Use of IHE DEC Profiles for Remote Monitoring Authors: Paul Schluter and Todd Cooper HITSP RMON discussion October 2, 2008

Upload: samuel-garcia

Post on 27-Mar-2015

215 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Use of IHE DEC Profiles for Remote Monitoring Authors: Paul Schluter and Todd Cooper HITSP RMON discussion October 2, 2008

Use of IHE DEC Profiles forRemote Monitoring

Authors: Paul Schluter and Todd CooperHITSP RMON discussion October 2, 2008

Page 2: Use of IHE DEC Profiles for Remote Monitoring Authors: Paul Schluter and Todd Cooper HITSP RMON discussion October 2, 2008

DeviceDevice

Intermediary

Device

EHR or Care Coordination Applications

Home Settin

g

Hospital Setting

Device Intermediary

IHE-DEC

Subset of IHE-DEC

IHE-DEC: Same protocol used in two settings …

Page 3: Use of IHE DEC Profiles for Remote Monitoring Authors: Paul Schluter and Todd Cooper HITSP RMON discussion October 2, 2008

IHE Device Enterprise Communication (DEC) and related Profiles (PCD, SPD, ACM).

Device Observation Consumer

(DOC)

Device Observation

Reporter (DOR)

Device

PCD-01 Transaction: Communicate Device Data

• Periodic Physiologic Data: Heart Rate, Respiration rate• Episodic Physiologic Data: NIBP, Weight, Cardiac tests• Primarily structured data but facilities non structured data as well• Must include Patient ID• > 1 minute sampling rate (but could be faster)• Waveform support is simple extension• Flexible extensions with:

• Alarms and alerts supported through related profile (ACM)• Subscribe to Patient Data (SPD) Profile• Security, privacy, confidentiality through related profile (ATNA)

• ISO/IEEE 11073-10101 nomenclature (does not rule out LOINC or SNOMED)• ISO/IEEE 11073 or UCUM units-of-measure• OBX-4 used for «mds.vmd.chan.metric», OBX-20 for measurement sites• Profile and simplify other aspects of HL7 V2.5 messaging

Page 4: Use of IHE DEC Profiles for Remote Monitoring Authors: Paul Schluter and Todd Cooper HITSP RMON discussion October 2, 2008

Mapping Preserves Measurement Context Medical Device System (MDS)

Virtual Medical Device (VMD)

Channel (Chan)

Numeric

Enumeration

1

*

1

*

1

*

1

0..*

Metric

Overall Device or System

Virtual Medical Device

Channel (only as needed)

A single observation

Page 5: Use of IHE DEC Profiles for Remote Monitoring Authors: Paul Schluter and Todd Cooper HITSP RMON discussion October 2, 2008

PCD-01 Message Example (simple heart rate)

MSH|^~\&|HL7^080019FFFF4F6AC0^EUI-64|MMS|||20070118133700||ORU^R01|a4e2e3:11036b5cdbb:-|P|2.5|20070118133700||NE|AL||

8859/1PID|||6O008D^^^A^PI||DAVIDSON^DAVID^^^^^LPV1||E|3WICU^305^305OBR|1|080019FFFF4F6AFE200701|080019FFFF4F6AC0200701|126.3.3.1^2000^MDC|||20070118133700OBX|1|NM|147842^MDC_ECG_HEART_RATE^MDC|1.6.1.1|80|/min^/min^UCUM|||||R‘Heart Rate’ ‘/min’ UOM

IEEE EUI-64

OBX-4 Value

Page 6: Use of IHE DEC Profiles for Remote Monitoring Authors: Paul Schluter and Todd Cooper HITSP RMON discussion October 2, 2008

PCD-01 Message Example (more complex)MSH|^~\&|HL7^080019FFFF4F6AC0^EUI-64|MMS|||20070118133700||ORU^R01|

a4e2e3:11036b5cdbb:-|P|2.5|20070118133700||NE|AL||8859/1PID|||6O008D^^^A^PI||DAVIDSON^DAVID^^^^^LPV1||E|3WICU^305^305OBR|1|080019FFFF4F6AFE200701|080019FFFF4F6AC0200701|126.3.3.1^2000^MDC|||20070118133700OBX|1|NM|147842^MDC_ECG_HEART_RATE^MDC|1.6.1.1|80|/min^/min^UCUM|||||ROBX|1|NM|147842^MDC_ECG_HEART_RATE^MDC|1.6.1.1|60|/min^/min^UCUM|||||ROBX|2|NM|131841^MDC_ECG_AMPL_ST_I^MDC|1.7.1.1|0.0|mm^mm^UCUM|||||ROBX|3|NM|131842^MDC_ECG_AMPL_ST_II^MDC|1.7.1.2|0.0|mm^mm^UCUM|||||ROBX|4|NM|131901^MDC_ECG_AMPL_ST_III^MDC|1.7.1.3|0.0|mm^mm^UCUM|||||ROBX|5|NM|131847^MDC_ECG_AMPL_ST_V5^MDC|1.7.1.4|0.0|mm^mm^UCUM|||||ROBX|6|NM|131902^MDC_ECG_AMPL_ST_AVR^MDC|1.7.1.5|0.0|mm^mm^UCUM|||||ROBX|9|NM|150035^MDC_PRESS_BLD_ART_MEAN^MDC|1.3.1.1|96|mm[Hg]^mm[Hg]^UCUM|||||ROBX|10|NM|150033^MDC_PRESS_BLD_ART_SYS^MDC|1.3.1.2|122|mm[Hg]^mm[Hg]^UCUM|||||ROBX|11|NM|150034^MDC_PRESS_BLD_ART_DIA^MDC|1.3.1.3|82|mm[Hg]^mm[Hg]^UCUM|||||ROBX|12|NM|149522^MDC_BLD_PULS_RATE_INV^MDC|1.2.1.1|60|/min^/min^UCUM|||||ROBX|13|NM|150047^MDC_PRESS_BLD_ART_PULM_MEAN^MDC|1.4.2.1|14|mm[Hg]^mm[Hg]^UCUM|||||ROBX|14|NM|150045^MDC_PRESS_BLD_ART_PULM_SYS^MDC|1.4.2.2|24|mm[Hg]^mm[Hg]^UCUM|||||ROBX|15|NM|150046^MDC_PRESS_BLD_ART_PULM_DIA^MDC|1.4.2.3|10|mm[Hg]^mm[Hg]^UCUM|||||ROBX|16|NM|151610^MDC_VENT_CO2_RESP_RATE^MDC|1.5.1.1|12|{breath}/min^{breath}/min^UCUM|||||ROBX|17|NM|151976^MDC_VENT_PRESS_AWAY_END_EXP_POS^MDC|1.5.1.2|4|cm[H20]^cm[H20]^UCUM|||||ROBX|18|NM|152008^MDC_VENT_VOL_MINUTE_AWAY^MDC|1.5.1.3|0.0|l/min^l/min^UCUM|||||ROBX|19|NM|151920^MDC_VENT_CONC_AWAY_O2_INSP^MDC|1.5.1.4|21|%^%^UCUM|||||ROBX|20|NM|151980^MDC_VENT_VOL_TIDAL^MDC|1.5.1.5|1096|ml^ml^UCUM|||||ROBX|21|NM|151957^MDC_VENT_PRESS_MAX^MDC|1.5.1.6|17|cm[H20]^cm[H20]^UCUM|||||ROBX|22|NM|151784^MDC_PRESS_RESP_PLAT^MDC|1.5.1.7|30|cm[H20]^cm[H20]^UCUM|||||ROBX|23|NM|151792^MDC_PRESS_AWAY^MDC|1.5.1.8|35|cm[H20]^cm[H20]^UCUM|||||ROBX|24|NM|151808^MDC_PRESS_AWAY_END_EXP_POS_INTRINSIC^MDC|1.5.1.9|1|cm[H20]^cm[H20]^UCUM|||||R

Page 7: Use of IHE DEC Profiles for Remote Monitoring Authors: Paul Schluter and Todd Cooper HITSP RMON discussion October 2, 2008

IHE DEC (and associated) profiles

Provides a Unified Interface that --• Supports the complete range of device data (except for

high-bandwidth continuous waveforms) found in health related devices today and in the future.

• Provides a consistent interface in Health Record, CareManagement Systems and Personal Health Records that is:- Secure, private- High performance even on low bandwidth links- Scalable, from a simple device intermediary (a few devices for the

same patient) to multiple devices for many patients. Trivial to implement.

- Straightforward mapping from proprietary and IEEE 11073 formats (uses same nomenclature as X73 Personal Health devices).

- Based on proven HL7 standards that are supported by all Care Management and EHRs products today