ultrsonography principle and application

36
ULTRASONOGRAPHY INTRODUCTION Ultrasonograpy is a non-invasive imaging modality using high frequency sound waves. • Used -For diagnostic - Tomographic imaging of soft tissues - to quantitate the functional movement of organ - to measure the size of the organ` - to study the patterns of blood flow - to assess fetal age - For therapeutic purpose— - HIFU- Knifeless surgery - Physiotherapy - Lithotripsy

Upload: suniu

Post on 07-May-2015

1.277 views

Category:

Education


2 download

TRANSCRIPT

Page 1: Ultrsonography Principle and application

ULTRASONOGRAPHY INTRODUCTION

• Ultrasonograpy is a non-invasive imaging modality using high frequency sound waves.

• Used

-For diagnostic - Tomographic imaging of soft tissues

- to quantitate the functional movement of organ

- to measure the size of the organ`

- to study the patterns of blood flow

- to assess fetal age

- For therapeutic purpose—

- HIFU- Knifeless surgery- Physiotherapy- Lithotripsy

Page 2: Ultrsonography Principle and application

  • The human ear can detect

frequencies in the range of

20 -20000 Hz.

• Sound above this range is known

as Ultrasound.

• Used for both diagnostic and therapeutic purposes.

• Most diagnostic instruments use sound in the range of 1-10 MHz

Page 3: Ultrsonography Principle and application

• THE IMAGE is displayed in various shades of grey depending on tissue density

e.g. bone appears white and fluid appears black.

• The number of shades of grey displayed by the machine is around 256, but the human eye can perceive only 16 shades of grey. This improves the resolution of the picture.

ULTRASONOGRAPHY DEMONSTRATES THE STRUCTURE OF TISSUE RATHER THAN TISSUE TYPE.

PULSE-ECHO ” PRINCIPLEPiezo- Electric crystal (Lead Zirconate

Titanate- PZT)

Page 4: Ultrsonography Principle and application

A ) ULTRASOUND WAVES NEED MEDIUM FOR TRAVELLING

• Velocity (V= ν ) of sound is constant in soft body tissue but increase with intensity of medium.

Page 5: Ultrsonography Principle and application

B) Attenuation / reduction in intensity• US beam is attenuated as it travels through tissues• Echoes reflected back towards the transducer are also

attenuated.• Factors contributing to Attenuation are:

1. Absorption:• Energy in ultrasound beam is absorbed by tissues and is

converted into heat. • It is the basis of therapeutic ultrasound.

Page 6: Ultrsonography Principle and application

2. Acoustic impedance- ‘Z’It is the resistance offered by tissues to the sound waves. Z tissue = Tissue density x Velocity of sound in that Tissue

Page 7: Ultrsonography Principle and application

3. Reflection: It is basis of diagnostic ultrasound• Sound Waves Are Reflected Back Towards transducers

giving rise to echo.• It Occurs – if two adjuscent tissues have different Z – occurs at patient transducer interface - poor coupling – incidence angle

– “Z” (TISSUE DENSITY) determines the % of the reflected beam as it passed from one tissue to another.

Page 8: Ultrsonography Principle and application

Sound reflection at various interfaces

Page 9: Ultrsonography Principle and application

3. Scattering: • It occurs when the beam encounters irregular interface.• Angle of US beam interacting with this interface results

in scattering in all direction.• CAUSES ARTIFACTS

Page 10: Ultrsonography Principle and application

4. Refraction:• sound waves bend at CURVED

interfaces

• Angle of incidence.

• Affects both transmission & reflection

• Can create artifacts

5. Transmission:When [Z] of tissues at interface are same- allow

penetration to depth

Page 11: Ultrsonography Principle and application

A.

COMPONENTS OF AN ULTRASONOGRAPHIC MACHINE• TRANSDUCERS• CONTROL PANEL• COMPUTER• PRINTER• COUPLING GEL

TRANSDUCER • Function: To send and receive signals• Piezo – electric crystal (Lead zerconate titanate)• FREQUENCY- Resolution Vs Penetration• Lower the frequency, lesser resolution, greater

penetration.• Greater the frequency, greater resolution, lesser

penetration.

Page 12: Ultrsonography Principle and application

Applications: superficial structuresVascular, Small Parts

LINEAR ARRAY TRANSDUCERS Thin rectangular clips lined side by side (60-256 crystal in line)

Beam produced is rectangular

TYPES OF TRANSDUCERS

Page 13: Ultrsonography Principle and application

CURVI-LINEAR TYPE Shaped in a curve – Trapezoidal view –

both superficial & deeper

structures are imaged.

Page 14: Ultrsonography Principle and application

SECTOR TYPE• Single crystal oscillates to

provide fan shaped beam• Small size, more

maneuverability• Used for thoracic and

abdominal organs through small contact area

ApplicationsCardiac , Abdomen , Transcranial , Vascular

Page 15: Ultrsonography Principle and application

MISCELLANEOUS TRANSDUCERS • US TRANSDUCER with a Biopsy probe

guide

• Endoscopic - Esophageal probe

Page 16: Ultrsonography Principle and application

Image Display Modes

A-Mode (Amplitude mode)Single Ultrasound beam is usedThe returning echoes are shown as peaks along the horizontal axis.The height of the peak is directly proportional to the strength of the echo. Gives information about the organ boundaries.

Page 17: Ultrsonography Principle and application

B-Mode (Brightness mode):• Multiple Ultrasound beams

are used.• Returning echo is depicted

as the dots.• Brightness of the dot is

directly proportional to the strength of the echo.

• So, is also called as the “Grey- Scale imaging”.

• 2D image – • most popular mode of

display

Page 18: Ultrsonography Principle and application

Real-Time B- Mode Ultrasonography

Real-time B-mode scanners display a moving gray scale image of cross sectional anatomy

Page 19: Ultrsonography Principle and application

M- Mode (Time-Motion mode)

• Single Ultrasound beam is used. • Returning echoes depicted as dots.• Position of dot will denote depth of

organ -- along the vertical axis. • Moving structure (Time) on

horizontal axis.• Brightness of dot denote the

strength of echo. • It is used – Echocardiography-

Page 20: Ultrsonography Principle and application

SCANNING PROCEDURE

• First the organ to be scanned is decided- case history, complaint, symptoms, clinical examination, lab examination, radiological examination etc

• With the knowledge of the

topographic anatomy, use the ‘acoustic window’ (Easiest and the nearest site for passing the ultrasound wave into the desired organs)

Page 21: Ultrsonography Principle and application
Page 22: Ultrsonography Principle and application

DORSAL RECUMBANCY - abdomen

Page 23: Ultrsonography Principle and application

Selection of transducer• It depends on – Size of animals – Depth of organ.

– Objective : Choose the highest frequency that will penetrate to the depth needed for the particular examination and gives highest resolution.

• Small dog, cats 7.5 to 10.0 MHz• Medium dogs 5 MHz• Large dogs 3.5 to 5.0 MHz• Large animals 2.0 to 3.5 MHz

Page 24: Ultrsonography Principle and application
Page 25: Ultrsonography Principle and application
Page 26: Ultrsonography Principle and application

Scanning controls• Near /Far Gain:- alter

amplification of returning echoes

• Time compensation gain:

• Depth- determine the depth

of US image

• Freeze- real time image can be temporarily frozen

Basic set of controls

Page 27: Ultrsonography Principle and application

IMAGE INTERPRETATION

• Hyper-echoic / Echogenic /Echo Rich- WHITE AREAS– Given by highly

reflective interface such as bone or air.

• Hypo- Echoic / Echo poor- GREY AREAS– Given by interface of

moderate reflection.

– Anechoic/echo lucent/echo free- BLACK AREAS • Denotes the complete

transmission of the sound as through the fluids.

– Iso echoic-

THE STRUCTURE OF TISSUE RATHER THAN TISSUE TYPE.

Page 28: Ultrsonography Principle and application

•TISSUES IN ORDER OF INCREASING ECHOGENECITY

•Bile, urine•Renal medulla•Muscle•Renal cortex•Liver•Storage fat•Spleen•Prostate•Renal sinus•Structural fat, vessel walls•Bone, gas, organ boundaries, calculi

Interpretation based on texture of organ

Uniform/ regular /homogenous

Non-uniform/ irregular/ non-homogenous

Fine Granular / Coarse Granular

Page 29: Ultrsonography Principle and application

objectives

– Presence or absence of a pregnancy– Identify the location of the pregnancy –

intrauterine or extrauterine. – Assess the growth and development of the fetus. – Placental localization-– Assess the amount of liquor amnii.– Assess the fetal age.

IMAGING OF REPRODUCTIVE TRACT

Page 30: Ultrsonography Principle and application

Time of Sonographic recognition of canine fetal structures

Page 31: Ultrsonography Principle and application

3.TESTES• Topography:

– Testicles are superficial structures- high freq. transducer used.– Non scrotal testicles search should begin in inguinal region.

• NORMAL APPEARANCE– well circumscribed, smooth outline, oval in shape– parenchyma moderately echogenic

Sagital image Transverse image

Page 32: Ultrsonography Principle and application

IMAGING OF THE THORACIC CAVITY

• Dis advantages: Not easy to image.– the rib cage – air within the lungs

• Transducer is placed – on Inter costal Spaces,– behind the xiphisternum – at the thoracic inlet.

Page 33: Ultrsonography Principle and application

DUPLEX DOPPLER ULTRASONOGRAPHY • Involves the simultaneous use of real time B-Mode imaging and pulsed doppler ultrasound

PRINCIPLE OF DOPPLER: Blood cells moving towards transducer give bright echoes & which move away from transducer give weak echoes.-Towards transducer – red / yellow / orange-Away from transducer – blue/green

Page 34: Ultrsonography Principle and application

FORMS OF DOPPLER• Pulsed Doppler:• Continuous Doppler:

Uses:-to identify structures by the presence or absence of

blood flow to organ-to detect thrombus or clot in blood vessel-to study direction of blood flow & associated

abnormality

Page 35: Ultrsonography Principle and application

Advantages of ultrasound• It is a non- invasive diagnostic technique• Differentiation of soft tissue abnormalities• Visualization of intra- organ anomaly • Technique can be performed on any patient

i.e., old, dyspnoeic and comatose patients• No known harmful effects • Visualiztion of Biopsy being taken.

Page 36: Ultrsonography Principle and application

Constraints• Ultrasound imaging is not suitable – Examination of tissues lying below bones –Air containing organs – lungs, intestine etc

• Clipping / shaving of hairs –aesthetic value• Application of coupling medium such as mineral

oil, aqueous gel - desquamation of cells leading to metastasis.

• High capital investment involved• Technical expertise