ANESTHETIC EQUIPMENT

Chapter 4 VAAA pp. 165-215

Page 2: Vets 238   Anesthetic Equipment Final

ENDOTRACHEAL TUBESPurpose:

To deliver anesthetic gas from the breathing circuit to the patient:

Physical Properties:• patient-end: beveled (slant-tip) to facilitate passage into the

trachea• machine- or circuit-end: connects to the breathing circuit

with an adapter

Made of:1.) red rubber: relatively inexpensive, potential problems:• a.) absorb disinfectant solutions >>> cause tubes to crack,

can irritate trachea• b.) extremely flexible >>> may kink or collapse easily

2.) vinyl plastic (transparent)• less porous, resist cracking, less flexible, become

stiff with age3.) silicone rubber• smooth, flexible, less irritating BUT

expensive

Page 3: Vets 238   Anesthetic Equipment Final

E-tubes (cont)

Variety of shapes:1) Murphy tube = has an eye near the bevel,

helps prevent complete obstruction of the tube

2) Magill tube = no eye near bevel3) Cole endotracheal tube = a narrowed

patient end, to facilitate passage into trachea

Sizes: • most commonly based on the internal

diameter (in millimeters)• size is written on the tube and matched

adaptor• Some are in “frenches”

Page 4: Vets 238   Anesthetic Equipment Final

E-tubes (cont)

Endotracheal Tubes with cuff:• cuff provides a seal between the tube and the trachea• Advantages of cuffed tubes

– cuff helps prevent leakage of waste gas around the tube– cuffed tubes reduce risks of aspiration of fluids (blood,

saliva, etc.) into lungs– animal prevented from breathing room air >>> better

able to regulate anesthesia

Use cuffed tubes with CAUTION:– excessive pressure in the cuff against the tracheal

mucosa can cause pressure necrosis of the mucosa – if the tube is too long, it will increase the “dead space

volume” or may be advanced into the animal too far = into one bronchus only

use caution with laser surgery >>> produces heat, may ignite the tube!!

Page 5: Vets 238   Anesthetic Equipment Final

Proper Tube Position

• Between larynx and bronchi

• Too far– Single side/lobe

entubated

• Too shallow– Laryngeal damage

Page 6: Vets 238   Anesthetic Equipment Final

ANESTHETIC MACHINES AND BREATHING CIRCUITS

Designed to deliver a gaseous (inhalant) anesthetic agent to and from the patient

• anesthetic agent is mixed with a carrier gas • usually oxygen alone or oxygen plus nitrous oxide

Functions of breathing circuit:1.) Delivers oxygen

– at controlled flow rate2.) Vaporizes liquid anesthetic

– at a carefully controlled concentration– mixes it with oxygen– delivers it to the patient

3.) Removes exhaled gases away from the patient for: – partial recirculation (with removal of carbon dioxide) or– elimination through a scavenging system

4.) Can be used in emergenciesto deliver only oxygen to a critically ill patient (vaporizer turned off)

Page 7: Vets 238   Anesthetic Equipment Final

COMPONENTS OF THE ANESTHETIC DELIVERY SYSTEM

When confused “follow the flow”trace O2 through system

Three major components of system:1. compressed gas supply2. anesthetic machine3. breathing circuit

Page 8: Vets 238   Anesthetic Equipment Final

1.) Compressed Gas Supply

– Source of oxygen • (room air = ?% O2)• Anesthesia machines can provide up to

100% O2

– Oxygen flow • must meet the metabolic needs of the

patient• carry vaporized inhalant anesthetic

agent to the patient

Page 9: Vets 238   Anesthetic Equipment Final

Compressed Gas Supply (cont)

a.) Gas Cylinders = compressed gas stored in metal cylinders

• Large volumes of gas stored in a relatively small container

• Under very high pressure ?• Outlet valve (at the top of the cylinder)

– controls outflow of gas• Attached directly to the anesthesia machine

via a yoke or• Free-standing tanks can be attached by hose

or by pipe• Safeguards make it almost impossible to

attach the wrong cylinder of gas can attach to the machine connections– Pin system – Color-coded

Page 10: Vets 238   Anesthetic Equipment Final

Compressed Gas Supply (cont)

b.) Tank Pressure Gauge O2: Directly related to portion of gas remaining in the tank

• change oxygen tank when pressure <100 psi• when to tell instructor in lab?

Nitrous Oxide also stored in compressed gas cylinders• much lower pressures• present in both liquid and gas states within the tank• gauge reads only the pressure of the gas phase• pressure of the gas phase does not decrease until all

of the liquid has evaporated (nearly empty)• weigh the tank to determine how much NO2 remains

in tank or change the tank when pressure falls below 500 psi

Page 11: Vets 238   Anesthetic Equipment Final

Compressed Gas Supply (cont)

c.) Pressure Reducing Valve • “Pressure regulator” reduces the

pressure from ? to ?• Provides a safe operating

pressure and allowing a constant flow of gas to the anesthesia machine

• Works regardless of the pressure in the tank (as long as it is above ?)

Page 12: Vets 238   Anesthetic Equipment Final

2.) Anesthetic Machine

Function: mixes oxygen with inhalant anesthetic agent and delivers the mixture (called “fresh gas” ) to the breathing circuit (i.e., the patient)

• a.) Flowmeter– Allows the anesthetist to regulate the rate

of gas flow to the patient• Measured in liters of gas per minute ( L/min)

– Gas passes through the flowmeter, the gas pressure is further reduced, from ? psi to ? psi

Page 13: Vets 238   Anesthetic Equipment Final

Anesthesia Machine (cont)

• b.) Vaporizer – oxygen coming out of the top of the

flowmeter flows to the vaporizer– the vaporizer converts the liquid anesthetic

to a vapor (gas) state AND carefully controls the concentration of the vaporized anesthetic added to the carrier gas(es).

– the carrier gas(es) plus the inhalant anesthetic are then carried to the “fresh gas outlet” and into the breathing circuit

Page 14: Vets 238   Anesthetic Equipment Final

VAPORIZERS (cont)The most complicated component of the whole machine, also

the most expensive part

Function: Adds inhalant anesthetic agent to the carrier gas

• Most vaporizers are designed to work with only one specific agent

– Color coded and labeled• Most vaporizers have an indicator window -- to show how

much fluid agent remains– Fill port

• If tipped over or shaken vigorously, vaporizers will leak liquid anesthetic agent into the bypass channels of the vaporizer. Thus, potentially lethal concentrations of anesthetic agent may be delivered to the next patient attached to this machine.

– Treatment: after the vaporizer has been tipped over or shaken vigorously, oxygen can be run through the vaporizer for 15 minutes with the vaporizer dial turned off.

• Need to be serviced regularly

Page 15: Vets 238   Anesthetic Equipment Final

1.) Precision Vaporizers

Designed to deliver an exact concentration of anesthetic agent

Dial is graduated in percent concentrations.

For inhalants with high vapor pressurei.e. Isoflurane, Sevoflurane

Page 16: Vets 238   Anesthetic Equipment Final

2.) Non-precision vaporizersFor inhalant agents with low vapor pressures

• i.e. Methoxyflurane will only produce a maximum concentration of about 4% in the carrier gas. This is relatively safe for most patients.

• Much more simple in design. Much less expensive.

• Often consist of only a glass jar and a cotton wick.

• Actually the concentration of the anesthetic agent delivered to the patient is not known– Dial does not correlate with %

Advantages: (a.) can be used with low flow rates, economical.(b.) lower initial cost than a precision vaporizer.

Page 17: Vets 238   Anesthetic Equipment Final

VOC versus VIC

VOC = “vaporizer out of circuit,” = the vaporizer is not placed within the breathing circuit.

Precision vaporizers are located out of the circuit because they offer high resistance to the flow of gases.

VIC = “vaporizer in circuit,” = the vaporizer is within the breathing circuit; the carrier gas(es) enter the circuit directly from the flowmeter; recycled exhaled gases pass through the vaporizer each time they flow through the circuit.

Non-precision vaporizers are usually placed within the circuit because they offer little resistance to the flow of gases through the circuit.

Page 18: Vets 238   Anesthetic Equipment Final

3 FACTORS INFLUENCE VAPORIZER FUNCTION

The concentration of inhalant anesthetic agent delivered by the anesthesia machine may be affected by several environmental factors:

a.) Temperature– Volatile anesthetic agents vaporize more

readily at higher temperatures.– cold room versus one in a warm or hot room.

Most precision vaporizers are “temperature compensated,” such that there is little variation in the output of anesthetic agent.

Non-precision or older precision vaporizers are not temperature compensated

• compensation can be done manually, using a thermometer and a temperature adjustment scale.

Page 19: Vets 238   Anesthetic Equipment Final

FACTORS THAT MAY INFLUENCE VAPORIZER FUNCTION

b.) Carrier Gas Flow Rate• The volume of carrier gas that flows through

the vaporizer may affect the efficiency of the vaporization of the agent.

• Most modern precision vaporizers are “flow compensated” designed to compensate for variations in carrier gas flow rates. These vaporizers will vaporize the concentration of agent indicated on the vaporizer dial over a wide range of carrier gas flow rates.

• However, even for these vaporizers, flow compensation is not unlimited. Flows that are very high (above 5 L/min) or very low (below 500 ml/min) may affect the vaporizer output even in a flow compensated precision vaporizer.

Page 20: Vets 238   Anesthetic Equipment Final

FACTORS THAT INFLUENCE VAPORIZER FUNCTION

c.) Back Pressure• Back pressure refers to an increase in

pressure within the breathing circuit, which may cause gas from the breathing circuit to flow back into the vaporizer under pressure. This may occur, for example, when an animal is “bagged.”

• Most precision vaporizers are “back pressure compensated” to minimize this risk.

Page 21: Vets 238   Anesthetic Equipment Final

3.) Breathing Circuit

Functions: carries “fresh gas” from machine to the patientcarries exhaled gases away from the patient

• Fresh gas inlet

• Unidirectional Valves– Inhalation unidirectional valve or inhalation flutter valve

allows gas flow in only one direction, toward the patient:

• when the patient inhales, this valve opens, allows fresh gas to flow through the inspiratory hose to the patient

– Exhalation unidirectional valve or exhalation flutter valve exhaled gas leaving the patient passes through the exhalation hose exhaled gas moves through this valve, can

move only away from the patient, through the carbon dioxide

absorption canister

Page 22: Vets 238   Anesthetic Equipment Final

3.) Breathing Circuit (cont)

b.) Reservoir Bag ( or “rebreathing bag” )• Functions:• (1.) reservoir bag stores excess gas• (2.) movement of the bag indicates that the endotracheal tube is in

place, that the patient is breathing: respiratory rate, tidal volume

• (3.) allows anesthetist to deliver oxygen to patient ( “bag” it prn )– Benefits of “bagging” the patient regularly:

•(a.) bagging reverses atelectasis

•(b.) anesthetized patients have decreased ability to

breath: have a reduced tidal volume -- Bagging flushes the airways.•

(c.) may be life-saving if patient stops breathing

Minimum volume of reservoir bag = 60 ml per kg BWKeep the reservoir bag properly inflated = about 3/4 full.Do not allow the bag to overfill.“Back pressure” = an increase in pressure in the breathing circuitBag should not empty completely with inspiration

Page 23: Vets 238   Anesthetic Equipment Final

3.) Breathing Circuit (cont)c.) Pop-off Valve (pressure relief valve)• can be turned fully open, partly open, or closed completely• when open, allows gas to escape from the breathing circuit into

scavenging system• prevents the build-up of excess pressure in the circuit• normally kept (at least partially) open when patient is breathing • adjust the valve to regulate amount of gas in the bag• if the patient is to be “bagged”: pop-off valve is closed, breath,

reopened

d.) Carbon dioxide absorber• Chemical crystals which absorb CO2• Granules of carbon dioxide absorber gradually become exhausted,

then will no longer absorb CO2. Need to check the granules:• (1.) Fresh granules crumble easily with finger pressure.• Granules saturated with CO2 become hard and

brittle.• “If hard, discard.”• (2.) White = good; Purple = Need to change

• Note: Color change may be transient.

Page 24: Vets 238   Anesthetic Equipment Final

3.) Breathing Circuit (cont)e.) Oxygen Flush Valve• allows oxygen to bypass the ? And ?• pure oxygen enters the breathing circuit: at 50 psi, at 50 to

70 L/min• should not be used with certain non-rebreathing systems

f.) Pressure Manometer (“pressure gauge”) (See Figure 4-10, page 186, VAAA.)

• measures the total pressure of gases in the breathing circuit• indicates the pressure of gases in the patient’s airways and

lungs• useful when bagging the animal: for dogs and cats with

healthy lungs• < 20 cm H2O or < 15 mm Hgg.) Negative Pressure Relief Valve• designed to open and admit room air into the breathing

circuit if a negative pressure develops within the circuit• e.g., an (over) active scavenging system or oxygen flow

rate is too low or tank runs out of oxygen• ensures that the patient always receives some oxygen

Page 25: Vets 238   Anesthetic Equipment Final
Page 26: Vets 238   Anesthetic Equipment Final
Page 27: Vets 238   Anesthetic Equipment Final

OPERATION OF THE ANESTHETIC MACHINE

decisions, decisions, decisions !!!!

A. Choice of Breathing Circuit (three major choices) (See Table 4-6, page 196, VAAA.)

• The choice of a breathing circuit will determine the following:– a.) where the exhaled gases go: partially or

totally recycled, eliminated entirely– b.) oxygen (and nitrous oxide) flow rate(s)– c.) status of the pop-off valve: open or closed– d.) type of breathing circuit apparatus: a Bain

system, Y-Piece tubing, Modified Y, etc •

Page 28: Vets 238   Anesthetic Equipment Final

1.) Rebreathing Systems

Allows recirculation of exhaled gases back to the patient = “rebreathing system”

• flow of gas through the circuit is circular, referred to as a “circle system”

• Can vary the amount of oxygen and anesthetic vapor (“fresh gas”) delivered to the patient and vary the amount of waste gases lost from the system

Page 29: Vets 238   Anesthetic Equipment Final

a.) Total Rebreathing System (aka “closed systems”)

• All of the exhaled gases remain in the circuit and are recirculated

• oxygen flow rate is very low, providing only enough oxygen to meet the metabolic needs of the patient

• only the volume of oxygen used by the patient is replaced through the oxygen flow meter (about 10 ml of 100% O2/kg/minute )

the pop-off valve is closed entirely

very economical = low flow rates of oxygen and anesthetic agent

Serious Risks of Total Rebreathing Systems:• a.) accumulation of CO2 >>> fully relying on an efficient

CO2 absorber• b.) increased pressure in circuit (“back pressure”) = must

balance O2 use by the patient with fresh gas input

Page 30: Vets 238   Anesthetic Equipment Final

b.) Partial Rebreathing System (aka “semiclosed system”)

some volume of gases exhaled by the patient remain in the circuit and are returned to the patient

some volume of gases exhaled by the patient are eliminated through the pop-off valve into the scavenger system

flow rate of fresh gas (oxygen plus anesthetic vapors) is much higher than for the total rebreathing system

pop-off valve is partially open, allowing some exhaled gases to flow into the scavenging system

Page 31: Vets 238   Anesthetic Equipment Final

2.) Non-Rebreathing Systems “NRB”

No exhaled gases are returned to the patient, evacuated into the scavenging system

Flow of oxygen: from tank >>> flowmeter >>> vaporizer >>> directly to patient (bypassing the flutter valves,Co2 scavenger, pop-off, etc)

exhaled gases go through another hose and may enter a reservoir bag, but do not go to a CO2 absorber canister, gases are eliminated through the scavenger system. Therefore, “non-rebreathing” system

Several of the components of the standard circle system are NOT used in the non-rebreathing system:

-both unidirectional valves, -the CO2 absorber canister-the pop off bag-O2 flush valve

Page 32: Vets 238   Anesthetic Equipment Final

2.) Non-Rebreathing Systems “NRB” (cont)

• Most anesthetic machines designed to be used with rebreathing systems can be converted to adapt to a non-rebreathing system.– a.) need a high oxygen flow rate = 200 to

300 ml/kg/minute or flow rates that match or exceed the patient’s “minute volume”

• ( = tidal volume X breaths per minute )

– b.) can use a variety of adapters to deliver fresh gas directly to patient and conduct exhaled gases to a scavenger:

• Bain system, Ayre’s T-piece, etc

Page 33: Vets 238   Anesthetic Equipment Final

Example: the Bain system (See Figures 4-13 and 4-14)

=Very common non-rebreathing system

consists of an inner tubing which conducts fresh gas to the patientallows incoming gases to be warmed by exhaled gases

Gas moving away from the patient through the outer corrugated tube enters a reservoir bag before leaving through the scavenging system

• This bag allows monitoring respirations and permits manual “bagging” if needed

– A Bickford valve attached to the side of the reservoir bag allows for the free flow of gases from the patient plus the overflow into the scavenger system

Normally the oxygen flow rate is set very high (at least 130 ml/kg/minute)Waste gases are flushed away to scavenging system

Low oxygen flow rates should be avoided because the Bain system does not remove any CO2; thus, it will accumulate within the system

Page 34: Vets 238   Anesthetic Equipment Final

Criteria for Choice of Rebreathing versus Non-Rebreathing System

1.) Patient size Non-rebreathing systems offer little resistance to the movement of gases. Use a non-rebreathing system for patients weighing less than 7 kg (15 lbs.)

2.) Convenience Non-rebreathing circuits are generally lighter, cause less pull on the endotracheal tube

3.) Cost Total rebreathing systems are the most economical; non-rebreathing systems are the most costly to use -- considering costs of O2 and the anesthetic agent.

4.) Control = ability to change the depth (planes) of anesthesia quickly

• rebreathing system>>>>relatively slow• non-rebreathing system >>>> much faster

Page 35: Vets 238   Anesthetic Equipment Final

Choosing which system (cont)

5.) Conservation of heat and moisture• Fresh gas coming from a vaporizer is cool

( 16 C) and dry (near 0% humidity). • Rebreathing systems warm and humidify the

gas in the circuit to the degree that the gases are recycled to the patient.

• With non-rebreathing systems, the warmed and humidified gases exhaled by the patient are lost through the scavenging system; the patient breathes only the dry, cool gas coming from the vaporizer, producing significant heat and water losses.

6.) Production of waste gases: less to more volume

produced• total rebreathing >> partial rebreathing>> non-rebreathing

Page 36: Vets 238   Anesthetic Equipment Final

B. Carrier Gas Flow Rates even more decisions!!

What flow rate of carrier gas is needed for each anesthetic procedure???

1. Flow Rates During Induction • Use higher flow rates during induction,

particularly if doing mask or chamber induction.

– For mask induction, use a flow rate of about 300 ml/kg/minute for cats and small dogs.

• Under 10 kg, use 1 to 3 L/minute.• Over 10 kg, use 3 to 5 L/minute.

– For chamber induction, • use a flow rate of 5 L/minute of oxygen.

Page 37: Vets 238   Anesthetic Equipment Final

Carrier Flow Rates (cont)

2. Flow Rates During Maintenance• Non-rebreathing systems, use relatively

high flow rates:• >130 ml/kg/minute for a Bain circuit• >200 ml/kg/minute for other circuits

• Partial rebreathing systems, use relatively low flow– use 25 to 50 ml/kg/minute (typically 1-2L/min)

Page 38: Vets 238   Anesthetic Equipment Final

CARE AND USE OF ANESTHETIC EQUIPMENT

Daily Setup each day before use, the anesthesia machine should be thoroughly checked for problems. ( See protocol for lab. )

•Ongoing Maintenance (periodic maintenance =proper performance)• 1. Oxygen (and Nitrous Oxide) Tanks• After use, turn off the outlet valve; remove pressure

remaining in machine by draining off oxygen via the oxygen flush button.

• 2. Oxygen FlowmeterTurn the dial to off position; do not overtighten.

• 3. Vaporizer– Before a procedure is begun, fill the vaporizer.– Turn off the vaporizer when it is not in use.

• These vaporizers need to be cleaned and recalibrated every one to two to three years depending on how much they are used.

– Signs of need of cleaning:• dial movement feels sticky, produces resistance• anesthetic in vaporizer turns brown• cannot maintain patient at surgical anesthetic plane even at high vaporizer

settings

Page 39: Vets 238   Anesthetic Equipment Final

Care of Equipment (cont)

4. Carbon Dioxide Absorber Canister• Check granules after each procedure for change of color.• Check granules for crushability before each day’s use of the

machine.• When the granules are replaced, minimize handling

granules; protect yourself and patient(s) from dust; do not pack granules tightly into canister; leave a cm or two of air space prevent dust from entering the tubing or hoses. Dust is corrosive to mucous membranes.

5. Cleaning Machine Parts (where water from the patient condenses)

• Some parts require periodic removal cleaning with a disinfectant to prevent buildup of water vapor, mucus, and dust:

• flutter valves, pop-off valve, Y hoses, modified F apparatus

• After each procedure, the removable parts of the breathing circuit should be washed in warm soapy solution, rinsed well, and allowed to air dry thoroughly.

Page 40: Vets 238   Anesthetic Equipment Final

Care of Equipment (cont)

6. Disinfecting Anesthetic Equipment• Equipment that contacts the patient’s airway

or oral cavity requires thorough disinfection.•• 7. Cleaning Endotracheal Tubes• No ideal agent for disinfection: chlorhexidine

= harmless to tissues, but does not kill all microorganisms and spores.

• All items exposed to disinfectants must be thoroughly rinsed with water and dried before use.

• Eventually such items deteriorate and must be replaced.