Peak Tailing
-
8/20/2019 Troubleshooting of GC Systems
11/16
Troubleshooting of GC Systems
Page 21
Loss of Resolution
Resolution is a function of separation and peak width
Separation
Peak Width
Troubleshooting of GC Systems
Page 22
Loss of ResolutionSeparation Decrease
Different column temperature
Column contamination
Co-elution
Different column?
-
8/20/2019 Troubleshooting of GC Systems
12/16
Troubleshooting of GC Systems
Page 23
Loss of ResolutionPeak Broadening
Change in carrier gas velocity
Column contamination
Column degradation
Injector (split flow, liner, installation)
Detector (flow, installation)
Troubleshooting of GC Systems
Page 24
Baseline Disturbances
Sudden Changes, Wandering, or Drifting
Column or detector not fully conditioned
Changes in carrier and detector gas flows
– Valves switching, leaks
Detector thermal or current instability
GC/MS scan range change
Contamination
WANDER
DRIFT
-
8/20/2019 Troubleshooting of GC Systems
13/16
Troubleshooting of GC Systems
Page 25
Noisy Baseline
Contamination, gas, detector, etc.
Column in detector flame
Incorrect detector gas flows
Air leak - ECD, TCD
Detector electronics malfunction
MILD
SEVERE
Troubleshooting of GC Systems
Page 26
Spiking Baseline
Particles entering the detector
Random: poor connection
Regular: nearby "cycling" equipment
-
8/20/2019 Troubleshooting of GC Systems
14/16
Troubleshooting of GC Systems
Page 27
Negative Peaks
Normal for TCD, ECD
All peaks - check recorderconnection and settings
Dirty or old cell if after largepeaks with ECD
Troubleshooting of GC Systems
Page 28
Flat-Topped Peaks
Maximum detector signal exceeded
Maximum recorder signal exceeded
Excessive sensitivity settings
-
8/20/2019 Troubleshooting of GC Systems
15/16
Troubleshooting of GC Systems
Page 29
Quantitation ProblemsSystem
Detector problem (stability)
Outside detector's linear range
Recorder settings
Activity (adsorption)
Troubleshooting of GC Systems
Page 30
Quantitation ProblemsTechnique or Conditions
Injection technique or conditions
Syringe worn
Baseline disturbances
Co-elution
Matrix effects
Sample evaporation – leaky vials
Sample decomposition
-
8/20/2019 Troubleshooting of GC Systems
16/16
Troubleshooting of GC Systems
Page 31
Troubleshooting Tips
1. Isolate the problem.
(Blank Run, Inject Un-retained Compound, Jumper Tube Test)
2. Change only one variable at a time.
3. Compare before/after chromatograms.
(Peak shape, response, retention, baseline rise, background, etc.)
4. Utilize Technical Support.
Troubleshooting of GC Systems
Page 32
TECHNICAL SUPPORT
1-800-227-9770, #4, #1
1-972-699-6423 (Daron)
1-877-874-0307 (toll free)
281-997-9107 (FAX)
E-mail:[email protected]