unified frame of reference improves inter-subject variability of ......the electrocardiography (ecg)...
TRANSCRIPT
Paukkunen et al. BioMedical Engineering OnLine (2015) 14:16 DOI 10.1186/s12938-015-0013-9
RESEARCH Open Access
Unified frame of reference improves inter-subjectvariability of seismocardiogramsMikko Paukkunen1*, Petteri Parkkila1, Raimo Kettunen2 and Raimo Sepponen1,3
* Correspondence:[email protected] of ElectricalEngineering and Automation, AaltoUniversity, P.O. BOX: FI-13340, 00076Helsinki, FinlandFull list of author information isavailable at the end of the article
Abstract
Background: Seismocardiography is the noninvasive measurement of cardiacvibrations transmitted to the chest wall by the heart during its movement. Whilemost applications for seismocardiography are based on unidirectional accelerationmeasurement, several studies have highlighted the importance of three-dimensionalmeasurements in cardiac vibration studies. One of the main challenges in usingthree-dimensional measurements in seismocardiography is the significant inter-subjectvariability of waveforms. This study investigates the feasibility of using a unified frameof reference to improve the inter-subject variability of seismocardiographic waveforms.
Methods: Three-dimensional seismocardiography signals were acquired from tenhealthy subjects to test the feasibility of the present method for improving inter-subjectvariability of three-dimensional seismocardiograms. The first frame of referencecandidate was the orientation of the line connecting the points representing mitralvalve closure and aortic valve opening in seismocardiograms. The second candidatewas the orientation of the line connecting the two most distant points in the threedimensional seismocardiogram. The unification of the frame of reference wasperformed by rotating each subject’s three-dimensional seismocardiograms so that thelines connecting the desired features were parallel between subjects.
Results: The morphology of the three-dimensional seismocardiograms varied stronglyfrom subject to subject. Fixing the frame of reference to the line connecting the MCand AO peaks enhanced the correlation between the subjects in the y axis from0.42 ± 0.30 to 0.83 ± 0.14. The mean correlation calculated from all axes increased from0.56 ± 0.26 to 0.71 ± 0.24 using the line connecting the mitral valve closure and aorticvalve opening as the frame of reference. When the line connecting the two mostdistant points was used as a frame of reference, the correlation improved to 0.60 ± 0.22.
Conclusions: The results indicate that using a unified frame of reference is a promisingmethod for improving the inter-subject variability of three-dimensional seismocardiograms.Also, it is observed that three-dimensional seismocardiograms seem to have latentinter-subject similarities, which are feasible to be revealed. Because the projections ofthe cardiac vibrations on the measurement axes differ significantly, it seems obligatoryto use three-dimensional measurements when seismocardiogram analysis is based onwaveform morphology.
Keywords: Seismocardiography, Ballistocardiography, Aaccelerometer, Three-dimensional,Inter-subject variability, Frame of reference
© 2015 Paukkunen et al.; licensee BioMed Central. This is an Open Access article distributed under the terms of the CreativeCommons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, andreproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedicationwaiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwisestated.
Paukkunen et al. BioMedical Engineering OnLine (2015) 14:16 Page 2 of 9
BackgroundSeismocardiography (SCG) is the noninvasive measurement of cardiac vibrations trans-
mitted to the chest wall by the heart during its movement [1]. The emergence of SCG
measurement can be traced back to 19th century when Gordon reported observing a
heartbeat while standing on a scale [2]. Gordon’s observation led off the development
of several measurement methods for cardiac induced vibrations such as ballistocardio-
graphy (BCG). BCG is a method in which the reaction forces acting on the whole hu-
man body are measured. Thus, SCG can be viewed as a derivative of BCG. The
emergence of SCG as a separate measurement technique dates back to 1960s when the
pioneering works of Baevsky and Bozhenko [3,4] were published. The advances in elec-
tronic accelerometer technology have resulted in multiple applications for SCG, such
as ischemic heart disease detection, gating of cardiac imaging and therapy, as well as
smartphone-based heart activity monitoring [5-9].
Most applications for SCG are based on unidirectional acceleration measurement,
usually done in the back-to-front direction. However, several studies have highlighted
the importance of three-dimensional (3-D) measurements in cardiac vibration studies
[10-13]. For example, Migeotte et al. concluded that the vibration waveforms observed
on the foot-to-head axis are poorly correlated to the magnitude of the maximum sys-
tolic force vector computed using all three axes of acceleration [10]. This finding indi-
cates that the vibrations of the heart might be projected disproportionately between
the three orthogonal axes of measurement and may thus lead to misinterpretation of
the waveforms and the underlying physiological phenomena. Another fairly recent
paper reported a method that was based on 3-D SCG vector trajectories which were
used for computing the IJ-waveform amplitudes [14]. The IJ-waveforms were proposed
to be related to cardiac output. The report found that that since the 3-D approach
takes into account the uneven distribution of vibrations between the measured axes,
the achieved results might be independent from slight misalignments of the SCG sen-
sor. This indicates that if slight misorientations of the SCG sensor are accountable, the
compensation of inter-subject variation due to physiological differences should also be
feasible.
Inter-subject variability of seismocardiograms, which hinders the deployment of
SCG in to clinical practice, is a widely observed phenomenon in SCG studies [15,16].
We hypothesize that a unified frame of reference would facilitate improvement of
inter-subject variability of seismocardiograms. The hypothesis relies on two assump-
tions. The first assumption is that asymptomatic hearts function similarly in general.
Thus, only modest inter-subject variability should be seen when measuring SCG from
healthy subjects. However, significant inter-subject differences in seismocardiograms
are continuously reported even in healthy subjects. This indicates that SCG measure-
ments are prone to variability due to other factors than differences in cardiac function.
These other factors could be anatomical differences such as the orientation of the
heart and the aortic arc or differences in the mechanical coupling of the myocardial
vibrations to the sternum. The second assumption is that 3-D acceleration measure-
ment captures the vibrations of the chest sternum in every direction. Thus, even if the
vibrations of the heart are projected differently between subjects, the use of a unified
frame of reference might straighten the vector trajectories and reveal latent similarity
in the seismocardiograms.
Paukkunen et al. BioMedical Engineering OnLine (2015) 14:16 Page 3 of 9
This study investigates the feasibility of using a unified frame of reference to improve
the inter-subject variability of SCG waveforms. To test the feasibility of the proposed
method, a data set from ten subjects is acquired and pre-processed.
MethodsEquipment
The 3-D SCG signals were measured with three orthogonally mounted accelerometers
(SCA610-C21H1A, Murata Electronics, Finland). The accelerometers and associated elec-
tronics are both described in detail in the authors’ previous work [17]. For this study, we
mounted the individual accelerometers to a smaller package than before using eyesight to
orient the sensors. Due to the mounting process, the orthogonality of the acceleration
sensors could not be guaranteed. The largest orientation error between the accelerometers
was measured between the xy plane and the z axis (5 degrees). Thus, in the worst case,
any single axis of acceleration measurement will include an error of 9% compared to
ideally orthogonal accelerometers. The acceleration sensors had a sensitivity of 2 V/g, true
DC response and rated output noise of approximately 60 μgrms in the frequency band of 1
to 50 Hz. Prior to use, the accelerometers were tested at VTI’s (now Murata Electronics)
laboratories where they were shown to have their -3 dB points at 47 Hz. The acceleration
signals were anti-alias filtered with 8th-order Bessel lowpass filters with a rated cut-off fre-
quency of 100 Hz and attenuation of at least 96 dB at 800 Hz. The filters are coupled in
the Sallen-Key topology using quad operational amplifiers (AD8630, Analog Devices,
USA). For this particular study, the gain stage of the acceleration measurement in [17] is
bypassed, resulting in a DC response and a gain of 0 dB for the acceleration measurement.
Three individual accelerometers were used because, to the best of the authors’ knowledge,
no integrated options with similar characteristics (high sensitivity, low noise, DC re-
sponse) were available during the time of implementation. In terms of accuracy, there is
no inherent benefit in using individual accelerometers compared to integrated 3-D accel-
erometers. In particular, recent research has applied integrated 3-D accelerometers with
similar noise and sensitivity performance to SCG but with no DC response [18]. A DC re-
sponse would be important in other SCG applications where the instantaneous inclination
of the sternum was of interest.
The electrocardiography (ECG) lead II was measured using a commercial wireless
dual-lead ECG system (BC-ECG2, BIOPAC Systems Inc, US). Respiratory efforts were
detect using a commercial wireless respiratory effort detection system (BN-RESP-
XDCR, BIOPAC Systems Inc, US). All data were captured using the MP150 Data
Acquisition System (BIOPAC Systems Inc, US). The acceleration signals were coupled
to the MP150 through a Universal Interface Module (UIM100C, BIOPAC Systems Inc,
US). Preprocessing of the signals was done in the Acknowledgment environment
(BIOPAC Systems Inc, US). All post-processing of the data was done in the Matlab
environment (MATLAB R2013b, Mathworks, US).
In vivo measurements
Human subject protocol
The present study was completely conducted in the premises of Aalto University,
Espoo, Finland. Ten male volunteers with an average age of 33 years (standard
Paukkunen et al. BioMedical Engineering OnLine (2015) 14:16 Page 4 of 9
deviation (SD) 8.5 years), an average weight of 79.5 kg (SD 14.9 kg) and an average
height of 175.2 cm (SD 7.1 cm) were measured as test subjects. During the measure-
ments, the subjects were at resting supine position. To minimize breathing-induced
variability, the subjects were instructed to hold their breath as long as comfortable.
The accelerometer package was positioned on the lower part of the sternum about
one centimeter above the xiphoid process using double-sided adhesive tape. The accel-
erometer package was mounted so that one accelerometer measured the acceleration in
the back-to-front (z axis), one in the right-to-left (x axis), and one in the foot-to-head
direction (y axis) (see Figure 1).
The contents of the test and the course of its events were explained individually to
the subjects before the study. In addition, a written consent was received. The study
did not contain any such intervention in the physical integrity of the test subjects, or
any other features that would require an ethical review as considered by the National
Advisory Board on Research Ethics in Finland.
Selecting the frame of reference
The rationale behind using a unified frame of reference lies in the assumption that
most hearts function similarly in general. Thus, the inter-subject differences might be
mainly due to anatomical differences and differences in the mechanical coupling of the
heart’s vibration to the sternum. Some methods that might be useful in managing the
variability of seismocardiograms have been presented in the literature. The approaches
include the division of heartbeats in to inspiration and expiration heartbeats [19] and
normalization by resampling each heartbeat to be of same length [13], for example.
Castiglioni et al., for example, proposed that the magnitude of the IJ-waveform com-
puted from 3-D seismocardiograms is independent of slight orientation differences of
Figure 1 Measurement axes. The x axis is the left-to-right direction, the y axis the foot-to-head directionand the z axis the back-to-front direction.
Paukkunen et al. BioMedical Engineering OnLine (2015) 14:16 Page 5 of 9
the SCG sensor [14]. However, the use of orientation information, which is readily ob-
tainable from 3-D SCG signal, is rarely reported.
The authors hypothesize that if some orientation-related representative feature could
be extracted from every seismocardiogram, this feature could be used to normalize the
orientation of seismocardiograms so that each seismocardiogram could be interpreted
in the same frame of reference. Thus, this approach would facilitate reducing the effect
of mechanical coupling and anatomic differences in the seismocardiograms. Two candi-
dates were selected to serve as frames of reference. The first candidate was the orienta-
tion of the line connecting the points that represent mitral valve closure (MC) and
aortic valve opening (AO) (see Figure 2). During this time interval, the ventricles con-
tract with no change in volume (i.e. isovolumic contraction) and no blood flows to the
aorta. Thus, only respiration efforts and the movement of the heart affect the SCG
waveforms. As the baseline shifts in SCG signal due to respiratory efforts (i.e. the in-
clination of the sternum changing) can be effectively eliminated with high-pass filtering,
the orientation of the filtered SCG signal during isovolumic contraction can be as-
sumed to depict the orientation of the heart’s mechanical axis. Recent research has sug-
gested that breathing manifests also in SCG morphology [20]. In this study, we attempt
to minimize these changes by restricting the analysis to epochs of breath holding. The
second candidate was the orientation of the line connecting the two most distant points
in the 3-D seismocardiogram (see Figure 2). This line was assumed to represent the
direction in which the heart could produce maximal systolic force.
Signal processing
Pre-processing
All SCG and ECG signals were digitally filtered in BIOPAC with a band-pass filter
using a 0.5 Hz lower and a 40 Hz upper cut-off frequency. Based on the respiratory
Figure 2 Frames of reference. The red asterisk marks the MC point and the green asterisk the AO point.The green and red circles are the two most distant points in the seismocardiogram. The red dashed lineconnects the MC and AO points, and the green dashed line connects the two most distant points in theseismocardiogram. The units are in milli-gs (1 mg = 9,81 mm/s2).
Paukkunen et al. BioMedical Engineering OnLine (2015) 14:16 Page 6 of 9
signal, the data where the breath hold takes place was manually selected for each sub-
ject. R peaks were detected using a simplified Pan-Tompkins algorithm [21] by first
finding all cardiac cycles with an arbitrary length and then seeking the maximum values
inside each cycle. Extraction of SCG cycles was then performed by segmenting the sig-
nals to 700-ms long windows starting 100 ms before and ending 600 ms after each
ECG R peak. We found that the timings of the relevant SCG features in relation to the
R peaks were not associated with variable heart rate, and thus normalization would
have drastically affected the averaging process. Therefore, in this study, no heart rate
normalization was performed. After segmentation, ensemble averaging was done by
using the largest number of heartbeats that was available from all subjects (11 beats).
These heartbeats were selected by minimizing the mean standard deviation of the
resulting ensemble averages. MC and AO peaks were manually detected from averaged
z axis SCG signals using the annotation scheme proposed by Crow et al. as a guideline
[22]. The MC point was identified as the positive peak on the z-axis seismocardiogram
following the onset of the QRS complex, whereas the AO point is the positive peak
after the MC point.
Computing the frame of reference
A 3-D seismocardiogram is defined as a vector function of time with magnitude offfiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiX2 þ Y 2 þ Z2
pwhere each of the SCG axes represents a Cartesian coordinate in 3-D
space. Ensemble averaged 3-D seismocardiograms for each subject were rotated sepa-
rately based on the two frames of reference, which were the line connecting the most
distant points (Frame 1) and the line connecting the MC and AO peaks (Frame 2), as
described previously. To unify the orientation of the original 3-D seismocardiograms,
they were rotated so that either frame 1 or frame 2 was parallel with the z axis
(Figure 3). Rotation was performed in a plane with a normal vector described by the
cross product of the frame of reference and the z axis.
Evaluation of the inter-subject variability of seismocardiograms
Squared Pearson correlation coefficients (R2) as well as root-mean-square error (RMSE)
and normalised root-mean-square error (NRMSE) in the MC-AO interval for each
SCG axis were calculated in order to quantify the inter-subject variability in the 3-D
seismocardiograms both before and after the rotations. For each SCG axis, this pro-
cedure resulted inX10 − 1
1
n ¼ 45 different correlation coefficients, RMSE, and NRMSE
values between the ten subjects.
Results and discussionThe effect of rotations on the inter-subject correlation, RMSE, and NRMSE is shown in
Table 1. The morphology of the 3-D SCG waveforms varied strongly from subject to
subject, as evidenced by the relatively weak inter-subject correlation (mean correlation
of 0.39 on the x axis, 0.42 on the y axis, and 0.86 on the z axis) of the unprocessed 3-D
seismocardiograms. Fixing the frame of reference to the line connecting the MC and
AO peaks (i.e., frame 2) considerably enhanced the correlation between the subjects in
the y axis (mean correlation of 0.52 and 0.83 using frame 1 and frame 2, respectively).
Figure 3 Rotation of seismocardiograms. The leftmost column shows the original unrotated signal of asubject, the plots in the middle show the original signal rotated so that the line connecting the two mostdistant points is parallel to the z axis (frame 1), and the rightmost column shows the original signal rotatedso that the line connecting the MC and AO points is parallel to the z axis (frame 2). The units are in mg.
Paukkunen et al. BioMedical Engineering OnLine (2015) 14:16 Page 7 of 9
Also, the NRMSE decreased from 0.43 to 0.29 on using frame 2, suggesting the rotation
process decreased the inter-subject variability. However, the impact of using a unified
frame of reference was not as prominent in the other axes. The x axis correlations were
only slightly improved and RMSE slightly decreased using either frame of reference.
The proposed method did not significantly affect the standard SCG z-axis features.
This is quantified by the high correlation between the original and rotated z axis signals
(subject-wise mean of 0.99 for frame 1 and 0.95 for frame 2) as well as the consistency
of the original and rotated z axis waveform RMSEs. Also, it can be visually confirmed
from Figure 4 that the z-axis waveforms remain similar despite the rotation process.
Using the frame 2, the mean correlation across all axes was improved from 0.56 to
0.71. Also, 44 out of 45 y axis correlation coefficients were improved using frame 2.
Despite the seemingly low amount of increased coefficients in X and Z axes, well over
half of the coefficients in separate axes did improve when both frames of reference
were considered (31, 44 and 27 in the x, y, and z axis, respectively).
Although the y axis correlations were improved remarkably when a unified frame of
reference was used, the x axis correlations remained weak. This suggests that the x axis
does not contain significant physiological information. As the accelerometers used in
this study were very sensitive (2 V/g), it seems that insufficient coupling of cardiac and
blood flow-induced vibrations to the sternum might partly cause the weak inter-subject
correlation. Migeotte et al. also observed that most of the time the cardiac function is
Table 1 Effect of rotation
RMSE [mg] Normalized RMSE Correlation
Axis x y z x y z x y z
Unrotated 4.91±1.76 3.63±1.59 7.26±3.40 0.43±0.12 0.43±0.17 0.13±0.05 0.39±0.30 0.42±0.31 0.86±0.13
Frame 1 3.90±1.56 3.62±1.81 7.47±3.51 0.55±0.19 0.45±0.16 0.13±0.05 0.44±0.33 0.52±0.31 0.85±0.13
Frame 2 4.40±1.68 4.60±2.49 7.27±3.63 0.47±0.21 0.29±0.13 0.14±0.05 0.43±0.31 0.83±0.14 0.85±0.14
Figure 4 Effect of rotation on individual seismocardiogram trajectories. The individual seismocardiogramtrajectories without rotation (left), with rotation using frame 1 (center), and with rotation using frame 2 (right)are shown. The gray plots are the mean seismocardiograms of individual subjects. The blue lines are the meanof the means of seismocardiograms of individual subjects.
Paukkunen et al. BioMedical Engineering OnLine (2015) 14:16 Page 8 of 9
projected almost entirely on the yz-plane [13]. To enhance 3-D SCG, the characteristics
of the x axis should be further studied.
While the present method was feasible to remarkably improve the y axis correlations,
the z axis correlations were virtually unaffected. The possibility of masking or modifica-
tion of individual biological issues cannot, however, be excluded and it is yet unclear
whether these biological issues are important. The proposed method does not wipe out
information but only displays it differently. The mounting of the accelerometers might
have affected the results. As the orientation error found in the accelerometers was rela-
tively small, the effect of this error might be negligible. However, the degree of the
effect must be quantified in future research. Also, we suggest that in future work
the accelerometers should be mounted using a repeatable process that will guarantee
the orthogonality of the accelerometers.
The z axis SCG has been proposed as feasible to be used in cardiac time interval
measurements [23]. Given the high z axis correlations, it seems that the capability of
the z axis to be used in measuring cardiac time intervals was not affected when a uni-
fied frame of reference was used. This observation further asserts the feasibility of using
a unified frame of reference to improve 3-D SCG measurements.
ConclusionsThe results achieved in this study indicate that the use of a unified frame of reference is a
promising method for improving the inter-subject variability of 3-D seismocardiograms.
A larger sample size is needed to reassert the validity of this method. More importantly,
these results show that 3-D seismocardiograms have latent inter-subject similarities,
which are feasible to be revealed. Because the projections of the cardiac vibrations on the
measurement axes differ greatly, it seems obligatory to use 3-D SCG measurements when
analysis is based on SCG waveform morphology rather than time intervals.
Competing interestsThe authors declare that they have no competing interests.
Authors’ contributionsMP designed the study, drafted the manuscript, and participated in data analysis. PP carried out the data analysis andparticipated in drafting the manuscript. RK and RS participated in designing the study and critically revised themanuscript. All authors read and approved the final manuscript.
Paukkunen et al. BioMedical Engineering OnLine (2015) 14:16 Page 9 of 9
AcknowledgementsMikko Paukkunen is funded by The Finnish Cultural Foundation and The Finnish Society of Electronics Engineers.
Author details1Department of Electrical Engineering and Automation, Aalto University, P.O. BOX: FI-13340, 00076 Helsinki, Finland.2School of Medicine, University of Eastern Finland, Kuopio, Finland. 3Health Factory, Aalto University,P.O. BOX: FI-13340, 00076 Helsinki, Finland.
Received: 8 August 2014 Accepted: 10 February 2015
References
1. Salerno D, Zanetti J. Seismocardiography: a new technique for recording cardiac vibrations. Concept, method, andinitial observations. J Cardiovasc Technol. 1990;9((2):111–8.2. Gordon J. Certain molar movements of the human body produced by the circulation of the blood. J Anat Physiol.
1877;11(Pt 3):533–6.3. Bozhenko BS. Seismocardiography–a new method in the study of functional conditions of the heart. Ter Arkh.
1961;33:55–64.4. Baevsky R, Egorov A, Kazarian L. Metodika seismokardiografii. Kardiologia. 1964;18:87–9.5. Dinh A. Heart activity monitoring on Smartphone. Int Proc Chem Biol Environ Eng. 2011;11(9):45–9.6. Korzeniowska-Kubacka I, Bilińska M, Piotrowicz R. Usefulness of seismocardiography for the diagnosis of ischemia
in patients with coronary artery disease. Ann Noninvasive Electrocardiol. 2005;10(3):281–7.7. Wick C, Su J, McClellan J, Brand O, Bhatti P, Buice A, et al. A system for seismocardiography-based identification of
quiescent heart phases: implications for cardiac imaging. IEEE Trans Inf Technol Biomed. 2012;16(5):869–77.8. Tavakolian K, Khosrow-Khavar F, Kajbafzadeh B, Marzencki M, Blaber A, Kaminska B. Precordial acceleration signals
improve the performance of diastolic timed vibrations. Med Eng Phys. 2013;35(8):1133–40.9. Becker M, Roehl A, Siekmann U, Koch A, de la Fuente M, Roissant R, et al: Simplified detection of myocardial
ischemia by seismocardiography. Herz 2013, [Epub ahead of print]:1–7.10. Migeotte P, De Ridder S, Tank J, Nathalie Pattyn, I I Funtova, Roman M Baevsky, et al.: Three dimensional
ballisto-and seismo-cardiography: HIJ wave amplitudes are poorly correlated to maximal systolic force vector.In Proceedings of the Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conferenceof the IEEE: 28 August - 1 September 2012; San Diego, California, USA. 2012.
11. McKay WPS, Gregson PH, McKay BWS, Militzer J. Sternal acceleration ballistocardiography and arterial pressurewave analysis to determine stroke volume. Clin Invest Med. 1999;22(1):4–14.
12. De Ridder S, Migeotte P F, Neyt X, Pattyn N and Prisk G: Three-dimensional ballistocardiography in microgravity:A review of past research. In Proceedings of the Engineering in Medicine and Biology Society,EMBC, 2011 AnnualInternational Conference of the IEEE: 30 August - 3 September 2011; Boston, Massachusetts, USA. 2011.
13. Migeotte P, Tank J, Pattyn N, Funtova I, Baevsky R, Neyt X, et al.: Three dimensional ballistocardiography:methodology and results from microgravity and dry immersion. In Proceedings of the Engineering in Medicineand Biology Society, EMBC, 2011 Annual International Conference of the IEEE: 2011.
14. Castiglioni P, Faini A, Parati G and Di Rienzo M: Wearable seismocardiography. In Proceedings of the Engineeringin Medicine and Biology Society, 2007. EMBS 2007. 29th Annual International Conference of the IEEE: 22–26August 2007; Lyon, France. 2007.
15. Pandia K, Ravindran S, Kovacs G T, Giovangrandi L and Cole R: Chest-accelerometry for hemodynamic trendingduring valsalva-recovery. In Proceedings of the Applied Sciences in Biomedical and Communication Technologies(ISABEL), 2010 3rd International Symposium On: 2010.
16. Di Rienzo M, Meriggi P, Vaini E, Castiglioni P and Rizzo F: 24 h seismocardiogram monitoring in ambulantsubjects. In Proceedings of the Engineering in Medicine and Biology Society (EMBC), 2012 Annual InternationalConference of the IEEE: 2012.
17. Paukkunen M, Linnavuo M, Haukilehto H, Sepponen R. A system for detection of three-dimensional precordialvibrations. IJMTIE. 2012;2(1):52–66.
18. Wiens A, Etemadi M, Roy S, Klein L, Inan O: Towards Continuous, Non-Invasive Assessment of Ventricular Functionand Hemodynamics: Wearable Ballistocardiography. IEEE Journal of Biomedical and Health Informatics2014, PP(99):1.
19. Tavakolian K, Vaseghi A, Kaminska B. Improvement of ballistocardiogram processing by inclusion of respirationinformation. Physiol Meas. 2008;29(7):771–8.
20. Pandia K, Inan OT, Kovacs GTA, Giovangrandi L. Extracting respiratory information from seismocardiogram signalsacquired on the chest using a miniature accelerometer. Physiol Meas. 2012;33(10):1643–60.
21. Pan J, Tompkins WJ: A real-time QRS detection algorithm. Biomedical Engineering, IEEE Transactions on1985, (3):230–236.
22. Crow RS, Hannan P, Jacobs D, Hedquist L, Salerno DM. Relationship between seismocardiogram andechocardiogram for events in the cardiac cycle. Am J Noninvasive Cardiol. 1994;8(1):39–46.
23. Tavakolian K, Ngai B, Blaber AP and Kaminska B: Infrasonic cardiac signals: Complementary windows tocardiovascular dynamics. In Proceedings of the Engineering in Medicine and Biology Society, EMBC, 2011 AnnualInternational Conference of the IEEE: Boston, Massachusetts, USA. 2011.