Ultrasound sensors for micromoulding
M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. DerdouriIndustrial Materials Institute, National Research Council, Quebec CanadaB. Whiteside, M.T Martyn ,E. Brown, P.D. CoatesIRC in polymer engineering, University of Bradford.
![Page 2: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/2.jpg)
Ultrasound basics
Ultrasound pulses ~ 3.6-30MHz (for polymers)Evaluation of
•Transit time•Amplitude
Transmitter
Detector
Polymer
![Page 3: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/3.jpg)
Equipment
T & P T & P inputsinputs
to ultrasonic transducersto ultrasonic transducers
Computer controlled Computer controlled data acquisitiondata acquisition
Ultrasound Ultrasound inputinput
1GHz 1GHz sampling sampling frequencyfrequency
Commercial Commercial pulser-receiverpulser-receiver
Digital oscilloscope Digital oscilloscope (free standing or (free standing or internal PC card)internal PC card)
![Page 4: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/4.jpg)
Ultrasound velocity change with elastic moduli and density
• Longitudinal velocity– Bulk Modulus– Shear Modulus– Density
• For the melt range tested, G <<K
3
41 GKCl
KCl
Sensitive to filler level, morphology, temperature, pressure
![Page 5: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/5.jpg)
Attenuation– Scattering
• Inclusions/impurities in the material
– Absorption• Wave energy is absorbed by the material as heat.
Attenuation coefficients:-
Air:10 dB/MHz/cm Polyethylene:0.25 dB/MHz/cm Tool Steel:0.002 dB/MHz/cm
![Page 6: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/6.jpg)
Pulse – echo mode
Transducer
Polymer
A single transducer acts as transmitter and detector
When polymer enters the cavity, the amplitude of the steel/cavity interface drops and echoes are seen from the far cavity wall
These echoes are seen to move due to the cooling and freezing of the polymer, which reduces the transit time
![Page 7: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/7.jpg)
Flow front detection
Multiple sensors can be employed to monitor cavity fillingCan be useful for detection of ‘jetting’ effects
Polymer presence is indicated by a rapid variation of amplitude
![Page 8: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/8.jpg)
Shrinkage detection
Polymer
PolymerPolymer
Air
Polymer
Transducer
Transducer
Transducer
![Page 9: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/9.jpg)
Shrinkage Detection
1.04
1.06
1.08
1.10
1.12
1.14
1.16
1.18
1.20
1.22
0 2 4 6 8 10
Time (s)
Pe
ak
he
igh
t (v
)
0
1
2
3
4
5
6
7
8
9
10
Trig
ge
r (v
)
100mm/s
75mm/s50mm/s10mm/s
![Page 10: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/10.jpg)
53.5
54.0
54.5
55.0
55.5
56.0
56.5
0 10 20 30 40 50 60Melt pressure (bar)
Transit time (s)
Temperature/Pressure dependence
180°C
200°C
220°C
![Page 11: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/11.jpg)
Summary• Advantages
– Non-invasive technique– Sensitive to temperature, pressure,
morphology, filler level (nanocomposites)– Can be used to monitor cavity filling, cooling
and shrinkage
• Disadvantages– Difficult to isolate actual temperature and
pressure values – other sensors required
![Page 12: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/12.jpg)
Micromoulding applications?
![Page 13: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/13.jpg)
Sol-Gel spray application
1µm Bizmuth Titanate powder dispersed into solutionPiezo films are deposited on external surfaceThickness up to 100µm – determines the resonant frequencyFilms are poledSilver paste electrodes added to form transducer
![Page 14: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/14.jpg)
Technique benefits
• Very small form factor – well suited for micromoulding applications
• Installation of sensors on any surface, including curved surfaces
• Transducer can operate in pulse-echo mode• Able to operate at temperatures in excess of
500C
![Page 15: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/15.jpg)
Extrusion monitor
Sensors installed on external surface of extrusion module on Battenfeld Microsystem50Allows evaluation of material variations, screw wear
![Page 16: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/16.jpg)
Extrusion monitor
All pulses/echoes reflected from steel/cavity interface
Centre frequency of transducer~13Mhz
![Page 17: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/17.jpg)
Extrusion monitor
Polyethylene material
Screw speed 100rpm
![Page 18: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/18.jpg)
Cavity sensors
![Page 19: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/19.jpg)
Cavity data
Runner (Thickness 1mm) Cavity (Thickness 0.3mm)
![Page 20: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/20.jpg)
Cavity data
Polyethylene material
Lower transit times for the thinner section
Can be used to study cooling of the material
![Page 21: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/21.jpg)
Cooling monitoring
Result agree well with static tests
![Page 22: Ultrasound sensors for micromoulding M. Kobayashi*, C.-K. Jen, C. Corbeil, Y. Ono, H. Hébert and A. Derdouri Industrial Materials Institute, National Research](https://reader036.vdocuments.net/reader036/viewer/2022070305/55141bb5550346e7488b5568/html5/thumbnails/22.jpg)
Conclusions
• Sol-Gel method great potential route for manufacture of ultrasound transducers suitable for micromoulding applications
• Sensors have been installed on Microsystem50 and data has been produced
• Technology allows characterisation of the entire process
• Sensor size to be scaled down further