xslms series light measurement systemsfiles.ctctcdn.com/1bc17c78001/0ccd72fe-c332-4fcb-a... ·...

6
High-speed analysis of LED and SSL lamps and assemblies XSLMS SERIES LIGHT MEASUREMENT SYSTEMS FEATURES: Wide Dynamic Range Allows a Single Sphere to Measure a Wide Range of Light Levels NIST Traceable Standards for In-house Recalibration Spectral Results in Milliseconds Comprehensive Light Measurement Software • Total Spectral Flux • Luminous Flux • Radiant Flux • Chromaticity • CCT • CRI • Peak Wavelength • Dominant Wavelength • I,V and Luminous Efficiency Spectraflect® Interior Sphere Coating IESNA Recommended Geometries Labsphere, Inc. • 231 Shaker St. • P.O. Box 70 • North Sutton, NH 03260 • USA • Tel: +1 (603) 927-4266 • www.labsphere.com 201008-001 Revision 00 ACCURATE The XSLMS Series Systems feature the widest dynamic range available and are easily customized with a choice of four wavelength ranges and a variety of interchangeable light measurement accessories. NIST traceable total spectral flux standards enable simple in-house system recalibration and verification of the system. The systems can be expanded to include an auxiliary lamp for the correction of absorption errors to ensure accurate test results. Sphere diameters from 10 inches to 3 meters are coated with Spectraflect® which combines a 98% reflective surface with nearly perfect Lambertian reflectance. Both durable and highly stable over time, this coating provides consistent integration of light over the lifetime of your sphere. The light measurement sphere design offers easy interior access to the device under test (DUT). The base port admits additional cables for assembled fixtures while the external port provides for the unobstructed positioning of board-mounted or heat-sinked devices for measurements. Quick connect internal and external electrical contacts allow for effortless substitution between DUTs. FAST The XSLMS Series Systems combine intuitive software with high-speed hardware to simplify complex measurements. The coupling of application specific software with our CDS-9800 Series CCD-based spectrometers, integrating sphere, and cosine receiver makes it possible to deliver total spectral flux, luminous flux, radiometric flux, and colorimetric results in milliseconds. EASY-TO-USE The Windows®-based XSLMS-SPEC software guides users through testing, providing straight-forward testing procedures and repeatable results to meet the needs of both new users and experienced researchers. SPECTRAFLECT REFLECTANCE DATA

Upload: others

Post on 18-Jul-2020

27 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: XSLMS SERIES LIGHT MEASUREMENT SYSTEMSfiles.ctctcdn.com/1bc17c78001/0ccd72fe-c332-4fcb-a... · Illuminant A 1.3 mlm 2.0 mlm 7.0 mlm 20 mlm 27 mlm Red LED Range: 0.20 mlm 0.33 mlm

High-speed analysis of LED and SSL lamps and assemblies

XSLMS SERIES LIGHT MEASUREMENT SYSTEMS

FEATURES: Wide Dynamic Range Allows a Single Sphere to Measurea Wide Range of Light Levels

NIST Traceable Standards for In-house Recalibration

Spectral Results in Milliseconds

Comprehensive Light Measurement Software• Total Spectral Flux• Luminous Flux• Radiant Flux• Chromaticity• CCT• CRI• Peak Wavelength• Dominant Wavelength • I,V and Luminous Efficiency

Spectraflect® Interior Sphere Coating

IESNA Recommended Geometries

Labsphere, Inc. • 231 Shaker St. • P.O. Box 70 • North Sutton, NH 03260 • USA • Tel: +1 (603) 927-4266 • www.labsphere.com 201008-001 Revision 00

ACCURATEThe XSLMS Series Systems feature the widest dynamic range available and are easily customized with a choice of four wavelength ranges and a variety of interchangeable light measurement accessories. NIST traceable total spectral flux standards enable simple in-house system recalibration and verification of the system. The systems can be expanded to include an auxiliary lamp for the correction of absorption errors to ensure accurate test results.

Sphere diameters from 10 inches to 3 meters are coated with Spectraflect® which combines a 98% reflective surface with nearly perfect Lambertian reflectance. Both durable and highly stable over time, this coating provides consistent integration of light over the lifetime of your sphere.

The light measurement sphere design offers easy interior access to the device under test (DUT). The base port admits additional cables for assembled fixtures while the external port provides for the unobstructed positioning of board-mounted or heat-sinked devices for measurements. Quick connect internal and external electrical contacts allow for effortless substitution between DUTs.

FASTThe XSLMS Series Systems combine intuitive software with high-speed hardware to simplify complex measurements. The coupling of application specific software with our CDS-9800 Series CCD-based spectrometers, integrating sphere, and cosine receiver makes it possible to deliver total spectral flux, luminous flux, radiometric flux, and colorimetric results in milliseconds.

EASY-TO-USEThe Windows®-based XSLMS-SPEC software guides users through testing, providing straight-forward testing procedures and repeatable results to meet the needs of both new users and experienced researchers.

SPECTRAFLECT REFLECTANCE DATA

Page 2: XSLMS SERIES LIGHT MEASUREMENT SYSTEMSfiles.ctctcdn.com/1bc17c78001/0ccd72fe-c332-4fcb-a... · Illuminant A 1.3 mlm 2.0 mlm 7.0 mlm 20 mlm 27 mlm Red LED Range: 0.20 mlm 0.33 mlm

XSLMS SERIES FEATURESFOUR HIGH SPEED SPECTROMETER OPTIONS OFFER WIDE DYNAMIC RANGE AND FLEXIBILITY TO MEET YOUR EXACT MEASUREMENT NEEDS.

The XSLMS Series Systems are available in a choice of four spectrometer wavelength ranges to give you flexibility in choosing a system that’s just right for your application.

• Wide dynamic range covers a variety of total flux measurement needs

• Suitable for UV and quantum efficiency evaluation due to 1/5 lower stray light effect compared to conventional spectrometers

• Exposure time range 5 msec to 20 sec (option: 5 msec to 65 sec) covers

various measurement needs, including extremely low light, high speed,

and production line

• High repeatability achieved with Quartz scrambled fiber bundles

• Lightweight and compact design

Wide dynamic rangeWide measurement range was achieved by expanding exposure time and combining ND filters.

Comparison of measurement range

1000000

100000

10000

1000

100

10

10 100 1000 10000 100000

Mea

sure

men

t ran

ge

exposure time (msec)

11

CDS-9800

Typical(standard type)

Typical(high-sensitivity type)

Comparison by stray light (Specification cutting filter: OG550)

0.0020

0.0015

0.0010

0.0005

0.0000300 400 500 600 700

Dar

k co

rrect

ion

valu

e

Wavelength(nm)

Low stray light functionWe can achieved about 1/5 decrease of the stray light effect compared with our conventional model by the lowering stray light function.

Sample:KT-krypton lamp(produced by Panasonic Co.,) LDS110V60W・K・T/E11

Typical(high-sensitivity type)

CDS-9800

Labsphere, Inc. • 231 Shaker St. • P.O. Box 70 • North Sutton, NH 03260 • USA • Tel: +1 (603) 927-4266 • www.labsphere.com 201008-001 Revision 00

Page 3: XSLMS SERIES LIGHT MEASUREMENT SYSTEMSfiles.ctctcdn.com/1bc17c78001/0ccd72fe-c332-4fcb-a... · Illuminant A 1.3 mlm 2.0 mlm 7.0 mlm 20 mlm 27 mlm Red LED Range: 0.20 mlm 0.33 mlm

Labsphere, Inc. • 231 Shaker St. • P.O. Box 70 • North Sutton, NH 03260 • USA • Tel: +1 (603) 927-4266 • www.labsphere.com 201008-001 Revision 00

Ordering Information and Specifications - 360 nm to 830 nm XSLMS-9800 Light Measurement System XSLMS-1098-VIS XSLMS-2098-VIS XSLMS-4098-VIS XSLMS-6598-VIS XSLMS-7698-VIS System Includes Light Measurement Sphere LMS-10 LMS-20 LMS-40 LMS-65 LMS-76 Spectrometer CDS-9800-VIS CDS-9800-VIS CDS-9800-VIS CDS-9800-VIS CDS-9800-VIS Power Supply, Lambda TDK GEN-150-10 GEN-150-10 GEN-150-10 GEN-150-10 GEN-150-10Light Measurement Software XSLMS-SPEC XSLMS-SPEC XSLMS-SPEC XSLMS-SPEC XSLMS-SPEC

Required, Sold SeparatelyLamp Standard TOCS-10 TOCS-20 TOCS-40 TOCS-65 TOCS-76

System Properties and Performance

Sphere Diameter: 10 in (25 cm) 20 in (50 cm) 40 in (1.02 m) 65 in (1.65 m) 76 in (1.93 m) Sphere Coating Reflectance: 98% 98% 98% 98% 98% Radiometric Range: 100 W (max) 400 W (max) 1,500 W (max) 4,000 W (max) 5,000 W (max) Minimum Lumen Level: Illuminant A 1.3 mlm 2.0 mlm 7.0 mlm 20 mlm 27 mlm Red LED Range: 0.20 mlm 0.33 mlm 2.1 mlm 4.0 mlm 5.7 mlm Green LED Range: 0.40 mlm 0.50 mlm 2.0 mlm 5.0 mlm 6.7 mlm Blue LED Range: 0.13 mlm 0.17 mlm 0.83 mlm 1.7 mlm 2.3 mlm Max Recommended DUT Dimension: 1 x 1 in (3 x 3 cm) 2 x 2 in (5 x 5 cm) 4 x 4 in (10 x 10 cm) 7 x 7 in (18 x 18 cm) 8 x 8 in (21 x 21 cm) Sphere Dimension: (W x D x H) 16.7 x 11.9 x 15.5 in 28.5 x 23.7 x 29.4 in 48.9 x 42.3 x 68.5 in 112.6 x 78.4 x 84.1 in 116.9 x 88.2 x 91.4 in (42.4 x 30.2 x 39.4 cm) (72.4 x 60.2 x 74.7 cm) (1.24 x 1.07 x 1.74 cm) (2.86 x 1.99 x 2.13 m) (2.97 x 2.24 x 2.32 m)2p External Port Dimension: 2 in (5.1 cm) 6 in (15.2 cm) 8 in (20.3 cm) 12 in (30.5 cm) 18 in (45.7 cm)

Spectrometer CDS 9800-VIS CDS 9800-VIS CDS 9800-VIS CDS 9800-VIS CDS 9800-VIS Detector: (*1) TE Cooled 1024 x 122 CCD TE Cooled 1024 x 122 CCD TE Cooled 1024 x 122 CCD TE Cooled 1024 x 122 CCD TE Cooled 1024 x 122 CCDSpectral Range: 360 - 830 nm 360 - 830 nm 360 - 830 nm 360 - 830 nm 360 - 830 nmGrating: f= 85.8 mm 448grooves/mm f= 85.8 mm 448grooves/mm f= 85.8 mm 448grooves/mm f= 85.8 mm 448grooves/mm f= 85.8 mm 448grooves/mmGate-Time: 5 msec - 20 msec 5 msec - 20 msec 5 msec - 20 msec 5 msec - 20 msec 5 msec - 20 msec (option: 5 msec to 65 sec) (option: 5 msec to 65 sec) (option: 5 msec to 65 sec) (option: 5 msec to 65 sec) (option: 5 msec to 65 sec)ND Filter: OD1-2 OD1-2 OD1-2 OD1-2 OD1-2Data Point Interval: 0.5 nm (1024ch) 0.5 nm (1024ch) 0.5 nm(1024ch)) 0.5 nm (1024ch) 0.5 nm (1024ch)Resolution: (nm/ch) 1.0 nm (512ch) 1.0 nm (512ch) 1.0 nm (512ch) 1.0 nm (512ch) 1.0 nm (512ch)Wavelength Accuracy: ± 0.3 nm ± 0.3 nm ± 0.3 nm ± 0.3 nm ± 0.3 nmStray Light: (B-band for Illuminant A) 2.5 - 10E-4 2.5 - 10E-4 2.5 - 10E-4 2.5 - 10E-4 2.5 - 10E-4Dynamic Range: 1000000:1 1000000:1 1000000:1 1000000:1 1000000:1LInearity: ± 0.8% ± 0.8% ± 0.8% ± 0.8% ± 0.8%A/D Converter: 16 bit 16 bit 16 bit 16 bit 16 bit A/D Rate: 1 MHz 1 MHz 1 MHz 1 MHz 1 MHzPC Interface: USB 2.0 USB 2.0 USB 2.0 USB 2.0 USB 2.0 Weight: 6 kg (18 lbs) 6 kg (18 lbs) 6 kg (18 lbs) 6 kg (18 lbs) 6 kg (18 lbs)Dimensions: (W x H x D) 105 x 230 x 282 mm 105 x 230 x 282 mm 105 x 230 x 282 mm 105 x 230 x 282 mm 105 x 230 x 282 mm

Lamp Standard SCL-600 SCL-600 SCL-1400 SCL-1400 SCL-1400 Approximate Luminous Flux: (lumens) 600 lm 600 lm 1400 lm 1400 lm 1400 lm Lamp Current: (amps) 2.600 2.600 2.680 2.680 2.680Rated Voltage: (volts) 12.8 12.8 28 28 28Rated Life Base: (hrs) 600 600 2000 2000 2000

Page 4: XSLMS SERIES LIGHT MEASUREMENT SYSTEMSfiles.ctctcdn.com/1bc17c78001/0ccd72fe-c332-4fcb-a... · Illuminant A 1.3 mlm 2.0 mlm 7.0 mlm 20 mlm 27 mlm Red LED Range: 0.20 mlm 0.33 mlm

Labsphere, Inc. • 231 Shaker St. • P.O. Box 70 • North Sutton, NH 03260 • USA • Tel: +1 (603) 927-4266 • www.labsphere.com 201008-001 Revision 00

Ordering Information and Specifications - 360 nm to 1100 nm XSLMS-9800 Light Measurement System XSLMS-1098-VIS-IR XSLMS-2098-VIS-IR XSLMS-4098-VIS-IR XSLMS-6598-VIS-IR XSLMS-7698-VIS-IR System Includes Light Measurement Sphere LMS-10 LMS-20 LMS-40 LMS-65 LMS-76 Spectrometer CDS-9800-VIS-IR CDS-9800-VIS-IR CDS-9800-VIS-IR CDS-9800-VIS-IR CDS-9800-VIS-IR Power Supply, Lambda TDK GEN-150-10 GEN-150-10 GEN-150-10 GEN-150-10 GEN-150-10Light Measurement Software XSLMS-SPEC XSLMS-SPEC XSLMS-SPEC XSLMS-SPEC XSLMS-SPEC

Required, Sold SeparatelyLamp Standard TOCS-10 TOCS-20 TOCS-40 TOCS-65 TOCS-76

System Properties and Performance

Sphere Diameter: 10 in (25 cm) 20 in (50 cm) 40 in (1.02 m) 65 in (1.65 m) 76 in (1.93 m) Sphere Coating Reflectance: 98% 98% 98% 98% 98% Radiometric Range: 100 W (max) 400 W (max) 1,500 W (max) 4,000 W (max) 5,000 W (max) Minimum Lumen Level: Illuminant A 1.3 mlm 2.0 mlm 7.0 mlm 20 mlm 27 mlm Red LED Range: 0.20 mlm 0.33 mlm 2.1 mlm 4.0 mlm 5.7 mlm Green LED Range: 0.40 mlm 0.50 mlm 2.0 mlm 5.0 mlm 6.7 mlm Blue LED Range: 0.13 mlm 0.17 mlm 0.83 mlm 1.7 mlm 2.3 mlm Max Recommended DUT Dimension: 1 x 1 in (3 x 3 cm) 2 x 2 in (5 x 5 cm) 4 x 4 in (10 x 10 cm) 7 x 7 in (18 x 18 cm) 8 x 8 in (21 x 21 cm) Sphere Dimension: (W x D x H) 16.7 x 11.9 x 15.5 in 28.5 x 23.7 x 29.4 in 48.9 x 42.3 x 68.5 in 112.6 x 78.4 x 84.1 in 116.9 x 88.2 x 91.4 in (42.4 x 30.2 x 39.4 cm) (72.4 x 60.2 x 74.7 cm) (1.24 x 1.07 x 1.74 cm) (2.86 x 1.99 x 2.13 m) (2.97 x 2.24 x 2.32 m)2p External Port Dimension: 2 in (5.1 cm) 6 in (15.2 cm) 8 in (20.3 cm) 12 in (30.5 cm) 18 in (45.7 cm)

Spectrometer CDS 9800-VIS-IR CDS 9800-VIS-IR CDS 9800-VIS-IR CDS 9800-VIS-IR CDS 9800-VIS-IR Detector: (*1) TE Cooled 1024 x 122 CCD TE Cooled 1024 x 122 CCD TE Cooled 1024 x 122 CCD TE Cooled 1024 x 122 CCD TE Cooled 1024 x 122 CCDSpectral Range: 360 - 1100 nm 360 - 1100 nm 360 - 1100 nm 360 - 1100 nm 360 - 1100 nmGrating: f= 85.8 mm 303grooves/mm f= 85.8 mm 303grooves/mm f= 85.8 mm 303grooves/mm f= 85.8 mm 303grooves/mm f= 85.8 mm 303grooves/mmGate-Time: 5 msec - 20 msec 5 msec - 20 msec 5 msec - 20 msec 5 msec - 20 msec 5 msec - 20 msec (option: 5 msec to 65 sec) (option: 5 msec to 65 sec) (option: 5 msec to 65 sec) (option: 5 msec to 65 sec) (option: 5 msec to 65 sec)ND Filter: OD1-2 OD1-2 OD1-2 OD1-2 OD1-2Data Point Interval: 0.8 nm (1024ch) 0.8 nm (1024ch) 0.8 nm(1024ch)) 0.8 nm (1024ch) 0.8 nm (1024ch)Resolution: (nm/ch) 1.6 nm (512ch) 1.6 nm (512ch) 1.6 nm (512ch) 1.6 nm (512ch) 1.6 nm (512ch)Wavelength Accuracy: ± 0.5 nm ± 0.5 nm ± 0.5 nm ± 0.5 nm ± 0.5 nmStray Light: (B-band for Illuminant A) 2.5 - 10E-4 2.5 - 10E-4 2.5 - 10E-4 2.5 - 10E-4 2.5 - 10E-4Dynamic Range: 1000000:1 1000000:1 1000000:1 1000000:1 1000000:1LInearity: ± 0.8% ± 0.8% ± 0.8% ± 0.8% ± 0.8%A/D Converter: 16 bit 16 bit 16 bit 16 bit 16 bit A/D Rate: 1 MHz 1 MHz 1 MHz 1 MHz 1 MHzPC Interface: USB 2.0 USB 2.0 USB 2.0 USB 2.0 USB 2.0 Weight: 6 kg (18 lbs) 6 kg (18 lbs) 6 kg (18 lbs) 6 kg (18 lbs) 6 kg (18 lbs)Dimensions: (W x H x D) 105 x 230 x 282 mm 105 x 230 x 282 mm 105 x 230 x 282 mm 105 x 230 x 282 mm 105 x 230 x 282 mm

Lamp Standard SCL-600 SCL-600 SCL-1400 SCL-1400 SCL-1400 Approximate Luminous Flux: (lumens) 600 lm 600 lm 1400 lm 1400 lm 1400 lm Lamp Current: (amps) 2.600 2.600 2.680 2.680 2.680Rated Voltage: (volts) 12.8 12.8 28 28 28Rated Life Base: (hrs) 600 600 2000 2000 2000

Page 5: XSLMS SERIES LIGHT MEASUREMENT SYSTEMSfiles.ctctcdn.com/1bc17c78001/0ccd72fe-c332-4fcb-a... · Illuminant A 1.3 mlm 2.0 mlm 7.0 mlm 20 mlm 27 mlm Red LED Range: 0.20 mlm 0.33 mlm

Labsphere, Inc. • 231 Shaker St. • P.O. Box 70 • North Sutton, NH 03260 • USA • Tel: +1 (603) 927-4266 • www.labsphere.com 201008-001 Revision 00

Ordering Information and Specifications - 300 nm to 950 nm XSLMS-9800 Light Measurement System XSLMS-1098-VISB XSLMS-2098-VISB XSLMS-4098-VISB XSLMS-6598-VISB XSLMS-7698-VISB System Includes Light Measurement Sphere LMS-10 LMS-20 LMS-40 LMS-65 LMS-76 Spectrometer CDS-9800-VISB CDS-9800-VISB CDS-9800-VISB CDS-9800-VISB CDS-9800-VISB Power Supply, Lambda TDK GEN-150-10 GEN-150-10 GEN-150-10 GEN-150-10 GEN-150-10Light Measurement Software XSLMS-SPEC XSLMS-SPEC XSLMS-SPEC XSLMS-SPEC XSLMS-SPEC

Required, Sold SeparatelyLamp Standard TOCS-10 TOCS-20 TOCS-40 TOCS-65 TOCS-76

System Properties and Performance

Sphere Diameter: 10 in (25 cm) 20 in (50 cm) 40 in (1.02 m) 65 in (1.65 m) 76 in (1.93 m) Sphere Coating Reflectance: 98% 98% 98% 98% 98% Radiometric Range: 100 W (max) 400 W (max) 1,500 W (max) 4,000 W (max) 5,000 W (max) Minimum Lumen Level: Illuminant A 1.3 mlm 2.0 mlm 7.0 mlm 20 mlm 27 mlm Red LED Range: 0.20 mlm 0.33 mlm 2.1 mlm 4.0 mlm 5.7 mlm Green LED Range: 0.40 mlm 0.50 mlm 2.0 mlm 5.0 mlm 6.7 mlm Blue LED Range: 0.13 mlm 0.17 mlm 0.83 mlm 1.7 mlm 2.3 mlm Max Recommended DUT Dimension: 1 x 1 in (3 x 3 cm) 2 x 2 in (5 x 5 cm) 4 x 4 in (10 x 10 cm) 7 x 7 in (18 x 18 cm) 8 x 8 in (21 x 21 cm) Sphere Dimension: (W x D x H) 16.7 x 11.9 x 15.5 in 28.5 x 23.7 x 29.4 in 48.9 x 42.3 x 68.5 in 112.6 x 78.4 x 84.1 in 116.9 x 88.2 x 91.4 in (42.4 x 30.2 x 39.4 cm) (72.4 x 60.2 x 74.7 cm) (1.24 x 1.07 x 1.74 cm) (2.86 x 1.99 x 2.13 m) (2.97 x 2.24 x 2.32 m)2p External Port Dimension: 2 in (5.1 cm) 6 in (15.2 cm) 8 in (20.3 cm) 12 in (30.5 cm) 18 in (45.7 cm)

Spectrometer CDS 9800-VISB CDS 9800-VISB CDS 9800-VISB CDS 9800-VISB CDS 9800-VISB Detector: (*1) TE Cooled 1024 x 122 CCD TE Cooled 1024 x 122 CCD TE Cooled 1024 x 122 CCD TE Cooled 1024 x 122 CCD TE Cooled 1024 x 122 CCDSpectral Range: 300 - 950 nm 300 - 950 nm 300 - 950 nm 300 - 950 nm 300 - 950 nmGrating: f= 85.8 mm 340grooves/mm f= 85.8 mm 340grooves/mm f= 85.8 mm 340grooves/mm f= 85.8 mm 340grooves/mm f= 85.8 mm 340grooves/mmGate-Time: 5 msec - 20 msec 5 msec - 20 msec 5 msec - 20 msec 5 msec - 20 msec 5 msec - 20 msec (option: 5 msec to 65 sec) (option: 5 msec to 65 sec) (option: 5 msec to 65 sec) (option: 5 msec to 65 sec) (option: 5 msec to 65 sec)ND Filter: OD1-2 OD1-2 OD1-2 OD1-2 OD1-2Data Point Interval: 0.7 nm (1024ch) 0.7 nm (1024ch) 0.7 nm(1024ch)) 0.7 nm (1024ch) 0.7 nm (1024ch)Resolution: (nm/ch) 1.4 nm (512ch) 1.4 nm (512ch) 1.4 nm (512ch) 1.4 nm (512ch) 1.4 nm (512ch)Wavelength Accuracy: ± 0.3 nm ± 0.3 nm ± 0.3 nm ± 0.3 nm ± 0.3 nmStray Light: (B-band for Illuminant A) 2.5 - 10E-4 2.5 - 10E-4 2.5 - 10E-4 2.5 - 10E-4 2.5 - 10E-4Dynamic Range: 1000000:1 1000000:1 1000000:1 1000000:1 1000000:1LInearity: ± 0.8% ± 0.8% ± 0.8% ± 0.8% ± 0.8%A/D Converter: 16 bit 16 bit 16 bit 16 bit 16 bit A/D Rate: 1 MHz 1 MHz 1 MHz 1 MHz 1 MHzPC Interface: USB 2.0 USB 2.0 USB 2.0 USB 2.0 USB 2.0 Weight: 6 kg (18 lbs) 6 kg (18 lbs) 6 kg (18 lbs) 6 kg (18 lbs) 6 kg (18 lbs)Dimensions: (W x H x D) 105 x 230 x 282 mm 105 x 230 x 282 mm 105 x 230 x 282 mm 105 x 230 x 282 mm 105 x 230 x 282 mm

Lamp Standard SCL-600 SCL-600 SCL-1400 SCL-1400 SCL-1400 Approximate Luminous Flux: (lumens) 600 lm 600 lm 1400 lm 1400 lm 1400 lm Lamp Current: (amps) 2.600 2.600 2.680 2.680 2.680Rated Voltage: (volts) 12.8 12.8 28 28 28Rated Life Base: (hrs) 600 600 2000 2000 2000

Page 6: XSLMS SERIES LIGHT MEASUREMENT SYSTEMSfiles.ctctcdn.com/1bc17c78001/0ccd72fe-c332-4fcb-a... · Illuminant A 1.3 mlm 2.0 mlm 7.0 mlm 20 mlm 27 mlm Red LED Range: 0.20 mlm 0.33 mlm

Labsphere, Inc. • 231 Shaker St. • P.O. Box 70 • North Sutton, NH 03260 • USA • Tel: +1 (603) 927-4266 • www.labsphere.com 201008-001 Revision 00

Ordering Information and Specifications - 240 to 800 nm XSLMS-9800 Light Measurement System XSLMS-1098-UV-VIS XSLMS-2098-UV-VIS XSLMS-4098-UV-VIS XSLMS-6598-UV-VIS XSLMS-7698-UV-VIS System Includes Light Measurement Sphere LMS-10 LMS-20 LMS-40 LMS-65 LMS-76 Spectrometer CDS-9800-UV-VIS CDS-9800-UV-VIS CDS-9800-UV-VIS CDS-9800-UV-VIS CDS-9800-UV-VIS Power Supply, Lambda TDK GEN-150-10 GEN-150-10 GEN-150-10 GEN-150-10 GEN-150-10Light Measurement Software XSLMS-SPEC XSLMS-SPEC XSLMS-SPEC XSLMS-SPEC XSLMS-SPEC

Required, Sold SeparatelyLamp Standard TOCS-10 TOCS-20 TOCS-40 TOCS-65 TOCS-76

System Properties and Performance

Sphere Diameter: 10 in (25 cm) 20 in (50 cm) 40 in (1.02 m) 65 in (1.65 m) 76 in (1.93 m) Sphere Coating Reflectance: 98% 98% 98% 98% 98% Radiometric Range: 100 W (max) 400 W (max) 1,500 W (max) 4,000 W (max) 5,000 W (max) Minimum Lumen Level: Illuminant A 1.3 mlm 2.0 mlm 7.0 mlm 20 mlm 27 mlm Red LED Range: 0.20 mlm 0.33 mlm 2.1 mlm 4.0 mlm 5.7 mlm Green LED Range: 0.40 mlm 0.50 mlm 2.0 mlm 5.0 mlm 6.7 mlm Blue LED Range: 0.13 mlm 0.17 mlm 0.83 mlm 1.7 mlm 2.3 mlm Max Recommended DUT Dimension: 1 x 1 in (3 x 3 cm) 2 x 2 in (5 x 5 cm) 4 x 4 in (10 x 10 cm) 7 x 7 in (18 x 18 cm) 8 x 8 in (21 x 21 cm) Sphere Dimension: (W x D x H) 16.7 x 11.9 x 15.5 in 28.5 x 23.7 x 29.4 in 48.9 x 42.3 x 68.5 in 112.6 x 78.4 x 84.1 in 116.9 x 88.2 x 91.4 in (42.4 x 30.2 x 39.4 cm) (72.4 x 60.2 x 74.7 cm) (1.24 x 1.07 x 1.74 cm) (2.86 x 1.99 x 2.13 m) (2.97 x 2.24 x 2.32 m)2p External Port Dimension: 2 in (5.1 cm) 6 in (15.2 cm) 8 in (20.3 cm) 12 in (30.5 cm) 18 in (45.7 cm)

Spectrometer CDS 9800-UV-VIS CDS 9800-UV-VIS CDS 9800-UV-VIS CDS 9800-UV-VIS CDS 9800-UV-VIS Detector: (*1) TE Cooled 1024 x 122 CCD TE Cooled 1024 x 122 CCD TE Cooled 1024 x 122 CCD TE Cooled 1024 x 122 CCD TE Cooled 1024 x 122 CCDSpectral Range: 240 - 800 nm 240 - 800 nm 240 - 800 nm 240 - 800 nm 240 - 800 nmGrating: f= 85.8 mm 405grooves/mm f= 85.8 mm 405grooves/mm f= 85.8 mm 405grooves/mm f= 85.8 mm 405grooves/mm f= 85.8 mm 405grooves/mmGate-Time: 5 msec - 20 msec 5 msec - 20 msec 5 msec - 20 msec 5 msec - 20 msec 5 msec - 20 msec (option: 5 msec to 65 sec) (option: 5 msec to 65 sec) (option: 5 msec to 65 sec) (option: 5 msec to 65 sec) (option: 5 msec to 65 sec)ND Filter: OD1-2 OD1-2 OD1-2 OD1-2 OD1-2Data Point Interval: 0.6 nm (1024ch) 0.6 nm (1024ch) 0.6 nm(1024ch)) 0.6 nm (1024ch) 0.6 nm (1024ch)Resolution: (nm/ch) 1.2 nm (512ch) 1.2 nm (512ch) 1.2 nm (512ch) 1.2 nm (512ch) 1.2 nm (512ch)Wavelength Accuracy: ± 0.3 nm ± 0.3 nm ± 0.3 nm ± 0.3 nm ± 0.3 nmStray Light: (B-band for Illuminant A) 2.5 - 10E-4 2.5 - 10E-4 2.5 - 10E-4 2.5 - 10E-4 2.5 - 10E-4Dynamic Range: 1000000:1 1000000:1 1000000:1 1000000:1 1000000:1LInearity: ± 0.8% ± 0.8% ± 0.8% ± 0.8% ± 0.8%A/D Converter: 16 bit 16 bit 16 bit 16 bit 16 bit A/D Rate: 1 MHz 1 MHz 1 MHz 1 MHz 1 MHzPC Interface: USB 2.0 USB 2.0 USB 2.0 USB 2.0 USB 2.0 Weight: 6 kg (18 lbs) 6 kg (18 lbs) 6 kg (18 lbs) 6 kg (18 lbs) 6 kg (18 lbs)Dimensions: (W x H x D) 105 x 230 x 282 mm 105 x 230 x 282 mm 105 x 230 x 282 mm 105 x 230 x 282 mm 105 x 230 x 282 mm

Lamp Standard SCL-600 SCL-600 SCL-1400 SCL-1400 SCL-1400 Approximate Luminous Flux: (lumens) 600 lm 600 lm 1400 lm 1400 lm 1400 lm Lamp Current: (amps) 2.600 2.600 2.680 2.680 2.680Rated Voltage: (volts) 12.8 12.8 28 28 28Rated Life Base: (hrs) 600 600 2000 2000 2000