what you need - enartis · f +61 3 5987 3303 e [email protected] .au w abn: 60 068 057 045...
TRANSCRIPT
WHAT YOU NEED
• Visible Spectrophotometer
• Micropipettes▫ 10 - 100µL▫ 100 - 1000µL
• Cuvettes
• Parafilm
• Timer
WHAT TESTS CAN BE DONE
• Acetic Acid
• Amino Acid Nitrogen
• Ammonia
• Citric Acid
• Gluconic Acid
• Glucose & Fructose
• Lactic Acid
• Malic Acid
• Sucrose
WHAT TESTS CAN BE DONE
• Acetaldehyde
• Ascorbic Acid
• Ethanol
• Formic Acid
• Glycerol
• Metals
• Mannitol
• Sorbitol
• Succinic Acid
• Sulphite
• Starch
• Tartaric Acid
• Many more…
HOW THE KITS WORK
• Colorimetric
▫ Amino Acid Nitrogen (NOPA)
• Enzymatic
▫ Reduction of NAD+ to NADH
Malic Acid, Glucose & Fructose, Acetic Acid
▫ Oxidation of NADH to NAD+
Ammonia, Citric Acid
HOW THE KITS WORK
GOOD LABORATORY PRACTICE
• Calibrate Spectrophotometer
• Calibrate Micropipettes
• Cuvettes
• Kit storage
• Sample preparation
• Correct technique
SPECTROPHOTOMETER
• Calibrate every 3 months according to AS 3753
▫ Photometric accuracy
▫ Linearity check
▫ Stray light check
▫ Cell match check
SPECTROPHOTOMETER
• Erratic readings or excessive drift
▫ Dust in the sample compartment
▫ Fluctuating power supply
▫ Filter problem Go to 550nm, select %T and zero instrument on air
Adjust to 340nm and record %T, <5% is a problem
▫ Lamp re-adjustment required Go to 550nm and place white paper in light path. A bright
green rectangular light spot should be observed.
MICROPIPETTES
• Calibrate every 3 months according to AS 2162.2
▫ Distilled water and 4-figure balance required
▫ Test micropipette at its maximum capacity and again at close to its minimal volume (no more than 20% of its total capacity)
MICROPIPETTES
• Before calibrating, clean and regrease the micropipette piston
• Avoid liquid leaking inside the micropipette
▫ Always hold the micropipette more than horizontal
▫ Store on a micropipette rack rather than laying on bench or in a drawer
CUVETTES
• Disposable cuvettes should be disposed
• Do not handle the optical surfaces
• Avoid wiping the optical surfaces with tissue
• Do not clean quartz or glass cuvettes with detergent
KIT STORAGE
• Store kits cold to avoid reduction of shelf life
• Bring all reagents and samples to be analysed to room temperature prior to analysis
• Follow kit manufacturers guidelines for individual reagent storage…
KIT STORAGE
32 Brasser Avenue Dromana Victoria 3936 Australia T +61 3 5987 2242 F +61 3 5987 3303 E [email protected] W www.vintessential.com.au
ABN: 60 068 057 045
ENZYMATIC ANALYSIS KIT FOR THE DETERMINATION OF L-MALIC ACID IN GRAPE JUICE AND WINE
PRODUCT Product no. 4A160, for 30 tests, for in vitro use only. PRINCIPLE OF MEASUREMENT L-malic acid is found in grape juice and wine and is determined enzymatically according to the following equations: MDH L-malate + NAD
+ ↔ Oxaloacetate + NADH + H
+
L-malic acid is oxidised by nicotinamide adenine dinucleotide (NAD) to oxaloacetate using L-malate dehydrogenase (MDH) enzyme as a catalyst. The equilibrium does not favour formation of oxaloacetate and so oxaloacetate is removed by a trapping enzyme. The amount of NADH formed is measured at 340 nm and is stoichiometrically related to the amount of L-malate consumed. In this method, glutamate oxaloacetate transaminase (GOT) is used as the trapping enzyme. In the presence of L-glutamate, the oxaloacetate is irreversibly converted to L-aspartate.
GOT Oxaloacetate + L-glutamate → L-aspartate + α-ketoglutarate CONTENTS The kit includes the following reagents:
Reagent No. Reagent Preparation Quantity Stability
1 Buffer Nil 33 mL 2 years at 4oC
2 NAD Add 6.6 mL of distilled water, mix to dissolve
6.6 mL 2 years at 4oC
(diluted: 1 year at 4oC,
2 years at -20 oC)
3 GOT Mix gently by inversion before use 0.4 mL 2 years at 4oC
4 MDH Nil 0.4 mL 2 years at 4oC
5 Standard Nil 3.3 mL 2 years at 4oC
The shelf life of Reagents 1 & 2 can be extended by placing aliquots in a freezer. Do not freeze enzyme reagents 3 & 4. Failure to store reagents at the recommended temperature will reduce their shelf life. For concentration of Standard, refer to label on bottle. SAFETY
Wear safety glasses Reagent 1 is mildly corrosive Do not ingest Buffer or Standard as they contain sodium azide as a stabilizer
SAMPLE PREPARATION• Bring sample to room temperature prior to analysis
• Decolourise highly coloured samples▫ PVPP▫ Activated carbon not recommended for some assays▫ Complete decolourisation not necessary
• Dilute samples to approximate concentration of kit standard
• Filter turbid samples
• Avoid excessive use of SO2
CORRECT TECHNIQUE
• Analyse a blank and kit standard in every run
• Zero spectrophotometer according to kit instructions, usually against air without cuvette in light path
• Be consistent in measuring reaction times between blank, standard and samples
• Spike & duplicate analysis
CORRECT TECHNIQUE
• Disposable micropipette tips are single use only
• When mixing cuvette contents invert gently, don’t shake. Same goes for reagent bottles
• Avoid cross-contamination between assays and especially contamination of reagents
• Purity of distilled water is important, decant fresh each run and discard after use
TROUBLESHOOTING
• Low recovery on kit standard
▫ Blank assay activity is a sign of contamination and will cause low standard and sample results
▫ Keep a logbook of absorbance values for the blank and standard assays for each kit
▫ Pipetting error
▫ Coenzymes may have perished
TROUBLESHOOTING
• High recovery on kit standard
▫ Check photometric accuracy of spectrophotometer or absorbance drift
▫ Volumetric accuracy of pipettes
▫ Check that the change in absorbance in the Blank assay is consistent with previous runs
▫ Human error (pipetted twice into the same cuvette?)
TROUBLESHOOTING
• Negative results
▫ If the sample contains negligible amounts of the compound being measured and a dilution factor has been applied, then negative values may be exaggerated.
▫ Excessive potassium metabisulphite interferes with malic acid assay (false negative results)
TROUBLESHOOTING
• Assay absorbance
▫ The higher the absorbance value, the lower the amount of light transmitted and measured by the photometer
▫ An absorbance of 1.000 occurs when 90% of the light has been absorbed
▫ At absorbance readings greater than 1.000 – 1.500 errors can occur because absorbance values increase exponentially
TROUBLESHOOTING
Or, simply call the manufacturer!
Details for Greg Howell
W: www.vintessential.com.au