vdi extensionmodules...

20
VDI Extension Modules Installation and Quick Start Guide Virginia Diodes, Inc. Agilent PNA/PNAX Family Network Analyzers PNA Series N522XA PNAX Series N524XA Document Number: VDI707.5 Print Date: May 2013

Upload: buihanh

Post on 10-Jun-2018

246 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: VDI ExtensionModules InstallationandQuickStartGuidena.support.keysight.com/mmwave/exampleVDIinstall.pdf · VDI!ExtensionModules! InstallationandQuickStartGuide! Virginia’Diodes,’Inc.’!

     

 VDI  Extension  Modules  

Installation  and  Quick  Start  Guide  Virginia  Diodes,  Inc.  

 Agilent  PNA/PNA-­‐X  Family  Network  Analyzers  

PNA  Series  N522XA  PNA-­‐X  Series  N524XA  

                         

                 

     

 Document  Number:  VDI707.5  

Print  Date:  May  2013    

Page 2: VDI ExtensionModules InstallationandQuickStartGuidena.support.keysight.com/mmwave/exampleVDIinstall.pdf · VDI!ExtensionModules! InstallationandQuickStartGuide! Virginia’Diodes,’Inc.’!

 

VDI707.5  –  May  2013     2  

Safety  Notes  The  following  terminology  is  used  throughout  this  manual.  Familiarize  yourself  with  each  safety  note  and  its  meaning  before  installing  and  operating  the  VNA  Extension  Modules.  

Required   Required  denotes  necessary  setup  instruction.  It  calls  attention  to  procedure  related  to  intrinsic  properties  of  the  system,  which  if  not  adhered  to  will  result  in  degraded  performance  due  to  incorrect  setup.  

 

CAUTION   Caution  denotes  a  hazard.  It  calls  attention  to  a  procedure  that  if  not  adhered  to  will  result  in  damage  to  the  system.  Do  not  proceed  beyond  a  caution  sign  until  the  indicated  conditions  are  fully  understood  and  met.  

                   

Page 3: VDI ExtensionModules InstallationandQuickStartGuidena.support.keysight.com/mmwave/exampleVDIinstall.pdf · VDI!ExtensionModules! InstallationandQuickStartGuide! Virginia’Diodes,’Inc.’!

 

3     May  2013  –  VDI707.5  

Table  of  Contents  

SAFETY  NOTES  ........................................................................................................................................  2  

1  RECOMMENDED  OPERATING  PROCEDURES  AND  PRECAUTIONS  ......................................  4  1.1  RF  DRIVE  LIMITATIONS  ......................................................................................................................  4  1.2  TEST  PORT  CARE  .................................................................................................................................  4  1.3  RF  CABLE  CARE  ...................................................................................................................................  4  1.4  ESD  PRECAUTION  ................................................................................................................................  4  1.5  GENERAL  RECOMMENDATIONS  ...........................................................................................................  5  

2  FIRST  TIME  PNA/PNA-­‐X  SOFTWARE  SETUP  .............................................................................  6  2.0.1  UPDATE  PNA/PNA-­‐X  FIRMWARE  ..................................................................................................................  6  2.0.2  LOAD  THE  MILLIMETER  WAVE  MACRO  ONTO  THE  PNA/PNA-­‐X  ............................................................  6  2.0.3  SET  PNA/PNA-­‐X  PREFERENCES  ....................................................................................................................  8  2.0.4  LOAD  CALIBRATION  KIT  FILES  .........................................................................................................................  8  

3  OPERATION  ..........................................................................................................................................  9  3.1  HARDWARE  SETUP  &  CONNECTIONS  ..................................................................................................  9  3.1.1  MECHANICAL  SETUP  ...........................................................................................................................................  9  3.1.2  RF/LO/IF  CABLE  CONNECTIONS  ..................................................................................................................  10  3.1.3  POWER  CONNECTIONS  .....................................................................................................................................  11  3.2  SOFTWARE  SETUP  ..............................................................................................................................  12  3.2.1  CONFIGURE  FRONT  PANEL  OPERATION  .......................................................................................................  12  3.2.2  LOAD  STATE  FILE  ..............................................................................................................................................  12  3.2.3  FREQUENCY  OFFSET  MODE  INITIALIZATION  ...............................................................................................  12  3.2.4  SAVE  STATE  FILE  ..............................................................................................................................................  13  3.3  OPERATOR’S  CHECK  ..........................................................................................................................  13  3.3.1  CHECK  WAVE  QUANTITIES  ..............................................................................................................................  13  3.3.2  CHECK  DR  ..........................................................................................................................................................  13  3.4  CALIBRATION  .....................................................................................................................................  14  3.4.1  TRL  CALIBRATION  PROCEDURE  ....................................................................................................................  14  3.4.2  SOLT  PROCEDURE  ............................................................................................................................................  17  3.4.3  SAVING  CAL  STATE  ...........................................................................................................................................  19  

4  CONTACT  .............................................................................................................................................  20      

   

Page 4: VDI ExtensionModules InstallationandQuickStartGuidena.support.keysight.com/mmwave/exampleVDIinstall.pdf · VDI!ExtensionModules! InstallationandQuickStartGuide! Virginia’Diodes,’Inc.’!

 

VDI707.5  –  May  2013     4  

1  Recommended  Operating  Procedures  and  Precautions  

1.1  RF  Drive  Limitations  • Power  inputs  to  the  [RF  Input]  and  [LO  Input]  ports  of  the  VNA  Extension  Modules  are  noted  

in  the  User’s  Guide  for  each  system.  Remaining  within  this  input  power  range  will  provide  optimal  performance  and  stability.  

CAUTION   Do  not  exceed  input  power  range  to  [RF  Input]  and  [LO  Input]  ports  by  more  than  3dB.  Power  inputs  3dB  over  the  high  end  of  the  recommended  range  can  cause  irreversible  damage  to  the  Extension  Modules.  

 

CAUTION   Input  saturation  and  damage  limits  exist  for  RF  power  into  the  test  ports.  For  specific  limits  by  frequency  band  please  refer  to  our  website.  VNAX  Specifications  

1.2  Test  Port  Care  • Do  not  remove  the  test  ports  from  the  systems.  The  test  ports  must  be  connected  with  care  

for  the  best  possible  RF  calibration  results.  Additionally  the  test  ports  protect  the  system  from  dust  and  debris.  

• Replace  the  provided  dust  caps  when  the  system  is  not  in  use,  this  will  prevent  dust  and  debris  from  entering  the  waveguide.  

• Torque  waveguide  connection  screws  with  between  20-­‐50cNm  when  connecting  a  component  to  the  test  port  of  the  Extension  Module.  

 

CAUTION   Using  torques  higher  than  50cNm  when  connecting  the  flanges  can  permanently  damage  the  interface.  

• Inspect  waveguide  flanges  prior  to  making  connections.  If  necessary,  clean  the  flange  with  pre-­‐dampened  TexWipe  wipes  or  swabs  (e.g.  Part  Number  TX1065).  If  these  are  not  available,  TexWipe  cloths  lightly  dampened  with  ethanol  will  also  work  (e.g.  Part  Number  TX604).  

CAUTION   Making  a  connection  with  metal  debris  between  the  two  waveguide  flanges  can  permanently  damage  the  waveguide  interface  and  prevent  repeatable  connections.  

1.3  RF  Cable  Care  • Use  a  torque  of  90cNm  when  making  coaxial  connections.  • Replace  provided  dust  caps  when  system  is  not  in  use.  • Avoid  sharp  bends  in  cables.  

1.4  ESD  Precaution  • The  internal  components  of  the  Extension  Modules  can  be  permanently  damaged  by  

ElectroStatic  Discharge  (ESD).  It  is  recommended  that  engineers  and  technicians  wear  a  special  grounded  wrist  strap  when  handling  this  component.  In  addition,  the  work  environment  around  the  component  should  be  well  grounded.  

   

Page 5: VDI ExtensionModules InstallationandQuickStartGuidena.support.keysight.com/mmwave/exampleVDIinstall.pdf · VDI!ExtensionModules! InstallationandQuickStartGuide! Virginia’Diodes,’Inc.’!

 

5     May  2013  –  VDI707.5  

1.5  General  Recommendations  • If  the  phase  of  the  system  is  not  stable,  check  the  cable  connections  as  described  in  Section  

3.1.2  to  verify  all  cables  are  properly  connected  and  torqued.  • For  best  calibration  results  minimize  the  bending  of  the  LO  Cables.  • After  completing  measurements  navigate  to  [Response]  >  [Power]  >  [on/off]  to  toggle  RF  

Power  off.  Then  turn  off  both  VDI-­‐175  power  supplies,  and  replace  dust  covers  on  the  Extension  Modules.  

• Dust  covers  included  on  the  Extension  Modules  are  only  intended  to  protect  against  dust.  

   

Page 6: VDI ExtensionModules InstallationandQuickStartGuidena.support.keysight.com/mmwave/exampleVDIinstall.pdf · VDI!ExtensionModules! InstallationandQuickStartGuide! Virginia’Diodes,’Inc.’!

 

VDI707.5  –  May  2013     6  

2  First  Time  PNA/PNA-­‐X  Software  Setup  The  following  procedure  will  set  up  the  Agilent  PNA/PNA-­‐X  family  Network  Analyzers  to  work  with  the  VDI  Extension  Modules.  The  procedure  only  needs  to  be  followed  once  on  a  particular  PNA/PNA-­‐X.  If  already  completed  on  this  PNA/PNA-­‐X  then  skip  to  section  3,  Operation.  

2.0.1  Update  PNA/PNA-­‐X  Firmware  VDI  recommends  installing  the  newest  PNA/PNA-­‐X  firmware.  

• Exit  the  PNA/PNA-­‐X  Firmware  [File]  >  [Exit].  • Obtain  the  most  recent  version  of  the  firmware  from  Agilent  at  

http://na.tm.agilent.com/pna/firmware/firmware.html.  Download  the  “Latest  Production  Release”  for  the  appropriate  PNA/PNA-­‐X  model.  

• Once  the  firmware  finishes  downloading,  right  click  on  the  firmware  file  and  choose  install.  • Restart  the  PNA/PNA-­‐X  and  the  new  firmware  will  start  automatically.  

2.0.2  Load  the  Millimeter  Wave  Macro  onto  the  PNA/PNA-­‐X  The  mmWave  Macro  simplifies  the  setup  of  the  Extension  Modules.  The  macro  will  be  used  to  set  the  power  levels  and  multiplication  factors  for  the  inputs  to  the  VDI  Extension  Modules.  It  is  possible  to  do  this  setup  manually  through  the  [Stimulus]  >  [Frequency]  >  [Frequency  Offset]  window,  but  it  is  not  recommended.  

• Minimize  the  PNA/PNA-­‐X  Firmware  [File]  >  [Minimize  Application].  • Obtain  the  most  recent  version  of  the  macro  “Direct  Connect  Millimeter  Wave  Macro”  from  

Agilent  at  http://na.tm.agilent.com/pna/apps/applications.htm  • Right  click  on  the  downloaded  “mmwave_setup.msi”  file  and  choose  install.  • Follow  the  on-­‐screen  instructions.  • Maximize  the  PNA/PNA-­‐X  Firmware.  • Navigate  to  [Utility]  >  [Macro]  >  [Macro  Setup].  • Select  an  empty  line  and  click  [Edit].  

     

Page 7: VDI ExtensionModules InstallationandQuickStartGuidena.support.keysight.com/mmwave/exampleVDIinstall.pdf · VDI!ExtensionModules! InstallationandQuickStartGuide! Virginia’Diodes,’Inc.’!

 

7     May  2013  –  VDI707.5  

• In  the  “Edit  Macro  Setup”  window  enter  “mmWave”  as  the  “Macro  Title”  and  click  [Browse…].  

   • Navigate  to  C:\Program  Files\Agilent\Network  Analyzer\Applications\mmWave\  

   • Select  mmWave.exe  and  click  [Open].  

   • Click  [OK]  twice  to  return  to  the  instrument  home  screen  

Page 8: VDI ExtensionModules InstallationandQuickStartGuidena.support.keysight.com/mmwave/exampleVDIinstall.pdf · VDI!ExtensionModules! InstallationandQuickStartGuide! Virginia’Diodes,’Inc.’!

 

VDI707.5  –  May  2013     8  

2.0.3  Set  PNA/PNA-­‐X  Preferences  • Navigate  to  [Utility]  >  [System]  >  [Configure]  >  [Preferences].  • Check  the  box  next  to  Cal:  For  Frequency  Offset,  use  Primary  Frequencies  

   • Click  [OK].  

2.0.4  Load  Calibration  Kit  Files  • Minimize  the  PNA/PNA-­‐X  Firmware  [File]  >  [Minimize  Application].  • Obtain  the  most  recent  version  of  the  calibration  kit  file  from  VDI  at  

http://vdi.bz/downloads  • Unzip  the  calibration  kit  files.  • Maximize  the  PNA/PNA-­‐X  Firmware.  • Navigate  to  [Response]  >  [Cal]  >  [Cal  Kit…].  • In  the  Edit  PNA  Cal  Kits  window  that  opens,  click  [Import  Kit].  

   • Navigate  to  the  unzipped  calibration  kit  files  and  select  the  relevant  file(s).  • Click  [OK].  

   

Page 9: VDI ExtensionModules InstallationandQuickStartGuidena.support.keysight.com/mmwave/exampleVDIinstall.pdf · VDI!ExtensionModules! InstallationandQuickStartGuide! Virginia’Diodes,’Inc.’!

 

9     May  2013  –  VDI707.5  

3  Operation  The  following  procedure  first  describes  mechanical  and  software  setup  of  the  VDI  Extension  Modules  with  the  Agilent  PNA/PNA-­‐X,  then  gives  instruction  on  how  to  ensure  the  system  is  working  as  intended.  This  section  assumes  that  this  particular  PNA/PNA-­‐X  has  been  initialized  by  following  the  instructions  in  section  2,  First  Time  PNA/PNA-­‐X  Software  Setup.  

3.1  Hardware  Setup  &  Connections  

3.1.1  Mechanical  Setup    

 Typical  VDI  bench  setup  showing  Extension  Modules  connected  for  front  panel  operation.  

 • Place  the  Extension  Modules  on  a  flat  surface  and  use  the  adjustable  feet  to  align  and  level  

the  test  ports.  When  leveled  properly  the  waveguide  flanges  will  slide  together  with  minimal  resistance.  

• For  best  system  stability  install  in  a  location  with  constant  temperature  and  minimal  airflow.  • The  adjustable  feet  can  be  removed  and  the  six  1/4-­‐20  threaded  holes  can  be  used  to  mount  

the  Extension  Modules.  Do  not  allow  screws  to  protrude  further  than  0.25”  into  the  Extension  Modules.  

CAUTION   Do  not  block  or  impede  airflow  into  the  air  vents  on  the  bottom  or  rear  of  the  Extension  Modules.  

• The  bolts  in  the  lid  can  be  removed  and  the  four  1/4-­‐20  threaded  holes  can  be  used  to  mount  the  Extension  Modules.  Do  not  allow  screws  to  protrude  further  than  0.25”  into  the  Extension  Modules.  

 

Page 10: VDI ExtensionModules InstallationandQuickStartGuidena.support.keysight.com/mmwave/exampleVDIinstall.pdf · VDI!ExtensionModules! InstallationandQuickStartGuide! Virginia’Diodes,’Inc.’!

 

VDI707.5  –  May  2013     10  

3.1.2  RF/LO/IF  Cable  Connections  • [Preset]  the  PNA/PNA-­‐X  • Remove  the  jumpers  attached  to  A,  R1,  B,  R2  receivers  on  the  front  panel  of  the  PNA/PNA-­‐X.  • The  [SOURCE  OUT  –  CPLR  THRU]  jumpers  must  remain  connected  for  front  panel  operation.  • Connect  cables  marked  “IF”  from  [Ref.  IF]  and  [Meas.  IF]  on  the  back  of  one  Extension  

Module  to  [RCVR  R1  IN]  and  [RCVR  A  IN]  respectively  on  the  front  panel  of  the  PNA/PNA-­‐X.  • Repeat  this  procedure  for  the  other  Extension  Module,  connecting  [Ref.  IF]  and  [Meas.  IF]  to  

[RCVR  R2  IN]  and  [RCVR  B  IN]  respectively.  • Connect  cables  marked  “RF/LO”  from  [RF  Input]  and  [LO  Input]  on  the  back  of  one  Extension  

Module  to  [Port  1]  and  [Port  3]  respectively  on  the  front  panel  of  the  PNA/PNA-­‐X.  • Repeat  this  procedure  for  the  other  Extension  Module,  connecting  [RF  Input]  and  [LO  Input]  

to  [Port  2]  and  [Port  4]  respectively.    

 Front  panel  of  PNA/PNA-­‐X  showing  typical  cable  connections.  

 Rear  panel  of  Extension  Module  showing  typical  cable  connections.  

Page 11: VDI ExtensionModules InstallationandQuickStartGuidena.support.keysight.com/mmwave/exampleVDIinstall.pdf · VDI!ExtensionModules! InstallationandQuickStartGuide! Virginia’Diodes,’Inc.’!

 

11     May  2013  –  VDI707.5  

3.1.3  Power  Connections  • Toggle  the  switch  on  the  front  of  the  VDI-­‐175  Power  Supplies  to  the  off  (down)  position.  • Plug  both  VDI-­‐175  into  the  wall  using  the  included  AC  power  cables.  • Connect  each  VDI-­‐175  to  one  Extension  Module  using  the  included  DC  power  cables.  

   

Page 12: VDI ExtensionModules InstallationandQuickStartGuidena.support.keysight.com/mmwave/exampleVDIinstall.pdf · VDI!ExtensionModules! InstallationandQuickStartGuide! Virginia’Diodes,’Inc.’!

 

VDI707.5  –  May  2013     12  

3.2  Software  Setup  

3.2.1  Configure  Front  Panel  Operation  • Navigate  to  [System]  >  [Configure]  >  [Millimeter  Module  Config…]  • Make  sure  “Standard  PNA”  is  selected  and  click  “OK”.  

3.2.2  Load  State  File  • If  no  state  file  has  previously  been  stored  for  these  Extension  Modules,  skip  to  section  3.2.3  

Frequency  Offset  Mode  Initialization.  • Otherwise,  navigate  to  [File]  >  [Recall…]  and  choose  the  previously  saved  state  file  • Skip  to  section  3.3  Operator’s  Check.  

3.2.3  Frequency  Offset  Mode  Initialization  • Navigate  to  [Utility]  >  [Macro]  >  [mmWave].  • Enter  Start  Frequency  and  Stop  Frequency.  These  correspond  to  Output  Frequency  Start  

and  Output  Frequency  Stop  in  the  VDI  User’s  Guide  for  the  Extension  Modules.  • Enter  Multiplier  RF  IN  and  Multiplier  LO  IN.  These  values  are  fixed  by  VDI  hardware  and  

correspond  to  Multiplier  RF  and  Multiplier  LO  in  the  VDI  User’s  Guide  for  the  Extension  Modules.  

• Enter  279  MHz  for  the  IF  Frequency.  • Check  the  box  labeled  mmWave  LO<mmWave  RF.  • Uncheck  the  box  labeled  Port  Powers  Coupled.  • Enter  Port1  Power  and  Port2  Power.  These  correspond  to  VNA  Output  Power  RF  in  the  

VDI  User’s  Guide  for  the  Extension  Modules.  • Enter  LO1  Power  and  LO2  Power.  These  correspond  to  VNA  Output  Power  LO  in  the  VDI  

User’s  Guide  for  the  Extension  Modules.  • Click  [Calculate]  then  [Apply]  then  [OK].  

 The  mmWave  Setup  window  showing  typical  values  for  WR1.5  Extension  Modules.  See  VDI  User’s  Guide  for  values  specific  to  the  Extension  Modules.  

Page 13: VDI ExtensionModules InstallationandQuickStartGuidena.support.keysight.com/mmwave/exampleVDIinstall.pdf · VDI!ExtensionModules! InstallationandQuickStartGuide! Virginia’Diodes,’Inc.’!

 

13     May  2013  –  VDI707.5  

3.2.4  Save  State  File  • Toggle  RF  Power  Off.  • Navigate  to  [File]  >  [Save  As].  Choose  a  save  location  and  filename  and  save  as  type  .sta.  • Loading  this  file  at  a  later  date  will  recall  the  Frequency  Offset  Settings  as  well  as  IF  

Bandwidth,  Number  of  Points,  Displayed  Traces,  etc.    

3.3  Operator’s  Check  The  operator’s  check  is  a  subset  of  tests  that  give  a  high  degree  of  confidence  that  the  Extension  Modules  and  PNA/PNA-­‐X  are  performing  properly.  This  test  should  be  performed  any  time  you  wish  to  have  confidence  that  the  system  is  working  well.  The  operator’s  check  does  not  verify  performance  to  all  specifications.  

3.3.1  Check  Wave  Quantities  • Navigate  to  [Stimulus]  >  [Sweep]  >  [Number  of  Points].  Set  this  to  1001.  • Navigate  to  [Response]  >  [Avg]  >  [IF  Bandwidth].  Set  this  to  1kHz.  • Add  traces  [A,1]  [B,2]  [R1,1]  and  [R2,2].  • Terminate  the  waveguide  test  ports  of  the  two  Extension  Modules  with  Short  Circuits.  • Turn  on  the  VDI-­‐175  Power  Supplies.  • Turn  on  PNA/PNA-­‐X  RF  Power.  • Compare  the  traces  on  screen  to  Receiver  Power  vs  Frequency  Graph  in  Section  3  of  the  VDI  

User’s  Guide.  The  shapes  of  the  curves  should  match  if  the  system  is  working  properly.  

3.3.2  Check  DR  • Navigate  to  [Stimulus]  >  [Sweep]  >  [Number  of  Points].  Set  this  to  201.  • Navigate  to  [Response]  >  [Avg]  >  [IF  Bandwidth].  Set  this  to  10Hz.  • Add  traces  [S12]  and  [S21].  • Connect  the  waveguide  test  ports  of  the  two  Extension  Modules  directly  together.  • Turn  on  the  VDI-­‐175  Power  Supplies.  • Turn  on  PNA/PNA-­‐X  RF  Power.  • After  one  sweep  completes,  [Marker/Analysis]  >  [Memory]  >  [Normalize]  on  both  traces.  

The  resulting  graph  on  screen  should  be  flat  at  0dB.  • Disconnect  the  waveguide  test  ports  of  the  Extension  Modules  and  terminate  them  with  

Short  Circuits.  • Compare  the  traces  on  screen  to  Dynamic  Range  vs  Frequency  Graph  in  Section  3  of  the  VDI  

User’s  Guide.  Note  that  the  User’s  Guide  graph  is  a  ten  sweep  RMS  average  so  the  data  shown  on  screen  will  have  larger  amplitude  variation.  

• The  average  values  of  the  User’s  Guide  graph  and  the  on  screen  data  will  match  if  the  system  is  working  properly.  

Page 14: VDI ExtensionModules InstallationandQuickStartGuidena.support.keysight.com/mmwave/exampleVDIinstall.pdf · VDI!ExtensionModules! InstallationandQuickStartGuide! Virginia’Diodes,’Inc.’!

 

VDI707.5  –  May  2013     14  

Typical  Single  Sweep  Dynamic  Range  data  for  a  WR1.5  system  (dB).  Measurement  was  made  using  procedure  described  above.  

3.4  Calibration  VDI  Calibration  Kits  from  WR15  through  WR3.4  are  designed  for  the  TRL  calibration  procedure  but  can  also  be  used  to  perform  SOLT  calibration.  VDI  Calibration  Kits  from  WR2.8  through  WR1.0  are  designed  for  the  SOLT  calibration  procedure.  The  VDI  Calibration  Kits  can  also  be  used  to  perform  a  variety  of  other  calibration  procedures,  however  these  are  not  covered  in  this  document.  

3.4.1  TRL  Calibration  Procedure  • VDI  typically  performs  calibrated  measurements  using  1kHz  RBW  and  1001  points.  These  

values  should  be  optimized  based  on  the  goals  of  the  measurement.  • Navigate  to  [Response]  >  [Cal  Wizard]  • Select  the  “SmartCal”  bullet  and  click  [Next  >]  

   

Page 15: VDI ExtensionModules InstallationandQuickStartGuidena.support.keysight.com/mmwave/exampleVDIinstall.pdf · VDI!ExtensionModules! InstallationandQuickStartGuide! Virginia’Diodes,’Inc.’!

 

15     May  2013  –  VDI707.5  

• Choose  “2  Port  Cal”  with  Ports  1  and  2  and  click  [Next  >]  

   • Select  the  appropriate  waveguide  band  and  check  “Modify  Cal:  Change  Cal  Method,  

standards”  then  click  [Next  >]  

   • Click  [Cal  Type/Stds…]  

   

Page 16: VDI ExtensionModules InstallationandQuickStartGuidena.support.keysight.com/mmwave/exampleVDIinstall.pdf · VDI!ExtensionModules! InstallationandQuickStartGuide! Virginia’Diodes,’Inc.’!

 

VDI707.5  –  May  2013     16  

• Choose  “TRL”  from  the  “Calibration  type”  dropdown  menu  and  click  [OK]  

   • Click  [Next  >]  • Follow  the  on  screen  instructions  and  measure  the  calibration  standards  when  prompted  • When  the  waveguide  calibration  procedure  is  complete,  verify  the  calibration  by  connecting  

a  straight  waveguide  between  the  two  modules  and  measuring  through  parameters.    

 S-­‐Parameters  for  a  1”  Straight  Waveguide  measured  at  WR3.4  (220-­‐330GHz)  after  a  typical  calibration.  

   

Page 17: VDI ExtensionModules InstallationandQuickStartGuidena.support.keysight.com/mmwave/exampleVDIinstall.pdf · VDI!ExtensionModules! InstallationandQuickStartGuide! Virginia’Diodes,’Inc.’!

 

17     May  2013  –  VDI707.5  

3.4.2  SOLT  Procedure  • VDI  typically  performs  calibrated  measurements  using  1kHz  RBW  and  1001  points.  These  

values  should  be  optimized  based  on  the  goals  of  the  measurement.  • Navigate  to  [Response]  >  [Cal  Wizard]  • Select  the  “SmartCal”  bullet  and  click  [Next  >]  

   

• Choose  “2  Port  Cal”  with  Ports  1  and  2  and  click  [Next  >]  

   • Select  the  appropriate  waveguide  band  and  check  “Modify  Cal:  Change  Cal  Method,  

standards”  then  click  [Next  >]  

     

Page 18: VDI ExtensionModules InstallationandQuickStartGuidena.support.keysight.com/mmwave/exampleVDIinstall.pdf · VDI!ExtensionModules! InstallationandQuickStartGuide! Virginia’Diodes,’Inc.’!

 

VDI707.5  –  May  2013     18  

• Click  [Cal  Type/Stds…]  

   • Choose  “SOLT”  from  the  “Calibration  type”  dropdown  menu  and  click  [OK]  

   • Click  [Next  >]  • Follow  the  on  screen  instructions  and  measure  the  calibration  standards  when  prompted  • When  the  calibration  procedure  is  complete,  verify  the  calibration  by  connecting  a  straight  

waveguide  terminated  with  a  short  circuit  to  each  modules  and  measuring  reflect  parameters.  

Page 19: VDI ExtensionModules InstallationandQuickStartGuidena.support.keysight.com/mmwave/exampleVDIinstall.pdf · VDI!ExtensionModules! InstallationandQuickStartGuide! Virginia’Diodes,’Inc.’!

 

19     May  2013  –  VDI707.5  

 S22  for  a  1”  Straight  Waveguide  terminated  with  a  short  circuit,  measured  at  WR1.5  (500-­‐750GHz)  after  a  typical  calibration.  

3.4.3  Saving  Cal  State  After  calibrating  it  is  possible  to  save  the  calibration  parameters.  This  file  can  later  be  loaded  to  provide  a  calibrated  state  and  allow  the  user  to  quickly  make  measurements.  Since  the  calibration  is  affected  by  cable  motion,  lab  temperature,  and  system  warm-­‐up,  results  will  not  be  as  precise  as  a  fresh  calibration.  • Navigate  to  [File]  >  [Save  As…]  • Choose  a  location  and  filename  and  save  as  type  “.csa”  

   

Page 20: VDI ExtensionModules InstallationandQuickStartGuidena.support.keysight.com/mmwave/exampleVDIinstall.pdf · VDI!ExtensionModules! InstallationandQuickStartGuide! Virginia’Diodes,’Inc.’!

 

VDI707.5  –  May  2013     20  

4  Contact    Website:  www.vadiodes.com  Technical  Questions:  [email protected]