Yue YanProf. Bhudatt Paliwal 10/24/2014
Background Motivation Preliminary studies
• Clinical comparison• Monte-Carlo (MC) simulation• Measurement test
Conclusion Questions and answers
2
Flattening filter (FF) has been applied as an essential part of Linac
system over decades.
The development of modern RT makes the FF no longer an
indispensable part of Linac system.
The potential clinical application of the FFF bremsstrahlung photon
beams generated by modern Linacs is becoming much more
apparent[1-12].
However, limitations still exist for the FFF beam.
3
Pros of FFF• Greatly increased dose rate (approx. 2-4 times compared with the
flattened beam)[2-5,9].
• Large decrease of external scatter from the gantry head[6].
• Greatly reduced neutron contamination for high energy photon beams[7].
• Reduced uncertainty in dose calculation due to scatter from the FF[8]. Cons of FFF
• High dose rate of the FFF beam is more likely to lead to unacceptable
dose deviation caused by intrafraction patient motion during treatment
delivery[1].
• Soft spectrum of the FFF beam tends to increase the internal patient
scatter to peripheral organs[9-12].
4
Internal patient scatter is the predominate factor
contributing to peripheral organ dose[12].
External head scatter and dose leakage from the gantry
head also contribute smaller amounts to peripheral
organ dose.
5
The hypothesis: Dose to OARs can be further reduced by attenuating the soft
photons in the FFF beam spectrum.
6
FFF Beam Flattened BeamLow head scatter High head scatter from the flattening filter
Soft spectrum Hard spectrum
High dose rate Low dose rate
Increased internal scatter due to soft photons Increased internal scatter due to dose normalization
Lower tail dose in the beam profile Higher tail dose in the beam profile
SPECTER
In general, lower mean dose to the OARs In general, higher mean dose to the OARs
Adjusting both the beam spectra and the profile to yield a more hardened SPECTER beam could reduce the internal patient scatter and reduce the dose to OARs for the FFF beam.
7
Clinical comparison (cont…):• Treatment plans study summary:
8
Category Beam ModalityDelivery
TechniqueTreatment Schedule OARs
Head and Neck
6 MV and 10 MVFlattened beam &
FFF Beam
Static IMRT&
VMAT
60 Gy/30 frac. (2 pts.)30 Gy/15 frac. (1 pt.)70 Gy/28 frac. (1 pt.)
spinal cord, brainstem,
brain, larynx, pharynx, parotid
Lung45 Gy/30 frac. (1 pt.)66 Gy/33 frac. (2 pts.)60 Gy/30 frac. (1 pt.)
lungs, spinal cord, heart,
larynx, thyroid, esophagus
Prostate 70 Gy/28 frac. (3 pts.)78 Gy/39 frac. (1 pt.)
rectum, bladder, hips
Breast(RTOG 1005)[13]
Static IMRT50 Gy/25 frac.
12 Gy/6 frac. (1 pt.) (sequential boost)
lungs, heart, thyroid, skin
Clinical comparison (cont…):• Normalization: 95% @95% to exclude lower-
mean-dose effect to the target of the FFF beam compared with the
flattened beam.
• TrueBeam (Varian, CA) system was commissioned on Eclipse®
(Varian, Palo Alto, Medical Systems) TPS workstation.
• Anisotropic Analytical Algorithm (AAA) was used to calculate the
dose.
• Treatment parameters (e.g. field size, arc number, beam angle)
were kept to be identical to exclude differences from these factors.
• Clinical objectives were used to simulate clinical application.
9
Clinical comparison (cont…): head and neck (case 7):o Dashed line: FFF beam; solid line: flattened beam
10
VMAT Plans
IMRT Plans
Clinical comparison (cont…):
11
Dose and NTCP Ratios(FFF/flattened) for Head and Neck Cancer CasesVMAT
6 MV 10 MV
Organs' name Mean dose ratio
Max dose ratio
Mean BED ratio NTCP ratio Mean dose
ratioMax dose
ratioMean BED
ratio NTCP ratio
Larynx 0.98±0.02 0.99±0.02 0.97±0.02 0.84±0.20 0.97±0.04 1.00±0.01 0.97±0.05 0.87±0.19R parotid 0.94±0.03 0.96±0.02 0.93±0.03 0.31±0.20 0.91±0.04 0.96±0.05 0.91±0.04 0.24±0.27
R submandibular 0.97±0.02 0.98±0.00 0.96±0.02 0.64±0.03 0.92±0.04 0.96±0.04 0.91±0.04 0.28±0.14L cochlea 0.97±0.04 0.98±0.04 0.97±0.05 0.71±0.48 0.94±0.06 0.95±0.06 0.93±0.07 0.43±0.40
IMRT
6 MV 10 MV
Organs’ name Mean dose ratio
Max dose ratio
Mean BED ratio NTCP ratio Mean dose
ratioMax dose
ratioMean BED
ratio NTCP ratio
Larynx 1.02±0.02 1.02±0.01 1.03±0.02 1.57±0.39 1.02±0.01 1.02±0.01 1.02±0.01 1.42±0.25R parotid 1.00±0.02 1.01±0.01 1.00±0.02 0.95±0.47 0.99±0.04 1.00±0.02 0.99±0.04 0.89±0.72
R submandibular 1.00±0.04 1.00±0.01 1.00±0.05 1.21±0.75 1.04±0.02 1.02±0.02 1.05±0.02 1.74±0.15L cochlea 0.98±0.03 1.00±0.07 0.98±0.04 1.32±1.29 1.00±0.03 1.02±0.08 1.00±0.05 2.18±2.25
Clinical comparison (cont…):
12
MU for Head and Neck Cancer CasesEnergy 6 MV 10 MV
Delivery Method VMAT IMRT VMAT IMRT
Beam Modality Flattened FFF Flattened FFF Flattened FFF Flattened FFF
Total MUs
Case 5 559 640 2117 3072 545 674 2001 3633
Case 7 812 959 1726 2647 780 1011 1655 3128
Case 8 605 757 965 1292 581 801 883 1380
Case 10 390 459 916 1323 373 483 873 1773
Case 24 389 430 803 1254 370 427 768 1758MU Ratio
(FFF/Flattened) 1.17±0.05 1.47±0.09 1.27±0.08 1.92±0.27
Clinical comparison (cont…):• Summary and discussion:
The differences between the FFF beam and the flattened beam are more
obvious in VMAT plans compared with the static IMRT plans.
Maximum dose rates may not be achievable for the FFF beam plans in
general due to MLC speed limitation.
In certain clinical cases with large field sizes (~16 20 ), the FFF
beam tends to provide higher dose to OARs compared with the flattened
beam in the static IMRT plans.
The FFF beam plans uses higher MU to provide the same target
coverage compare with the flatted beam plans.
In general, FFF beam provides lower mean dose to OARs compared with
the flattened beam.
13
Monte-Carlo study:• Detailed geometry information is necessary to accurately simulate
the TrueBeamTM system (Varian, CA).
• Varian provided the IAEA (International Atomic Energy
Agency)phase space data above the jaws inside the gantry head.
14(http://sudentas.com/tag/truebeam) (TrueBeam Monte Carlo Data Package, Varian Medical System)
Monte-Carlo study (cont…):• VirtuaLinac web interface based on Amazon Web Services (AWS) system enables
us to consider the geometry of the jaws inside the gantry head[17,18].
• Based on the new IAEA phase space data, BEAMnrc MC software package was
used to simulate the TrueBeam system.
15
Monte-Carlo Study (cont…):• Excellent agreements were obtained between the MC simulation
and the standard measurement data.
16
10
Monte-Carlo study (cont…):• Dose profile normalization: 6 MV beams
17
Monte-Carlo study (cont…):• Dose normalization: 6 MV and 10 MV beams
18
6 MV
Monte-Carlo study (cont…):• A soft-spectrum-filter (SPECTER) is proposed to attenuate the soft
photons in the FFF spectrum.
• The low density of the SEPCTER in the central axis (filled with air)
enables us to keep the high dose rate of the FFF beam.
19
Monte-Carlo study (cont…):• Different designs of the SPECTER: (gantry head z=58 cm.)
20
21 cm18.3 cm
15.4 cm
12.5 cm
9.6 cm
6.7 cm
21 cm
18.3 cm
15.4 cm
12.5 cm
9.6 cm
6.7 cm
18.3 cm
15.4 cm
12.5 cm
9.6 cm
6.7 cm
18.3 cm
15.4 cm
12.5 cm
9.6 cm
6.7 cm
Monte-Carlo study (cont…): • Steps to calculate the internal scatter[12,19]:
.
• Boundary of the central region was slightly larger than the field
size.
21
Monte-Carlo study (cont…): beam fluence
22
All MC simulations after this slice were based on the lead circularcross section SPECTER.
Monte-Carlo study (cont…):• The SPECTER beam reduced the total dose and the internal scatter compared
with the FFF beam.
23
25 25 , 10 40 40 , 10
Measurement results:• Two SPECTERs were made using lead and tin.
• Cross sections were square.
• The heights were chosen to be 6 for lead and 1.5 for tin.
24
Measurement results (cont…):• Lead SPECTER provides better dose reduction effect compared
with tin, due to higher attenuation of the soft photons of lead.
25
Field sizes from left to right: 3 3, 5 5, 10 10, 15 15,20 20, 25 25 . 10 . Solid line: FFF beam,dashed lines with and without dots: SPECTER beamswith square cross section. Dashed line without dots: leadSPECTER; dashed line with dots: tin SPECTER.
Measurement Results
Field size is chosen to be 25 25 . Solid line: FFF beam,dashed lines with and without dots: SPECTER beamswith square cross section. Dashed line without dots: leadSPECTER; dashed line with dots: tin SPECTER.
Measurement results (cont…):• However, the MC simulation did not agree with the measurement
data in the tail region.
• It can be explained by the over-responding of the Profiler(Sun
Nuclear, Melbourne, FL)[21].
26
Conclusion:• The SPECTER beam has potential to reduce the internal scatter
and improve the dose sparing effect to OARs.
• The low density in the central axis of SPECTER provides high
dose rate compared with the flattened beam.
• Attenuating the soft photons in the FFF beam spectrum may not
lead to prolonged treatment time compared with the FFF beam due
to the speed limitation of the MLCs.
• Geometry and material are important for the design of the
SPECTER.
27
1. Ong CL, Dahele M, Slotman BJ and Verbakel FFAR. (2013) Dosimetric Impact of the Interplay Effect During Stereotactic Lung Radiation Therapy Delivery Using Flattening Filter-Free Beams and Volumetric Modulated Arc Therapy. Int J Radiat Oncol Biol Phys 86(4): 743-748.
2. Huang YF, Siochi RA, Bayouth JE. (2012) Dosimetric properties of a beam quality-matched 6 MV unflattened photon beam. J Appl Clin Med Phys 13(4):71-81.
3. Vassiliev ON, Titt U, Kry SF, Pönisch F, Gillin MT and Mohan R. (2006) Monte Carlo study of photon fields from a flattening filter-free clinical accelerator. Med Phys 33(4):820-827.
4. Reggiori G, Mancosu P, Castiglioni S, Alongi F, Pellegrini C, Lobefalo F, Catalano M, Fogliata A, Arcangeli S, Navarria P, Cozzi L and Scorsetti M. (2012) Can volumetric modulated arc therapy with flattening filter free beams play a role in stereotactic body radiotherapy for liver lesions? A volume-based analysis. Med Phys 39(2):1112-1118.
5. Titt U, Vassiliev ON, Pönisch F, Dong L, Liu H and Mohan R. (2006) A flattening filter free photon treatment concept evaluation with Monte Carlo. Med Phys 33(6):1595-1602.
6. Cashmore J, Ramtohul M and Ford D. (2011) Lowering whole-body radiation doses in pediatric intensity-modulated radiotherapy through the use of unflattened photon beams. Int J Radiat Oncol BiolPhys 80(4):1220-1227.
7. Kry SF, Titt U, Pönisch F, Vassiliev ON, Salehpour M, Gillin M, Mohan R. (2007) Reduced neutron production through use of a flattening-filter-free accelerator. Int J Radiat Oncol Biol Phys 68(4):1260-4.
8. Georg D, Knöös T and McClean B. (2011) Current status and future perspective of flattening filter free photon beams. Med Phys 38(3):1280-1293.
9. Kry SF, Vassiliev ON and Mohan R. (2010) Out-of-field photon dose following removal of the flattening filter from a medical accelerator. Phys Med Biol 55 2155-2166.
10. Kragl G, Baier F, Lutz S, Albrich D, Dalaryd M, Kroupa B, Wiezorek T, Knöös T, Georg D. (2011) Flattening filter free beams in SBRT and IMRT: Dosimetric assessment of peripheral doses. Z Med Phys21 91-101.
11. Wang YN, Khan MK, Ting JY, Easterling SB. (2012) Surface dose investigation of the flattening filter-free photon beams. Int J Radiation Oncol Biol Phys 83(2):e281-e285.
28
12. Chofor N, Harder D, Willborn KC and Poppe B. (2012) Internal scatter, the unavoidable major component of the peripheral dose in photon-beam radiotherapy. Phys Med Biol 57 1733-1743.
13. Vicini FA, Freedman GM, White JR, Arthur DW, Haylard M, Taghian A, Hayman J, Rosenstein B, Bentzen SM, Li XA, Bleicher RJ and Winter K. A Phase III trial of accelerated whole breast irradiation with hypofractionation plus concurrent boost versus standard whole breast irradiation plus sequential boost for early-stage breast cancer RTOG 1005 (unpublished).
14. Dhabaan A, Elder E, Schreibmann E, Crocker I, Curran WJ, Oyesiku NM, Shu H-K, Fox T. (2010) Dosimetricperformance of the new high-definition multileaf collimator for intracranial stereotactic radiosurgery. J Appl ClinMed Phy 11(3):197-211.
15. Zhang GG, Ku L, Dilling TJ, Stevens CW, Zhang RR, Li WQ and Feygelman V. (2011) Volumetric modulated arc planning for lung stereotactic body radiotherapy using conventional and unflattened photon beams: a dosimetric comparison with 3D technique. Radiat Oncol 6:152.
16. Gay HA, Niemierko A. (2007) A free program for calculating EUD-based NTCP and TCP in external beam radiotherapy. Phys Medica 23:115-125.
17. J. F. Fowler. (1989) The linear-quadratic formula and progress in fractionated radiotherapy The British journal of radiology 62:679-694.
18. Constantin M, Perl J, LoSasso T, Salop A, Whittum D, Narula A, Svatos M and Keall PJ. (2011) Modeling the TrueBeam linac using a CAD to Geant4 geometry implementation: Dose and IAEA-compliant phase space calculations. Med Phys 38(7):4018-4024.
19. Constantin M, Constantin D E, Keall P J, Narula A, Svatos M, and Perl J. (2010) Linking computer-aided design (CAD) to Geant4-based Monte Carlo simulations for precise implementation of complex treatment head geometries, Phys Med Biol 55 N211-20.
20. Chofor N, Harder D, Willborn K, Rühmann, A and Poppe B. (2011) Low-energy photons in high-energy photon fields – Monte Carlo generated spectra and a new descriptive parameter. Z Med Phys 21 183-197.
21. Sun Nuclear Cooperation, Diodes versus Ion Chambers. 22. Agostinelli S et al. (2003) Geant4—a simulation toolkit. Nucl Instrum Methods Phys Res A 506 250-