1 components of optical instruments. 2 spectroscopic methods are based on either: 1. absorption 2....

86
1 Components of Optical Instruments

Upload: brent-murphy

Post on 24-Dec-2015

228 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

1

Components of Optical Instruments

Page 2: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

2

Spectroscopic methods are based on either:

1. Absorption

2. Emission

3. Scattering

Spectroscopic instruments dependent on any of the above mechanisms encompass common components which can be listed as:

Page 3: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

3

1. A stable source of radiation2. A wavelength selector to choose a single

wavelength necessary for a certain absorption, emission or scattering process.

3. A radiation detector (transducer) that can measure absorbed, emitted or scattered radiation.

4. A signal processor that can change the electrical signal (current, voltage, or resistance) to a suitable form like absorbance, fluorescence, etc.

Page 4: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

4

Source SampleCell

DetectorWavelength Selector

Detector

Source

An Absorption Instrumental Setup

An Emission or Scattering Instrumental Setup

SampleCell

Wavelength Selector

Processor

Processor

Page 5: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

5

Page 6: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

6

Sources of Radiation

A source to be used in a selected range of wavelength should have the following properties:

1 .It should generate a beam of radiation covering the wavelength range in which to be used. For example, a source to be used in the visible region should generate light in the whole visible region (340-780 nm).

2 .The output of the source should have enough radiant power depending on the technique to be used.

3 .The output should be stable with time and fluctuations in the intensity should be minimal .

Page 7: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

7

This necessitates the use of good regulated power supply. Sometimes, a double beam instrument is used to overcome fluctuations in the intensity of the beam with time. In such instruments, the beam from the source is split into two halves one goes to the sample while the other travels through a reference. Any fluctuations in the intensity of the beam traversing the sample will be the same as that traversing the reference at that moment. Subtraction of the reference beam from that of the sample results in excellent correction for fluctuations in the intensity of the beam.

Page 8: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

8

Classifications of Sources

There can be several classifications of sources. One classification can be based on where their output is in the electromagnetic spectrum. A second classification can be based on whether the source is a thermal or gas filled lamps, etc. A third method of classification can be based on whether the source is a continuous or a line source. Other classifications do exist but the one which is easier to use is the method which divide sources into either continuous or line sources

Page 9: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

9

Continuous Sources

A continuous source is a source, which has an output in a continuum of wavelengths range. An example is deuterium source in the ultraviolet (UV), which has an output in the range from 180-350 nm. Another example is the familiar tungsten lamp covering the range from 340-2500 nm, thus its output extends through the whole visible and near infrared (IR) regions.

Page 10: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

10

Line Sources

A line source is a source, which has a line output at definite wavelengths, rather than a range of wavelengths. Hollow cathode and electrodeless discharge lamps are examples of line sources which produce few sharp lines in the UV and visible (Vis). These will be discussed in details in Chapter 9. Another category of line sources is the laser

Page 11: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

11

Lasers

The term LASER is an acronym for Light Amplification by Stimulated Emission of Radiation. The first laser was introduced in 1960 and since then too many, highly important applications of lasers in chemistry were described.

Page 12: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

12

Page 13: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

13

Wavelength SelectorsWavelength Selectors

• FiltersFilters

• PrismsPrisms

• GratingsGratings

• Michelson InterferometerMichelson Interferometer

Page 14: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

14

Wavelength Selectors

Wavelength selectors are important instrumental components that are used to obtain a certain wavelength or a narrow band of wavelengths. Three types of wavelength selectors can be described:

I. Filters Filters are wavelength selectors that usually allow

the passage of a band of wavelengths and can be divided into three main categories:

Page 15: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

15

Absorption Filters

This type of filters absorbs most incident wavelengths and transmits a band of wavelengths. Sometimes, they are called transmission filters. Absorption filters are cheap and can be as simple as colored glasses or plastics. They transmit a band of wavelengths with an effective bandwidth (the effective band width is the width of the band at half height) in the range from 30-250 nm. Their transmittance is usually low where only about 10-20% of incident beam is transmitted.

Page 16: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

16

Page 17: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

17

Cut-off Filters

In this type of filters, transmittance of about 100% is observed for a portion of the visible spectrum, which rapidly decreases to zero over the remainder of the spectrum.

Page 18: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

18

Page 19: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

19

Usually, cut-off filters are not used as wavelength selectors but rather in combination of absorption filters to decrease the bandwidth of the absorption filter or to overcome problems of orders, to be discussed later. Only the combination of the two filters (common area) will be transmitted which has much narrower effective bandwidth than absorption filters alone.

Page 20: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

20

Filters

Simple, rugged (no moving parts in general)

Relatively inexpensive

Can select some broad range of wavelengths

Most often used in

field instruments

simpler instruments

instruments dedicated to monitoring a single wavelength range.

Page 21: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

21

Interference Filters

These filters are sometimes called Fabry-Perot filters and are dependent on the concept of light interference. An interference filter is composed of a transparent dielectric, like calcium fluoride, sandwiched between two semitransparent metallic films. The array is further sandwiched between two glass plates to protect the filter. The thickness of the dielectric is carefully controlled, as it is this factor, which defines the resulting wavelength. Incident polychromatic radiation hits the filter at right angles and the transmitted beam will have a very narrow bandwidth. The structure of the interference filter can be depicted as in the figure below:

Page 22: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

22

Polychromatic Radiation

Narrow Band of Radiation

Glass Plate

Metallic Film

Dielectric Material

Page 23: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

23

Page 24: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

24

Interference Wedges

It is clear from the discussion above that several interference filters are necessary to, for example, cover the visible range of the spectrum. This is not convenient as we would have to interchange filters according to wavelength of interest. To overcome this problem, a wedge machined dielectric was used. The dielectric in this case has different thicknesses and thus can transmit a wide range of wavelengths accordingly. The figure below is a schematic of a wedge interference filter:

Page 25: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

25

SlitWedge

Movement

Output Wavelengths

Glass PlateDielectric Metallic Film

Incident Radiation

Page 26: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

26

Prisms

A prism is a wavelengths selector that depends on the dispersion ability of the incident radiation by the prism material. Dispersion, as discussed earlier, is the variation of refractive index with wavelength, or frequency. Since a beam of a polychromatic light is composed of several wavelengths, the dispersion of these wavelengths will be different when they are transmitted through the prism. One can see the following dispersion pattern for white light:

Page 27: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

27

Incident beam

Red

Orange

Yellow

Green

Blue

Page 28: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

28

Two common types of prisms can be identified:

Cornu Prism: It is a 60o prism which is made either from glass or quartz. When quartz is used, two 30o prisms (one should be left handed and the other is right handed) are cemented together in order to get the 60o prism. This is necessary since natural quartz is optically active and will rotate light either to right or left hand. Cementing the left and right handed prisms will correct for light rotation and will transmit the beam in a straight direction

Page 29: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

29

Littrow Prism: A littrow prism is a 30o prism which uses the same face for input and dispersed radiation. The beam is reflected at the face perpendicular to base, due to presence of a fixed mirror. A littrow prism would be used when a few optical components are required.

Page 30: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

30

Mirror

Cornu Littrow

Page 31: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

31

It should be always remembered that glass is nontransparent to UV radiation. Therefore, when radiation in the ultraviolet is to be dispersed, a quartz prism, rather than a glass, prism should be used. Quartz serves well in both UV and Vis. It should also be appreciated that the dispersion of a prism is nonlinear since it is dependent on wavelength. Dispersion increases for shorter wavelength. Prisms are very good wavelength selectors in the range from may be 200-300 nm but are bad ones for wavelength selection above 600 nm. The nonlinear dispersion of prisms also imposes problems on the instrumental designs which will be discussed later.

Page 32: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

32

200 250 300 350 500 800

Wavelength (dispersion ability)

Page 33: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

33

Gratings

A grating is an optically flat polished surface that has dense parallel grooves. Two types of gratings are usually encountered, transmission and reflection (diffraction) gratings. Transmission gratings are seldom used in spectroscopic instruments and almost all gratings, which are used in conventional spectroscopic instruments, are of the reflection type. The groove density can be as low as 80 to several thousand (6000) lines/mm. Two common types of reflection gratings can be identified:

Page 34: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

34

1. Echellette Gratings: Typical echellette gratings contain from 300 to 2000 lines/mm but an average line density of about 1200 to 1400 lines/mm is most common. The echellette grating uses the long face for dispersion of radiation. It is the grating of choice for molecular spectroscopic instruments. As will be shortly explained, and in contrast to prisms, gratings usually have linear dispersion of radiation.

2. Echelle Gratings: These have relatively coarse grooves (~80-300 lines/mm). They use the short face for dispersion of radiation and are characterized by very high dispersion ability.

Page 35: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

35

Dispersion by Gratings

We can visualize what is going on when radiation hits the surface of a grating. Our discussion will be focused on echellete gratings but conclusions are fully applicable to all reflection gratings as well.

Page 36: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

36

d

1

X

i

r

2'

B A

DC

2 1'

Page 37: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

37

AB is the spacing between two consecutive blazes = d, combination between (2) and (3) and substitution in (1) gives:

n = d sin i + d sin r

n = d(sin i +sin r)

This relation suggests that there can be several wavelengths for each diffraction angle. For example:

Page 38: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

38

Page 39: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

39

Monochromators

A monochromator is the part of instrument responsible for producing monochromatic radiation. It is an essential component of any spectroscopic instrument and is composed of a prism or grating, as the selector, in addition to focusing elements; like mirrors or lenses. All these components are contained in a box that has an entrance and an exit slit. Two common types of monochromators can be described:

Page 40: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

40

Czerney-Turner Grating Monochromator

This is composed of a grating, two concave mirrors and two slits. The following setup can be associated with this monochromator system:

Page 41: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

41

Exit SlitEntrance Slit

Grating

FocalPlane

Page 42: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

42

Page 43: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

43

Bunsen Prism Monochromators

This type of monochromators uses a prism as the dispersion element in addition to two focusing lenses and two slits. The setup can be depicted as in the figure below:

Page 44: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

44

Focal Plane

Exit Slit

Entrance Slit

Collimating Lens Focusing Lens

Prism

Page 45: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

45

Page 46: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

46

Performance Characteristics of Grating Monochromators

Four main properties can assess the performance of grating monochromators. These include the following:

1. Spectral Purity

If the exiting beam is thoroughly studied, it will always be observed that it is contaminated with small amounts of wavelengths far from that of the instrumental setting. This is mainly due to the following reasons:

Page 47: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

47

a. Scattered radiation due to presence of dust particulates inside the monochromator as well as on various optical surfaces. This drawback can be overcome by sealing the monochromator entrance and exit slits by suitable windows.

b. Stray radiation which is radiation that exits the monochromator without passing through the dispersion element. This problem as well as all other problems related to spurious radiation, including scattering, can be largely eliminated by introducing baffles at appropriate locations inside the monochromator, as well as painting the internal walls of the monochromator by a black paint.

c. Imperfections of monochromator components, like broken or uneven blazes, uneven lens or mirror surfaces, etc, would lead to important problems regarding the quality of obtained wavelengths.

Page 48: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

48

Page 49: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

49

Page 50: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

50

Page 51: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

51

Page 52: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

52

Page 53: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

53

Page 54: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

54

Monochromator Slits

Slits of a monochromator are very important for its performance. It may be primitive to say that multiple wavelengths hitting the focal plane can emerge from the exit slit if the exit slit is too wide. On the other hand, a beam of very low power can emerge from the exit slit when the slit is too narrow. The first case leads to bad wavelength selection (bad resolution) as a mixture of wavelengths is obtained, while the other case may make it impossible for the detector to sense the low power beam (bad detectability). Therefore, the width of the slits should be carefully adjusted, where some instruments allow such adjustments.

Page 55: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

55

However, many instruments have fixed slit monochromators optimized for general purpose applications. A slit is machined from two pieces of metal to give sharp edges that are exactly aligned (same plane) and parallel.

The entrance slit of a monochromator can be

looked at as a radiation source with an image that will exactly fill the exit slit at a particular grating setting. Images from other wavelengths will align at the focal plane of the monochromator. An image of interest can be brought to focus by appropriate rotation of the grating.

Page 56: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

Choice of Slit Width

Since the effective bandwidth of a monochromator is dependent on its dispersion (eff = wD-1) and the slit width, careful choice of the slit width must be done. In most cases, monochromators are equipped with a mechanism for the adjustment of the slit width. From the discussion above, it should be appreciated that a narrower slit should be preferred for best wavelengths resolution. However, it should be clear that as the slit width gets narrower and narrower, the radiant power reaching the detector will decrease

Page 57: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

As the slit width gets narrower and narrower which is too bad for quantitative analysis. Therefore, it can be stated that the slit width should be kept as narrow as possible but with enough radiant power reaching the detector, especially in the case of qualitative analysis where we are interested in the features of the spectrum. On the other hand, wider slits can be used for quantitative analysis since, in such applications, we do not look at the fine features of the spectrum. Overall, adjustment of the slit width is a compromise between detectability and resolution; an analyst should use his own judgment according to the problem on hand.

Page 58: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

Sample Containers

Sample containers should be transparent to incident radiation. Glass is not a good material in the UV region while quartz or fused silica can be used for both UV and visible. Also, all windows and optical components through which radiation should be transmitted in a spectroscopic instrument should not absorb incident radiation.

Page 59: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

Radiation Transducers

There are several types of radiation detectors or transducers. Each detector or class of detectors can be used in a specific region of the electromagnetic spectrum. There are no universal detectors that can be used for radiation of all frequencies. The purpose of radiation transducers is to convert radiant energy into an electrical signal (current or voltage).

Page 60: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

Properties of an Ideal Transducer

An ideal transducer should have the following properties1. High sensitivity: The transducer should be capable of

detecting very small signals2. Signal to noise ratio (S/N): A high signal to noise ratio is

an important characteristic of a good transducer3. Constant response: When radiation of different

wavelengths but of the same intensity are measured, the transducer should give a constant response

4. Fast response: A short response time is essential especially for scanning instruments.

Page 61: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

5. Zero dark current: In absence of illumination, the detector output should read zero

6. Zero drift: If radiation of constant intensity hits the transducer, signal should be constant with time

7. Signal (S) should be proportional to intensity of incident radiation

S = kIHowever, in practice, a fixed value (called dark

current, Kd) is usually added to signalS = KI + Kd

We will concentrate our discussion to transducers in the UV-Vis range which are referred to as photon transducers.

Page 62: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

Photon Transducers

Several transducers can be introduced under the class of photon transducers; these include the following:

1. Photovoltaic or Barrier CellsThese are simple transducers that operate in the visible

region (350-750 nm) with maximum sensitivity at about 550 nm. The cell is composed of a copper or iron base on which a selenium semiconducting layer is deposited. The surface of semiconductor is coated with a thin semitransparent film of a metal like silver or gold. The whole array is covered with a glass plate to protect the array. The copper base and silver thin film are the two electrodes of the cell.

Page 63: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent
Page 64: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

Electrons, from selenium, are released due to breakdown of covalent bonds as a result of incident radiation and thus an equivalent number of holes is formed. The electrons migrate towards the metallic film while holes move towards the copper base. Electrons move through the external circuit towards the base and thus a current can be measured, which is dependent on the intensity of incident radiation. Barrier cells are simple, rugged, and cheap.

Page 65: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

They have the extraordinarily important advantage that they do not require an external power supply, which make them the transducers of choice for portable instruments and remote applications. However, they have some important drawbacks including low sensitivity except for intense radiation, they suffer from fatigue (signal decreases with time although the intensity is constant). They have low resistance which makes amplification of the signal difficult to achieve.

Page 66: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

2 .Vacuum Phototubes

A photo tube transducer is one of the most common and wide spreading transducers that are formed from an evacuated glass or quartz envelope that houses a semicylindrical cathode and a wire anode assembly. The cathode surface is coated with a layer of a photoemissive materials like Na/K/Cs/Sb but other formulations exist which have various sensitivities and wider wavelength ranges. The voltage difference between the cathode and the anode is usually maintained at about 90 V.

Page 67: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

The incident beam hitting the cathode surface generates electric current that is proportional to radiation intensity. This detector has better sensitivities than the barrier cell and does not show fatigue. The detector is good for the general detection of radiation intensity n the UV-Vis region and is used in most low cost instruments. The transducer is also rugged and reliable. However, a small dark current is always available.

Page 68: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent
Page 69: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

3 .Photomultiplier Tubes

A photomultiplier tube (PMT) is one of the most sensitive transducers, which can measure radiant powers of very low intensities. The operational mechanism of the PMT is similar to the vacuum phototube described above but with extra electrodes (dynodes: same surface composition as cathode) for signal amplification. When a photon hits the photo emissive cathode surface, electrons are released and are accelerated to the first dynode at a positive potential to cathode (about 90 V).

Page 70: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

Extra electrons are generated since accelerated electrons from cathode strongly hit the more positive dynode surface. Electrons are further released from this first dynode to the more positive second dynode (90 V more positive than the first dynode) resulting in release of more electrons. This process continues as electrons are accelerated to other more positive dynodes and thus huge amplification of signal results (~106 electrons for each photon).

Page 71: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent
Page 72: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

Photomultiplier tubes are limited to measurement of low radiant power radiation since high radiant powers would damage the photoemissive surfaces, due to very high amplification. It is the very high amplification, which imposes a relatively important high dark current value of the PMT. Dark current may arise due to electronic components or an increase in the temperature.

Page 73: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent
Page 74: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

A release of a single electron from the cathode surface will generate a cascade of electrons from consecutive dynodes. Cooling of the PMT is suggested to increase sensitivity where cooling to -30 oC Can practically eliminate dark current. PMTs have excellent sensitivities, fast response time and operational capabilities in both UV and visible regions of the electromagnetic spectrum.

Page 75: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

4 .Silicon Diode Transducers

A semiconductor material like silicon can be doped by an element of group V (like arsenic and antimony) would have more electrons as a group V atom (5e) replaces a silicon atom (4e). The thus doped semiconductor is called an n-type semiconductor. In contrast, when a group III element (3e) is doped in a silicon matrix, replacement of a silicon atom (4e) with a group III atom (like indium or gallium, 3e) results in the formation of a less electrons semiconductor or a p-type.

Page 76: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

A diode is a device that has a greater conductance in one direction than the other. A diode is manufactured by forming adjacent n-type and p-type regions within the same silicon or germanium single crystal. The term pn junction refers to the interface between these two regions.

A diode can be connected to a power supply (a battery) in one of two modes:

Page 77: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent
Page 78: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

A silicon diode transducer consists of a reverse-biased pn junction formed on a silicon crystal. The application of a reverse bias creates a depletion layer that will ultimately result in zero current. When a beam of radiation hits silicon diode, holes and electrons will be formed in the depletion layer thus producing a current proportional to the intensity of incident radiation.

Silicon diodes are more sensitive than phototubes but far less sensitive than photomultiplier tubes. They can be used in both UV and visible regions

Page 79: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

5 .Multichannel Photon Transducers

The simplest multichannel transducer ever made is the photographic film where the full image can be captured in one shot. However, the time required for handling and developing the film makes it difficult to practically use it in conventional instruments, although it is still in use in some techniques like x-ray diffraction spectroscopy. There are two other major classes of multichannel photon transducers, which find important applications and use in spectroscopic instruments.

Page 80: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

a. Photodiode arrays (PDA)

these are simply linear arrays of silicon diodes described above. The number of linear diodes used in each photodiode array usually 64 to 4096 with 1024 silicon diodes as the most common. One can imagine the complexity of the electronic circuitry used in such an array as well as the data handling and manipulation requirements. The entrance slit is usually fixed at a size enough to fill the surface area of one silicon diode. The entire spectrum can thus be instantaneously recorded. The arrays are also called diode array detectors (DAD).

Page 81: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent
Page 82: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent
Page 83: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

b. Charge Transfer and Charge Coupled Transducers

The photosensitive elements are, in contrary to PDAs, arranged in two dimensions in both charge injection devices (CID) and charge-coupled devices (CCD). Therefore, these are very similar to photographic films. For example, a commercially available transducer is formed from 244 rows with each row containing 388 detector elements. This will add up to a two-dimensional array holding 16672 detector elements (pixels) on silicon chip that is 6.5 mm by 8.7 mm.

Page 84: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent
Page 85: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent
Page 86: 1 Components of Optical Instruments. 2 Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering Spectroscopic instruments dependent

The full description of the system and its mechanism will not be covered here as this is behind the scope of this course. However, we should qualitatively know that these important transducers function by first collecting the photogenerated charges in different pixels and then measuring the quantity of the charge accumulated in a brief period. Measurement is accomplished by transferring the charge from a collection area to a detection area.