Page 1: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

1

®

user manual

Page 2: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

2

Page 3: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

3

Contents

Chapter 1 1 Intended Use, Contraindications and General Precautions

Chapter 2 3 Description of the Inogen One® G3 Oxygen Concentrator 3 ImportantPartsoftheInogenOne®G3OxygenConcentrator 4 UserInterfaces 5 Input/OutputConnections 6 PowerSupplyOptions 8 InogenOne®G3Accessories

Chapter 3 11 Operating Instructions 11 GeneralInstructions 15 AdditionalOperatingInstructions 18 BatteryOperatingInstructions 19 BatteryCareandMaintenanceChapter 4 21 Inogen One® G3 Oxygen Concentrator Audible and Visible Signals (including Alarms)

Chapter 5 29 Troubleshooting Chapter 6 31 Cleaning, Care and Maintenance 31 CannulaReplacement 31 CaseCleaning 32 FilterCleaningandReplacement 33 InogenOne®G3ColumnChangeProcedure 36 OtherServiceandMaintenance 36 DisposalofEquipmentandAccessories

Chapter 7 37 Symbols Used on Concentrator and Accessories

Chapter 8 39 Inogen One® G3 System Specifications

English

Page 4: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

4

Page 5: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

1

1 Intended Use, Contraindications and General Precautions

IntendedUse,ContraindicationsandGeneralPrecautionsChapter1

Intended Use

TheInogenOne®G3OxygenConcentratorisusedonaprescriptivebasisbypatientsrequiringsupplementaloxygen.Itsuppliesahighconcentrationofoxygenandisusedwithanasalcannulatochanneloxygenfromtheconcentratortothepatient.TheInogenOne®G3maybeusedinhome,institution,vehicle,onanairplaneandvariousmobileenvironments.

CAUTION USAFederallawrestrictsthisdevicetosalebyorontheorderofaphysician.Mayalsobeapplicableinothercountries.

CAUTION Nasalcannulashouldberatedfor5litersperminute(e.g.Salter16SOFT)toensureproperpatientusageandoxygendelivery.

WARNING Availabilityofanalternatesourceofoxygenisrecommendedincaseofpoweroutageormechanicalfailure.Consultyourequipmentproviderfortypeofback-upsystemrecommended.

CAUTION Itistheresponsibilityofthepatienttomakeback-uparrangementsforalternativeoxygensupplywhentraveling;Inogenassumesnoliabilityforpersonschoosingnottoadheretomanufacturerrecommendations.

TheexpectedlifefortheInogenOne®G3OxygenSystemis5years,withtheexceptionofthesievebeds(metalcolumns)whichhaveanexpectedlifeof1yearandthebatteries,whichhaveanexpectedlifeof500fullcharge/dischargecycles.

! C

English

Page 6: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

2

Contraindications

WARNING ThisdeviceisNOTINTENDEDtobelifesustainingorlifesupporting. CAUTION Undercertaincircumstances,theuseofnon-prescribedoxygentherapy

canbehazardous.Thisdeviceshouldbeusedonlywhenprescribedbyaphysician.

CAUTION Additionalmonitoringorattentionmayberequiredforpatientsusingthisdevicewhoareunabletohearorseealarmsorcommunicatediscomfort.Ifthepatientshowsanysignsofdiscomfort,aphysicianshouldbeconsultedimmediately.

CAUTION TheInogenOne®G3isnotdesignedorspecifiedtobeusedinconjunctionwithahumidifier,nebulizerorconnectedwithanyotherequipment.Useofthisdevicewithahumidifier,nebulizerorconnectedwithanyotherequipmentmayimpairperformanceand/ordamagetheequipment.DonotmodifytheInogenOne®G3Concentrator.Anymodificationsperformedontheequipmentmayimpairperformanceordamageequipmentandwillvoidyourwarranty.

! C

! C

! C

General Precautions

WARNING Thedeviceproducesenrichedoxygengaswhichacceleratescombustion.DoNOTALLOWSMOKINGOROPENFLAMESwithin

10feetofthisdevicewhileinuse.

WARNING DonotsubmersetheInogenOne®G3oranyoftheaccessoriesinliquid.Donotexposetowaterorprecipitation.Donotoperateinexposedrain.Thiscouldleadtoelectricalshockand/ordamage.

CAUTION Donotuseoil,grease,orpetroleum-basedproductsonorneartheInogen

One®G3.

CAUTION NeverleavetheInogenOne®G3inanenvironmentwhichcanreachhightemperatures,suchasanunoccupiedcarinhightemperatureenvironments.Thiscoulddamagethedevice.

Page 7: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

3

Important Parts of the Inogen One® G3 Oxygen Concentrator

Description of the Inogen One® G3 Oxygen Concentrator2

On/Off Button

Display Back-light ButtonBreath

Detection Light

Display

Audible Alarm Button

Alert/Alarm Light

Flow Control

DescriptionoftheInogenOne®G3OxygenConcentratorChapter2

English

DescriptionoftheInogenOne®G3OxygenConcentratorChapter2

Page 8: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

4

User Controls

ON / OFF ButtonPressoncetoturn“ON”;Pressandholdforonesecondtoturn“OFF”.

Audible Alarm Button PressingthisbuttonwilltoggletheInogenOne®G3’sbreathdetectionaudiblealertonandoff.

BreathDetectionAlertMode.TheInogenOne®G3willalertwithaudibleandvisualsignalsfor“nobreathdetected”whenthismodeisenabledandnobreathhasbeendetectedfor60seconds.At60seconds,thedevicewillenterintoautopulsemodeandonceanotherbreathisdetected,thedevicewillexitautopulsemodeanddelivernormallyoninspiration.Thedisplay’smodeindicationareawillshowabellicon,flashingredoryellowlightdependingonmodel,anddisplaymessagewhenthealertisenabled.

Ifpowerislost,thebreathdetectionaudiblealertremainssetintheuserpreferredmode.Flow Setting Control Buttons Usethe–or+flowsettingcontrolbuttonstoselectthesettingasshownonthedisplay.Flowsettingswillvarybetween1-4or1-5dependingonmodel.

Display Backlight ButtonPushon;automaticallyturnsoffafter15seconds.

User Interfaces

DisplayThisscreendisplaysinformationregardingflowsetting,powerstatus,batterylifeanderrors.IfyouwouldliketochangethelanguageontheInogenLCDScreencontactyourEquipmentProviderforinstructions.

Page 9: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

5DescriptionoftheInogenOne®G3OxygenConcentratorChapter2

User Interfaces (continued)

Indicator LightsAredoryellowlightindicateseitherachangeinoperatingstatusoraconditionthatmayneedresponse(alarm).Coloroflightwillvarydependingonmodel.Aflashinglightishigherprioritythannon-flashing.

Audible SignalsAnaudiblesignal(beep)indicateseitherachangeinoperatingstatusoraconditionthatmayneedresponse(alarm).Morefrequentbeepsindicatehigherpriorityconditions.

Input / Output Connections

Particle FiltersThefiltersmustbeinplaceattheintakeendoftheconcentratorduringoperationtokeepinputairclean.

Cannula Nozzle FittingThenasalcannulaconnectstothisnozzleforInogenOne®G3outputofoxygenatedair.

DC Power InConnectionforexternalpowerfromtheACpowersupplyorDCpowercable.

USB PortUsedforservicingonly.

English

Page 10: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

6

Power Supply Options

Single and Double Rechargeable Lithium Ion BatteriesThebatterywillpowertheInogenOne®G3withoutconnectiontoanexternalpowersource.Whenfullycharged,asinglebatterywillprovideupto4.7hoursofoperation;adoublebatterywillprovideupto10hoursofoperation.ThebatteryrechargeswhenproperlyinstalledintheInogenOne®G3andtheconcentratorisconnectedtoACorDCpower.Rechargingtimeisupto5hoursforasinglebatteryand8hoursforadoublebattery.Seeinformationinthe“BatteryCareandMaintenance”section.

Power SupplyOverviewTheInogenOne®G3ACpowersupply(BA-301)isusedtopowertheInogenOne®G3concentratorfromanACpowersource.

DescriptionTheInogenOne®G3ACpowersupplyisspecificallydesignedforusewiththeInogenOne®G3OxygenConcentrator(IO-300).TheACpowersupplyprovidestheprecisecurrentandvoltagerequiredtosafelypowertheInogenOne®G3andisdesignedtooperatefromspecifiedACpowersources.WhenusedwithACpowersources,thepowersupplyautomaticallyadaptstoinputvoltagesfrom100Vto240V(50-60HZ)permittingusewithmostpowersourcesthroughouttheworld.

TheACpowersupplywillchargetheInogenOne®G3BatterieswhenusedwithACinputpower.Duetoaircraftpowerlimitations,theACPowerSupplycannotbeusedtochargetheInogenOne®G3Batterywhenusedonanaircraft.TheACpowersupplyisusedwiththefollowingcomponents:

•PowersupplywithattachedpoweroutputcabletoconnecttotheInogenOne®G3•ACpowerinputcable

TheDCpowercable(BA-306)isspecificallydesignedforusewiththeInogenOne®G3OxygenConcentrator(IO-300).TheDCpowerinputcableconnectsdirectlytotheautomobilecigarettelighterorauxiliaryDCpowersupply.RefertoChapter3foradditionaloperatinginstructionsforusewithDCpowersource.

Page 11: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

7DescriptionoftheInogenOne®G3OxygenConcentratorChapter2

English

*Actualproductappearancemayvary.

AC Power Plug*

Power Output Plug (Plug into Concentrator)

LED Indicator Light*AC Input

InogenOneACPowerSupplyModel#BA-301

InogenOneG3DCPowerCableModel#BA-306

DC cigarette lighter Power Plug for use in Automobile/RV/Boat/Aircraft

Power Output Plug (Plug into Concentrator)

WARNING Donotusepowersuppliesorpowercablesotherthanthosespecifiedinthisusermanual.Theuseofnon-specifiedpowersuppliesorpowercablesmaycreateasafetyhazardand/orimpairequipmentperformance.Donotwrapcordsaroundpowersupplyforstorage.Donotdrive,dragorplaceobjectsovercord.Doingsomayleadtodamagedcordsandafailuretoprovidepowertotheconcentrator.Toavoiddangerofchokingorstrangulationhazard,keepcordsandtubingawayfromchildrenandpets.

! C

Page 12: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

8

Inogen One® G3 Accessories

WARNING Donotusepowersupplies/adaptersoraccessoriesotherthanthosespecified inthisusermanual.Theuseofnon-specifiedaccessoriesmaycreateasafety hazardand/orimpairequipmentperformance.

Nasal CannulaAnasalcannulamustbeusedwiththeInogenOne®G3toprovideoxygenfromtheconcentrator.Asinglelumencannulaupto25feetinlengthisrecommendedtoensureproperbreathdetectionandoxygendelivery.

WARNING Toavoiddangerofchokingandstrangulationhazard,keeptubingawayfromchildrenandpets.

NOTE Increasingthecannulalengthmayreducetheperceivednoiseduringoxygenbolusdelivery.

NOTE Whenusingacannula25ft.inlengthwiththeInogenOne®G3,anincreaseinflowsettingmayberequired.

Carry Bag (CA-300)ThecarrybagprovidesprotectionwhilecarryingtheInogenOne®G3withahandleandadjustableshoulderstrap.Theunitisinsertedintoplacefromthebottomofthecarrybag.TakecaretoinserttheInogenOne®G3sothatthedisplayisseenthroughtheplasticwindowandthecannulabarbisaccessiblethroughtheopencutoutatthetopofthebag.InogenOne®G3intakeandexhaustventsshouldbevisiblethroughtheopenmeshpanels.TheInogenOne®G3canbeoperatedusingbatterypowerduringtransport.Theexpandablezipperedbottomisdesignedwithanadjustablezipperforusewithasingleordoublebatteryandprovideseasyaccessforbatterychanges.Theexpandablesectionisnotintendedforuseasastoragecompartment.

! C

! C

Page 13: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

9

Inogen One® G3 Optional Accessories

External Battery Charger (BA-303)1.PlugtheExternalBatteryChargerACpowersupplycord

intoanelectricaloutlet.2.PlugtheExternalBatteryChargerACpowersupplyinto

thebatterycharger.3.SlideyourchargerontotheInogenOneG3Battery byclickingandlockingintothecharger.4.Whenthebatteryisinthecorrectposition,asolid redlightwillindicatethatthebatteryischarging.5.Whenthegreenlightilluminates,thebatteryis fullycharged.

CAUTION AvoidtouchingtherecessedelectricalcontactsoftheExternalBatteryCharger;damagetocontactsmayaffectchargeroperation.

NOTE Thesecontactsarenotpoweredunlessabatteryisinplaceandcharging.

NOTE TocompletelyremovepowerfromtheExternalBatteryCharger, removetheplug.

Inogen One G3 Backpack (CA-350)Alternative/optionalwayofcarryingyourInogenOne®G3,handsfree,morecomfort,outofyourwaywithextrapocketsforadditionalaccessories.

DescriptionoftheInogenOne®G3OxygenConcentratorChapter2

English

Page 14: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

10

Page 15: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

11OperatingInstructionsChapter3

English

Operating Instructions3

! C

General Instructions

1. Place the Inogen One® G3 in a well ventilated location.

Airintakeandexhaustmusthaveclearaccess.LocatetheInogenOne®G3insuchawaythatanyauditoryalarmsmaybeheard.

WARNING AvoiduseoftheInogenOne®G3inpresenceofpollutants,smokeorfumes.DonotusetheInogenOne®G3inpresenceofflammableanesthetics,cleaningagentsorotherchemicalvapors.

CAUTION Donotobstructairintakeorexhaustwhenoperatingthedevice.

Blockageofaircirculationorproximitytoaheatsourcemayleadtointernalheatbuildupandshutdownordamagetotheconcentrator.

CAUTION TheInogenOne®G3Concentratorisdesignedforcontinuoususe.Foroptimalsievebed(columns)life,theproductshouldbeusedfrequently.

2. Ensure both particle filters are in place.

CAUTION DonotoperatetheInogenOne®G3withoutbothparticlefiltersinplace.Particlesdrawnintothesystemmaydamagetheequipment.

Exhaust Intake

Exhaust

Intake

5SettingModel 4SettingModel

Page 16: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

12

3. Install the battery. InserttheInogenOne®G3batterybyslidingbatteryintoplaceuntilthelatchreturnstotheupperposition.

CAUTION TheInogenOne®G3batteryactsasasecondarypowersupplyintheevent ofaplannedorunexpectedlossoftheACorDCexternalpowersupply.When

operatingtheInogenOne®G3fromanACorDCexternalpowersupply,aproperlyinsertedInogenOne®G3batteryshouldbemaintainedintheunit.Thisprocedurewillensureuninterruptedoperationandwilloperateallalarmsandalertsintheeventofalossoftheexternalpowersupply.

4. Connect the Power Supply. ConnecttheACinputplugtothepowersupply.ConnecttheACpowerplugtothepowersourceandconnectthepoweroutputplugtotheInogenOne®G3.ThegreenLEDonthepowersupplywillbeilluminatedandabeepwillsoundfromtheconcentrator.

CAUTION Ensurethepowersupplyisinawellventilatedlocationasitrelieson aircirculationforheatdissipation.Thepowersupplymaybecomehotduring operation.Makesurethepowersupplycoolsdownbeforehandling.

CAUTION Thepowersupplyisnotwaterresistant.

CAUTION Donotdisassemblethepowersupply.Thismayleadtocomponentfailure and/orsafetyrisk.

CAUTION Donotplaceanythinginthepowersupplyportotherthanthesuppliedwallcord.AvoidtheuseofelectricalextensioncordswiththeInogenOne®G3.

Ifanextensioncordmustbeused,useanextensioncordthathasanUnderwritersLaboratory(UL)Markandaminimumwirethicknessof

18gauge.Donotconnectanyotherdevicestothesameextensioncord.

Page 17: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

13OperatingInstructionsChapter3

NOTE Undercertainconditions(seeTechnicalSpecifications)thepower supplymayshutdown.ThegreenLEDwillblinkorwillnolongerbe

illuminated.Ifthisoccurs,disconnectthepowersupplyforatleast10secondsandreconnect.

NOTE WhenthepowersupplyisdisconnectedfromtheACoutlet,alsodisconnectitfromtheconcentratortoavoidunnecessarybatterydischarge.

5. Connect the nasal cannula tubing to the nozzle fitting.NozzlefittingislocatedonthetopoftheInogenOne®G3.Useofasinglelumencannulaupto25feetinlengthisrecommendedtoensureproperbreathdetectionandoxygendelivery.Additionaltitrationmaybeneededtoensureproperoxygendeliverywhenusingaparticularcannula.

CAUTION Toensureoxygenflow,ensurethatthenasalcannulaisproperlyconnectedtothenozzlefittingandthatthetubingisnotkinkedorpinchedinanyway.

CAUTION Replacethenasalcannulaonaregularbasis.Checkwithyourequipmentproviderorphysiciantodeterminehowoftenthecannulashouldbereplaced.

6. Turn on your Inogen One® G3 by pressing the ON/OFF Button.AsingleshortbeepwillsoundaftertheInogenlogoisdisplayed.“PleaseWait”willappearwhiletheconcentratorstartsup.Thedisplaywillindicatetheselectedflowsettingandpowercondition.Followingabriefstart-upsequence,awarmupperiodupto2minuteswillinitiate.Duringthistimeperiodtheoxygenconcentrationisbuildingtobutmaynothavereachedspecification.AdditionalwarmuptimemaybeneededifyourInogenOne®G3hasbeenstoredinextremelycoldtemperatures.

7. Set the Inogen One® G3 Concentrator to the flow rate prescribed by your physician or clinician. Usethe+or–settingbuttonstoadjusttheInogenOne®G3tothedesiredsetting.Thecurrentsettingcanbeviewedonthedisplay.

English

Page 18: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

14

! C

8. Position the nasal cannula on your face and breathe through your nose. TheInogenOne®G3willsensetheonsetofinhalationanddeliveraburstofoxygenataprecisetimewhenyouinhale.TheInogenOne®G3willsenseeachbreathandcontinuetodeliveroxygeninthismanner.Asyourbreathingratechanges,theInogenOne®G3willsensethesechangesanddeliveroxygenonlyasyouneedit.Attimes,ifyouinhaleveryquicklybetweenbreathstheInogenOne®G3mayignoreoneofthebreaths,givingtheappearanceofamissedbreath.ThismaybenormalastheInogenOne®G3sensesandmonitorsthechangesinyourbreathingpattern.TheInogenOne®G3willnormallysensethenextbreathanddeliveroxygenaccordingly.

Agreenlightwillflasheachtimeabreathisdetected.Makecertainthatthenasalcannulaisproperlyalignedonyourfaceandyouarebreathingthroughyournose.

WARNING Ifyoubegintofeelillorareexperiencingdiscomfortwhileusingthisdevice,consultyourphysicianimmediately.

CAUTION TheInogenOne®G3isdesignedtoprovideaflowofhighpurityoxygen.

Anadvisoryalarm,“OxygenLow”,willinformyouifoxygenconcentrationdrops.Ifalarmpersists,contactyourequipmentprovider.

GeneralToremovepower,unplugtheinputcordfromitssource(i.e.,ACwalloutlet,DCcarcigarettelighteradapter)anddisconnectitfromtheInogenOne®G3.

CAUTION Ensurethepowersupplyispoweredfromonlyonepowersource(ACorDC)atanygiventime.

Page 19: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

15OperatingInstructionsChapter3

Additional Operating Instructions

For Use In Home with AC PowerForoperationofthepowersupplyusinganACpowersource,followtheseinstructions:

1. ConnecttheACinputtothepowersupply.2. ConnecttheACpowerplugtothepowersourceandconnectthepoweroutputplug

totheInogenOne®G3.ThegreenLEDwillbeilluminated,indicatingthe powersupplyhasinputpower.

Traveling with your Inogen One® G3 System

TheInogenOne®G3Systemmakestravelbyboat,carortrainmoreconvenientforoxygenusersthaneverbefore.Nowyougetthesamequalityperformanceandconveniencewhileonthegothatyou’reusedtoreceivingfromyourInogenOne®G3athome.HerearesomeusefulandimportantinstructionsformaximizingperformanceandconveniencewhenusingyourInogenOne®G3whiletraveling.

Youshouldbeginplanningforatripwithachecklistofitemstoremember.Thischecklistshouldinclude:

✓ ACpowersupplyandDCpowercable✓ ExtraBattery(s)ifrequired✓ Importantphonenumbers,suchasthoseofyourphysicianandhomehealthcare provider,orprovidersintheareayouwillbetraveling✓ Planforbackupoxygenintheeventofaprolongedpoweroutageor mechanicalfailure

For Use In Automobile/RV/Boat/AircraftForoperationusingaDCpowersource,followtheseinstructions:

1.ConnecttheDCpoweroutputplugtotheInogenOne®G3.2.ConnecttheDCpowerplug(CigaretteLighterAdapter)to

thepowersource.3. Theplugshouldinsertintothesocketwithoutexcessive

forceandsecurelyremaininplace.

English

Page 20: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

16

4. Checkthedevicedisplayscreentoconfirmthatanexternalpowersupplyisconnected. AniconofabatterywithalightningboltoranACpowerplugwillbedisplayedonthe

screen,indicatingthatexternalpowerisconnectedandfunctioningproperly.

WARNING EnsurethattheautomobilepowersocketisadequatelyfusedfortheInogenOne®G3powerrequirements(minimum15Amp).Ifthepowersocketcannotsupporta15Ampload,thefusemaybloworthesocketmaybedamaged.

WARNING ThetipoftheCigaretteAdapterPlugbecomesHOTwheninuse.Donottouchthetipimmediatelyafterremovalfromanautocigarettelightersocket.

CAUTION Ensuretheautomobilepowersocketiscleanofcigaretteashandtheadapterplugfitsproperly,otherwiseoverheatingmayoccur.

CAUTION Donotusethepowersupplywithacigaretteplugsplitterorwithan extensioncable.ThismaycauseoverheatingoftheDCpowerinputcable.

CAUTION DonotjumpstarttheautomobilewiththeDCpowercableconnected.Thismayleadtovoltagespikeswhichcouldshutdownand/ordamagetheDCpowerinputcable.

CAUTION WhenpoweringtheInogenOne®G3inanautomobileensurethevehicle’sengineisrunningfirst,beforeconnectingDCcableintocigarettelighteradapter.Operatingthedevicewithouttheenginerunningmaydrainthevehicle’sbattery.

CAUTION Achangeinaltitude(forexample,fromsealeveltomountains)mayaffecttotaloxygenavailabletothepatient.Consultyourphysicianbeforetravelingtohigherorloweraltitudestodetermineifyourflowsettingsshouldbechanged.

Traveling By AirTheFAAallowstheInogenOne®G3onboardallU.S.aircraft,hereareafewpointstomakeairtraveleasy.

Planning Your FlightWhenflyingwiththeInogenOne®G3,youmustinformtheairlinethatyouwillbeusingyourInogenOne®G3onboardtheaircraft.Youmustalsokeepasignedstatementfromyourphysicianwithyouthatincludes:

! C

! C

Page 21: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

17OperatingInstructionsChapter3

English

•Yourabilitytosee/hearalarmsandappropriatelyrespond.

•Whenoxygenuseisnecessary(alloraportionofthetrip).

•Maximumflowratecorrespondingtothepressureinthecabinundernormaloperatingconditions.

•Someairlinesmayequiptheiraircraftwithonboardelectricalpower.YoumayhaveanopportunitytorequestaseatwithapowerportwhichcanbeusedtopoweryourInogenOne®G3.However,availabilityvariesbyairline,typeofaircraftandclassofservice.Youshouldcheckwithyourairlinesforavailabilityandalwaysplanonhavingsufficientbatterypowerfornolessthan150%oftheexpecteddurationoftheflight.Airlinesmayhavespecificrequirementsforbatterylifeduration,socheckwiththeairlinebeforetraveling.

•YourDCPowerCableisequippedwithacommonlyusedcigarettelighteradapter.However,aircraftsusedifferentpowerportconfigurationsanditisdifficulttodeterminewhichtypeofpowerportcompatibilityyouraircraftmaysupply.It’sagoodideatopurchaseanadapter,foundatelectronicandtravelstores.

Before Your FlightHerearesomethingstokeepinmindthedayyourflightdeparts:

•EnsureyourInogenOne®G3isclean,ingoodconditionandfreefromdamageorothersignsofexcessivewearorabuse.

•BringenoughchargedbatterieswithyoutopoweryourInogenOne®G3fornolessthan150%oftheexpecteddurationoftheflight,plusaconservativeestimateofunanticipateddelays.

•Regional/Commuterairlinesdonotofferonboardelectricalpower.Ifyourtravelplanscallforflightsonregionalairlines,bringenoughchargedbatterieswithyoutopoweryourInogenOne®G3fornolessthan150%oftheexpecteddurationoftheflight,plusaconservativeestimateofunanticipateddelays.

•Arriveattheairportearly.AirportsecurityscreeningpersonnelmayrequireextratimetoinspectyourInogenOne®G3.

Page 22: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

18

•Whilewaitingtoboardyourflight,youmaybeabletoconservebatterypowerbyusingtheACPowerSupplytopoweryourInogenOne®G3fromanelectricaloutletintheairportterminalifavailable.

•YoushouldinformtheairlineyouwillbeusingyourInogenOne®G3.Haveyourphysicianletterwithyouandreadyforinspectionifrequested.

During Your Flight1. Ifusingairlinepowerport,removetheBatteryfromtheInogenOne®G3Oxygen

Concentrator.Duetoaircraftpowerlimitations,theACPowerSupplycannotbeusedtochargetheInogenOne®G3Batterywhenonboardaircraft.

2. ConnecttheDCPowerPlugforavailableairlinepower.Checkwithairlinepersonneltoensurecompatibility.

•Duringtaxi,takeoffandlanding,stowyourInogenOne®G3undertheseatinfrontofyou.YourInogenOne®G3willfituprightundermostairlineseats.However,ifitdoesn’tfityoumayturnitonitssidewithventsfacingup.

•ItisnotnecessarytoturnoffyourInogenOne®G3duringtaxi,takeoffandlandingifyourphysician’swrittenstatementrequiresyoureceiveoxygenduringtheseperiods.

CAUTION Achangeinaltitude(forexample,fromsealeveltomountains)mayaffecttotaloxygenavailabletothepatient.TheInogenOne®G3hasbeenverifiedtoprovideoxygentospecificationupto10,000ft.(3048m).Consultyourphysicianbeforetravelingtohigherorloweraltitudestodetermineifyourflowsettingsshouldbechanged.

After Your Flight•Remembertorechargeadditionalbatteriesyoumayhaveusedpriortoyournextflight.

Traveling By Bus, Train or BoatContactyourcarriertofindoutaboutpowerportavailability.

Battery Operating InstructionsEnsurethatthebatteryisinplaceandcharged.DisconnecttheInogenOne®G3fromitspowersource.WhiletheInogenOne®G3isoperatingonbatterypower,thebatterywilldischarge.Thedisplaywillindicatetheestimatedremainingpercentage(%)orminutesofuse.

Page 23: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

19

! C

English

OperatingInstructionsChapter3

Whentheconcentratordetectsthatthebatterylifeislow,withlessthan10minutesremaining,alowpriorityalertwillsound.Whenthebatteryisempty,thealertwillchangetoahighpriority.

Whenbatterylifeislow,dooneofthefollowing:• PlugtheInogenOne®G3intoanACorDCpowersourceusingtheACpowersupplyor DCcable.•ReplacethebatterywithachargedbatteryafterturningofftheInogenOne®G3 (bypressingtheON/OFFbutton).Toremovebatterypressandholdthebatterylatch buttonandslidebatteryofftheconcentrator.• Ifthebatteryisdrained,chargethebatteryorremoveitfromtheconcentrator.

IftheInogenOne®G3isbeingpoweredbytheACpowersupplyorDCpower,batterieswillchargeduringoperation.LeavingyourInogenOne®G3pluggedinpastthefullchargetimewillnotharmtheconcentratororthebattery. WARNING Itistheresponsibilityofthepatienttoperiodicallycheckthebatteryand replaceasnecessary.Inogenassumesnoliabilityforpersonschoosingnotto adheretomanufacturersrecommendations.

Normal Battery ChargingToensurethatyourbatteryisproperlycharging,inspectthatthecorrectACandDCpoweroutputplugadapterisbeingusedandthattheadapterisproperlyinsertedintothepoweroutlet.Observethedisplayorlightsthatindicatechargingstatus.

NOTE Whenstartingtochargeafullydischargedbattery,thechargingprocessmay startandstopduringthefirstfewminutes.

Battery Care and MaintenanceYourInogenOne®G3LithiumIonBatteryrequiresspecialcaretoensureproperperformanceandlonglife.UseonlyInogenOne®G3BatterieswithyourInogenOne®G3Concentrator.

Keep DryAlwayskeepliquidsawayfrombatteries.Ifbatteriesbecomewet,discontinueuseimmediatelyanddisposeofbatteryproperly.

Page 24: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

20

Effect of Temperature on Battery PerformanceTheInogenOne®G3singlebatterypowerstheInogenOne®G3Concentratorupto4.7hoursundermostenvironmentalconditions.Toextendtherun-timeofyourbattery,avoidrunningintemperatureslessthan41˚F(5˚C)orhigherthan95˚F(35˚C)forextendedperiodsoftime.

Battery Time Remaining ClockTheInogenOne®G3continuouslydisplaysbatterytimeremaining.Thisdisplayedtimeisonlyanestimateandtheactualtimeremainingmayvaryfromthisvalue.

Please Follow These Important Guidelines to Maximize Battery Performance and Life:• Storebatteryinacool,dryplace.Storewithachargeof40-50%.• Ifusingmultiplebatteries,makesurethateachbatteryislabeled(1,2,3orA,B,C,etc.) androtateonaregularbasis.Batteriesshouldnotbeleftdormantformorethan 90daysatatime.

Page 25: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

21

Mode

PowerText

Display Icons

TheInogenOne®G3displayisdividedintothreeareas.Theupperleftcornerofthedisplayshowsthebreathdetectionalertstatus.Thelowerleftcornerindicatespowersourceandbatterychargelevel.Therightsideofthedisplaycontainstextinformation,suchasflowsetting,batterytimeremaininganderrornotifications.

Power Status Icons

Theseiconsareexamplesofthoseshowninthedisplay’spowerstatuswindowwhentheInogenOne®G3isoperatingonbatterypower.

Inogen One® G3 Oxygen Concentrator Audible and Visible Signals

AudibleandVisibleSignalsChapter4

! C

! C

! C

! C

Meaning

Batteryisempty.

Batteryhaslessthan10%chargeremaining.Thisiconflashes.

Batteryhasapproximately40%to50%chargeremaining.

Batteryisfull.

Icon

4English

Page 26: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

22

Power Status Icons (continued)TheiconsbelowareexamplesofthoseshownwhentheInogenOne®G3isoperatingfromanexternalpowersupplyandchargingthebattery.Thelightningboltindicatesthatanexternalpowersupplyisconnected.

! C

! C

! CMeaning

Batteryischargingwithchargelevelbetween60%and70%.

Thebatteryisfullychargedandischargingasnecessarytomaintainitscharge.

Batteryischargingwithchargelevellessthan10%.

TheInogenOne®G3isoperatingfromanexternalpowersourcewithnobatterypresent.

Icon

Mode IconsThesearetheiconsshowninthedisplay’smodewindow.

Display Text

NOTE Whentwoconditionsoccursimultaneously,theconditionwiththehigherprioritywillbedisplayed.

Meaning

Thebreathdetectionaudiblealarmhasbeenenabled.

Thebreathdetectionaudiblealarmisdisabled.Thisisthedefaultcondition.

Icon

! C

Page 27: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

23AudibleandVisibleSignalsChapter4

English

Informational MessagesThefollowinginformationdisplaysarenotaccompaniedbyanyaudiblefeedbackoranyvisualchangeintheindicatorlights.

Condition/Action/Explanation

TheInogenlogoisdisplayedatstartup.

Displayedduringwarmup.“X”representstheselectedflowsetting(eg.,Setting2).

Defaultdisplaywhenoperatingonbatterypower.“X”representstheselectedflowsetting(e.g.,Setting2).“HH:MM”representstheapproximatetimeremainingonthebatterycharge(e.g.,1:45).

Defaultdisplaywhenoperatingonanexternalpowersupplyandthebatteryischarging.“xx%”representsthepercentbatterycharge(e.g.,86%).

Defaultdisplaywhenthebatteryisnotchargingorwhenthetimeremainingisnotavailablefromthebattery.

Displayedwhentheconcentratorispluggedinandbeingusedtochargeabattery(notbeingusedforoxygenproduction).Itisnormaltoseeafullychargedbatteryreadbetween95%and100%whenexternalpowerisremoved.Thisfeaturemaximizestheusefullifeofthebattery.

Message Display & Text

SettingXPleaseWait

SettingXBatteryHH:MM

SettingXChargingxx(or)BatteryFull

SettingXBatteryxx%

Chargingxx%(0r) BatteryFull

Notifications

! C

WARNING Audiblenotifications,rangingfrom55dbato65dbadependingontheusersposition,aretowarntheuserofproblems.Toinsurethataudiblenotificationsmaybeheard,themaximumdistancethattheusercanmoveawayfromitmustbedeterminedtosuitthesurroundingnoiselevel.

Page 28: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

24

Low Priority AlertsThefollowinglowpriorityalertmessagesareaccompaniedbyadouble beepandasolid red or yellow light.Coloroflightwillvarydependingonmodel.

Condition/Action/Explanation

Batterypowerislow,withlessthan10minutesremaining.Attachexternalpowersupplyorpowerdownandinsertafullychargedbattery.

Columnmaintenanceisrequiredwithin30days.Contactyourequipmentprovidertoarrangeforservice.

Message Display & Text

BatteryLow AttachPlug

ReplaceColumnsorO2ServiceSoon(dependingonmodel)

Notifications (continued)TheInogenOne®G3monitorsvariousparametersduringoperationandutilizesanintelligentalarmsystemtoindicateamalfunctionoftheconcentrator.Mathematicalalgorithmsandtimedelaysareusedtoreducetheprobabilityoffalsealarmswhilestillensuringpropernotificationofanalarmcondition.Ifmultiplealarmconditionsaredetected,thehighestpriorityalarmwillbedisplayed.Thefollowingnotificationmessagesareaccompaniedbyasingle, short beep.

Condition/Action/Explanation

On/Offbuttonhasbeenpressedfortwoseconds.Concentratorisperformingsystemshutdown.

AudibleAlarmbuttonhasbeenpressedforfiveseconds.

Message Display & Text

PleaseWait ShuttingDown

HH:MM Vx.x:SerialNumber

Page 29: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

25AudibleandVisibleSignalsChapter4

English

Low Priority Alerts (continued)Condition/Action/Explanation

Batteryerrorhasoccurred.Checktheconnectionofyourbatteryandensurethatitisproperlyattachedandlatchedonconcentrator.Ifbatteryerrorrecurswithsamebattery,stopusingthebatteryandswitchtoanewbatteryorremovebatteryandoperateconcentratorusingexternalpowersupply.

Concentratorisproducingoxygenataslightlylowlevel(<82%)foraperiodof10minutes.Ifconditionpersists,contactyourequipmentprovider.

Batteryhasexceededitschargingtemperatureandcharginghasstopped.Thebatterywillnotchargewhilethisalertispresentbutwillbegintochargewhenthebatterytemperaturereturnstothenormaloperatingrange.Ifbatterychargingisdesiredsooner,removethebatteryfromtheconcentratorandallowittocoolinanopenareaforapproximately10-15minutes.Then,re-insertthebatteryintotheInogenOne®G3.Iftheproblemstillpersists,contactyourequipmentprovider.

Concentratorisproducingoxygenbutcannotreportbatterystatus.Replacebattery.Ifconditionpersists,contactyourequipmentprovider.

Theconcentratorrequiresservicingattheearliestconvenience.Theconcentratorisoperatingtospecificationandmaycontinuetobeused.Contactyourequipmentprovidertoarrangeforservice.

Theconcentrator’soxygensensorhasmalfunctioned.Youmaycontinuetousetheconcentrator.Iftheconditionpersists,contactyourequipmentprovider.

Message Display & Text

CheckBattery

OxygenLow

RemoveBatterytoCool

CommError

ServiceSoon orServiceNeeded(dependingonmodel)

SensorFail

Page 30: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

26

Condition/Action/Explanation

Concentratorhasnotdetectedabreathfor60seconds.Checkthatcannulaisconnectedtoconcentrator,therearenokinksintubingandcannulaispositionedproperlyinyournose.

Oxygenoutputconcentrationhasbeenbelow50%for10minutes.Ifconditionpersists,switchtoyourbackupoxygensourceandcontactyourequipmentprovidertoarrangeforservice.

Abreathhasbeenrecognized,butproperoxygendeliveryhasnotbeendetected.

Batteryhasexceededtemperaturelimitwhileconcentratorisrunningonbatterypower.Ifpossible,moveconcentratortoacoolerlocationorpowerunitwithanexternalpowersupplyandremovebattery.Ifconditionpersists,contactyourequipmentprovider.

Concentratorisexperiencinganerrorbutiscapableofcontinuedoperation.Removeandre-insertbatteryand/orcheckexternalpowersupplyconnections.Ifconditionpersists,contactyourequipmentprovider.

Message Display & Text

NoBreathDetect CheckCannula

OxygenError

O2DeliveryError

BatteryHOT Warning

SystemError

Medium Priority AlertsThefollowingmediumpriorityalertmessagesareaccompaniedbyatriple beep,repeatedevery25seconds,andaflashing red or yellow light.Coloroflightwillvarydependingonmodel.

Page 31: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

27

Condition/Action/Explanation

Concentratorhasinsufficientbatterypowertoproduceoxygen.Attachexternalpowersupplyorexchangebattery,thenrestartunitifnecessarybypressingOn/Offbutton.

Batteryhasexceededtemperaturelimitwhileconcentratorisrunningonbatterypower.Concentratorhasstoppedproducingoxygen.Ifpossible,moveconcentratortoacoolerlocation,thenturnpoweroffandbackon.Ensureairintakeandoutletventshaveclearaccessandparticlefiltersareclean.Ifconditionpersists,switchtoabackupsourceofoxygenandcontactyourequipmentprovider.

Concentratortemperatureistoohighandoxygenproductionisshuttingdown.Ensureairintakeandoutletventshaveclearaccessandparticlefiltersareclean.Ifconditionpersists,switchtoabackupsourceofoxygenandcontactyourequipmentprovider.

Message Display & Text

BatteryEmpty AttachPlug

BatteryHOT

SystemHOT

High Priority Alerts

CAUTION IfyouarenotneartheInogenOne®G3youmaynotbeabletohearorseethehighpriorityalerts.MakesuretheInogenOne®G3isinalocationwherethealertsandalarmswillberecognizediftheyoccur.

Thefollowinghighpriorityalertmessagesareaccompaniedbyafive beep pattern,repeatedevery10secondsandaflashing red or yellow light.Coloroflightwillvarydependingonmodel.

AudibleandVisibleSignalsChapter4

English

Page 32: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

28

Condition/Action/Explanation

Thismayresultfromtheconcentratorbeingstoredinacoldenvironment(below0ºC(32ºF)).Movetoawarmerenvironmenttoallowtheunittowarmupbeforestartingit.Ifconditionpersists,switchtoabackupsourceofoxygenandcontactyourequipmentprovider.

Concentratorhasstoppedproducingoxygenandisshuttingdown.Youshould:

1. Switchtobackupoxygensource2.Contactyourequipmentprovider

Message Display & Text

SystemCOLD

SystemError orError(###)(dependingonmodel)

High Priority Alerts (continued)

Page 33: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

29

Troubleshooting

TroubleshootingChapter5

Useexternalpowersupplyorreplacebatterywithonethatisfullycharged

Checkpowersupplyconnectionandverifygreenlightissolid

CheckDCCableconnectionattheConcentratorandatcigarettelighterorauxiliaryDCpowersource

Contactyourequipmentprovider

Batteryisdischargedornobatteryispresent

ACPowersupplyisnotconnectedproperly

DCCableisnotconnectedproperly

Malfunction

ConcentratordoesnotpoweronwhenOn/Offbuttonispressed

Recommended Solution

RefertoChapter4

Possible Cause

RefertoChapter4

Problem

Anyproblemaccompaniedbyinformationonconcentratordisplay,indicatorlightsand/oraudiblesignals

Solutionstosomepossibleissuesyoumayencounteraredescribedinthissection.

Inogen One® G3 Oxygen Concentrator

5English

Page 34: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

30

PressOn/Offbuttontopowerconcentrator

Checkcannulaanditsconnectiontoconcentratornozzle

Concentratorisnotpoweredon

Cannulaisnotconnectedproperlyoriskinkedorobstructed

Nooxygen

Recommended SolutionPossible CauseProblem

Troubleshooting (continued)

Page 35: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

31

Cannula Replacement

Yournasalcannulashouldbereplacedonaregularbasis.Consultwithyourphysicianand/orequipmentproviderand/orcannulamanufacturer’sinstructionsforreplacementinformation.Asinglelumencannulaupto25feetinlengthisrecommendedtoensureproperbreathdetectionandoxygendelivery.

CAUTION Nasalcannulashouldberatedfor5litersperminute(e.g.Salter16SOFT)toensureproperpatientusageandoxygendelivery.

Case CleaningYoumaycleantheoutsidecaseusingaclothdampenedwithamildliquiddetergent(suchasDawnTM)andwater.

WARNING DonotsubmersetheInogenOne®G3oritsaccessoriesinwaterorallowwatertoenterintothecase;thismayleadtoelectricalshockand/ordamage.

WARNING DonotusecleaningagentsotherthanthosespecifiedinthisUserManual.Donotusealcohol,isopropylalcohol,ethylenechlorideorpetroleumbasedcleanersonthecasesorontheparticlefilters.

Cleaning, Care and Maintenance

Cleaning,CareandMaintenanceChapter6

6

! C

! C

English

Page 36: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

32

Filter Cleaning and ReplacementBothparticlefiltersmustbecleanedweeklytoensuretheeaseofairflow.Removefiltersfromthefrontofthedevice.Cleantheparticlefilterswithamildliquiddetergent(suchasDawnTM)andwater;rinseinwateranddrybeforereuse.

NOTE Itmaybenecessarytocleantheparticlefiltersmoreoftenindustyenvironments.

TopurchaseadditionalparticlefilterscontactyourequipmentproviderorInogen.

Output FilterTheoutputfilterisintendedtoprotecttheuserfrominhalationofsmallparticlesintheproductgasflow.TheInogenOne®G3includesanoutputfilterconvenientlylocatedbehindtheremovablecannulanozzlefitting.Inogenrequiresthatthisfilterbereplacedbetweenpatients.

TheoutputfiltermaybereplacedbytheequipmentproviderorbytheownerusingtheOutputFilterReplacementKit(RP-107).

TheInogenOneG3Concentratormustbecleanedanddisinfectedaspertheaboveinstructionsforeachnewpatient.Nospecialmaintenanceneedstobecarriedoutbythepatient.YourequipmentproviderperformsmaintenanceoperationstoassurecontinuedreliableservicefromyourInogenOneG3.Themanufacturer’sinstructionsforthepreventativemaintenanceofthedevicesaredefinedintheservicemanual.Allworkmustbeperformedbytrainedtechnicianscertifiedbythemanufacturer.

DC Input Cable Fuse ReplacementTheCigaretteLighterDCpowerplugcontainsafuse.IftheDCinputcableisbeingusedwithaknowngoodpowersourceandtheunitisnotreceivingpower,thefusemayneedtobereplaced. Toreplacethefuse,followtheseinstructionsandrefertothephotographbelow.

1. Removethetipbyunscrewingtheretainer.Useatoolifnecessary.2. Removetheretainer,tipandfuse.3. ThespringshouldremaininsidetheCigaretteLighterAdapterhousing.Ifthespring

isremoved,makesuretoreplacethespringfirstbeforeinsertingthereplacementfuse.

Page 37: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

33Cleaning,CareandMaintenanceChapter6

English

4. Installareplacementfuse,InogenRP#125(BUSSMDA-12)andreassemblethetip.Ensuretheretainerringisproperlyseatedandtightened.

Inogen One® G3 Column Change Procedure

NOTE Columnchangeinstructionsareonlytobeusedwhenmaintenanceis requiredandarenotintendedforpracticepurpose.

1. TurnofftheInogenOne®G3concentratorbypressingthepowerbuttontoshutdownthedevice.

2.RemovetheInogenOne®G3concentratorfromthecarrybag.

3.RemovethebatteryfromtheInogenOne®G3concentrator.

4.TurnInogenOne®G3oversothatitisstandingintheupsidedownpositionwherethedisplaypanelis

facingdownandthebottomoftheInogenOne®G3isfacingup.

5.Therearetwocolumns(metaltubes)onthe InogenOne®G3concentratorandoneislocatedon eachsideofthedevice.

FuseCigarette Adapter Plug Retainer Tip Retainer

CAUTION Forcontinuedprotectionagainstriskoffire,onlyusefusespecified.

Page 38: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

34

Inogen One® G3 Column Change Procedure (continued) 6.Removecolumnbypressingthelatchbutton withthumborfinger.

7. Whileholdingthebuttonopen,slidethecolumn(metaltube)outbygrippingtheexposedalumi-numsurfacewithotherhand.

8.Removethecolumn(metaltube)completelyfromtheInogenOne®G3.

9.Repeatsteps6-8toremovetheothercolumn(metaltube).

Column(metaltube)Installation10.Removedustcapsofnewcolumn(metaltube).

Makesurethereisnodustordebriswherethedustcapswerelocated.

Openandunlocked

Page 39: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

35Cleaning,CareandMaintenanceChapter6

English

Inogen One® G3 Column Change Procedure (continued)

11.Insertcolumn(metaltube)intotheInogenOne®G3concentrator.Donotleavethecolumnexposed;

itshouldbeinsertedintotheInogenOne®G3as

soonasthedustcapshavebeenremoved.

12.Pushthecolumn(metaltube)intothedevicesuchthatthecolumn(metaltube)isfullyseatedintotheInogenOne®G3concentrator.Thespringloadedlatchbuttonshouldfullyreturntotheclosedposition.

13.Repeatsteps10-12toinstalltheothercolumn(metaltube).

14.ConnecttheACpowersupplycordtotheInogen One®G3andplugthepowersupplyACcordintoan electricaloutlet.Do Not Power on the Inogen One® G3 concentrator.

15.Pressandholdthelightbuttonfor10secondsandscreenwilldisplay“sievereset”

message.Releasebuttononcemessageisdisplayedonscreen.

16.Pressthebellbuttononceandscreenwilldisplay“sieveresetsuccess”.

17.PressthepowerbuttontoturnontheInogenOne®G3,andusenormally.

WARNING Donotuseanycolumnsotherthanthosespecifiedinthisusermanual.Theuseofnon-specifiedcolumnsmaycreateasafetyhazardand/orimpairequipmentperformanceandwillvoidyourwarranty.

! C

Closedandlocked

Page 40: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

36

Disposal of Equipment and AccessoriesFollowyourlocalgoverningordinancesfordisposalandrecyclingoftheInogenOne®G3andaccessories.IfWEEEregulationsapply,donotdisposeofinunsortedmunicipalwaste.WithinEurope,contacttheEUAuthorizedRepresentativefordisposalinstructions.Thebatterycontainslithiumioncellsandshouldberecycled.Thebatterymustnotbeincinerated.

Maintenance Items List• InogenOne®G3singlebattery(model#BA-300)• InogenOne®G3doublebattery(model#BA-316)• Replacementintakeparticlefilters(model#RP-300)-4settingmodel• Replacementintakeparticlefilters(model#RP-301)-5settingmodel• OutputFilterReplacementKit(model#RP-107)• InogenOne®G3columns(model#RP-320)-4settingmodel• InogenOne®G3columns(model#RP-321)-5settingmodel

Forassistance,ifneeded,insettingup,using,maintaining,ortoreportunexpectedoperationorevents,contactyourequipmentprovider,ormanufacturer.

Other Service and Maintenance WARNING DonotdisassembletheInogenOne®G3oranyoftheaccessoriesor

attemptanymaintenanceotherthantasksdescribedinthisusermanual;disassemblycreatesahazardofelectricalshockandwillvoidyourwarranty.Donotremovethetamperevidentlabel.Foreventsotherthanthosedescribedinthismanual,contactyourequipmentproviderforservicingbyauthorizedpersonnel.

CAUTION DonotuselubricantsontheInogenOne®G3oritsaccessories.

! C

Page 41: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

37

Symbols Used On Concentrator and Accessories

SymbolsUsedonConcentratorandAccessoriesChapter7

7English

Meaning

Awarningindicatesthatthepersonalsafetyofthepatientmaybeinvolved.Disregardingawarningcouldresultinsignificantinjury.

Acautionindicatesthataprecautionorserviceproceduremustbefollowed.Disregardingacautioncouldleadtoaminorinjuryordamagetoequipment.

SeeUserManualforInstructions.

U.S.FederalRegulationRestrictsthisDevicetoSalebyOrderofPhysician.MayalsobeapplicableinotherCountries.

ACPower

DCPower

NoSmokingwhiledeviceisinuse.

NoOpenFlames(Concentrator);Donotincinerate(Battery).Refertoinstructionmanual/booklet.

ManufacturerAuthorizedRepresentativeintheEuropeanCommunity

Symbol

WARNING

CAUTION

ONLY

ONLY

ONLY

ONLY

ONLY

! C

EC REP

Page 42: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

38

Meaning

KeepDry

IndoororDryLocationUseOnly,DoNotGetWet

UseNoOilorGrease

DoNotDisassemble(contactyourequipmentproviderforservicingbyauthorizedpersonnel)

DoNotDisposeofInUnsortedMunicipalWaste

TypeBFAppliedPart,NotIntendedforCardiacApplication

ClassIIDevice

ElectricalSafetyAgencyCertificationLogo

CompliesWithApplicableEUDirectivesIncludingMedicalDeviceDirective

Symbol

ONLYONLY

ONLY

ONLYONLY

0123

93-01213-00-00 rev1

0123

COLOR SUMMARY: 1. Black

User Interface Label

Meaning

ON/OFFButton

DisplayBacklightButton

IncreaseFlowSetting

DecreaseFlowSetting

AudibleAlarmButton

Symbol

Page 43: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

39InogenOne®G3SystemSpecificationsChapter8

English

Inogen One® G3 ConcentratorDimensions:Withsinglebattery

L/W/H:8.75in.(22.2cm.)/3.0in.(7.6cm.)/7.25in.(18.4cm.)L/W/H:8.75in.(22.2cm.)/3.0in.(7.6cm.)/8.25in.(21cm.)

Weight: 4.8pounds(2.19Kg.)(includessinglebattery)Noise: 39dBA(onsetting2)for5settingmodel

42dBA(onsetting2)forthe4settingmodelWarm-UpTime: 2minutesOxygenConcentration: 90%-3%/+6%atallsettingsFlowControlSettings: 4settings:1to4or5settings:1to5Power:ACPowerSupply:

DCPowerCable:RechargeableBattery:

ACInput: 100to240VAC50to60Hz

Auto-Sensing: 1.0A

DCInput: 13.5-15.5VDC,10AMax.Voltage: 12.0to16.8VDC

BatteryDuration: Upto4.7hourswithsinglebatteryUpto10hourswithdoublebattery

BatteryChargingTime: Upto5hoursforasinglebatteryUpto8hoursforadoublebattery

EnvironmentalRangesIntendedforUse:

Temperature:41to104˚F(5to40˚C)Humidity:0%to95%,non-condensingAltitude:0to10,000ft(0to3048meters)

EnvironmentalRangesIntendedforShippingandStorage:

Temperature:-13to158˚F(-25to70˚C)Humidity:0%to95%,non-condensingStoreinadryenvironmentAltitude:0to10,000ft(0to3048meters)

Transportation: KeepDry,HandleWithCare

Inogen One® G3 System Specifications8

Page 44: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

40

Inogen One® G3 Concentrator (continued)

Classifications

ModeofOperation: ContinuousDutyTypeofProtectionAgainstElectricalShock:

ClassII

DegreeofProtectiontoConcentratorComponentsAgainstElectricalShock:

TypeBFNotintendedforcardiacapplication

DegreeofProtectiontoConcentratorComponentsAgainstIngressofWaterWhileUsedOutsideofCarryBag:

IP20-Notprotectedfromdrippingwater.Protectedagainstingressofsolidobjects>12.5mm.

DegreeofProtectiontoConcentratorComponentsAgainstIngressofWaterWhileUsedInsideofCarryBag:

IP22-Verticallydrippingwatershallhavenoharmfuleffect&protectagainstingressofsolidobjects>12.5mm.diameterwhentheenclosureistiltedatanangleupto15˚fromitsnormalposition

DegreeofProtectiontoConcentratorExteriorProvidedbytheCarryBag:

IP02-Verticallydrippingwatershallhavenoharmfuleffectwhentheenclosureistiltedatanangleupto15˚fromitsnormalposition

DegreeofSafetyforApplicationinPresenceofAnestheticGases:

Notsuitableforsuchapplication

TestedbyIndependentLaboratory:

Safety:IEC60601-1CAN/CSAC22.2No.60601-1ElectromagneticCompatibility:IEC60601-1-2RTCADO160

ELECTROMAGNETIC COMPATIBILITY ThisCEMarkedequipmenthasbeentestedandfoundtocomplywiththeEMClimitsfortheMedicalDeviceDirective93/42/EEC[EN55011ClassBandEN60601-1-2].Theselimitsaredesignedtoprovidereasonableprotectionagainstharmfulinterferenceinatypicalmedicalinstallation.

Page 45: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

41

PortableandmobileRFcommunicationsequipmentshouldbeusednoclosertoanypartofthedevice,includingcables,thantherecommendedseparationdistancecalculatedfromtheequationapplicabletothefrequencyofthetransmitter.

Recommendedseparationdistance:d=1.2√P150kHzto80MHzd=1.2√P80MHzto800MHzd=2.3√P800MHzto2.5GHz

WherePisthemaximumoutputpowerratingofthetransmitterinwatts(W)accordingtothetransmittermanufactureranddistherecommendedseparationdistanceinmeters(m).

FieldstrengthsfromfixedRFtransmitters,asdeterminedbyanelectromagneticsitesurveya,shouldbelessthanthecompliancelevelineachfrequencyrangeb.

Interferencemayoccurinthevicinityofequipmentmarkedwiththefollowingsymbol:

3Vrms150kHz to80MHz

3V/m80MHz to2.5GHz

3Vrms

3V/m

ConductedRFIEC61000-4-6

RadiatedRFIEC61000-4-3

Compliance Level

IEC 60601 Test Level

Immunity Test Electromagnetic Environment - Guidance

Guidance and Manufacturer’s Declaration - Electromagnetic Immunity: TheConcentratorisintendedforuseintheelectromagneticenvironmentspecifiedbelow.TheuseroftheConcentratorshouldmakesureitisusedinsuchanenvironment.

Floorsshouldbewood,concreteorceramictile.Iffloorsarecoveredwithsyntheticmaterial,therelativehumidityshouldbeatleast30%.

Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.

Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.

Electrostaticdischarge(ESD)

IEC61000-4-2

Electricalfasttransient/burst

EC61000-4-4

Surge

IEC61000-4-5

±6kVcontact

±8kVair

±6kVcontact

±8kVair

±2kVforpowersupplylines

±1kVforinput/outputlines

±2kVforpowersupplylines

±1kVforinput/outputlines

±1kVline(s)toline(s)

±2kVline(s)toearth

±1kVline(s)toline(s)

±2kVline(s)toearth

Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.Iftheuserofthe[MEEQUIPMENTorMESYSTEM]requirescontinuedoperationduringpowermainsinterruptions,itisrecommendedthatthe[MEEQUIPMENTorMESYSTEM]bepoweredfromanuninterruptedpowersupplyorabattery.

Powerfrequencymagneticfieldsshouldbeatlevelscharacteristicofatypicallocationinatypicalhospitalorhomeenvironment.

Voltagedips,shortinterruptionsandvoltagevariationsonpowersupplyinputlines

IEC61000-4-11

Powerfrequency(50/60Hz)magneticfield

IEC61000-4-8

<5%UT(>95%dipinUT)for0.5cycle

40%UT(60%dipinUT)for5cycles

70%UT(30%dipinUT)for25cycles

<5%UT(>95%dipinUT)for5sec

<5%UT(>95%dipinUT)for0.5cycle

40%UT(60%dipinUT)for5cycles

70%UT(30%dipinUT)for25cycles

<5%UT(>95%dipinUT)for5sec

3A/m 3A/m

English

Page 46: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

42

NOTE At80MHzand800MHz,thehigherfrequencyrangeapplies.

NOTE Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreflectionfromstructures,objects,andpeople.

NOTE UTisthea.c.mainvoltagepriortoapplicationofthetestlevel.

a:Fieldstrengthfromfixedtransmitters,suchasbasestationsforradio(cellular/cordless)telephonesandlandmobileradios,amateurradio,AMandFMradiobroadcastandTVbroadcastcannotbepredictedtheoreticallywithaccuracy.ToassesstheelectromagneticenvironmentduetofixedRFtransmitters,anelectromagneticsitesurveyshouldbeconsidered.IfthemeasuredfieldstrengthinthelocationinwhichtheconcentratorisusedexceedstheapplicableRFcompliancelevelabove,theconcentratorshouldbeobservedtoverifynormaloperation.Ifabnormalperformanceisobserved,additionalmeasuresmaybenecessary,suchasre-orientingorrelocatingthedevice.

b:Overthefrequencyrange150kHzto80MHz,thefieldstrengthsshouldbelessthan3V/m.

Recommended Separation Distances between Portable and Mobile RF Communications Equipment and This Device:

ThisconcentratorisintendedforuseinanelectromagneticenvironmentinwhichradiatedRFdisturbancesarecontrolled.TheuseroftheconcentratorcanhelppreventelectromagneticinterferencebymaintainingaminimumdistancebetweenportableandmobileRFcommuni-cationsequipment(transmitters)andthisconcentratorasrecommendedbelow,accordingtothemaximumoutputpowerofthecommuni-cationsequipment.

150 kHz to 80 MHzd=1.2√P

0.120.381.23.812

0.120.381.23.812

0.230.732.37.323

0.010.11

10100

80 MHz to 800 MHzd=1.2√P

800 MHz to 2.5 GHzd=2.3√P

Separation Distance According to Frequency of Transmitter (M)Rated Maximum Power Output of Transmitter (W)

Fortransmittersratedatamaximumoutputpowernotlistedabove,therecommendedseparationdistancedinmeters(m)canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter,wherePisthemaximumoutputpowerratingofthetransmitterinwatts(W)accordingtothetransmittermanufacturer.

Guidance and Manufacturer’s Declaration – Electromagnetic EmissionsTheconcentratorisintendedforuseintheelectromagneticenvironmentspecifiedbelow.Theuseroftheconcentratorshouldassurethatitisusedinsuchanenvironment.

NOTE At80MHzand800MHz,theseparationdistanceforthehigherfrequencyrangeapplies.

NOTE Theguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreflectionfromstructures,objects,andpeople.

Emissions Test Compliance Electromagnetic Environment - Guidance

RFemissionsCISPR11

Group1 TheconcentratorusesRFenergyonlyforitsinternalfunction.ThereforeitsRFemissionsareverylowandnotlikelytocauseanyinterferenceinnearbyequipment.

RFemissionsCISPR11

ClassB Theconcentratorissuitableforuseinallestablishments,includingdomesticestablishmentsandthosedirectlyconnectedtothepubliclow-voltagepowersupplynetworkthatsuppliesbuildingsusedfordomesticpurposes.HarmonicEmissions

IEC61000-3-2ClassA

Voltagefluctuations/flickeremissionsIEC61000-3-3

Complies

Page 47: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3
Page 48: user manualagedcare.airliquidehealthcare.com.au/wp-content/uploads/2018/08/User... · 18 Battery Operating Instructions 19 Battery Care and Maintenance Chapter 4 21 Inogen One® G3

PN 96-07335-00-01 A

Inogen, Inc.326 Bollay DriveGoleta, CA 93117Toll Free: 877-466-4362 +1-805-562-0515 (Outside the USA) E-mail: [email protected] www.inogen.com

®

©2017 Inogen. All rights Reserved.