welcome to section e, principles of disinfection and ... · welcome to section e, principles of...
TRANSCRIPT
![Page 1: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/1.jpg)
1
Welcome to Section E, Principles of Disinfection and Sterilization in the
Out-Patient Practice Settings. This is the fifth of a seven part state
approved course required to meet compliance with the NC Rule .0206
Infection Control in Healthcare Settings. This course is specifically
designed for the Out-Patient practice Settings.
![Page 2: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/2.jpg)
2
Every year millions of invasive medical procedures are performed in the
outpatient settings. Each of these procedures involves contact by a
medical device or surgical instrument with a patient’s sterile tissue or
mucous membranes. A major risk of all such procedures is the
introduction of pathogens which can lead to infection. Failure of
appropriate disinfection and sterilization can also lead to person to
person trans mission of BBP or MTb.
The objectives of this lecture are to have the participants be able to
state the principles of cleaning, disinfection and sterilization; and be
able to list the current methods for disinfection and sterilization, per the
CDC HICPAC Guideline for Disinfection and Sterilization in Healthcare
Facilities published in 2008. These Guidelines should be used for
developing policies and procedures in all healthcare practice settings.
It is critical that the facility designated infection preventionist be familiar
with the concepts of cleaning, disinfection, and sterilization to ensure
appropriate methods, monitor and evaluate practice, and monitor
patient outcomes in order to identify failures in practice or processes.
![Page 3: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/3.jpg)
3
Over 35 years ago, Dr Earle Spaulding devised a rational approach to
disinfection and sterilization of patient care items based on the item’s
intended use. He placed every patient care item into one of three
categories based on the risk of infection involved with its use. The 3
categories are critical, semi-critical, and noncritical. The level of
disinfection or sterilization then is dependent on the category the item is
placed in: critical are items that contact sterile tissue such as surgical
instruments and must be sterilized; semi-critical are items that contact
mucous membranes such as endoscopes or non-intact skin such as
wound therapy equipment and must be high-level disinfected because it
kills all but high numbers of bacterial spores; and noncritical devices
that contact only intact skin such as stethoscopes which must be low-
level disinfected.
![Page 4: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/4.jpg)
4
To understand why certain disinfection or sterilization products and
process are recommended, one must understand that microorganisms
have differing abilities or characteristics for resistance to disinfectants
and sterilants. In terms of most to least resistant, Prions, the causative
agent of Mad Cow and CJD disease, are the most resistant and do not
respond to conventional chemical and physical decontamination
methods. Next, in terms of resistance, are spores of which Clostridium
difficile, or C diff is common, and has a protective covering that forms in
about a day after it is expelled in fecal material into the environment.
Then MTB which also has a protective outer covering and makes it
more difficult to kill. The non-enveloped viruses which include norovirus
are resistant to alcohol and this is problematic because of the lack of
effectiveness associated with the widely used alcohol-based hand rubs
in healthcare. Next are the fungi like Candida, followed by the common
vegetative bacteria like Staph, strep, enterococci, and the gram
negatives like pseudomonas, E coli, Serratia. Finally, the easiest to kill
are the enveloped viruses like HIV and HBV.
![Page 5: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/5.jpg)
5
Healthcare workers (HCWs) should be aware that there are factors that
are known to influence the efficacy of the disinfection and sterilization
process outcome to include: one of the most important is the cleaning of
the object. Cleaning reduces the bioburden and removes foreign
material (organic and inorganic salts) that interfere with sterilization
process by acting as barriers to the sterilization agent. Precleaning in
patient-care areas may be needed on items that are heavily soiled with
feces, sputum, blood, and so on. Items sent to Central Processing
without removing gross soil may be difficult to clean because of dried
secretions and excretions. One of the common failures of reprocessing
is not testing to confirm that the concentration and exposure time to the
disinfectant or sterilant follows the manufacturer’s directions. Another
factor is the nature of the object, meaning does the object have hinges,
crevices, or lumens that could create difficulties in reprocessing. Some
factors that influence disinfectant processes are temperature and water
hardness. For example the activity of disinfectants increases as the
temperature increases, but some exceptions exist. Furthermore, too
great an increase in temperature causes the disinfectant to degrade
and weakens its germicidal activity. An increase in pH improves the
effectiveness of some disinfectants (glutaraldehydes, quaternary
ammonium compounds) but decreases the antimicrobial activity by
altering the disinfectant molecule of phenolics, hypochlorites, and
iodine. Also, water hardness reduces the rate of kill of certain
disinfectants.
![Page 6: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/6.jpg)
6
The first of the three categories we will review is the processing of
critical instruments and patient care items.
![Page 7: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/7.jpg)
7
Critical patient care items or instruments involve a high risk of infection.
These items penetrate the bloodstream or other normally sterile tissues.
They have the highest risk of transmitting infection and should be heat-
sterilized between patient uses. Alternatively, use sterile, single-use
disposable devices.
The goal of sterilization processing is to kill all microorganisms.
• Examples of items included in this category are surgical
instruments, for example scalpels, cardiac catheters, all
intravenous devices, and urinary catheters,
![Page 8: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/8.jpg)
8
The next category is semi-critical instrument processing. Semi-critical
items and equipment are very commonly used in the out-patient setting.
One of the most commonly used items in this category is endoscopes
used for millions of procedures in the US each year for diagnostic
screening, treatment, and surgery. However, more healthcare
associated outbreaks have been linked to contaminated endoscopes
than to any other medical device. This is because of the complexity of
the equipment, with multiple channels and difficulty of adequately
reprocessing. Endoscopes have been associated with numerous
breeches including: inadequate cleaning, improper selection of cleaning
and disinfecting agents, failure to follow automatic endoscopic
reprocessor (AER) manufacturer’s recommendations for disinfection, or
lack of maintenance resulting in healthcare associated infections and
outbreaks.
![Page 9: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/9.jpg)
9
Semi-critical items are those items that contact only mucous
membranes or non-intact skin and do not penetrate soft tissues. Intact
mucous membranes, such as those in the lungs or the GI tract,
generally are resistant to infection by common bacterial spores, but
susceptible to other organisms, such as bacteria, mycobacteria, and
viruses. As such, they have a lower risk of transmission than critical
items.
Semi-critical instruments require high level disinfection to kill all
microorganisms except high numbers of bacterial spores.
Items in this category that are heat-tolerant should be heat sterilized
between patient uses. For respiratory therapy equipment that is heat
sensitive, facilities may use the heat automated pasteurization method.
However, for other heat-sensitive instruments immersion in chemical
sterilants will achieve high level disinfection (HLD).
Examples of semi-critical instruments include respiratory therapy
anesthesia and equipment, GI endoscopes and bronchoscopes, ear and
vaginal speculums, and diaphragm fitting rings.
![Page 10: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/10.jpg)
10
There are many products on the market that can be used for high level
disinfection of semi-critical items. Healthcare providers should follow the
manufacturers’ directions for use of products to achieve high level
disinfection. The times for immersion to achieve high level disinfection
will vary according to the germicide used and the concentration. The
exposure times may be as short as 10 minutes or as long as 45
minutes, and at in-use temperatures from 20o to 25oC.
Glutaraldehyde, hydrogen peroxide, ortho-phthalaldehyde, peracetic
acid with hydrogen peroxide and chorine are cleared by the Food and
Drug Administration (FDA) provided all factors recommended are met.
Compatibility of some sterilants with patient equipment should be
considered when choosing products for high level disinfection; for
example some scopes may be damaged when using hydrogen peroxide
formulations. Importantly, outbreaks continue to occur particularly in the
outpatient settings when ineffective disinfectants, including iodophors,
alcohol, and over-diluted glutaraldehyde are used for high level
disinfection.
![Page 11: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/11.jpg)
11
The last category is noncritical items. In contrast to critical and some
semi-critical items, most noncritical items may be decontaminated
where they are used and do not need to be transported to a central
processing area. There is no direct risk of infection from patient contact
with contaminated noncritical items used on patients’ intact skin.
However, there is a risk of a subsequent secondary infections to
patients from contact with contaminated noncritical items used during
patient care. This can happen by secondary transmission, when HCWs
handle contaminated noncritical items (blood pressure cuffs) without
performing adequate hand hygiene or are wearing gloves that are not
changed. Or, the patients themselves touching contaminated surfaces
in their environment ( bedrails, exam tables) colonize their skin or
mucous membranes with MDROs that could become a subsequent
secondary infection with that pathogen, especially if the patient has
areas of non-intact skin, invasive procedures or devices, or have
underlying immunocompromising conditions.
![Page 12: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/12.jpg)
12
There are non-critical surfaces like the walls, floors, windows, room
curtains. And then there are noncritical devices and instruments that are
directly on the patient. Noncritical instruments and devices only contact
intact (unbroken) skin, which serves as an effective barrier to
microorganisms. However, these devices, if contaminated with
epidemiological important organisms (MDROs), could result in
colonization of the patient or HCWs hands.
These items carry a lower risk of transmitting infections and usually
require only cleaning and low-level disinfection.
Intermediate-level disinfectants destroy vegetative bacteria,
mycobacteria, most viruses, most fungi, but not bacterial spores. Low-
level disinfectants destroy vegetative bacteria, some fungi and viruses,
but not mycobacteria or spores.
When using a low-level disinfectant and there is visible blood on the
item, according to OSHA it must have a label claim for killing HIV and
HBV or have tuberculocidal activity. Alternatively, if an item is visibly
bloody, it may be cleaned and disinfected using an intermediate-level
disinfectant before use on another patient.
• Examples of instruments in this category include blood
pressure cuffs, EKG leads, pulse oximeters, stethoscopes, and
bedpan.
![Page 13: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/13.jpg)
13
Intermediate-level EPA-registered hospital disinfectants are those with label claims regarding tuberculocidal activity such as chlorine-based products with a 1:10 or 1:100 dilution, phenolics, and some quaternary ammonium compounds. Low-level EPA- registered hospital disinfectants with no tuberculocidal claim may include chlorine at a 1:500 dilution,( or 100 ppm concentration), quaternary ammonium compounds, or 70 to 90% alcohol.
A relatively new product available is accelerated hydrogen peroxide which contains very low levels of anionic and nonionic surfactants that act with hydrogen peroxide to produce microbicidal activity. These ingredients are considered safe for humans and benign for the environment. This product is prepared and marketed in several concentrations from 0.5% to 7%. The lower concentrations (0.5%) are designed for the disinfection of hard surfaces while the higher concentrations (2%) are recommended for use as high-level disinfectants. A 0.5% accelerated hydrogen peroxide demonstrated bactericidal and virucidal activity in 1 minute, and mycobactericidal and fungicidal activity in 5 minutes. It is more costly than other low-level disinfectants such as quaternary ammonium compounds. The product is claimed to have an excellent antimicrobial performance and a favorable safety profile.
Generally, HCWs should follow the manufacturer's ―use dilution‖ for all products to achieve the level of disinfection required. Exposure times vary but are minimally 60 seconds to allow for drying and completion of the disinfectant activity.
One major controversy of disinfecting non-critical surfaces is that the CDC Disinfection and Sterilization Guidelines recommended a minute exposure for all non-critical surface disinfection, which is in conflict with manufacturer’s instructions. In order to get EPA clearance of the CDC Guideline, it was necessary to insert the sentences, "By law, all applicable label instructions on EPA-registered products must be followed. If the user selects exposure conditions that differ from those on the EPA-registered product label, the user assumes liability from any injuries resulting from off-label use and is potentially subject to enforcement action under FIFRA". There are several points that should be made about this apparent disconnect between label instructions and what studies show. Multiple scientific studies have demonstrated the efficacy of hospital disinfectants against pathogens causing healthcare-associated infections with a contact time of at least 1 minute; the only way an institution can achieve a contact time of 10 minutes is to reapply the surface disinfectant 5-6 times to the surface as the typical dry time for a water-based disinfectant is 1.5-2 minutes and currently, healthcare facilities like UNC Health Care are achieving surface disinfection of non-critical patient care items and environmental surfaces by one application of a disinfectant and requiring a >1 minute dry time; equally important as disinfectant contact time is the application of the disinfectant to the surface or equipment to ensure that all contaminated surfaces and non-critical patient care equipment are wiped as current studies demonstrate that only approximately 50% of high-risk objects are cleaned at terminal cleaning. There are no data that demonstrate improved infection prevention by a 10 minute contact time versus a 1 minute contact time. Thus, the CDC Disinfection and Sterilization Guideline authors, who are from UNC Healthcare, believe the guideline allows them to continue the use of low-level disinfectants for noncritical environmental surfaces and patient care equipment with a 1 minute contact time Additionally, all healthcare facilities should emphasize the thoroughness of cleaning to ensure that all contaminated surfaces are wiped.
![Page 14: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/14.jpg)
14
Many Ambulatory Care facilities are challenged with space limitations
for complying with CDC recommendations when reprocessing reusable
patient care items. In the next section, we will review the steps for
instrument processing. The Central Processing area ideally should be
divided into at least 3 separate areas. Physical barriers should
separate the decontamination area from the other sections to contain
contamination on used items. In the decontamination area reusable
contaminated supplies are received, sorted, and decontaminated. The
recommended airflow pattern should contain contaminates within the
decontamination area and minimize the flow of contaminates to clean
areas. The American Institute of Architects recommends negative
pressure and no fewer than 6 air exchanges per hour in the
decontamination area; AAMI recommends 10 air exchanges per hour in
the decontamination room and positive pressure in the sterilizer
equipment room. The packing area is for inspecting, assembling, and
packaging clean but not sterile items. The sterile storage area should be
a limited access area except for sterile storage, where the relative
humidity should not exceed 70%. The walls, floors, and ceiling should
be made of non-shedding materials.
![Page 15: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/15.jpg)
15
Most instrument cleaning, disinfecting, and sterilization should occur in a
designated central processing area to control both quality and personnel
safety.
To prevent cross-contamination, the instrument processing area should
be physically or spatially divided into 3 regions for cleaning, packaging,
sterilization, and storage.
• In the cleaning area, reusable contaminated instruments are
received, sorted, and cleaned.
• The packaging area is for inspecting, assembling, and
packaging clean instruments in preparation for final sterilization.
• The sterilization and storage area contains the sterilizers and
related supplies, incubators for analyzing spore tests (if
performed in office—although some states require using a
testing service), and can contain enclosed storage for sterile
items and disposable (single-use) items.
![Page 16: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/16.jpg)
16
Cleaning is the first step in all decontamination processes. Cleaning
involves the physical removal of debris by scrubbing and using water
with detergents or enzymatic cleaners, or by an energy based process
such as using washer-decontaminators or ultrasonic cleaners.
![Page 17: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/17.jpg)
17
Automated or mechanical cleaning equipment, such as ultrasonic
cleaners, instrument washers, and washer-disinfectors, are commonly
used to clean instruments. Automated cleaners improve the efficacy of
the cleaning process and reduce the handling of sharp instruments. A
more alkaline detergent is generally used with mechanical equipment to
compensate for the lack of friction used with manual cleaning. The
manufacturer’s recommendations for dilution, temperature, water
hardness, and use (designed for use in washer/decontaminators) should
be followed. After cleaning, instruments should be rinsed with water to
remove chemical or detergent residue.
Photo credit: Chris Miller, PhD, Indiana University School of Dentistry.
![Page 18: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/18.jpg)
18
If manual cleaning is necessary, soak instruments in a rigid container filled with a
detergent/disinfectant, enzymatic cleaner, or even plain water. This step prevents
drying of patient material and makes cleaning easier and less time consuming. Manual
cleaning requires a nearly neutral detergent. Enzymes, usually proteases are
sometimes added to neutral pH detergent solutions to assist in removal of organic
material. Cleaning solutions can also contain lipases (enzymes active on fats) and
amylases (enzymes active on starches). Enzyme detergents are cleaners, not
disinfectants, and some disinfectants may inactivate enzymes. Neutral pH detergents
that are compatible with medical instruments and endoscopes may be the best choice
for cleaning, especially delicate instruments. Some data show that enzymatic
detergents are more effective than neutral pH detergents. New formulations of non
enzymatic hydrogen peroxide have demonstrated efficacy comparable to enzymatic
detergents in removing protein and endotoxin materials in lab testing.
Other issues include:
• Do not use high-level disinfectants/sterilants (e.g., glutaraldehyde) as
instrument-holding solutions.
To avoid injury from sharp instruments, personnel should wear puncture-resistant,
heavy-duty, utility gloves (i.e., not patient care gloves,) proceed slowly and carefully
when handling or manually cleaning contaminated instruments and devices to avoid a
percutaneous injury.
Submerge instruments in detergent solution and scrub with long handled brush to
avoid spattering
• To protect against splashes, a facemask, eye protection or face shield, and a
gown or jacket should be worn.
Photo credit: Lt. Col. Jennifer Harte, U.S.A.F. Dental Investigation Service, Great
Lakes, IL.
![Page 19: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/19.jpg)
19
After thorough cleaning and drying of instruments, critical and semi-
critical instruments that will be stored before use should be wrapped or
placed into container systems prior to heat sterilization. This step
protects items from contamination after the sterilization cycle and during
storage.
Open or unlock hinged instruments so that all surfaces are exposed.
Place a chemical indicator inside each wrapped package. If the
indicator cannot be seen from the outside, place another indicator (e.g.,
indicator tape) on the outside of the package.
Always wear heavy-duty, puncture-resistant utility gloves while
inspecting and packaging instruments.
![Page 20: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/20.jpg)
20
After items are packaged, they are ready to undergo the sterilization
process, the goal of which is the complete elimination or destruction of
all forms of microbial life by either physical or chemical processes.
![Page 21: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/21.jpg)
21
There are two types of heat sterilization methods commonly used in
outpatient settings.
1) Steam under pressure (autoclaving). There are two types of tabletop
steam autoclaves:
In most commonly used gravity displacement sterilizers,
steam enters the chamber and unsaturated air is forced out
of the chamber through a vent in the chamber wall.
In contrast, pre-vacuum sterilizers are fitted with a vacuum
pump to create a vacuum in the chamber and ensure air
removal from the sterilizing chamber and load before the
chamber is pressurized with steam. This method improves
the speed and efficiency of the sterilization process.
2) Dry heat sterilizers are either static air (convection or FDA-approved
oven type) or forced air (rapid heat-transfer).
With all of these methods, always use FDA-approved devices and
closely follow the manufacturer’s instructions for proper use.
![Page 22: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/22.jpg)
22
Each sterilization process method has time and temperature
parameters that must be followed depending on the type of sterilizer.
Gravity Displacement typical operating temperatures are 30 minutes at
121-123oC (250-254oF) and at 132-135oC (270-275oF).
Tabletop steam sterilizers used most commonly in doctor’s offices and
out-patients settings are compact gravity displacement steam sterilizers
that have a chamber volume of not more than 2 cubic feet and that
generate steam when distilled or deionized water is added.
The prevacuum sterilizer type, because of higher operating
temperatures ranging from 132–135oC (270-275oF) to 141-144oC (285-
291oF) and rapid removal of air, make for a shorter cycle time of 4
minutes.
Dry Heat sterilization has a typical standard cycle of one to two hours
at 170 to 190oC or 320 to 375oF.
![Page 23: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/23.jpg)
23
Steam sterilization has several advantages over other types of
sterilization processes including that it is nontoxic; the cycle is easy to
control and monitor with a huge safety margin. It is inexpensive, rapidly
microbicidal and the least affected by contamination with organic or
inorganic soils and penetrates medical packaging and device lumens.
For all these reasons, it is the first choice for use with heat tolerant
medical instruments.
![Page 24: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/24.jpg)
24
There are several disadvantages of steam sterilization as well. The
most serious of which is that steam sterilization will destroy heat or
moisture–sensitive instruments and likely damage most heat labile
instruments, which are often complex, and very expensive instruments.
Even heat tolerant items may have dulling or rusting from the steam
sterilization process.
Another important disadvantage is the potential of HCWs getting burns
when handling or removing the instruments from the sterilizer.
![Page 25: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/25.jpg)
25
Dry heat sterilizers work by transferring heat energy from air inside the
oven to the instruments. The lack of vapor pressure requires higher
temperatures to achieve sterilization.
Dry heat is good for items that are likely to dull or rust in the autoclave,
and it is good for items like powders, cellulose or ink that require
sterilization.
However, the packaging used must be able to withstand high
temperatures to prevent an accidental fire or charring of contents that
could happen inside the sterilizer.
![Page 26: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/26.jpg)
26
There are two types of dry heat sterilizers, static air and forced air. The
static air sterilizer has heating coils in the bottom of the chamber. It
works by natural convection as hot air rises from the coils in the
chamber to reach adequate temperatures. The forced air sterilizer
creates rapid air transfer by circulating heated air throughout the
chamber. The advantage is that the forced air sterilizer has a shorter
cycle time of about 12 minutes as compared to an hour or two for the
static air sterilizer.
![Page 27: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/27.jpg)
27
Heat-sensitive instruments that are require sterilization or high-level
disinfection can alternatively achieve this by soaking them in a liquid
chemical sterilant cleared by the FDA. However, exposure to these
powerful and toxic chemicals can be harmful to health care personnel
and patients if the manufacturer’s instructions for use and safety
precautions are not followed precisely. For these reasons, CDC
encourages the use of heat-tolerant or disposable alternatives.
A list of the FDA cleared sterilants and high level disinfectants with label
claims for reprocessing reusable medical devices is available at:
http://www.fda.gov/cdrh/ODE/germlab.html
Photo credit: Col. Shannon Mills, United States Air Force.
![Page 28: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/28.jpg)
28
A summary of important recommendations when facilities are
implementing sterilization:
First, steam is preferred for critical items not damaged by heat.
Second, follow the operating parameters recommended by the
manufacturer for the equipment or products used.
Third, use low temperature sterilization technologies for reprocessing
critical items that may be damaged by heat.
And fourth, immediately use critical items that have been sterilized by
an immersion in a chemo-sterilizer solution like glutaraldehyde or OPA.
There should not be long term storage because these items are not
wrapped for protection.
![Page 29: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/29.jpg)
29
To review, there are 3 key issues of sterilization . The first, that all
sterilization processes are effective in killing spores, the hardest of
microorganisms to kill.
Second, that cleaning removes salts and proteins and precede all high
level disinfection and sterilization.
And third, that failure to clean or ensure exposure of microorganisms to
sterilants could interfere with the adequacy of the sterilization process.
![Page 30: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/30.jpg)
30
Next, we will discuss monitoring. Monitoring the effectiveness of your
sterilization equipment, procedures, and recordkeeping is essential for
patient safety, and mandated by several regulatory agencies (TJC,
CMS, and NC rule .0206).
![Page 31: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/31.jpg)
31
The sterilization procedure should be monitored routinely by using a
combination of mechanical, chemical, and biological indicators to
evaluate the sterilizing conditions and indirectly the microbiologic status
of the processed items. The mechanical monitors for steam include the
daily assessment of cycle time and temperature by examining the
temperature record charts or computer readout, and an assessment of
pressure via the pressure gauge.
Chemical indicators usually are either heat or chemical sensitive inks
that change color when one or more sterilization parameters ( e.g.
steam-time, temperature, and or saturated steam) are present.
Biological indicators are recognized by most authorities as being the
closest to the ideal monitors of the sterilization process; because they
measure the sterilization process directly by using the most resistant
microorganisms (i.e. Bacilllus spores) and not merely by testing the
physical and chemical conditions necessary for sterilization.
![Page 32: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/32.jpg)
32
Chemical indicators are convenient, are inexpensive, and indicate that the item has been exposed to the sterilization process. Based on current studies, chemical indicators should be used in conjunction with biologic indicators but should not replace them because they indicate sterilization at marginal sterilization time and because only a biologic indicator consisting of resistant spores can measure the microbial killing power of the sterilization process.
Chemical indicators are affixed to the outside of each pack to show that the package has been processed through a sterilization cycle, but these indicators do not prove sterilization has been achieved. Preferably a chemical indicator also should be placed on the inside of each pack to verify sterilant penetration.
Chemical indicators have been grouped into five classes based on the their ability to monitor one or multiple sterilization parameters. If the internal or external indicator suggests inadequate processing, the item should not be used.
If using a prevacuum steam sterilizer, an air-removal test called the Bowie-Dick Test must be performed daily in the empty sterilizer to ensure air removal.
![Page 33: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/33.jpg)
33
Biological Indicator are the only process that directly monitor the
lethality of a given sterilization process. Spores used to monitor a
sterilization process have demonstrated resistance to the sterilizing
agent and are more resistant than the bioburden found on medical
devices. Geobacillus stearothermophilus spores (105) are used to
monitor steam sterilization, hydrogen peroxide, gas plasma , and liquid
peracetic acid sterilizers. Bacillus atrophaeus formerly known as
Bacillus subtilis at 106 are used to monitor ETO and dry heat.
Geobacillus stearothermophilus is incubated at 55-60C
![Page 34: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/34.jpg)
34
An ideal biological monitor of the sterilization process should be easy to use, be inexpensive, not be subject to exogenous contamination, provide positive results as soon as possible after the cycle so that corrective action may be accomplished, and provide positive only when the sterilization parameters (e.g. steam-time, temperature, and/or saturated steam, ETO time, temperature, relative humidity) are inadequate to kill microbial contamination.
Originally, spore strip biological indicators required up to 7 days of incubation to detect viable spores from marginal cycles. The next generation of biological indicators was self contained in plastic vials containing a spore-coated paper strip and a growth media in a crushable ampoule. This indicator had a maximum incubation of 48 hours but significant failures could be detected in <24 hours. Rapid readout indicators that have been available for more than 10 years, and have been shown to be comparable to that of the conventional biologic indicators.
The standard biological indicator used for monitoring full-cycle steam sterilizers does not provide reliable monitoring of flash sterilizers. Biological indicators specifically designed for monitoring flash sterilization are now available, and studies comparing them have been published.
![Page 35: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/35.jpg)
35
Steam and low temperature sterilizers (peracetic acid and hydrogen gas
plasma) should be monitored at least weekly (with the appropriate
commercial preparation of spores). Weekly biological monitoring is a
requirement in the NC Rule .0206
If a sterilizer is used frequently (e.g. several loads per day), daily use of
biological indicator allows for earlier discovery of equipment
malfunctions or procedure errors and thus minimizes the extent of
patient surveillance and product recall needed in the event of a positive
biological indicator.
Each load should be monitored with a biological indicator if it contains
an implantable object. If feasible, implantable items should not be used
until the results of spore tests are known to be negative.
![Page 36: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/36.jpg)
36
Since sterilization failure can occur (about 1% for steam) a procedure to follow
in the event of positive spore tests with steam sterilization has been provided
by the CDC and AORN. The positive spore test may occur for reasons such
as slight variations in the resistance of the spore, or improper use of the
sterilizer.
After a single positive biologic indicator test with steam sterilization, remove
and retest the sterilizer. Objects other than implantable object, do not need to
be recalled unless the sterilizer or the sterilization procedure is defective as
determined by maintenance personnel or inappropriate cycle settings are
identified. If additional spore tests remain positive, consider the items non-
sterile and recall. Then reprocess the items from the implicated load(s).
If patient care items were used before retrieval, the infection preventionist
should assess the risk of infection in collaboration with the physician, and if
needed, consult an outside reprocessing specialist. The margin of safety in
steam sterilization is sufficiently large that there is a minimal risk associated
with items in a load that show spore growth, especially if the item was properly
cleaned and the temperature was achieved as demonstrated by a chemical
and readout chart.
After a single positive biological indicator used with a method used other than
steam sterilization, treat as nonsterile all items that have been processed in
that sterilizer, dating from the last sterilization cycle having the last negative
biological indicator. Retrieve items and reprocess.
![Page 37: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/37.jpg)
37
As part of a quality control program maintain sterilization records
(mechanical, chemical and biological) for a time period that complies
with standards, for example 3 to 5 years to meet the Joint Commission
visitation schedule and the statutes of limitations, and some state and
federal regulations. Cat II
For each sterilization cycle record the type of sterilizer and cycle used;
the load identification number; the load contents, the exposure
parameters (e.g time and temperature): the operator’s name or initials;
and the results of mechanical, chemical, and biological monitoring. Cat
II
![Page 38: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/38.jpg)
38
Once items are cleaned, dried, and inspected, those requiring
sterilization must be wrapped or placed in rigid containers and should
be arranged in instrument trays or baskets according to the guidelines
provided by AAMI and other professional organizations. These
guidelines state that hinged instruments should be opened; items with
removable parts should be disassembled unless the device
manufacturer or research suggest otherwise. Follow manufacturer’s
instructions for preparing instruments such as concave surfaces placed
downward and heavy items placed below light items to avoid damage.
The wrapping should be done in a manner to prevent gaps and tenting.
There are several choices in products to maintain sterility of surgical
instruments, including the most popular in ambulatory care, the peel
open pouches, also rigid containers, and roll stock that will self seal to
form a pouch and sterile wraps woven and nonwoven. Healthcare
facilities can choose any of these options. The key is the packaging
material must allow penetration of the sterilant, must be compatible with
the type of sterilizer, and must be puncture resistant and durable, and
have received FDA clearance.
![Page 39: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/39.jpg)
39
All items to be sterilized should be arranged so that all surfaces will be
directly exposed to the sterilizing agent. Thus the loading procedures
must allow for free circulation of steam (or another agent around each
item. Due to the variety of textiles and containers on the market, the
manufacturers of the sterilizer and wrap or container product should be
consulted on pack preparation and density parameters.
There are several important basic principles for loading a sterilizer:
allow for proper sterilant circulation; perforated trays should e placed on
their edge (e.g basins); small items should be loosely placed in wire
baskets; and peel packs should be placed on edge in perforated or
mesh bottom racks or baskets.
![Page 40: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/40.jpg)
40
A summary of recommendations for sterilization include that steam is
preferred for critical and semi-critical items that are not damaged by
exposure to high heat.
It is vital to always follow the manufacturer’s operating instructions for
sterilizers and products, and equipment.
It is also important to use ―FDA cleared‖ containers, wrapping, or
packaging system that is compatible with the type of sterilization
process used.
And it is critical for successful sterilization to not overload the sterilizer
chamber.
![Page 41: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/41.jpg)
41
Next we will review CDC recommendations for storage of sterile items.
The sterile storage area should be a well-ventilated area that provides
protection against dust, moisture, and temperature and humidity
extremes. Sterile supplies should be stored far enough from the floor (8
to 10 inches), the ceiling (5 inches unless near a sprinkler head or18
inches from the sprinkler head), and outside walls (2 inches) to allow for
adequate circulation, ease of cleaning, and compliance with fire codes.
Store sterile items so the packaging is not compromised (e.g.,
punctured, bent). Sterilized items should be labeled with a load number
that indicates the sterilizer used, the cycle or load number, the date of
sterilization, and the expiration date (if applicable).
Closed or covered cabinets are ideal but open shelving may be used for
storage. Any package that has fallen on the floor must be inspected for
damage to the packaging and contents. If the package is heat-sealed in
impervious plastic and the seal is still intact, the package should be
considered not contaminated. If undamaged, items in packaged plastic
need not be reprocessed.
![Page 42: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/42.jpg)
42
The shelf life of a packaged sterile item depends on the quality of the
wrapper, the storage conditions, the conditions during transport, the
amount of handling, and other events like exposure to moisture that
compromises the integrity of the package. If event-related storage of
sterile items is used, then packaged sterile items can be used
indefinitely unless the packaging is compromised. Packages should
always be inspected and evaluated before use for loss of integrity (e.g.
torn, punctured, or wet) If the integrity of the package is compromised,
repack and reprocess the pack before use.
Time related storage considers items remain sterile for varying periods
depending on the type of material used to wrap the item/tray. If time-
related shelf life (less common) is used, label the pack at the time of
sterilization with an expiration date. Once the expiration date is
exceeded the pack should be reprocessed.
![Page 43: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/43.jpg)
43
Recommendations for quality control are to provide comprehensive and
intensive training for all staff assigned to reprocess semi-critical and
critical medical/surgical devices to ensure they understand the
importance of reprocessing these instruments. To ensure and maintain
competency, 1)train each member of the staff who reprocesses semi-
critical and/or critical instruments as follows; 2) provide hands on
training according to facility policy for reprocessing; supervise all work
until competency is documented for each reprocessing task, 3) conduct
competency testing at the beginning of employment and regularly
thereafter (for example annually), and 4) review the written
reprocessing instructions regularly to ensure they comply with the
scientific literature and manufacturer’s instructions.
To assist NC healthcare facilities, a sample competency monitoring tool
for reprocessing developed by UNC Healthcare System is included in
the handouts.
![Page 44: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/44.jpg)
44
As part of the quality control program, conduct infection control rounds
periodically, minimally annually, in high risk reprocessing areas ( e.g.
Gastroenterology or Endoscopy Clinics) to ensure reprocessing
instructions are current and accurate and are correctly implemented.
Include a sterilizer maintenance contract with records of service, a
system of process monitoring, visual inspection of packing materials,
and traceability of load contents. Perform preventive maintenance on
sterilizers by qualified personnel who are guided by the manufacturer’s
instructions.
Periodically review policies and procedures for sterilization to ensure
that they are current with guidelines.
Consult AAMI and or the manufacturer for correct preparation and
wrapping materials that may be used with sterilizer.
![Page 45: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/45.jpg)
45
In the past two decades, newly emerging or remerging diseases have
created some concerns as to whether healthcare facilities should do
more or do something different with disinfection and sterilization of
equipment and instruments used on patients known or suspected of
having these emerging pathogens.
![Page 46: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/46.jpg)
46
This is a fairly comprehensive list of the emerging pathogens reported
or that may have a potential to be in the US. There is hepatitis C that
can survive in the environment for at least a couple weeks, and that it is
not more resistant to cleaning and disinfection than hepatitis B.
However, Clostridium difficile because of its ability to be a spore former
in the environment is resistant to alcohol-based hand rubs and nearly all
surface disinfectants except those containing a 1:10 dilution of bleach.
Cryptosporidium and Helicobacter pylori and E. coli 0157:H7 have
contaminated water systems from farm animal runoff. There are
multidrug resistant pathogens prevalent in healthcare. The pathogens
that have been the source of outbreaks in the community and
healthcare facilities are SARS, coronavirus, avian influenza, and
norovirus. And there is even the potential of the bioterrorism agents
including anthrax, plague, and smallpox.
![Page 47: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/47.jpg)
47
For all the these emerging pathogens just listed, healthcare facilities
can use standard disinfection and sterilization procedures for patient
care equipment which is adequate to sterilize or disinfect instruments or
devices contaminated with blood and other body fluids from persons
infected with these emerging pathogens.
![Page 48: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/48.jpg)
48
Only one disease is an exception to being eradicated by normal
disinfection, and that is Creutzfeldt-Jakob Disease or CJD. There are
other neurologic diseases that are also caused by prions such as
scrapie.
Creutzfeldt-Jacob Disease (CJD) and other prion diseases are rare,
fatal diseases of the central nervous system characterized by rapid,
progressive dementia and other sensory and motor disturbances.
CJD is thought to be caused by small, abnormal pieces of protein
referred to as prions. Prions occur naturally in mammals and birds but
turn destructive when they become folded in a way that changes how
they react with other brain chemicals.
CJD has an extremely long incubation period and is fairly rare with only
one case per million worldwide reported.
![Page 49: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/49.jpg)
49
A new variant of CJD (vCJD), the human version of Bovine Spongiform
Encephalopathy (BSE), was recognized after what has become known
as Mad Cow Disease occurred from changes made in the processing of
food for bovines in Great Britain. Several hundred people to date have
been reported, mostly in the UK, Italy, France, Hong, and Canada which
is waning in number of new cases, since the food processing practices
were changed. One case reported in the US is a former UK resident.
![Page 50: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/50.jpg)
50
Because of the devastating nature of this rare degenerative neurologic
disorder caused by a prion. The CJD does not appear to be
epidemiologically related by disease by age or sex, has no seasonal
distribution, and no geographic aggregation that cannot be accounted
for from a reservoir. Although the disorder has a very long, probably
several years in length, incubation period, once the individual becomes
symptomatic there is a rapid progression of serve symptoms and death.
Importantly to this discussion prions, although extremely rare, are
resistant to conventional disinfection and sterilization reprocessing.
![Page 51: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/51.jpg)
51
High risk tissues include: brain, spinal cord, and posterior eye (retina
and optic nerve). All other tissues are considered low or no risk. If the
patient is identified with a neurologic disorder of unknown etiology and
has a procedure involving the high risk tissues with critical or semi-
critical equipment, then keep the equipment in a closed container or bag
until consultation from reprocessing experts or SPICE can be obtained
for the appropriate evaluation and methods for handling and
reprocessing.
![Page 52: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/52.jpg)
52
Thank you for your time and attention. This concludes Section E,
Sterilization and Disinfection. The next slide contains important
references that were used in the development of this Section.
![Page 53: Welcome to Section E, Principles of Disinfection and ... · Welcome to Section E, Principles of Disinfection and Sterilization in the Out-Patient Practice Settings. This is the fifth](https://reader033.vdocuments.net/reader033/viewer/2022052722/5f0cfe587e708231d43825f2/html5/thumbnails/53.jpg)
53