vaporlab™ 700 gas analyzer

14
© 2006 MSA VaporLab™ 700 Gas Analyzer

Upload: joy

Post on 23-Jan-2016

63 views

Category:

Documents


0 download

DESCRIPTION

VaporLab™ 700 Gas Analyzer. Product Description. Rack mount, turn-key gas analyzer Analyzes a complex matrix for trace quantities of toxic gases Gas chromatography is used to separate the gas of interest from background gases. Product Description. Rugged, long life Metal Oxide Detector - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: VaporLab™ 700 Gas Analyzer

© 2006 MSA

VaporLab™ 700 Gas Analyzer

VaporLab™ 700 Gas Analyzer

Page 2: VaporLab™ 700 Gas Analyzer

© 2006 MSA

Product DescriptionProduct Description

Rack mount, turn-key gas analyzer Analyzes a complex matrix for

trace quantities of toxic gases Gas chromatography is used to

separate the gas of interest from background gases

Page 3: VaporLab™ 700 Gas Analyzer

© 2006 MSA

Product DescriptionProduct Description

Rugged, long life Metal Oxide Detector

Specific chemical analysis Accurate and reproducible

measurements High Sensitivity - Detection

capability down to 5 ppb

Page 4: VaporLab™ 700 Gas Analyzer

© 2006 MSA

Product DescriptionProduct Description

Gas Chromatography Made Simple No support gas cylinders required

Scrubbed room air used as support gas No hazardous source or detector gas

Solid-state Metal-Oxide sensor No complex calculations or

adjustments Optional auto-calibrator

Ensures greatest accuracy without any customer intervention

Page 5: VaporLab™ 700 Gas Analyzer

© 2006 MSA

Product DescriptionProduct Description

Uses the power of Gas Chromatographic Analysis without the complex operational or support requirements associated with this technology

Low Cost of Operation & Maintenance

Page 6: VaporLab™ 700 Gas Analyzer

© 2006 MSA

VaporLab 700 Flow Diagram

VaporLab 700 Flow Diagram

CO

NC

EN

TRATO

R

VALVE #2

PRESSURE SENSOR

N.O

.

N.C.

CO

M

AMPLIFIERCARRIER

PRESSURE SIGNAL

VENTCARRIER

PUMPSCRUBBER

CARRIER INLET

ANALYTICAL COLUMN

VENT

OVEN HEATER

THERMISTOR

COLUMN TEMP.

SIG.

DETECTOR

GC DETECTOR

SIG.

#3N.O

.

N.C

.C

OM

N.O.

N.C.COM

VALVE #4N

.O.

N.C.

CO

M

SAMPLE PUMP

N.O

.

N.C.

CO

M

N.O

.

N.C.

CO

M

VALVE #5

VALVE #6

SAMPLE INLET

CAL GAS INLET

#1

FAN

MANIFOLD HEATER

VENT

CALIBRATION GAS DILUTION ACCESSORY

N.O.

N.C.COM

N.O.

N.C.COM

FLOW CONTROLLER

ZERO AIR

CAL GAS

MIXER

FLOW CONTROLLER

100% TO 0.1% DILUTED CALIBRATION GAS

(P> 1 ATM)

VENT

N.C.

N.C.

Page 7: VaporLab™ 700 Gas Analyzer

© 2006 MSA

Auto-Calibration/Dilutor Option

Auto-Calibration/Dilutor Option

Allows self-calibration in the field Built into the unit Reduces cost of operation

Less expensive gas standards at ppb levels

Allows for simpler field maintenanceRecalibration of the detector is easy!Multi-point calibration improves accuracy

and dynamic range

Page 8: VaporLab™ 700 Gas Analyzer

© 2006 MSA

How Does It WorkHow Does It Work

Sample is introduced into the unit

Pre-concentrator to provide greatest sensitivity

Pre-concentrator is heated to produce a plug of gas on analytical column

Gas of interest is separated from background gases on GC column

After a reproducible period of time, gas of interest is detected with solid state detector

Page 9: VaporLab™ 700 Gas Analyzer

© 2006 MSA

Optional DisplayOptional Display

Menu driven keypad

Provides easy access to menus including system status, calibration and alarms

LED’s provide quick indication of system status

Page 10: VaporLab™ 700 Gas Analyzer

© 2006 MSA

ApplicationsApplications

Industrial Monitoring Process Control Monitoring OHSA Occupational Exposure

Monitoring Perimeter Monitoring (EPA)

Page 11: VaporLab™ 700 Gas Analyzer

© 2006 MSA

BTEX DetectionBTEX Detection

BTEX = Benzene, Toluene, Ethyl Benzene and Xylenes

Oil and Gas Industry Process and storage applications

Food and Beverage Industry Approved for all ISBT Method 12 criteria

for selective measurement of BTEX in CO2

Page 12: VaporLab™ 700 Gas Analyzer

© 2006 MSA

BTEX SeparationBTEX SeparationBTEX @ 1000ppb

20000

30000

40000

50000

60000

70000

80000

90000

0 100 200 300 400

Time (Seconds)

Sig

nal

(H

z)

Benzene 33353 Hz

Toluene50726 Hz

Ethyl Benzene39361 Hz

o-Xylene26394 Hz

Retention TimesBenzene=92 secToluene=142Ethyl Benzene=206O-Xylene=316

Page 13: VaporLab™ 700 Gas Analyzer

© 2006 MSA

BTEX SeparationBTEX Separation

BTEX @ <10ppb

24000

24500

25000

25500

26000

26500

0 100 200 300 400

Time (Seconds)

Sig

na

l (H

z)Benzene831 Hz

Toluene1335 Hz

Ethyl Benzene941 Hz

o-Xylene608 Hz

Water

Injection Peak

Retention TimesBenzene=92 secToluene=142Ethyl Benzene=206O-Xylene=316

Page 14: VaporLab™ 700 Gas Analyzer

© 2006 MSA

VaporLab 700 Gas Analyzer

VaporLab 700 Gas Analyzer