Welcome

Chris Stamper

Technical Field Engineer

www.pace

ricd.c

om

Page 2: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

2

The Fundamentals of Cardiac Devices Module 2

www.pace

ricd.c

om

Page 3: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

3

Objectives

After completing this introduction to Cardiac Devices, you will be able to:

• Identify the various cardiac device systems

• Identify the basic indications for implantation for each device

• Recognize each device type on a chest x-ray

www.pace

ricd.c

om

Page 4: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

4

Cardiac Devices

• Designed to:

– Restore or maintain a rhythm and rate sufficient to meet metabolic needs

– Provide diagnostic information about

• Device operation

• The patient

www.pace

ricd.c

om

Page 5: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

5

Cardiac Devices

• Pacemakers or Implantable Pulse Generators (IPG)

– Provide a rate to support metabolic needs

– Provide various diagnostics

– Single and dual chamber

– About 8-10 years longevity

– Some of the newer pacemakers include therapies which pace-terminate AT/AF

www.pace

ricd.c

om

Page 6: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

6

Cardiac Devices

• Implantable Cardioverter Defibrillators (ICDs)

– Restore sinus rhythm in the presence of tachycardia

• Defibrillate

• Cardiovert

• Anti-Tachy Pace (ATP)

– Provide a rate to support metabolic needs

• Includes single or dual chamber pacing

– Provide various diagnostics

– About 6-9 years longevity

– There are also devices designed to terminate AF via cardioversion/ATP in addition to standard ICD therapieswww.pa

ceric

d.com

Page 7: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

7

Cardiac Devices

• Cardiac Resynchronization Therapy (CRT)

– IPG with CRT, or an IPG + ICD with CRT

– Restore ventricular synchrony

• Uses a specially designed lead placed usually on the posterior-lateral wall of the LV via the Coronary Sinus circulation

• LV epicardial lead placement is an option

• Provides RV and LV synchronous pacing

– May restore rhythms in presence of lethal tachycardia

• CRT pacing +ICD (High-Power CRT)

– Provides a rate to support metabolic needs

• CRT pacing only (Low-Power CRT)

– Provides various diagnostics

www.pace

ricd.c

om

Page 8: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

8

Cardiac Devices

• Implantable Loop Recorders (ILR)

– Provides rate-based monitoring

• Fast rates

• Slow rates

– Provides EGM during patient triggered events

www.pace

ricd.c

om

Page 9: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

9

IndicationsPacemakers

• The AHA and ACC have defined the indications for pacing by the underlying arrhythmia

• Very detailed, but to simplify:

– Symptomatic bradycardia refractory to any treatment

– Typical diagnoses:

• Sinus Node Disease (SND), or Sick Sinus Syndrome

• Complete Heart Block

• Chronotropic Incompetence

• Vaso-vagal syncope

• Carotid sinus hypersensitivity

– Usually excludes “low grade” blocks (Mobitz I and 1st degree)www.pa

ceric

d.com

Page 10: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

10

IndicationsDefibrillators

• Primary vs. Secondary Prevention for SCA

– Primary

• Patients who have experienced a previous SCA or ventricular arrhythmia

• Studies such as AVID1, CIDS2, CASH3 support the use of ICDs in this population

– Secondary

• Patients who have not previously experienced SCA/VA, but are at risk

• Studies, such as MADIT II4 and SCD-HeFT5, have demonstrated the use of ICDs in these patients

www.pace

ricd.c

om

Page 11: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

11

IndicationsDefibrillators (cont.)

Well defined by Heart Rhythm Society (HRS) and include:

• Cardiac Arrest

– Due to VT or VF, not transient or reversible

– Spontaneous sustained VT with structural heart disease

• Syncope of undetermined origin with:

– Sustained VT or VF induced during EP

• Nonsustained VT with:

– Coronary disease or prior MI and LV Dysfunction

– Inducible VF or sustained VT (non-suppressible by antiarrhythmic drugs)

• Spontaneous sustained VT

– Not amenable to other treatmentswww.pace

ricd.c

om

Page 12: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

12

IndicationsDefibrillators (cont.)

• ICD Class I Recommendation

• Patients at least 40 days post-MI

• LVEF ≤ 30 – 40%

• NYHA class II or III

• Non-ischemic patients

• LVEF ≤ 30 – 35%

• NYHA class II or III

• Patients at risk of SCA due to genetic disorders

• Long QT syndrome

• Brugada syndrome

• Hypertrophic cardiomyopathy (HCM)

• Arrhythmogenic right ventricular dysplasia (ARVD)

Note: This list includes the current major indications for an ICDwww.pa

ceric

d.com

Page 13: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

13

IndicationsCardiac Resynchronization Therapy

• NYHA Class III or IV heart failure

• On optimal medical therapy

• QRS > 120 ms wide

– Or Echo evidence of ventricular dyssynchrony

• Ejection Fraction of < 30%

• Who are candidates for an ICD with CRT, and who for an IPG with CRT?

– Most get an ICD with CRT (also called “High-Power CRT”) because of the risk of SCD in patients with LV dysfunctionwww.pa

ceric

d.com

Page 14: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

14

IndicationsLoop Recorders

• Transient, infrequent but recurrent syncope

www.pace

ricd.c

om

Page 15: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

15

Recognizing SystemsPacemakers

Dual Chamber Pacemaker

RV Lead at the Apex

RA Lead in Appendage

www.pace

ricd.c

om

Page 16: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

16

Recognizing SystemsICD or High Power

Right Atrial Lead

Approximate position outlined

Right Ventricular Lead

With RV and SVC coils

www.pace

ricd.c

om

Page 17: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

17

Recognizing SystemsCRT Low-Power

Right Atrial Lead

Left Ventricular Lead

Placed on the surface of the LV via the Coronary Sinus

Right Ventricular Lead

www.pace

ricd.c

om

Page 18: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

18

Recognizing SystemsCRT High-Power

Right Atrial Lead

Left Ventricular Lead

Placed on the surface of the LV via the Coronary Sinus

Right Ventricular Lead

Note the 2 high voltage defib coils

Surface ECG leads www.pace

ricd.c

om

Page 19: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

19

Pacing Therapy

• Senses underlying heart rate

– Delivers low energy electrical pulses when rate falls below programmed limit

• Low energy pulses capture and depolarize the heart muscle causing it to contract

– Dual Chamber IPGs provide A-V synchrony

www.pace

ricd.c

om

Page 20: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

20

• Atrium and Ventricle

– Sensing

– Pacing

• Ventricle

– Antitachycardia pacing (ATP)

ICD Pacing Therapy

Atrial Lead

Ventricular Lead

www.pace

ricd.c

om

Page 21: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

21

High-Voltage Therapy

Cardioversion and Defibrillation are delivered in biphasic waves

For example: First, from the SVC coil and the Can to the RV coil, and then reverse.

The device must detect – charge – confirm – and deliver the shock.

Fast, accurate detection, and fast charge times are criticalwww.pa

ceric

d.com

Page 22: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

22

ICD Therapy

Brady and Anti-Tachy pacing (ATP)

• The ICD lead is designed to carry both high voltage and pacing therapies

– Brady

– ATP

– Cardioversion

– Defibrillation

www.pace

ricd.c

om

Page 23: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

23

CRT Therapy

• Video of Dyssynchronous Heart • Video of CRT in Action

By pacing from both the right and left ventricles, CRT may improve EF and reduce patient symptoms.

(Click above to play video) (Click above to play video)

www.pace

ricd.c

om

Page 24: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

24

Implant Evolution

Pacemakers—Yesterday

• First Implants in early 1960s

• Single Chamber, non programmable

• About 2 year longevity

• About 200 cc

• Abdominal implants, sternotomy for epicardial leads

www.pace

ricd.c

om

Page 25: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

25

Implant Evolution

Pacemakers—Today

• Pectoral implants

• 6-7 F transvenous lead placement

– Outpatient/overnight stay

• < 30 cc

• 8-10 years longevity

• Dual chamber, multi-programmable

• Advanced diagnostics and trending informationwww.pa

ceric

d.com

Page 26: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

26

Implant Evolution

ICD Past: ~10-12 years ago ICDs

• Required major surgery

– Abdominal implants

– Median sternotomy to suture defib patches on heart

– Length of hospital stay > 1 week

• Nonprogrammable

• High-energy shock only

• Indicated for patients who survived cardiac arrest - twice

• 1 ½ year longevity

• 209 cc

• < 1,000 implants/year

www.pace

ricd.c

om

Page 27: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

27

Implant Evolution

ICDs Today

• Similar to a pacemaker implant

• Transvenous, single incision

– Pectoral implant

– Overnight stay

• Local anesthesia, conscious sedation

• Programmable therapy options

• Single, dual and triple chamber (CRT)

• Up to 9 years longevity

• About 35 cc

• > ~100,000 implants/yearwww.pa

ceric

d.com

Page 28: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

28

Status Check

• Given the following patient

– 67 YO male patient, had an anterior MI 2 years ago

– EF 20%

– History includes

• Tobacco 2 packs/day • Hyperlipidemia

• CAD

– Medications

• Vasopril (ACE Inhibitor) • Lipitor (Statin)

• Coreg (ARB) • Aldosterone (ADH)

• Atenolol (Beta Blocker)

– C/O

• Severe shortness of breath at rest, fatigue, unable to perform ADL, vertigo, rare syncope

www.pace

ricd.c

om

Page 29: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

29

Status Check

• Patient’s ECG:

NSR

Rate 78

LBB

What device system (if any) is the patient likely to get?Click for Answer

High-Power CRT

Wide QRS, EF < 30% on max

medical therapy and is S/P MI. = ↑ risk of

SCDwww.pace

ricd.c

om

Page 30: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

30

Status Check

• Could this be the previous patient’s chest x-ray?

– Is it CRT?

– Is it High-Power?

Click for Answer

Yes, we see an RV lead with RV and SVC coils, an atrial pacing lead, and an LV pacing lead

www.pace

ricd.c

om

Page 31: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

31

Status Check

• Consider the following patient

– Elderly female patient

– Medical history includes

• Ex-smoker – quit 10 years ago

• Mild exertional angina

– Cardiac cath shows mild disease right coronary artery

• 3 adult children

– Medications

• Nitroglycerine PRN • Aspirin 81 mg. 1/day

– C/O

• Fatigue, unable to perform normal activitieswww.pa

ceric

d.com

Page 32: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

32

Status Check

• Patient’s stress test indicates:

– Stopped after 4 minutes for fatigue

– ECG immediately after stress test: Click for Answer

Likely a pacemaker, as the patient has chronotropic incompetence – her heart rate does not increase with exercise.

What device system would you recommend?

NSR, Rate 57

www.pace

ricd.c

om

Page 33: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

3333

Basic Concepts—Electricity and Pacemakers

Module 3

www.pace

ricd.c

om

Page 34: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

3434

Objectives

Upon completion you will be able to:

• Describe the relationship between voltage, current, and resistance

• Describe the clinical significance of alterations in voltage, current, and resistance

www.pace

ricd.c

om

Page 35: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

3535

Characteristics of an electrical circuit:Including a pacemaker circuit

• Voltage

• Current

• Impedance

www.pace

ricd.c

om

Page 36: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

36

Voltage

• Voltage is the force, or “push,” that causes electrons to move through a circuit

• In a pacing system, voltage is:

– Measured in volts (V)

– Represented by the letter “V”

– Provided by the pacemaker battery

– Often referred to as amplitude or pulse amplitude

www.pace

ricd.c

om

Page 37: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

37

Current

• The flow of electrons in a completed circuit

• In a pacing system, current is:

– Measured in milliamps (mA)

– Represented by the letter “I”

– Determined by the amount of electrons that move through a circuit

www.pace

ricd.c

om

Page 38: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

38

Impedance

• The opposition to current flow

• In a pacing system, impedance is:

– Measured in ohms (Ω)

– Represented by the letter “R”

– The measurement of the sum of all resistance to the flow of current

www.pace

ricd.c

om

Page 39: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

39

Voltage, Current, and Impedance are Interdependent

• The interrelationship of the three components is analogous to the flow of water through a hose

– Voltage represents the force with which . . .

– Current (water) is delivered through . . .

– A hose, where each component represents the total impedance:

• The nozzle, representing the electrode

• The tubing, representing the lead wire

www.pace

ricd.c

om

Page 40: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

4040

Voltage, Current, and ImpedanceRecap

• Voltage: The force moving the current (V)

– In pacemakers it is a function of the battery chemistry

• Current: The actual continuing volume of flow of electricity (I)

– This flow of electrons causes the myocardial cells to depolarize (to “beat”)

• Impedance: The sum of all resistance to current flow (R or Ω or sometimes Z) – Impedance is a function of the characteristics of the conductor (wire),

the electrode (tip), and the myocardium

www.pace

ricd.c

om

Page 41: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

41

Voltage and Current FlowElectrical Analogies

Spigot (voltage) turned up, lots of water flows (high current drain)

Spigot (voltage) turned low, little flow (low current drain)

Water pressure in system is analogous to voltage –

providing the force to move the current

www.pace

ricd.c

om

Page 42: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

42

Resistance and Current FlowElectrical Analogies

• Normal resistance – friction caused by the hose and nozzle

More water discharges, but is all of it going to the nozzle?

• High resistance – a knot results in low total current flow

• Low resistance – leaks in the hose reduce the resistance

www.pace

ricd.c

om

Page 43: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

43

Ohm’s Law

• Describes the relationship between voltage, current, and resistance

• V = I X R

• I = V / R

• R = V / I

VV

II RR

V

R

I

V

R

I

R

V

I =

=

=X

www.pace

ricd.c

om

Page 44: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

44

Ohm’s law tells us:

1. If the impedance remains constant, and the voltage decreases, the current decreases

2. If the voltage is constant, and the impedance decreases, the current increases

So What?

www.pace

ricd.c

om

Page 45: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

4545

Status Check

• Start with:

– Voltage = 5 V

– Impedance = 500 Ω

– Current = 10 mA

• Solve for Current (I):

– I = V/R

– I = 5 V ÷ 500 Ω = 0.010 Amps

– Current is 10 mA

• Reduce the voltage to 2.5 V

– Voltage = 5 V

– Impedance = 500 Ω

– Current = ?

• Is the current increased/ decreased or unchanged?

– I = V/R

– V = 2.5 V ÷ 500 Ω =

0.005 Amps or 5 mA

• The current is reduced

What happens to current if the voltage is reduced but the impedance is unchanged?

www.pace

ricd.c

om

Page 46: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

4646

Status Check

• Start with:

– Voltage = 5 V

– Impedance = 500 Ω

– Current = 10 mA

• Solve for Current (I):

– I = V/R

– I = 5 V ÷ 500 Ω = 0.010 Amps

– Current is 10 mA

• Reduce impedance to 250 Ω– Voltage = 5 V

– Impedance = 250 Ω

– Current = ?

• Is the current increased/ decreased or unchanged?

– I = V/R

– V = 2.5 V ÷ 250 Ω =

0.02 Amps or 20 mA

• The current is increased

What happens to current if the impedance is reduced but the voltage is unchanged?

www.pace

ricd.c

om

Page 47: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

4747

Other terms

• Cathode: A negatively charged electrode

– For example, the electrode on the tip of a pacing lead

• Anode: A positively charged electrode

– Examples:• The “ring” electrode on a bipolar lead

• The IPG case on a unipolar system

– More on this later (see: Pacemaker Basics)

Anode

Cathodewww.pace

ricd.c

om

Page 48: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

4848

Battery BasicsSo where does the current come from?

• A battery produces electricity as a result of a chemical reaction. In its simplest form, a battery consists of:

– A negative electrode (anode)

– An electrolyte, (which conducts ions)

– A separator, (also an ion conductor) and

– A positive electrode (cathode)

Negative terminal

Anode

Separator

Cathode

Positive terminal

www.pace

ricd.c

om

Page 49: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

4949

Applying Electrical Concepts to Pacemakers

Module 4

www.pace

ricd.c

om

Page 50: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

5050

Objectives

• Upon completion you will be able to:

– Recognize a high impedance condition

– Recognize a low impedance condition

– Recognize capture threshold

– Determine which sensitivity value is more (or less) sensitive

www.pace

ricd.c

om

Page 51: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

5151

Electrical Information

• Why is this electrical information relevant?

• A pacemaker is implanted to:

– Provide a heart rate to meet metabolic needs

• In order to pace the heart, it must capture the myocardium

• In order to pace the heart, it must know when to pace, i.e., it must be able to sense

• A pacemaker requires an intact electrical circuit

www.pace

ricd.c

om

Page 52: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

52

Ohm’s LawRelevance to Pacemaker Patients

• High impedance conditions reduce battery current drain

– Can increase pacemaker battery longevity

– Why?

• R = V/I If “R” increases and “V” remains the same, then “I” must decrease

• Low impedance conditions increase battery current drain

– Can decrease pacemaker battery longevity

– Why?

• R = V/I If “R” decreases and “V” remains the same, then “I” must increase

www.pace

ricd.c

om

Page 53: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

5353

The Effect of Lead Performance on Myocardial Capture

What would you expect to happen if a lead was partially fractured?

• Impedance (or Resistance) would rise

• Current would decrease and battery energy conserved

- but -

Could you guarantee that enough current (I) can flow through this fractured lead so that each time the pacemaker fired the myocardium would beat?

www.pace

ricd.c

om

Page 54: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

54

Lead wire fracture

Increased resistance

High Impedance ConditionsA Fractured Conductor

• A fractured wire can cause Impedance values to rise

– Current flow from the battery may be too low to be effective

• Impedance values may exceed 3,000 Ω

Other reason for high impedance: Lead not seated properly in pacemaker header.www.pa

ceric

d.com

Page 55: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

55

Lead Impedance Values Change as a Result of:

• Wire fractures

• Insulation breaks

Typically, normal impedance reading values range from 300 to 1,000 Ω

– Some leads are high impedance by design. These leads will normally show impedance reading values greater than 1,000 ohms

• Medtronic High Impedance leads are:

– CapSure® Z

– CapSure® Z Novuswww.pace

ricd.c

om

Page 56: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

56

Low Impedance Conditions

• Insulation breaks expose the lead wire to the following

– Body fluids, which have a low resistance, or

– Another lead wire (in a bipolar lead)

• Insulation break that exposes a conductor causes the following

– Impedance values to fall

– Current to drain through the insulation break into the body, or into the other wire

– Potential for loss of capture

– More rapid battery depletion

Current will follow the path of LEAST resistancewww.pa

ceric

d.com

Page 57: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

57

Capture Threshold

• The minimum electrical stimulus needed to consistently capture the heart outside of the heart’s own refractory period

Ventricular pacemaker 60 ppm

Capture Non-Capture

www.pace

ricd.c

om

Page 58: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

58

Effect of Lead Design on Capture

• Lead maturation

– Fibrotic “capsule” develops around the electrode following lead implantation

– May gradually raise threshold

– Usually no measurable effect on impedance

www.pace

ricd.c

om

Page 59: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

59

Steroid Eluting Leads

• Steroid eluting leads reduce the inflammatory process

– Exhibit little to no acute stimulation threshold peaking

– Leads maintain low chronic thresholds

Silicone rubber plugcontaining steroid

Tines forstablefixation

Porous, platinized tipfor steroid elution

www.pace

ricd.c

om

Page 60: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

60

Effect of Steroid on Stimulation Thresholds

Pulse Width = 0.5 msec

03 6

Implant Time (Weeks)

Textured Metal Electrode

Smooth Metal Electrode

1

2

3

4

5

Steroid-Eluting Electrode

0 1 2 4 5 7 8 9 10 11 12

Vol

ts

References: Pacing Reference Guide, Bakken Education Center, 1995, UC199601047aEN. Cardiac Pacing, 2nd Edition, Edited by Kenneth A. Ellenbogen. 1996.

www.pace

ricd.c

om

Page 61: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

61

Myocardial Capture

• Capture is a function of:

– Amplitude—the strength of the impulse expressed in volts

• The amplitude of the impulse must be large enough to cause depolarization (i.e., to “capture” the heart)

• The amplitude of the impulse must be sufficient to provide an appropriate pacing safety margin

– Pulse width—the duration of the current flow expressed in milliseconds

• The pulse width must be long enough for depolarization to disperse to the surrounding tissue

www.pace

ricd.c

om

Page 62: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

62

Comparison5.0 Volt Amplitude at Different Pulse Widths

Amplitude5.0 V

0.25 ms 1.0 ms0.5 mswww.pace

ricd.c

om

Page 63: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

63

The Strength-Duration Curve

• The strength-duration curve illustrates the relationship of amplitude and pulse width

– Any combination of pulse width and voltage, on or above the curve, will result in capture

Pulse Width

.50

1.0

1.5

2.0

.25V

olts

0.5 1.0 1.5

Capture

No Capture

Rheobase

Chronaxie

www.pace

ricd.c

om

Page 64: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

64

DurationPulse Width (ms)

Clinical Utility of the Strength-Duration Curve

• By accurately determining capture threshold, we can assure adequate safety margins because:

– Thresholds may differ in acute or chronic pacing systems

– Thresholds fluctuate slightly daily

– Thresholds can change due to metabolic conditions or medications

0.5 1.0 1.5

.50

1.0

1.5

2.0

.25

Sti

mu

lati

on

Th

resh

old

(V

olt

s)

X Programmed Output

www.pace

ricd.c

om

Page 65: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

65

Programming Outputs

• Primary goal: Ensure patient safety and appropriate device performance

• Secondary goal: Extend the service life of the battery

– Typically program amplitude to < 2.5 V, but always maintain adequate safety margins

• A common output value might be 2.0 V at 0.4 ms

– Amplitude values greater than the cell capacity of the pacemaker battery (usually about 2.8 V) require a voltage multiplier, resulting in markedly decreased battery longevity

www.pace

ricd.c

om

Page 66: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

6666

Pacemaker Sensing

• Refers to the ability of the pacemaker to “see” signals

– Expressed in millivolts (mV)

• The millivolts (mV) refers to the size of the signal the pacemaker is able to “see”

0.5 mV signal

2.0 mV signalwww.pace

ricd.c

om

Page 67: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

67

SensitivityThe Value Programmed into the IPG

Time

5.0 mV

2.5 mV

1.25 mV

www.pace

ricd.c

om

Page 68: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

68

Time

5.0 mV

2.5 mV

1.25 mV

5 mV sensitivity

At this value the pacemaker will not see the 3.0 mV signal

SensitivityThe Value Programmed into the IPG

www.pace

ricd.c

om

Page 69: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

69

SensitivityThe Value Programmed into the IPG

At this value, the pacemaker can see both the 3.0 mV and the 1.30 mV signal. So, is “more sensitive” better, because the pacemaker sees smaller signals?

1.25 mV Sensitivity

Time

5.0 mV

2.5 mV

1.25 mV

But what about this?

www.pace

ricd.c

om

Page 70: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

70

Sensing Amplifiers/Filters

• Accurate sensing requires that extraneous signals are filtered out

– Because whatever a pacemaker senses is by definition a P- or an R-wave

– Sensing amplifiers use filters that allow appropriate sensing of P- and R-waves, and reject inappropriate signals

• Unwanted signals most commonly sensed are:

– T-waves (which the pacemaker defines as an R-wave)

– Far-field events (R-waves sensed by the atrial channel, which the pacemaker thinks are P-waves)

– Skeletal muscle myopotentials (e.g., from the pectoral muscle, which the pacemaker may think are either P- or R-waves)

– Signals from the pacemaker (e.g., a ventricular pacing spike sensed on the atrial channel “crosstalk”)www.pa

ceric

d.com

Page 71: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

71

Sensing Accuracy

• Affected by:

– Pacemaker circuit (lead) integrity

• Insulation break

• Wire fracture

– The characteristics of the electrode

– Electrode placement within the heart

– The sensing amplifiers of the pacemaker

– Lead polarity (unipolar vs. bipolar)

– The electrophysiological properties of the myocardium

– EMI – Electromagnetic Interferencewww.pace

ricd.c

om

Page 72: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

72

Lead Conductor Coil IntegrityAffect on Sensing

• Undersensing occurs when the cardiac signal is unable to get back to the pacemaker

– Intrinsic signals cannot cross the wire fracture

• Oversensing occurs when the severed ends of the wire intermittently make contact

– Creates signals interpreted by the pacemaker as P- or R-waves

www.pace

ricd.c

om

Page 73: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

73

Lead Insulation IntegrityAffect on Sensing

• Undersensing occurs when inner and outer conductor coils are in continuous contact

– Signals from intrinsic beats are reduced at the sense amplifier, and amplitude no longer meets the programmed sensing value

• Oversensing occurs when inner and outer conductor coils make intermittent contact

– Signals are incorrectly interpreted as P- or R-waves

www.pace

ricd.c

om

Page 74: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

74

Unipolar Pacemaker

• Where is the sensing circuit?

_Cathode

Anode

Lead tip to canThis can produce a large potential difference (signal) because the cathode and anode are far apart

Click for Answer

www.pace

ricd.c

om

Page 75: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

75

Bipolar Pacemaker

• Where is the sensing circuit?

Anode and Cathode

Click for Answer

Lead tip to ring on the leadThis usually produces a smaller potential difference due to the short inter-electrode distance

• But, electrical signals from outside the heart (such as myopotentials) are less likely to be sensed

www.pace

ricd.c

om

Page 76: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

7676

Cardiac Conduction and Device Sensing

By now we should be familiar with the surface ECG and its relationship to cardiac conduction. But, how does this relate to pacemaker sensing?www.pa

ceric

d.com

Page 77: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

7777

Vectors and Gradients

Sense

The wave of depolarization produced by normal conduction creates a gradient across the cathode and anode. This changing polarity creates the signal.

Once this signal exceeds the programmed sensitivity – it is

sensed by the device.

Click for More

2.5 mV

www.pace

ricd.c

om

Page 78: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

7878

Changing the Vector

Sense

2.5 mV

A PVC occurs, which is conducted abnormally. Since the vector relative to the lead has changed, what effect

might this have on sensing?

Click for More

In this case, the wave of depolarization strikes the anode and cathode almost simultaneously. This

will create a smaller gradient and thus, a smaller signal.

www.pace

ricd.c

om

Page 79: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

7979

Putting It All Together

• Appropriate output programming can improve device longevity

– But, do not compromise patient safety!

• Lead design can improve device longevity via

– Steroid eluting leads

• Can help keep chronic pacing thresholds low by reducing inflammation and scarring

– High Impedance leads

• Medtronic CapSure Z and Medtronic CapSure Z Novus

• Designed so electrode Ω is high, but V low so current (I) is low as well, reducing battery drain

• Control of manufacturing

– Batteries, circuit boards, capacitors, etc., specific to needs, can lead to improved efficiencies and lowered static current drain

– Highly reliable lead design

www.pace

ricd.c

om

Page 80: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

8080

Putting It All Together

• Pacemaker Longevity is:

– A function of programmed parameters (rate, output, % time pacing)

– A function of useful battery capacity

– A function of

• Static current drain

• Circuit efficiency

• Output Impedance

• The lower the programmed sensitivity the MORE sensitive the device

– Lead integrity also affects sensingwww.pace

ricd.c

om

Page 81: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

8181

Status Check

• Determine the threshold amplitude

Capture threshold = lowest value with consist capture

This is at 1.25 V

Click for Answer

1.25 V 0.05 V0.75 V1.00 V

www.pace

ricd.c

om

Page 82: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

8282

Status Check

OR

Click for Answer

Pacemaker A is able to “see” signals as small as 0.5 mV. Thus, it is more sensitive.

Programmed Sensitivity 0.5 mV

Pacemaker A

Programmed Sensitivity 2.5 mV

Pacemaker B

• Which of these pacemakers is more sensitive?

www.pace

ricd.c

om

Page 83: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

8383

Status Check

• A pacemaker lead must flex and move as the heart beats. On average, how many times does a heart beat in 1 year?

Click for Answer

35 MILLION times. It is not a simple task to design a lead that is small, reliable, and lasts a lifetime.

www.pace

ricd.c

om

Page 84: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

8484

Status Check

Click for Answer

Lead Fracture:

• High Impedance

• Possible failure to capture myocardium

Do you notice anything on this x-ray?

www.pace

ricd.c

om

Page 85: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

8585

Status CheckWhat would you expect?

• Which value is out of range?

• What could have caused this?

Pacemaker Interrogation Report

Mode: DDDR

Lower: Rate 60 ppm

UTR: 130 ppm

USR: 130 ppm

Atrial Lead Impedance: 475 Ohms

Ventricular Lead Impedance: 195 Ohms

Insulation failure

Click for Answer

www.pace

ricd.c

om

Page 86: Welcome Field Engineerpacericd.com/documents/IBHRExAM/Fundamentals Of Cardiac Device… · – Sustained VT or VF induced during EP ... • 209 cc • < 1,000 implants/year . 27 Implant

86

Thank You