18th International Symposium on the Application of Laser and Imaging Techniques to Fluid MechanicsLISBON | PORTUGAL JULY 4 – 7, 2016 Understanding Arterial Pressure from PIV data Patrick H. Geoghegan 1,* , Paul D. Docherty 1 , Mathieu Sellier 1 , Mark C. Jermy 1 1: Dept. of Mechanical Engineering, University of Canterbury, New Zealand * Correspondent author: [email protected] Keywords: Biomedical flow, Arterial Hemodynamics, Compliant, Pressure ABSTRACT Cardiovascular disease is a major cause of disease around the world and is estimated to cost developed countries roughly 2% of their GDP annually. Critically ill patients with hemodynamic dysfunction exhibit reduced stroke volume, compliance, and diffusion of oxygenated blood to the organs, which leads to organ failure. Current treatment techniques include inotrope therapy and fluid resuscitation which are treated with independently selected doses based on mean arterial pressure and heart rate. This lacks concordance amongst clinicians regarding the optimal dosing levels. If the spatial variance in the pressure and flow fields in hemodynamic dysfunction could be determined, the confidence intervals for model-based capture and titration of therapy could be developed. In this work stereoscopic particle image velocimetry is used to obtain the velocity fields in a simplified model of a stenosis (a phantom). This thin walled compliant phantom is manufactured so that the mechanical response is equivalent to in- vivo conditions. The experimental set-up imparts a pulsatile flow on the phantom ensuring in-vivo conditions by matching both Womersley and Reynolds number. The pressure fields are obtained via integration of the Navier-Stokes equation and correlated to the upstream pressure signal. By applying the Navier-Stokes equations to the velocity field data from the PIV analysis and in contrast to typical computational fluid dynamics (CFD) there was no need to calibrate and iterate between pressure and velocity domains to satisfy a series of boundary conditions. The transient pressure domain information was obtained by the simple determination of the partial derivatives of the velocity. The pressure values were stable, smooth, in the expected order of magnitude, and intuitively correct. The results show that a pressure difference across the stenosis can be observed, and that this is related to the severity of the stenosis. This provides a pathway into stenosis diagnosis using pressure measurement equipment currently used in medical practice, after some further analysis and investigation has been carried out. 1. Introduction Cardiovascular Disease (CVD) caused 30% of all deaths in New Zealand in 2012 (Ministry of Health 2015). CVD has been estimated to cost ~2% of GDP in developed countries (Adeyi et al. 2007). Critically ill patients with hemodynamic dysfunction exhibit reduced stroke volume, compliance, and diffusion of oxygenated blood to the organs (Davies et al. 2007). This ultimately

Upload: others

Post on 17-Jun-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Understanding Arterial Pressure from PIV dataltces.dem.ist.utl.pt/lxlaser/lxlaser2016/final... · 3) 1060 1140 Young’s Modulus (N/m 2) 776 000 1 320 000 Time Period Oscillation

18th International Symposium on the Application of Laser and Imaging Techniques to Fluid Mechanics・LISBON | PORTUGAL ・JULY 4 – 7, 2016

Understanding Arterial Pressure from PIV data

Patrick H. Geoghegan1,*, Paul D. Docherty1, Mathieu Sellier1, Mark C. Jermy1 1: Dept. of Mechanical Engineering, University of Canterbury, New Zealand

* Correspondent author: [email protected]

Keywords: Biomedical flow, Arterial Hemodynamics, Compliant, Pressure

ABSTRACT

Cardiovascular disease is a major cause of disease around the world and is estimated to cost developed countries roughly 2% of their GDP annually. Critically ill patients with hemodynamic dysfunction exhibit reduced stroke volume, compliance, and diffusion of oxygenated blood to the organs, which leads to organ failure. Current treatment techniques include inotrope therapy and fluid resuscitation which are treated with independently selected doses based on mean arterial pressure and heart rate. This lacks concordance amongst clinicians regarding the optimal dosing levels. If the spatial variance in the pressure and flow fields in hemodynamic dysfunction could be determined, the confidence intervals for model-based capture and titration of therapy could be developed. In this work stereoscopic particle image velocimetry is used to obtain the velocity fields in a simplified model of a stenosis (a phantom). This thin walled compliant phantom is manufactured so that the mechanical response is equivalent to in-vivo conditions. The experimental set-up imparts a pulsatile flow on the phantom ensuring in-vivo conditions by matching both Womersley and Reynolds number. The pressure fields are obtained via integration of the Navier-Stokes equation and correlated to the upstream pressure signal. By applying the Navier-Stokes equations to the velocity field data from the PIV analysis and in contrast to typical computational fluid dynamics (CFD) there was no need to calibrate and iterate between pressure and velocity domains to satisfy a series of boundary conditions. The transient pressure domain information was obtained by the simple determination of the partial derivatives of the velocity. The pressure values were stable, smooth, in the expected order of magnitude, and intuitively correct. The results show that a pressure difference across the stenosis can be observed, and that this is related to the severity of the stenosis. This provides a pathway into stenosis diagnosis using pressure measurement equipment currently used in medical practice, after some further analysis and investigation has been carried out.

1. Introduction Cardiovascular Disease (CVD) caused 30% of all deaths in New Zealand in 2012 (Ministry of Health 2015). CVD has been estimated to cost ~2% of GDP in developed countries (Adeyi et al. 2007). Critically ill patients with hemodynamic dysfunction exhibit reduced stroke volume, compliance, and diffusion of oxygenated blood to the organs (Davies et al. 2007). This ultimately

Page 2: Understanding Arterial Pressure from PIV dataltces.dem.ist.utl.pt/lxlaser/lxlaser2016/final... · 3) 1060 1140 Young’s Modulus (N/m 2) 776 000 1 320 000 Time Period Oscillation

18th International Symposium on the Application of Laser and Imaging Techniques to Fluid Mechanics・LISBON | PORTUGAL ・JULY 4 – 7, 2016

leads to organ failure, and exacerbation of the disease state. Typical therapies for hemodynamic dysfunction in critical care include inotrope therapy (Bhattacharya et al. 2010) and fluid resuscitation (Durairaj and Schmidt 2008; Rivers et al. 2010). These therapies are currently titrated against the patient condition using independently selected doses based on indicators such as mean arterial pressure and heart rate. This lacks concordance amongst clinicians regarding the optimal dosing levels. It is possible that model-based analysis could be used to determine patient-specific optimal doses of inotropes and fluid resuscitation (Marik et al. ; Hoff et al.). However, before model-based analysis can form the basis of decision support frameworks, patient-specific models must be calibrated to aortic pressure measurements that are taken as the standard of care in most ICU settings. To date, some preliminary hemodynamic modelling has been undertaken (Chase et al. 2010; Hann et al. 2010; Hann et al. 2011; Revie et al. ; Sundaresan et al. 2010; Kamoi et al.). If the spatial variance in the pressure and flow fields in hemodynamic dysfunction could be determined, the confidence intervals for model-based capture and titration of therapy could be developed. Clinicians are able to interpret some elements of hemodynamic pressure signals to ascertain certain information about the hemodynamic dysfunction a particular patients has (Marik et al. 2011; Hoff et al. 2009; Mayer et al. 2010; Marik et al. 2009). However, the ambiguity of the location of the pressure sensor in the artery leads to variance in interpretation and confounds some efforts to standardize care. Pulse pressure and mean venous pressure are generally measured in the Intensive Care Unit (ICU), and may contain the information needed to locate and diagnose certain arterial geometry deformations. A stenosis is a buildup of plaque and inflammation that alters the internal geometry of an artery. Stenoses generally reduce internal diameter of the artery and induce local high velocity and pressure differentials (Geoghegan et al. 2013; Pielhop et al. 2012; Kung et al. 2011; Geoghegan et al. 2009). The high velocity leads to higher wall shear stress. Wall shear stress has been noted as a major source of damaged for the endothelium structure. MRI scanning is often used to diagnose, measure and locate stenosed arteries. Since, arterial pressure signals are more accessible than MRI scans, they may offer a new tool to diagnose and locate stenosis. This work will determine whether a simple linear pressure response model can locate a stenosis using only pressure signals. Pulsatile fluid flow through a compliant axisymmetric stenosis model was determined using particle image velocimetry (PIV). Pulsatile pressure fields were obtained using velocity field data via integration of the Navier-Stokes equation and correlated to the upstream pressure signal to observe whether the stenosis affected the pressure in a distinct manner from the healthy regions of the artery.

Page 3: Understanding Arterial Pressure from PIV dataltces.dem.ist.utl.pt/lxlaser/lxlaser2016/final... · 3) 1060 1140 Young’s Modulus (N/m 2) 776 000 1 320 000 Time Period Oscillation

18th International Symposium on the Application of Laser and Imaging Techniques to Fluid Mechanics・LISBON | PORTUGAL ・JULY 4 – 7, 2016

2. Experimental Methodology The thin walled compliant flow phantom (Fig. 1) was produced by the method developed in Geoghegan et al. (2012) and was constructed at a scale of 3.2 times life size from silicone (Dow Corning Sylgard 184). The main properties of the phantom (with a symmetric stenosis of 50% by diameter) along with that of the common carotid artery (CCA) are given in Tab. 1. The phantom internal geometry and the region of interest (ROI) studied is also shown. The physiological flow wave was obtained from phase contrast magnetic resonance imaging (MRI) measurements of the time dependent in-vivo flow rate in the CCA of a healthy male volunteer (Buchmann and Jermy 2010) with maximum, minimum and mean Re of 939, 379 and 632, respectively and a Womersley number α = 4.54 (Fig. 2). The phantom was ported to a previously developed flow system (Geoghegan et al. 2013; Geoghegan et al. 2010). The working fluid was an aqueous glycerine solution of 39:61 ratio by weight, which provided a refractive index that matched the silicone phantom (n=1.141) (Geoghegan et al. 2012). The fluid had a kinematic viscosity (ν) of 10.2×10-6 m2s-1 and a density (ρ) of 1150 kgm-3 at 20°C. The physiological flow wave was produced by means of a piston pump driven by a stepper motor attached to a ball screw. An in-house National Instruments LabVIEW code controlled the piston system with an electromagnetic flow meter (Krohne Optiflux 1300) providing a real-time feedback of flow rate to the CPU. A pressure transducer was located upstream to record the inlet pressure variation.

Fig. 1 (left) Silicone flow phantom representing the common carotid artery with symmetric stenosis. (right) Magnification of the

flow phantom ROI with a 50 % symmetric stenosis (Geoghegan et al. 2012)

Page 4: Understanding Arterial Pressure from PIV dataltces.dem.ist.utl.pt/lxlaser/lxlaser2016/final... · 3) 1060 1140 Young’s Modulus (N/m 2) 776 000 1 320 000 Time Period Oscillation

18th International Symposium on the Application of Laser and Imaging Techniques to Fluid Mechanics・LISBON | PORTUGAL ・JULY 4 – 7, 2016

Tab. 1 Typical CCA (Buchmann 2010; Riley et al. 1992) and flow phantom properties

Property CCA Flow Phantom

Wall Thickness (m) 0.000617 0.00128 Internal Diameter (m) 0.0062 0.02 Density(blood, glycerol) (kg/m3) 1060 1140

Young’s Modulus (N/m2) 776 000 1 320 000 Time Period Oscillation (s) 1 2.92 Propagation wavelength (m) 8.25 27.26

Length (m) 0.1 0.33 Symmetric Stenosis (% by diameter) N/A 50%

Fig. 2 Phase averaged in-vitro inlet physiological flow waveform (Buchmann 2010)

Images were recorded using two Dantec Flowsense 2MP cameras with a CCD array of 1608×1208pixel2. The 60mm Nikon lenses had an aperture size set to f#=8 and a magnification of 0.2 to 0.15. This gives an estimated depth of field of 3 to 4.9mm and a diffraction limited particle size of 12.5 to 11.9µm. The flow was illuminated with a 40mm high, 0.5mm thick light sheet from a New Wave Solo 120 XT laser with a wavelength of 532nm at 120mJ/pulse using a series of spherical and cylindrical lenses. The flow was seeded with nominally 10 µm diameter hollow glass spheres with a density of 1.1g/cm3. The PIV measurements were phase locked, recording 22 image pairs per waveform over 50 waveforms. For time oscillating flow waves the laser pulse time delay selection is a compromise due to the difference in maximum pixel displacement for maximum and minimum flow rate. For this analysis a time delay of 650µs was selected, which produced <3% invalid vectors at each phase. After minimum background subtraction, intensity normalisation

Page 5: Understanding Arterial Pressure from PIV dataltces.dem.ist.utl.pt/lxlaser/lxlaser2016/final... · 3) 1060 1140 Young’s Modulus (N/m 2) 776 000 1 320 000 Time Period Oscillation

18th International Symposium on the Application of Laser and Imaging Techniques to Fluid Mechanics・LISBON | PORTUGAL ・JULY 4 – 7, 2016

and masking, the images were processed with an in-house code implementing 2D FFT cross-correlation (Buchmann and Jermy 2007) with iterative window sizing and displacement. The data was validated using the signal-to-mean ratio filter and normalised median test (Westerweel and Scarano 2005). The internal pressure derivative fields were determined using simplified cylindrical Navier stokes equations assuming full rotational symmetry. 𝜕𝑝𝜕𝑟 = −𝜇

1𝑟𝜕𝜕𝑟 𝑟

𝜕𝑢)𝜕𝑟 +

𝜕+𝑢)𝜕𝑧+ −

𝑢)𝑟+ − 𝜌

𝜕𝑢)𝜕𝑡 + 𝑢)

𝜕𝑢)𝜕𝑟 + 𝑢/

𝜕𝑢)𝜕𝑧 (1)

𝜕𝑝𝜕𝜙

= 0 (2) 𝜕𝑝𝜕𝑧 = −𝜇

1𝑟𝜕𝜕𝑟 𝑟

𝜕𝑢/𝜕𝑟 +

𝜕+𝑢/𝜕𝑧+ − 𝜌

𝜕𝑢/𝜕𝑡 + 𝑢)

𝜕𝑢/𝜕𝑟 + 𝑢/

𝜕𝑢/𝜕𝑧 (3)

where p is the local pressure; r and z are the radial and axial directions, respectively; 𝑢2 is the velocity component in the 𝑖45 direction; 𝜌 is the density; 𝜇 is the dynamic viscosity. The pressure at some location k in the field can thus be calculated by integrating the Navier Stokes equation with respect to the radial and axial directions:

𝑝 𝑟6, 𝑧6 = 𝑝 𝑟8, 𝑧8 +𝜕𝑝𝜕𝑟 𝑑𝑟

):

);+

𝜕𝑝𝜕𝑧 𝑑𝑧

/:

/; (4)

𝑝 𝑟6, 𝑧6 = 𝑝 𝑟8, 𝑧8 + −𝜇1𝑟𝜕𝜕𝑟 𝑟

𝜕𝑢)𝜕𝑟 +

𝜕+𝑢)𝜕𝑧+ −

𝑢)𝑟+ − 𝜌

𝜕𝑢)𝜕𝑡 + 𝑢)

𝜕𝑢)𝜕𝑟 + 𝑢/

𝜕𝑢)𝜕𝑧 𝑑𝑟+. . .

):

);

+ −𝜇1𝑟𝜕𝜕𝑟 𝑟

𝜕𝑢/𝜕𝑟 +

𝜕+𝑢/𝜕𝑧+ − 𝜌

𝜕𝑢/𝜕𝑡 + 𝑢)

𝜕𝑢/𝜕𝑟 + 𝑢/

𝜕𝑢/𝜕𝑧 𝑑𝑧

/:

/;

(5) If 𝑝 𝑟8, 𝑧8 is declared to be at the wall beside the inlet, the gauge pressure at this point can be determined by linking the local pressure to the distension of the elastic wall:

𝑝 𝑟8, 𝑧8 =2𝑟8 − 𝐷?𝐷?

2𝐸ℎ𝐷?BC (6)

where 𝑟8is the distended radius of the phantom at z=0, E is the Young’s modulus of the phantom wall, h is the wall thickness, DN is the natural internal diameter of the unstretched phantom.

Page 6: Understanding Arterial Pressure from PIV dataltces.dem.ist.utl.pt/lxlaser/lxlaser2016/final... · 3) 1060 1140 Young’s Modulus (N/m 2) 776 000 1 320 000 Time Period Oscillation

18th International Symposium on the Application of Laser and Imaging Techniques to Fluid Mechanics・LISBON | PORTUGAL ・JULY 4 – 7, 2016

Analysis was performed using MATLAB (Mathworks, Natick, MA, United States of America). The derivatives were calculated using the function ‘gradient.m’. Integrals were calculated using the trapezium rule within the function ‘cumtrapz.m’. The trapezium rule effectively utilizes the noise mitigation effects of integrals to remove some of the sharp behaviors that occur due to the differentiation of noisy experimental data. Hence, there was no need to using velocity smoothing routines to mitigate the effect of noise on the outcome. 4. Results and Discussion The instantaneous velocity fields and in-plane streamlines at selected points in the flow waveform through the stenosis throat is shown in Fig. 3. Between -2 ≤ x/D ≤ -1 there was an increase in centerline (y/D = 0) velocity due to the decrease in cross sectional area of the phantom geometry. As flow passes through the stenosis throat two distinct flow regimes were observed; a central high velocity jet, and a low velocity recirculation region either side of the jet. These two flow patterns were separated by a shear layer. The peak flow velocity (1.9 m/s) observed is 2.6 times larger than the 0.72 m/s previously observed in a straight (i.e. unstenosed, healthy) geometry (Geoghegan et al. 2016).

Fig. 3 Instantaneous absolute velocity fields (m/s) and velocity vectors at selected phase locations

The velocity fields were combined with the Navier-Stokes equations in their cylindrical form (Eq.1-3) to calculate the pressure profiles. The corresponding phase locations for the data in Fig. 3 are shown in Fig. 4. The pressure profiles are cyclic due to the cyclic experimental input that was calibrated to achieve a particular flow wave. The history of input pressure and pressure at selected

Page 7: Understanding Arterial Pressure from PIV dataltces.dem.ist.utl.pt/lxlaser/lxlaser2016/final... · 3) 1060 1140 Young’s Modulus (N/m 2) 776 000 1 320 000 Time Period Oscillation

18th International Symposium on the Application of Laser and Imaging Techniques to Fluid Mechanics・LISBON | PORTUGAL ・JULY 4 – 7, 2016

locations are shown in Fig. 5. p4 and p5 lie in the high velocity regions post-stenosis jet. These pressures are identical at all times i.e. there is no change in pressure over a distance of one diameter downstream from the stenosis throat. p1 and p2 lie in the low velocity regions near the wall, outside the jet, one upstream and one downstream of the stenosis. These pressure histories match each other, but for a slight difference in amplitude. The pressure history on the centerline one diameter upstream of the stenosis (p3) is very similar to that at p1 and p2, with a close match to p2. In a complaint, straight, healthy artery, there will be no pressure variation axially, except for phase lag, which is minimal over the distances experienced in the arterial system. From the results shown here, this is not the case with a stenosed artery. With the use of a pressure transducer attached to a catheter, a stenosis could potentially be detected by pressure drop alone. The exact relationship between pressure difference and stenosis severity will require an extensive analysis of different artery shapes and stenosis to provide the sensitivity required for pressure drop.

Fig. 4 Pressure (kPa) assuming radial symmetry at selected phase locations (Note the non-zero scale limit – Blue=~5kPa)

Page 8: Understanding Arterial Pressure from PIV dataltces.dem.ist.utl.pt/lxlaser/lxlaser2016/final... · 3) 1060 1140 Young’s Modulus (N/m 2) 776 000 1 320 000 Time Period Oscillation

18th International Symposium on the Application of Laser and Imaging Techniques to Fluid Mechanics・LISBON | PORTUGAL ・JULY 4 – 7, 2016

Fig. 5 The relationship between pressure input-output at selected points in the stenosis. P1-3 follow a similar profile and P4-P5

completely match

By applying the Navier-Stokes equations to the velocity field data from the PIV analysis and in contrast to typical computational fluid dynamics (CFD) there was no need to calibrate and iterate between pressure and velocity domains to satisfy a series of boundary conditions. The transient pressure domain information was obtained by the simple determination of the partial derivatives of the velocity. The pressure values were stable, smooth, in the expected order of magnitude, and intuitively correct. This analysis used an experimental input flow rate that was indicative of the flow wave observed in the carotid artery (Geoghegan et al. 2013). However, to achieve this flow wave in the experimental setup, the input pressure wave did not closely match pressure curves observed in humans, thus indicating complex upstream flow-geometry behaviors were not replicated in this experimental setup. However, this experimental setup rendered an accurate flow profile at the stenosis, and thus achieved the original goals of the study. Future validation should incorporate more physiological aortic structures so that the upstream pressure is transformed in a more physiological way when it reaches the stenosis. This work was performed on a straight axis-symmetric geometry, which is representative of many arterial geometries. However, other stenosis geometries can include curvature or asymmetry. For such geometries, the method would require a different implementation of the Navier-Stokes analysis. 5. Conclusions

Page 9: Understanding Arterial Pressure from PIV dataltces.dem.ist.utl.pt/lxlaser/lxlaser2016/final... · 3) 1060 1140 Young’s Modulus (N/m 2) 776 000 1 320 000 Time Period Oscillation

18th International Symposium on the Application of Laser and Imaging Techniques to Fluid Mechanics・LISBON | PORTUGAL ・JULY 4 – 7, 2016

This work obtained pressure fields from PIV experimental results by integrating the cylindrical form the Navier-Stokes equations. The results show that a pressure difference across the stenosis can be observed, which provides a pathway into stenosis diagnosis using current medical techniques with further analysis and investigation. A regressive cross-correlation of the input-output signals will be performed to determine if pressure response through the stenosis can be used as a marker of stenosis location. Further validation in a linear geometry is required to ascertain the effect of stenosis severity and stenosis wall angle on the pressure field. If flow separation downstream of the stenosis does not occur there would be a different pressure distribution downstream of the stenosis affecting data analysis. The assumptions used in the cylindrical form of the Navier-Stokes equations applied here are only applicable in axis-symmetric geometries. Further development of the analysis is required to provide a pressure analysis of more complex geometries. 6. References Adeyi O, Smith O, Robles S (2007) Public policy and the challenge of chronic noncommunicable

diseases. World Bank Publications, Bhattacharya S, Abebe K, Simon M, Saba S, Adelstein E (2010) Role of Cardiac Resynchronization

in End-Stage Heart Failure Patients Requiring Inotrope Therapy. Journal of Cardiac Failure 16 (12):931-937. doi:http://dx.doi.org/10.1016/j.cardfail.2010.07.253

Buchmann N (2010) Development of Particle Image Velocimetry for In Vitro Studies of Arterial Haemodynamics. University of Canterbury, Christchurch, New Zealand

Buchmann NA, Jermy MC Particle image velocimetry measurements of blood flow in a modeled carotid artery bifurcation. In: 16th Australasian Fluid Mechanics Conference, Gold Coast, Australia, 2007.

Chase JG, Starfinger C, Hann CE, Revie JA, Stevenson D, Shaw GM, Desaive T (2010) Model-based Prediction of the Patient-specific response to Adrenaline. Open Med Inform J 4:149-163

Davies JE, Hadjiloizou N, Leibovich D, Malaweera A, Alastruey-Arimon J, Whinnett ZI, Manisty CH, Francis DP, Aguado-Sierra J, Foale Ra, Malik IS, Parker KH, Mayet J, Hughes AD (2007) Importance of the aortic reservoir in determining the shape of the arterial pressure waveform – The forgotten lessons of Frank. Artery Research 1 (2):40-45. doi:10.1016/j.artres.2007.08.001

Durairaj L, Schmidt GA (2008) Fluid Therapy in Resuscitated Sepsis*Less Is More. CHEST Journal 133 (1):252-263. doi:10.1378/chest.07-1496

Geoghegan PH, Buchmann N, Jermy M, Nobes D, Spence C, Docherty PD (2010) SPIV and image correlation measurements of surface displacement during pulsatile flow in models of compliant, healthy and stenosed arteries. Paper presented at the 15th International Symposium of Laser Techniques to Fluid Mechanics, Lisbon, Portugal, 5th-8th July

Geoghegan PH, Buchmann NA, Soria J, Jermy MC (2013) Time-resolved PIV measurements of the flow field in a stenosed, compliant arterial model. Experiments in Fluids 54 (5):1-19. doi:10.1007/s00348-013-1528-0

Geoghegan PH, Buchmann NA, Spence CJT, Moore S, Jermy M (2012) Fabrication of rigid and flexible refractive-index-matched flow phantoms for flow visualisation and optical flow measurements. Experiments in Fluids 52 (5):1331-1347. doi:10.1007/s00348-011-1258-0

Page 10: Understanding Arterial Pressure from PIV dataltces.dem.ist.utl.pt/lxlaser/lxlaser2016/final... · 3) 1060 1140 Young’s Modulus (N/m 2) 776 000 1 320 000 Time Period Oscillation

18th International Symposium on the Application of Laser and Imaging Techniques to Fluid Mechanics・LISBON | PORTUGAL ・JULY 4 – 7, 2016

Geoghegan PH, Jermy MC, Buchmann NA, Spence CJ, Freitag T (2009) Experimental investigation

of flow in a compliant tube using particle image velocimetry. Paper presented at the 8th International Symposium on Particle Image Velocimetry, Melbourne Australia,

Geoghegan PH, Nobes DS, Jermy MC (2016) A PIV comparison of the flow field and wall shear stress in rigid and compliant models of healthy carotid arteries. Journal of Mechanics in Medicine and Biology Accepted awaiting publication

Hann CE, Chase JG, Desaive T, Froissart CB, Revie J, Stevenson D, Lambermont B, Ghuysen A, Kolh P, Shaw GM (2010) Unique parameter identification for cardiac diagnosis in critical care using minimal data sets. Computer Methods and Programs in Biomedicine 99 (1):75-87. doi:10.1016/j.cmpb.2010.01.002

Hann CE, Revie J, Stevenson D, Heldmann S, Desaive T, Froissart CB, Lambermont B, Ghuysen A, Kolh P, Shaw GM, Chase JG (2011) Patient specific identification of the cardiac driver function in a cardiovascular system model. Computer Methods and Programs in Biomedicine 101 (2):201-207. doi:10.1016/j.cmpb.2010.06.005

Hoff R, Rinkel G, Verweij B, Algra A, Kalkman C (2009) Blood Volume Measurement to Guide Fluid Therapy After Aneurysmal Subarachnoid Hemorrhage. Stroke 40 (7):2575-2577. doi:10.1161/strokeaha.108.538116

Kamoi S, Pretty C, Docherty PD, Squire D, Revie J, Chiew YS, Desaive T, Shaw GM, Chase JG (2014) Continuous Stroke Volume Estimation from Aortic Pressure Using Zero Dimensional Cardiovascular Model: Proof of Concept Study from Porcine Experiments. PLoS ONE 9 (7):e102476. doi:10.1371/journal.pone.0102476

Kung E, Les A, Figueroa C, Medina F, Arcaute K, Wicker R, McConnell M, Taylor C (2011) In Vitro Validation of Finite Element Analysis of Blood Flow in Deformable Models. Annals of Biomedical Engineering 39 (7):1947-1960. doi:10.1007/s10439-011-0284-7

Marik P, Monnet X, Teboul J-L (2011) Hemodynamic parameters to guide fluid therapy. Annals of Intensive Care 1 (1):1-9. doi:10.1186/2110-5820-1-1

Marik PE, Cavallazzi R, Vasu T, Hirani A (2009) Dynamic changes in arterial waveform derived variables and fluid responsiveness in mechanically ventilated patients: A systematic review of the literature *. Critical Care Medicine 37 (9):2642-2647

Mayer J, Boldt J, Mengistu AM, Röhm KD, Suttner S (2010) Goal-directed intraoperative therapy based on autocalibrated arterial pressure waveform analysis reduces hospital stay in high-risk surgical patients: a randomized, controlled trial. BioMed Central Ltd.,

Ministry of Health (2015) Mortality and Demographic data 2012. Ministry of Health, Wellington Pielhop K, Klaas M, Schröder W (2012) Analysis of the unsteady flow in an elastic stenotic vessel.

European Journal of Mechanics - B/Fluids (0). doi:10.1016/j.euromechflu.2012.01.010 Revie JA, Stevenson DJ, Chase JG, Hann CE, Lambermont BC, Ghuysen A, Kolh P, Shaw GM,

Heldmann S, Desaive T (2012) Validation of subject-specific cardiovascular system models from porcine measurements. Computer Methods and Programs in Biomedicine (in press). doi:10.1016/j.cmpb.2011.10.013

Riley W, Barnes R, Evans G, Burke G (1992) Ultrasonic measurement of the elastic modulus of the common carotid artery. The Atherosclerosis Risk in Communities (ARIC) Study. Stroke 23 (7):952-956

Rivers EP, Jaehne AK, Eichhorn-Wharry L, Brown S, Amponsah D (2010) Fluid therapy in septic shock. Current Opinion in Critical Care 16 (4):297-308

Sundaresan A, Chase JG, Hann CE, Shaw GM (2010) Cardiac output estimation using pulmonary mechanics in mechanically ventilated patients. Biomedical engineering online 9:80. doi:10.1186/1475-925X-9-80

Westerweel J, Scarano F (2005) Universal outlier detection for PIV data. Experiments in Fluids 39 (6):1096-1100