unified-drug-analysis-pittcon-2016
TRANSCRIPT
![Page 1: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/1.jpg)
Unified Drug Testing by Online SPE-LC/MS/MSFocus on Productivity Achieved Through Ease of Use:
One totally automated method measures ALL the drugs in urine and/or oral Fluids
Mark Hayward,2 Rick Youngblood,2 Kim Gamble,2 Martin Johnson,1 and Matthew T. Hardison1
1Assurance Scientific Laboratories, 727 Memorial Dr. Suite 103, Bessemer AL 35022
2ITSP Solutions Inc., 10 South Carolina St., Hartwell GA 30643
SPE cartridge
SyringePositive pressure micro scale SPE
Automated like this!
![Page 2: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/2.jpg)
Measurement of drugs of abuse in urine and/or oral fluids (OF) is common
• Pre-employment screening• DOT / federal mandated testing• Law enforcement• Compliance / diagnostic determinations by physicians
– Latter two are growing rapidly!• Fastest-growing & preferred approach is LC/MS/MS
– Due to high degree of certainty for simultaneous identification and concentration determination
• Continued growth in the use of LC/MS/MS for the measurement of drugs of abuse in urine and OF seems certain
![Page 3: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/3.jpg)
Still, there are several technical challenges that need to be met
• Easily measure low dose drugs at/near 1 ng/g– For medical purposes [Pesce, et. al. 2012 AACC conference] as well
as for zero tolerance testing• Simplicity for performing measurements with lab technicians
– Automation: load samples / run list (prioritized queuing for STAT samples)
– All LC/MS/MS peaks sufficiently intense for automatic integration• Ability to achieve high productivity for all work
– Needs apply to ALL testing! [not just high volume tests]– Investing 90% effort in 10% of tests is not acceptable
• Minimizing the labor and number of workflows required– Simplicity has always been crucial– As volume and number of tests continues to grow, simplicity grows
in importance!
![Page 4: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/4.jpg)
Pain Management Drug Monitoring• Required to prevent abuse, addiction, diversion,
mortality and morbidity (urine drug testing)• Nevertheless, must meet patient needs first (and not
penalize low, irregular dosing)• Production environment: assembly line • Needs to be easy, robust, and have low labor
requirements to measure all relevant drugs at all relevant concentrations
• Needs sufficient capacity relative to capital investment: ≥100 reimbursable reports per day per LC/MS/MS
How does this impact one's approach toward the measurement methodology?
![Page 5: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/5.jpg)
Measurement range (defining the challenge):low single digit ng/g for some opioids and benzos
Pesce, et. al.
2012 AACC conference
Dilute and shoot is insufficientMeasurement at these levels usually requires some
pre-concentration of the sample SPE, LLEAt the same time, this needs to be easy! automation
![Page 6: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/6.jpg)
Approach chosen: SPE performed with LC/MS/MS autosampler
• Automated (serial) pre-concentration of samples, so that all drugs can be measured in one method
• Remove salts, proteins, and cells for robust LC/MS/MS operation– Very little high LogP (greasy) interferences– Primary role is to filter, pre-concentrate, and remove salts / other H2O soluble
interferences– Reverse phase should be sufficient
• Modest capital investment: must buy LC autosampler regardless choose one that does more of the work CTC/PAL ITSP
ITSP = Instrument Top Sample Prep
![Page 7: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/7.jpg)
What is ITSP?Micro-SPE
performed by a CTC/PAL
ITSP SPE cartridge10-45 mg sorbent
10 mg most common and has 32 l internal volume
ITSP SPE cartridgebeing discarded after use
ITSP SPE: overall system, AKA your autosampler!
Photos: Assurance Scientific Laboratories
Precise syringe driven flow
ITSP = Instrument Top Sample Prep
![Page 8: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/8.jpg)
Technology Capabilities SPE Technologies
ITSP syringe flow SPE Parallel tube and plate based SPE
Pipette based SPE (packed sorbent)
Pipette based SPE (loose sorbent, dispersive)
HPLC Pump flow SPE (2D HPLC)
Total automation YY - but separate workflow flow
from LC/MS and GC/MS [extra robots = extra $$$]
Y - but separate workflow flow from LC/MS and GC/MS [extra
robots = extra $$$]
Y - but separate workflow flow from LC/MS and GC/MS [extra
robots = extra $$$]
Y - requires additional pumps, valves, and fixed columns [extra LC
hardware = extra $$$]
On-line operation w/ LC/MS/MS Y N N N Y
On-line operation w/ GC/MS/MS Y N N N N
Accurate Flow Control Y N N N YSingle use sorbent Y Y Y Y N
All sorbents available Y Y N N NPre-concentrate
w/o dry down Y N N N Y
Fast Y N Y N YMethod development
automation Y N N N N
Off-line operation to feed multiple instruments Y Y Y Y N
Full chain of custody Y - bar code reading possible at every step
Y - with manual labeling and recording of each step N N N
Works with any MS software
Y - SW pre-embedded in all major MS brands
N - separate work flow N - separate work flow N - separate work flow Y - each MS manufacturer has some unique form of 2D HPLC
Ease of use & maintenance
Y - the PAL is a HPLC/GC autosampler & requires same skills & maintenance
Conceptually simple, but laborious and flow control is
challenging. Robots similar to pipette based SPE.
Easy to do once one has bought, set up, and learned how to use separate robotic
pipette systems ($$$)
Easy to do once one has bought, set up, and learned how to use separate robotic
pipette systems ($$$)N
Pros Automation, accuracy, & multiple tests on single system set up Many methods in literature
Easy to do once one has bought, set up, and learned how to use separate robotic
pipette systems ($$$)
Easy to do once one has bought, set up, and learned how to use separate robotic
pipette systems ($$$)Single workflow automation
Cons None Laborious and lack of accurate flow control
Limited sorbent choices and bed masses. Flow control not
proven.
Not proven to improve sample condition and no control of
flow over sorbent
Requires significant expertise and one never knows
when/how SPE column will fail (limits applications)
Comparison of SPE technologies
![Page 9: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/9.jpg)
Understanding ITSP and how it differs from other approaches
• ITSP has precise flow control allowing separations to be performed at their Van Deemter optimum velocity
• This yields separation performance that is difficult (if not impossible) to achieve with other SPE approaches
• ITSP also is a completely unique form of SPE in that it is a truly low volume device (32 l)
• This allows elution to performed precisely with l volumes (ca. 50 l ready for direct injection on LC/MS or GC/MS)
• This allows ITSP to easily pre-concentrate samples by loading 1-10 ml of sample while eluting with <100 l of solvent
• Other forms of SPE require larger volumes for elution (10 to 50x) and significant pre-concentration can only be achieved by adding a separate slow, laborious dry-down step
• No other SPE approach can achieve the precise chromatographic separations, pre-concentration of sample, robust operation, and total automation achieved simultaneously by ITSP
![Page 10: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/10.jpg)
0 1 2 3 4 5 6 7 8
80
85
90
95
100
SCX %-Recovery in 20% WaterSCX %-Recovery in 80% WaterReverse Phase %-Recovery
SPE is Chromatography!Optimized outcomes require accurate flowAre you optimized? Or just guessing?
OxycodoneVan Deemter curves
for SPE by ITSP
SCX optimum velocity = 0.37 mm/s (1.2 l/s by ITSP) with little room for error!
RP optimum velocity = 1.5 mm/s
(4x higher than SCX!)
SPE flow driven pneumatically or by vacuum cannot achieve and maintain optimal flow!
ITSP
= T
OTA
L CO
NTR
OL
OF
YOU
R SP
E! ITSP = TOTAL CO
NTRO
L OF YO
UR SPE!
Flow (l/s) [5 l/s = 1.5 mm/s]
% R
ecov
ery
![Page 11: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/11.jpg)
SPE-LC/MS/MS method developmenthelicopter view of strategy
• Focus on simplicity and minimization of steps• Prioritize hardest to measure drugs (lowest concentration,
low dose opioids and benzos) over the easy to measure drugs for recovery optimization and pre-concentration
• Focus on relative recoveries for elution pre-concentrate• Choose balanced LC/MS conditions that allow separation
and measurement of both acidic and basic drugs as well as polar and non-polar drugs (1 method, all drugs!)
• Establish linear scalability and stoichiometry in sample loading as a data driven way to establish the validation readiness of the method
• Leverage automation to achieve rapid method development and execution ITSP with the CTC/PAL
• Develop the method for urine first, then adapt to oral fluid
![Page 12: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/12.jpg)
Serially automated SPE method development Parallel testing of C18 and DVB SPE phases (3x): each step is a sample list!
Hands on view of strategy
• Test SPE cartridge wash with various solvents (3x cartridge volume) and no wash: rinse cartridge with water, load spiked urine & measure drug breakthroughs (choose wash solvent)
• Test SPE cartridge conditioning & loaded sample wash with buffers (at 3x cartridge volume): load spiked urine & measure drug breakthroughs (choose conditioning/wash buffer)
• Test SPE cartridge elution with various solvents and measure drug recoveries (choose elution solvent and measure optimal flow)
• Test SPE cartridge elution at multiple volumes with various buffers in chosen elution solvents and measure drug recoveries (pre-concentrate) while monitoring LC separation (choose elution buffer based on LC separation first, then recoveries)
• Vary sample amount over a range of at least 10x and measure drug recoveries. If linear stoichiometry is not observed, re-optimize above steps based on data. If linear, re-optimize LC/MS/MS, choose sample amount, then validate!
![Page 13: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/13.jpg)
Automated method development• Run each of the 5 lists sequentially (described in previous
slide) transferring optima measured into next list• Optima results:– Sorbent: C18 end-capped (gives higher recovery for benzos /
opioids)– Conditioning: MeOH, then NH4OAc buffer– Loading: ≥0.5% NH4OAc in sample, Flow 5 l/s– Wash: aqueous NH4OAc (≥0.5%)– Elution: 0.2% NH4OAc in MeOH, 75 l at 5 l/s (pre-
concentrates lowest concentration drugs most)– Sample load range: 100 – 500 l is linear at 0.5% NH4OAc in
sample (can be increased with higher %NH4OAc in sample)
Minimizes breakthrough
![Page 14: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/14.jpg)
Test compatibility of SPE eluent with LC separation (SPE – LC interfacing)
5x10
00.025
0.050.075
0.10.125
0.150.175
0.20.225
0.250.275
0.30.325
0.350.375
0.40.425
0.450.475
0.50.525
0.550.575
0.60.625
0.650.675
0.70.725
0.750.775
0.80.825
0.850.875
0.90.925
0.950.975
11.025
1.051.075
1.11.125
1.151.175
1.21.225
1.251.275
1.31.325
1.35
Cpd 9: Codeine: +ESI MRM Frag=140.0V [email protected] (300.0000 -> 199.0000) C18 80% ACN elution 0.175% NH4 Acetate 2ul Loop.d
1.395
1.179
Counts vs. Acquisition Time (min)0.88 0.9 0.92 0.94 0.96 0.98 1 1.02 1.04 1.06 1.08 1.1 1.12 1.14 1.16 1.18 1.2 1.22 1.24 1.26 1.28 1.3 1.32 1.34 1.36 1.38 1.4 1.42 1.44 1.46 1.48 1.5 1.52 1.54 1.56 1.58 1.6 1.62 1.64 1.66 1.68 1.7 1.72 1.74 1.76 1.78
Codeine
Hydrocodone
80% ACN
Buffered 80% ACN
Of course, chemical presentation of the sample from SPE to LC is important
Just like with SPE, control of the pH (ionization state) controls retention
Buffer: NH4OAcLC column: C18
B = ACN
Elution in 80% ACN limits LC injection volume to 2 l (2.1 x 50 mm column). Elution in 100% MeOH (buffered) allows 5 l LC injection. Viscosity has an equally important role in LC injection along with pH.
SPE eluent
![Page 15: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/15.jpg)
SPE elution volumeAllows optimization for drug classes
40 50 60 70 80 90 1000.4
0.6
0.8
1
THCA6-MAMBuprenorphineCodeineDiazepamSecobarbitalPhencyclidine
Elution volume (ul)
Nor
mal
ized
resp
onse
Favors opiates, metabolites, and other illicits
Favors opioids, benzos, barbs, and THCABest for PM
C18 SPE data shown with MeOH elutionDVB with MeOH elution favors low volume elution for all drugs
k’ = 1.5 - 2 k’ = 2 - 3
Gains in sensitivity from lower volume elution using DVB do not outweigh the absolute recoveries observed with C18 SPE
k’ > 3
Dilutes all drugs
![Page 16: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/16.jpg)
0 100 200 300 400 500 600 700 800 900 10000
0.5
1
MDMAFentanylAmphetamineDiazepamOxymorphonePentobarbitalTHCA
Sample loading: defining SPE cartridge capacity and linear working range
Linear range
Current TQs
Older TQs
500 l syringe loading sample at 5 l/s
Volume (l) of urine loaded on SPE cartridge
ROI optimumPre-concentration 1.5-3x
Nor
mal
ized
LC/M
S/M
S re
spon
se
C18 SPE, MeOH elution, 0.5% NH4OAc in sample
Linear response observed within 100 to 500 l sample load range for all PM drugsCurrent triple quads (TQs) can measure all PM drugs in the lower half of this rangeOpiates, metabolites, and other illicits saturate cartridge first at 500 to 1000 l sample loadPerformance below 100 l can be improved with smaller syringe and blowing out cartridge with air at each step (RTC) [also dilution to 200 l with PAL works nicely]ITSP cartridge volume is 32 l
Opioids and benzos are most optimized by design (recoveries >90%)
Opiates, metabolites, & illicits fully functional (recoveries 80-90%) Linear range can be
extended with higher %NH4OAc in sample
![Page 17: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/17.jpg)
Focus on the LC/MS/MS also can be productive
6x10
0
0.05
0.1
0.15
0.2
0.25
0.3
0.35
0.4
0.45
0.5
0.55
0.6
0.65
0.7
0.75
0.8
0.85
0.9
0.95
1
1.05
1.1
1.15
1.2
1.25
1.3
1.35
1.4
1.45
1.5
1.55
Cpd 99: 11-nor-9-Carboxyl-THC: +ESI MRM Frag=120.0V [email protected] (345.0000 -> 327.2000) C1850 ul Elution.d
4.40
2
Counts vs. Acquisition Time (min)0.7 0.8 0.9 1 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 2 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8 2.9 3 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 4 4.1 4.2 4.3 4.4
6x10
00.10.20.30.40.50.60.70.80.9
11.11.21.31.41.51.61.71.81.9
22.12.22.32.42.52.62.72.82.9
33.13.23.33.43.53.63.73.83.9
44.14.24.34.44.54.64.74.84.9
55.15.25.35.45.55.65.7
Cpd 96: 11-nor-9-Carboxyl-THC: +ESI MRM Frag=120.0V CF=0.000 DF=0.000 [email protected] (345.0000 -> 327.2000) 2000B.d
4.067
Counts vs. Acquisition Time (min)0.8 0.9 1 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 2 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8 2.9 3 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 4 4.1
Original LC/MS/MSIt works4.5 min
Heat column, increase flow, add gradient segmentsImprove separation where peaks are crowdedDecrease time between well separated peaks4.1 minColumn switching / conditioning also saves overhead time
![Page 18: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/18.jpg)
Acidic drugs can be measured under LC conditions used for basic drugs
4x10
00.10.20.30.40.50.60.7
0.80.9
11.11.21.31.41.51.61.71.81.9
22.1
2.22.32.42.52.62.72.82.9
33.13.23.33.43.53.6
3.73.83.9
44.14.24.34.44.54.6
Cpd 96: 11-nor-9-Carboxyl-THC: +ESI MRM Frag=120.0V CF=0.000 DF=0.000 [email protected] (345.0000 -> 327.2000) 2000B.d 1
Counts vs. Acquisition Time (min)2.4 2.45 2.5 2.55 2.6 2.65 2.7 2.75 2.8 2.85 2.9 2.95 3 3.05 3.1 3.15 3.2 3.25 3.3 3.35 3.4 3.45 3.5 3.55 3.6 3.65 3.7 3.75 3.8 3.85 3.9 3.95 4 4.05 4.1 4.15 4.2 4.25 4.3 4.35
THCA
Pentobarbital Secobarbital
![Page 19: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/19.jpg)
Workflow: minimizing cycle timePAL operation in the inject ahead mode
SPE 1 SPE 2 SPE 3LC/MS/MS 1 LC/MS/MS 2
4.5 min 4.5 min
…………
Total cycle time (SPE + LC/MS/MS) = 4.5 min
![Page 20: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/20.jpg)
Summary: Drugs in urine• An online and automated SPE-LC/MS/MS method has been
developed for pain management (PM) monitoring in urine samples• The method is scalable and can be adapted to any LC/MS/MS simply
by adjusting the volume of urine loaded for SPE• Rather than using multiple panel focused methods, this single
method is used to measure all PM related drugs• Method development focus has been on the lowest dose, hardest to
measure prescribed drugs (suggests prioritization works!)
• Rework is limited to only the highest dose drugs in the highest dosed patients (inject less, bring into linear range)
• Robust operation and a cycle time (SPE + LC/MS/MS) of 4.5 min has been achieved
• The use of ITSP with the CTC/PAL for serial automation is a very efficient way to perform SPE method development: this method required 3 lab days using 1 SPE-LC/MS/MS system
![Page 21: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/21.jpg)
Adapt method developed for urine to function equally well for oral fluid (OF)• First things first: choosing a sampling device• Many considered / criticized, primary concerns:• Sample stability• Volumetric sampling accuracy
• We added need for forensic acceptance and success when challenged at trial
• Quantisal sampling device chosen for its:• Highest volumetric accuracy (+/-10%)• Built in filtering (important: all OF samples require either
filtering or centrifugation)• >1 week sample stability• Greatest acceptance in clinical and forensic use
![Page 22: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/22.jpg)
Adapt / test method for oral fluid (OF)• SPE and sample buffering increased to 2% NH4OAc to overcome built in
Quantisal buffering • High opiate / illicit drug recovery requires formation of drug-OAc ion
pair• Sample volume increased to 1 ml to achieve 0.1 ng/g cut offs• Drug concentrations lower in OF 13x pre-concentration
• Linear response for 71 drugs achieved• Method tested with 20 multi medication / not so good prognosis patient
samples where both urine and OF collected in parallel and tested with both LC/MS/MS and immunoassay– LC/MS/MS in complete agreement for urine and OF (no false neg) and “partying”
successfully detected in one sample set– Included patients dosed 1 mg/day opioids / benzos: easily measured / auto-
integrated at approx 0.5 ng/g in OF (0.1 ng/g cutoffs)– Immunoassay showed one of its deficiencies by not being able to detect /
confirm 1 mg/day dosed drugs (urine and OF)
![Page 23: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/23.jpg)
Unified urine and oral fluid method results
• Lab time to adapt urine method to OF: 2 days• Again, leveraging the automation
• Method (urine & OF) validated to both clinical and forensic standards in multiple labs (cv 3-5%)
• Urine and OF can be measured in the same run• Still need to include blanks, calibrators, and QCs for
both• Method (urine & OF) in use for production work and
delivering 100-200 results overnight for each LC/MS/MS• High ROI readily achieved for top end LC/MS/MSs
![Page 24: Unified-Drug-Analysis-PittCon-2016](https://reader038.vdocuments.net/reader038/viewer/2022110107/5886cdd71a28abcc7d8b7bd9/html5/thumbnails/24.jpg)
What’s next? Larger range of tests ready to run with a single LC/MS/MS• We have successfully added Vit D (blood) measurement to individual
PAL systems already measuring drugs in urine / OF– These 3 tests are currently considered highest ROI in clinical lab– 2 cartridge types, 2 LC column types, all on-board, all the time
• We have drug methods for blood/serum/plasma/DPS/DBS samples to add to these systems (ion exchange SPE)– Includes DPS/DBS disk extraction and IS addition
• We are implementing SPE solvent sourcing to Fast Wash stations (up to 4) as an alternative to ink wells– Up to 8 solvents total using minimal PAL rail– Allows 2-4 weeks of solvent capacity (0.5-2 liter bottles)
• We are implementing 4 way valve cleaning solvent sourcing for DLW (multiple cleaning chemistries for multiple methods)
• This should allow 4 tests to be setup/ready to run on single PAL-LC/MS/MS systems requiring only method selection in MS software and perhaps loading color coded SPE cartridges