Version: November 16, 2009

EPMA of Particles

Electron probe microanalysis

EPMA

Page 2: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

What’s the point?EPMA is traditionally done for bulk material. What are the

issues for particles?

How precise/accurate are such analyses?

Page 3: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

Bulk vs particle

Normal EPMA assumes that the electron beam is exciting a homogeneous volume, i.e. there is no difference either laterally or vertically. Thus, the matrix correction is being applied in a uniform matter, and there is one applicable (z) profile for each element.However, particles createmany difficulties, including:• Their size (small)• Their venue (location, surroundings)• Their surface (not flat nor smooth)

Page 4: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

The Good, the Bad, the UglyMost operators of electron

microprobes, using WDS, understand the difficulties of trying to do EPMA on non-flat surfaces (particles).

However, the real problem is that there are probably 100 EDS systems on SEMs for every 1 WDS on an electron microprobe. And with SEMs and EDS systems so easy to operate, a large number of users make major errors without knowing it, assuming whatever the software spits out as a composition is really what it is.

After these slides, you should know the difference between the GOOD, the BAD, and the UGLY… particle analyses.

Page 5: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

Size: “Mass Effect”

Goldstein et al, 1992, p. 479, 481

Consider a 2 um diameter sphere (say of NIST glass K412).

Mass effect/error: electrons escape from sides of small particles if the E0 is great enough so that electrons scatter out of the body before using up all their energy.

Page 6: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

Size: Absorption Effect

Absorption effect of non-flat upper surface: there is a different path length from the normal flat geometry. In this case, the emergent x-rays will travel ~50% shorter distance in the material, and thus have ~50% of the absorption, and all else being equal, will yield a K-ratio twice as large as it would be if it were larger and had a flat polished surface.

Goldstein et al, 1992, p. 479, 481

Page 7: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

Size - Detector-sample Geometry

EDS (shown to right): Variable effect of geometry of trajectory between beam impact area on non-uniform surface and the location of the detector. K-ratios from point 1 could be significantly different than those from point 3, despite the sphere being of uniform composition.

WDS: Could be as bad, with up to 5 spectrometers positioned at 5 different orientations

Goldstein et al, 1992, p. 479, 481

Page 8: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

EDS vs WDS

For all the problems cited, EDS has an advantage over WDS for “particle analysis” in that the spectrometer is in a constant position relative to the sample and beam “landing spot”, whereas WDS spectrometers would be at different positions, creating various different absorption path lengths to each spectromter.

WDS 1 WDS 2

Page 9: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

How Do People Try to Do EPMA of

Particles?1. Take whatever values come off the EDS

with no correction other than built in ‘normalization to 100 wt%’;

2. Overscan the sample;

3. Put the beam on a top “flat” surface and hope for the best;

4. Do #3 but apply Armstrong and Buseck’s geometric correction factors;

5. Peak-to-background method;

6. Do Monte Carlo simulations with DTSA-2 or PENEPMA and compare with actual K-ratios (time consuming);

7. If large enough, mount, polish flat and do traditional WDS

Page 10: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

Two approaches

• Traditional approach: (a) “overscan” and (b) normalize numbers – but this is not very good (above left table)• Armstrong and Buseck (1975) developed a procedure based upon a regular geometric shape factor, where the different path length and other effects could be used. Method is based on bracketing particle and beam overscanning during collection of spectrum by EDS and modeling of electron path and x-ray propagation out through several shapes – sphere, hemisphere, squared pyramid, and rectangular, tetragonal, cylindrical and right triangular prisms. Correction factors are given in terms of predicted k-factors for pairs of particular elements, vs particle thickness along e beam. This is not easy, takes much trial and error, but apparently can yield fairly good results (see table above).

Goldstein et al, 1992, p. 488, 489

Page 11: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

Armstrong-Buseck correction

Goldstein et al, 1992, p. 490

Here is another useful way to view a lot of data: error histograms. The top one shows the “conventional” particle analysis procedure to be very inaccurate, whereas using their geometric-correction approach, the errors shrink greatly.

Though there still are problems, which the analyst must recognize and report with any writeup.

Page 12: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

Other approach: Peak/Background

It is known that whereas geometric effects can create large differences in measured x-ray intensities, the ratio of a characteristic x-ray to the nearby background (=continuum of same energy range) intensity is less sensitive to geometric effects.

Thus, an ‘automatic internal normalization factor’ can be created for both a standard and an unknown. This has been shown to work well. The issue then is to find software that has this implemented.

This approach is discussed in Goldstein et al, 2nd edition, pp. 490-493.

(It is also used in biological EPMA, where the beam can destroy the cellular material, so mass “disappears”.)

Page 13: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

Other approaches: Peak/Background

Results from using P/B approach for 2 minerals. Excellent results for the pyrite, better but not within the 1% error we would like to see.

Goldstein et al, 1992, p. 491

Page 14: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

Other approaches: Monte Carlo Simulations

Another approach is to create a similar geometric shape, give it a composition which is what you believe the unknown to be (based upon initial EPMA results).

In the model, you place the beam on a particular spot corresponding to where the beam hits it, and then “run the simulation” to create x-ray counts (and similarly for a standard), and then compare the resulting simulated K-ratio with the experimental K-ratio.

If the K-ratios match, then the composition is correct; if not, modify the composition and run another simulation. Iterate until the simulated and the experimental K-ratios match.

Page 15: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

Other approaches: Monte Carlo Simulations

Nicholas Ritchie at NIST has upgraded the DTSA-II EDS program to include Monte Carlo simulation of EDS in various simple shapes

http://nist.gov/dtsa

Page 16: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

DTSA-II You then compare the EDS spectrum of the simulated material with the actual experimental acquired EDS spectrum

http://nist.gov/dtsa

Page 17: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

PENEPMA

Another high powered EPMA simulation program is PENEPMA, developed by Salvat and Llovet at U of Barcelona. It includes secondary fluorescence, though is a bit more complicated than DTSA-II. It can be used to create X-ray spectra in complicated geometries.

Page 18: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

Other Approaches to Particle Analysis

Another use of SEM to identify particles is electron backscatter diffraction (EBSD) which can use the diffraction pattern (Kikuchi lines) to match a database of crystal lattice data to ID a particular phase, though not the exact chemical composition. For example, it can distinguish hematite from magnetite, but not the amount of say Ti or Mn present.

Page 19: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

EBSD of Fe-oxide particlesElectron backscatter diffraction (EBSD) allows diffraction (”Kikuchi”) patterns to be collected in an SEM, upon a sample surface tilting (70°) sharply away from the electron beam, collected with a special phosphor or cc detector, and then the pattern is compared with a set of possible matching crystal structures (all computerized, of course).

A best fit (MAD < 0.5-1) is calculated if one of the structures can match for any possible orientation direction.

Page 20: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

Here 6 bands were used to match a particle from the smoke stack of a steel mill, to hematite, with a fit (MAD) of .55

Page 21: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

Other concerns regarding particles

• If very fine (e.g. powder), difficult to disperse on a substrate, so different particle sizes and potentially different compositions are clearly separate

• Difficult to apply conductive coat to side walls and underhangs, so that insulating materials will charge in high vacuum. Need to use rotating/swiveling coating attachment.

Page 22: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

Dispersion Technique-1

Ingredients:

• A small container with tight-fitting lid (here, 8 oz)

• Can of compressed air

• Mount with sticky tape (here, 1 cm Al SEM mount)

• Optional: something for mount to sit in/on so surface facing up (here: a large nut)

Step 1: Place a small amount of the powder/ small grains to be dispersed, along wall. Above, amount shown is really too much, use only a couple of grams of fine powder.

Page 23: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

Dispersion Technique-2

Step 2: Have already poked a small hole in the top, just large enough for the compressed air nozzle to stick through

Step 3: Poke nozzle thru hole and give very brief (1/10 second) blast. Wait 10-30 seconds before opening the top to investigate (10 seconds for coarse particles, 30+ seconds for fine powder)

Page 24: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

Dispersion Technique-3

Results: nice dispersal of <1 - 15 um fly ash particles on carbon tape. If not dispersed, they would all be in a thick mass and difficult to clearly see the individual particles.

Page 25: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

Inclusions = particles

Plan View

Cross section

In a polished mount, there is a large flat surface, but we would consider small inclusions (at or less than the incident electron interaction volume) to be “particles”. Here, instead of being surrounded on the sides by air (or vacuum), and on the bottom by some substrate, the complication of having x-rays potentially from 2 masses is obvious.

Page 26: Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

Conclusion

EPMA of particles

• is increasingly being applied in many situations;

• is not exactly straight forward;

• does yield results with potentially large inaccuracies/ errors;

• can only be considered qualitative, unless several careful steps are included