Page 1: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

X-Ray Production & Emission

Page 2: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Objectives: Review x-ray production

requirements

X-ray tube interactions

X-ray emission spectrum

Page 3: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Principle Parts of the X-ray Imaging System Operating Console

High-voltage generator

X-ray tube

The system is designed to provide a large number of e- with high kinetic energy focused to a small target

Page 4: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

E- traveling from cathode to anode

Projectile e- interacts with the orbital e- of the target atom. This interaction results in the conversion of e- _______ energy into ________ energy and ________ energy.

Page 5: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Tube Interactions 3 possible tube interactions

Tube interactions are generated from _____ slamming into ________?

Heat (99%), EM energy as infrared radiation (heat) & x-rays (1%)

X-rays = Characteristic (20%) or Bremsstrahlung (80%)

Page 6: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Heat Most kinetic energy of projectile e- is

converted into heat – 99%

Projectile e- interact with the outer-shell e- of the target atoms but do not transfer enough energy to the outer-shell e- to ionize

Page 7: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Heat is an excitation rather than an ionization

Page 8: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Heat production Production of heat in the anode increases

directly with increasing x-ray tube current & kVp

Doubling the x-ray tube current doubles the heat produced

Increasing kVp will also increase heat production

Page 9: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Characteristic Radiation – 2 steps Projectile e- with high enough energy to

totally remove an inner-shell electron of the tungsten target

Characteristic x-rays are produced when outer-shell e- fills an inner-shell void

All tube interactions result in a loss of kinetic energy from the projectile e-

Page 10: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

It is called characteristic because it is characteristic of the target elementin the energy of the photon produced

Page 11: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Only K-characteristic x-rays of tungsten are useful for imaging

Page 12: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Bremsstrahlung Radiation Heat & Characteristic produces EM energy

by e- interacting with tungsten atoms e- of the target material

Bremsstrahlung is produced by e- interacting with the nucleus of a target tungsten atom

Page 13: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Bremsstrahlung Radiation A projectile e- that completely avoids the

orbital e- as it passes through a target atom may pass close enough to the nucleus of the atom to convert some of the projectile e- kinetic energy to EM energy

Because of the electrostatic force?

Page 14: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Bremsstrahlung

is a germanword meaningslowed-downradiation

Page 15: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

X-ray energy Characteristic x-rays have very specific

energies. K-characteristic x-rays require a tube potential of a least 70 kVp

Bremsstrahlung x-rays that are produced can have any energy level up to the set kVp value. Brems can be produced at any projectile e- value

Page 16: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Discrete spectrum Contains only specific values

Page 17: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Continuous Spectrum Contains all possible values

Page 18: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Characteristic X-ray Spectrum Characteristic has discrete energies based

on the e- binding energies of tungsten

Characteristic x-ray photons can have 1 of 15 different energies and no others

Page 19: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Characteristic x-ray emission spectrum

Page 20: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Bremsstrahlung X-ray Spectrum Brems x-rays have a range of energies and

form a continuous emission spectrum

Page 21: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Factors Affecting the x-ray emission spectrum Tube current, Tube voltage, Added

filtration, Target material, Voltage waveform

The general shape of an emission spectrum is always the same, but the position along the energy axis can change

Page 22: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Quality The farther to the right the higher the

effective energy or quality

Page 23: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Quantity The more values in the curve, the higher

the x-ray intensity or quantity

Page 24: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

mAs A change in mA or s or both results in the

amplitude change of the x-ray emission spectrum at all energies

The shape of the curve will remain the same

Page 25: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

mA increase from 200 to 400

Page 26: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

kVp A change in voltage peak affects both the

amplitude and the position of the x-ray emission spectrum

Page 27: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Filtration Adding filtration is called hardening the x-

ray beam because of the increase in average energy

Characteristic spectrum is not affected & the maximum energy of x-ray emission is not affected

Page 28: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Filtration Adding filtration to the useful beam

reduces the x-ray beam intensity while increasing the average energy

Added filtration is an increase in the average energy of the x-ray beam (higher quality) with a reduction in x-ray quantity Lowering the amplitude and shifting to the

right

Page 29: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

What kVp does this graph indicate?

Page 30: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Target Material The atomic number of the target affects

both the quantity and quality of x-rays

Increasing the target atomic number increases the efficiency of x-ray production and the energy of characteristic and bremsstrahlung x-rays

Page 31: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Target material

Page 32: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Voltage Waveform 5 voltage waveforms: half-wave

rectification, full-wave rectification, 3-phase/6-pulse, 3-phase/12-pulse, and high-frequency.

Maintaining high voltage potential

Page 33: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Voltage generators

Page 34: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

X-ray Quantity or Intensity What units of measurement is used for

radiation exposure or exposure in air?

Milliampere-seconds (mAs) – x-ray quantity is proportional to mAs

Kilovolt Peak (kVp) – If kVp were doubled the x-ray intensity would increase by a factor of four or kVp2

Page 35: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

X-ray Quantity or Intensity Distance – x-ray intensity varies inversely

with the square of the distance from the x-ray target

When SID is increased, mAs must be increased by SID2 to maintain constant OD

Page 36: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Filtration 1 to 3 mm of aluminum (Al) added to the

primary beam to reduce the number of low-energy x-rays that reach the patient, reducing patient dose

Filtration reduces the quantity of x-rays in the low-energy range

Page 37: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Reducing low-energy photons

Page 38: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

X-ray Quality or Penetrability As the energy of an x-ray beam is

increased, the penetrability is also increased

High-energy photons are able to penetrate tissue farther than low-energy photons

High-quality = high-penetrability Low-quality = low-penetrability

Page 39: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

HVL = Half-Value Layer What is the HVL

HVL is affected by the kVp and added filtration in the useful beam

Photon quality is also influenced by kVp & filtration

HVL is affected by kVp

Page 40: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

HVL In radiography, the quality of the x-rays is

measured by the HVL

The HVL is a characteristic of the useful x-ray beam

A diagnostic x-ray beam usually has an HVL of 3 to 5 mm Al

Page 41: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

HVL 3 to 5 mm Al = to 3 to 6 cm of soft tissue

HVL is determined experimentally and a design specification of the equipment

Page 42: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

X-ray Quality Kilovolt Peak (kVp) = increasing the kVp

increased photon quality and the HVL

Page 43: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Types of Filtration Diagnostic x-ray beams have two filtration

components – inherent filtration and added filtration

Inherent filtration – The glass enclosure of the tube (the window) – approximately 0.5 mm Al equivalent

Page 44: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Added Filtration 1 or 2 mm sheet of aluminum between the

tube housing and the collimator

The collimator contributes an additional 1mm Al equivalent added filtration

Page 45: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum
Page 46: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Compensating filter A filter usually made of Al, but plastic can

be used to maintain OD when patient anatomy varies greatly in thickness

Are useful in maintaining image quality. They are not radiation protection devices

Page 47: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Wedge filter

Page 48: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum

Compensating Filter What is an aspect of the tube design that

works as a compensating filter?

What causes this?

Page 49: X-Ray Production & Emission. Objectives:  Review x-ray production requirements  X-ray tube interactions  X-ray emission spectrum