trz03te he hw1 - rmg messtechnik gmbh...all rmg turbine meters have passed the disturbance...

65
by Honeywell Serving the Gas Industry Worldwide Electronic Turbine Meter TRZ 03-TE( L) OPERATING INSTRUCTIONS STATUS MARCH 2013

Upload: others

Post on 09-Jun-2020

8 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

by Honeywell

Serving the Gas Industry Worldwide

Electronic Turbine Meter TRZ 03-TE(L)

OPERATING INSTRUCTIONS

STATUS MARCH 2013

Page 2: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

Note: Unfortunately, paperwork does not automatically update itself but technical developments are constantly being made. Therefore, we reserve the right to change the descriptions and statements contained in our operating instructions without prior notice. However, you can conveniently download the most recent version of this manual (and those of other devices) from our website www.rmg.com. RMG Messtechnik GmbH Otto-Hahn-Str. 5 Phone numbers: 35510 Butzbach (Germany) Switchboard: +49 (0)6033 897-0 Fax: +49 (0)6033 897-130 Customer Service: +49 (0)6033 897-127 E-mail: [email protected] Spare Parts: +49 (0)6033 897-173

Page 3: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

CONTENTS ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

INTRODUCTION ................................................................................... 1 

Scope of application ............................................................................................. 1 

Method of operation ............................................................................................. 2 

Meter head ............................................................................................................ 3 

Approvals .............................................................................................................. 3 

Standards / Guidelines ......................................................................................... 3 

Validity of meter proving ...................................................................................... 3 

Measuring ranges ................................................................................................. 4 

Extension of the measuring range........................................................................ 4 

Accuracy of measurement .................................................................................... 5 

Permissible temperature ranges .......................................................................... 5 

Pressure loss ......................................................................................................... 6 

Pressure tap .......................................................................................................... 7 

Using gas meters with different types of gases .................................................. 7 

SAFETY INSTRUCTIONS ...................................................................... 8 

Instructions for the installer ................................................................................. 9 

INSTALLATION AND COMMISSIONING ............................................ 11 

Installation .......................................................................................................... 11 

Perforated-plate straighteners ........................................................................... 15 

Seals .................................................................................................................... 16 

Screws ................................................................................................................. 17 

Block diagram ..................................................................................................... 18 

Power Supply ...................................................................................................... 19 

Electrical connections ........................................................................................ 19 

Installing the remote totalizing unit ................................................................... 24 

Temperature measurement ................................................................................ 25 

COMMISSIONING .............................................................................. 26 

Filling with oil ...................................................................................................... 26 

Connecting the gas flow ..................................................................................... 26 

Initializing the totalizing unit .............................................................................. 26 

Page 4: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

CONTENTS ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

OPERATION ....................................................................................... 27 

Display ................................................................................................................. 27 

Programming ....................................................................................................... 29 

Parameters and Modes of the TRZ 03-TE / TEL ................................................. 32 

Influences on the error of measurement due to operation ............................... 38 

Lubrication .......................................................................................................... 40 

Changing the battery .......................................................................................... 43 

Maintenance instructions ................................................................................... 44 

Labelling .............................................................................................................. 44 

SPECIFICATIONS ............................................................................... 46 

Measuring range ................................................................................................. 46 

Measuring error .................................................................................................. 46 

Temperature ranges ............................................................................................ 46 

Inputs and outputs .............................................................................................. 46 

Materials used for the meter case ..................................................................... 50 

Examples of connection ...................................................................................... 51 

Overview of Parameters for the TRZ 03 TE / TEL .............................................. 54 

Measuring ranges/Dimensions/Pressure classes of TRZ 03-TE ...................... 55 

Measuring ranges/Dimensions/Pressure classes of TRZ 03-TEL .................... 56 

Overview of the materials used .......................................................................... 57 

ANNEX ............................................................................................... 58 

Matrix of the TRZ 03-TE / TEL ............................................................................ 58 

Converting the TRZ 03-TE/TEL into the version with external supply and current output ..................................................................................................... 60 

Cable type abbreviations .................................................................................... 61 

Page 5: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INTRODUCTION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

1

Introduction

Scope of application The TRZ 03-TE and TRZ 03-TEL turbine meters are flow meters which can be used for custody transfer metering. Unlike these instruments, the TERZ 94 turbine meter can only be used for secondary metering. All three types of turbine meters measure the quantity of gas flowing through them in units of volume at prevailing pressure and temperature. Therefore, the units of volume are determined at flowing conditions. The volume of the gas flowing through and the flow rate are indicated by a mechanical totalizer in cubic meters at flowing conditions or m3/h respectively. In addition, the turbine meters can be fitted with HF or LF pulse transmitters and with reed contacts. In this way, pulses are obtained whose number is proportional to the volume which has flowed through. These pulses can be further processed by volume correctors. The major differences between types TRZ 03-TE, TRZ 03-TEL or TERZ 94 are as follows: TRZ 03-TE

Approved for custody transfer metering in compliance with DIN 33800. Installation length: 3 x DN Accuracy 0.5% (above 0.2 Qmax) 2-channel measurement

TRZ 03-TEL

Approved for custody transfer metering in compliance with TR G13 / OIML. Requires no additional inlet pipe even in the case of heavy flow disturbances. Accuracy 0.5% (above 0.2 Qmax) 2-channel measurement

TERZ 94

For secondary metering. Installation length: 1.5 x DN Accuracy 1% (above 0.2 Qmax)

The following applies to all meter types:

Maximum operating pressure: 100 bar (not for gas meters with a plastic turbine wheel) Meter sizes from G 40 to G 16000 Measuring range 1:20; 1:30 or 1:50 (see data sheets) Connections in compliance with DIN or ANSI are available. Special designs for aggressive gases are available. Available for low temperatures (< 10°C) All gas meters can be operated in any position up to the nominal diameter of DN 200. Flow monitoring Alarm output Meter head removable

Page 6: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INTRODUCTION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

2

Method of operation The method of operation of the electronic turbine meter is based on the measurement of the gas velocity. The velocity of the gas flowing through the gas meter is increased in the flow straightener and the gas strikes the turbine wheel in a defined flow cross section. In the flow straightener, unwanted vortices, turbulences and asymmetries are removed or reduced. The turbine wheel is mounted axially, while the blades of the turbine wheel are arranged at a certain angle to the gas flow. Within the measuring range (Qmin - Qmax), the rotational speed of the turbine wheel is almost proportional to the mean gas velocity and, therefore, to the rate of flow. The number of rotations is a measure of the volume that has flowed through. Two pulse-wire sensors pick up the rotational frequency of the turbine wheel by a non-contact measurement. For this purpose a permanent magnet is fixed at the end disc of the turbine shaft which induces one voltage pulse per rotation. These pulses are further processed by the electronic system of the meter head.

Turbine wheel

Lubrication pump

Thermowell(optional)

Index head

Sensor

SetscrewFlow straightener

Pressure tap pr

Permanent magnet

Sensor sleeve

O ringIP65

Current output 4-20mA yes no

24 VDC - 0085

EEx ib[ia] IIC T3/T4

EG-type examination certificate!

II2G TÜV 02 ATEX 1970 -20°C T +60°C/40°C amb

Manufacturing Ser. No.:

Fig. 1: Sectional drawing of a turbine meter For output – e.g. to a volume corrector or for control purposes – a HF pulse output (unchanged sensor frequency) or a NF pulse output (programmable scaling factor) are available.

Page 7: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INTRODUCTION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

3

Meter head Inside the meter head, the number of pulses is divided by the meter factor (number of pulses per m3) and the result is used to calculate the volume at measurement conditions. In the main totalizer, the sum of the volume at measurement conditions with flowed through the meter per time unit on the flow rate display.

At the HF output, the unchanged signal frequency of the sensor element is outputted, whereas, for the LF output, this HF frequency can be reduced by one programmable scaling factor.

Approvals Types TRZ 03-TE and TRZ 03-TEL has been approved for custody transfer metering.

The following approvals have been obtained: German approval No. 7.211/02.13 DVGW registration No. CE-0085BN0291

Standards / Guidelines All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy flow disturbances. Therefore, this meter design meets the requirements for installation in compliance with Technical Guideline G 131), Sec. 1. Test specifications are as laid down in PTB Testing Instructions, Vol. 4, Volume gas meters, 2nd revised edition of 1992.

The RMG turbine meters of type TRZ 03-TE and TRZ 03-TEL comply with DIN 33800.

Validity of meter proving The turbine meters of type TRZ 03 which are suitable for custody transfer metering must be subjected to subsequent meter proving at regular intervals. The validity of meter proving is determined in the German Metering and Calibration Rules issued in 1988 (with amendment of September 24,1992) and extends to five years for electronic turbine meters without a lubricator.

Note: these periods are valid for Germany, in other countries the periods are normally different!

In the case of subsequent meter proving, the gas meter must be removed and tested on a test rig.

1) can be ordered at PTB, Bundesallee 100, D-38116 Braunschweig (www.ptb.de)

Page 8: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INTRODUCTION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

4

Measuring ranges The measuring ranges are between 10 and 25,000 m3/h (flowing conditions). A measuring range is specified for each meter size. It is limited by the minimum flow rate Qmin and the maximum flow rate Qmax (see tables on pages 55 to 56).

For types TRZ 03-TE and TRZ 03-TEL, this is the flow range where the gas meter must indicate correct values within the error limits specified by the German Metering and Calibration Rules.

Turbine meters of type TRZ 03-TE have measuring ranges up to 1:30 even under atmospheric pressure. If a high-pressure test is conducted in compliance with Technical Guideline G 71) (PTB), the measuring range can be extended to 1:50. Then the minimum flow rate Qmin HP is the lowest test point during high-pressure testing. Types TRZ 03-TE and TRZ 03-TEL may then be used for billing purposes within the specified HP flow and density ranges.

Extension of the measuring range In the range of 0.2 Qmax to Qmax, the measuring behaviour of turbine meters is determined by the aerodynamic conditions in the flow channel and the measuring cross section. By means of many series of tests conducted both under atmospheric pressure and under higher pressures and with an appropriate rating for such ranges, it is possible to achieve a deviation of the calibration curve under atmospheric conditions and under high-pressure conditions of < 0.5% in the flow range of 0.2 Qmax to Qmax. In the lower flow range, the measuring behaviour is determined by the relationship between the gas flow driving the measuring wheel and the slowing-down torques due to drags (bearings and totalizer). The driving torques increase linearly with the density and quadratically with the velocity of the gas to be measured. Due to the physical conditions, the measuring range is therefore enlarged in relation to the density. The lower flow limit shifts in the direction of smaller loads.

Use the formula below as approximate equation:

)h/m(2.1

QQ 3minmfl

The density can be determined with the following approximate formula:

)m/kg(x)1p( 3nm

The influence of the temperature is not taken into account in this formula. Qmfl: Minimum flow rate at flowing conditions Qmin: Minimum flow rate of the gas meter pm: Operating pressure in bar : Density in kg/m3 n: Standard density of the gas n = standard density of natural gas 0.8 kg/m3 1.2 Density of air at 20oC and 1.01325 bar (in kg/m3)

1) can be ordered at PTB, Bundesallee 100, D-38116 Braunschweig (www.ptb.de)

Page 9: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INTRODUCTION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

5

Accuracy of measurement The following error limits apply within the permissible measuring range: Measuring range: Qmin to 0.2 Qmax 0.2 Qmax to Qmax Calibration limit1) 2 % 1 % TRZ 03-TE, TRZ 03-TEL 1 % 0.5 % TERZ 94 2 %

(DN 25 - DN 80: 3 %) 1 %

(DN 25: 2 %, DN 40/50: 1.5 %)

1) Permissible maximum error pursuant to the German Metering and Calibration Rules It will be checked whether these limits are observed. They also apply to the high-pressure range.

+ 2

- 2

- 1

% 0

+ 1

10 50 100 Fig. 1: Calibration curve of a turbine meter The reproducibility, i.e. the difference between the results of two measurements under identical conditions, is as follows:

TRZ 03-TE, TRZ 03-TEL, TERZ 94: 0.1%

Permissible temperature ranges For the standard design, the following fluid temperature and ambient temperature ranges are permitted: Fluid temperature range: -10°C to +50°C Ambient temperature range: -10°C to +60°C

Page 10: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INTRODUCTION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

6

Pressure loss The pressure loss of RMG turbine meters has been reduced to a minimum thanks to modifications with regard to design. The measuring points for pressure loss are located 1 x DN upstream and downstream of the gas meter. The pressure loss is calculated using the following formula:

4

2m

p DNQ

Zp

where: p is the pressure loss [mbar] Zp is the pressure loss coefficient is the density [kg/m³] Qm is the volume flow rate at measurement conditions [m³/h] DN is the nominal diameter of the gas meter [mm]

Device type Zp

TRZ 03 / TRZ 03-K turbine meter 3000 TRZ 03-L turbine meter 3600 L1 perforated-plate straightener as per ISO/DIN 3150 L2 perforated-plate straightener as per ISO/DIN 6300 L3 perforated-plate straightener as per ISO/DIN 9450 LP-35 perforated-plate straightener as per RMG standard 1260 RB 19 tube-bundle straightener as per ISO/DIN 1260

The values for Zp are approximate mean values. The exact value is calculated from the pressure loss which is determined on testing the volumeter. Example of calculation:

Calculation of the pressure loss for a turbine meter with upstream perforated-plate straightener.

TRZ 03, Qm = 650 m3/h, DN 150, = 1.3 kg/m3 (natural gas) From the table: Zp(TRZ03) = 3000, Zp(LP-35) = 1260

Calculation:

Zp(total) = 3000 + 1260 = 4260 mbar6.4150

6503.14260p

4

2

Page 11: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INTRODUCTION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

7

Pressure tap A pressure tap has been provided to connect the pressure transducer of a volume corrector or a pressure gauge to read the pressure at measurement conditions prevailing inside the gas meter. This pressure tap is identified by “pr”.

Using gas meters with different types of gases

Gas type Symbol Density Meter case Comments at 0°C 1.013 bar Acid gas Special Special measuring element Air 1.29 Standard Ammonia NH3 0.77 Standard O-rings / lubrication Argon Ar 1.78 Standard Biogas Special Special measuring element Butane C4H10 2.70 Standard Carbon dioxide CO2 1.98 Standard Exception: food industry Carbon monoxide CO 1.25 Standard Ethane C2H6 1.36 Standard Ethylene (gaseous) C2H4 1.26 Standard Special design Freon (gaseous) CCI2F2 5.66 Standard O-rings / lubrication Helium He 0.18 Standard Reduced measuring range Hydrogen H2 0.09 Special Reduced measuring range Hydrogen sulphide (0.2%) H2S 1.54 Special Special measuring element Methane CH4 0.72 Standard Natural gas 0.8 Standard Nitrogen N2 1.25 Standard Oxygen (100%) O2 1.43 Standard Special design Pentane C5H12 3.46 Standard Propane C3H8 2.02 Standard Propylene (gaseous) C3H6 1.92 Standard Special measuring element Sulphur dioxide SO2 2.93 Special Special design Town gas Standard

Page 12: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

SAFETY INSTRUCTIONS ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

8

Safety Instructions The TRZ 03-TE and TRZ 03-TEL turbine meters are used for measuring the volume at measurement conditions of non-aggressive gases and fuel gases. Measurements of aggressive gases are only permitted if the special designs are used which have been developed for this purpose. These gas meters are not suitable for measuring liquids, otherwise they will be destroyed. The TRZ 03-TE and TRZ 03-TEL turbine meters comply with currently applicable standards and regulations. However, failure to operate them properly may cause hazards. Persons who install or operate the TRZ 03-TE and TRZ 03-TEL turbine meters in areas subject to explosion hazards, must be familiar with the currently applicable standards and regulations with regard to explosion protection. The electronic totalizing unit of the explosion-protected design has been approved for use in areas subject to explosion hazards and its code is:

II 2 G EEx ib[ia] IIC T3, T4

The appropriate certificate of conformity can be found in the annex and its reference number is: TÜV 02 ATEX 1970

Please observe the following signs:

Danger of explosion In the manual, this symbol warns you of an explosion hazard. Please observe the instructions given next to this symbol. As to the danger of explosion, please observe the following in particular: Only the explosion-protected design of the TRZ 03-TE / TRZ 03-TEL may be used in

areas subject to explosion hazards. Connect the pulse outputs of these devices only to intrinsically safe circuits.

The battery must be changed in an area without explosion hazards. The specifications on cable type and cable length in this manual and in the ex-

approval must be observed. Modifications to the device make the Ex approval invalid and are therefore prohibited.

Damage to property In the manual, this symbol warns you of possible damage to property. The instructions given next to this symbol inform you about what you can do to avoid damage to the turbine meter.

It is essential to observe the warning information in these operating instructions and the generally applicable safety rules. No warranty claims can be asserted if there is unauthorized interference with the device!

Page 13: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

SAFETY INSTRUCTIONS ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

9

Instructions for the installer

Marking

II 2 G Ex ib[ia] IIC T4 or T3

TÜV 02 ATEX 1970 / IECEx TUN 09.0002 Ta = -25°C ..... +40 / 60°C For data, see the EG type examination certificate (see annex)

0044 0085

Manufacturer

RMG Messtechnik GmbH Otto-Hahn-Straße 5 D-35510 Butzbach (Germany)

Use

Device type Description

TERZ 94*** Volumeter EA **** Volumeter EZ **** Volumeter TRZ 03-*** Turbine meter ENCO-* Encoder index EC 21 Temperature corrector EC 24 Volume corrector TEC 21 Volumeter + temperature corrector TEC 24 Volumeter + volume corrector ETC **** Volumeter + temperature corrector EVC **** Volumeter + volume corrector This device is an apparatus for areas subject to explosion hazards.

Page 14: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

SAFETY INSTRUCTIONS ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

10

Installation or removal

When installing the device, care must be taken that the degree of protection of the casing is complied with. Exposure to direct sunlight has to be avoided. When removing the device, de-energize the signal circuits connected and remove the relevant cables properly. Installation

If one or more than one signal circuit is used, make sure when choosing the cables that the permissible limits as per the EC type examination certificate are not exceeded. Each explosion-protected signal circuit has to be installed in a separate cable which is to be taken through the relevant high-strength cable gland. It is absolutely necessary that the intrinsically safe cables are permanently installed. Make sure that the connecting cables are provided with wire-end sleeves. Commissioning

Before commissioning this equipment, make sure that all lines have been correctly installed and connected in the terminal compartment. The casing must be completely closed. For installation and commissioning, the IEC 60079-14 of 2004 / EN 60079-14 of 2006 standard has to be complied with. National norms and/or regulations also need to be taken into consideration. Servicing and maintenance

The battery may only be changed in a non-incendive area. This device may be repaired only by RMG Messtechnik. Modifications to the product

Changes or modifications made to this product by unauthorized persons may lead to incorrect malfunctions and are prohibited for safety reasons! In this case the [ATEX] approval will be no longer valid. This device may be repaired only by RMG Messtechnik!

Page 15: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INSTALLATION AND COMMISSIONING ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

11

Installation and Commissioning

Installation Caution: Please read these installation instructions through and make sure that you have under-stood the procedure before you start to install the RMG turbine meter or put it into operation. Turbine meters are precise measuring instruments and must be handled carefully during transport, storage and operation. When you install the turbine meter, please observe the direction of flow indicated by an arrow on the meter case.

Installation errors may cause physical injuries or result in the destruction of the turbine meter.

It is essential to follow the instructions below:

Remove the yellow protective film at the flanges completely. Rests of this plastic film affect the flow profile and cause measuring errors!

The RMG TRZ 03-TE, TRZ 03-TEL and TERZ 94 turbine meters can be operated in any position up to the nominal diameter of DN 200. From the nominal diameter of DN 250, however, only a horizontal direction of flow is possible. If a particular position was indicated when ordering the turbine meter, you must install the turbine meter in this position. In addition, you must make sure that the filling opening of the lubricator faces upwards.

Any components affecting the gas flow must be avoided directly upstream of the turbine meter (see DVGW Guideline G 492 II1) and PTB Guideline G 132), exception: TRZ 03-TEL).

An inlet pipe of a minimum of 2 x DN is required upstream of the RMG TRZ 03-TE turbine meter. The inlet pipe must be designed as a straight pipe section of the same nominal diameter as the gas meter. If there is a heavy flow disturbance, it is necessary to use flow straighteners (see table on next page). A pipe or fitting (bend) of the same nominal diameter as the gas meter and an overall length of 2 x DN must be installed downstream of the gas meter. Temperature measuring instruments may only be installed at a distance of 1 x DN or in the case of nominal diameters of DN 300 at a minimum distance of 300 mm. If there are flow disturbances (e.g. due to a gas pressure regulator) upstream of the inlet pipe, it is also necessary to use a perforated-plate straightener. You can use perforated-plate straighteners complying with ISO 5167-1 or of type RMG LP-35, the latter resulting in a pressure loss which is 2.5 times lower than that of the standardized flow straightener.

1) can be ordered at DVGW, Josef-Wirmer-Str. 1-3, D-53123 Bonn (www.dvgw.de) 2) can be ordered at PTB, Bundesallee 100, D-38116 Braunschweig (www.ptb.de)

Page 16: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INSTALLATION AND COMMISSIONING ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

12

2 DN

2 DN

1 DN

Perforated-plate straightener

d

e

d

DN

DN

d = e = 0.13 DN•

LP-35 perforated-plate straightener

Upstream of the RMG TRZ 03-TEL turbine meter, no inlet pipe is required, even in the case of

heavy flow disturbances such as those caused by a gas pressure regulator. It has been tested without an inlet pipe in compliance with Technical Guideline G13 (which corresponds to the OIML Guideline IR-32/89). A pipe or fitting (bend) of the same nominal diameter as the gas meter and an overall length of 2 x DN must be installed downstream of the gas meter. Temperature measuring instruments may only be installed at a distance of 1 x DN or in the case of nominal diameters of DN 300 at a minimum distance of 300 mm.

The opening angle of reducers or expansion fittings installed upstream of a turbine meter of

type TRZ 03-TE, TRZ 03-TEL or TERZ 94 must not exceed 30°.

In order to obtain precise measurement results, the turbine meter must be installed in the gas line in such a way that no seals protrude from the flanges into the pipeline.

A protective screen should be installed on the intake side of the gas meter to protect the turbine meter against any foreign particles possibly contained in the gas flow. The protective screen can be a perforated plate of sheet metal with a hole diameter of 3 mm which is available from RMG.

The pr tap which is located on the RMG turbine meter is the pressure-measuring point which was used for taking the relevant pressure at measurement conditions during meter proving. This pressure-measuring point is used for connecting pressure-measuring instruments such as flow computers or volume correctors. Other connection options (e.g. for temperature measurement) can be supplied on a pipe section on the output side of the gas meter.

Page 17: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INSTALLATION AND COMMISSIONING ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

13

Caution: Protect the turbine meter from damage caused by heavy flow variations, e.g. if the downstream pipeline system must be filled or blown off.

Caution: If it is necessary to do welding work on the gas line, such work can only be performed at a safe distance from the gas meter. Extreme temperatures prevailing in the gas line in the proximity of the gas meter can result in a permanent damage to the gas meter.

Caution: All electrical connections between the gas meter and the amplifiers or the flow computer must be carried out in compliance with the installation instructions. Make sure that these connections are intrinsically safe.

Caution: Any liquids remaining in the gas line after hydrostatic testing can damage the interior parts of the gas meter. If hydrostatic testing is necessary, the turbine meter must be replaced by a pipe section. Make sure that no liquid remains in the gas line upstream of the gas meter after hydrostatic testing.

Operating data

Recommended threshold values for maximum service life and maximum accuracy:

Maximum overload: < 20% above Qmax, for a short time (< 30 sec)Maximum flow changes or shock loads

< 0.01·Qmax/sec ̂ 1% of Qmax/sec e.g. starting up 0 to 100%: > 100 sec

Maximum pressure change: < 0.1 bar/sec Maximum flow pulsation: < 5% Particle size in the gas flow: < 5 mm Lubrication of bearings: See chapter on “Lubrication”.

Lubrication intervals depend on the condition of the gas (condensate, rust or dust).

Vibration / mechanical shock: < 1 mm/sec (vibration velocity) These specified data must be determined and checked during commissioning, prior to filling and during the start-up and running-in phases of the gas meters. If more than one threshold value occurs at the same time, appropriate action must be taken in the station to improve the measuring conditions. The operator must record all measuring data (gas meter and operating data) during the whole period of operation in order to detect the causes of a possible destruction of the gas meter at an early stage and take corrective action in good time.

Page 18: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INSTALLATION AND COMMISSIONING ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

14

Corrective action or reduction of the critical operating conditions can be achieved by the following measures for example: Start-up screens (mesh width < 0.15 mm) Filters Perforated plates (Ø 3 to 4 mm) protecting the gas meter Valves with control drive mechanism (flow variation) Non-return valves (pulsation and return flow)

Technical Guideline G 13 The installation conditions for new stations complying with TRG G 13 and the simpler installation conditions for RMG turbine meters are compared in the table below.

Type of flow disturbance

Installation cond. as per TR G13

Installation cond. for RMG gas meters Type TRZ 03-TE

Comments

Inlet pipe 5 DN Outlet pipe 2 DN

Inlet pipe 2 DN Outlet pipe 2 DN

The outlet pipe can also be designed as a bend.

None Inlet pipe 10 DN Flow disturbances upstream of this inlet pipe need not be considered, if the require-ment for an alternating and pulsating flow is fulfilled.

Bend Inlet pipe 5 DN Inlet pipe 2 DN Space bend Inlet pipe 5 DN

and additionally 2 perforated-plate straighteners or one tube-bundle straightener

Inlet pipe 2 DN

Gas pressure regulator with sound attenuator

Inlet pipe 5 DN Inlet pipe 2 DN and additionally one perforated-plate straightener

Gas pressure regulator without sound attenuator

Inlet pipe 5 DN and additionally two perforated-plate straighteners

Inlet pipe 2 DN and additionally one perforated-plate straightener

Diffuser Inlet pipe 5 DN and additionally one perforated-plate straightener

Inlet pipe 2 DN

Diffuser with swirling flow

Inlet pipe 5 DN and additionally two perforated-plate straighteners

Inlet pipe 2 DN

Page 19: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INSTALLATION AND COMMISSIONING ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

15

Perforated-plate straighteners

There are the following options for flow straighteners:

RMG L1 to L3 perforated-plate straighteners complying with ISO 5167-1 and DIN 1952

RMG LP-35 perforated-plate straightener

d

e

D

a

d

e

D

D

Features ISO / DIN L1 to L3 RMG LP-35 Hole diameter d d 0.05 D 0.04 D 0.13 D Plate thickness e e d e = d 0.13 D Plate clearance a 0.5 D a 1 D 0.5 D - Opening ratio m 0.2 m 0.4 0.3 0.6 Pressure loss, dyn. p 5 - 15 (c2 / 2) 2 - 15 (c2 / 2)

In conjunction with RMG turbine meters, these flow straighteners fulfil the requirements of Technical Guideline G 13 and are approved under EU approval No. D 81 / 7.211.10 for turbine meters.

Page 20: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INSTALLATION AND COMMISSIONING ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

16

Seals It must be guaranteed that flange seals of RMG turbine meters do not protrude from the flange into the gas line.

All seals approved as per DVGW can be used depending on the requirements for stability and reli-ability.

We recommend seals with the following maximum material characteristic values according to the AD2000 rules: − gaskets: k0 x KD = 20 x bD | k1 = 1.3 x bD [N/mm] − grooved seals: k0 x KD = 15 x bD | k1 = 1.1 x bD [N/mm] − spiral seals: k0 x KD = 50 x bD | k1 = 1.4 x bD [N/mm] − octagonal ring joint seal: KD = 480 N/mm2 For recommended dimensions, see the tables below.

Gaskets PN 10 PN 16 ANSI 150 PN 25 PN 40 DN d1 d2

50 2" 77 107 107 105 107 107 80 3" 90 142 142 137 142 142

100 4" 115 162 162 175 168 168 150 6" 169 218 218 222 225 225 200 8" 220 273 273 279 285 292 250 10" 274 328 330 340 342 353 300 12" 325 378 385 410 402 418 400 16" 420 490 497 514 515 547 500 20" 520 595 618 607 625 628 600 24" 620 695 735 718 730 745

Grooved seals ANSI 300 / ANSI 600 PN 64

DN d1 d2 d1 d2 50 2" 107 107 107 107 80 3" 142 142 142 142

100 4" 162 162 168 168 150 6" 218 218 225 225 200 8" 273 273 285 292 250 10" 328 330 342 353 300 12" 378 385 402 418 400 16" 490 497 515 547 500 20" 595 618 625 628 600 600 695 735 730 745

1.5 up to 5 mm d1

d2

Page 21: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INSTALLATION AND COMMISSIONING ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

17

Spiral seals ANSI 300 PN 64 ANSI 600 DN d1 d2 D1 d2 d1 d2

50 2" 69.9 85.9 66 84 69.9 85.9 80 3" 101.6 120.7 95 119 101.6 120.7

100 4" 127.0 149.4 120 144 120.7 149.4 150 6" 182.6 209.6 174 200 174.8 209.6 200 8" 233.4 263.7 225 257 225.6 263.7 250 10" 287.3 317.5 279 315 274.6 317.5 300 12" 339.9 374.7 330 366 327.2 374.7 400 16" 422.4 463.6 426 466 412.8 463.6 500 20" 525.5 577.9 530 574 520.7 577.9 600 24" 628.7 685.8 630 674 628.7 685.8

Screws

Temperature ranges for screws and nuts

-10°C to +80°C -40°C to +80°C

Pressure Variant 1 Variant 2 Variant 3

up to and including 40 bar

Screws complying with DIN EN ISO 4014 made of material 5.6,

Nuts complying with DIN EN ISO 4032 made of material 5-2

Screws complying with DIN EN ISO 4014 made of material 25CrMo4,

Nuts complying with DIN EN ISO 4032 made of material 25CrMo4

from 40 bar Screw bolts complying with ANSI B1.1 made of material ASTM A193 Grade B7,

Nuts complying with ANSI B1.1 made of material ASTM A194 Grade 2H

Screw bolts complying with ANSI B1.1 made of material ASTM A320 Grade L7,

Nuts complying with ANSI B1.1 made of material ASTM A320 Grade L7

Screw bolts complying with ANSI B1.1 made of material 42CrMo4,

Nuts complying with ANSI B1.1 made of material 42CrMo4

Anti-fatigue bolts complying with DIN 2510 made of material 25CrMo4,

Nuts complying with DIN 2510 made of material 25CrMo4

Page 22: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INSTALLATION AND COMMISSIONING ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

18

Block diagram 2-channel (TRZ 03-TE and TRZ 03-TEL)

2-channel

Channel 1

Channel 2

1K

12345678 m3

HF output

LF output

1n

fi

Current output

1n

Alarm contact

Option:EZD or RS485 interface

With 2 channels:pulse

monitoring

10 / 10000180 °

You can install the meter head also vertically. To do this, you must proceed as follows:

Unscrew the cover of the meter head. Disconnect the two wires from terminal X5 (remember the polarity!). Unscrew the fastening screws of the meter board.Remove the board carefully from the case. If

necessary, disconnect the two wires from the button. Loosen the set screw (2). Now you can remove the case of the meter head and an aluminium

assembly part will appear. On the side of the case, there is a hole which is covered with a screw. Mount the retainer part

to this hole and close the hole in the rear panel by means of the screw. Thread the signal wires through the retainer part, place the meter head onto the turbine case,

turn it into the desired position and retighten the set screw (2). Insert the board again into the case, connect the signal wires again to terminal X5 (make sure

that the polarity is correct!) and screw on the cover again.

Page 23: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INSTALLATION AND COMMISSIONING ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

19

Power Supply Depending on the type of the device, there are different options of power supply.

1. The basic TRZ 03-TE model is supplied by an internal lithium battery which enables to operate the device continuously for a period of approx. six years. This, however, is conditional on the device being read once a week or “woken up” by pressing the external button.

2. If an interface module is used for transmitting data, such as the ENCO with external supply, the service life of the battery is more than 10 years.

3. The TRZ 03-TE / TEL (current transmitter) is completely supplied through a current loop. If pulse processing is to be maintained even in the case of a power failure of the current loop, a standby battery (available as an option) should be installed which will bridge the power supply during a period of up to 6 months.

Electrical connections The LF and HF volume pulses and the alarm output are led out of the meter head by a 7-pin connector (Binder). All other connections have to be made at the terminals on the board. Standard connection battery version and mains powered version:

The supply/current-loop (mains powered version only) is connected to the terminals X22 on the transmitter board TERZ94trm.

In areas subject to explosion hazards, the TRZ 03-TE / TEL must only be connected to certified intrinsically safe circuits.

Make sure that the limiting values from the conformity certificate (see appendix) for the devices, which have to be connected, are not exceeded.

1

2

34

5

PE (Abschirmung)

6

1 - / 4 + LF signal 2 - / 5 + Alarm 3 - / 6 + HF signal

PE (Screening)

Page 24: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INSTALLATION AND COMMISSIONING ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

20

If one or more electric circuits are used, it is to ensure that with the cable selection the permissible limiting values according to the EEC design inspection certificate are not exceeded.

Each ex signal circuit has to be wired in its own cable, which is to be led through the appropriate cable gland.

A fixed installation of the intrinsically safe cables is mandatory.

The connecting cables have to be provided with core-end sleeves. To reach the electrical connections on the board, you must first remove the cover of the meter head.

To convert from the meter head directly mounted on the meter case to the remote totalizing unit and vice versa it is necessary to change the positions of some jumpers (see inputs and outputs in the annex). X_S1 (and X_S2*) X_TERZ90 Meter head mounted on meter case 3-5 and 4-6 all open Remote totalizing unit 3-5 and 4-6 1-2 (and 3-4*) * for 2-channel measurement only

Jumpers X_S1 and X_S2 for

configuration of pulse inputs

Plug X3 Service module

Pin connector button on the case

Reset (soldering pads)

Terminal X5 Pulse inputs

Plug X2_0 for current module TERZ94trm and supply

Terminal block X4 Pulse outputs

Jumper X_TERZ90

Programming keys

Page 25: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INSTALLATION AND COMMISSIONING ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

21

Battery powered version: Controlling the start-stop totalizer or resetting the resettable totalizer (depending on the programming of the electronic totalizing unit) is performed through input X5 terminals 1 and 2. As soon as input X5 terminals 1 / 2 has been short-circuited through an external contact, interruption or resetting is performed.

For this purpose, set jumpers at the positions identified with XS 2 to the “reed contact” function.

Mains powered version: In the case of the TERZ 94-S, terminal X22 (on current module TERZ94trm) is used as current-loop connection to supply the device and as output current (4-20 mA). To connect cables to the spring terminals, you need a screwdriver with a blade width of a maximum of 2.5 mm. Introduce the blade into the intended slot and press down the screwdriver to open the spring terminal. Earthing

To prevent measuring errors caused by electromagnetic interference, you must not fail to earth the case of the meter head via the earthing screw on the left side of the case.

Thereby make sure that you have an electrical link between TERZ 94 and the piping, as it is shown in the drawing below.

Minimum cable cross section: up to a length of 10 m: 6 mm2 from a length of 10 m: 10 mm2

Earthing screw

Page 26: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INSTALLATION AND COMMISSIONING ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

22

potentialmeasuringstation

potentialmeasuringstation

Outputs

P

equipotential bondingline, min. 6 mm2

- 0085

IP65

Current output 4-20mA yes no

24 VDC

Cables

Use 2-core or multicore shielded cables which are twisted together in pairs (type LIYCY) for the signal lines (LF output, HF output, current-loop connection, control input). The shielding must always be connected to earth on both sides. In the case of the battery powered version, you must proceed as described in the chapter “Cable glands”. We recommend that cable cross sections between 0.25 mm2 and 0.5 mm2 are used. Due to the cable gland, the outside diameter must be between 4.5 mm and 6.5 mm.

When the device is used in areas subject to explosion hazards, the maximum cable length is limited by the limit values for intrinsically safe circuits and depends on the inductance and capacitance of the cable!

In particular: The maximum cable length between the TRZ 03-TE and the transmitter supply unit of type KFD2-STC3-Ex1, in connection with the cable type LiYCY 2 x 0.75 mm2, is 250 m for the current output. You can also use other cable types, if the following cable values are not exceeded: Capacitance < 200 nF/km inductance < 1 mH/km

Page 27: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INSTALLATION AND COMMISSIONING ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

23

Cable glands

Clamp the shielding on both sides, as shown in the picture below, into the cable glands located on the outside of the case:

Unscrew the cap nut. Pull out the plastic clamping piece. Push the cable end through the cap nut and the clamping piece and bend the shielding

backwards. Put the clamping piece back into the connecting piece. Screw on the cap nut again.

Cap nut

Clamping piece

O-ring

Connecting piece

Page 28: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INSTALLATION AND COMMISSIONING ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

24

Installing the remote totalizing unit

Sensor

Sensor

ZählwerkZählwerk

-

-

+

+Index InputIndex Input

-

+

Sensor

Sensor

ZählwerkZählwerk

-

-

+

+Index InputIndex Input

-

+

ServicemodulTERZ94trm

TERZ94p

Impulsausgänge

Datenschnittstelle

Datenschnittstelle

EZD/RS-485 Boot

TP-GND

Impulseingang

X2_0

X4

X15

X9

X1

51

X5

CX

2

CX

3

CX

6

CX

1

L1

X6_0

X152

X_S1

XLP1

Ree

d 1+

3 /

2+4

Se

nsor

vers

tärk

er 3

+4

Wie

gan

d 3+

5 / 4

+6

4

2 6

531X_S2 X10-0

TEMP_LM7X

EZSENS01

PT1000

5313

2

1

C21

C1

V12

V13

V14

C22

Q1

X7

X16_0X1X3 Taste

r

RE

SE

T

X_TERZ90

S3 S1 S2 S4P

+

+-

-

4

5

1

2

- 0085

IP65

Current output 4-20mA yes no

24 VDC

If your TRZ 03-TE is designed for remote totalizing, you can install the totalizing unit at a distance of up to 50 m from the meter case. Usually, the cable has already been connected to the totalizing unit when the TERZ 94 is delivered. Should this not be the case, you will have to connect the connecting cable to Input-terminal X5, clamp S1+ and clamp S1- of the board. For 2-channel meters connect the second sensor to the clamps S2+ and S2-. Use only shielded cables of the type:

LIYCY - 2 x 0.75 blue (1-channel version) LIYCY - 2 x 2 x 0.75 blue (standard) Maximum cable length: 50 m Furthermore, you must check the plug-in jumpers XS1, XS2 and XTERZ90 on the display board. Settings at XS1 and XS2: jumpers 3-4

Page 29: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

INSTALLATION AND COMMISSIONING ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

25

Temperature measurement In order to measure the gas temperature, a resistance thermometer can be used in a thermowell in a fitting downstream of the gas meter. It is recommended that a second thermowell, e.g. for a monitoring thermometer, be used. If such thermowells are not provided for in the meter case, temperature measurement must be performed at a distance of up to 3 x DN or a maximum of 600 mm downstream of the turbine meter. All turbine meters of types TRZ 03-TE and TRZ 03-TEL from the nominal diameter of DN 80 (3") upwards can be fitted with a thermowell for a PT100 resistance thermometer. Due to their short size, turbine meters of type TERZ 94 cannot be fitted with a thermowell. Operating temperatures prevailing at outdoor metering stations in winter or at natural gas metering stations downstream of regulating stations are usually in the range of -5C to +10C. By comparison, higher operating temperatures prevail at metering stations located downstream of compressor stations. For this reason, the measuring elements of the temperature sensors located outside the gas line must be appropriately insulated against the weather. In order to achieve optimum thermal conduction, oil as a heat transfer liquid should be filled in the thermowells in any case.

Page 30: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

COMMISSIONING ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

26

Commissioning

Filling with oil When the turbine meters are delivered, the lubricators do not contain any oil. The oil pumps must not be filled with oil until they are at the installation site prior to commissioning! A small bottle of oil is supplied with each gas meter. For detailed lubrication instructions, see chapter “Lubrication” from page 40.

Connecting the gas flow

Do not fill any downstream pipelines or station sections through the turbine meter. This may speed up the turbine wheel and lead to excessively high loads with resultant damage.

Short-time overload operation of 20% above the maximum flow rate Qmax is permissible. Such load conditions should be avoided, however, since in most cases they cannot be controlled and exceed the officially approved limits for custody transfer metering. Moreover, such overloads reduce the service life of the gas meter. The gas flow must be free of foreign particles, dust or liquids. Otherwise it is recommended that filters or separators be installed. Before starting up the device, make sure that all lines have been correctly installed and connected in the terminal compartment. The case must be completely closed.

Initializing the totalizing unit When you start up the device for the first time, set the totalizing unit in the desired condition. Check the programming of the pulse width, the LF pulse scaler, etc. Mains powered version: Check the programming of the current output mode. Note: All parameters can only be changed after the cover of the meter has been removed.

Page 31: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

OPERATION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

27

Operation

Display In normal operating mode, the main totalizer is displayed. When you press the control button on the right side of the case, all values will be shown. After the preset display wake-up time has elapsed, the main totalizer will appear again. If the display of the TRZ 03-TE / TEL does not show anything, the device is in energy-saving mode. In this mode, the display is switched off. However, incoming pulses are processed and the outputs are set. As soon as you press the control button, the display values will appear again. Display switched off:

0 0 0 0 5 8 3 1 m3 Main totalizer Vm

1 0. 0 0 0 m3/h Flow rate

0 0 0 0 0 0 0 0 m3 Start / stop totalizer

0 0 0 0 0 0 0 0 m3 Resettable totalizer

5 0. 0 0 P/m3 Meter factor

Page 32: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

OPERATION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

28

1 -- Pulse scaler

000 0 0 0 0 1 MOD Configuration

0000008000 S.No. Serial number

Extra values for version with external supply

80.38 Hz Frequency (sensor)

0 0 0 0 5 8 3 1 m3 Main totalizer

An overview of parameters and the complete matrix are shown in the annex. NOTE:

The button located externally on the case reacts when it is released and not when it is pressed. The duration of pressing the button is evaluated and results in different reactions of the device. Pressing the button for a short time causes the display to show the next display value, while pressing the button for a longer time (more than 2 seconds) will activate all segments of the display (segment test) and then the display will switch over to showing the Vm totalizer. The following text describes procedures where the button is normally pressed for a short time.

Page 33: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

OPERATION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

29

Programming For programming the TRZ 03-TE / TRZ 03-TEL, there are four buttons on the rear side of the board. Alternatively, you can program the device with the programming module (available as an accessory). The programming module is to be connected to the pin connector indicated in the picture below.

Programmingmodule

Programmingkeys

P

Ser

vice

mo

dul

TE

RZ

94

trm

TE

RZ

94p

Impu

lsa

usg

än

ge

Da

ten

sch

nitts

telle

Dat

ensc

hnitt

stel

le

EZ

D/R

S-4

85B

oot

TP

-GN

D

Imp

uls

ein

gan

g

X2_

0

X4

X1

5

X9

X151

X5

CX2

CX3

CX6

CX1

L1

X6_0

X152

X_S

1

XLP

1

Reed 1+3 / 2+4Sensorverstärker 3+4Wiegand 3+5 / 4+6

4

26

53

1X

_S2

X10

-0T

EM

P_L

M7

XEZ

SE

NS

01

PT

1000

53

13

2 1

C21

C1

V12

V13

V14

C2

2

Q1

X7

X16

_0X

1

X3

Taster

RESET

X_

TE

RZ

90

S3

S1

S2

S4

P

Programming with the programming module is to be performed in the same way as with the buttons.

P

+

-

Page 34: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

OPERATION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

30

The external and internal buttons correspond to each other in the following way:

Int. button Ext. button Meaning

P P Display mode: Switch over to programming mode (Press the button longer than 2 seconds) Programming mode: Set the decimal point.

+ Display mode: Switch over to the next higher level. Programming mode: - Increase the decimal by 1. - Scroll in the list. (Value identified with “L“)

- Display mode: Switch over to the next lower level. Programming mode: - Decrease the decimal by 1. - Scroll in the list. (Value identified with “L“)

Display mode: Short-time view of the coordinate (e.g. A01) Programming mode: Go to the right by one decimal place (if the last decimal has been reached: Quit programming mode).

Page 35: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

OPERATION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

31

Principle of programming

For programming, you must always proceed as follows:

First change over to the display value to be modified. To do this, press either the control button (only forwards) or the internal buttons “+” and “-” or the external buttons “” and “” (forwards and

backwards).

Change to programming mode by pressing “P” for at least 2 seconds. On the left side of the display, a flashing character or cursor will appear. If necessary press the “” key so many times until a flashing character appears.

0 1 2 3.2 3 P/m3

You can now modify the flashing decimal by pressing either “+”or “” (+1) or “-” or “” (-1). Example: If you press the ““ button three times, the first decimal will be increased from 0 to 3. If an “L” appears on the far left side of the display, this value is a list. With a list, you can only scroll in the specified values.

After you have completed your programming of the first decimal, press “” once and the next character will start to flash. Now proceed with your programming until you have reached the last decimal place.

Then you must press “” once again to have the set value accepted and quit programming mode.

Press the “P“ button to set the decimal point behind the flashing digit. With totalizers, modes and integers, no decimal point is permitted.

Press the control button if you have made an error or if you want to discontinue inputting data.

Display values

Measured values, such as flow rate, frequency, etc. are display values and cannot be directly modified. However, there are many parameters which influence the formation of these measured values. These parameters are described in the following section. Examples of display values: flow rate, version number, year of construction, serial number, value of the current output in mA, etc.

Flashing character

Page 36: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

OPERATION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

32

Parameters and Modes of the TRZ 03-TE / TEL In the following text, the individual parameters are described in detail.

Meter factor (pulse value)

With the meter factor (pulse value), the relevant flow rate at measurement conditions is calculated from the signal frequency of the sensor element in the electronic totalizing unit:

h

m3600

K

fQ

3

M

f: Signal frequency (Hz) K: Meter factor (pulses/m3) QM: Flow rate at measurement conditions (m3/h) The meter factor has been calibrated in the factory in such a way that working cubic metres are directly displayed. Its value can be read from a label located on the left side of the case. Any modification of this adjustment is within the sphere of responsibility of the operating company. NOTE! The new value is immediately used for all calculations performed after each modification of the meter factor. The uninfluenced signal frequency of the sensor element is available at the HF output. The frequency range can be determined from the meter factor K and the minimum and maximum flow rates at measurement conditions of the meter in accordance with the following formulae:

K3600

Qf minMmin K

3600

Qf maxMmax

QMmin: Minimum flow rate at measurement conditions QMmax: Maximum flow rate at measurement conditions K: Meter factor (pulse value)

Example:

QMmin = 16 m3/h QMmax = 250 m3/h K = 2362 pulses/m3

Hz5.10Hz23623600

16fmin Hz164Hz2326

3600

250fmax

Page 37: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

OPERATION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

33

LF pulse scaler

The frequency of the LF pulses is set by a pulse scaler (A07). There are the following setting options: 0.01; 0.1; 1; 10; 100. Other values can not be programmed, it is only possible to scroll in the list of these values (‘t’ key resp. ‘_’ key). This pulse scaler has only effect on the frequency of the LF pulses not on the display. The number of displayed digits after decimal point or a multiplier are programmed as an operating mode (A08) Recommended values (= values set in the factory)

DN

[mm]

Pulse scaler

[pulses/m3]

LF decade scaler

Pulse rate

[pulses/m3]

Maximum output frequency

[Hz] 1)

25 10 1 10 0.1

40 10 10 1 0.1

50 10 10 1 0.1

80 10 10 1 0.1

100 10 10 1 0.1

150 1 1 1 0.3

200 1 1 1 0.6

250 1 1 1 0.7

300 1 1 1 1.1

400 1 10 0.1 0.2

500 1 10 0.1 0.3

600 1 10 0.1 0.4

1) Rounded values

Page 38: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

OPERATION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

34

Configuration

The basic configuration of the TRZ 03-TE / TEL is implemented via an 8-character display value (A08). This “TERZ mode” defines the operating mode of the device. Some parameters require a specific hardware or hardware configuration. NOTE! Hardware and software configurations must coincide. Any modifications of these parameters can cause malfunctions. Let’s start on the left side of the display: (operating alignment italic) Internal parameters

For testing and diagnostics (no influence on the metering result). 0: Menu for standard operation. 1: Menu for service 9: All parameters visible.

Main totalizer mode

This parameter defines the unit for the gas volume 0: m3 (cubic meter). 1: cf (cubic foot).

LF pulse width

The LF pulse width can be set at 125 ms or 250 ms. 0: 125 ms 1: 250 ms

Cut-off time of the display

A few minutes after you have operated the device for the last time, the display will switch off to expand the service life of the battery of the TRZ 03-TE / TEL. There are the following times for selection:

0: 1 minute 1: 5 minutes 2: 10 minutes 3: 15 minutes 4: 1 hour (for test purpose only!)

Interface protocol

0: OFF 1: EZD protocol 0x00 / Standard Vo protocol (NAMUR) 2: EZD protocol 0x7f / Vo protocol for external devices (NAMUR) 3: Test protocol

Page 39: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

OPERATION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

35

Pulse scale In this operating mode the display is set. To get the real totalizer readings the displayed value must be multiplied with the multiplier.

0: 2 digits after decimal points, multiplier: 1 1: 1 digit after decimal points, multiplier: 1 2: no digits after decimal points, multiplier: 1 3: no digits after decimal points, multiplier: 10 4: no digits after decimal points, multiplier: 100

Current output

0: OFF 1: Current output is active 2: Current output is active and faults are signalled (3,5 mA) 3: Current output is active and faults are signalled (21,8 mA)

The current values in brackets are outputted in case of a fault. TERZ operating modes

There are the following operating modes at the moment: 0: ENCO-F 1: ENCO-M 2: TERZ 94 1 channel 3: TERZ 94 1 channel with start / stop totalizer 4: TERZ 94 1 channel with reset totalizer 5: TRZ03-TE/TEL 2 channels with pulse monitoring (SW), Vm Stop 6: TRZ03-TE/TEL 2 channels with pulse monitoring (SW), Vm Run 7: TRZ03-TE/TEL 2 channels with pulse monitoring (HW), Vm Stop 8: TRZ03-TE/TEL 2 channels with pulse monitoring (HW), Vm Run 9: TRZ03-TE/TEL 2 channels without pulse monitoring

SW means “pulse monitoring by software”, HW by hardware; Vm Stop means totalizer is stopped in case of a fault whereas the counting continnes in the mode Vm Run.

Example:

5: TRZ 03-TE/TEL – 2 channel, pulse monitoring (SW) active, Vm Stop 1: Current output is active, faults are not signalled. 1: Display of 1 digit after decimal point 0: Interface protocol off 1: Display cut-off time: 5 minutes 1: LF pulse width: 250 ms 0: Display in m3 0: Operating mode: No diagnostics mode is activated

00110115 MOD

Page 40: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

OPERATION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

36

Pulse comparator

The 2-channel version of the TERZ-94 (TRZ03 TE / TEL) provides for pulse monitoring. Pulse monitoring requires two input pulses which are out of phase. If a pulse occurs at the same time on channel 1 and channel 2, a fault counter is increased. If there is a missing pulse on one of the two channels, a fault counter is increased, too. If this fault counter reaches the value X(A10) before Y(A11) undisturbed pulses have entered one of the two channels, a fault will be outputted. This fault will not be erased until Y undisturbed pulses have been counted. Flow rate at 4 mA

With this parameter, you set the flow rate (in m3/h) which corresponds to an outputted current of 4 mA. Generally, the minimum flow rate Qmin of the meter is set here. (This setting is made in the factory.) Flow rate at 20 mA

With this parameter, you set the flow rate (in m3/h) which corresponds to an outputted current of 20 mA. Generally, the maximum flow rate Qmax of the meter is set here. (This setting is made in the factory.) Then the current which will be outputted for flow rates within the programmed limits will be between 4 and 20 mA and results from linear conversion. Current default value (calibration current)

With this parameter, you set the value of the calibration current. A constant calibration current is outputted if the current output mode parameter is set to default. The value of the calibration current must be within the following limits:

Minimum value: 4 mA Maximum value: 20 mA

Page 41: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

OPERATION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

37

Current output mode

Here you define the measured or default value which is to be outputted. 0: Default value (see above) 1: Flow rate Qm All other values are not permitted! Current damping

Here you define how the current output is to be damped. Permissible values for D are 0 to 0.9 ! Output current = (1-D) x value + D x old value Correction factors for current offset and current rise

The current output is adjusted in the factory through two correction factors (offset and rise). In the event that a recalibration should really be necessary, there is a PC-based program available which will define the new offset and rise values for you.

Page 42: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

OPERATION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

38

Influences on the error of measurement due to operation

Intermittent operation

Rapid changes of the gas flow should be avoided, since the turbine wheel can follow these changes only with a certain delay due to its inertia. In this way, errors of measurement occur, in particular, if the gas flow is shut off. Since in this case the turbine wheel slows down after the flow has stopped, a gas volume is measured which is always higher than that which has actually flowed through the gas meter. In the supplement to PTB Guideline G 13, an instrument is stipulated which records the flow during the slow-down time of the turbine wheel for intermittent operation (i.e. if the gas flow is permanently switched on and shut off). If control signals of a shut-off valve are available, this volume can be recorded for custody transfer metering purposes for example by the TAZ 9, an RMG device recording the gas volume during the slow-down time of turbine meters after flow stop (for custody transfer metering only in conjunction with the TRZ 03-TE or TRZ 03-TEL). As soon as the shut-off valve is closed, the TAZ 9 starts to record the gas volume during the slow-down time of the turbine meter.

Influence of pulsations

The gas flow must be free of shocks or pulsations. A gas metering station can be affected by flow pulsations, however, if the following equipment has been installed upstream or downstream in the system:

reciprocating piston compressors, rotary displacement meters, gas pressure regulators lacking steadiness of operation, pipes where no gas flows (siphons). Volume flow pulsation is the decisive quantity for evaluating the performance of gas meters under the influence of pulsations. Volume flow pulsation is physically always associated with pressure variations. The following relation is established in a first approximation:

Q pDN

QKrel rel

2

where:

Q̂rel: is the relative volume flow pulsation (peak-to-peak)

Q̄: is the mean volumetric flow

p̂rel: is the relative pressure pulsation (peak-to-peak) DN: is the nominal diameter of the gas meter K: is a constant depending on standard density, velocity of sound, compressibility, pressure at

base conditions, temperature, and station-specific parameters.

Page 43: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

OPERATION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

39

With this relation, it is possible to estimate volume flow pulsation on the basis of pressure pulsation, which can be measured more easily. Direct measurement of volume flow pulsation is preferable, however, since the results are more reliable. The crucial factor is the pulsation at the location of measurement.

Consequences

In the case of pulsating flow, the turbine meter shows a measured value which is too high. Due to the pulse applied by the flow on the impeller which increases quadratically with the flow velocity, the resultant rotational speed is higher than the mean value of the flow velocity. The influence with high gas density is less, but increases with a high mass moment of inertia (heavy impeller) or fast-running wheels. Furthermore, high pulsation amplitudes may result in premature wear of shaft bearings due to increased load.

Limiting values

Frequency ranges It is generally unlikely that measured values will be distorted in the frequency range above

100 Hz. In practice, it is hardly possible to initiate any significant flow variations at such frequencies. Disturbances are to be expected most frequently in the frequency range between 0.1 Hz and

100 Hz, since with typical station dimensions, the gas column can be expected to produce resonances. Flow variations with a high relative amplitude may occur.

In the frequency range below 0.1 Hz, there is a quasi-steady flow which will not cause any distortion with the gas meters.

Pulsation amplitudes

Studies have shown that no disturbances are to be expected in the case of relative flow pulsations below 5% (peak-to-peak) and relative pressure pulsations of less than 0.1% to 0.5% (peak-to-peak). These data should be regarded as approximate values depending on the flow rate and pulsation frequency involved.

Page 44: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

OPERATION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

40

Lubrication Lubricator For lubrication, either an oil gun or a permanently fitted oil pressure pump is used. The various designs are given in the tables below.

Type of oil pump Container volume Delivered quantityDO Oil gun (oil nipple) 150 cm3 0,6 cm3/stroke KO Pushbutton pump 8 cm3 0,114 cm3/stroke GO Lever pump 150 cm3 1,5 cm3/stroke DS Permanent lubrication -- --

TERZ 94 TRZ 03-TE and TRZ 03-TEL

Pressure classes Pressure classes DN PN 10, 16

ANSI 150 PN 25, 40, 64, 100

ANSI 300, 600 PN 10, 16 ANSI 150

PN 25, 40, 64, 100 ANSI 300, 600

50 DS DS (KO) 1) KO 80 DS DS (KO) 1) KO

100 DS DS (KO) KO 150 DS DS (KO) KO 200 KO KO 250 KO GO GO 300 GO GO 400 GO GO 500 GO GO 600 GO GO

1) meters manufactured before November 2007 with DO Types in brackets are options

Specification for lubricating oils For lubrication purposes, we recommend that you use only Shell Tellus S2 MA 10 or another lubricating oil with 2 to 4°E at 25°C complying with MIL-L-6085-A in order to avoid damage to the shaft bearings. You can order your supply of lubricating oil from us in 1-liter containers under Ref. No. 82.11.148.00. Shelf life of oil: the shelf life depends on the operating conditions (e.g. UV light, humidity etc.). In principle the oil suffers no quality loss during the first 3-4 years. Initial lubrication When the turbine meters are delivered, the lubricators do not contain any oil. The oil pumps must not be filled with oil until they are at the installation site prior to commissioning. A small bottle of oil is supplied with each gas meter. For the initial lubrication more pump strokes are necessary than for the relubrications because first the oil pipes have to be filled with oil.

Pump DN 50 – DN 200 DN 250 – DN 600DO (oil gun) 20 strokes*) - KO (pushbutton pump) 40 strokes 45 strokes GO (lever pump) - 10 strokes

*) after venting the oil gun!

Page 45: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

OPERATION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

41

Relubrication

The relubrication intervals are stated on a indicating plate on the meter case. See also “Lubrication procedure”. Example:

In the case of unfavourable operating conditions, such as condensate forming because of water or hydrocarbon, or dust-laden gas, or service temperatures above 50°C, we recommend that lubrication be performed at shorter intervals, even daily under extreme conditions (permanent formation of condensate).

If the gas meter is operated under the above-mentioned conditions, it is likely that its service life will be reduced. If you have questions to the relubrication in such cases please contact RMG. Lubrication procedure

To fill the pumps and lubricate the turbine meters, proceed as follows: Oil gun

To fill with oil: 1. Unscrew cover. 2. Pull piston by means of the chain. 3. Fill with oil. 4. Reinsert piston by pushing it as far as

possible. 5. Screw on cover again tightly. 6. Vent the oil gun! To lubricate: 1. Plug oil gun onto oil nipple. 2. Depress oil gun towards oil nipple

once for each stroke.

Every 3 months 2 strokes

Relubrication! Every 3 months 2 strokes Lubricating oil: 2-4°E at 25°C conforming to MIL-L-6085-A see operating instructions

Cover

Piston

Page 46: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

OPERATION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

42

Pushbutton pump

To fill with oil: 1. Unscrew cover. 2. Fill with oil. 3. Screw on cover again. To lubricate: 1. Unscrew knurled cover (or hexagon

cover in the case of older designs). 2. Depress pushbutton, which is now

visible, once for each stroke. 3. Screw on knurled (or hexagon) cover

again. Every 3 months 6 strokes

Lever pump

To fill with oil: 1. Pull off cover upwards (in the case of

lever pumps with transparent container, unscrew cover).

2. Fill with oil. 3. Screw (or plug) on cover again tightly. To lubricate: 1. Move lever once for each stroke as far

as it will go. Up to DN 400: every 3 months 2 strokes From DN 500: every 3 months 3 strokes

Page 47: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

OPERATION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

43

Changing the battery For the battery version use only lithium batteries of the type SL 770/P Size C, 3.6 V produced by Sonnenschein or type LS 33600 produced by Saft.

These batteries have a service life of approx. 6 years and can be ordered ready for installation from RMG under the following order number:

For Ex and Non-Ex devices: 207626 Emergency battery for version with external supply: 207658

If you use other batteries than the replacement batteries from RMG, you run the

risk of the TRZ 03-TE / TEL not being approved for areas subject to explosion hazards any more.

In areas subject to explosion hazards, you must never insert or remove the battery. Remove the display board from the case and change the battery in an area without explosion hazards.

To change the battery, you must proceed as follows:

Remove the cover of the meter head – on the rear side, there is the battery fixed with cable ties. If the meter is located in an area subject to explosion hazards, disconnect the sensor wires from

terminal X5 (see chapter “Electrical connections” – Remember the terminals and colours!) and take the display board to an area without explosion hazards.

Remove the connecting wires of the battery from the contacts on the bottom of the terminal board.

Screw off the four screws with witch the battery board is fixed. Remove the cable ties with which the battery is fixed and take out the battery. Insert the new battery and fix it with cable ties. Plug the connecting wires of the battery on the contacts of the terminal board. Make sure that

the polarity is correct. Reinsert the battery-board into the case and screw it on. In case you have previously

disconnected the wires of the sensor and the control button, you must connect them again (make sure that the polarity of the sensor wires is correct!).

Close the case.

Page 48: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

OPERATION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

44

Maintenance instructions RMG turbine meters are maintenance-free except that they require lubrication and change of battery at regular intervals. Since all gas meters fitted with oil pumps are shipped with an empty oil storage tank, it is absolutely essential to fill the storage tank of the oil pump with oil and perform lubrication before the gas meter is put into service.

The operator should nevertheless check the turbine meter at regular intervals. In this connection, please refer also to DVGW code of practice G 495. (Gas pressure regulators for monitoring and servicing large-scale gas metering systems)

The gas meter should be checked for accuracy approximately every two years depending on whether it is possible or necessary to perform such checks. This can be done as follows: In the station itself by connecting two gas meters in series. In an officially acknowledged testing laboratory for gas measuring instruments. In the factory.

Labelling All important data which are necessary for operating the gas meter are stated on the plates which are affixed to the meter case, totalizer or pulse transmitters.

Further datapress button

EG-type examination certificate!

II2G EEx ib ia IIC T3/T4TÜV 02 ATEX1970

-20°C +60°C/40°C Ta

008524 VDC

IP65Germany

max

min

Main data plate of the TRZ 03-TE turbine meter

Page 49: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

OPERATION ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

45

0085Ser.-No.

Year

Further data see data plates on the remote totalizingunit.

For Ex devices see EC Type-Approval Certificate No. TÜV 02 ATEX 1970

Data plate of the TRZ 03-TE measuring device

Pulse OutputsTRZ03-TE/TEL

- +

imp NF

Alarm

imp HF

1 4

2 5

3 6

I

I

1

2

I3

1m =3

1m =3

Connectors assignment

Re - LubricationAt 3 month intervals, 2 strokes.Lubricant: 2-4°E at 25°Ccorresponding to MIL-L-6085-A.Supplementary informationsee operating instructions.

Indicating plate for relubrication of the main bearings

Re- LubricationAt 3 month intervals, 6 strokes.Lubricant: 2 - 4°E at 25°Cconforming to MIL-L-6085-A.The containervolume of 0,15 litre issufficient for 3 months. Supplementaryinformation see operating instructions.

Indicating plate for relubrication of the main bearings

Arrow indicating the direction of flow

pr Indication of the reference pressure tap

Page 50: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

SPECIFICATIONS ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

46

Specifications

Measuring range Under atmospheric pressure: 1:20 or 1:30

If the pressure increases, the measuring range of the meter extends. With a high pressure calibration measuring ranges of up to 1:50 can be reached.

Measuring error

For Qmin to 0.2 · Qmax: ±1% For 0.2 · Qmax to Qmax : ±0.5%

Reproducibility: ±0.1%

Temperature ranges Fluid temperature range: -10°C to +50°C

Ambient temperature range: TRZ 03-TE/TEL II2 G EEx ib IIC T4 -20°C to +40°C II2 G EEx ib IIC T3 -20°C to +60°C

with external supply II2 G EEx ib IIC T4 -20°C to +60°C

In the case of special designs, even higher or lower fluid temperatures are possible in non-Ex versions.

Inputs and outputs

Sensor input S1 (Measuring input)

(For Ex connected loads, see approval certificate)

Terminals: X5,1 (+) X5,2 (-) Sensor type: Wiegand sensor, direct Line length < 50 cm Jumpers X_S1 / 3-5 and 4-6

Wiegand sensor, remote totalizer Line length < 30 m Jumpers: X_S1 / 3-5 and 4-6 Jumpers: X_TERZ90 / 1-2

Reed contact Jumpers X_S1 / 1-3 and 2-4

Page 51: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

SPECIFICATIONS ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

47

Sensor input S2 (Comparative input or start/stop/reset)

(For Ex connected loads, see approval certificate) Terminals: X5,3 (+) X5,4 (-) Sensor type: Wiegand sensor, direct Line length < 50 cm Jumpers X_S2 / 3-5 and 4-6 Wiegand sensor, remote totalizer Line length < 50 m Jumpers: X_S2 / 3-5 and 4-6 Jumpers: X_TERZ90 / 3-4 Reed contact Jumpers X_S2 / 1-3 and 2-4 HF output

In the case of the explosion-protected (Ex) design, the device can only be connected to a certified intrinsically safe circuit. Terminals: X4,4 (+) X4,3 (-) Output: Open-drain transistor Tpulse: 1 ms ± 10% Fmax: 300 Hz Ex Non-Ex Umin: 4.0 V 4.0 V Umax: 28 V 30 V Imax: 60 mA* (total current) 400 mA External inductance 1 H External capacitance 25 μF

* maximum total current of the 3 pulse outputs (LF + HF + Alarm)

Page 52: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

SPECIFICATIONS ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

48

LF output

In the case of the explosion-protected (Ex) design, the device can only be connected to a certified intrinsically safe circuit. (For Ex connected loads, see approval certificate.) Terminals: X4,2 (+) X4,1 (-) Output: Open-drain transistor Tpulse: 125 ms ± 10% (Fmax: 4 Hz) 250 ms ± 10% (Fmax: 2 Hz) Ex Non-Ex Umin: 4.0 V 4.0 V Umax: 28 V 30 V Imax: 60 mA* (total current) 400 mA External inductance 1 H External capacitance 25 μF

* maximum total current of the 3 pulse outputs (LF + HF + Alarm) Alarm output

In the case of the explosion-protected (Ex) design, the device can only be connected to a certified intrinsically safe circuit. (For Ex connected loads, see approval certificate.) Terminals: X4,6 (+) X4,5 (-) Output: Open-drain transistor Ex Non-Ex Umin: 4.0 V 4.0 V Umax: 28 V 30 V Imax: 60 mA* (total current) 400 mA External inductance 1 H External capacitance 25 μF

* maximum total current of the 3 pulse outputs (LF + HF + Alarm)

Page 53: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

SPECIFICATIONS ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

49

Vo or RS-485 data interface

In the case of the explosion-protected (Ex) design, the device can only be connected to a certified intrinsically safe circuit. (For Ex connected loads, see approval certificate.) Vo data interface: Internal Connector (Binder) Terminals: X15,4 (+) 4 X15,3 (-) 1 X15,1 and X15,2 open Umin: 7.0 V Ui: 13.5 V Ii: 15 mA Pi: 210 mW External capacitance: 2,5 μF External inductance: 1 H Ex separating unit: KFD2-ST2-Ex1.LB (Pepperl+Fuchs) RS-485 data interface: Terminals: X15,4 (+ supply) X15,3 (- supply) X15,2 (line A) X15,1 (line B) Umin: 7.0 V Ui: 10.5 V Ii: 428 mA Pi: 900 mW Internal capacitance: 1,32 μF Internal inductance: 600 μH External capacitance: 23,7 μF External inductance: 1 H Ex separating unit: 17-21S1-S111 / EExi (Bartec) Note: If the EZD protocol or the RS485 bus is used, the device is supplied via the data interface.

Page 54: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

SPECIFICATIONS ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

50

Temperature input (hardware and software options)

For Ex connected loads, see approval certificate. Terminals: X9,2 (+) X9,1 (-) Sensor type: PT 1000 Measuring range: -20°C to 60°C Resolution: ± 0.2°C Current-loop connection (version with external supply only!)

Terminals: X22.1 (+) X22,2 (-) Uext (min): 12 V Uext (max): 28 V Imin: 3.5 mA Imax: 23 mA External load (max.): RLmax = (Uext - 10 V) / Imax (in ) e.g. Uext = 16 V RLmax = (16 V - 10 V) / 23 mA = 260 Current output with - minimum flow rate: 4 mA - maximum flow rate: 20 mA - Warning: 3.5 mA - Alarm: 23 mA Accuracy, current output: Better than 1% of the final value Data for the use in hazardous areas:

Ui = 28 V Ii = 110 mA Pi = 770 mW Ci = 2.2 nF Li = 110 μH

Materials used for the meter case Spheroidal cast iron or steel, depending on the pressure class and the nominal diameter.

Page 55: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

SPECIFICATIONS ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

51

Examples of connection

Battery-operated device

Hazardous area Safe area

Isolating amplifiers*KFD2-SOT-Ex2 (24 V/DC)

Pulse outputsLIYCY 2x2x0.75 mm blue2

24 V/DC

PE

3

6

3

1

4

1

7

9

7

11

11

12

12

PE

PE

8

10

8

Alarm

HF

LF

1112PE

Isolating amplifierKFD2-SOT-Ex1 (24 V/DC)

Alarm outputLIYCY 2x0.75 mm blue2

6

43

2

1

5

Pin assignment (Binder):

1= LF - 4= LF +2= Alarm - 5= Alarm +3= HF - 6= HF +

* If 2-channel isolating amplifiers are used, the two inputs must be separated!

Page 56: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

SPECIFICATIONS ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

52

Mains-powered device (Ex design)

(Connection only via the current module with built in backup battery)

Hazardous area Safe area

Isolating amplifiers*KFD2-SOT-Ex2 (24 V/DC)

Pulse outputsLIYCY 2x2x0.75 mm blue2

24 V/DC

PE

3

6

3

1

4

1

7

9

7

11

11

12

12

PE

PE

8

10

8

Alarm

HF

LF

1112PE

Isolating amplifierKFD2-SOT-Ex1 (24 V/DC)

Alarm outputLIYCY 2x0.75 mm blue2

Current outputLIYCY 2x0.75 mm blue

2

Supply unitKFD2-STC3-Ex1 (24 V/DC)

6

43

2

1

5

Pin assignment (Binder):

1= LF - 4= LF +2= Alarm - 5= Alarm +3= HF - 6= HF +

* If 2-channel isolating amplifiers are used, the two inputs must be separated!

Page 57: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

SPECIFICATIONS ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

53

Mains-powered device (Non-Ex design)

(Connection only via the current module with built in backup battery)

Safe area

Supply unitKFD2-CR-1.300 (or similar)

6

43

2

1

5

HF, LF and Alarm output:U : 24 V/DCMaximum sum of the currents (HF + LF + Alarm) < 400 mA

max

Alarm :

HF pulses:

LF pulses:

Pin assignment (Binder):

1= LF - 4= LF +2= Alarm - 5= Alarm +3= HF - 6= HF +

Page 58: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

SPECIFICATIONS ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

54

Overview of Parameters for the TRZ 03 TE / TEL

Parameter Setting Example (display)

A01 Main totalizer (Vm) 0 to 99999999 12345678 m³

A02 Flow rate (Qm) 12.3 m³/h

A03 Disturbing quantity totalizer 0 to 99999999 123 m³

A04 Start-stop totalizer 0 to 99999999 00000000 m³

A05 Resettable totalizer 0 to 99999999 00000000 m³

A06 Meter factor in all: six digits 12.3456 I/m³

A07 Pulse scaler 0.01; 0.1; 1; 10; 100 L 1 --

A08 Configuration TERZ operating mode 00011021 MOD

A09 Error Rolling display 0 ERR

B01 Serial number 1.004 VER

C01 Current output Main supply design 12.345 mA

C02 Flow rate at 4 mA 0 to 999999.99 20.000 --

C03 Flow rate at 20 mA 0 to 999999.99 650.000 --

C04 Calibration current 4.000 to 20.000 4.000 mA

C05 Current output mode 0 – Default value 1 – Flow rate Qm

3 MOD

C06 Correction factor - Offset 0 to 99999999 1234.5678 --

C07 Correction factor - Rise 0 to 99999999 1234.5678 --

C08 Damping display / current output

0 to 0.9 0.4 --

Z06 Frequency display 651 Hz

Page 59: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

SPECIFICATIONS ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

55

Measuring ranges/Dimensions/Pressure classes of TRZ 03-TE

DN Size Measuring range Ua DN Pressure classes / Approx. weight Dimensions

mm Qmin-Qmax m3/h mm

in 1:10/1:20 1:30 in PN kg ANSI kg L H C

50 2"

G 40 G 65

13-65* 10-100

0,1 0,1

50 2"

10/16 13 150 13 - 25/40 21 300 13 150 210 60- 64/100 21 600 21

80 3"

G 100 16-160 - 1 80 3"

10/16 20 150 20 G 160 13-250 - 1 25/40 25 300 25 240 230 96G 250 20-400 - 1 64/100 34 600 36

100 4"

G 160 13-250 - 1 1004"

10/16 25 150 30 240 G 250 20-400 - 1 25/40 32 300 35 300 260 120G 400 32-650 20-650 1 64/100 45 600 55 270

150 6"

G 400 32-650 - 1 1506"

10/16 50 150 50 265 G 650 50-1000 32-1000 1 25/40 60 300 65 450 265 180

G 1000 80-1600 50-1600 10 64/100 70/90 600 100 285

200 8"

G 1000 G 1600

80-1600 130-2500

50-1600 80-2500

10 10

2008"

10/16 75 150 100 300 25/40 95 300 120 600 320 240

64/100 150/160 600 160 320

250 10"

G 1000 80-1600 - 10 25010"

10/16 100/110 150 110 G 1600 130-2500 80-2500 10 25/40 135/150 300 160 750 330 300G 2500 200-4000 130-4000 10 64/100 180/225 600 260

300 12"

G 2500 G 4000

200-4000 320-6500

130-4000 200-6500

10 10

30012"

10/16 138/150 150 155 25/40 225/265 300 230 900 360 360

64/100 275/290 600 310

400 16"

G 4000 G 6500

320- 6500 500-10000

200-6500 320-10000

10 10

40016"

10/16 200/290 150 350 25/40 350/440 300 460 1200 400 480

64/100 525/580 600 575

500 20"

G 6500 G 10000

500-10000 800-16000

320-10000 500-16000

10 100

50020"

10/16 560/610 150 620 25/40 640/700 300 650 1500 450 600

64/100 830/1060 600 1075

600 24"

G 10000 G 16000

800-16000 1300-25000

500-16000 800-25000

100100

60024"

10/16 900/940 150 950 25/40 980/1075 300 1000 1800 500 720

64/100 1230/1570 600 1600

* (Measuring range 1:5)

PN pressure classes complying with DIN 2401, part 1, ANSI pressure classes complying with B16.5. From DN 80 to DN 300, the turbine meters of the PN 10/16 pressure class can be fitted with a thermowell to accommodate a resistance thermometer.

L

H

DN

C

Page 60: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

SPECIFICATIONS ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

56

Measuring ranges/Dimensions/Pressure classes of TRZ 03-TEL

DN Size Measuring range Ua DN Pressure classes / Approx. weight Dimensions

mm Qmin-Qmax m3/h mm

in 1:10/1:20 1:30 in L L L kg L H C

50 2"

G 40 G 65

13-65* 10-100

0.1 0.1

50 2"

10/16 13 150 13 235 - 25/40 21 300 13 150 235 28- 64/100 21 600 21 250

80 3"

G 100 16-160 - 1 80 3"

10/16 20 150 20 260 G 160 13-250 - 1 25/40 25 300 25 240 260 40G 250 20-400 - 1 64/100 34 600 36 270

100 4"

G 160 13-250 - 1 1004"

10/16 25 150 30 270 G 250 20-400 - 1 25/40 32 300 35 300 275 50G 400 32-650 20-650 1 64/100 45 600 55 290

150 6"

G 400 32-650 - 1 1506"

10/16 50 150 50 290 G 650 50-1000 32-1000 1 25/40 60 300 65 450 300 80

G 1000 80-1600 50-1600 10 64/100 70/90 600 100 325

200 8"

G 1000 G 1600

80-1600 130-2500

50-1600 80-2500

10 10

2008"

10/16 75 150 100 25/40 95 300 120 600 290 120

64/100 150/160 600 160

250 10"

G 1000 80-1600 - 10 25010"

10/16 100/110 150 110 G 1600 130-2500 80-2500 10 25/40 135/150 300 160 750 330 165G 2500 200-4000 130-4000 10 64/100 180/225 600 260

300 12"

G 2500 G 4000

200-4000 320-6500

130-4000 200-6500

10 10

30012"

10/16 140/155 150 160 25/40 230/270 300 235 900 360 200

64/100 280/295 600 315

400 16"

G 4000 G 6500

320- 6500 500-10000

200-6500 320-10000

10 10

40016"

10/16 290/300 150 360 25/40 360/450 300 470 1200 400 300

64/100 535/590 600 585

500 20"

G 6500 G 10000

500-10000 800-16000

320-10000 500-16000

10 100

50020"

10/16 575/625 150 635 25/40 655/715 300 665 1500 450 385

64/100 845/1075 600 1090

600 24"

G 10000 G 16000

800-16000 1300-25000

500-16000 800-25000

100100

60024"

10/16 925/965 150 975 25/40 1000/1100 300 1025 1800 500 480

64/100 1250/1590 600 1625

* (Measuring range 1:5)

PN pressure classes complying with DIN 2401, part 1, ANSI pressure classes complying with B16.5. From DN 80 to DN 300, the turbine meters of the PN 10/16 pressure class can be fitted with a thermowell to accommodate a resistance thermometer.

L

H

DN

C

Page 61: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

SPECIFICATIONS ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

57

Overview of the materials used

Designation Material

Meter case GGG40, cast steel or welded steel Flow straightener Delrin, aluminium or steel Turbine wheel Delrin or aluminium Measuring element case Aluminium Measuring element bearings Aluminium and/or stainless steel Ball bearings Stainless steel Shafts Stainless steel Set screw Steel Permanent magnet Oerstit 500 Sensor sleeve with sensor Nirosta steel O-rings 83FKM592

Page 62: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

ANNEX ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

58

Annex

Matrix of the TRZ 03-TE / TEL Coordinate

Specification

Terz94 TRZ03-TE/TEL(Batt.)

Terz94-S TRZ03-TE/TEL(main supply)

1 Channel 2 Channel 1 Channel 2 Channel

A01 Main Totalizer Vm X X X XA02 Flow rate (Qm) X X X XA03 Disturbing quantity totalizer NO X NO XA04 Start / Stop totalizer X NO X NO A05 Resettable totalizer X NO X NO A06 Meter factor X X X XA07 Pulse scaler X X X XA08 Configuration X X X XA09 Error X X X XA10 Operating mode Pulse X from X:Y NO X NO XA11 Operating mode Pulse Y from X:Y NO X NO X B01 Serial number X X X X C01 Current output NO NO X XC02 Flow rate at 4 mA NO NO X XC03 Flow rate at 20 mA NO NO X XC04 Calibration current NO NO X XC05 Current output mode NO NO X XC06 Correction factor – offset NO NO X XC07 Correction factor – Rise NO NO X XC08 Damping display / current output NO NO X X Z01 Number of digit positions following a

decimal pointNO NO NO NO

Z02 Factor NO NO NO NO Z03 Divisor NO NO NO NO Z04 Direction of rotation for ENCO NO NO NO NO Z05 Totalizer NO NO NO NO Z06 Frequency X X X XZ07 Maximum flow value 0 0 0 0Z08 Frequency-mode 0 0 0 0Z09 Error state 0 0 0 0Z10 Number of pulses X X X XZ11 Version number X X X X

Not available NO Available XAvailable in service mode 0

In case only a part of the matrix is displayed. It any depends on the operating mode, which fields are displayed. Fields, which have no function in the selected operating mode are blanked out.

Page 63: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

ANNEX ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

59

TRZ 03-TE/TEL button assignments

TRZ 03-TE/TEL 94 Display Mode

12345678 m3

P P

Scrolling in the matrix (upwards)

Scrolling in the matrix (downwards)

> 2 sec.: switch to the programming mode< 2 sec.: switch to the coordinate

Button at the rear of the board

< 2 sec.: change of the unit in the coordinateafterwards scrolling in the matrix (upwards

> 2 sec.: segment test and switching to Vm

TRZ 03-TE/TEL Programming Mode

12345678 m3

PP

Button at the rear of the board

Flashing digit +1

Flashing digit -1

Set decimal point / reset (if float)

Abortion of programming

Move to the next digit

Page 64: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

ANNEX ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

60

Converting the TRZ 03-TE/TEL into the version with external supply and current output The battery powered version can be converted into the mains powered version by installing another board. Remove the battery (see chapter “Changing the battery”). Remove the jumper for battery-powered mode (1-2) on the socket X2_0 of the board. Srew the 4 short stay bolts onto the present stay bolts on the board. Plug the current module TERZ94trm with a 12-pin connector (X2_1) into the socket X2_0. Apply the supply voltage (24V) to terminal X22 (1 +, 2 - ). Switch on the power supply (24 V). Check the current output:

Set the current output in channel A08 to I-Out (mode 1 or 2). Channel C02: enter flow rate at 4 mA. Channel C03: enter flow rate at 20 mA. Set channel C04 (default value) to 10.0 mA. Set channel C05 (output mode) to 1 (current output is proportional to the flow rate). Channel C06: enter correction value for the current offset (see value on the board) Channel C07: enter correction value for the current rise (see value on the board) In the event that a recalibration should be necessary, there is a PC-based program available

which will define the new offset and rise values. Now the mains powered version is operational.

Page 65: TRZ03TE he hw1 - RMG Messtechnik GmbH...All RMG turbine meters have passed the disturbance measurements in compliance with OIML Recommendation IR-32/89, Annex A, with slight and heavy

ANNEX ............................................................................................................................... ................................................................................

............................................................................................................................... ................................................................................ Manual TRZ 03-TE · EN06 · 2013-03

61

Cable type abbreviations Basic type: Control cable (twisted cable)

Example

Li Y C Y 2x 0.75 blue

Sheath colour

Conductor cross section / mm2

Number of cores

Sheath material Thermoplastic resins

Special features Screen braiding

Insulation material Thermoplastic resins

Cable designation Stranded core