· web viewthe shape of a transient is related to local kinetics of the associated corrosion...
TRANSCRIPT
An integrated approach in the time, frequency and time-frequency domain for
the identification of corrosion using electrochemical noise
A.M. Homborga, R.A. Cottisb, J.M.C. Molc
aNetherlands Defence Academy, P.O. Box 505, 1780AM Den Helder, The
Netherlands, [email protected], tel. +31 223 65 75 58
bCorrosion and Protection Centre, School of Materials, University of Manchester,
Oxford Road, Manchester, M13 9PL, United Kingdom,
cDelft University of Technology, Department of Materials Science and
Engineering, Mekelweg 2, 2628CD Delft, The Netherlands, [email protected]
Abstract
Transients in electrochemical noise (EN) signals that are associated with
localized corrosion typically contain frequency information that is quite localized
in time. The study of these transients is therefore preferably performed using
analysis procedures with high discrimination ability in both time and frequency
simultaneously. The present work studies the combination of the Hilbert-Huang
transform with the continuous wavelet transform, as data analysis methods that
operate in time as well as in frequency, for the identification of localized
corrosion. Additionally it incorporates fast Fourier transform as a technique to
determine the distribution of power over frequency. Finally, it integrates the
study of transient shape into the analysis of (time-) frequency properties. It is
1
shown that the maxima in instantaneous frequency amplitudes of transients with
otherwise different shapes can be similar. In the case of the presence of multiple
transients in the EN signal, the cut-off frequency obtained from power spectral
density plots is related to maxima in instantaneous frequency amplitudes of only
a limited number of transients.
Keywords
Electrochemical noise; Hilbert-Huang transform; Wavelet transform; Spectral
analysis; Transient shape
1 Introduction
Electrochemical noise (EN) measurements are principally intended to record
charge transfer that originates from spontaneous corrosion, while keeping
perturbation of the process to a minimum. Whereas measuring EN may appear
straightforward, in order to obtain meaningful mechanistic information from the
chemo-physical process it is essential to employ the appropriate data analysis
techniques [1, 2]. Amongst these, conceivably direct visual characterization of EN
transients in the raw data signal is the most straightforward. However, the
analysis of transient shape requires the possibility to visually locate and identify
transients individually. This necessity potentially decreases the suitability of
visual identification for this purpose. In the case of localized corrosion processes
that occur more or less at the same time, the resulting transient overlap may
impede the investigation of transient shape.
2
Operating in the frequency domain, for the analysis of EN signals fast Fourier
transform (FFT) is most often used to determine the distribution of power over
frequency, usually in the form of the power spectral density, or PSD [1]. An
important limitation of FFT is the requirement for a stationary signal, which is
often not met for EN signals.
The PSD shows specific characteristics that can indicate specific types of
corrosion: Firstly, the overall level of the PSD is reported to be related to the
intensity of the process [3, 4]. Secondly, the cut-off-, roll-off- or knee frequency,
which indicates the transition between the horizontal low-frequency part and
the slope at higher frequencies, can be used to discriminate between different
types of corrosion [4, 5]. As a third parameter, the slope in the higher frequency
part can be considered to differentiate between different corrosion types [3, 4, 6-
8]. It should however be noted that in some cases the discrimination ability of
the PSD can vary between the electrochemical current (ECN) or potential noise
(EPN) signal, and that it may be difficult to indicate the presence of a low-
frequency plateau or cut-off frequency [2].
A more recent development is the use of data analysis procedures operating in
the time-frequency domain for the investigation of EN [5, 9]. As an example,
corrosion studies using the Hilbert-Huang transform (HHT) have been reported
to effectively identify microbiologically influenced corrosion (MIC) [10] and
corrosion inhibition [11]. The HHT was first introduced by Huang et al. [12] and
is based on the extraction of characteristic scales, or intrinsic mode functions
from the EN signal, using a procedure called empirical mode decomposition
(EMD). After EMD, instantaneous frequencies are calculated from these intrinsic
oscillation modes [12]. The differentiation of those instantaneous frequencies in
3
time allows the analysis of nonlinear and non-stationary EN signals. To a certain
degree, it combines advantages of analysis procedures operating in the
frequency- as well as in the time domain. A different approach is to describe the
original EN signal using a linear combination of oscillations of a limited time
span, called wavelets. By scaling and translation of these wavelets, these allow
analysis of the signal at different timescales. The energy of each timescale can be
plotted in an energy distribution plot (EDP), where the dominant process is
associated with the timescale with the highest relative energy [5, 7]. Aballe et al.
[7] were the first to compare (discrete) wavelet transform (DWT) with FFT from
the perspective of their ability to identify different corrosion mechanisms. The
PSD was found to indicate the dominant timescale in the EN signal,
corresponding with observations from the EDP. However, it appeared that the
PSD could hardly discriminate between features in the EN signal related to
processes occurring in different timescales, whereas the DWT could. The PSD is
however closely related to the EDP determined by DWT [13]. Combined with the
apparently superior frequency resolution as provided by continuous wavelet
transform (CWT) with respect to DWT [13], it is interesting to investigate
whether information obtained by CWT can be confirmed by FFT. This is
therefore one of the primary objectives of the present work, together with the
comparison between the use of CWT- and Hilbert spectra for the analysis of EN
signals.
In this work, the combination of transient shape analysis, investigation of the
PSD and time-frequency spectra obtained from CWT and HHT is investigated.
Localized corrosion processes involving AISI304, carbon steel and AA2024-T3,
4
which were already analysed in detail by introducing transient analysis through
Hilbert spectra in earlier work [10, 11, 14], serve as case studies.
2 Experimental
All EN measurements involved a classical, open circuit configuration with two
working electrodes and a reference electrode. The experimental details are
identical to the ones described earlier for AISI304 [9], AA2024-T3 [11] and MIC
[10].
The AISI304 and AA2024-T3 working electrode surfaces were pre-treated
through wet grinding with up to 4000-grit SiC paper and rinsing with
demineralized water. Subsequently, the working electrodes were kept in a dry
environment at 20 °C for 1 day and analysed for imperfections prior to exposure.
Each working electrode covered an area of 0.05 cm2 and was either exposed to
an aqueous 10-4 M HCl (AISI304) or 10-1 M NaCl (AA2024-T3) electrolyte. The
electrolytes were produced from reagent of analytical quality, dissolved in
demineralized water. The experiments were carried out under aerated
conditions. A Red Rod REF201 reference electrode (Ag/AgCl/sat. KCl: 0.207 V vs.
SHE), from Radiometer Analytical, served as reference electrode.
The MIC experiments involved working electrodes that were made from carbon
steel, with working electrode areas of 19,6 mm2 each. Here, a platinum mesh
with an area of approximately 100 mm2 served as reference electrode. A
Postgate C solution [15] with additionally 2,5 wt.% NaCl, produced from reagent
of analytical quality that was dissolved in demineralized water, was used both as
electrolyte and as medium for the bacteria.
5
A Faradaic cage protected the measurement setup from electromagnetic
distortions. The atmospheric temperature of the surrounding area was regulated
at 20 °C. After each experiment, the working electrodes were analysed using an
optical microscope (Reichert MEF4 M) with a maximum magnification of 1000x.
Each experiment was repeated at least two times.
The ECN and EPN signals were measured using an Ivium Technologies
Compactstat potentiostat. For the experiments on AISI304 a sampling rate of 5
Hz was selected and for the measurements on AA2024-T3 and MIC 20 Hz was
used, all of which included the application of a low-pass filter of 10 Hz.
The CWT was computed with an analytic Morlet wavelet using the cwtft function
in Matlab, which uses the discrete Fourier transform algorithm [16]. Whatever
the algorithm used, a key decision when computing the CWT is the way in which
the signal is extended at the beginning and at the end of the time record (known
as ‘padding’). This is necessary because the wavelet must overlap the ends of the
time record by half its duration when computing the first and last points of the
spectrum. Various padding methods can be used (see [17] for the methods
available in Matlab) and the method chosen has a significant effect on the
artefacts produced at the beginning and at the end of the spectrum. For this work
symmetric padding was found to provide the best results. 32 Voices per octave
(i.e. 32 logarithmically spaced frequencies were computed for each factor of two
frequency range) were used to calculate the CWT. The Hilbert–Huang transform
was calculated by the application of a Matlab program developed by Rilling et al.
[18]. The analysis of transients was executed analogously to the procedure as
described in [11, 14]. In order to be able to calculate a reliable PSD, the DC drift
of each EN signal was first identified through discrete wavelet transform and
6
subsequently removed from the signal before application of the FFT. Further
details on this trend removal procedure were published earlier in [19].
3 Results and discussion
This section starts with the analysis of single transients. Subsequently EN signals
containing transients that occur over different timescales and that in some cases
appear overlapped, are treated. Finally, a corrosion process that gradually
changes over time is discussed.
3.1 Single transient analysis
EN signals consisting of only one transient are perhaps the most straightforward
to investigate. Both transient shape and frequency characteristics are relatively
easy to analyse. In the absence of multiple (different) transients which all
contribute to the overall frequency spectrum, it is expected that the overall
frequency characteristics visible in the PSD are similar to those determined by
the HHT and CWT.
Figure 1 shows the ECN and EPN time signal of AISI304 exposed in a 10-4 M HCl
solution for a duration of 1000 s. Figures 2 and 3 show the CWT- and Hilbert
spectra, respectively.
Figure 1
Figure 2a
7
Figure 2b
Figure 3a
Figure 3b
One transient occurred, at approximately t = 800 s. The shape of a transient is
related to local kinetics of the associated corrosion process. In the case of pitting
corrosion, the characteristics of EN transients directly reflect processes
occurring in the associated pits [20]. Identification of an individual transient is
facilitated in cases where only one transient is present in the EN signal, or
otherwise when pits do not arise simultaneously and consequently their
associated transients do not overlap. For the ECN and EPN transients shown in
Figure 1 the gradually increasing amplitude reflects the metastable growth phase
of a pit. Subsequently, a rapid decay in the ECN transient indicates fast
repassivation and a slow decrease in magnitude of the EPN transient (i.e. a slow
recovery of the potential value in the positive direction) is associated with the
slow discharge of interfacial capacitance [20]. The increasing slope towards the
peak of the transients can be explained by an increase in the charge transfer rate:
apparently, the metastable growth phase is characterized by an increase in pit
growth rate. As a result of this, different steeper and shallower sloping regions
can be identified in the transients during the metastable growth phase.
The instantaneous frequency information provided by the Hilbert- and CWT
spectra of the ECN signal is comparable, with maxima around 6•10-2 Hz. The
Hilbert spectrum of the EPN signal shows a maximum at a lower frequency (2•10-
2 Hz), which can be attributed to the influence of the discharge of interfacial
8
capacitance in the overall EPN transient, which is a relatively slow process [9].
The increased amplitude of instantaneous frequencies around 4•10-3 Hz ahead of
the transient are likely to be artefacts arising from the empirical mode
decomposition that can be disregarded by the use of a proper transient analysis
procedure [14]. The CWT spectrum shows maxima around the lowest frequency
ranges of the spectrum (approximately 5•10-3 Hz).
As a comparison, Figure 4 shows a different kind of transient present in the ECN
and EPN time signal of again AISI304 working electrodes exposed in the same
10-4 M HCl solution for a duration of 1000 s. Figures 5 and 6 show the CWT- and
Hilbert spectra, respectively.
Figure 4
Figure 5a
Figure 5b
Figure 6a
Figure 6b
Although both working electrode sets were composed of AISI304 and received
similar pre-treatment, the transient visible in the ECN and EPN signal in Figure 4
lasts considerably longer than the one of the signal in Figure 1. The rate of the
repassivation phase is comparable, whereas the preceding metastable growth
phase is considerably longer in this case. In addition, the transient shown here
9
clearly indicates that, instead of a steady metastable growth, this phase is
characterized by consecutive bursts of charge. Each of those bursts generates a
fast increase of current locally, associated with steep parts in the slope of the
ECN transient. The accumulation of those processes however results in an
overall metastable growth phase that lasts considerably longer, and is in addition
slower than in the previous case. These differences in kinetics are visible in the
Hilbert- and CWT spectra. The Hilbert spectrum of the ECN signal shows a
maximum between 10-2 and 2•10-2 Hz, which is considerably lower than the 6•10-
2 Hz of the spectra of Figures 2a and 3a. The CWT spectrum of the ECN signal has
an overall maximum at 5•10-3 Hz, although here an additional peak is visible
between 10-2 and 2•10-2 Hz, at the time instant of the transient. The Hilbert- and
CWT spectrum of the EPN signal both show a maximum at 2•10-2 Hz, although the
overall maximum of the CWT spectrum is located at 5•10-3 Hz. It is worth noting
that, although the transients in Figure 1 and 4 share different durations, the
Hilbert- and CWT spectra of the EPN signals have local maxima in a similar
frequency range. Moreover, especially in the Hilbert spectra these maxima are
located at the same timeframe as the final part of their corresponding EPN
transient. By this way these spectra confirm the observation from transient
analysis that the rate of the repassivation phase is comparable in both cases,
regardless of the duration of the metastable growth phase.
Although less apparent, the part of the ECN transient in Figure 1 associated with
metastable growth also starts and ends with a steep part. It can be assumed that
this is generated by a comparable mechanism, as is the case for the transient in
Figure 4. Because the rate of the total metastable growth phase is higher than in
the case of the transient in Figure 4, the higher instantaneous frequencies
10
associated with these steep parts probably have a larger relative influence in the
Hilbert spectrum and, as a consequence, the dominating instantaneous
frequencies are in a higher range than compared to those in the Hilbert spectrum
of the ECN transient in Figure 4. Note that, with 16 nA, the absolute magnitude of
the brief transient in Figure 1 is even approximately 3 nA larger than that of the
longer lasting transient in Figure 4.
Figure 7 shows the PSDs of both the EN signals from Figure 1 (1) and 4 (2). By
this way, data analysis in the frequency domain can be combined with the
previous results from HHT and CWT. A vertical line indicates the cut-off
frequency of each PSD graph.
Figure 7a
Figure 7b
Comparing the cut-of frequencies visible in the PSD plots of Figure 7 reveals that
these correspond with the dominating instantaneous frequencies of the Hilbert
spectra and that, for the CWT spectra, these correspond only for the ECN signals
and for the EPN signal of Figure 4. In Figure 7a, the PSD of the transient from
Figure 1 shows a cut-off frequency of approximately 6•10-2 Hz and the PSD of the
transient from Figure 4 has a cut-off frequency of 10-2 Hz. The PSDs of the EPN
signals both show a cut-off frequency at 10-2 Hz.
3.2 Multiple transient analysis
11
Transient shape analysis can be useful to detect the nature of pitting corrosion.
In most cases, EN signals contain multiple transients. This subsection focuses on
the analysis of such signals. Figure 8 shows the ECN time signal of (a) again
AISI304 exposed in a 10-4 M HCl solution for a duration of 1000 s and (b) pitting
corrosion of carbon steel due to MIC after 13 days of immersion for a duration of
1000 s. In the case of the MIC experiments, detection of variations in EPN was
hindered by the use of a platinum mesh as reference electrode, which was
necessary for sterility considerations. Therefore, the EPN signal could only be
used for determination of the average open corrosion potential. In order to
facilitate transient analysis, Figure 9 shows an enlargement of the transients of
these signals, after a transient identification procedure as presented in earlier
work [11, 14]. Determination of transient duration can be achieved by
observation where the transient amplitude is nonzero. The procedure is based
on instantaneous frequency information in Hilbert spectra and is primarily
intended to avoid the influence of artefacts from those spectra [14]. Therefore,
differences with visual transient start and end points may occur. The transients
in the original ECN and EPN signal are indicated from left to right with 1-5
(Figure 9a) or 1-6 (Figure 9b).
Figure 8a
Figure 8b
Figure 9a
Figure 9b
12
Analogous to the transients from the EN signals in Figures 1 and 4, the transients
shown in Figure 9a exhibit an increasing slope towards their maximum absolute
value, which again indicates that the metastable growth phase is characterized
by an increase in pit growth rate. The duration of transient 3 is the largest,
approximately 23 s. This is the only transient in the positive direction in the
original time signal shown in Figure 8a. Transients 1, 2, 4 and 5 are in the range
of 8 to 15 s. For transient 3, the amplitude is almost 9 nA and transients 1, 2, 4
and 5 range between 3 and 6 nA. It appears here that a large transient duration is
not necessarily related to large amplitudes: the transient with the largest
timespan (transient 5, duration 15 s) shows a maximum amplitude of only 3 nA.
It is interesting to note the difference in transient shapes between Figures 9a and
9b. Where ECN transients in Figure 9a are related to initiation, slow metastable
growth and fast repassivation of a pit, those in Figure 9b are produced by
acidification at the interface between metal substrate and biofilm due to the
metabolism of SRB that causes a local corrosion attack [21, 22]. The SRB reduce
sulphate, which is present in the electrolyte and acts as terminal electron
acceptor [21, 23, 24]:
(1)
The resulting sulphide reacts with iron ions that are present due to the anodic
reaction, hence resulting in acidification of the anodic site along with the
production of iron sulphide [21, 22]:
13
(2)
Subsequently, migration of the H+ ions through the biofilm and away from the
substrate generates a decrease in transient magnitude [10]. All transients show a
similar characteristic shape with a more or less horizontal segment at the right
hand side, in which the decrease in charge transfer rate ceases for a moment.
Contrary to metastable pitting of AISI304, the largest transient also shows the
largest duration here, which can be explained by the activity of SRB. In case of a
thicker biofilm, the H+ production due to their metabolism can be high locally,
generating a larger pit in the metal substrate. The thick biofilm consequently
prevents fast migration of those H+ ions, thereby contributing to a larger
transient duration.
In order to be able to investigate the time-frequency characteristics, Figure 10
shows the CWT- (a) and Hilbert (b) spectrum of the ECN time signal of AISI304
exposed in a 10-4 M HCl solution for a duration of 1000 s and Figure 11 shows the
CWT- (a) and Hilbert (b) spectrum of the ECN time signal of pitting corrosion of
carbon steel due to MIC after 13 days of immersion for a duration of 1000 s.
Figure 10a
Figure 10b
Figure 11a
Figure 11b
14
Both the Hilbert- and CWT spectra of the ECN signal from Figure 10 show that
the largest transient, occurring after 200 s, is characterized by maxima in the
instantaneous frequency range between 2•10-2 and 3•10-2 Hz. In addition, the
remaining smaller transients are characterized by maxima in the order of 10-1 Hz
in both spectra, whereas those of the first transient are in a lower instantaneous
frequency region, around 4•10-2 Hz.
In Figure 11, the last transient (6 in Figure 9b), that was considerably longer
than the other five, is also associated with the lowest instantaneous frequencies
in the CWT- and Hilbert spectra. In the CWT spectrum, the largest amplitude for
this transient is at 4•10-2 Hz, whereas for the other transients maxima are
approximately present at 2•10-1 Hz. In the Hilbert spectrum, at 2•10-1 Hz local
maxima are also distinguishable, however the sharp transient peaks dominate
the spectrum to a larger extent, with maxima above 1 Hz for all transients. The
contribution of the longer second part of the last, large transient is visible in
increased instantaneous frequency amplitudes around 10-2 Hz.
Whether or not assisted by separate transient analysis figures, in the above cases
transient shapes can be relatively easily distinguished. This is however not
always the case. Figure 12 shows the ECN signal from a measurement on the
same measurement cell at the same day as the data presented in Figure 8b.
Figure 13 shows the CWT- (a) and Hilbert (b) spectra, respectively.
Figure 12
Figure 13a
15
Figure 13b
In the ECN signal of Figure 12, transients are increasingly overlapping. This is
made visible by a magnification of the transients occurring between 640 s and
660 s. The use of smaller working electrode surfaces can prevent this
phenomenon to a certain extent, however in most cases the semi-stochastic
nature of corrosion determines the amount of localized processes that occur
simultaneously. The sole use of transient shape analysis as identification
parameter for this specific type of localized corrosion becomes insufficient in
those cases. Despite the overlap in transients however, local frequency
characteristics remain largely preserved. The time-frequency techniques shown
in Figure 13 are therefore still able to visualize the instantaneous frequency
characteristics. The dominant instantaneous frequency ranges are comparable
with those of the spectra shown in Figure 11, which is expected since both ECN
signals originate from the same measurement cell under similar experimental
conditions.
Analogous to single transient analysis, also for the EN signals discussed here data
analysis in the frequency domain is combined with the results from HHT and
CWT. Figure 14a shows the PSD of the ECN signal from Figure 8a and Figure 14b
shows the PSDs of the ECN signals from Figure 8b (1) and 12 (2). A vertical line
again indicates the cut-off frequency of each PSD graph.
Figure 14a
Figure 14b
16
The cut-off frequency of the PSD graph of Figure 14a lies between 2•10-2 and
3•10-2 Hz. This corresponds with the transient characteristics of the large
transient occurring at 200 s in the CWT- and Hilbert spectrum of Figure 10. It
appears that this frequency information dominates the overall PSD to a large
extent, since the other transients show different frequency characteristics in the
CWT- and Hilbert spectrum that remain invisible in the PSD. The cut-off
frequencies of the PSD graphs of Figure 14b are approximately 10-1 Hz. This
value reflects the instantaneous frequency characteristics of the majority of the
transients: although the large transient at approximately t = 900 s in Figure 8b
has a lower dominant instantaneous frequency range (around 4•10-2 Hz), the
other five transients influence the PSD to a larger extent, resulting in an overall
PSD that is comparable with the one from the ECN signal in Figure 12, hence in a
loss of information. Contrary to the PSD in Figure 14a, in this case the largest
transient does not influence the PSD to the largest extent. This effect could be
explained by the smaller relative difference between transient sizes, and hence
the increased relative overall energy contribution of the smaller transients in
this case (compare the transients in Figure 9b with those in Figure 9a). It should
be noted that both PSD graphs of Figure 14b are similar, which reflects the
comparable corrosion characteristics in both cases, regardless of transient
overlap.
3.3 Overlapping transients and non-stationarity
17
In some cases, corrosion processes change significantly over time. Here, non-
stationarity is not only limited to the occurrence of transients, but transients also
change significantly in shape and duration over time. In addition, transients may
increasingly overlap as the corrosion process proceeds. As an example, Figure 15
shows the ECN and EPN time signal of AA2024-T3 exposed in a 10-1 M NaCl
solution for a duration of 1000 s.
Figure 15
Visual inspection of the ECN signal reveals a significant change in transient
characteristics, with larger transients of shorter duration occurring in an
increasing rate towards the end of the signal. Their number and overlapping
nature hinders clear visual identification. HHT and CWT both provide a means to
obtain clear time-frequency information in those cases, without the loss of time-
dependent features in the signal. Figures 16 and 17 show the CWT- and Hilbert
spectra of these EN signals, respectively.
Figure 16a
Figure 16b
Figure 17a
Figure 17b
The CWT spectrum in Figure 16a shows increasing contribution of frequencies
around 10-1 Hz in time, associated with the increasing number of short transients
18
in the signal. However, along the entire signal frequencies below 10-2 Hz
dominate the spectrum. The EPN CWT spectrum displayed in Figure 16b shows a
peak at 2•10-2 Hz. Compared to the CWT spectrum of Figure 16a, in the Hilbert
spectrum shown in Figure 17a the increasing magnitude of frequencies around
10-1 Hz in time is even more apparent. In addition, throughout the signal also a
considerable amount of low-frequency information is present, below 10-2 Hz.
Analogous to the EPN CWT spectrum in Figure 16b, in the EPN Hilbert spectrum
in Figure 17b each of the two large transients occurring after 550 s generates a
peak in instantaneous frequencies at 2•10-2 Hz. It must be appreciated here that
the resolution in time and frequency of the CWT are a function of the selected
wavelet. The selection of a Morlet wavelet with a lower sine frequency can
increase the resolution in time (while decreasing the resolution in frequency)
[13]. Also it must be appreciated that the Heisenberg Uncertainty Principle
implies a limitation on the simultaneous resolution in time and frequency.
Implicitly, the resolution of intrinsic mode functions (IMFs), resulting from the
empirical mode decomposition, is indeed decreasing with increasing IMFs. This
is a result from the interpolation between local maxima and minima that are
located further apart after each iteration, as the highest frequency components
are removed each time [9, 12].
Finally, data analysis in the frequency domain is again combined with the results
from HHT and CWT. Figure 18 shows the PSDs of the EN signals from Figure 15.
A vertical line indicates the cut-off frequency of each PSD graph.
Figure 18a
19
Figure 18b
For the ECN signal, the cut-off frequency is approximately 10-2 Hz, whereas for
the EPN signal the cut-off frequency is 2•10-2 Hz. This agrees well with
observations in the time-frequency spectra for the overall signals. Again, the lack
of differentiation in time is important to note in this case. Information like the
increasing magnitude of frequencies around 10-1 Hz in time that was observed in
the time-frequency spectra, originating from an increasing density of fast
transients, is lost in the PSD. Due to considerable transient overlap, this
phenomenon is also not obvious using visual identification. In addition, the
origin of the increased magnitude of the instantaneous frequencies around 2•10-2
Hz in the time-frequency spectra of the EPN signal, which could be associated
with the two large and slow transients occurring after 550 s, cannot be
determined using only the PSD.
3.4 Overall discussion
In the case of the presence of only one transient or multiple similar transients in
an EN signal, the cut-off frequency obtained from the PSD plot corresponds well
with the highest amplitude instantaneous frequencies in CWT- or Hilbert
spectra. However, once differences in transient magnitude or duration emerge
within the signal (i.e. the corrosion process changes over time), this correlation
becomes less straightforward. Here, the time-frequency techniques are powerful
tools to locate differences in kinetics over time. The visual characterization of
transient shape can be a quite useful additional tool, provided that the transients
20
can be detected accurately. This is challenging, considering the semi-stochastic
nature of the underlying process and possible transient overlap. Provided that
transient start and end can be determined, its shape allows localization of the
anodic process in time, the (change in) charge transfer rate, peak current or
potential and duration of the first part towards the maximum amplitude and
from there towards the transient end. Together with time-frequency
information, this ability aids the understanding of the physicochemical process
that generated the transient.
4 Conclusions
In the time domain, transient shape analysis can be used as a first global
corrosion identification step with respect to its charge transfer kinetics, prior to
further data analysis. It is a suitable technique in order to enable the
differentiation of corrosion processes in a broad sense, without the necessity of
more sophisticated data analysis tools.
In the frequency domain, a certain agreement exists between the cut-off
frequency in a PSD plot and maxima observed in CWT- and Hilbert spectra. In
some cases however, only a small part of the EN signal can have a significant
impact on the PSD. This increases the risk of loosing essential details about the
corrosion process during a measurement. If a corrosion process is more or less
in a steady state this is acceptable, and the PSD can provide a good overall
impression of the frequency characteristics of the investigated EN signal.
However, for more dynamic systems that change over time, differentiation in
time is required.
21
Finally, in the time-frequency domain, CWT spectra as derived in this work
provided similar information as Hilbert spectra. The preference for the use of
either of both time-frequency techniques could depend on the required
application and further data analysis.
5 References
[1] R.A. Cottis, Corrosion, 57 (2001) 265-285.[2] A.M. Homborg, T. Tinga, E.P.M. van Westing, X. Zhang, G.M. Ferrari, J.H.W. de Wit, J.M.C. Mol, Corrosion, 70 (2014) 971-987.[3] T. Zhang, Y. Shao, G. Meng, F. Wang, Electrochim. Acta, 53 (2007) 561-568.[4] A.-M. Lafront, F. Safizadeha, E. Ghali, G. Houlachi, Electrochim. Acta, 55 (2010) 2505-2512.[5] A. Aballe, M. Bethencourt, F.J. Botana, M. Marcos, Electrochem. Commun., 1 (1999) 266-270.[6] K. Hladky, J.L. Dawson, Corros. Sci., 22 (1982) 231-237.[7] A. Aballe, M. Bethencourt, F.J. Botana, M. Marcos, Electrochim. Acta, 44 (1999) 4805-4816.[8] G. Du, J. Li, W.K. Wang, C. Jiang, S.Z. Song, Corros. Sci., 53 (2011) 2918-2926.[9] A.M. Homborg, E.P.M. van Westing, T. Tinga, X. Zhang, P.J. Oonincx, G.M. Ferrari, J.H.W. de Wit, J.M.C. Mol, Corros. Sci., 66 (2013) 97-110.[10] A.M. Homborg, C.F. Leon Morales, T. Tinga, J.H.W. de Wit, J.M.C. Mol, Electrochimica Acta, 136 (2014) 223-232.[11] A.M. Homborg, E.P.M. van Westing, T. Tinga, G.M. Ferrari, X. Zhang, J.H.W. de Wit, J.M.C. Mol, Electrochim. Acta, 116 (2014) 355-365.[12] N.E. Huang, Z. Shen, S.R. Long, M.C. Wu, H.H. Shih, Q. Zheng, N.C. Yen, C.C. Tung, H.H. Liu, Proc. R. Soc. London, 454 (1998) 903-995.[13] R.A. Cottis, A.M. Homborg, J.M.C. Mol, Electrochimica Acta, 202 (2015) 277-287.[14] A.M. Homborg, T. Tinga, X. Zhang, E.P.M. van Westing, P.J. Oonincx, G.M. Ferrari, J.H.W. de Wit, J.M.C. Mol, Electrochim. Acta, 104 (2013) 84-93.[15] J.R. Postgate, The sulphate-reducing bacteria, Ed. Cambridge University Press, Cambridge, 1984.[16] The MathWorks, Inc., (2016). Wavelet Toolbox: User's Guide (R2016a). Retrieved July 1, 2016 from http://www.mathworks.com/help/pdf_doc/wavelet/wavelet_ug.pdf.[17] The MathWorks, Inc., (2016). Image Processing Toolbox: User's Guide (R2016a). Retrieved July 1, 2016 from http://www.mathworks.com/help/pdf_doc/wavelet/wavelet_ug.pdf.[18] G. Rilling, P. Flandrin, P. Goncalves, IEEE-EURASIP workshop on Nonlinear Signal and Image Processing NSIP-03, Grado, Italy, (2003) 1-5.
22
[19] A.M. Homborg, T. Tinga, X. Zhang, E.P.M. van Westing, P.J. Oonincx, J.H.W. de Wit, J.M.C. Mol, Electrochim. Acta, 70 (2012) 199-209.[20] G. Berthomé, B. Malki, B. Baroux, Corros. Sci., 48 (2006) 2432-2441.[21] B.W.A. Sherar, I.M. Power, P.G. Keech, S. Mitlin, G. Southam, D.W. Shoesmith, Corros. Sci., 53 (2011) 955-960.[22] W.A. Hamilton, Biofouling, 19 (2003) 65-76.[23] T. Gu, K. Zhao, S. Nesic, Corrosion, (2009) 1-12.[24] T. Gu, D. Xu, Demystifying MIC Mechanisms, in: NACE International, 2010.
Figure captions
Figure 1 ECN and EPN time signal of AISI304 exposed in a 10-4 M HCl solution for
a duration of 1000 s
Figure 2 CWT spectrum of the ECN (a) and EPN (b) signal of AISI304 exposed in
a 10-4 M HCl solution for a duration of 1000 s
Figure 3 Hilbert spectrum of the ECN (a) and EPN (b) signal of AISI304 exposed
in a 10-4 M HCl solution for a duration of 1000 s
Figure 4 ECN and EPN time signal of AISI304 exposed in a 10-4 M HCl solution for
a duration of 1000 s
Figure 5 CWT spectrum of the ECN (a) and EPN (b) signal of AISI304 exposed in
a 10-4 M HCl solution for a duration of 1000 s
Figure 6 Hilbert spectrum of the ECN (a) and EPN (b) signal of AISI304 exposed
in a 10-4 M HCl solution for a duration of 1000 s
23
Figure 7 PSDs of the ECN (a) and EPN (b) signals of Figure 1 (1) and 4 (2) of
AISI304 exposed in a 10-4 M HCl solution for a duration of 1000 s
Figure 8 ECN time signal of (a) AISI304 exposed in a 10-4 M HCl solution for a
duration of 1000 s and (b) pitting corrosion of carbon steel due to MIC after 13
days of immersion for a duration of 1000 s
Figure 9 Transients of the ECN signal of (a) AISI304 exposed in a 10-4 M HCl
solution for a duration of 1000 s and (b) pitting corrosion of carbon steel due to
MIC after 13 days of immersion for a duration of 1000 s, from left (1) to right (5
or 6)
Figure 10 CWT- (a) and Hilbert (b) spectrum of the ECN signal of AISI304
exposed in a 10-4 M HCl solution for a duration of 1000 s
Figure 11 CWT- (a) and Hilbert (b) spectrum of the ECN signal of pitting
corrosion of carbon steel due to MIC after 13 days of immersion for a duration of
1000 s
Figure 12 ECN time signal of pitting corrosion of carbon steel due to MIC after 13
days of immersion for a duration of 1000 s
Figure 13 CWT- (a) and Hilbert spectrum (b) of the ECN signal of pitting
corrosion of carbon steel due to MIC after 13 days of immersion for a duration of
1000 s
24
Figure 14 (a) PSD of the ECN signal of AISI304 exposed in a 10-4 M HCl solution
for a duration of 1000 s and (b) PSDs of the ECN signals of pitting corrosion of
carbon steel due to MIC after 13 days of immersion for a duration of 1000 s
Figure 15 ECN and EPN time signal of AA2024-T3 exposed in a 10-1 M HCl
solution for a duration of 1000 s
Figure 16 CWT spectrum of the ECN (a) and EPN (b) signal of AA2024-T3
exposed in a 10-1 M HCl solution for a duration of 1000 s
Figure 17 Hilbert spectrum of the ECN (a) and EPN (b) signal of AA2024-T3
exposed in a 10-1 M HCl solution for a duration of 1000 s
Figure 18 PSDs of the ECN (a) and EPN (b) signal of Figure 15 of AA2024-T3
exposed in a 10-1 M HCl solution for a duration of 1000 s
25