spiral.imperial.ac.uk  · web view2020. 7. 1. · the number of electrons accelerated towards the...

34
Image reconstruction: Part 1 – understanding filtered back projection, noise and image acquisition. Schofield R 1 , King L 2 , Tayal U 1 , Castellano I 2 , Stirrup J 3 , Pontana F 4 , Earls J 5 , Nicol E 1 1 Department of Cardiovascular CT, Royal Brompton Hospital, London, UK 2 Joint department of Physics, The Royal Marsden, London, UK 3 Department of Cardiology, Royal Berkshire Hospital, Reading, UK 4 Department of Cardiovascular Imaging, Lille University Hospital, France 5 George Washington University Hospital, Washington DC, USA Keywords: Cardiovascular CT, Filtered Back Projection, Sinogram, Iterative Reconstruction Authors Rebecca Schofield MBChB [email protected] Laurence King MSc [email protected] Upasana Tayal PhD [email protected] Elly Castellano PhD [email protected] Jim Stirrup MD(Res) [email protected] 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

Upload: others

Post on 22-Jun-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

Image reconstruction: Part 1 – understanding filtered back projection, noise and image acquisition.

Schofield R1, King L2, Tayal U1, Castellano I2, Stirrup J3, Pontana F4, Earls J5, Nicol

E1

1Department of Cardiovascular CT, Royal Brompton Hospital, London, UK

2Joint department of Physics, The Royal Marsden, London, UK

3 Department of Cardiology, Royal Berkshire Hospital, Reading, UK

4 Department of Cardiovascular Imaging, Lille University Hospital, France

5 George Washington University Hospital, Washington DC, USA

Keywords: Cardiovascular CT, Filtered Back Projection, Sinogram, Iterative

Reconstruction

Authors

Rebecca Schofield MBChB [email protected]

Laurence King MSc [email protected]

Upasana Tayal PhD [email protected]

Elly Castellano PhD [email protected]

Jim Stirrup MD(Res) [email protected]

Francois Pontana PhD [email protected]

James Earls MD [email protected]

Ed Nicol* MD [email protected]

*corresponding author: Dr Ed Nicol, Department of Cardiovascular CT, Royal

Brompton Hospital, London, SW3 6NP, UK. Tel: 02073528121, email:

[email protected]

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

16

17

18

19

20

21

22

23

24

25

26

Page 2: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

Word count: 3697

Conflict of Interest: The authors have no conflicts of interest to declare.

27

28

29

30

Page 3: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

Abstract

Image reconstruction is an increasingly complex field in CT. Iterative Reconstruction

(IR) is at present an adjunct to standard Filtered Back Projection (FBP)

reconstruction, but could become a replacement for it. Due to its potential for

scanning at lower radiation doses, IR has received a lot of attention in the medical

literature and all vendors offer commercial solutions. Its use in cardiovascular CT

has been driven in part due to concerns about radiation dose and image quality.

This paper is the first manuscript of a pair. It aims to review the basic principles of

CT scanning, to describe image reconstruction using Filtered Back Projection, and to

identify the physical processes that contribute to image noise which IR may be able

to compensate for. The aim is to enable cardiovascular imagers to understand what

happens to the raw data prior to the reconstruction process so they may have a

better appreciation of the strengths and weaknesses of the various reconstruction

techniques available.

The second manuscript of this pair will discuss the various vendor permutations of IR

in more detail, including the most recent machine learning based offerings, and

critically appraise the current clinical research available on the various IR techniques

used in cardiovascular CT.

31

32

33

34

35

36

37

38

39

40

41

42

43

44

45

46

47

48

49

50

51

52

53

Page 4: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

Introduction

The future of CT will, almost inevitably, involve shorter scanning times and reduced

radiation dose (following the “As low as reasonably achievable (ALARA) consistent

with the imaging task” principle). The radiation dose required to image the patient is

determined by the degree to which image noise can be tolerated by the imager: the

lower the image noise required, the higher the radiation dose that must be used. The

standard method of image reconstruction in CT is known as Filtered Back Projection

(FBP). FBP is computationally efficient, but the filtering step enhances the noise in

the CT image. Iterative reconstruction (IR) techniques have existed since the origins

of CT scanning1 but a lack of computing power initially prevented their use in

standard clinical practice. By virtue of the iterative nature of the reconstruction, IR

images are less noisy than FBP ones. Thus IR potentially enables reduced radiation

dose imaging compared to FBP, although it is not in itself a dose reduction tool.

Improvements in processing power have enabled IR to be re-introduced into clinical

imaging over recent years. The implementation of IR has involved intense

commercial competition and vendors have developed specific proprietary

techniques. This hinders detailed understanding of the modelling processes used

and, as a result, IR techniques are often difficult to understand with confusing

nomenclature adding to the challenge for clinicians.

Undoubtedly cardiovascular CT has benefited greatly from advances in CT

technology such as prospective ECG gating and ECG based tube modulation to

reduce doses, and weight/BMI based reductions in tube voltage. The cardiovascular

CT community is primed to accept the next technological advancements. Although

IR techniques are deemed acceptable in thoracic imaging2 -9, we would advocate

caution before transferring these findings into routine cardiovascular CT clinical

54

55

56

57

58

59

60

61

62

63

64

65

66

67

68

69

70

71

72

73

74

75

76

77

78

Page 5: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

practice, particularly when the imaging tasks are different, such as the assessment

of non-calcified plaque (NCP) burden in coronary arteries. The ability of IR

techniques to detect minor irregularities in small structures (such as coronary

arteries) is an example of an area that requires robust validation and has been

identified as a potential pitfall of several IR techniques10.

To understand IR techniques, it is essential to have a robust understanding of the

basic imaging physics of CT. To highlight what the different IR techniques offer and

‘do’ to the images we first present a basic introduction to the physics of CT imaging,

an explanation of FBP reconstruction techniques, and a review of sources of image

noise.

Physics of CT data acquisition

Within the CT x-ray tube, electrons are emitted from a cathode and accelerated

across a vacuum to a target anode by a high voltage (the tube voltage or kV)

between the cathode and anode. Typically, in CT the tube potential ranges between

70-140kV. The number of electrons accelerated towards the anode determines the

tube current (mA). The electrons strike the anode at the focus and release their

energy as heat and x-ray photons. Due to the various physical processes involved,

and the stochastic nature of the electrons interacting in the target anode, the x-ray

photons are emitted with a range of x-ray energies described by the x-ray spectrum

(Figure 1). The peak energy of the spectrum is determined by the set tube voltage.

Lower energy x-rays are removed from the x-ray beam by physical filters before

reaching the patient; this is desirable as lower energy x-rays have a high probability

79

80

81

82

83

84

85

86

87

88

89

90

91

92

93

94

95

96

97

98

99

100

101

Page 6: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

of attenuation in the patient before reaching the detector and so contribute to patient

dose but not the image.

Pre-patient collimators restrict the shape and direction of the ‘primary’ x-ray beam

emanating from the x-ray focus so that it just covers the active detector array, thus

reducing unnecessary irradiation of areas not being imaged. Post-patient anti-scatter

grids absorb x-rays that have undergone scatter in the patient and have diverged

from the path of the primary x-ray beam, as these x-rays would provide erroneous

positional and attenuation information.

X-ray photons that are transmitted through the patient are absorbed in the CT

detector. All modern CT detectors are made from scintillators (i.e. materials that

release light photons via fluorescence following excitation by x-ray photons). The

current generation of commercial detectors are made from rare earth ceramic

materials. An ideal detector has high atomic density to give a high x-ray photon

interaction probability, is predictable (releases a number of light photons linearly

proportional to the x-ray intensity over a wide range) and is rapidly responsive (has

an instantaneous fluorescent response with no after-glow). The light photons are

then converted to electrical signal in photodiodes. This electrical signal is digitised

and collected to form the raw projection data.

A CT detector array consists of a series of rows of detector elements aligned axially

across the patient. There are typically 16-320 detector rows, and 800-1000 detector

elements in each detector row. During a CT acquisition some or all of the detector

rows may be active. The individual detector elements do not respond in a perfect

way to x-ray photons and individual ones will have slightly different responses to x-

ray signal. This may be corrected by calibration under controlled conditions. In the

102

103

104

105

106

107

108

109

110

111

112

113

114

115

116

117

118

119

120

121

122

123

124

125

Page 7: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

future, photon counting detectors, currently in development, may overcome some of

the shortfalls of use of scintillators.

On modern CT scanners, the x-ray source and detector array are mounted opposite

one other in a rotating gantry. During a CT acquisition, the gantry rotates around the

patient whilst the active detector elements are sampled about 1000 times per

second. Each time the detector is sampled a projection is acquired. The projection is

akin to an x-ray of a thin slab of the patient (Figure 2). In order to reconstruct the CT

images, a set of projections acquired over a little more than half a gantry rotation, or

a full gantry rotation, is needed.

In order to scan the required anatomy, the patient is moved through the gantry.

During axial or ‘step and shoot’ scanning the gantry makes one revolution around the

stationary patient while projection data is acquired, followed by an incremental table

movement, and this repeats until the entire desired anatomy is imaged. During

helical scanning, the scanner continuously acquires projection data as the patient

table moves through the gantry. Helical scanning allows a volume scan to be

acquired much faster. In addition, helical scanning allows axial images to be

reconstructed at overlapping intervals. The scan pitch is defined as the table

movement for a 360-degree rotation of the x-ray source divided by the nominal x-ray

beam width. Typical ranges of clinical pitches are 0.7 - 1.4, although 0.1 – 3.2 are

possible using specialist CT scanners. During dynamic scanning, the scanner

continuously acquires projection data at one patient table position. This allows the

selected patient anatomy, e.g. the heart, to be scanned over a period of time.

126

127

128

129

130

131

132

133

134

135

136

137

138

139

140

141

142

143

144

145

146

147

148

149

Page 8: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

CT Image Reconstruction

Line integrals, attenuation profiles and the sinogram

When an X-ray beam passes through a patient, the intensity of the beam decreases

exponentially with distance travelled due to the x-ray photons interacting with the

tissues in the body (Beer’s Law). On exiting the patient, the remaining x-ray photons

are absorbed by the CT detector array and are converted into an electronic signal.

The attenuation of the x-ray beam along a given ray path is determined by the sum

over all the tissues along that ray path of the product of the length of the ray path

through each tissue (x) and the effective linear attenuation coefficient () of that

tissue averaged over the x-ray spectrum. This is called a line integral. The x-ray

projection (or attenuation profile) is made up of the set of line integrals along all the

ray paths in the x-ray beam. Figure 2 illustrates two x-ray projections of an

anthropomorphic chest phantom at right angles to each other. The resultant

attenuation profiles appear quite different.

The CT raw data for an axial slab through the patient is a set of x-ray projections

acquired at multiple gantry angles for just over half a gantry rotation or for a full

gantry rotation. The CT raw data for an axial slab is thus the full set of line integrals

through the patient for all ray paths in the x-ray beam at all gantry angles. This is

called the Radon Transform.

In modern CT, attenuation profiles are typically acquired a thousand times per 360°

rotation of the gantry. The sinogram is a way of showing the raw acquisition data as

a function of gantry angle in a two-dimensional matrix. Figure 3 illustrates the

sinogram for a single detector row. The x-axis of the sinogram represents the gantry

angle. The y-axis shows the attenuation profile measured at that gantry angle. Each

150

151

152

153

154

155

156

157

158

159

160

161

162

163

164

165

166

167

168

169

170

171

172

173

Page 9: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

point in the attenuation profile is provided by a detector element along the detector

row. For a given coordinate (x,y) in the patient, the contribution of that coordinate to

each attenuation profile will manifest itself as a sinusoidal signal. This behaviour

gives the name to this particular representation of the raw acquisition data. Note that

the sinogram can be thought of as the graphical representation of the Radon

transform. The sinogram for a modern CT scanner with a multi-row detector will be a

three-dimensional matrix, with the z-axis showing the sinogram from each active

detector row.

Filtered Back Projection

The mathematical problem of recreating a 2D image from a series of line integrals is

complex and beyond the scope of this manuscript. For the interested reader, the

mathematical challenges are discussed with relation to tomographic imaging by

Cormack11 and other authors. Instead, we use words and pictures to explain how the

standard CT reconstruction algorithm, Filtered Back Projection, works.

In simple Back Projection (BP), the attenuation profile at each gantry angle is ‘back-

projected’ across image space. The attenuation value in the sinogram is divided by

the number of image pixels along the direction of the projection from x-ray source to

detector, and the average attenuation value thus obtained is assigned to these

pixels; if the pixel values were to be added up along the direction of the projection,

the original attenuation value in the sinogram would be recovered. This is repeated

for each gantry angle. The final back-projected image is then the sum of all the back-

projected attenuation profiles. The simple BP process is graphically represented in

174

175

176

177

178

179

180

181

182

183

184

185

186

187

188

189

190

191

192

193

194

195

196

Page 10: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

Figure 4. Whilst simple BP is conceptually straight forward to understand, it does not

give clear images.

The BP ‘blur’ can be counteracted by applying a spatial frequency filter to the

attenuation profiles prior to back projection. The same steps of image reconstruction

are shown in Figure 5, but this time a simple ramp filter has been applied to the

attenuation profiles making up the sinogram before back-projection. This is Filtered

Back Projection. Applying the mathematical filter to data before back-projection is

computationally easy to do in a step-wise fashion as each new attenuation profile is

acquired. This is the reason that filtered back projection can be used to reconstruct

axial CT images in very short times: large volume scans take tenths of a second to

reconstruct on modern scanners.

The ramp filter is a mathematical function that suppresses low spatial frequency

components of the attenuation profiles. Rapid spatial changes in the attenuation

profile, for example where a boundary between high and low density anatomical

structures exists, are enhanced. This has the effect of suppressing blurring and

enhancing edges in the image, as is observable in Figures 4 and 5. However, this

filter also enhances image noise, which exists in the raw signal primarily at high

spatial frequencies. This gives FBP CT images their characteristically mottled or

speckled appearance.

The ramp filter is mathematically mandated to remove the blur from back-projection,

but it is possible to combine it with different strength filters (kernels) to enhance the

spatial resolution of the final image to varying degrees depending on the application

(e.g. high-resolution vascular kernels, or smoother soft tissue kernels); greater image

noise is the penalty of achieving higher spatial resolution.

197

198

199

200

201

202

203

204

205

206

207

208

209

210

211

212

213

214

215

216

217

218

219

220

Page 11: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

Limitations of Filtered back Projection

In FBP reconstruction, it is assumed that the x-rays travel in straight lines, the x-ray

photons all have the same energy, and the x-ray intensity attenuates exponentially in

the body (Beer’s Law). It is also assumed that the x-ray source is an infinitely small

focal spot, and that the x-ray interactions occur along a line between the focal spot

and the geometric centre of the detector element, rather than continuously

throughout the patient slab being imaged and the whole detector element.

The CT image

The CT image is a two-dimensional attenuation map of the imaged slab through the

patient. Each point in the map is assigned a linear attenuation coefficient relative to

water, which is defined as the Hounsfield Unit (HU). The Hounsfield Unit for air and

water are -1000 and 0 respectively by definition. The image typically consists of a

matrix of 512×512 pixels.

The CT image is not a perfect rendition of the linear attenuation coefficients of the

patient’s anatomy. There is an uncertainty associated with each pixel value arising

from the x-ray detection process, the assumptions made about the interactions of x-

rays with the body, and the image reconstruction process. The uncertainty can

manifest itself as local fluctuations in CT number, referred to as image noise, or

artefacts such as streaks or banding. However manifested, such image noise is

unwelcome because it reduces the quality of the CT image, potentially rendering it

non-diagnostic.

Image noise is usually measured as the standard deviation of CT numbers (in

Hounsfield units) in a region of interest. Noise can also be evaluated by a more

221

222

223

224

225

226

227

228

229

230

231

232

233

234

235

236

237

238

239

240

241

242

243

244

Page 12: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

subjective “quality” or “texture” which is more difficult to quantify but which can be an

issue with higher order iterative reconstruction.

Contributions to image noise from the x-ray detection process

There are several mechanisms that introduce noise within the data acquisition

process. In basic terms image noise can be divided into ‘statistical noise’, ‘structural

noise’ and ‘electronic noise’ which are discussed in detail below.

Statistical Noise

Statistical noise is caused by the fluctuation in the raw x-ray signal due to the

probabilistic nature of x-ray production at the x-ray tube, attenuation of x-rays by the

patient, and absorption in the detector. These variations arise from the “quantum”

nature of x-rays and so statistical noise is also known as ‘quantum noise’. The

statistical variation is inversely proportional to the square root of the number of x-ray

photons. The noise component therefore becomes a smaller proportion of the signal

when the number of x-rays increases. Increasing the tube potential or tube current-

time product (mAs) increases the number of x-ray photons produced and as a result

the statistical noise is reduced.

Structural Noise

Structural noise is typically created by variations in the structure of the detector and

the differing response of individual elements. Structural noise is not usually a major

component of image noise and is minimised by regular calibration of the detector

under controlled conditions.

Scattered X-rays

245

246

247

248

249

250

251

252

253

254

255

256

257

258

259

260

261

262

263

264

265

266

267

Page 13: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

Interactions of the x-ray photons within the body alter both the energy and direction

of travel of photons within the patient. This creates scattered x-rays. Scattered

photons create spurious signal if they are absorbed by the detector, adding to the

image noise. More scatter is created by low-density objects (i.e. those with a low

atomic number, such as fat). The amount of scatter produced increases with the tube

potential. Focused anti-scatter grids at the detector reduce scatter and are routinely

used in CT.

Electronic noise

Electronic circuits in the detector system are the main cause of electronic noise. This

noise carries no diagnostic information and is a constant low-amplitude signal,

independent of the x-ray signal, so that is only a significant contributor to image

noise when the amplitude of the measured x-ray signal in the detector is also low.

This occurs when the number of x-rays reaching the detector is low, for example in

very low dose protocols or when scanning obese patients who provide significant

attenuation of the x-ray beam.

The electronic noise can be minimised by reducing the amount of electronic coupling

between the photodiode and the digital read-out. Detectors featuring a fully

integrated electronic system (integrated detectors) in which the scintillator,

photodiode, pre-amplifier and analogue-to digital converter are all in the same chip

have been shown to have significantly reduced noise and superior image quality12.

The main constraints to universal use of single chip systems are manufacturing

ones.

To demonstrate these types of noise visually, Figures 6a – 6d were created from

simulated raw projection data from an axial CT image of an anthropomorphic chest

268

269

270

271

272

273

274

275

276

277

278

279

280

281

282

283

284

285

286

287

288

289

290

291

Page 14: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

phantom. The data was manipulated before filtered back projection to demonstrate

the presence of different types of noise.

Contributions to image noise from the image reconstruction process

In FBP, the projection data is filtered using convolution techniques. These are

computationally efficient, but convolution enhances the differences between

neighbouring data points in the attenuation profile, and hence magnifies noise in the

data. This noise is propagated through the reconstruction process into the final

image. Image noise and image texture are both highly dependent on the

reconstruction kernel selected: smooth kernels suppress noise at the expense of

spatial resolution, and sharp kernels enhance small and linear features such as

coronary arteries or stents but also increase image noise.

Contributions to image noise from beam hardening and photon starvation

In order to reconstruct axial images using FBP, it is assumed that the x-ray beam

has a constant effective photon energy along the ray path through the patient. In

reality the x-ray beam contains x-rays of many different energies (i.e. a spectrum of

x-ray energies). The attenuation in the patient is preferentially greater for lower

energy x-rays. The x-ray beam spectrum further along the ray path therefore

preferentially contains higher energy x-rays with greater penetration (a ‘harder’ x-ray

beam), resulting in lower attenuation and lower measured CT numbers along the

remaining beam path. Attempts to overcome this problem include physical pre-

filtering of the x-ray beam to obtain a pre-hardened beam, or additional post-

processing of the FBP image. In cardiovascular CT, beam hardening causes

292

293

294

295

296

297

298

299

300

301

302

303

304

305

306

307

308

309

310

311

312

313

314

315

Page 15: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

artefacts adjacent to densely calcified atherosclerotic plaques, prosthetic heart

valves, intravascular wires, implanted devices and intravascular stents. The size and

shape of the dense object becomes distorted as regions of artificially low attenuation

that can be misinterpreted as low attenuation plaques or intravascular thrombus can

be created. Beam hardening accounts for much of the false positive findings in CT

coronary angiography.

When x-rays pass through an object of high attenuation, i.e. large bones or metallic

devices such as permanent pacemakers or implantable cardiac defibrillators, the

resultant x-ray signal at the detector can become so low that there is essentially no

information to contribute to the image from that projection; this photon starvation can

result in a streak artefact with complete loss of signal downstream of the increased

attenuation structure making images difficult to interpret. This can account for false

positive perfusion defects in the myocardium adjacent to the dense thoracic vertebral

bodies, decreased attenuation within a coronary stent, or even HU changes adjacent

to high concentration of iodine enhanced blood pool in the great vessels or cardiac

chambers.

Challenges of FBP and how we can address them

The disadvantages of FBP are that it assumes that the sinogram represents a

perfect representation of the object being imaged, and the filtering step amplifies

noise in the acquired signal, producing the characteristic mottled axial images that

we are used to. Statistical noise aside, any other aberrations in the acquisition data

can produce dramatic image artefacts, such as streak or banding artefacts.

However, although these FBP images can be quite grainy and noisy, radiologists

316

317

318

319

320

321

322

323

324

325

326

327

328

329

330

331

332

333

334

335

336

337

338

339

Page 16: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

who have ‘grown up’ looking at them often feel more comfortable reviewing these

images rather than IR ones even as the concern for patient radiation dose has driven

the acceptable signal-to-noise ratio progressively lower and lower. In part two of this

paper, we will examine Iterative Reconstruction techniques, which provide the

opportunity to reduce image noise for a given radiation dose by manipulating the raw

acquisition data and / or processing the reconstructed image and thus increase the

acceptability of low dose CT images.

340

341

342

343

344

345

346

347

348

349

350

351

352

353

354

355

356

357

358

359

360

361

Page 17: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

References

1. Hounsfield, GN. Computerized transverse axial scaning (tomography). Part I.

description of the system. Br J Radiol 1973; 46:1016-22.

2. Pontana F, Duhamel A, Pagniez J, Flohr T, Faivre JB, Hachulla AL, Rmey J,

Remy-Jardin M. Chest computed tomography using iterative reconstruction vs

filtered back projection (part2): image quality of low-dose CT examinations in

80 patients. Eur Radiol 2011;21:636-643.

3. Prakash P, Kalra M, Digumarthy S, Hsieh J, Pien H, Singh S, Gilman M,

Shepard JA. Radiation dose reduction with chest computed tomography using

adaptive statistical iterative reconstruction technique: initial experience. J

Comp Assit Tomography 2010;34(1):40-45.

4. Kim Y, Yoon K, Bo LE, Seok JL, Yon JR, Jin HL, Chang JH. Ultra-low dose

CT of the thorax using iterative reconstruction: evaluation of image quality and

radiation dose reduction. AJR 2015;204(6):1197-1202.

5. Baumueller S, Winklehner A, Karlo C, et al. Low-dose CT of the lung:

potential value of iterative reconstructions. Eur Radiol 2012; 22:2597–2606

6. Lee SW, Kim Y, Shim SS, et al. Image quality assessment of ultra low-dose

chest CT using sinogram-affirmed iterative reconstruction. Eur Radiol 2014;

24:817–826

7. Prakash P, Kalra MK, Ackman JB, et al. Diffuse lung disease: CT of the chest

with adaptive statistical iterative reconstruction technique. Radiology 2010;

256:261–269

8. Pontana F, Paigniez J, Duhamel A, Flohr T, Faivre JB, Murphy C, Remy J,

Remy-Jardin M. Reduced-Dose Low-Voltage chest CT angiopraphy with

362

363

364

365

366

367

368

369

370

371

372

373

374

375

376

377

378

379

380

381

382

383

384

385

Page 18: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

sinogram-affirmed iterative reconstruction versus standard dose filtered back

projection. Radiology 2013;267(2):609-618.

9. Pontana F, Moureau D, Schmidt B, Duhamel A, Faivre JB, Yasunga K, Remy

J, Remy-Jardin M. CT pulmonary angiogram with 60% dose reduction:

influence of iterative reconstructions on image quality. Diagnostic and

interventional imaging 2015;96:487-493.

10.Hou Y, Liu X, Xv S, Guo W, Guo Q. Comparisons of image quality and

radiation dose between iterative reconstruction and filtered back projection

reconstruction algorithms in 256- MDCT coronary angiography. AJR Am J

Roentgenol. 2012;199(3):588e594.

11.Cormack, AM. Reconstruction of Densities from the projections, with

applications in radiological physics, Phys. Med. Biol. 1973;18(2):195-207.

12.Duan X, Wang J, Leng S, Schmidt B, Allmendinger T, Grant K, Flohr T,

McCollough CH.Electronic Noise in CT Detectors: Impact on Image and Noise

Artefacts Duan X et al AJR 2013);201(4)626-632.

386

387

388

389

390

391

392

393

394

395

396

397

398

399

400

401

402

403

404

405

406

407

408

409

410

411

412

Page 19: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

Figure 1: Filtered X-ray Bremsstrahlung spectra produced by X-ray tube potentials of 50 and 100 kV

keV= kilo-electron volts, energy of photons.

413

414

415416

417

418

419

420

421

422

423

424

425

426

427

428

429

430

Page 20: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

Figure 2: the charts show the integrated attenuation profile (arbitrary units) of an

anthropomorphic chest phantom against detector position, for an AP and lateral

projection respectively. The attenuation profiles appear to be quite different in these

directions, although some rotational symmetry exists. Ray paths (a) and (b) for each

projection correspond to a low and high attenuation signal respectively.

ba

b

a

431

432

433

434

435

436

437

438

439

440

441

442

443

444

445

446

447

448

449

450

451

452

453

454

455

456

457

458

459

460

461

Page 21: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

Figure 3: Two example CT slices and their simulated sinograms over a 180°

acquisition are given. The first image is a cylindrical water phantom with simulated

high and low-attenuation objects: a uniform cylindrical phantom looks identical from

all projection angles but the simulated objects in this example move position relative

to one another, giving a sinogram with sinusoidal traces. The second image is an

anthropomorphic phantom with clinical detail, with more complex attenuation profiles

and thus a more complex sinogram.

Attenuation

Projection

462

463

464

465

466

467

468

469

470

471

472

473

474

475

476

477

478

479

480

481

482

483

484

485

486

487

488

489

490

491

Page 22: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

Figure 4: Simple Back projection (a) a slice of an anthropomorphic chest phantom

is shown, with a single attenuation profile in the AP direction. When this single profile

is projected back across reconstructed image space, the projection image (b) is

formed. Image (c) is a back-projection of two perpendicular projection profiles; image

(d) is formed from back-projection of four profiles; (e) from eight profiles; and (f) from

720 profiles. At this point the image is starting to resemble the subject that was

imaged, but with considerable blurring. The simple back-projection process

inherently produces blurred images.

fed

cba

492

493

494

495

496

497

498

499

500

501

502

503

504

505

506

507

508

509

510

511

512

513

514

515

516

517

518

519

520

Page 23: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

Figure 5: Filtered back projection - taking the same attenuation profiles from an

imaged slice of an anthropomorphic chest phantom (a), the corresponding filtered

back projected images to the simple back projected images in Figure 4 (b-f) are

shown here, created from (b) one, (c) two, (d) four, (e) eight and (f) 720 filtered back-

projected attenuation profiles. The filtering process recovers spatial resolution in the

final image.

fed

cba

521

522

523

524

525

526

527

528

529

530

531

532

533

534

535

536

537

538

539

540

541

542

543

544

545

546

547

Page 24: spiral.imperial.ac.uk  · Web view2020. 7. 1. · The number of electrons accelerated towards the anode determines the tube current (mA). The electrons strike the anode at the focus

Figure 6: Filtered Back Projection (FBP) reconstruction (a) was recreated from raw

data generated from a CT image of an anthropomorphic phantom with no added

noise. Reconstruction (b) was carried out after addition of Gaussian noise to the raw

data, simulating increased statistical noise in the acquisition. Figure (c) was created

after altering a single data point in the raw data to a value of zero, representing the

case where a single detector element has failed for a single projection during

acquisition, resulting in a line artefact (high-lighted by white arrow). Figure (d) was

created after changing just 5% of the raw data points to minimum or maximum

values (“salt & pepper noise”), indicating a large degree of structural noise.

a b

c d

548

549

550

551

552

553

554

555

556

557

558

559