Catalog Number 52004-88 Start Up Instructions for 1720 Series Turbidimeters and the AquaTrend ® Network System © Hach Company, 1997–2000, 2002. All rights reserved. Printed in the U.S.A. dd/dp 03/02 3ed

Upload: moises-fernandes-goncalves

Post on 28-Nov-2014

193 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Turbidimetro 1720 Aqua Tratada

Catalog Number 52004-88

Start Up

Instructions for 1720 Series Turbidimetersand the AquaTrend® Network System

© Hach Company, 1997–2000, 2002. All rights reserved. Printed in the U.S.A. dd/dp 03/02 3ed

Page 2: Turbidimetro 1720 Aqua Tratada

Table of Contents

Section 1 Basic System Information ..................................................................................................................... 51.1 How To Use This Manual .................................................................................................................................... 51.2 System Components........................................................................................................................................... 6

Section 2 Mounting the Devices ............................................................................................................................ 92.1 Mounting the Turbidimeter................................................................................................................................... 92.2 Mounting the Power Supply, AquaTrend®, and Signal Output Module (SOM) .................................................. 10

2.2.1 Power Supply, AquaTrend, and SOM Pole Mounting............................................................................... 102.2.2 Power Supply, AquaTrend, and SOM Wall Mounting............................................................................... 132.2.3 Power Supply, AquaTrend, and SOM Panel Mounting............................................................................. 14

2.3 Mounting Devices on the 1720 Series Turbidimeter Stand ............................................................................... 152.3.1 1720 Series Turbidimeter Stand Mount ................................................................................................... 162.3.2 AquaTrend, SOM, or PS1201 Power Supply Stand Mount...................................................................... 162.3.3 Portable AquaTrend Stand Mount............................................................................................................ 162.3.4 Digital Display Module Stand Mount........................................................................................................ 16

Section 3 Wiring the Devices ............................................................................................................................... 193.1 Network Wiring.................................................................................................................................................. 19

3.1.1 1720D Series Turbidimeter Connections to PS1201 Power Supply ........................................................ 193.1.2 1720D/L Series Turbidimeter Connections to PS1201 Power Supply ..................................................... 223.1.3 AquaTrend with Signal Output Module (SOM) to PS1201 Power Supply Connections ........................... 243.1.4 AquaTrend without SOM to PS1201 Power Supply Connection.............................................................. 263.1.5 SOM to Power Supply Connections ........................................................................................................ 29

3.2 Power Supply to Power Supply Network Interconnect ...................................................................................... 323.3 Other Device Wiring .......................................................................................................................................... 333.4 Network Termination ......................................................................................................................................... 34

3.4.1 Network Termination at the AquaTrend/SOM or the SOM....................................................................... 343.4.2 Network Termination at the AquaTrend without SOM (Model 51200)...................................................... 35

Section 4 Plumbing the Turbidimeter.................................................................................................................. 374.1 Hydraulic Connections ...................................................................................................................................... 374.2 Starting Sample Flow........................................................................................................................................ 37

Section 5 Power and Startup of the System ....................................................................................................... 395.1 AC Input Power ................................................................................................................................................. 395.2 About the AquaTrend ........................................................................................................................................ 395.3 Identifying the Master AquaTrend ..................................................................................................................... 40

Section 6 Setting Up the Network........................................................................................................................ 416.1 Setting the Time and Date ................................................................................................................................ 41

6.1.1 Setting the Time ...................................................................................................................................... 416.1.2 Setting the Date....................................................................................................................................... 42

6.2 Network Setup .................................................................................................................................................. 426.2.1 Installing a Single Sensor System ........................................................................................................... 436.2.2 Installing Multiple Sensors or Devices ..................................................................................................... 436.2.3 Attaching a Measurement to a Channel .................................................................................................. 45

6.3 Display Setup .................................................................................................................................................... 466.4 Attaching the SOM............................................................................................................................................ 466.5 Attaching the SIO.............................................................................................................................................. 46

Page 2Table of Contents 52004-88TOC.fm

Page 3: Turbidimetro 1720 Aqua Tratada

Table of Contents

Section 7 Adding a Digital Display Module ........................................................................................................ 477.1 Attached Cordset Connection ........................................................................................................................... 477.2 Network Setup .................................................................................................................................................. 477.3 Adding the Digital Display Module to the Network ............................................................................................ 48

7.3.1 Assigning the Digital Display Module to a Sensor ................................................................................... 48

Section 8 Using the Portable AquaTrend............................................................................................................ 498.1 The Portable AquaTrend ................................................................................................................................... 498.2 Mounting the Portable AquaTrend..................................................................................................................... 50

Section 9 Adding a Serial Input/output Module ................................................................................................. 539.1 Network Wiring Connections............................................................................................................................. 539.2 Power Connections ........................................................................................................................................... 539.3 Adding the SIO to the Network ......................................................................................................................... 53

General Information .............................................................................................................................................. 55How to Order........................................................................................................................................................... 57Repair Service ........................................................................................................................................................ 58Limited Warranty..................................................................................................................................................... 59

Page 352004-88TOC.fm Table of Contents

Page 4: Turbidimetro 1720 Aqua Tratada

Visit http: //www.hach.com

Page 5: Turbidimetro 1720 Aqua Tratada

Section 1 Basic System Information

An AquaTrend® Network System uses a single AquaTrend Module to interface a variety of instruments. The network is built by wiring a sensor, power supply, and any other peripheral device with LonWorks® digital Network capabilities or 4–20 mA network capability using the signal Input Module and establishing communications via the AquaTrend Interface. Power and network connections are required for all AquaTrend devices operating on the network.

After physically installing the devices, you must add each device to the AquaTrend network via the network menus in the Master AquaTrend. After adding all devices to the network, you must designate which devices communicate with one another. For example, once a sensor is attached to an AquaTrend, you must instruct the sensor to attach its reading to a designated AquaTrend display channel.

Because many underlying network connections occur when performing network operations, it is important to have all existing devices on the network powered and operating normally when network operations are conducted. Careful attention to network management operations and the interconnection of devices will ensure successful operation of the AquaTrend Network System.

1.1 How To Use This ManualThis manual is intended to be used as a quick-start guide to efficiently get your system up and running. The instructions for installing the basic components of a single or two sensor system are presented in the main body of the manual with the instructions for optional accessory devices presented in the appendices. It is written to be followed sequentially, enabling all the tasks to be performed in the most efficient order.

Please take the time to familiarize yourself with the manuals supplied with the individual devices that comprise your system. Important safety information, additional detail on the tasks outlined in this manual, and other more advanced features are presented in the specific device manuals. Figure 1 illustrates the major steps for installing the system. Figure 2 shows the components of a basic Single Sensor System.

Figure 1 Installing the System

Page 552004-88 basic system information.fm Basic System Information

Page 6: Turbidimetro 1720 Aqua Tratada

Section 1

Figure 2 Single Sensor System Example

1.2 System Components

AquaTrend

• Master AquaTrend — Acts as the main network data control center for setup and operation of the network. The network can contain as many as nine AquaTrends but only one can be the Master AquaTrend. Only through the Master AquaTrend can network operations be performed.

• Remote AquaTrend — Any AquaTrend on a network that is not a Portable or Master AquaTrend. Remote AquaTrends can control and monitor all sensors on the network. Remote AquaTrends cannot perform network operations.

• Portable AquaTrend — Mobile device designed for set up and calibration of sensors already on the network. Portable AquaTrends can communicate with one sensor at a time and cannot perform network operations.

Power Supply 12V DC

Network Connection Module

ON

OFF

EXIT

ENTERMENU

DISPLAY

AquaTrend™/SOM

COM

NONCCOM

NONC

}}

52010-60(2 relays and 2 analog outputs)

Analog Output #2

PS1201

Analog Output #1

AquaTrend with SOM 51350-60

115 / 230 VACPower In

1720 Series

20-foot Cable(Attached to 1720D Turbidimeter)

Turbidimeter 52000-60

Customer SuppliedON/OFF Switch

Spare -open for

expansion(Typical)

Relay #1

Relay #2

Page 6System Components 52004-88 basic system information.fm

Page 7: Turbidimetro 1720 Aqua Tratada

Section 1

PS1201 — A power supply that provides power and network communication to network devices (i.e., 1720 Series Turbidimeter, AquaTrend, Signal Output Module, etc.).

Sensor — Any one of the Hach devices designed to take measurements for specific parameters and used with the AquaTrend Network System (e.g., the 1720D Turbidimeter).

Serial I/O Module (SIO) — Communication device for two-way communication with a PC or one-way communication with a printer.

Signal Output Module (SOM) — Network device that provides two relays and two analog outputs. The SOM can be purchased as a stand-alone device (Model 51250) or as an integral part of the AquaTrend/SOM (Model 51350).

Digital Display Module (DDM) — An eight-character display that can be used with sensors or analyzers to display a single measurement.

Repeater Board — Device used to extend the length of the network.

Signal Input Module — This module provides two 4–20mA analog inputs which can be used to bring measurements from non-networked sensors into the network.

Page 752004-88 basic system information.fm System Components

Page 8: Turbidimetro 1720 Aqua Tratada

Visit http: //www.hach.com

Page 9: Turbidimetro 1720 Aqua Tratada

Section 2 Mounting the Devices

2.1 Mounting the TurbidimeterThe turbidimeter body is designed for wall-mounting (although it may be mounted on the optional floor stand). The turbidimeter sensor must be mounted within 20 feet of the power supply. Refer to Figure 3 for specific mounting dimensions.

Determine the best mounting location using the information given below:

• Locate the turbidimeter as close to the sampling point as possible. Slotted mounting brackets are integral parts of the turbidimeter body. Hardware to mount the body is not supplied with the instrument.

• Allow at least 22 cm (approximately 10 inches) clearance for removal of the head assembly and bubble trap cover from the top of the turbidimeter body.

• Leave enough room below the turbidimeter body to remove the bottom plug and to place a container under the drain when calibrating or cleaning the turbidimeter.

Refer to Figure 4 and install the turbidimeter body as follows:

Note: Make sure the top of the turbidimeter body is level.

Install two ¼-20 bolts 10-¾ inches apart (on center). Leave at least ¼ inch of the bolt head exposed. Make sure the bolts are installed level. Hang the slotted mounting brackets of the turbidimeter body onto the bolts.

Figure 3 Turbidimeter Dimensions

Page 952004-88 mounting the devices.fm Mounting the Devices

Page 10: Turbidimetro 1720 Aqua Tratada

Section 2

Figure 4 Turbidimeter Body Installation

2.2 Mounting the Power Supply, AquaTrend®, and Signal Output Module (SOM)

These devices were designed to mount easily to a wall, pole, or panel. Allow at least one inch clearance from other objects on the top and sides and a minimum of six inches on the bottom. Disassembly is not required for pole or wall mounting, and minimal disassembly is required for panel mounting. Remove all necessary knockouts from the modules before mounting them.

A mounting bracket (see Figure 9 on page 13) for securing the devices to a horizontal or vertical pole (or similar structure) or to a wall is supplied. Customer supplied U-bolts provide a channel for a ¾- to 2-inch pole (outside diameter). The pole must be capable of supporting at least five pounds.

2.2.1 Power Supply, AquaTrend, and SOM Pole Mounting

1. Obtain two ¼-20 x 2.843 U-bolts and nuts.

2. Remove all knockouts necessary for electrical and network wiring before mounting the modules.

3. Press the back of the mounting bracket against a pole in the proper mounting position, either vertically or horizontally.

4. Place the curve of one U-bolt behind the pole, and insert the threaded ends of the bolt through two of the center holes on the mounting bracket. See Figure 6.

5. Secure the two nuts with a wrench, and repeat Steps 1-3 on the opposite end of the bracket.

Page 10Mounting the Power Supply, AquaTrend®, and Signal Output Module (SOM) 52004-88 mounting the devices.fm

Page 11: Turbidimetro 1720 Aqua Tratada

Section 2

Figure 5 Power Supply, AquaTrend and SOM Dimensions

Figure 6 Pole Mounting the Power Supply, AquaTrend, and SOM

Page 1152004-88 mounting the devices.fm Mounting the Power Supply, AquaTrend®, and Signal Output Module (SOM)

Page 12: Turbidimetro 1720 Aqua Tratada

Section 2

Figure 7 Portable AquaTrend Dimensions

Figure 8 Portable AquaTrend Bracket Dimensions

1.99"(50.49 mm)

1.86"(47.25 mm)

2.03" (51.55 mm)2.14" (54.38 mm)

.50" (12.70 mm)5.20"(132.08 mm)

2.14"(54.38 mm)

2.40"(60.85 mm)

6.20"(157.48 mm)

3.32"(84.42 mm)

For vertical pipe rangingfrom .75 (19.05 mm) ODto 2.00 (50.80 mm) OD

For horizontal pipe rangingfrom .75 (19.05 mm) ODto 2.00 (50.80 mm) OD

Page 12Mounting the Power Supply, AquaTrend®, and Signal Output Module (SOM) 52004-88 mounting the devices.fm

Page 13: Turbidimetro 1720 Aqua Tratada

Section 2

2.2.2 Power Supply, AquaTrend, and SOM Wall Mounting

1. Remove all knockouts necessary for electrical and network wiring before mounting the modules.

2. Use a Phillips screwdriver to remove the four screws holding the mounting bracket onto the back of the device.

3. Use the bracket as a template to mark positions on the wall for four of the eight screw holes — either the vertical or horizontal pair (see Figure 9).

4. Reconnect the mounting bracket to the device.

5. Drill pilot holes in the marked positions for four ¼" screws.

6. Use four ¼" hex-head or pan-head screws to secure the bracket and device to the wall.

Figure 9 Mounting Bracket Dimensions for the Power Supply, AquaTrend, and SOM

Page 1352004-88 mounting the devices.fm Mounting the Power Supply, AquaTrend®, and Signal Output Module (SOM)

Page 14: Turbidimetro 1720 Aqua Tratada

Section 2

2.2.3 Power Supply, AquaTrend, and SOM Panel Mounting

1. Remove all knockouts necessary for electrical and network wiring before mounting the modules.

2. Turn off all power to the device and disconnect all cables if necessary.

3. Use a Phillips screwdriver to remove the four screws holding the mounting frame onto the back of the device. The wall mount frame is not needed for panel mounting.

4. Use a flat-blade screwdriver to loosen the screws of the device face.

5. Lift the face off the device and set it aside. Or, store it by screwing the right side of the face to the left side of the back enclosure. If installing the AquaTrend with SOM, unplug the ribbon cable from the circuit board in the back enclosure.

6. Use a punch and hammer to remove the three knock-outs on the inside rim of the bottom enclosure. Center the punch on the knockout for best results.

7. Measure the panel hole using the dimension information given in Figure 9. Mark the panel for placement of three #10 pan-head screws.Do not install a screw on the high-voltage side of the power supply.

8. Cut out the panel hole and drill three pilot holes for the screws (see Figure 10).

9. Fit the device into the panel hole and secure it with three #10 pan-head screws.

10. If power and network connections are to be made, refer to the appropriate section in this manual or in the individual device manual. If no other connections need to be made, proceed with Step 11.

11. Ensure that the gasket is properly seated in the device face and tighten the screws to 10 in-lb (1.1 n-m) maximum. Do not overtighten the screws.

Page 14Mounting the Power Supply, AquaTrend®, and Signal Output Module (SOM) 52004-88 mounting the devices.fm

Page 15: Turbidimetro 1720 Aqua Tratada

Section 2

Figure 10 Panel Mounting the Power Supply, AquaTrend, and SOM

2.3 Mounting Devices on the 1720 Series Turbidimeter StandThe 1720 Series Turbidimeter Stand shown in Figure 11 is designed for mounting a 1720 Series Turbidimeter, an AquaTrend or AquaTrend with SOM, and a PS1201 Power Supply. A Digital Display Module (DDM) may also be mounted if desired. The stand is assembled when shipped and hardware for mounting the DDM and turbidimeter is included. Mounting U-bolts must be supplied by the customer. Other devices may be added and must be mounted using the pole mounting instructions in the device manual (refer to the Expanded System example shown in Figure 12). Remove all knockouts necessary for electrical and network wiring before mounting the modules on the stand.

Page 1552004-88 mounting the devices.fm Mounting Devices on the 1720 Series Turbidimeter Stand

Page 16: Turbidimetro 1720 Aqua Tratada

Section 2

Figure 11 1720 Series Turbidimeter Stand

2.3.1 1720 Series Turbidimeter Stand MountMount the 1720D on the lower cross support of the stand. Slide the slots on the rear of the turbidimeter onto the two screws on the lower cross support of the stand. Tighten the screws to secure the turbidimeter to the cross support (see Figure 11).

2.3.2 AquaTrend, SOM, or PS1201 Power Supply Stand MountMount the AquaTrend or SOM on the same side of the stand as the turbidimeter, leaving enough room below the AquaTrend for easy access to the wiring knock-outs. Mount the PS1201 Power Supply on the opposite side of the stand, facing away from the turbidimeter (see Figure 11).

Obtain two ¼-20 x 2.843 U-bolts, washers, and nuts. Press the rounded portion of the U-bolt against the pole with the threaded ends through the mounting holes in mounting plate. Tighten the nuts to secure the module to the pole (see Figure 11 and Figure 12).

2.3.3 Portable AquaTrend Stand MountThe Portable AquaTrend may be permanently mounted to the stand or to a wall or pole using the mounting bracket supplied with the module. Detailed instructions are given in Section 8.2 on page 50 or refer to the AquaTrend Interface Manual.

2.3.4 Digital Display Module Stand MountMount the DDM to the stand at the top cross support. Remove the screws, insert them through the mounting holes in the DDM, and tighten until secure (see Figure 12).

Page 16Mounting Devices on the 1720 Series Turbidimeter Stand 52004-88 mounting the devices.fm

Page 17: Turbidimetro 1720 Aqua Tratada

Section 2

Figure 12 Basic and Expanded System Mounting

Page 1752004-88 mounting the devices.fm Mounting Devices on the 1720 Series Turbidimeter Stand

Page 18: Turbidimetro 1720 Aqua Tratada

Visit http: //www.hach.com

Page 19: Turbidimetro 1720 Aqua Tratada

Section 3 Wiring the Devices

Figure 13 shows the available wiring knock-outs and/or through-holes on the bottom of the power supply module. The other modules in the system have similarly positioned knock-outs for power and network connections. For easiest installation, determine which knock-outs will be used and if necessary, use a punch and hammer to remove them before mounting the module.

Figure 13 PS1201 Power Supply Knock-outs with Conduit Fittings

3.1 Network WiringNetwork wiring for sensors and modules is made at one of the terminal strips of the PS1201 Power Supply (shown in Figure 14). Connect only one type or color of wire to each of the six terminal positions as shown in the illustration.

Note: The module is shipped with four internal wires attached to the PS1201 terminal block. These wire colors may not correspond to the Hach supplied two- and four-conductor cable wire colors designated in Figure 20.

3.1.1 1720D Series Turbidimeter Connections to PS1201 Power Supply

Note: The cable is already wired to the turbidimeter head. Wire the unterminated end of the cable to the PS1201 Power Supply.

1. Turn off all power to the PS1201 Power Supply and disconnect all cables if necessary.

2. Unscrew the four screws securing the PS1201 face using a flat-bladed screwdriver.

3. Lift the face off the PS1201 and set it aside, or, store the front cover by screwing the right side of the front cover to the left side of the back enclosure.

Page 1952004-88 wiring the devices.fm Wiring the Devices

Page 20: Turbidimetro 1720 Aqua Tratada

Section 3

Figure 14 PS1201 Terminal Connection Positions and Corresponding Wire Colors

4. If necessary, use a punch and hammer from the outside of the enclosure to remove one of the left three unused knock-outs on the bottom of the Power Supply enclosure. Center the punch on the knock-out for best results.

Note: To assure the PS1201 continues to meet NEMA 4X, I.P.66 enclosure ratings, use only the supplied NEMA-style enclosure strain relief fittings.

5. Remove the washer from the strain relief on the cable attached to the turbidimeter head. Remove the fitting and route the cable through the hole created in Step 4.

6. Strip ¼-inch of insulation from each wire.

7. Insert each wire into the correct terminal block position until the wire insulation is seated against the terminal block. Tighten the screw and gently tug on each wire to ensure a solid connection. Refer to Figure 15 and Table 1 for specific information.

8. Tighten the strain relief fitting using the supplied washer and locknut until the cable no longer slides in the strain relief.

9. If other power or network connections are to be made at this time, refer to the appropriate section in this manual or the individual device manual. If no other connections need to be made, continue with the remainder of this procedure.

Page 20Network Wiring 52004-88 wiring the devices.fm

Page 21: Turbidimetro 1720 Aqua Tratada

Section 3

10. Ensure that the gasket is properly seated in the PS1201 face and tighten the screws to 10 in-lb (1.1 n-m) maximum. Do not overtighten the screws.

Figure 15 1720D Connections to PS1201 Power Supply

Table 1 1720D to PS1201 Wiring Connections

Terminal Position Wire Color

Pin 1 Red

Pin 2 Orange

Pin 3 Black

Pin 4 Shield/Clear

Pin 5 White

Pin 6 Green

Page 2152004-88 wiring the devices.fm Network Wiring

Page 22: Turbidimetro 1720 Aqua Tratada

Section 3

3.1.2 1720D/L Series Turbidimeter Connections to PS1201 Power Supply

Note: The cable is already wired to the turbidimeter head. Wire the unterminated end of the cable to the PS1201 Power Supply.

1. Turn off all power to the PS1201 Power Supply and disconnect all cables if necessary.

2. Unscrew the four screws securing the PS1201 face using a flat-bladed screwdriver.

3. Lift the face off the PS1201 and set it aside, or, store the front cover by screwing the right side of the front cover to the left side of the back enclosure.

4. If necessary, use a punch and hammer from the outside of the enclosure to remove one of the left three unused knock-outs on the bottom of the Power Supply enclosure. Center the punch on the knock-out for best results.

Note: To assure the PS1201 continues to meet NEMA 4X, I.P.66 enclosure ratings, use only the supplied NEMA-style enclosure strain relief fittings.

5. Remove the lock nut from the strain relief on the cable attached to the turbidimeter head. Remove the fitting and route the cable through the hole created in Step 4.

6. Strip ¼-inch of insulation from each wire if necessary.

7. Insert each wire into the correct position of the terminal block until the wire insulation is seated against the terminal block. Tighten the screw and gently tug on each wire to ensure a solid connection. Refer to Figure 16 and Table 2 for specific information.

8. Tighten the strain relief fitting using the supplied washer and locknut until the cable no longer slides in the strain relief.

9. If other power or network connections are to be made at this time, refer to the appropriate section in this manual or the individual device manual. If no other connections need to be made, continue with the remainder of this procedure.

10. Ensure that the gasket is properly seated in the PS1201 face and tighten the screws to 10 in-lb (1.1 n-m) maximum. Do not overtighten the screws.

Page 22Network Wiring 52004-88 wiring the devices.fm

Page 23: Turbidimetro 1720 Aqua Tratada

Section 3

Figure 16 1720D/L Connections to PS1201 Power Supply

Table 2 1720D/L Connections to the PS1201 Power Supply

Terminal Position Wire Color

Pin 1 Red

Pin 2 No Connection

Pin 3 Black

Pin 4 Shield/Clear

Pin 5 White

Pin 6 Green

Page 2352004-88 wiring the devices.fm Network Wiring

Page 24: Turbidimetro 1720 Aqua Tratada

Section 3

3.1.3 AquaTrend with Signal Output Module (SOM) to PS1201 Power Supply Connections

Order enough cable (Cat. No. 52158) from Hach to suit the specific installation parameters. Connections are made at both ends of the cable; at the AquaTrend using the connection information in the first two columns of Table 3, and at the PS1201 Power Supply using the information presented in the third and fourth column of Table 3.

Note: The maximum cable length from the PS1201 to the most distant device is 100 ft. (30 m).

Make the connections at the PS1201 end as follows:

1. Turn off all power to the PS1201 Power Supply and disconnect all cables if necessary.

2. Unscrew the four screws securing the PS1201 face using a flat-bladed screwdriver.

3. Lift the face off the PS1201 and set it aside, or, store the front cover by screwing the right side of the front cover to the left side of the back enclosure.

4. If necessary, carefully use a punch and hammer from the outside of the enclosure to remove one of the left three unused knock-outs on the bottom of the PS1201 enclosure. Center the punch on the knock-out for best results.

5. Insert the cable from the AquaTrend through the strain relief fitting provided in the hole described in Step 4, above. Make sure the outer cable insulation is just captured by the strain relief fitting.

6. Strip six inches (15 cm) of outer insulation from the cable (Cat. No. 52158).

7. Strip ¼ inch of insulation from each wire and insulate the shield wire with tubing (Cat. No. 40825-00).

8. Connect the cable wires to the six-position terminal block of the PS1201 (Cat. No. 52143-00) in the positions shown in Figure 17.

9. Tighten the strain relief fitting using the supplied washer and locknut until the cable no longer slides in the strain relief.

Note: To ensure the PS1201 continues to meet NEMA 4X, I.P.66 enclosure ratings, use only the supplied NEMA-style enclosure strain relief fittings.

10. Make connections at the Analog/Recorder Output Terminal Block and the Relay Terminal Block (see Figure 17) using the information supplied in Table 5 and Table 6 on page 30.

11. If other power or network connections are to be made at this time, refer to the appropriate section in this manual or the individual device manual. If no other connections need to be made, continue with the remainder of this procedure.

12. Ensure that the gasket is properly seated in the PS1201 face and tighten the screws to 10 in-lb (1.1 n-m) maximum. Do not overtighten the screws.

Page 24Network Wiring 52004-88 wiring the devices.fm

Page 25: Turbidimetro 1720 Aqua Tratada

Section 3

Figure 17 AquaTrend with SOM to PS1201 Power Supply Connections

Note: Refer to Table 5 and Table 6 on page 30 for Analog/Recorder Output and Relay Connection information.

Table 3 AquaTrend with SOM to PS1201 Power Supply Connections

AquaTrend with SOM J2 Terminal Connections in the AquaTrend with SOM

PS1201Connections to PS1201 Terminal Block

Position on J2 Terminal Wire Color/FunctionPosition on PS1201

Terminal Wire Color

Pin 1 Clear/shield Pin 1 Red

Pin 2 Green/network A Pin 2 No Connection

Pin 3 White/network B Pin 3 Black

Pin 4 No Connection Pin 4 Shield1

1 Insulate the Shield wire with the appropriate tubing, Cat. No. 40825-00.

Pin 5 Red/+12 Pin 5 White

Pin 6 Black/ Pin 6 Green

Page 2552004-88 wiring the devices.fm Network Wiring

Page 26: Turbidimetro 1720 Aqua Tratada

Section 3

To connect the cable at the AquaTrend with SOM end:

1. Make sure no power is being supplied to the power supply.

2. Use a flat-blade screwdriver to loosen the screws of the AquaTrend face.

3. Lift the face off the AquaTrend. Disconnect the ribbon cable from the circuit board in the back enclosure and set the face aside, or, store it by screwing the right side of the face to the left side of the back enclosure.

4. Use a punch and hammer to remove one knock-out on the inside rim of the bottom housing. Center the punch on the knock-out for best results.

Note: To assure the AquaTrend continues to meet NEMA 4X, IP66 enclosure ratings, use only the supplied NEMA-style enclosure strain relief fittings.

5. Refer to the first two columns in Table 3 and connect the cable wires to the six-position terminal block (J2) as shown in Figure 17. Use a small flat-blade screw driver to depress each nylon lever, then insert the appropriate wire into the terminal block.

Note: If any color wire is not present in the terminal block, skip that connection.

Note: Although conduit is generally not required for low voltage network wiring, conduit, if used, must be bonded together by the customer. This product does not provide automatic conduit bonding.

6. Make other wiring connections at this time if necessary, then attach the AquaTrend face with the four screws removed earlier. Tighten the screws to 10 in-lb (1.1 n-m) maximum. Do not overtighten the screws.

3.1.4 AquaTrend without SOM to PS1201 Power Supply ConnectionOrder enough cable (Cat. No. 52158) from Hach to suit the specific installation parameters. Connections are made at both ends of the cable; for the AquaTrend use the connection information in the first two columns of Table 4 on page 28, and for the PS1201 Power Supply use the information presented in the third and fourth column of Table 4 on page 28.

Make the connections at the PS1201 end as follows:

1. Turn off all power to the PS1201 Power Supply and disconnect all cables if necessary.

2. Unscrew the four screws securing the PS1201 face using a flat-bladed screwdriver.

3. Lift the face off the PS1201 and set it aside, or, store the front cover by screwing the right side of the front cover to the left side of the back enclosure.

4. If necessary, carefully use a punch and hammer from the outside of the enclosure to remove one of the left three unused knock-outs on the bottom of the Power Supply enclosure. Center the punch on the knock-out for best results.

5. Install the cable from the AquaTrend with the strain relief fitting provided in the hole described in Step 4, above. Make sure the outer cable insulation is just captured by the strain relief fitting.

Page 26Network Wiring 52004-88 wiring the devices.fm

Page 27: Turbidimetro 1720 Aqua Tratada

Section 3

6. Strip six inches (15 cm) of outer insulation from the cable (Cat. No. 52158).

7. Strip ¼-inch of insulation from each wire and insulate the shield wire with tubing (Cat. No. 40825-00).

8. Connect the cable wires to one of the six-position terminal blocks on the PS1201 using the position information given in the third and fourth column of Table 4 on page 28.

9. Tighten the strain relief fitting using the supplied washer and locknut until the cable no longer slides in the strain relief.

Note: To ensure the PS1201 continues to meet NEMA 4X, I.P.66 enclosure ratings, use only the supplied NEMA-style enclosure strain relief fittings.

10. If other power or network connections are to be made at this time, refer to the appropriate section in this manual or the individual device manual. If no other connections need to be made, continue with the remainder of this procedure.

11. Ensure that the gasket is properly seated in the PS1201 face and tighten the screws to 10 in-lb (1.1 n-m) maximum. Do not overtighten the screws.

Make the connections at the AquaTrend without SOM as follows:

1. Make sure no power is being supplied to the power supply.

2. Use a flat-blade screwdriver to loosen the screws of the AquaTrend face.

3. Lift the face off the AquaTrend and set it aside, or, store it by screwing the right side of the face to the left side of the back enclosure.

4. Use a punch and hammer to remove one knock-out on the inside rim of the bottom housing. Center the punch on the knock-out for best results.

Note: To assure the AquaTrend continues to meet NEMA4X, IP66 enclosure ratings, use only the supplied NEMA-style enclosure strain relief fittings.

5. Refer to the first and second column of Table 4 and connect the cable wires to the six-position terminal block (TB1) as shown in Figure 18. Use a small flat-blade screw driver to depress each nylon lever, then insert the appropriate wire into the terminal block.

Note: If any color wire is not present in the terminal block, skip that connection.

Note: Although conduit is generally not required for the low voltage network wiring, conduit, if used, must be bonded together by the customer. This product does not provide automatic conduit bonding.

6. Make other wiring connections at this time if necessary, then attach the AquaTrend face using the four screws.

Page 2752004-88 wiring the devices.fm Network Wiring

Page 28: Turbidimetro 1720 Aqua Tratada

Section 3

Figure 18 AquaTrend without SOM Network and Power Connections

Table 4 AquaTrend without SOM Network and Power Connections

AquaTrend without SOMConnections to TB1 Terminal

PS1201Connections to PS1201 Terminal Block

Position on TB 1 Terminal

Wire Color/FunctionPosition on

PS1201 Terminal Wire Color

Pin 6 Black/ground Pin 1 Red

Pin 5 Red/ +12 Pin 2 No Connection

Pin 4 No Connection Pin 3 Black

Pin 3 Green/ Network B Pin 4 Shield1

1 Insulate with appropriate tubing, Cat. No. 40825-00.

Pin 2 White/network A Pin 5 White

Pin 1 Clear/shield Pin 6 Green

Page 28Network Wiring 52004-88 wiring the devices.fm

Page 29: Turbidimetro 1720 Aqua Tratada

Section 3

3.1.5 SOM to Power Supply ConnectionsOrder enough cable (Cat. No. 52158) from Hach to suit the specific installation parameters. Connections are made at both ends of the cable; for the SOM use the connection information in the first two columns of Table 7 on page 31, and for the PS1201 Power Supply, use the information presented in the third and fourth column of Table 7 on page 31.

Note: The maximum cable length from the PS1201 to the most distant device is 100 ft. (30 m).

Make the connections at the PS1201 end as follows:

1. Turn off all power to the PS1201 Power Supply and disconnect all cables if necessary.

2. Unscrew the four screws securing the PS1201 face using a flat-bladed screwdriver.

3. Lift the face off the PS1201 and set it aside, or, store the front cover by screwing the right side of the front cover to the left side of the back enclosure.

4. If necessary, carefully use a punch and hammer from the outside of the enclosure to remove one of the left three unused knock-outs on the bottom of the PS1201 enclosure. Center the punch on the knock-out for best results.

5. Install the cable from the SOM with the strain relief fitting provided in the hole described in Step4, above. Make sure the outer cable insulation is just captured by the strain relief fitting.

6. Strip six inches (15 cm) of outer insulation from the cable (Cat. No. 52158).

7. Strip ¼-inch of insulation from each wire and insulate the shield wire with tubing (Cat. No. 40825-00).

8. Connect the cable wires to the six-position terminal block on the PS1201 (Cat. No. 52143-00) in the positions shown in Figure 19 on page 31 and in the third and fourth columns of Table 7 on page 31.

9. Tighten the strain relief fitting using the supplied washer and locknut until the cable no longer slides in the strain relief.

Note: To ensure the PS1201 continues to meet NEMA 4X, I.P.66 enclosure ratings, use only the supplied NEMA-style enclosure strain relief fittings.

10. If other power or network connections are to be made at this time, refer to the appropriate section in this manual or the individual device manual. If no other connections need to be made, continue with the remainder of this procedure.

11. Ensure that the gasket is properly seated in the PS1201 face and tighten the screws to 10 in-lb (1.1 n-m) maximum. Do not overtighten the screws.

To connect the cable at the SOM end:

1. Make sure no power is being supplied to the power supply.

2. Use a flat-blade screwdriver to loosen the screws of the SOM face.

Page 2952004-88 wiring the devices.fm Network Wiring

Page 30: Turbidimetro 1720 Aqua Tratada

Section 3

3. Lift the face off the SOM and set it aside, or, store it by screwing the right side of the face to the left side of the back enclosure.

4. Use a punch and hammer to remove one knock-out on the inside rim of the bottom housing. Center the punch on the knock-out for best results.

Note: To assure the SOM continues to meet NEMA 4X, IP66 enclosure ratings, use only the supplied NEMA-style enclosure strain relief fittings.

5. Refer to the first two columns in Table 7 and connect the cable wires to the six-position terminal block (J2) as shown in Figure 19. Use a small flat-blade screw driver to depress each nylon lever, then insert the appropriate wire into the terminal block.

Note: If any color wire is not present in the terminal block, skip that connection.

Note: Although conduit is generally not required for the low voltage network wiring, conduit, if used, must be bonded together by the customer. This product does not provide automatic conduit bonding.

6. Make connections at the Analog/Recorder Output Terminal Block and the Relay Terminal Block using the information in Table 5 and Table 6.

7. Make other wiring connections at this time if necessary, then attach the SOM face using the four screws. Tighten the screws to 10 in-lb (1.1 n-m) maximum. Do not overtighten the screws.

Table 5 SOM Analog/Recorder Output Connections

Analog/Recorder Output Terminal Block(J3 Terminal)

Function

J3-1 Analog Out #2 –

J3-2 Analog Out #2 +

J3-3 Analog Out #1 –

J3-4 Analog Out #1 +

Table 6 SOM Relay Connections

Relay Terminal Block (J1 Terminal) Function

1 Relay 1 normally closed

2 Relay 1 normally open

3 Relay 1 common

4 Relay 2 normally closed

5 Relay 2 normally open

6 Relay 2 common

Page 30Network Wiring 52004-88 wiring the devices.fm

Page 31: Turbidimetro 1720 Aqua Tratada

Section 3

Figure 19 SOM to PS1201 Power Supply Connections

Table 7 SOM to PS1201 Power Supply Connections

SOM J2 Terminal Connections in the SOM

PS1201Connections to PS1201 Terminal Block

Position on J2 Terminal

Wire Color/FunctionPosition on

PS1201 Terminal Wire Color

Pin 1 Clear/shield Pin 1 Red

Pin 2 Green/network A Pin 2 No Connection

Pin 3 White/network B Pin 3 Black

Pin 4 No Connection Pin 4 Shield1

1 Insulate with appropriate tubing, Cat. No. 40825-00

Pin 5 Red/+12 Pin 5 White

Pin 6 Black Pin 6 Green

Page 3152004-88 wiring the devices.fm Network Wiring

Page 32: Turbidimetro 1720 Aqua Tratada

Section 3

3.2 Power Supply to Power Supply Network Interconnect

Note: The steps that follow must be duplicated for each power supply. The only exception is the connection of the shield. When power supplies are connected in a series of two or more, the shield is connected at only one end for each pair of power supplies.

Each turbidimeter must have a PS1201 to provide ac power. The power supplies may be interconnected via a two conductor, shielded cable with a length less than 400 meters. Although these types of connections are usually made between two PS1201 Power Supplies, information for interconnecting between other modules is presented in the instruction manuals for those modules.

For connections between two Power Supplies, refer to Figure 20 and proceed as follows:

Figure 20 Power Supply Interconnect Wiring

1. Make sure no power is being supplied to either PS1201 Power Supply.

2. Unscrew the four screws securing each PS1201 face using a flat-bladed screwdriver.

3. Lift the faces off both of the Power Supplies and set them aside, or, store the front cover by screwing the right side of the front cover to the left side of the back enclosure.

4. If necessary, use a punch and hammer (from the outside of the enclosure) to remove one of the left three unused knock-outs on the bottom of each PS1201 enclosure. Center the punch on the knock-out for best results.

Page 32Power Supply to Power Supply Network Interconnect 52004-88 wiring the devices.fm

Page 33: Turbidimetro 1720 Aqua Tratada

Section 3

5. Route the interconnect cable (Cat. No. 52157) to each PS1201 enclosure through the holes created in Step 4, using the strain relief fittings provided. Make sure the outer cable insulation is just captured by the strain relief fitting.

6. Strip six inches (15 cm) of outer insulation from the cable (Cat. No. 52157).

7. Strip ¼-inch of insulation from each wire and insulate the shield wire with appropriate tubing. Insulate the non-connected end of the shield wire with electrical tape.

8. Connect the ends of the cable as shown in Table 8. See Figure 20 on page 32.

Note: Connect the shield only at one end of the interconnect cable (choose either end). Insulate the non-connected end with tape or heat-shrink tubing.

9. Tighten the strain relief fitting using the supplied washer and locknut until the cable no longer slides in the strain relief.

Note: To assure the PS1201 continues to meet NEMA 4X, I.P.66 enclosure ratings, use only the supplied NEMA-style enclosure strain relief fittings.

10. If other power or network connections are to be made at this time, refer to the appropriate section in this manual or the individual device manual. If no other connections need to be made, continue with the remainder of this procedure.

11. Ensure the gaskets are properly seated and replace the faces on both power supplies. Replace the corner screws and tighten to 10 in-lb (1.1 n-m) maximum to maintain the NEMA rating.

3.3 Other Device WiringWiring information for the Serial Input/Output (SIO) Module is presented in Section 9.1 on page 53 or in the SIO Manual (Cat. No. 52074-18).

Refer to the Digital Display Module (DDM) Manual, Cat. No. 52400-18, for information on wiring the DDM to the AquaTrend network.

Table 8 Interconnecting Network Wiring Between Two Power Supplies

Terminal Position Wire Color

POSITION 1 No Connection

POSITION 2 No Connection

POSITION 3 No Connection

POSITION 4SHIELD (Insulate with appropriate

tubing, Cat. No. 40825-00,)

POSITION 5 WHITE (with 52143-00)

POSITION 6 GREEN (with 52143-00)

Page 3352004-88 wiring the devices.fm Other Device Wiring

Page 34: Turbidimetro 1720 Aqua Tratada

Section 3

3.4 Network Termination

Note: To avoid damaging the instrument, remove power to the instrument before removing the front cover.

The network signals require a terminator for proper operation. As shipped from the factory, the terminator is disabled on all devices.

Important Note: Failure to properly terminate the network may cause total network failure. Only one device must have the network terminator enabled.

3.4.1 Network Termination at the AquaTrend/SOM or the SOM

1. Identify the AquaTrend/SOM or SOM the network termination will be performed on.

2. Remove the front face of the AquaTrend/SOM or SOM and locate J4 (a white, three-pin connector shown in Figure 21) on the circuit board in the back enclosure of the device. A jumper connected only to pin 2 should be present as shown in Figure 22.

3. To enable the terminator, lift the jumper off of pin 2 and move the jumper so that it connects pins 2 and 3 as shown in Figure 21.

4. Make sure all cable strain reliefs are tightened properly and replace the front panel.

Figure 21 Network Termination at the AquaTrend/SOM or SOM

Page 34Network Termination 52004-88 wiring the devices.fm

Page 35: Turbidimetro 1720 Aqua Tratada

Section 3

Figure 22 Network Termination

3.4.2 Network Termination at the AquaTrend without SOM (Model 51200)

Follow the steps below to enable the terminator and complete the installation of the AquaTrend.

1. Remove the front face of the AquaTrend and locate J6 on the circuit board in the back housing of the device. J6 is a white, three-pin connector. A jumper should be present and connected only to pin 2.

2. To enable the terminator, lift the jumper off of pin 2 and move the jumper so that it connects pins 2 and 3 as shown in Figure 23.

3. Replace the AquaTrend front face and tighten the four corner screws to 10 in-lb (1.1 n-m) maximum. Do not overtighten the screws.

Figure 23 Network Termination at the AquaTrend

Page 3552004-88 wiring the devices.fm Network Termination

Page 36: Turbidimetro 1720 Aqua Tratada

Visit http: //www.hach.com

Page 37: Turbidimetro 1720 Aqua Tratada

Section 4 Plumbing the Turbidimeter

4.1 Hydraulic ConnectionsSample inlet and drain connections are made on the turbidimeter body. The sample inlet fitting installed in the body is a ¼-inch NPT x ¼-inch compression fitting. One additional fitting supplied with the instrument is a ½-inch NPT-to-hose fitting for use with ½-inch ID flexible plastic tubing on the drain.

Figure 24 Sample Flow

One-fourth inch OD rigid or semirigid tubing is recommended for sample lines. Run them as directly as possible between the turbidimeter body and the sampling point to minimize sample flow lag time.

The optimum flow rate is 250-750 mL/minute (4.0 to 11.9 gal/hour). Flow rate into the turbidimeter may be controlled with a flow restriction device on the inlet line. Flow rates below 250 mL/min. will reduce response time and cause inaccurate readings. Flow rates above 750 mL/min. will cause the turbidimeter to overflow into the drain. Inaccurate readings may be seen if the flow rate is too high.

4.2 Starting Sample FlowStart sample flow through the instrument by opening the sample supply valve. Allow the turbidimeter to run long enough for the tubing and body to become completely wetted and the reading on the display to stabilize. One to two hours may be required initially for complete stabilization.

Page 3752004-88 plumbing the turb.fm Plumbing the Turbidimeter

Page 38: Turbidimetro 1720 Aqua Tratada

Visit http: //www.hach.com

Page 39: Turbidimetro 1720 Aqua Tratada

Section 5 Power and Startup of the System

5.1 AC Input Power Please refer to the PS1201 Power Supply Manual for all necessary information for wiring the PS1201 Power Supply to AC power.

When all network and AC wire connections have been completed, make sure all modules are closed and enclosure gaskets and covers are securely in place to maintain the NEMA rating on the equipment. Supply sample to the turbidimeter and initially allow it to flush the turbidimeter body for one to two hours. Supply power to all PS1201 Power Supplies. Check the status of all modules and sensors. If the module or sensor has an external configure light, the light should be flashing. (The configure light is internal to the Signal Output Module (SOM) and AquaTrend with SOM.) The configure light on each module and sensor will be extinguished when the modules are configured by the software.

5.2 About the AquaTrendThe AquaTrend keypad, shown in Figure 25, enables the user to navigate the AquaTrend network. Key functions are explained in Table 9.

Figure 25 AquaTrend Keypad

Table 9 AquaTrend Key Functions

Number Key Function of Keys

1Allows for navigating through the various menus, changing of settings, and scrolling through characters and across text lines.

2Displays active data in numeric or graphic format. Pressing the DISPLAY key toggles between data presented in numerical form and graphic form.

3 Displays the main menu from any menu level or from the data display.

4From the Main Menu, provides access to the data display screen. From any other menu level, displays the previous menu screen. The EXIT key can also be used to abandon any alphanumeric entry or selection process.

5 In any menu, accepts menu options and confirms entries.

Page 3952004-88 power and startup.fm Power and Startup of the System

Page 40: Turbidimetro 1720 Aqua Tratada

Section 5

5.3 Identifying the Master AquaTrendEvery AquaTrend Network System requires a network manager to perform network operations, such as adding or removing devices. The AquaTrend and AquaTrend with SOM can perform network management functions when set to Master AquaTrend status through the AquaTrend software; however, only one Master AquaTrend can be present on the network at a time. All other AquaTrends must be set to Remote AquaTrend status. Care should be taken when determining the location of the Master AquaTrend.

Every AquaTrend is shipped in an undefined state. When connecting only one AquaTrend to the network, it must be designated as the Master AquaTrend.

Set the system AquaTrend to Master status as follows:

1. Upon initial power-up, select MASTER and press the ENTER key.

2. Press ENTER to select Network Manager to ON. After a brief delay, the display will show the data display screen (CHANNEL NOT ASSIGNED).

3. Press the MENU key to proceed to the Main Menu.

Page 40Identifying the Master AquaTrend 52004-88 power and startup.fm

Page 41: Turbidimetro 1720 Aqua Tratada

Section 6 Setting Up the Network

The Master AquaTrend is designed to enable the user to easily perform setup and network functions for all attached devices from one central location. The Main Menu structure is shown in Figure 26. All network functions initiate from these menu categories. Return to the main menu screen from any menu level by pressing the MENU key. Back-up one menu level with each press of the EXIT key.

Figure 26 Master AquaTrend Main Menu

6.1 Setting the Time and DateVerify the time and date are correct. If necessary, change them as instructed in the following sections.

6.1.1 Setting the Time

1. Press the MENU key on the AquaTrend to access the Main Menu.

2. Use the up or down arrow key to move the pointer to AQUATREND MENU and press ENTER.

3. Select the SET TIME/DATE option and press ENTER.

Note: The AquaTrend uses a 24-hour clock (military time), so there is no selection for a.m. and p.m. For example, if the current time is 4:00 p.m., enter the 24-hour format equivalent of 16:00.

a. Select SET TIME.

b. A pop-up box will prompt you to enter the new time. The numbers representing the hour will be highlighted.

c. Press the up arrow key to increase the hour setting and press the down arrow key to decrease the setting.

d. When the proper hour is displayed, press the right arrow key and set the minutes as in Step c above.

Page 4152004-88 setting up the network.fm Setting Up the Network

Page 42: Turbidimetro 1720 Aqua Tratada

Section 6

e. Press ENTER to accept the displayed value, or press EXIT to cancel.

6.1.2 Setting the Date

1. Press the MENU key to access the Main Menu.

2. Use the up or down arrow key to move the pointer to AQUATREND MENU and press ENTER.

3. Select the SET TIME/DATE option and press ENTER.

a. Select the SET DATE option and press ENTER.

b. A pop-up box will prompt you to enter the new date.

c. Press the up and down arrow keys to increment through the available digits. Press the right or left arrow key to move to another character field within the date.

d. When the proper date is displayed, press ENTER to accept or press EXIT to cancel.

6.2 Network SetupA “network” is the connection of one or more devices (at least one must be a sensor) to one or more AquaTrends. A “device” can be an additional 1720D sensor, Digital Display Module (DDM), Serial I/O Module (SIO), Signal Output Module (SOM), or a remote AquaTrend.

Up to eight sensors, eight DDMs, eight remote AquaTrends, eight SOMs, two SIO Modules, and all necessary power supplies may be attached to the network.

The two procedures for adding device(s) to the network are presented in section 6.2.1 and section 6.2.2.

Page 42Network Setup 52004-88 setting up the network.fm

Page 43: Turbidimetro 1720 Aqua Tratada

Section 6

6.2.1 Installing a Single Sensor SystemUse this feature when installing a complete system with no more than (1) Sensor (the 1720D Turbidimeter is being used in this example), (1) Master AquaTrend, (1) SOM (as an integral part of an AquaTrend or as a separate device), (1) SIO, (1) DDM, and (1) remote AquaTrend. During this procedure, the default device settings are accepted and sensor measurements are automatically assigned to appropriate channels.

1. Press the MENU key to access the Main Menu.

2. Choose NETWORK MENU

3. Select SINGLE SENSOR SYSTEM. A pop-up box will appear stating SINGLE SENSOR SYSTEM SEARCHING FOR UNCONFIGURED DEVICES.

4. If the system is correctly installed, the display will next show a pop-up box listing all devices found. Press enter to install the single sensor system.

5. After configuration, the display will show OPERATION COMPLETE.

6. Press the DISPLAY key to show sensor data.

6.2.2 Installing Multiple Sensors or DevicesTable 10 presents available device types. Follow the procedure below to add a device. Repeat as many times as is necessary to add all needed devices to the network.

1. Press the MENU key to start from the AquaTrend Main Menu.

2. Move the pointer to NETWORK MENU and press ENTER.

3. Select ADD DEVICE and press ENTER. A pop-up box will show: SEARCHING FOR UNCONFIGURED DEVICES...PLEASE WAIT.

Table 10 AquaTrend Device Types

Device Type Devices

Sensor1720 Series Turbidimeter, Signal Input Module, or any APA 6000 Analyzer

Communications Serial I/O Module (SIO)

Outputs Signal Output Module (SOM)

Aquatrend AquaTrend or Remote AquaTrend

Display Digital Display Module (DDM)

Page 4352004-88 setting up the network.fm Network Setup

Page 44: Turbidimetro 1720 Aqua Tratada

Section 6

4. If only one unconfigured device type is connected to the network it will be selected automatically; if more than one type of unconfigured device is found, select the device type to be added.

5. If only one unconfigured device of the selected type is available, it will be selected automatically; if two or more devices of the same type are available, the AquaTrend will ask you to press the configure button on the device you want to add.

6. Press the configure button on the device to be added, if necessary. Refer to the appropriate device manual if you do not know where to find its configure button.

Note: The configure button for the SOM and SOM within the AquaTrend is located inside the device enclosure.

7. Press ENTER to confirm ADD DEVICE.

8. The AquaTrend will configure the device and assign it a unique name. Change the name if necessary using the arrow keys (right and left arrows to move to a different character in the name and up and down arrows to change the character). Press ENTER to confirm the name.

Page 44Network Setup 52004-88 setting up the network.fm

Page 45: Turbidimetro 1720 Aqua Tratada

Section 6

6.2.3 Attaching a Measurement to a ChannelAlthough sensors may have been added to the network, readings will not be displayed until a measurement is attached to a channel. Attach each sensor measurement to one of the eight AquaTrend channels when setting up a new sensor.

1. Press the MENU key to begin at the Main Menu.

2. Use the up or down arrow key to select the AQUATREND MENU and press ENTER.

3. Select DISPLAY SETUP and press ENTER.

4. Select MEASUREMENT TO CHANNEL and press ENTER.

5. Select ATTACH, and press ENTER. A list of available sensors will appear.

6. Select the sensor of interest and press ENTER.

7. Select the channel of interest and press ENTER.

8. Select ATTACH and press ENTER to continue.

The message OPERATION COMPLETE will appear. To see the data on the AquaTrend screen, press the DISPLAY key on the front of the AquaTrend (see Figure 26).

Page 4552004-88 setting up the network.fm Network Setup

Page 46: Turbidimetro 1720 Aqua Tratada

Section 6

6.3 Display Setup The AquaTrend has the capability to display data from a single sensor or up to eight sensors at one time. Viewing additional channels reduces the amount of information shown for each channel.

Access the display mode by pressing the DISPLAY key. Use the right arrow key to increase the number of channels displayed on the screen. Press the left arrow key to decrease the number of channels displayed. The up and down arrow keys moves the active channel designation.

6.4 Attaching the SOMAlthough the SOM is now physically attached and configured on the network, it is still necessary to configure necessary alarms and recorders for all 1720 Series Turbidimeters connected to the network. Information for making those network connections is presented in the respective 1720 Series Instruction Manual.

Network connection information for relays and outputs is presented in the SOM Manual.

6.5 Attaching the SIOPlease refer to the SIO Manual for attachment information.

Page 46Display Setup 52004-88 setting up the network.fm

Page 47: Turbidimetro 1720 Aqua Tratada

Section 7 Adding a Digital Display Module

Note: Even though the Digital Display Module is physically connected to a sensor, it will not display readings from the sensor until it is configured by the AquaTrend (see section 7.2).

Installing and adding the Digital Display Module (DDM) to the network requires three steps:

1. Mount the DDM on a wall or a panel. Refer to the DDM Manual (Cat. No. 52400-18) for mounting information.

2. Physically plug the DDM into a sensor to provide power and network connections for the display (Figure 27).

3. Configure the DDM from the AquaTrend and assign it to a sensor (section 7.2).

7.1 Attached Cordset ConnectionTo attach the DDM to sensors with “Remote” jacks, use the factory-attached 2-meter cordset. Plug the silver barrel connector of the cord into the “Remote” connection of the Hach sensor and tighten the barrel. The module then receives power and is connected to the network through the sensor. See Figure 27.

The DDM may show random characters after it is plugged into a sensor. These will disappear when the module is configured by the network (see section 7.2).

Figure 27 Cordset Connection

7.2 Network SetupThe network setup consists of three steps:

1. Adding the DDM to the AquaTrend network (section 7.3).

2. Assigning the DDM to a sensor (section 7.3.1).

3. If necessary, assigning a sensor measurement to the DDM (section 7.3.1).

Page 4752004-88 adding a digital display.fm Adding a Digital Display Module

Page 48: Turbidimetro 1720 Aqua Tratada

Section 7

The DDM must be added to the AquaTrend network before the AquaTrend can recognize and route data to the module. After the module is added, it must be assigned to a specific sensor.

The yellow LED on the Digital Display Module flashes when the module is initially plugged in to indicate the module is not configured. After the module is configured by the AquaTrend, the yellow LED turns off. (The light shines steadily while the CONFIGURE button is being pushed.)

7.3 Adding the Digital Display Module to the NetworkRefer to Section 7.3 on page 48 to add the DDM to the network, then continue with Section 7.3.1. Adjust other Network Menu settings according to instructions in the AquaTrend Interface Manual.

7.3.1 Assigning the Digital Display Module to a SensorPerform this procedure only after adding the DDM to the network.

Once a (Cat. No. 52400-18) is assigned to a sensor, it displays data from that sensor even if it is physically connected to a different sensor on the network. Refer to the DDM Manual (Cat. No. 52400-18) for additional information.

1. Press the MENU key to start from the AquaTrend Main Menu.

2. Select NETWORK MENU. Press ENTER.

3. Select SENSOR TO DDM. Press ENTER.

4. Select ATTACH. Press ENTER.

5. The AquaTrend will display the available sensors and their names. Select the desired sensor and press ENTER.

6. The AquaTrend will display the available DDMs and their names. Select the desired module. Press ENTER.

7. The AquaTrend will confirm the DDM has been assigned to the sensor.

Page 48Adding the Digital Display Module to the Network 52004-88 adding a digital display.fm

Page 49: Turbidimetro 1720 Aqua Tratada

Section 8 Using the Portable AquaTrend

8.1 The Portable AquaTrend The Portable AquaTrend communicates directly to the sensor it is attached to without interrupting network commands from the Master AquaTrend. The Portable AquaTrend functions primarily as a means of accessing sensor menus for calibration and to display and log data in the immediate vicinity of the sensor.

The rotary handle on the Portable AquaTrend provides a stand for placement on a flat surface. To position the handle, pull the handle out on both sides of its base and rotate the handle to the position desired (see Figure 28). Place the Portable AquaTrend on a surface near the sensor where the display is visible and the keypad is easily accessible.

Figure 28 Portable AquaTrend Handle/Stand

The Portable AquaTrend can be easily connected to a sensor by plugging the cord jack on the Portable AquaTrend into the receptacle provided on the sensor. See Figure 29.

Figure 29 Portable AquaTrend Jack and Receptacle

Page 4952004-88 using the portable aquatrend.fm Using the Portable AquaTrend

Page 50: Turbidimetro 1720 Aqua Tratada

Section 8

8.2 Mounting the Portable AquaTrendThe Portable AquaTrend’s mounting kit provides an easy way to mount the device to a wall or pole. See Figure 30 and Figure 31.

For wall mounting:

1. Use the bracket as a template to mark positions on the wall for four of the eight screw holes – either the vertical or horizontal pair.

2. Drill the marked positions for four ¼-20 screws.

3. Use four ¼-20 hex-head or pan-head screws to secure the bracket to the wall (10 in-lb [1.1 n-m] maximum).

4. Fit the bracket lip into the slot on the back of the Portable AquaTrend and push down until the device fits securely in the bracket (see Figure 30).

Figure 30 Wall Mount for the Portable AquaTrend

Page 50Mounting the Portable AquaTrend 52004-88 using the portable aquatrend.fm

Page 51: Turbidimetro 1720 Aqua Tratada

Section 8

For pole mounting:

1. Obtain two ¼-20 x 2.843 U-bolts and nuts.

2. Press the back of the mounting bracket against a pole in the proper mounting position (indentations on the back of the bracket fit the pole either horizontally or vertically).

3. Place the curve of one U-bolt behind the pole, and insert the threaded ends of the bolt through two of the holes on the mounting frame. The center holes are for vertical mounting and outside holes are for horizontal mounting.

4. Secure the two nuts with a wrench, and repeat Steps 1-3 for the second U-bolt.

5. Fit the bracket lip into the slot on the back of the Portable AquaTrend and push down until the device fits securely in the bracket (as in Figure 30).

Figure 31 Pole Mount for the Portable AquaTrend

Page 5152004-88 using the portable aquatrend.fm Mounting the Portable AquaTrend

Page 52: Turbidimetro 1720 Aqua Tratada

Visit http: //www.hach.com

Page 53: Turbidimetro 1720 Aqua Tratada

Section 9 Adding a Serial Input/output Module

9.1 Network Wiring Connections Attach the supplied connector (Cat. No. 52097) to the remaining unterminated end of the network cable attached to the PS1201 Power Supply. Refer to Figure 32 and make the connections as follows:

• White to position 1 on connector

• Green to position 2 on connector

• Shield to position 3 on connector

When the connector is attached to the cable, plug the connector into the rectangular hole labeled “NETWORK” on the front of the Serial Input/Output (SIO) Module.

Figure 32 Network Connections

9.2 Power ConnectionsThe SIO uses an AC adapter for power, either 120V (Cat. No. 46079-00) or 240V (Cat. No. 46080-00). Plug the adapter into the power jack on the SIO and then into an appropriate electrical socket.

9.3 Adding the SIO to the NetworkRefer to Section 9.3 on page 53 to add the SIO Module to the network. Adjust other Network Menu settings according to instructions in the AquaTrend Interface Manual.

After the SIO Module has been named and identified by the AquaTrend, changes can be made to the SIO Module settings, if needed. See the SIO Manual (Cat. No. 52074-18) for more information.

Page 5352004-88 adding an SIO.fm Adding a Serial Input/output Module

Page 54: Turbidimetro 1720 Aqua Tratada

Visit http: //www.hach.com

Page 55: Turbidimetro 1720 Aqua Tratada

General Information

At Hach Company, customer service is an important part of every product we make.

With that in mind, we have compiled the following information for your convenience.

Page 5552004-88 general info stopper.fm General Information

Page 56: Turbidimetro 1720 Aqua Tratada

Visit http: //www.hach.com

Page 57: Turbidimetro 1720 Aqua Tratada

How to Order

By Telephone:

By Fax:

6:30 a.m. to 5:00 p.m. MSTMonday through Friday(800) 227-HACH (800-227-4224)

(970) 669-2932

By Mail:

Ordering information by E-mail:

Hach CompanyP.O. Box 389Loveland, Colorado 80539-0389 U.S.A

[email protected]

Technical and Customer Service (U.S.A. only)Hach Technical and Customer Service Department personnel are eager toanswer questions about our products and their use. Specialists inanalytical methods, they are happy to put their talents to work for you.Call or E-mail .1-800-227-4224 [email protected]

International CustomersHach maintains a worldwide network of dealers and distributors.To locate the representative nearest you, send E-mail to

or contact:[email protected]

Loveland, Colorado, U.S.A.Telephone: (970) 669-3050; Fax: (970) 669-2932Hach Company World Headquarters;

Information Required

U.S.A. Customers

• Hach account number (if available) • Billing address

• Your name and phone number • Shipping address

• Purchase order number • Catalog number

• Brief description or model number • Quantity

Page 5752004-88 general information.fm How to Order

Page 58: Turbidimetro 1720 Aqua Tratada

Repair Service

In the United States:Hach Company100 Dayton AvenueAmes, Iowa 50010(800) 227- (U.S.A. only)Telephone: (515) 232-2533FAX: (515) 232-1276

4224

In Canada:Hach Sales & Service Canada Ltd.1313 Border Street, Unit 34Winnipeg, ManitobaR3H 0X4(800) 665-7635 (Canada only)Telephone: (204) 632-5598FAX: (204) 694-5134E-mail: [email protected]

In Latin America, the Caribbean, the Far East, theIndian Subcontinent, Africa, Europe, or the Middle East:Hach Company World Headquarters,P.O. Box 389Loveland, Colorado, 80539-0389 U.S.A.Telephone: (970) 669-3050FAX: (970) 669-2932E-mail: [email protected]

Authorization must be obtained from Hach Company before sendingany items for repair. Please contact the HACH Service Center servingyour location.

Page 58Repair Service 52004-88 general information.fm

Page 59: Turbidimetro 1720 Aqua Tratada

Limited Warranty

Hach Company warrants this product against any defects that are due to faulty material or workmanship for a period of one year from the date of purchase.

In the event that a defect is discovered during the warranty period, Hach Company agrees that, at its option, it will repair or replace the defective product or refund the purchase price of the defective product less any charges for shipping and handling.

This warranty does not apply to: (i) damage caused by Act of God; (ii) damage caused by misuse, neglect, accident or improper application or installation; (iii) damage caused by any repair or attempted repair not authorized by Hach Company; (iv) any product not used in accordance with the instructions furnished by Hach Company.

Any product repaired or replaced under this warranty will itself be warranted only for the remainder of the warranty period of the original product being repair or replaced.

This warranty does not apply to short-life or consumable components of a product, such as, without limitation, light bulbs and chemical reagents.

How To Obtain ServiceBecause Hach Company manufactures and warrants a variety of products, you must receive a Return Authorization and details as to how warranty service may be obtained from Hach Company by calling the appropriate number for your product, see Repair Service on page 58.

Modifications and LimitationsThis warranty contains the sole express warranty made by Hach Company in connection with its products. All implied warranties, including without limitation, the warranties of merchantability and fitness for a particular purpose, are expressly disclaimed.

Some states do not allow the disclaimer of implied warranties and if this is true in your state the above limitation may not apply to you. This warranty gives you specific rights, and you may also have other rights that vary from state to state.

This warranty constitutes the final, complete and exclusive statement of warranty terms and no person is authorized to make any other warranties or representations on behalf of Hach Company.

Limitation of RemediesThe remedies of repair, replacement or refund of purchase price as stated above are the exclusive remedies for the breach of this warranty. In no event shall Hach Company be liable for any incidental or consequential damages of any kind for breach of warranty, negligence, on the basis of strict Liability or under any other legal theory.

Geographical LimitationThis warranty is extended only to products purchased, delivered and used in the 50 United States.

Page 5952004-88 general information.fm Limited Warranty