2 nd annual ardent workshop

31
2 nd ANNUAL ARDENT WORKSHOP Quality assessment of hadrontherapy fields with TEPCs Paolo Colautti, INFN 2 nd Annual ARDENT Workshop, Milano, 14-18 October 2013

Upload: ayasha

Post on 24-Feb-2016

60 views

Category:

Documents


0 download

DESCRIPTION

2 nd Annual ARDENT Workshop. Quality assessment of hadrontherapy fields with TEPCs Paolo Colautti, INFN. 2 nd Annual ARDENT Workshop, Milano, 14-18 October 2013. The radiation quality concept . - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: 2 nd Annual  ARDENT Workshop

2nd ANNUAL ARDENT WORKSHOP

Quality assessment of hadrontherapy fields with

TEPCs

Paolo Colautti, INFN

2nd Annual ARDENT Workshop, Milano, 14-18 October 2013

Page 2: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

The radiation quality concept

The radiation field quality is meant as a physical measurable quantity, which is significant for primary effects on a biological system

We define the measured y-spectrum in 1-2 µm sites as the radiation-field

microdosimetric quality

Some detectors measure only , which is used as quality mark of the radiation field.

The couple of values and is a better quality mark of the radiation field.

RBEµ, the microdosimetric assessment of RBE, can be an accurate quality mark for a given biological end-point.

Page 3: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Therapeutic beam constrains

Fluence rates

Proton continuos beam

BNCT facility

1 cm TEPC ought to measure at counting rates bigger than 106 s-1

1 mm TEPC will measure at counting rates bigger than 104 s-1

CNAO protons per spill

CNAO carbon ions per spill

Page 4: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Energy calibration technique in F ~ 1cm cylindrical TEPCs

Vacuum tide Al cylinder

Tissue-equivalent conductive plastic

Page 5: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Energy calibration technique in F ~ 1mm cylindrical TEPCs

a-source

a-source

Gas IN

Gas OUT

Cathode H.V.

Anode pulse

Guard tubes H.V.

Page 6: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Measurements at the Nice therapeutic proton beam

Page 7: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Insulator surfaces damage without electric-field tubes

75 µm

Page 8: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Solution: a cavity inside the insulator to switch off the the electronic avalanche

0.9 mm

0.9

mm

0.150 mm

0.200 mm

Page 9: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Mini TEPC of 2.7 mm of external diameter

Page 10: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Vacuum and gas flow apparatus

Membrane pump

Pressure readout

Exhaust

GasMFCMFC controller

Turbopump

Solenoid valve

Page 11: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Electronic chain

HVanode

HVcathode

Pulser

+18/-6 volt

Low LET amplifier

Medium LET amplifier

High LET amplifier

Low LET MCA

Medium LET MCA

High LET MCA

Page 12: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Electronic chain

Page 13: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Before measuring: linearity checks

Integral linearity

High LET

Page 14: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Initial data from the 3 MCA

Page 15: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Volt calibration

Page 16: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Logaritmic compaction

1st 2nd Bth

log yi log yi+1

1st

….. …..Equal log intervals

Example: B = 60

First decade y1 = 1.0391, y2 = 1.0798, y3 …. Δ1 = 0.04065

Second decade y1 = 10.0391, y2 = 10.0798, y3 …. Δ1 = 0.4065

Original countsCount density in compacted intervals

Page 17: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Sub-spectra junction

Sub-spectra junction is not feasible if they do not superimpose

Page 18: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

The pulse-height frequency distribution

Page 19: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

The pulse-height weighted distribution

Page 20: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Lineal energy calibration

ICRU 49 data

142 keV/µm

Calibration factor: 142/475 = 0.299

Page 21: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

∆E calculation in CSDA

Page 22: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Maximum ∆E-lost calculation in CSDA

Page 23: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Lineal-energy dose distribution

Fractional part of the visual area is the fractional contribution to the the absorbed dose by the corresponding y-events

Blue points are used to linearly extrapolate the frequency y-values down to 0.01 keV/µm

Page 24: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Weighting the dose distribution with a biological-significant function

Fractional part of the visual area is the fractional contribution to the effective-dose of corresponding y-

events

The weight is 1 for y-values < 10 keV/µmThe weight is < 1 for y-values ≈> 200

keV/µm

Page 25: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Splitting the dose spectrum in its neutron and gamma components

Page 26: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Quality of the conjunctive-melanoma proton-therapeutic beam

Page 27: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

The therapeutic beam quality Iso dose map

Iso effective-dose map

Iso RBEµ map

Page 28: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Two mini TEPCs for quality measurements of BNCT radiation fields

Page 29: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

The twin TEPC for BNCT

Page 30: 2 nd Annual  ARDENT Workshop

30

BNCT dose components twin TEPC measurements

Page 31: 2 nd Annual  ARDENT Workshop

P.Colautti 2nd ARDENT workshop. Milano 14-18 October 2013

Suggested lectures

• Glenn F.Knoll. Radiation Detection and Measurements. John Wiley & Sons, NY, 1979

• ICRU Report 36. Microdosimetry. 1983