Wissner-Bosserhoff Movita Bed Instruction & Safety Manual Please read this manual before operation, for maintenance instructions and safe usage The enabling power of applied knowledge

Upload: others

Post on 12-Apr-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

1T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

Wissner-Bosserhoff Movita BedInstruction & Safety Manual

Please read this manual before operation, for maintenance instructions and safe usage

The enabling power of applied knowledge

Page 2: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

2

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

Index1. Table of Contents ..........................................................................................................................2 Important Safety Note ...................................................................................................................32. Notes on Instructions for use / Explanation of Symbols ...............................................................43. Area of Use / Intended Use ..........................................................................................................44. Setting up the Bed ........................................................................................................................55. Technical Data for the Bed / Nameplate .......................................................................................55.1. Electromotor Patient Surface .............................................................................................65.2. Electric motor height adjustment........................................................................................66. Product Safety / Extract from applied Standards ..........................................................................67. Product Overview (Example Wood Décor T10) ............................................................................78. Description of electrical Functions ................................................................................................88.1. Backrest Function (BR) ......................................................................................................98.2. Legs up position (LU) .........................................................................................................98.3. Comfortable sitting position (CS) .....................................................................................108.4. Auto contour Function (AC) .............................................................................................108.5. Height adjustment Function (HA) .....................................................................................109. Safety Notes for Height Adjustment ............................................................................................ 1110. Control box (Optional).................................................................................................................1211. Delivery .......................................................................................................................................1212. Assembly ....................................................................................................................................1312.1. Patient Surface ................................................................................................................1312.2. Patient Surface Motor ......................................................................................................1313. Transport / Moving Bed ..............................................................................................................1314. Function Test ..............................................................................................................................1414.1. Handset “Carisma 07 JPF“ Function Description.............................................................1414.2. Handset – Rear Side with Locking Function ....................................................................1514.3. Calf rest / Mechanical ......................................................................................................1614.4. Braking / Moving ..............................................................................................................1614.5. Adjusting Side rails (Model 06) ........................................................................................1714.6. Adjusting Side rails (Release Catch 09) ..........................................................................1714.7. Adjusting Side rails (Model 09) ........................................................................................1815. Permitted Accessories ................................................................................................................1915.1. Lifting Pole (Accessories) ................................................................................................1915.2. Infusion Holder (Accessories) ..........................................................................................2015.3. Patient Surface Extension (Option)..................................................................................2015.4. Mattress (Accessories) ....................................................................................................2015.5. Urine Bottle Basket / Urine Bag Container Model 09 (Accessory) ......................................................................................................................2115.6. Handset holder (Accessory) .......................................................................................................2115.7. Side rail exntension for 06/09 Side rail (Accessories) ................................................................2216. Cleaning and Disinfection ...........................................................................................................2316.1. Initial Cleaning .................................................................................................................23

Page 3: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

3T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

16.2. Disinfection / Sterilisation ..................................................................................................2416.3. Care Instructions for Wooden Parts ..................................................................................2416.4. Use of Cleaning Agents.....................................................................................................2516.5. Use of Disinfection Agents ................................................................................................2516.6. Damage to Wooden Parts .................................................................................................2516.7. Environmental Compatibility..............................................................................................2516.8. Quality ...............................................................................................................................2517. Help in case of Malfunctions .......................................................................................................2618. Storage .......................................................................................................................................2619. Safety Notes and Residual Risks ...............................................................................................2620. Maintenance and servicing .........................................................................................................3020.1. Completeness and Accessories ........................................................................................3120.2. Wear and Tear ...................................................................................................................3120.3. Functionality ......................................................................................................................3120.4. BGV A3/VDE 0751– Inspections of the Electrical Components of he Handset ....................................................................................................................3120.5. Castors ..............................................................................................................................3221. Useful Life ...................................................................................................................................3222. Disposal ......................................................................................................................................3223. Spare parts, Spare part lists and Service manuals.....................................................................3224. Nameplate and Labels ................................................................................................................3325. Guarantee and Service ...............................................................................................................3326. Contact Partner...........................................................................................................................3327. EC Declaration of Conformity .....................................................................................................35

Appendixes:Appendix A: Spare parts list Movita 09/10 .............................................................................................36Appendix B: Nameplate – explanation of details using an example ......................................................39Appendix C: Important safety and usage notes for nursing home beds ................................................39Basic Functions and uses of nursing care beds ....................................................................................40Appendix D: Battery operation (option) - additional notes .....................................................................45IMPORTANT SAFETY NOTE:Please make sure that you read the safety-relevant passages marked with a in the individual chapters of the instructions for use very carefully.

Page 4: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

4

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

Dear CustomersWe thank you for the trust that you have shown in purchasing one of our products. We are convinced that our product is fully able to cope with the requirements of everyday nursing activities. The careful choice of materials, the use of highly modern technologies and not least, the thoroughness and team spirit of our employees, mean that our products have a large benefi t and a solid quality. We are sure that this will mean care is made easier for you.

2. Notes On Instructions For Use / Explanation Of Symbols This symbol indicates safety-relevant instructions, the non-observance of which could have serious consequences (e.g. damage, injuries). For this reason it is essential that such instructions and notes are observed.

These instructions for use are intended for trained nursing staff or for those persons who are in charge of the commissioning and operation of this product. It is essential that you follow the instructions.

Caution! A precise knowledge and observation of these instructions for use is a prerequisite for handling this product. They should therefore be kept safely until the product is disposed of. To avoid operating errors and to guarantee operation without malfunction, the instructions for use must be available to operating personnel at all times. It is recommended that they are stored in the bed or in the bedside cabinet.

The texts and graphic representations in these instructions for use do not necessarily match the exact scope of delivery, as they are designed for all variations of the model series. The drawings and graphics are not to scale.

3. Area Of Use / Intended UseThe nursing home bed is normally used in the area of nursing and care for the elderly. This bedis used exclusively for this purpose.

It is used to compensate and ease handicaps for people in nursing institutes and carehomes for the elderly, as well as to simplify nursing tasks.Any other applications must be agreed beforehand in writing with the company Wissner-BosserhoffGmbH. The product must be used as a utility for nursing, and is subject to the regulations ofthe relevant professional organisation.

Page 5: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

5T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

This nursing bed is a medical product in the meaning of the regulations and standards named in the following chapter, “Product Safety / Excerpt of Applied Standards”. Accordingly this product may only be used under medical supervision. The decisive factor in whether nursing is medically supervised or not depends on whether this nursing takes place on the instructions of medical personnel or not. (cf. DIN EN 60601-2-38/A1:2000)

4. Setting Up The Bed• Make sure that the fl oor surface is suitable, so that no damage can occur to the fl oor. Floors that

are, for example, too soft, unsealed or with imperfections are not suitable. These fl oors are often not suitable: soft wooden fl oors, porous and soft stone fl oors, carpets with foam backing, soft linoleum or similar fl oor covering. If in doubt, please contact your fl ooring company and the Technical Customer Service of Wissner-Bosserhoff GmbH.

• Ensure that the bed stands freely in the room so that there is a suffi cient safety clearance all around the bed for all adjustment positions (including the tilting function). For the extreme positions of the bed, an additional safety clearance of 2.5 cm from objects, walls and installations must be taken into account, as protection from entrapment. The safety clearance of the head end from objects, walls and installations is at least 20 cm.

• Look for the existence of mechanical and electrotechnical protective measures, such as wallmounted bumpers and corner guards, RCB switches etc.

• Pay attention to the whereabouts of useful wall sockets for the mains connection for the bed. Positions are unsuitable that could lead to collisions or stretching when the bed is adjusted, or in ...which the mains cable stretches too far across the fl oor or even under the bed.

• In general, the use of extension cables or multiple adapters that may lie unsecured on the fl oor of the room should be avoided.

• Magnetic objects or magnetic fi elds must be kept at a distance

5. Technical Data for the Bed / Nameplate1

• Outer dimensions: approx. 208 x 106 cm (depending on the wooden lining)

• Nominal dimension of patient surface: 90 x 200 cm• Height adjustment: 36 to 80 cm depending on the model

(measured central section)• Backrest adjustment: 65.5°/ backrest retraction SBR 11.5 cm• Thigh section adjustment: 32°/ thigh rest retraction 4 cm• Safe workload patient surface: up to 205 kg (155kg patient + 30kg mattress + 20kg

linen and accessories) depending on model and confi guration

• Tilting: none• Bed weight: approx. 120 kg depending on model and confi guration• Protection type IPX4• Noise level: approx. 57 dB/A

Page 6: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

6

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

• Application environment: • the environment temperature lies in the range from + 10°C to + 40°C • the relative humidity lies in the range from 30% to 75% • the atmospheric pressure lies in the range from 700 hPa to 1060 hPa

5.1 Electric Motor Patient Surface• Electrical connection: 230 V 1.25 A 50 Hz• Protection type IPX4• Class of protection II• Output voltage 24 V / 32 V depending on model variant (please

read nameplate)• Operating time of motors max. 2 min. adjustment operation, 18 min. pause• Power consumption Up to approx. 320 VA peak load• Adjustment speed approx. 5.4 mm/sec.2

• Pressure force before redirection: max. 3,500N

1 All technical data are nominal data. Construction and manufacturing tolerances are in accordance with EN ISO 14253-1 and GUM-1:952 Approximate values according to the model design and load value

5.2. Electric Motor Height Adjustment• Max. voltage in idling mode 2 4 V DC• Current limiting (discharger): max.10 A• Max. lifting force, direct: 1 x 6,000 N• Max. tensile force 3,000N• Adjustment speed 10 – -14 mm/s depending on model, fi ttings and load• Protection type2: IPX4• Class of protection III protective extra-low voltage• Operating time of motors max. 2 min. adjustment operation, 18 min. pause

6. Product Safety / Extract From Applied StandardsThe product bears the mark and thus complies with the German and European safety regulations to be applied to the product. The product was designed and produced in accordance with and DIN EN ISO 13485: 2003.

Designation Comment Designation CommentMedical product acc. to

93/42/EECMDD (EU directive) DIN 33 402 Body mass of people

Law on Medical Devices MPG(national implementation)

DIN 50,967 Nickel-chrome and copper-nickelchrome

coating

EN ISO 14 971 Risk management appliedto medical products

DIN 68861-1/ EN12720

Furniture surfaces

DIN EN 12,182(relevant sections)

Technical aids for handicapped

people

EN 60 529 IP code (protection from humidity)

Page 7: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

7T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

Designation Comment Designation CommentIn compliance with EN

1970 A1:2004(relevant sections)

Adjustable beds forhandicapped people

DIN EN 60 601-1-2 EMV electromagnetic compatibility

In compliance with DINEN 60,601-1

(relevant sections)

Medical electrical appliances

BfArM – recommendationfrom

07-2000

Recommendations of German Federal

Institute for Drugs and Medical Products

(BfArM)

In accordance withDIN EN

60601-2-38/A1:2000(relevant sections)

Medical electrical appliances

and accessories

EU directive2002/96/EC

Disposal of waste, old electrical and

electronic equipment. Mandatory identifi cationsince August 13th, 2005

In accordance withDIN EN

60601-2-52(relevant sections)

Medical electrical appliances

and accessoriesMedical beds

DIN EN 980 Graphical symbols for marking medical

products

In compliance with EN12 530

Castors for nursing homebeds

Caution! A precise knowledge and observation of these instructions for use is a prerequisite for handling this product. They should therefore be kept safely until the product is disposed of. To avoid operating errors and to guarantee operation without malfunction, the instructions for use must be available to operating personnel at all times. It is recommended that they are stored in the bed or in the bedside cabinet.

2 In assembled state

7. Product Overview (Example Wood Décor T10)1. Head section2. Side guard3. Undercarriage4. Handset5. Foot section6. Backrest7. Thigh rest8. Calfrest9. Central castor brake

Page 8: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

8

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

8. Description Of Electrical FunctionsDepending on the model, the nursing home bed can be moved to the following positions by adjusting the backrest, the thigh rest and calf rest, and the height position:

Overview of handset pictogramsFig. 2

Caution! All electrical functions could have faults - unless you lock them and just release them for the duration of the function.

Caution! Other devices that emit electromagnetic radiation (e.g. radio telephones) can have an infl uence on electrical bed functions. For this reason in general it applies that bed functions not required have to be locked for reasons of safety. The use of such devices in the vicinity should be avoided as far as possible.

Caution! When moving the bed to different positions, it must be ensured that no persons, body parts or objects are in the adjustment area or between the head / foot end and the fl oor.

Caution! Since the system can be run via battery operation (optional), it is not enough just to pull out the plug in case of the motors functioning incorrectly. If an incorrect function occurs, please switch off functions via the control box and contact our Customer Service (battery operation option).

Page 9: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

9T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

8.1. Backrest Function (Br)

The adjustment is made via the corresponding buttons of the handset.The BR moves towards the head end of the bed during the upward motion. This function enables patients/residents to fi nd a comfortable position without the feeling of becoming restricted in the upper body area.The natural movement of the body is thus supported by this technology.

Caution in case of emergency lowering! Do not operate under load. When operating the mechanical emergency lowering (optional), the backrest must always be held at the mattress restricting bar and guided down slowly.

Recommendation: Handling can be simplifi ed by having the patient/resident fi rmly hold the lifting pole trapeze (if available) so that there is no load at all on the backrest.

Handling: Hold the BR in the set position and remove the load from it slightly. Turn the red lever (on the side of the motor unit covering) in the direction of the backrest.

Handling: Hold the TR in the set position and remove the load from it slightly. Turn the red lever (on the side of the motor unit covering) in the direction of the thighrest.

8.2. Legs Up Position (Lu) +The adjustment is made via the above mentioned buttons of the handset.

Caution! For reasons of safety, this position may only be released by medical staff, or only “under medical supervision“! The handset keys are released via the safety key!

Due to the intended use, a Trendelenburg position is not necessary!

Caution! The legs up position is not used for elderly care beds for acute care in an emergency/shock position. Primarily this function is used only for effortlessly setting the legs up position for therapeutic reasons!

To avoid the risk of getting into an unintended shock position, it is strongly recommended that the (BR) is raised approx. synchronously with the legs up position (preset). It is important here that the head of the patient remains higher than the orthostatic point of the body (the liver).

Page 10: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

10

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

8.3. Comfort Seating (Cs) (Function Not Available) +

The adjustment is made via the above mentioned buttons of the handset.

Caution! In conjunction with a bed extension there is a danger of entrapment between extension and fl oor.

Prior to each movement into the comfort seating position, the bed surface must be adjusted to a reasonable height! The same applies when a linen holder is used!

Caution! For reasons of safety, this position may only be released by medical staff, or only “under medical supervision”! The handset keys are released via the safety key! (Chapter “Rear Side of Handset with Locking Function“)

8.4. Auto Contour Function (Ac) +The adjustment is made via the above mentioned buttons of the handset.

8.5. Height Adjustment Function (Ha) +The height adjustment is made via the above mentioned buttons of the handset.

Caution! The height of the patient surface must be adjusted to the size of the resident/patient to avoid the risk of a fall.

Page 11: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

11T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

9. Safety Notes For Height Adjustment

Caution! Prior to every height adjustment of the bed, you must make sure that there are no objects, persons, animals or anything else in the danger area! Switch the handset to locking mode so that if handled by unauthorised persons there will

be no consequences (see chapter 14.2)!

Caution! If the occupied bed is unsupervised, it must be moved into the lowest position to reduce the risk of injury through falling when getting in or out of the bed or changing from the lying position!

Caution! Make sure that the handset cable does not lie within the moving range of the scissor-lifting mechanism, to avoid damage to the cable from it getting squashed! For this reason the handset cable (1) should always be guided over the longitudinal cross beam of the upper frame (2).

Page 12: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

12

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

10. Control Box (Optional)The electrical system represents state-of-the-art technology and meets the principles of initial fault tolerance. One of these safety precautions to guarantee initial fault tolerance is the control box. When using initial fault tolerance handsets, such as the Carisma 07 NPF with integrated locking mode, the control box is not necessary. Handling the locking device with the handset is described in the Handset chapters. Depending on requirements, the control box is located to the right or left of the patient surface.In case of malfunction of the electrical system, functions can be stopped via the corresponding rotary switch.

Handling:1. By turning the rotary switch to the “lock open“ symbol, the handset functions are released.2. By turning the rotary switch to the “lock closed“ symbol, the handset functions are locked.

11. DeliveryThe nursing home bed is generally completely assembled when delivered, or is assembled by specialists on-site.

The delivery papers should be used to make sure that all components have been delivered! Possible defects or damage should immediately be reported to the haulage company and to the supplier, and noted on the delivery note.

Caution! The handset should be in the so-called “Resident Mode“. This means that the comfort sitting position (CS) and legs up position (LU) are locked (Functions unavailable)!

To release the above-mentioned positions, rotate the locking switch on the rear of the handset (cf. chapter 14.2).

Page 13: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

13T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

12. AssemblyIf required, for example for the purpose of replacement, simple assembly procedures can be carried out by technically authorised persons.

12.1. Patient Surface1. Position the undercarriage.2. With the help of a second person, place the patient surface on to the lifting elements so that the

screws are at the height of the frame.3. Twist the remaining screws into the holes intended for them and screw them fi rmly to the frame.

12.2. Patient Surface Motor1. Placed the motor beneath the patient surface into the sockets provided for it and fi x in place with the

slider (Fig.4).2. Connect the slot connections according to the diagram shown below. (Fig. 5)

13. Transport / Moving BedBring bed into mobile position and move bed (see Chapter Brakes / Moving). Hold only the head and foot ends when moving the bed (see Fig. 2). Under no circumstances may the bed be pulled/pushed with the side rails!

Caution! When transporting the bed, do not push it over the handset cable or over the mains connection cable. Use the cable hook fi xed to the cable! When moving the bed, make sure that the feet do not get caught in the moving castors.

Page 14: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

14

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

14. Function TestBefore a nursing home bed is used for the fi rst time, several function tests must be performed and checks be made to ensure that the adjustments corresponding to the model have been achieved and are made without interruption.

These are explained in more detail in the following chapter.

14.1. Handset “Contempora 05“ Functional Description

Caution: set the locking switch before you start. (13.4)

Function 1 Backrest (BR) up / down

Function 2 Thighrest (TR) up / down

Function 3 Patient surface height

(HA) up / down

Arrow 4 Adjustment direction up

Arrow 5 Adjustment direction down

Diode (B) = green light fl ashing indicates currently activated function key Function key remains activated for approx. 10 sec. before moving automatically to locking mode

Diode (A) = Initial fault tolerance check If locking areas are released => LED lights up green

If power units are in operation => LED lights up red, as long as one of the action keys up/down (key 4 or 5) is pressed.

Error function is displayed if

a) red LED is permanently on

b) when key is pressed, if no LED display / beep occurs

c) Continuous beep

LED does not light up + beep when key is pressed (end position has been reached)

Page 15: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

15T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

14.2. Handset – Rear Side With Locking Function

Locking switch for various modes:

By turning the safety key, 3 modes can

be adjusted.

1. Locking mode = All functions are locked

2. Care mode = All functions are free to be used

3. Locking switch = Turn in desired direction

4. Resident mode = Functions 1 to 3 are free to be used

Replacing the handset holding hook:

Press holding straps inwards at points A and B and twist out handset holding hook upwards.

Caution! Be aware of any loose parts when replacing the handset holding hook.

Caution! Clip safety key into provided notch! Do not clip onto the ring below, since this could destroy the key!

Page 16: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

16

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

Initial fault tolerance of the handset:When operating the keys, a short signal can be heard. This means that the handset function is being performed correctly. If a continuous sound is heard when the key is not pressed, either the handset function is locked by the control box (optional) or there is an error. If there is an error here, please disconnect the bed from the mains supply (remove plug) and contact our Customer Service!.

14.3. Calfrest / MechanicalHeight adjustment of the calfrestThe height of the calfrest can be adjusted viaratchets.

Handling:

Raising:Lift the calfrest on the mattress restrictingbar (possibly on both sides) and allow it toclick into the desired position.

Lowering:Pull calfrest to highest position and let it downslowly.

14.4. Braking Capacity And MovabilityOperating the castor brake

Position (A) “STOP“The castor brake is in the braked position

Position (B) “RELEASED“The castor brake is in the released position

Position (C) “DIRECTION“The castor brake is set in a specifi c directionso that it is easy to manoeuvre the bed

Page 17: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

17T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

14.5. Adjusting Side Rails (Model 06)Side rails up/downOperation:

Lowering (A):Raise and hold top side rail in the recessedgrips as far as it will go. Pressslider (C) towards the middle of the bed,hold it and let down the side rail at thesame time.

Raising (B):Raise top side rail in the recessed gripsuntil slider (C) clicks into position.

Caution! The central locking mechanism of the side rail does not provide suffi cient side protection according to the standard. It only facilitates the resident’s orientation.

14.6. Adjusting Side Rails (Release Catch 09)Side rails up/down

Lowering (A):Pull up side rail (C) and simultaneously pull slider(D) towards the middle of the bed.Slowly let side rail glide into the lowest position (A).Repeat at footend of the bed.

Raising (B):Take side rail (C) from below and raise until slider(D) clicks audibly into place.

Caution! The central locking mechanism of the side rail does not provide suffi cient side protection according to the standard. It only facilitates the resident’s orientation.

Caution! Ensure that the locking mechanism of the side rail is in place by shaking it to and fro after adjustment

Page 18: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

18

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

Caution! When raising/lowering the side rails, make sure that no fi ngers, hands, or other body parts of the resident are in the adjustment area.

Caution! Restless or disoriented patients/residents could get trapped in the side rail. For this reason we strongly recommend that side rail cushioning is used. This protects the patient/resident from potential injury risks.

Caution! When using raised mattress systems, extensions to increase the height as well as the side protection cushioning available for this (accessories) as an important additional safety and protection feature are recommended.

14.7. Adjusting Side Rails (Model 10)Side rails up/down

Operation:Lowering (A):Push release buttons (E) simultaneously upwards.Then lower the handrail (D) of the side guard fromthe highest position (‘400 mm’) until it can be heardto click into the ‘360 mm‘ position, or until it engagesin the middle position. Stop pushing the release buttonsbefore this is done.

By pressing the release buttons (E) again, the sideguard can be lowered into the lowest position.

Raising (B):Pull the side guard upwards with the handrail (D) untilit can be heard to click into the middle position.Then press the release buttons (E) upwards simultaneouslyand raise the side guard until it can be heardto click into the desired position. Stop pushing therelease buttons before this is done.

Caution! Ensure that the locking mechanism of the side guard is in place by shaking the side guard to and fro.

Caution! If the bed is equipped with the Comfort patient surface, the side guard must always be set to the highest position (‘400 mm’).

Page 19: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

19T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

Caution! When raising/lowering the side rails, make sure that no fi ngers, hands, or other body parts of the resident are in the adjustment area.

Caution! Restless or disoriented patients/residents could get trapped in the side rail. For this reason we strongly recommend that side rail cushioning is used. This protects the patient/resident from potential injury risks.

15. Approved3 Accessories4

15.1. Lifting Pole (Accessories)The patient lifting pole is attached to the head end on the patient surface frame.

Handling:

1. Place lifting pole in the direction of the arrow into the socket (A) provided on the patient surface frame.2. Lock lifting pole by twisting it so that the bolt lies in one of the notches provided in the socket.

Adjustment of the lifting pole is possiblevertically (with tilting range) and horizontally.

Caution! The max. load capacity of the lifting pole is 750 N.

Caution! Since plastics are not unconditionally resistant to ageing, we strongly recommend that safety-relevant parts (in this case the grab handle and belt) are replaced with new ones after 4-5 years at the latest. The year of manufacture is imprinted on the product.

3 No liability is assumed for damage caused by the use of accessories not approved4 Approved accessories are those accessories recommended by the manufacturer

Page 20: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

20

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

15.2. Infusion Holder (Accessories)The infusion holder (like the lifting pole) is fi xedto the socket provided on the patient surfaceframe.

The infusion stand is only intended for attachinginfusions, not for attaching infusion pumps orsimilar.

Caution! The max. load is 8 kg (=2 kg per hook).

15.3. Patient Surface Extension (Option)The patient surface can be extended by 10 or 20cm at the footend. An additional extension can beattached to the calfrest.

Caution! The patient surface extension may only be assembled in the factory or by authorised specialist personnel!

15.4. Mattress (Accessories)Our Universal mattresses (dimensions: 87 x 200 x12 cm) have special incisions on their lower side.When the mattress is laid on the bed, it must beensured that the incisions match the rotationpoints of the patient surface. Only in this way canan optimum lying contour be guaranteed.

Caution! For safety reasons, the mattress must not be higher (thicker) then 12 cm and should not be thinner than 10 cm. The safety distances to the side rails must be maintained!

Caution! We recommend using mattresses that are manufactured with a min. volume weight of 40 kg/m3 (VW 40)!.

Page 21: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

21T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

15.5. Urine Bottle Basket / Urine Bag Container Model 09 (Accessory)The accessories holder SD09 (A) is attachedto the side rail or to the side panel (SD10).SD09 is additional secured from pulling offwith a rubber band.

Urine bag holders, bottle holders or handsetholders can be placed in this accessoriesholder.

Caution! Before you fi x the accessories holder, make sure that it cannot collide with other bed parts or fi ttings in the room.

Caution! The maximum permitted load on the accessories holder is 2 kg !

15.6. Handset Holder (Accessory)The handset holder (B) is used to place thehandset at a defi ned place. It is inserted intothe accessories holder (A), positioned on theupper side rail where it is suitable and fi xedwith the rubber band (C).

Caution! The handset holder is fl exibel. For this reason, it must not be used as assistance in getting up or as a handle.

The handset holder (B) is used to place thehandset at a defi ned place. It is placed intothe respective accessories holder (A) (SD10).

Possibilities for positioning see adjacenttable.

Caution! The handset holder is fl exibel. For this reason, it must not be used as assistance in getting up or as a handle.

Page 22: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

22

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

15.7. Side Rail Extension For 06/09 Side Rail (Accessories)1.Side rail2.Swivel arm3.Mushroom head bolt4.Clamping lever

Assembly:1. Pull up the side rails of the nursing home bed to the highest position.2. Release the clamping lever (4) by turning it 90° and twisting out the mushroom head bolt.3. Place the side rail extension onto the side rail of the bed as shown in the following fi gure.

Caution! The side safety clearance from the head and foot ends must each be approx. 6 cm. Rule of thumb: The clearance must be equally distributed at both ends. For nursing home beds of special length (longer than 2 m) a side rail extension of special length must be used.

4. Twist in the mushroom head bolt and screw the clamping levers tight (4) by turning them 90°.5. Perform a function test of the side rail, including the assembled side rail extension.If there are problems with the function, please contact our Customer Service.

Caution! You must pay special attention when adjusting the side rail. The risk of entrapment of body parts of the resident is especially high during adjustment procedures.

Caution! Placing the side rails at an inclination does not provide suitable or suffi cient protection from falling out of the bed.

Caution! For side rails with extensions, please use the available side protection cushioning (accessory), which is an important additional safety and protection feature

Page 23: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

23T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

16. Cleaning And Disinfection16.1. Initial CleaningPrior to fi rst-time use the nursing home bed must be cleaned in accordance with the IP protection level.

IP protection types in accordance with DIN 40050 / EN 60529The digits following the letters IP specify how the current-carrying or moving interior parts of an electricaldevice are protected from solids or water penetrating from outside.

Codes Protection level for protection fromcontact and impurities

Protection level for protection from water Cleaning and

Disinfection

IP

1. codenumber

Designation of extent ofprotection

2. codenumber

Designation of extent of protection

0 No protection 0 No protectionCleaning with slightly damp cloth, spray and

wipe disinfection

1 Protection from large (>50mm Ø ) impurities 1

Protection from vertically

dripping water

Cleaning with slightly damp cloth, spray and

wipe disinfection

2 Protection from mediumsized(>12 mm Ø ) impurities 2

Protection from obliquely dripping

water. (up to15° from vertical)

Cleaning with slightly damp cloth, spray and

wipe disinfection

3 Protection from small(>2.5 mm Ø ) impurities 3

Protection from spray water (up to 60° from vertical)

Cleaning with slightly damp cloth,spray and

wipe disinfection

4 Protection from grainshaped(>1 mm Ø ) impurities 4 Protection from

splashes of water

Cleaning with slightly damp cloth, spray and

wipe disinfection

5 Protection fromdust deposits 5 Protection from

hose water

With water and brush, without

hard jet of water

6 Protection from infi ltrationof dust 6 Protection from

strong water jets

With water and brush; stronger

jet of water; no high-pressure

cleaner; cleaning in washing units

permitted, depending on bed licence.

7Protection when

temporarily dipped in water

With water and brush; cleaning in washing

units permitted, depending

on bed approval.

Page 24: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

24

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

Caution! For safety reasons it is essential that the penetration or appliance of fl uids (e.g. water, solvents, oils, lyes, cleaning agents, etc.) into or on parts of the nursing home bed are avoided.

Caution! Housing that provides protection from the penetration of water does not necessarily provide the same protection from other fl uids!

Caution! For this reason the nursing home bed must generally be protected from any kind of dampness! Cleaning should only be carried out with a slightly damp cloth.

16.2. Disinfection / SterilisationThe nursing home bed must be disinfected regularly, and at least before every change of resident. The bed may not be disinfected in automatic washing units or with spray water. The following agents are recommended for the disinfection of the nursing home bed, and may only be replaced by equivalent agents: Terralin, Perform and Sagrotan Med (Schülke & Mayr GmbH, Norderstedt, Germany).

These agents may only be used in accordance with the manufacturer’s regulations! These refer particularly to the dosage of the agents and possible hazards in combination with other agents. the wooden lining must not come into contact with these agents.

16.3. Care Instructions For Wooden PartsWe have used materials of the highest quality to create a product for you with a high level of benefi ts and homely real wood character.

In order that you can gain pleasure for as long as possible from your Wissner-Bosserhoff product, we recommend that you clean with a slightly damp, soft cloth.

Each real wood material (veneer, solid wood) is subject to a natural ageing process, even when varnished. Over the course of time, environmental infl uences such as UV radiation, humidity or warmth cause changes to the colouring of the real wood surfaces. In addition, wood as a natural material has, by its nature, a wide variety of shadings. The textures of veneers can basically only be generally specifi ed, since each trunk of a particular species of tree has its own characteristics. Slight variations in colour and contrast in deliveries from Wissner-Bosserhoff are natural, and technically unavoidable. This is particularly the case for refurnishing or replenishment deliveries.

For the reasons mentioned, relative differences in tone, contrast and colour between veneer surfaces, solid wood parts or patterned surfaces (melamin, HPL), as well as growth-dependent inclusions in real wood materials, do not represent defects, and Wissner-Bosserhoff GmbH cannot assume any liability or guarantees for this.

Through careful selection, consulting and colour coordination of veneers and varnish, Wissner- Bosserhoff guarantees a high-quality in the coloration of products.

Page 25: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

25T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

16.4. Use Of Cleaning AgentsFor particularly stubborn stains, commercially available, environmentally friendly all-round household cleaners can of course be used. The furniture surfaces of Wissner-Bosserhoff products meet the requirements of standard DIN 68861-1A for furniture surfaces subjected to chemicals, and thus have an excellent resistance to normal pollution.

No cleaners or cleaning agents may be used, however, that contain mechanically abrasive componentsand that have the effect of scouring agents.

It must be ensured that no wetness or fl uid stains remain on the real wood and veneer surfaces when in use and when being cleaned. Wood is a natural product which has the property of absorbing moisture, and which expands in doing so. This can cause damage to the surface sealing.

For this reason the use and storage of wooden products under the infl uence of high levels of humidity must be avoided.

After cleaning, it is recommended that the real wood and veneer surfaces are wiped down with a dry cloth to avoid damage to the product through moisture as far as possible.

16.5. Use Of Disinfection AgentsOnly disinfection agents may be used that correspond to the disinfection agents described in Appendix 1 of DIN 68861 Part 1:

a) Phenolderivates, 0.5-% aqueous solution: Chlorinated alkyl aryl-phenol, cycloalkyl- aryl-phenolb) Chloramin T, 2.5-% aqueous solution: p-Toluolsulfonchloramide-sodium

16.6. Damage To Wooden PartsIf the veneer or surface sealing is damaged by bumps, scratches or incisions, the surface must immediately be protected from the penetration of humidity by suitable repair materials. Please contact the Wissner-Bosserhoff Customer Service or a specialist company.

16.7. Environmental CompatibilityAll wooden materials used comply with strict German laws and regulations and are harmless in terms of health. The focus of our operation is on people and their environment.

16.8. QualityOur products are subject to strict controls through our quality assurance system in compliance with international standards DIN EN ISO 9001:2000 and DIN EN ISO 13485: 2003 for the manufacture and distribution of medical products.

For all properties guaranteed, Wissner-Bosserhoff GmbH assumes a one-year manufacturing warranty. Claims that originate from improper or non-standard use or care are excluded from this. If you have grounds for complaint, please contact our Customer Service.

Page 26: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

26

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

17. Help In Case Of Malfunctions

Malfunction Possible Cause RemedyElectric motor adjustments notmade correctly • Mechanics are locked Check moving parts and remove

any foreign bodies

Electric motor adjustments arenot performed • Handset is defective Contact our Customer Service

and request spare part

• Bed not plugged in to power supply Connect plug to power supply

• The system was overloaded and temperature safeguard has switched off

Wait for cooling down phase (at least 20 min., depending on overload). If the system still does not function, contact our Customer Service

The side rail cannot be properlyadjusted • Mechanics are locked Check moving parts and remove

any foreign bodies

• Mechanical parts are bent Contact our Customer Service

Castors do not brake or donot roll

• Impurities have become caught in castors over time Remove impurities

• The castor system is defective Contact our Customer Service

18. Storage

The following activities must be performed when storing the bed:1. Remove mains plug.2. Remove accessories such as lifting pole, serving tray, etc.3. Cover nursing home bed so that damage or scratches to paint cannot occur.4. Apply date of storage (for the purpose of maintenance intervals).

Caution! For storage, the same conditions apply as for the working environment.

19. Safety Notes And Residual Risks1) These instructions for use must be read and observed before the nursing home bed is used.2) The details of the nameplate must be observed.3) The bed may not be operated if it displays malfunctions that could cause a hazard to residents,

employees or third parties.4) The bed may only be operated by persons who can guarantee correct handling because of their

training or knowledge and practical experience.5) Before operating the bed, the person doing this must be satisfi ed with the functional safety and the

proper condition of the bed.6) The bed must be subject to a function test, possibly daily or with each change of shift, so that it is

guaranteed that the bed can be used as intended without danger for the patient or the operator.

Page 27: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

27T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

7) Do not repair and reuse lifting poles that have become distorted, but replace them.

8) The bed may only be moved on level and fi rm fl oor covering.

9) If the bed is occupied the castors must always be in braked position. Otherwise there is a risk of the patient falling when getting in or out, since he may support himself on the bed.

10) The height of the patient surface must be adjusted to the size of the patient to avoid the risk of a fall.

11) When operating the mechanical emergency lowering of the backrest, (optional) this must always be fi xed to the mattress restricting bar or its pole frame to avoid a sudden lowering.

12) When moving the bed to different positions, it must be ensured that no children, body parts or other objects are in the adjustment area.

13) Adjustments to the side guards (side rails) may only be made by nursing staff. When adjusting the patient surface positions, nursing staff must be satisfi ed that the patient will not come into contact with the side rails, to avoid entrapment of hands, legs or other body parts.

14) Extreme leaning out of the bed sideways with the upper body must be avoided.

15) Do not put a load of more than 75 kg in a vertical direction and more than 50 kg in a horizontal direction on the side rails .

16) Depending on the physical and mental state of the patient, functions must be locked or released on the handset via a control box. The effectiveness of the locking functions must be tested with the handset.

17) Since the system can be run via battery operation, it is not enough just to pull out the plug in case of the motors functioning incorrectly. If an incorrect function occurs, please switch off functions via the control box and contact our Customer Service (battery operation option).

18) For adjustments via the handset, it is recommended that the operator remains in the vicinity of the control box so that possible incorrect functions can be immediately switched off.

19) All electrical functions can be faulty – unless they are locked through the control box.

20) For the power source, a 220 / 230 volt, 50 Hz mains plug in accordance with VDE stipulations must be used.

21) Do not touch the plug with wet hands. Only hold the mains connection cable on the plug and pull out from the mains socket.

22) Lay the mains connection cable without loops, do not bend or pull around sharp edges.

23) Generally protect connection cable and supply cable from mechanical stress.

24) Position the mains connection cable and supply cable to the handset so that they are safe from being squashed and do not place any objects on the connection or extension cable.

25) When setting up the bed in the room provided for it, safety clearances from the wall, windowsills and other objects must be maintained. Safety clearances depend on the design and model of the bed, for which extreme height adjustment and tilting movements are performed and at least 25 mm space for fi ngers must be maintained.

26) If the bedding and height adjustment functions are performed, it must be ensured that the safety key is not in the lock or in the vicinity of the handset (in the range of the patient/ resident). Through improper release via the safety key, positions can be achieved that may represent a risk to the patient / resident, such as a wrongly applied Trendelenburg position.

Page 28: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

28

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

27) Place the handset in the holder provided for this (accessory) or on the bed so that the keys can not be pressed accidentally, e.g. between side rail and bedside cabinet.

28) Cleaning of the bed may only be carried out with a damp cloth.

29) If the bed is not used properly, dangers can occur through:

• Unauthorised operation of electrical functions during the uncontrolled positioning of the bed,

• Using the bed for children under the age of 12,

• Operating functions of the bed by the patient without previous instruction,

• Simultaneous operation of electrical functions by different people,

• Connecting electrical devices to the bed that are not intended,

• Moving the bed by pulling the mains cable or side rails,

• Pulling the mains cable to remove the mains plug from the power supply,

• Lack of protection for the connection and supplied cables from sharp edges or other mechanical strains (e.g. running over the mains cable),

• Moving the bed on an unstable surface,

• Using the bed on a sloping surface with an angle of 10° or more,

• Devices that create strong electromagnetic fi elds and that may possibly infl uence the control of the bed are not permitted in the direct vicinity of the bed (e.g. mobile phones).

• Before cleaning or servicing the bed the mains plug must always be removed from the power supply.

• For safety reasons the lifting handle (trapeze) together with the strap must be completely replaced after 5 years at the latest,

• The mattress to be used must meet valid safety standards and may not be thicker than 120 mm,

• Continuously dripping fl uid in the motor unit area (e.g. in case of incontinence),

• Improper repair work on electrical installations.

30) Safety notes – measures for the safe operation of hospital and nursing home beds:

• Electrical components on hospital and nursing home beds may only be repaired and serviced by the manufacturer or by specialist personnel trained specifi cally by the manufacturer. Otherwise there is a danger that specifi c procedures will not be performed directly through lack of knowledge of product functionalities, leading to an increased risk of electrical shock or fi re.

• Electrical components on hospital and nursing home beds may be replaced only with the original spare parts from the manufacturer, since replacement with incorrect components can lead to a highly increased risk of electrical shock or fi re.

• A safe location on the bed must be specifi ed for the hand controls, so that playing with the hand controls can be prevented.

Page 29: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

29T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

• Cleaning the bed may only be carried out in accordance with the specifi cations in section “Cleaning and Disinfection” of these instructions for use, since increased risks could otherwise occur for which the manufacturer can take no responsibility.

• If incontinent patients occupy the bed, an incontinence protection for the mattress must always be used.

• If immobile patients/residents lie in the bed permanently, bed sores can occur if additional lying aids (e.g. pressure sore mattresses) and other special nursing measures for pressure sore prevention are not employed. Under no circumstances is this the responsibility of the manufacturer of the bed.

• The product is explicitly not suitable for permanent use. If the specifi ed operating time is exceeded, the power unit can heat up, causing it to switch off automatically. The power unit must then cool down for at least two hours, and can only be put into operation again after this time has expired.

• It is absolutely essential that a blockage of the bed or of parts of the bed mechanics is avoided, since this can lead to damage and to the total failure of the power unit through overheating.

• It is absolutely essential that the safe working load is not exceeded, since this can lead to damage and to the total failure of the power unit through overheating.

31) Servicing notes – Technical Safety Inspections – notes on rectifying errors:

• Electrically-operated hospital and nursing home beds are medical products and are subject to the so-called Technical Safety Inspections (TSI) in accordance with the German Law on Medical Products (MPG) and the Medical Products Operator Decree § 6 (MedProdBetrV). This regular TSI should be carried out at least once a year. When this is done the functional and electrical safety in accordance with VDE0751, including a visual and function inspection (inspection of the resource in operation) must be performed, as well as an inspection of possible measuring functions.

• In general the accessories and other devices that are combined with the product must also be inspected for technical safety. It must be ensured when doing this that only accessories authorised by the manufacturer and permitted combinations of appliances are used. These TSIs may only be undertaken by specialised electricians or persons instructed in electrotechnical matters using a special measuring and test device, who have been trained specifi cally by the manufacturer for the product and the permitted accessories and combinations of appliances. The person performing the inspections may not accept any instructions from the operator. Special test records must be written for each of the TSIs. The product itself should be marked with a test sticker displaying the next test date.

• Furthermore, electrically-operated hospital and nursing home beds are electrical resources, for the safety of which the employer is responsible. The monitoring function for this obligation is the responsibility of the German professional association for health service and welfare care (BGW) and the German industrial inspectorate. The regulations of the professional associations for Health and Safety at the Workplace (abbreviated BGV, formerly VBG) apply. In particular, BGV A3 (formerly

Page 30: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

30

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

• VBG “Electrical Facilities and Resources”(BGV A2)) applies to repeat inspections of movable electrical resources with a guideline of six months, but at least once per year. These inspections may only be performed by a qualifi ed electrician or a person instructed in electrotechnical matters, using a special measuring and testing device. The inspections in accordance with BGV A3 may also be performed by specialist staff trained by the manufacturer, in the context of the Technical Safety Inspections for Medical Devices, since these BGV inspections form a part of the Technical Safety Inspections.

• Electrically-operated hospital and nursing home beds are active medical products and must be maintained, in accordance with the German Medical Products Operator Directive (MedProdBetrV) in an inventory sheet for each premises. It is recommended that the correct implementation of the stipulated TSIs is also documented in this inventory sheet and that the next test date is specifi ed. The required records for the technical safety inspections already performed should be attached to the inventory sheet.

• The proper implementation and traceable documentation of the technical inspections, maintenance and service work specifi ed by the manufacturer, as well as the technical safety inspections, is a necessary prerequisite on the part of the purchaser that he may acquire rights of guarantee. If the operator of a medical product does not meet his obligations, risk of damage and accident may arise from this that are explicitly not the responsibility of the manufacturer.

• Servicing work must be performed in accordance with VDE0751-1 and be concluded with a documented technical safety inspection.

20. Maintenance And ServicingThe nursing home beds of Wissner-Bosserhoff GmbH are generally designed for a useful life of 10 - 15 years. The beds here are extremely low in maintenance. This is because attention was paid during the development of the products to ensure that maintenance work was reduced to a minimum, with low operating costs.

In daily use, however, experience shows that there is also careless handling of products, and rough handling can also contribute to them ageing quicker and to the wear of certain components, on which the manufacturer can have no direct infl uence. For this reason, routine servicing examinations should be performed by the operator, also to guarantee the availability of beds, which is in his own interest.

The Technical Customer Service of Wissner-Bosserhoff GmbH offers servicing and necessary training covering all aspects of the product. Wissner-Bosserhoff certainly places great emphasis on training for customers, since only trained operators and maintenance technicians can ensure proper handling and use of the beds.

Maintenance should be performed exclusively by competent specialists who have been trained for the product by Wissner-Bosserhoff.The manufacturer assumes liability for the safety and reliability of the product only if it is regularly maintained and used in accordance with the operating, usage and safety notes in these instructions for use. If a function control, an inspection, measurement or maintenance reveals serious defects that cannot be rectifi ed, the product must be suspended from further use.

Page 31: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

31T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

Wissner-Bosserhoff GmbH recommends a simple mechanical servicing at yearly intervals. In general the manufacturer assumes a servicing in accordance with the currently valid rules of technology, since a nursing home bed is equipment that is subject to the accident prevention regulations of the professional association. For this reason, BGV A3 in conjunction with the VDE 0751 standard applies for safety inspections.

Our Customer Service is available to you for any enquiries and to provide training and test lists for individual products.

20.1. Completeness And AccessoriesIt must be determined, by means of the delivery note, the instructions for use and a visual inspection, whether parts are missing, and therefore need to be ordered or replaced. In general the accessories and other devices that are combined with the product must also be checked. It must be ensured when doing this that only accessories approved by the manufacturer and permitted combinations of appliances are used.

20.2. Wear And TearThe individual components of the nursing home bed must be examined for scouring, scratches or other types of wear in the context of an inspection and function control, and the cause must be determined and rectifi ed by the replacement of the individual parts involved. It must be possible to perform all locking mechanisms and adjustments in the correct manner. Pay particular attention to safety-relevant parts of the nursing home beds.

20.3. FunctionalityTest of whether all adjustments can be made to their maximum positions. Perform a function control using the sections “Function Control” and “Description of Nursing Home Beds - Functions” in these instructions for use.

20.4. Bgv A3/Vde 0751– Inspections Of The Electrical Components And The Handset • Test of whether the power units move in and out correctly, and whether the unit and handset

remain in fl awless condition with respect to: • Checking of mechanics (screws, bolts etc. are tight, parts, labels etc. undamaged) • Condition of cables, housing, etc. (cables crossed, squashed, etc.) • Function Test • Visual inspection • Suffi cient cable relief and protection from buckling • Discharge current test • PE conductor test for models in class of protection I • Insulation test if necessary

Page 32: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

32

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

20.5. CastorsTest of functionality of castors (moving – braking) and the removal of possibly accumulated dust, threads, hair or similar that may have been caught in the castors. If necessary, the castor brakes must be readjusted. Please contact the Wissner-Bosserhoff Customer Service.

21. Useful LifeIf used, cleaned, serviced and repaired correctly, a useful life for our products of 10 - 15 years can be expected. Excluded from this are wearing parts such as castors, pneumatic springs, electrical components, etc.

22. DisposalThe bed may optionally contain lead batteries, metal, wood and plastic parts made of ABS, PA, PUR, PE. (cf. scope of delivery and model).The metal and plastic parts that accumulate during servicing and repairs must be disposed of properly and professionally in accordance with valid laws and regulations. Specifi cally the electric motors and electrical controls may only be disposed of through specialist fi rms authorised to do this.

Caution! Electrical Waste!Follow legal requirements for disposal of old equipment

23. Spare Parts, Spare Part Lists And Service ManualsOnly Wissner-Bosserhoff original parts may be used.

Customer Service, Sales and the technical consultation department provide information concerning spare part deliveries, etc. (for the address, see Contact Partners).

Spare part lists, current price lists and servicing instructions with exploded diagrams can be requested if required from Wissner- Bosserhoff, Technical Customer Service, by specifying the nameplate data of the bed or the relevant article number, order number and delivery date.

Customer Service Phone: 08 9351 7500

Page 33: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

33T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

24. Nameplate And LabelsIn order to handle enquiries or spare part orders, details of the nameplate are required.

An example explanation of the details on the nameplate can be found in Appendix B of these instructions for use.

The nameplates and labels of individual components (e.g. motor, pneumatic spring, hydraulics, handset) are also very useful

25. Guarantee And ServiceWith the “movita“ nursing home bed you have bought a Wissner-Bosserhoff quality product. This bed was manufactured with great care using high-quality materials and modern production techniques.

The “movita” nursing home bed has a 36-month guarantee

from the date of purchase. This guarantee covers all malfunctions and defects in materials and manufacture. Malfunctions and defects arising from incorrect handling and external infl uences are excluded from this. If there should be cause for justifi ed complaints within the guarantee period, however, these will be rectifi ed free of charge. This guarantee can be claimed with the purchase slip, which bears the date of purchase. Our general terms and conditions of delivery apply.

26. Contact PartnerManufacturer:Wissner-Bosserhoff GmbHHauptstr. 4-658739 Wickede (Ruhr), Germany

For enquiries we can be reached at the following:Tel.: 08 9351 7500E: [email protected]

If you make enquiries, we would like to ask you to prepare for a phone conversation by noting the details on the delivery notes and the nameplates (see Appendix B of these instructions for use) so that the Wissner-Bosserhoff customer consultant can be notifi ed of these. If these data are not available, guarantee services in particular cannot be treated free of charge.

Page 34: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

34

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

Service PartnersPlease stick reminder sticker here

Page 35: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

35T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

Page 36: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

36

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

Appendix A: Spare Parts List Movita 09/10

Page 37: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

37T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

Enter important ordering details from the bed nameplate!Article No.Series No.Enter important ordering details from the control motor name plate!Article No.Series No.

Item Designation Parts No.

Electronic components

9.1 Motor Quadromat 7 care 32 V 01-004644

9.2 Motor Quadromat 7 care 32 V with emergency lowering BR and TR 01-004766

9.3 Motor Quadromat 7 care 24V with charging PCB for battery 01-003274

9.4 Motor Quadromat 7 care 24 V with emergency lowering BR and TR 01-004765

9.5 Motor Quadromat 7 care 32 V with emergency lowering model Switzerland 01-006057

9.6 Battery for 24 V motor 01-002582

8 Handset with 5 keys 01-005203

10 Megamat 2 (undercarriage motor) 01-007656

Wood linings

“movita“ (divided side guards)

1 Head end W10 divided side guards 04-0833

2 Foot end W10 divided side guards 04-0818

1 Head end K10 divided side guards 04-0666

2 Foot end K10 divided side guards 04-0668

1 Head end T10 divided side guards 04-0669

2 Foot end T10 divided side guards 04-0670

1 Head end S10 divided side guards 04-0675

2 Foot end S10 divided side guards 04-0676

1 Head end G10 divided side guards 04-0685

2 Foot end G10 divided side guards 04-0686

1 Head section D10 divided side guards 04-0950

2 Foot section D10 divided side guards 04-0951

“movita“ (continuous side guards)

1 Head end W06/09 continuous side guards 04-0834

2 Foot end W06/09 continuous side guards 04-0816

1 Head end K06/09 continuous side guards 04-0651

2 Foot end K06/09 continuous side guards 04-0652

1 Head end T 06/09 continuous side guards 04-0635

Page 38: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

38

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

2 Foot end T 06/09 continuous side guards 04-0636

1 Head end S 06/09 continuous side guards 04-0671

2 Foot end S 06/09 continuous side guards 04-0672

1 Head end G 06/09 continuous side guards 04-0680

2 Foot end G 06/09 continuous side guards 04-0681

1 Head section D 06/09 continuous side guards 04-0944

2 oot section D 06/09 continuous side guards 04-0945

Divided side guards

3 Hand rail, divided side guard 04-0565

5 ide panel, divided side guard 04-0769

Side guard 06

3 Top side rail 06 04-0508

4 Lower side rail 06 04-0509

Side guard 09

3 Top side rail 09 04-0637

4 Lower 09 side rail 04-0509

Other components

11 Spring element 72 Shore, complete 02-000316

12 Spring element 68 Shore, complete 02-000315

13.1 Backrest lattice contempora 05-0389

13.2 Plastic supports, backrest contempora 01-005658

14 Seat section contempora 05-0385

15.1 Thigh rest lattice contempora 05-0390

15.2 Plastic supports, thighrest contempora 01-005659

16.1 Calfrest lattice contempora 05-0391

16.2 Plastic supports, calfrest contempora 01-005660

17 Side guard SD 10 without hand rail, left-head end/right-foot end 01-000447

18.1 Brake lever right 30° 01-007709

18.2 Brake lever left 30° 01-007710

19.1 Castor Ø100 Decora, directional castor 01-007657

19.2 Castor Ø100 Decora, without directional castor 01-007658

19.1 Castor Ø125 cover design, directional castor 01-007659

19.2 Castor Ø125 cover design, without directional castor 01-007660

19.1 Castor Ø150 cover design, directional castor 01-007783

19.2 Castor Ø150 cover design, without directional castor 01-007784

19 Castor Ø100 individual braking capacity 01-003742

Spare parts sets

Set of 5 handset spanners 10-0392

Page 39: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

39T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

Appendix C: Important Safety And Usage Notes For Nursing Home BedsINTENDED USE AND PURPOSE OF NURSING HOME BEDS A nursing home bed may only be operated and used in accordance with its intended purpose,

in compliance with the regulations of the German Law on Medical Products (MPG), the accompanying orders, the generally recognised rules of technology and the regulations on work safety and accident prevention. A nursing home bed may not be operated in a defective condition in which the bed could be a hazard to patients, nursing staff or third parties.

The conditions of the respective instructions for use of the product shall apply.

The following points must be observed: The nursing home beds of Wissner-Bosserhoff GmbH are intended exclusively for the occupancy

of persons from the age of 12 and above. This must be ensured, since persons with a body size or weight deviating from the normal average are exposed to an increased risk under certain circumstances. As the recommended standard range, persons in a range from a woman of 146 cm in height to a man of 185 cm in height (with a non-critical state of health and capable of acting independently at least to a certain extent) can occupy the bed without additional risks or restrictions in performance (in accordance with Standard EN 60601-2-38:1996 + A1:2000). Beyond this range, additional accessories must be used if necessary, or use is not possible. Bed extensions

Appendix B: Nameplate – Explanation Of Details Using An Example

Page 40: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

40

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

(accessories) should be used for larger persons, and it is essential that additional protective measures such as side rail protection covers (accessories) are used for emaciated persons. Individual application cases that deviate from this must be co-ordinated with Wissner-Bosserhoff GmbH.

• In general, the professional judgment of the nursing person responsible (medical supervision) is necessary to guarantee safe use of the nursing home beds. The physical, psychological and medical needs and conditions of patients must be observed. Accordingly the bed functions must be individually released or locked for each patient/ resident. This applies particularly to products with special tilting options for the patient surface, and when using side guards.

• The maximum workload of this nursing home bed is 205 kg. Higher loads are possible and must be taken from the instructions for use for the respective product! The bed may not be used for patients with a weight of more than 155 kg. The safe working load of 205 kg includes the patient weight, the mattress and the accessories located on the bed (lifting pole, linen etc.)!

• The operation of nursing home beds is allowed exclusively from mains sockets installed in the building that are up-to-date and that are inspected regularly for the correct function of safety installations (RCD protection, insulation error detectors, power fuse, etc.) in accordance with valid regulations.

• The nursing home beds are not protected from explosions and may not be used in application rooms in which fl ammable anaesthetics, cleaning or solvent agents or possible ignitable substances mixed with air and can be found. The regulations of the professional associations must be observed in this respect.

• Possible use deviating from this intended purpose represents a non-intended use and is excluded from all claims of liability.

Basic Functions And Uses Of Nursing Home BedsModern nursing home beds have numerous bedding functions, which are described in detail in the instructions for use for the respective product. The full benefi ts of a nursing home bed can only be achieved through careful training of the users and inhabitants of a nursing home. Please repeat these training sessions as regularly as possible and pay particular attention to the necessary safety regulations that must be observed.

In general, there is always a risk of unintended or wrong adjustment of the bed. For example, pressing switches accidentally or unintended release of functions can be a hazard to certain patients/residents! To ensure a high degree of protection, it is absolutely essential to use locking facilities and/or the release keys selectively. Under no circumstances may release keys remain on the handset or on the bed! Release keys are intended exclusively for use by nursing staff.

The use of the locking facilities and if required, the intentional separation of the bed from the power supply – as an absolute locking measure for the maximum protection of a possibly jeopardised patient / resident– lies exclusively at the discretion of and in the responsibility of the medical supervision and nursing staff in charge. Corresponding measures should be noted in the shift report to guarantee a correct shift handover.

Page 41: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

41T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

The use of special positions, such as Trendelenburg, anti-Trendelenburg, layered bedding or relative lowered head position as a shock position is explicitly allowed only by medical supervision staff. These functions must be locked for patients and residents, since this could otherwise result in hazards.

The safety notes in the specifi c product instructions for use must be observed, since hazards could otherwise occur.

When using the side guards, particular care and caution should be taken:

- Side guards are used fi rst and foremost to protect the patient from rolling out. On no account may the side rails be used as a means to hold the patient fi rm.

- As soon as the mobility of the patient/resident is restricted, a court order is required to use the side guards. Only divided side guards or side guards that can be individually lowered, in which only the head and section is assembled, and which can be used as a help in getting in and out, are just one permissible exception for mobile residents/patients here.

- Overloading caused by sitting, leaning or pulling on the side rail can result in serious damage to this and cause hazards. It is essential that such overloading is avoided.

- When using side guards, their suitability should be checked, taking into account the peculiarities of the respective patient/resident. In particular the clearances between the panels and supports in relation to the body shape of the patient/resident must be taken into account, so that entrapment or slipping through is ruled out. If necessary, additional side protection cushioning must be used which reduces the size of the gap or closes it. This decision is exclusively the responsibility of the medical supervision in charge. The details on the restriction of the intended use in these instructions for use are referred to as an orientation, as well as the specifi c product instructions for use for the bed and for the side rail.

- Only side guards may be used on the beds which are also permitted for that particular bed type, since increased risk of accident could otherwise occur.

- The side rails/side guards must be checked each time before they are used for possible damage, and checked that they are securely attached to the bed and that the safety lock functions correctly! Details on the correct handling of the side guard can be found in the respective product instructions for use.

The use of the manual remote control by the resident/patient himself lies exclusively at the discretion of and in the responsibility of the medical supervision and nursing staff in charge. Corresponding measures should be noted in the shift report or in the nursing documentation to guarantee a correct shift handover and traceability of usage rights.

If the manual remote control is not to be used by the resident/patient himself, then it must be protected from being accessed by this person

Page 42: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

42

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

The manual remote control must not fall on the ground, be subject to heavy weight or have liquids poured on it. It should be cleaned with cleaning cloths used for computer keyboards. If there is damage to the mechanics or from dampness, the remote control must be replaced immediately. The cables must also be checked at frequent regular intervals.

The mains connection cables must be protected from damage during operation and when the bed is transported/moved:

- The cable must be safely prevented from being badly twisted, pulled, bent, folded or even squashed. When moving the height adjustment of a bed, make sure that the connection cable is free and can freely follow the movement of the bed at all times. The cable may not be stretched tight. The cable may not become trapped, for example under or behind another item of furniture or under the legs/castors of the bed. Otherwise damage could be caused to the mains supply cable or to parts of the covering. For this purpose, unused mains cable should be folded together in a fi gure of eight loop and held together with a strong rubber band. Alternatively the Wissner-Bosserhoff GmbH mains cable hook can be used (see images). Under no account may fi xed cable binders be used, since these could damage the mains cable if it is subject to unintentional tension.

- The mains cables must undergo a precautionary visual inspection at brief intervals, for example if mechanical load has been detected, if the cable is relayed or monthly, so that any hazards can be ruled out. The checklist in Appendix A can be used here as a guideline. In particular the critical points of inserting the mains plug, the mains plug cover and protecting the cable from traction or houding openings on the bed must be observed.

Page 43: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

43T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

It is essential that the cable is prevented from being dragged across the fl oor or even being squashed or run over with the bed castors when the bed is transported. Alternatively the Wissner-Bosserhoff GmbH mains cable hook can be used. (see following image).

Page 44: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

44

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

Use Of Original Accessories And Released Product Combinations The use of accessories and combinations of products not authorised for the product can

lead to serious hazards. This is why only accessories released by Wissner-Bosserhoff GmbH and products released for combination may be used.

Examples:

• Using lifting poles with an insuffi cient diameter can lead to the failure of the lifting pole sockets.

• The use of very heavy water mattresses leads to the safe working load being exceeded. This can cause failure of components, including the power units. Hazards arise from this that can lead to accidents.

Setting Up The Nursing Home Bed (Safety Clearances) The bed must be set up in such a way that when implementing all adjustment functions

of the bed, collision of bed parts with other objects in the room or with parts of the room is not possible (e.g. mains cable, windowsills, slanting ceilings, pipes, surface-mounted installations, wall boards etc.). A safety clearance from walls and other objects of at least 2.5 cm should be maintained in all possible bed positions when doing this. Caution: Equally no collision may occur for tilting and height adjustments!

Cleaning And Disinfection Of The BedThe nursing home bed must be cleaned and disinfected regularly, and at least before every change of resident. The nursing home bed may not be cleaned or disinfected in automatic washing units, nor with splash water, jet water or water spray. Only spray and wiping disinfection is permitted.

The following agents are recommended for the disinfection of the nursing home bed, and may only be replaced by equivalent agents: Terralin, Perform and Sagrotan Med (Schülke & Mayr GmbH, Norderstedt, Germany). These agents may only be used in accordance with the manufacturer’s regulations! These refer particularly to the dosage of the agents and possible hazards in combination with other agents.

The wooden lining must not come into contact with these agents. For particularly stubborn stains, commercially available, environmentally friendly all-round household cleaners can of course be used.

No cleaners or cleaning agents may be used, however, that contain mechanically abrasive components and that have the effect of scouring agents.

It must be ensured that no wetness or fl uid stains remain on the real wood and veneer surfaces when in use and when being cleaned. Wood is a natural product which has the property of absorbing moisture, and which expands in doing so. This can cause damage to the surface sealing. Damage to varnished services should be rectifi ed after cleaning with appropriate repair varnish. Please ask Wissner-Bosserhoff Customer Service.

Page 45: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

45T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

Manual remote controls may only be cleaned with cleaning cloths for computer keyboards. If dampness has penetrated, if the unit is heavily soiled or if the keys stick due to dirt particles, the hand controls must be replaced if necessary.

Appendix D: Battery Operation (Option) - Additional NotesAs a special fi tting, the bed power unit can be equipped with an additional battery function allowing the bed to be moved multiple times with its electrical functions independently of the mains supply.

The number of adjustments possible depends on the charge status of the battery. To prevent the batteries from excessive discharge, the bed should remain connected to the mains if possible.

Battery charging will then take place fully automatically via a specially integrated PCB.

Normally the battery is delivered fully loaded and mounted on the main power unit; assembly by an untrained operator is not intended.

Important features:

• The battery is maintenance-free • Permanently installed in the connection box • Lead gel battery – can be operated irrespective of its location • Nominal data: 24V, 1.2Ah • Lifetime of battery: at least 5 years • Disposal permitted only at special collection points

For safety reasons, the battery box must not be opened.

If you want to acquire the battery later on, we urge you to contact Technical Customer Service, since not every power unit is suitable for combination with the battery. Replacement of the battery box should only be performed by trained personnel to avoid possible damage through incorrect handling.

The lifetime of the battery is strongly dependent on usage and on the fact that possible harmful excessive discharge may have occurred.

It is also possible to use a fourfold external charging unit (accessory for parallel charging of four battery boxes), which can be obtained separately: We strongly urge you to observe the separate instructions for use for the battery charging unit in such a case.

Page 46: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

46

Wissner-Bosserhoff Movita Bed, Instruction & Safety Manual

T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

Product Identifi cation

General Warnings!

Follow the instructions for use!

Caution! Possible cable entrapment hazard!

Caution! Possible entrapment hazard – Feet!

Caution! Possible entrapment hazard – Hands!

Caution! Electrical Waste!

Page 47: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning

47T 08 9351 7500 E [email protected] W www.unicarehealth.com.au

Page 48: Wissner-Bosserhoff Movita Bed Instruction & Safety Manual and Safety Man… · T 08 9351 7500 E info@unicarehealth.com.au W 5 This nursing bed is a medical product in the meaning