2
User Manual RC-18 Single-Use Temperature Data Logger S/N:EF116A100001 LED indicator Barcode label USB port Waterproof plastic bag Configuration Info Button(Start/Status/Mark/Stop) Start/Stop Info Mark Info Status Request Paste Area Note: The picture is for reference only. Any discrepancy in future, please adhere to the final product. Back Front Note: It is suggested storing the logger at room temperature before started. Overview: The temperature data logger is mainly used for monitoring and logging temperature in storage and ship- ment of food and pharmaceuticals, such as cooler Boxes, refrigerated trucks, containers, etc. The logger can connect to a computer via its usb port and export a pdf report. It features an internal sensor and a cr2450 lithium battery with whole Protection grade up to ip67. A barcode is prepared for scanning. Illustration Technical Parameters All the parameters are preconfigured before the logger leaves the factory. Some can be customized based on your needs. Parameter Instructions Memory capacity: 16000 points at maximum. Actual record capacity may be different due to different log cycles. Start delay: When the logger is started, it delays for a period of time to start logging. The delay time can be customized. Alarm thresholds: Alarm zone can be set according to the alarm thresholds. The logger supports one upper limit and one lower limit. H Ultra-high temperature alarm zone: the value is greater than H L Ultra-low temperature alarm zone: the value is less than L Alarm type: Single The logger records the single time for continuous out-of-limit events Cumulative The data logger records the cumulative time of all the out-of-limit events Alarm delay: The logger does not alarm immediately when the temperature is within the alarm zone. It begins to alarm only when the alarm delay time elapse Operating Instructions Temperature measuring range: -30°C~70°C Resolution: 0.1°C Alarm threshold: default (customizable) Storage temperature: -30°C~70°C Data interface: USB2.0 Temperature accuracy: 0.5F/0.9F(-20°C~40°C); 1F/1.8F(others) Report type: PDF format Sensor type: internal NTC Battery: built-in CR2450 wide temperature lithium battery Memory capacity: 16000 points (MAX) Battery life: 2 years (stored and used under normal temperature environment) Log cycle: 6, 15, 30, 60, 90, 120, 180days (customizable) Protection grade: IP67 Size: 80mm (L) x 48mm (W) x 5.8mm (H) Alarm type: single or cumulative Weight: approx. 15g 12 S/N:EF116A100001 S/N:EF116A100001 Press and hold the button for 5 seconds to start logging. Green LED will flash 5 times. Peel othe barcode label for further use. Postbus 318 2180 AH Hillegom Nederland Tel: +31 (0)252-52 34 34 Fax: +31 (0)87-784 77 32 www.cargotemp.nl [email protected]

Upload: others

Post on 18-Jun-2020

23 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: User Manual RC-18 Single-Use Temperature Data Logger User ... · User Manual RC-18 Single-Use Temperature Data Logger Press and hold the button for 5 seconds to stop logging. Red

User Manual RC-18 Single-Use Temperature Data Logger

12

Note: The picture is for reference only. Any discrepancy in future, please adhere to the final product.

Note: It is suggested storing the logger at room temperature before started.

boxes, refrigerated trucks, containers, etc.

All the parameters are preconfigured before the logger leaves the factory. Some can be customized based on your needs.

Memory capacity: 16000 points at maximum. Actual record capacity may be di�erent due to di�erent log cycles.

Start delay:

Alarm thresholds: Alarm zone can be set according to the alarm thresholds. The logger supports one upper limit and one lower limit.

Alarm delay:

Ultra-high temperature alarm zone: the value is greater than H

Ultra-low temperature alarm zone: the value is less than L

H

L

Single

Alarm type:

Peel o� the barcode label for further use.

S/N:EF116A100001

User Manual RC-18 Single-Use Temperature Data Logger

Overview

Technical Parameters

LED indicator

Barcode label

Back Front

Temperature measuring range: -30℃~70℃ ℃

Alarm threshold: default (customizable)

Storage temperature: -30℃~70℃

Data interface: USB2.0

Temperature accuracy: 0.5℃/0.9℉(-20℃~40℃); 1℃/1.8℉(others)

Report type: PDF format

Sensor type: internal NTC

Memory capacity: 16000 points (MAX)

2 years

(stored and used under normal temperature environment)

Log cycle: 6, 15, 30, 60, 90, 120, 180days (customizable)

IP67

Size: 80mm (L) x 48mm (W) x 5.8mm (H)

Alarm type:

Weight: approx. 15g

S/N

:EF116A

100001

S/N

:EF116A

100001

USB port

Start/Stop Info Mark Info

Status Request

Paste AreaS/N:EF116A100001

LED indicator

Barcode label

USB port Waterproof plastic bag

Con�guration Info

Button(Start/Status/Mark/Stop) Start/Stop Info Mark Info

Status Request

Paste Area

Note: The picture is for reference only.Any discrepancy in future, please adhere to the final product.

Back Front

Note: It is suggested storingthe logger at room temperaturebefore started.

Overview:The temperature data logger is mainly used for monitoring and logging temperature in storage and ship-ment of food and pharmaceuticals, such as cooler Boxes, refrigerated trucks, containers, etc. The logger can connect to a computer via its usb port and export a pdf report. It features an internal sensor and a cr2450 lithium battery with whole Protection grade up to ip67. A barcode is prepared for scanning.

Illustration

Technical Parameters

All the parameters are preconfigured before the logger leaves the factory. Some can be customized based on your needs.

Parameter Instructions

Memory capacity: 16000 points at maximum. Actual record capacity may be different due to different log cycles.Start delay: When the logger is started, it delays for a period of time to start logging. The delay time can be customized.Alarm thresholds: Alarm zone can be set according to the alarm thresholds. The logger supports one upper limit and one lower limit.

H Ultra-high temperature alarm zone: the value is greater than HL Ultra-low temperature alarm zone: the value is less than L

Alarm type: Single The logger records the single time for continuous out-of-limit eventsCumulative The data logger records the cumulative time of all the out-of-limit events

Alarm delay: The logger does not alarm immediately when the temperature is within the alarm zone. It begins to alarm only when the alarm delay time elapse

Operating Instructions

Temperature measuring range: -30°C~70°C Resolution: 0.1°CAlarm threshold: default (customizable)Storage temperature: -30°C~70°CData interface: USB2.0Temperature accuracy: 0.5F/0.9F(-20°C~40°C); 1F/1.8F(others)Report type: PDF formatSensor type: internal NTCBattery: built-in CR2450 wide temperature lithium battery

Memory capacity: 16000 points (MAX)Battery life: 2 years(stored and used under normal temperature environment)Log cycle: 6, 15, 30, 60, 90, 120, 180days (customizable)Protection grade: IP67Size: 80mm (L) x 48mm (W) x 5.8mm (H)Alarm type: single or cumulativeWeight: approx. 15g

12

S/N

:EF116A

100001

S/N

:EF116A

100001

Press and hold the button for5 seconds to start logging. Green LED will �ash 5 times.

Peel o� the barcodelabel for further use.

Postbus 318 • 2180 AH Hillegom • Nederland • Tel: +31 (0)252-52 34 34 • Fax: +31 (0)87-784 77 32 www.cargotemp.nl • [email protected]

Page 2: User Manual RC-18 Single-Use Temperature Data Logger User ... · User Manual RC-18 Single-Use Temperature Data Logger Press and hold the button for 5 seconds to stop logging. Red

User Manual RC-18 Single-Use Temperature Data Logger

Insert the logger into a computer’s USB port to view the PDF report.Press and hold the button for 5 seconds to stop logging. Red LEDwill �ash 5 times. Tear the bag open and let USB port uncovered.

S/N

:EF116A

100001 S/N:EF116A100001

RC-18EF16A100001

EF16A100001-0000001

EF16A100001-0000001

The first page Other pages

Note:The data above is only used as an

example of the report. Please refer to theactual document for speci�c con�guration and

information.

Illustration

Status IndicationPress and release the button to judge the logger’s working status through the flash of the red or green LED indicator.

Report

Status

Not started

Start delay

Started-normal

Started-alarm

Started-mark

Started-out of mark limits

Stopped-normal

Stopped-alarm

Action of the indicators

Red and green LEDs flash once simultaneously.

Red and green LEDs flash once alternatively.

Green LED flashes once (It flashes once 10 seconds automatically).

Red LED flashes once (It flashes once 10 seconds automatically).

Red and green LEDs flash 3 times simultaneously.

Red and green LEDs flash 3 times simultaneously.

Green LED flashes 2 times.

Red LED flashes 2 times.

Postbus 318 • 2180 AH Hillegom • Nederland • Tel: +31 (0)252-52 34 34 • Fax: +31 (0)87-784 77 32 www.cargotemp.nl • [email protected]

1.Basic information

2.Description of the usage (customizable)

3.Configuration information

4.Alarm threshold and related statistics

5.Statistical information

6.Temperature data curve

7.Temperature data details

A.Document creation time(record stop time)

B.Alarm(Alarm status as shown in the figure)

C.The set stop mode

D.Alarm status of the alarm zone

E.Total times of exceeding alarm thresholds

F.Total time of exceeding alarm thresholds

G.Alarm delay and alarm type

H.Alarm thresholds and alarm zones

I.Stop time (actual stop mode)

J.Vertical coordinate unit of the data graph

K.Alarm threshold line (corresponding to the item L)

L.Alarm threshold

M.Data curve (Note: Red indicates ultra-high temperature;

blue indicates ultra-low temperature; black indicates others.)

N. Document name (serial number & description of usage ID)

O. Record time range in the current page

P. Records when date changes (date & temperature)

Q Records when the date does not change (time & temperature)