VARIO abutment 17°for screw-retained solutions
2 © Thommen Medical
Notes
EXPLANATION OF SYMBOLS
Batch code
Use by date
Date of manufacture
Sterilized using irradiation
Upper limit of temperature
Do not re-use
Non-sterile
Caution
Catalogue number
Conformity symbol as specified by EU Directive MDD 93/42/EEC
Consult instructions for use
Do not resterilize
Do not use if package is damaged
Atmospheric pressure limitation
Manufacturer
COLOUR CODE Each implant platform dia-
meter has a colour code, which can be found
on all implant packs, on the impression items
and on most diameter-specific instruments.
Yellow = platform � 3.5 mm
Green = platform � 4.0 mm
Blue = platform � 4.5 mm
Grey = platform � 5.0 mm
Violet = platform � 6.0 mm
RESPONSIBILITY/LIABILITY As part of an overall scheme, Thommen implants may be
used only with the ori ginal components and instruments in accordance with the manu-
facturer’s instructions. The use of unapproved components from other systems can
cause malfunctioning of the implants and abutments and lead to implant failure. The use
of the product is the responsibility of the user and, as such, beyond the control of
Thommen Medical AG. We refuse to accept any re sponsibility or liability for any damage
due to incorrect utilization of the product.
Products labeled “Do not re-use” may not be refurbished and/or reused. The refurbish-
ment and/or reuse of these products can affect their function (fitting and/or cutting
properties) as well as their safe use (risk of infection, disease transmission, fading of the
laser or colour marks, corrosion). Detailed information about the possible conse quences,
which may result from negligence to follow this information, is avail able from your dealer.
Caution: Federal law (USA) restricts this device to sale by or on the order of a dentist of
physician.
GUARANTEE OF STERILITY OF STERILE PRODUCTS Products of the Thommen
Implant System supplied in sterile packaging must not be resterilized. If the sterile pack-
aging is damaged during transport or storage, the product must not be used. Products
that have been opened and have not been immediately used for the intended operation
must not be used thereafter. After resterilization, the safety, function and efficacy of the
product cannot be guaranteed by the man ufacturer.
STORAGE Plastic parts and biomaterials must be stored at room temperature and
protected from exposure to strong light or high heat. Temperature limitations as specified
on the packaging must be observed.
INSTRUCTIONS FOR USE The following information is not intended as com prehensive
for the Thommen Implant System. New customers are advised to undergo training by a
specialist experienced in the use of this system.
VALIDITY This product catalogue replaces all previous editions.
PRODUCT DOCUMENTATION You can find detailed information on the han dling of
Thommen implants in our brochures. Ask our national representatives for product
brochures and instructions for use.
AVAILABILITY NOTE Not all products shown in this brochure are available in all
countries. For further information please contact our subsidiary or distributor in your
country.
COPYRIGHT©/REGISTERED® SPI®, INICELL®, APLIQUIQ® and Remotis® are registered
trademarks of Thommen Medical AG. Publication or reproduction is permitted only with
the written consent of Thommen Medical AG. LOCATOR® is a registered trademark of
Zest Anchors Inc., CA, USA.
COLOURED WARNING STICKER
Application was changed – follow the instructions in the corresponding docu mentation.
New design – the application has not been changed.
© Thommen Medical 3
Contents
VARIO abutment 17° 4
Indications/Contraindication 5
Application 6
Product overview 9
Torque values 11
4 © Thommen Medical
New 17° angled VARIO abutments are ideally suited for occlusal screw-
retained single or multiple unit restorations where implants have been
placed with pronounced angulation to the planned prosthetics. The abut-
ments made from CP Titanium, are available with two different rotational
orientations of the cone with respect to the hexagon to allow optimal
alignment of the abutment. The prefabricated gold alloy screw seat ensures
a precise connection between the finished crown or bridge and abutment.
Advantages at a glance
· Axial angulation of up to 17° can be managed
· Two abutment versions for an exact alignment
· Prefabricated gold alloy screw seat for optimal precision
· Can be combined with existing VARIO abutments for
bridge
VARIO abutment 17°for screw-retained solutions
© Thommen Medical 5
Indications
Thommen SPI®VARIO angled abutments are intended to be used in con-
junction with SPI® System dental implants in the maxillary and/or mandibular
arch to provide support for crowns or bridges.
Contraindications, warnings, and precautions for use with the SPI® Dental
Implant System are included in the SPI® dental implant instructions for use.
Note
Angled abutments on small diameter implants are not recommended for
use in the molar region, unless they are splinted to larger diameter implants.
Abutment type A Abutment type B Abutment type A with assembled screw seat
Platform �
5.6 mm
Platform �
7.6 mm (minimum construct height)
Platform �
5.6 mm
6 © Thommen Medical
Hex headtype A
Hex headtype B
30°
7.6 mm
Implant shoulder
Screw seat occlusal
Maximumshort-
able
2.0 mm
Application
SELECTION abutments
VARIO abutments 17° SELECTION are available for planning purposes.
They include hexagon orientation type A (0°) or B (30°) for trial alignment.
The positioning handle may be inserted into the SELECTION abutment
occlusal screw channel to visualize the orientation of the cone and also
for placing the SELECTION abutments.
The SELECTION abutment shape and height of 7.6 mm above implant
shoulder corresponds to the minimum possible assembled height of a
crown or bridge construction.
Processing instructions for the fabrication of the suprastructure
The VARIO abutment 17° is ideally suited for small bridges having two implants
with minimal divergence and one or two intermediary units. The maximum
angulation between two implants which can be managed is as follows:
PF � 3.5 mm 6.0°
PF � 4.0 mm 7.0°
PF � 4.5 mm 8.5°
PF � 5.0 mm 11.0°
For bridge restorations with greater implant divergences, it is recommended
to use the VARIOmulti abutment system.
The prefabricated VARIO screw seat must be used for fabrication of the
superstructure. The plastic burn-out tube may be shortened to the upper
edge of the screw seat.
© Thommen Medical 7
Place the abutment on the analog and secure it using the appropriate
abutment screw.
Place the VARIO screw seat on the VARIO abutment 17° and secure it
with an occlusal screw.
Shorten the plastic cylinder to allow for maximum intercuspation. Avoid
occlusal contact of the plastic cylinder with the antagonist tooth. Maintain a
minimum construct height of 7.6 mm.
The VARIO abutment 17° must not be modified in any way as this may affect
the precision fit of the screw seat.
Gold 60 %
Platinum 24 %
Palladium 15 %
Iridium 1 %
Material specification
VARIO screw seat
Nonoxidizing precious metal cast-on alloy
Melting interval 1400–1460 °C
WAK 25–600 °C 12.8 μm/mk
8 © Thommen Medical
A
A
A
Before wax-up, the abutment screw channel must be blocked out with a
suitable material (A) that can be removed without leaving residues. In the
case of single crowns, block out the surface of the screw channel such that
an antirotation element is generated on the inside of the crown.
Wax up a corresponding crown or bridge framework onto the abutment in-
cluding the screw seat using standard procedures. Remove the occlusal
screw in order to remove the wax-up.
Invest and cast the superstructure following the procedures described in
detail in the brochure, VARIO/RETAIN Fo_22d010 or VARIOmulti Fo_22d117.
After casting, devest the framework and clean it carefully by using stand-
ard methods such as ultrasound, water steam, pickling and fi berglass
brush.
The screw (A) seat must not be sandblasted or reworked under any
circumstances as this may affect the precision fi t of the components.
Process and fi nish the construction in additional working steps according
to standard procedures.
We recommend fabrication of a transfer key.
New abutment and occlusal screws must be used for permanent inser-
tion of the restoration in the mouth.
Always follow the recommended torque values of 15 Ncm for abutment
screws PF � 3.5 mm, and 25 Ncm for abutment screws PF � 4.0–
6.0 mm. The occlusal screw requires a torque value of 15 Ncm.
Sterilization
VARIO angled abutments are provided non-sterile and should be sterilized
by standard gravity autoclave at 134 ºC (274 ºF) for 40 minutes with a
drying time of 20–60 minutes or pre-vacuum autoclave at 134 ºC (274 ºF)
for 18 minutes with a drying time of 20–60 minutes.
© Thommen Medical 9
Product overview
PF � 3.5 mm PF � 4.0 mm PF � 4.5 mm PF � 5.0 mm
VARIO abutment 17° set
Height (H) Shoulder height (h)
5.6 mm 0.5 mm Type A 1.03.160 1.03.161 1.03.162 1.03.163
Type B 1.03.180 1.03.181 1.03.182 1.03.183
Set content
· 1 VARIO abutment 17°
· 1 VARIO screw seat
· 1 Occlusal screw
· 1 Abutment screw
VARIO abutments 17° have a type A (0°) or B (30°) hex.
This enables an abutment to be aligned in 30° steps.
VARIO screw seat, NOA/burn-out plastic
� Height
2.9 mm 2.0/11.3 mm 2.03.610
Occlusal screw, 4-lobe, Titanium alloy
Length
3.5 mm 4.03.507
Abutment screw, 4-lobe, Titanium alloy
Length
5.8 mm 4.03.500
7.2 mm 4.03.501
VARIO SELECTION abutment 17°, Aluminum-anodized/plastic (PPSU)
Height (H) Shoulder height (h)
7.6 mm 0.5 mm Type A S.4.03.160 S.4.03.161 S.4.03.162 S.4.03.163
Type B S.4.03.180 S.4.03.181 S.4.03.182 S.4.03.183
Positioning handle for angled VARIO and VARIOmulti abutments, Stainless steel
� Length
4.0 mm 25.0 mm 3.03.521
10 © Thommen Medical
© Thommen Medical 11
Torque values
* Abutment screw for customizable gingiva former immediately after implantation � 10 Ncm.
This overview can be ordered in the form of a laminated single sheet under article no. Fo_22d123
(available in English only). See website for current version www.thommenmedical.com
PF
� 3
.5 m
m
PF
� 4
.0 m
m
PF
� 4
.5 m
m
PF
� 5
.0 m
m
PF
� 6
.0 m
m
• • • • •
• • • •
•
•
• • •
• • • • •
• • • •
• • • •
• • •
+
Healing cap/Gingiva former *
VARIO Ttransversal screw
VARIOmulti protective cap
Abutment screw for PF � 3.5 mm *(for all restoration materials)
VARIOmulti abutment
DIRECT closure screw
Occlusal screw for bar
Occlusal screw
Abutment screw in combination with ceramic restorations(except PF � 3.5 mm)
Abutment screw *(except PF � 3.5 mm)
VARIOmulti abutment (except PF � 3.5 mm)
Retentive anchor
ZEST LOCATOR® abutment
20 Ncm
10 Ncm
15 Ncm
25 Ncm
30 Ncm • • • • • •
• • • •
20 Ncm30 Ncm
ZEST LOCATOR® PF 3.5 mmZEST LOCATOR® PF 4.0–6.0 mmN
EW
SWISS PRECISION AND INNOVATION. www.thommenmedical.com
Fo
_0
2d
169
.02
12
/11
ca
na
rin
i c
om
mu
nic
ati
on
s