DxC 600i, Customer, Daily Start Up Job Aids Page 1 of 30 Version C, 061812 Brea Training Center UniCel ® DxC 600i Daily Start Up For Training Purposes Only The job aids included here are shortened versions of the procedures. Please refer to the sources listed below for the complete detailed procedures and all precautions. Information included here is correct as of the date published. Always verify you have current information before proceeding. Source: UniCel ® DxC 600i Synchron ® Access Clinical Systems Instructions for Use PN A93695 Version AA, October 2010 UniCel ® DxC Synchron ® Clinical Systems Instructions for Use PN A93719 Version AA, October 2010 Access 2 Help Version i6.0.05

Upload: others

Post on 05-Jul-2020

5 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 1 of 30 Version C, 061812 Brea Training Center

UniCel® DxC 600i Daily Start Up

For Training Purposes Only The job aids included here are shortened versions of the procedures. Please refer to the sources listed below for the complete detailed procedures and all precautions. Information included here is correct as of the date published. Always verify you have current information before proceeding.

Source:

UniCel® DxC 600i Synchron® Access Clinical Systems Instructions for Use

PN A93695 Version AA, October 2010

UniCel® DxC Synchron® Clinical Systems

Instructions for Use PN A93719

Version AA, October 2010

Access 2 Help Version i6.0.05

Page 2: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 2 of 30 Version C, 061812 Brea Training Center

Page 3: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 3 of 30 Version C, 061812 Brea Training Center

DxC 600i Daily Start Up

Print/View DxC 600i Supply, Reagent, and Calibration StatusPage 4

Load Access SuppliesPages 5-11

Perform Daily Maintenance procedures (Pages 12-16):

~ Check System StatusCheck System BackupCheck Access Zone Temperatures

~ Inspect Fluidics~ Clean Wash Carousel Probe Exteriors~ Prime Substrate

Load CTA SuppliesPages 17-20

Load MC ReagentsLoad/Unload CC Reagents

Pages 21-23

Process Daily Clean SystemRoutine on Access

Page 24

Calibrate DxC chemistriesPage 26

Clear DxC RacksPage 28

Process Quality ControlPages 29 & 30

Calibrate Access assaysPage 27

Access Supplies

Required?

CTA Supplies

Required?

DxC Calibration required?

Access Calibration required?

No

Yes

Yes

Yes

No

No

No

No

DxC Reagents Required?

Yes

Yes

Initial and Date Electronic Maintenance LogPage 25

Page 4: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 4 of 30 Version C, 061812 Brea Training Center

Print DxC 600i Supply, Reagent and Calibration Status

Page 5: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 5 of 30 Version C, 061812 Brea Training Center

Change Access Wash Buffer

Page 6: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 6 of 30 Version C, 061812 Brea Training Center

Empty Access Liquid Waste

Page 7: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 7 of 30 Version C, 061812 Brea Training Center

Change Substrate

Page 8: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 8 of 30 Version C, 061812 Brea Training Center

Load Access RVs

Select the <RV> button from the Access monitor

Open the RV Supply and RV Load doors

Press the RVs onto the spine, then removethe protective covering from the RV cartridge

Close the RV Load and RV Supply doors

When the screen prompts you to load the RVs, place the RV cartridge in the RV supply area.

Press down firmly on the RV load door to release the RVs from the spine

Open the RV load door and remove the spine. Verify that all vessels are present, standing

upright, and not protruding upward

Select Load RVs [F4]

Select Done [F1]

Reference: DxC 600i IFUChapter 2, Pre-Test ActivitiesLoad the Reaction Vessel (RV) Cartridges onto Access Analyzer

Note: If quantity of RVs is <28,

Access must be in READY

mode to load RVs

Page 9: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 9 of 30 Version C, 061812 Brea Training Center

Change Access Reaction Vessel Waste Bag

Page 10: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 10 of 30 Version C, 061812 Brea Training Center

Load Access Reagent

Select Main Menu [F9] or the Menu tab from the Access monitor

Mix the new unpunctured reagent pack by gently inverting to remove any particles

from the top of the pack. (Do not invert a punctured pack)

Wait for the Load Reagent Pack window to display

Close the reagent carousel door and sample wheel cover

Open the sample wheel cover and reagent carousel door

Insert the reagent pack at a downward angle and press down until it snaps into place

Select Supplies [F3]

Select Done [F1]

Select Load Reagent Pack [F1]

Scan the reagent pack bar code with the hand held bar code reader. Verify the system displays the reagent information on the Access monitor

Reference: DxC 600i IFUChapter 2, Pre-Test ActivitiesLoad a Reagent Pack onto Access Analyzer

Warning: Do NOT move a

partially used reagent pack

from one Access system

to another. It will be

inventoried as a full pack and

inaccurate results may

occur.

Page 11: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 11 of 30 Version C, 061812 Brea Training Center

Unload Access Reagent

Page 12: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 12 of 30 Version C, 061812 Brea Training Center

Check Access System Backup

Page 13: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 13 of 30 Version C, 061812 Brea Training Center

Check Access Zone Temperatures

Select Main Menu [F9] or the Menu tab from the Access monitor

If the cover has recently been opened, wait for the temperature to stabilize. If

temperature remains out of range contact Technical Support

Select Maintenance Review [F6]

Zone temperatures OK?

Yes

No

Procedure complete

Reference: Help button (Access monitor)Help Topics (“Contents” tab)Operator’s ProceduresRoutine MaintenanceDaily MaintenanceChecking the System Status

Page 14: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 14 of 30 Version C, 061812 Brea Training Center

Inspect Access Fluidics

Page 15: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 15 of 30 Version C, 061812 Brea Training Center

Clean Wash Carousel Probe Exteriors

Page 16: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 16 of 30 Version C, 061812 Brea Training Center

Prime Substrate

Page 17: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 17 of 30 Version C, 061812 Brea Training Center

Load AVs at the CTA

Page 18: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 18 of 30 Version C, 061812 Brea Training Center

Change CTA AV Waste Container

Page 19: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 19 of 30 Version C, 061812 Brea Training Center

Change the CTA Wash Buffer

Page 20: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 20 of 30 Version C, 061812 Brea Training Center

Change the CTA Auto-Gloss

Page 21: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 21 of 30 Version C, 061812 Brea Training Center

Load MC Reagent

Select Rgts/Cal from the menu bar

Open the MC reagent compartment door

Use the hand held bar code reader to scan the bar code label on the new reagent bottle/box.

OR

Select reagent to manually loadSelect Clear [F1]Using the picture at the right asa reference, enter the following:

Lot numberSerial numberExpiration date

Select the Tab key on the keyboard to save the entries

Verify the new reagent information is displayed on the monitor.

Select MC Rgts

Select Load [F1]

Select Done [F10]

Close the MC reagent compartment door

Prepare reagent(s) if required

Remove the cap and inlet tubing from the existing reagent and place it in the new reagent. Place the new bottle into the

reagent storage area. Discard the old bottle according to laboratory procedures

Note: Take care not to contaminate the inlet tubing.

Reference: DxC 600i IFUChapter 2, Pre-Test ActivitiesCheck and Load Supplies at DxC AnalyzerLoad/Unload MC Reagents onto DxC Analyzer

Select DxC tab

Page 22: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 22 of 30 Version C, 061812 Brea Training Center

Load CC Reagent

Page 23: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 23 of 30 Version C, 061812 Brea Training Center

Unload CC Reagent Select Rgts/Cal from the DxC monitor

Page down to locate the position(s) of the reagent(s) tobe unloaded on the CC carousel.

Note: CC reagents are displayed on pages 3-8 of the DxC tab, Rgt/Cal screen

Open the Reagent Carousel door

Procedure Complete

Grasp the cartridge in the load position (located at the blinking bar code reader) and move it in front of the bar

code reader beam. Verify the reagent information is removed from the DxC monitor. Remove the cartridge

Close the reagent carousel door

Additional reagents selected?

Yes

No

Select Load [F1]

Select the DxC tab

Select all positions to unload reagent cartridges

Reference: DxC 600i IFUChapter 2, Pre-Test ActivitiesLoad/Unload CC Cartridges at DxC Analyzer

Page 24: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 24 of 30 Version C, 061812 Brea Training Center

Process Daily Clean System Routine

Page 25: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 25 of 30 Version C, 061812 Brea Training Center

Initial and Date Electronic Maintenance Log Select Utils from the DxC monitor

Enter your initials

View super scripts for each task performed to determine further action required

Select OK

Select 2 Maintenance

Select Log [F2]

Reference: DxC 600i IFUChapter 6, Maintenance ProceduresElectronic Maintenance LogAccessing the Scheduled Log Feature

Select the desired frequency tab

Select individual tasks in the Initial and Date column or click on the Select All button

Select the Initial and Date button

Page 26: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 26 of 30 Version C, 061812 Brea Training Center

Calibrate DxC Chemistries

Page 27: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center

Calibrate Access Assays

No

Yes

Select Main Menu [F9] or the Menu tab from the Access monitor

When the screen prompts you to load the rack, place the rack in the Access sample carousel, and

close the sample wheel cover

Use the load list displayed on the Access screen to place the appropriate calibrators in AVs and load

in the correct positions on the rack. Note: AV dead volume = 100uL

Select Sample Manager [F1]

Procedure Complete

Select Test Request [F3]

Select Request Calibration [F6]

*Verify the calibrator lot number in use

matches the calibrator lot number

selected.

Select calibrator set*

When Calibration is complete view calibration status on Rgt/Cal screen

Calibration successful?

Consult references for troubleshooting

assistance

Select the Rack ID field, then type the rack number in the field and select <ENTER>. Verify the new rack

is displayed in the Off-Board list

Select OK [F1]

Select Load Rack [F1]

Select <RUN>

Select Done [F1]

Calibration successful?

Reference: Help button (Access monitor)Help Topics (“Contents” tab)Operator’s ProceduresSample ManagementCalibration Test RequestsEntering Calibration Test Requests

Page 28: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 28 of 30 Version C, 061812 Brea Training Center

Clear DxC Racks

Select Samples from the DxC monitor

Select Clear [F7]

Select OK

Highlight the Racks Field. Type the range of rack number to be cleared*. Select ENTER.

Note: Enter 1-999 to clear all racks

Select OK to confirm

*Calibrator rack assignments and racks in progress are not cleared with this step.

Reference: DxC 600/800 IFUChapter 6, Sample Programming and ProcessingClear SamplesProcedure for Clearing Samples

Page 29: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 29 of 30 Version C, 061812 Brea Training Center

Process Quality Control using Auto Generation

Page 30: UniCel DxC 600i Daily Start Up - Cornerstone OnDemand · DxC 600i, Customer, Daily Start Up Job Aids Page 27 of 30 Version C, 061812 Brea Training Center Calibrate Access Assays No

DxC 600i, Customer, Daily Start Up Job Aids Page 30 of 30 Version C, 061812 Brea Training Center

Manually Program Quality Control