Physics & Measurement

Peter C A KamPeter C A KamProfessor of Anaesthesia, UNSWProfessor of Anaesthesia, UNSW

St George HospitalSt George Hospital

Page 2: 20275003 Physics Measurement 2

The Gas Laws

Boyle’s LawBoyle’s Law

Charles’ LawCharles’ Law

The Third Perfect Gas LawThe Third Perfect Gas Law

The Ideal Gas EquationThe Ideal Gas Equation

Henry’s LawHenry’s Law

Dalton’s LawDalton’s Law

Page 3: 20275003 Physics Measurement 2

Boyle’s Law or First Gas Law

Boyle's LawBoyle's Law at a constant at a constant temperaturetemperature,,

the volume of a given mass of gasthe volume of a given mass of gasvaries inversely with its absolute pressure,varies inversely with its absolute pressure,

or,or,

PV = k1PV = k1

Page 4: 20275003 Physics Measurement 2

Charles’ Law or Second Gas Law

Charles' LawCharles' Law at a constant at a constant pressurepressure,,

the volume of a given mass of gasthe volume of a given mass of gasvaries proportionately to its absolute temperature,varies proportionately to its absolute temperature,

or,or,

V/T = k2V/T = k2

Page 5: 20275003 Physics Measurement 2

The Third Gas Law

The Third Perfect Gas LawThe Third Perfect Gas Law at a constant at a constant volumevolume,,

the absolute pressure of a given mass of gasthe absolute pressure of a given mass of gasvaries proportionately to its absolute temperature,varies proportionately to its absolute temperature,

or,or,

P/T = k3P/T = k3

Page 6: 20275003 Physics Measurement 2

Ideal Gas Equation

for 1 mol of any perfect gas,for 1 mol of any perfect gas,the the universal gas constantuniversal gas constant

R = PV/TR = PV/T

or, where n = number of mol of gas,or, where n = number of mol of gas,

PV = nRTPV = nRT

Page 7: 20275003 Physics Measurement 2

A Mole

the quantity of any substance containing the the quantity of any substance containing the

same number of particles as there are atoms in same number of particles as there are atoms in

0.012 kg of 0.012 kg of 1212CarbonCarbon

1 mol ~ 6.0223 x 101 mol ~ 6.0223 x 102323

Avogadro's numberAvogadro's number

Page 8: 20275003 Physics Measurement 2

Avogadro’s Hypothesis

equal volumes of gases, at the same equal volumes of gases, at the same temperature and pressure contain equal temperature and pressure contain equal numbers of moleculesnumbers of molecules

STPSTP T = T = 273.15 K273.15 K (0°C)(0°C) P = P = 101.325 kPa101.325 kPa (760 mmHg)(760 mmHg) for any gas at STP, 1 mol ~ for any gas at STP, 1 mol ~ 22.4 litre22.4 litre

Page 9: 20275003 Physics Measurement 2

Henry’s Laws

Henry's LawHenry's Law

at a constant at a constant temperaturetemperature, the amount of a gas , the amount of a gas dissolveddissolved in a liquid is directly proportional to the in a liquid is directly proportional to the partial pressurepartial pressure of that gas in equilibrium with that of that gas in equilibrium with that liquidliquid

Page 10: 20275003 Physics Measurement 2

Dalton's Law of Partial Pressures

Dalton's Law of Partial PressuresDalton's Law of Partial Pressures

in a mixture of gases, the pressure exerted by each in a mixture of gases, the pressure exerted by each

gas is equal to the pressure which would be exerted if gas is equal to the pressure which would be exerted if

that gas alone were presentthat gas alone were present

Page 11: 20275003 Physics Measurement 2

Solubility

Bunsen Solubility CoefficientBunsen Solubility Coefficient

the the volumevolume of gas, corrected to of gas, corrected to STPSTP, which dissolves , which dissolves

in one in one unit volumeunit volume of the liquid at the temperature of the liquid at the temperature

concerned, where the partial pressure of the gas concerned, where the partial pressure of the gas

concerned is concerned is 1 atmosphere1 atmosphere

Page 12: 20275003 Physics Measurement 2

Solubility

Ostwald Solubility CoefficientOstwald Solubility Coefficient the the volumevolume of gas which dissolves in one of gas which dissolves in one unit unit

volumevolume of the liquid at the of the liquid at the temperaturetemperature concerned concerned the temperature must be specifiedthe temperature must be specified it is it is independentindependent of pressure of pressure

as the pressure rises the number of molecules of gas as the pressure rises the number of molecules of gas in the liquid phase increases,in the liquid phase increases, however, when measured at the higher pressure the however, when measured at the higher pressure the volume is the samevolume is the same

Page 13: 20275003 Physics Measurement 2

Partition Coefficient

Partition CoefficientPartition Coefficient the the ratioratio of the amount of a substance present in one of the amount of a substance present in one

phase as compared with than in anotherphase as compared with than in another

the two phases being of the two phases being of equal volumeequal volume

the the temperaturetemperature must be specified, and must be specified, and

the phases being in the phases being in equilibriumequilibrium eg. blood:gas and tissue:bloodeg. blood:gas and tissue:blood

Page 14: 20275003 Physics Measurement 2

Diffusion

the spontaneous movement of molecules or the spontaneous movement of molecules or other particles in solution, owing to their random other particles in solution, owing to their random thermal motion, to reach a uniform concentration thermal motion, to reach a uniform concentration throughout the solventthroughout the solvent

the constant random thermal motion of the constant random thermal motion of molecules, in gaseous or liquid phases, which molecules, in gaseous or liquid phases, which leads to the net transfer molecules from a region leads to the net transfer molecules from a region of higher concentration to a region of lower of higher concentration to a region of lower concentration (concentration (thermodynamic activitythermodynamic activity))

Page 15: 20275003 Physics Measurement 2

Fick’s Law of Diffusion

the the rate of transferrate of transfer of a gas through a sheet of of a gas through a sheet of tissue is,tissue is, proportional to the proportional to the areaarea available for transfer available for transfer proportional to the gas proportional to the gas tension differencetension difference inversely proportional to the tissue inversely proportional to the tissue thicknessthickness

V gas = k.A (P gas 1 - Pgas 2) T

Page 16: 20275003 Physics Measurement 2

Determinants of diffusion

Characteristics of the GasCharacteristics of the Gas

Pressure GradientPressure Gradient

Membrane CharacteristicsMembrane Characteristics

Page 17: 20275003 Physics Measurement 2

Gas Characteristics

Molecular WeightMolecular Weight VV 1/ 1/MWMW

Graham's LawGraham's Law: relative rate of diffusion is inversely : relative rate of diffusion is inversely proportional to the square root of the gas molecular proportional to the square root of the gas molecular weightweight thus, lighter gases diffuse faster in gaseous media thus, lighter gases diffuse faster in gaseous media

than heavier gasesthan heavier gases lighter molecules for given energy have faster lighter molecules for given energy have faster

velocitiesvelocities therefore, Otherefore, O22 diffuses more rapidly than CO diffuses more rapidly than CO22 in the in the

gas phase (1.17 : 1)gas phase (1.17 : 1)

Page 18: 20275003 Physics Measurement 2

Gas Characteristics: Solubility

Henry's LawHenry's Law the amount of a gas which dissolves in unit the amount of a gas which dissolves in unit

volume of a liquid, at a given temperature, volume of a liquid, at a given temperature, is directly proportional to the partial is directly proportional to the partial pressure of the gas in the equilibrium phase pressure of the gas in the equilibrium phase

Page 19: 20275003 Physics Measurement 2

Gas Characteristics: Solubility

relative solubilities of COrelative solubilities of CO22 & O & O22 in water ~ in water ~

24:124:1 combining this with Graham's Law from combining this with Graham's Law from

above,above,the relative rates of diffusion the relative rates of diffusion

from alveolus to rbc for COfrom alveolus to rbc for CO22:O:O22 ~ ~ 20.7 : 120.7 : 1 solubility determines the limitation to the rate of solubility determines the limitation to the rate of

diffusion, gases being eitherdiffusion, gases being either

• diffusion limiteddiffusion limited, as for CO, as for CO

• perfusion limitedperfusion limited, as for N, as for N22OO

Page 20: 20275003 Physics Measurement 2

Diffusion, k

further, the diffusion of gas across a further, the diffusion of gas across a membrane, or into or out of a liquid, is membrane, or into or out of a liquid, is proportional to the gases proportional to the gases solubilitysolubility in the in the liquidliquid COCO22 being more soluble than O being more soluble than O22 diffuses far more diffuses far more

rapidly across the alveolar membrane and into the rapidly across the alveolar membrane and into the RBCRBC

NN22O being far more soluble than NO being far more soluble than N22 may diffuse may diffuse

into and expand closed cavities during induction of into and expand closed cavities during induction of anaesthesiaanaesthesia

Page 21: 20275003 Physics Measurement 2

Osmotic Pressure

Page 22: 20275003 Physics Measurement 2

Osmosis & pressure

OsmosisOsmosis : the movement of solvent across : the movement of solvent across a semipermeable membrane, down a a semipermeable membrane, down a thermodynamic activity gradient for that thermodynamic activity gradient for that solventsolvent

Osmotic PressureOsmotic Pressure : the pressure which : the pressure which would be required to prevent the movement would be required to prevent the movement of solvent across a semipermeable of solvent across a semipermeable membrane, down a thermodynamic activity membrane, down a thermodynamic activity gradient for that solventgradient for that solvent

Page 23: 20275003 Physics Measurement 2

Osmolality

the number of osmotically active particles the number of osmotically active particles (osmoles) per (osmoles) per kilogramkilogram of solvent of solvent

depression of depression of freezing pointfreezing point of a solution is of a solution is directly proportional to the osmolalitydirectly proportional to the osmolality 1 mol of a solute added to 1 kg of water 1 mol of a solute added to 1 kg of water

depresses the freezing point by depresses the freezing point by 1.86°C1.86°C presence of increased amounts of solute presence of increased amounts of solute

also lowers the also lowers the vapour pressurevapour pressure of the of the solvent, viz…….solvent, viz…….

Page 24: 20275003 Physics Measurement 2

Raoult’s Law

the depression or lowering of the the depression or lowering of the vapour vapour pressurepressure of a solvent is proportional to the of a solvent is proportional to the molar concentration of the solutemolar concentration of the solute

as the presence of a solute decreases the as the presence of a solute decreases the vapour pressure, making the solvent less vapour pressure, making the solvent less volatile, so the volatile, so the boiling pointboiling point is raised is raised

Page 25: 20275003 Physics Measurement 2

Raoult’s Law

These phenomena, These phenomena, depression of freezing point, depression of depression of freezing point, depression of

vapour pressure vapour pressure and elevation of boiling point, being related and elevation of boiling point, being related

to osmolarity to osmolarity are termed are termed colligative propertiescolligative properties of a of a

solutionsolution

Page 26: 20275003 Physics Measurement 2

Osmotic Pressure

1 mol1 mol of any solute dissolved in of any solute dissolved in 22.4 litres22.4 litres of of solution at solution at 0°C0°C will generate an osmotic will generate an osmotic pressure of pressure of 1 atmosphere1 atmosphere

in mixed solutions the osmotic pressure is the in mixed solutions the osmotic pressure is the sum of the individual molalitiessum of the individual molalities

Page 27: 20275003 Physics Measurement 2

Osmotic Pressure

> 99% of the plasma osmolality is due to electrolytes> 99% of the plasma osmolality is due to electrolytes

contribution of the plasma proteins: contribution of the plasma proteins: 1 mosmol/l 1 mosmol/l

normal rbc's lyse at osmolalities ~ 200 mosmol/lnormal rbc's lyse at osmolalities ~ 200 mosmol/l

as capillaries are relatively impermeable to protein, as capillaries are relatively impermeable to protein,

this generates an osmotic pressure difference this generates an osmotic pressure difference

between the plasma and the interstitial fluid, the between the plasma and the interstitial fluid, the

plasma oncotic pressure plasma oncotic pressure ~ 26 mmHg~ 26 mmHg

Page 28: 20275003 Physics Measurement 2

COLLIGATIVE PROPERTIES OF A SOLVENT

• Presence of solute stabilises solvent molecules

• More stable solvent molecules cause

(a) Increase in boiling point

(b) Increase in osmotic pressure

(c) Decrease in freezing point

(d) Decrease in vapour pressure

of solvent.

Page 29: 20275003 Physics Measurement 2

COLLIGATIVE PROPERTIES

1 Osmole of solute leads to ;

a) Boiling point of water increase by 0.52oC

b) Osmotic pressure increase by 2267kPa (17000)

c) Freezing point 1.85oC depression

d) Vapour pressure 0.04kPa (0.3 mmHg)

Page 30: 20275003 Physics Measurement 2

FREEZING POINT OSMOMETER

- 7OC

Sample

Thermometer

EthyleneGlycol

Thermocouple

StirringWire

Page 31: 20275003 Physics Measurement 2

MEASUREMENT OF OSMOLALITY

Time

Supercooling

True freezing pointTemperature

Page 32: 20275003 Physics Measurement 2

GAS OR LIQUID FLOW

Page 33: 20275003 Physics Measurement 2

Hagen-Poiseuille

where flow is where flow is laminarlaminar,, etaeta (h) (h) = = viscosityviscosity of the fluid in pascal seconds of the fluid in pascal seconds there are no eddies or turbulencethere are no eddies or turbulence flow flow

• is greatest at the is greatest at the centrecentre, being ~ twice the mean, being ~ twice the mean

• near the wall near the wall 0 0

• is directly proportional to the driving pressureis directly proportional to the driving pressure

Q = r4P

8l

Page 34: 20275003 Physics Measurement 2

Laminar Flow

but as R = dP/Q, sobut as R = dP/Q, so

thus, resistance in inversely proportional thus, resistance in inversely proportional to the to the (radius)(radius)44

R = 8nl r4

Page 35: 20275003 Physics Measurement 2

Turbulent Flow

the velocity profile across the lumen is lostthe velocity profile across the lumen is lost flow becomes directly proportional to the flow becomes directly proportional to the

square rootsquare root of the driving pressure of the driving pressure pressure flow is not linear and resistance is not pressure flow is not linear and resistance is not

constantconstant flow at which R is measured must be specifiedflow at which R is measured must be specified

other factors in turbulent flow follow,other factors in turbulent flow follow,

((rho) = densityrho) = density of the fluid in kg.m of the fluid in kg.m-3-3

Q = k r2 P l

Page 36: 20275003 Physics Measurement 2

Turbulent Flow

thus, radius has less effect, cf. laminarthus, radius has less effect, cf. laminar likelihood of the onset of turbulent flow is likelihood of the onset of turbulent flow is

predicted by the predicted by the Reynold's numberReynold's number

d =d = the the diameterdiameter of the tube of the tube v =v = the the velocityvelocity of flow of flow == rho, the rho, the densitydensity of the fluid in kg.m of the fluid in kg.m-3-3

== eta, the eta, the viscosityviscosity of the fluid in of the fluid in pascal secondspascal seconds

Re = vd

Page 37: 20275003 Physics Measurement 2

Turbulent Flow

empirical studies show that for empirical studies show that for cylindrical tubes, if cylindrical tubes, if Re > 2000Re > 2000 turbulent turbulent flow becomes more likelyflow becomes more likely

for a given set of conditions there is a for a given set of conditions there is a critical velocitycritical velocity at which Re = 2000 at which Re = 2000

the breakpoint for turbulent flow versus the breakpoint for turbulent flow versus Re also varies with the nature of the Re also varies with the nature of the fluidfluid eg. for blood turbulent flow at eg. for blood turbulent flow at Re > 1000Re > 1000

Page 38: 20275003 Physics Measurement 2

Viscosity

for a given set of conditions, flow is inversely proportional for a given set of conditions, flow is inversely proportional to viscosityto viscosity

blood viscosity increases with,blood viscosity increases with, low temperatureslow temperatures increasing ageincreasing age cigarette smokingcigarette smoking increasing haematocritincreasing haematocrit abnormal elevations of plasma proteinsabnormal elevations of plasma proteins

the viscosity of blood is anomalous due to the presence of the viscosity of blood is anomalous due to the presence of cellscells behavior is said to be behavior is said to be non-newtoniannon-newtonian

Page 39: 20275003 Physics Measurement 2

Tension

Laplace's LawLaplace's Law

P = T.h.(1/rP = T.h.(1/r11 + 1/r + 1/r22))

T T = the tangential force in N/m= the tangential force in N/m acting along a length of wallacting along a length of wall

h h = the thickness of the wall (usually = the thickness of the wall (usually smallsmall))

Page 40: 20275003 Physics Measurement 2

Laplace’s Law

thus, for straight tubes,thus, for straight tubes,

P = T.h./rP = T.h./r

and, for and, for spheresspheres,,

P = 2T.h/rP = 2T.h/r

Page 41: 20275003 Physics Measurement 2

Laplace’s Law

thus, as vessel diameter becomes thus, as vessel diameter becomes smaller, the collapsing force becomes smaller, the collapsing force becomes greater greater

this can lead to vessel closure at low this can lead to vessel closure at low pressures, the pressures, the critical closing critical closing pressurepressure

seen in alveoli, leading to instability with seen in alveoli, leading to instability with small alveoli tending to fill larger ones small alveoli tending to fill larger ones major action of surfactant is to maintain major action of surfactant is to maintain

alveolar stabilityalveolar stability

Page 42: 20275003 Physics Measurement 2

Measurement of Gas Volumes and Flows

Direct methodsDirect methods

Indirect methodsIndirect methods

Page 43: 20275003 Physics Measurement 2

WET SPIROMETER

••

••

••

••

••

••

Recorder

CO2 Absorber

Disadvantages1. High inertia2. Inaccurate at high respiratory rate or

FVC

Page 44: 20275003 Physics Measurement 2

VITALOGRAPH

Bellows

Recorder

Patient

Disadvantage : Patient effort dependent Bellows collect expired gas.

Only measures forced expiratory volumes and flows.

Page 45: 20275003 Physics Measurement 2

WRIGHT RESPIROMETER

Vane

Channels

Gas Flow

1. Gas stream directed by tangential slits to vane

2. Gas flow drives spinning vane

3. Spinning vane activates gears to record flow

4. Over reads at peak flow

Under reads at continuous flow

Page 46: 20275003 Physics Measurement 2

DRAGER VOLUMETER

Gasflow

1. Consists of 2 interlocking dumb bell rotors2. More accurate3. Affected by water vapour

Page 47: 20275003 Physics Measurement 2

INDIRECT MEASUREMENT OFGAS VOLUMES

1. Magnetometers

2. Pneumographs

3. Capacitance spirometry

4. Respiratory inductance plethysmograph

Page 48: 20275003 Physics Measurement 2

MAGNETOMETERS

1. Electromagnets attached to chest wall andabdomen

2. Electromagnetic field generated .

3. Chest and abdominal diameter changes –alter magnetic filed.

4. Disadvantage : Inaccurate ++

Page 49: 20275003 Physics Measurement 2

PNEUMOGRAPH

Pressure Transducer

PressureTransducer

Disadvantage : frequent recalibration required

Chest wall

Page 50: 20275003 Physics Measurement 2

CAPACITANCE SPIROMETRY

C

Top plate

Bottom plate

Used for apnoeic monitoring

Chest wall

Page 51: 20275003 Physics Measurement 2

INDUCTANCE SPIROMETRY

Oscillator

RECORDER COMPUTER

Chest wall

Page 52: 20275003 Physics Measurement 2

GAS FLOW MEASUREMENT

1. Variable orifice (constant pressure drop)flowmeter eg. rotameter

2. Variable pressure – fixed orifice flowmeter

Page 53: 20275003 Physics Measurement 2

ROTAMETER

1. Variable orifice flowmeter

2. Gas flow controlled by control value at bottom of rotameter

3. Vertical tube with tapering internal diameter- wider at the top- narrower at the bottom

4. Bobbin - acts as indicator of flow

5. Pressure drops across the annular space around bobbin opposes downward pressure produced by weight of bobbin.

Page 54: 20275003 Physics Measurement 2

ROTAMETER

W

Pressure drop [P1 – P2] balances weight [W] of bobbin

Bobbin

Rotameter tube

Gas Flow

P1

P2

Page 55: 20275003 Physics Measurement 2

ROTAMETER

1. Non – linear scale

2. At lower flow;- bobbin length > distance between bobbin and glass (d)- Laminar flow

3. At high flows;- bobbin length < d- turbulent flow

4. Accuracy + 2%

Page 56: 20275003 Physics Measurement 2

PNEUMOTACHOGRAPH

1. Measure gas flows.

2. Types(a) Fixed Resistance

Gas flow across fixed resistance differentialpressure signal & flow eg. screen and fleisch pneumotachograph.

(b) Hot wireSignal produced by gas flow cooling a heatedresistance wire.

(c) Pitot tube

Page 57: 20275003 Physics Measurement 2

SCREEN PNEUMOTACHOGRAPH

ScreenGas

P1 P2

Pressure difference flow

Page 58: 20275003 Physics Measurement 2

FLEISCH PNEUMOTACHOGRAPH

HEATING COIL

HEATING COIL

Fine bore parallel tubeEnsure laminar flow

P1 - P2

Heating coil to prevent water condensation

GAS FLOW

Page 59: 20275003 Physics Measurement 2

PITOT TUBE PNEUMOTACHOGRAPH

Upstream Downstream P2

P1 (total) (static P)

P1 - P2 velocity of gas

GAS FLOW

Page 60: 20275003 Physics Measurement 2

Heat & Temperature

Page 61: 20275003 Physics Measurement 2

Heat & Temperature

HeatHeat : a form of energy, being the state : a form of energy, being the state of of thermal agitationthermal agitation of the molecules of of the molecules of a substance, which may be transferred a substance, which may be transferred by,by,

conductionconduction through a substance through a substance

convectionconvection by a substance, and by a substance, and

radiationradiation as electromagnetic waves as electromagnetic waves

Page 62: 20275003 Physics Measurement 2

Heat & Temperature

TemperatureTemperature : is the physical state of a : is the physical state of a

substance which determines whether or not the substance which determines whether or not the

substance is in thermal equilibrium with its substance is in thermal equilibrium with its

surroundings, surroundings, heat energyheat energy being transferred being transferred

from a region of higher temperature to a region of from a region of higher temperature to a region of

lower temperaturelower temperature

Page 63: 20275003 Physics Measurement 2

Heat & Temperature

KelvinKelvin

the SI unit of thermodynamic temperaturethe SI unit of thermodynamic temperature

equal to 1/273.16 of the absolute temperature of equal to 1/273.16 of the absolute temperature of

the the triple pointtriple point of water of water

the temperature at which ice, water and water the temperature at which ice, water and water

vapour are all in equilibriumvapour are all in equilibrium Celsius scaleCelsius scale

Temperature (K) = Temperature (°C) + 273.15Temperature (K) = Temperature (°C) + 273.15

in Celsius the triple point of water is in Celsius the triple point of water is 0.01 °C0.01 °C

Page 64: 20275003 Physics Measurement 2

Critical Temperature

the temperature above which a gas cannot be the temperature above which a gas cannot be

liquified by pressure aloneliquified by pressure alone NN22OO = 36.5 °C= 36.5 °C

OO22 = -119 °C= -119 °C

GasGas:: a substance in the gaseous phasea substance in the gaseous phaseaboveabove its critical T its critical T

VapourVapour:: a substance in the gaseous phasea substance in the gaseous phasebelowbelow its critical its critical TT

Page 65: 20275003 Physics Measurement 2

Critical Pressure

the pressure at which a gas liquifies at the pressure at which a gas liquifies at its critical Tits critical T

NN22O ~ O ~ 73 bar 73 bar @@ 36.5 °C36.5 °C NN22O ~ 52 bar O ~ 52 bar @ 20.0 °C@ 20.0 °C

Pseudo-Critical TemperaturePseudo-Critical Temperature for a mixture of gases at a specific pressure, the for a mixture of gases at a specific pressure, the

specific temperature at which the individual gases specific temperature at which the individual gases may separate from the gaseous phase may separate from the gaseous phase

NN22O 50% / OO 50% / O22 50% 50% = - 5.5 °C= - 5.5 °C

• for cylinders (most likely at 117 bar)for cylinders (most likely at 117 bar) NN22O 50% / OO 50% / O22 50% 50% = - 30 °C= - 30 °C

• for piped gasfor piped gas

Page 66: 20275003 Physics Measurement 2

Adiabatic Change

the change of the change of physical statephysical state of a gas, without the of a gas, without the transfer of heat energy to or from the surrounding transfer of heat energy to or from the surrounding environmentenvironment rapid rapid expansionexpansion & energy required to overcome Van & energy required to overcome Van

der Waal's forces of attraction, as this energy cannot be der Waal's forces of attraction, as this energy cannot be

gained from the surroundings, it is taken from the kinetic gained from the surroundings, it is taken from the kinetic

energy of the molecules energy of the molecules basis of the basis of the cryoprobecryoprobe

rapid rapid compressioncompression, the energy level between molecules , the energy level between molecules

is reduced, as this energy cannot be dissipated to the is reduced, as this energy cannot be dissipated to the

surroundings, it is transferred to the kinetic energy of the surroundings, it is transferred to the kinetic energy of the

moleculesmolecules

Page 67: 20275003 Physics Measurement 2

T Measurement: Non-electrical

mercurymercury thermometers thermometers

accurate, reliable, cheapaccurate, reliable, cheap

readily made into a thermostat, or max. reading readily made into a thermostat, or max. reading

formform

Angulated constriction at base of stem prevents Hg Angulated constriction at base of stem prevents Hg

column returning to bulb via surface tension forcescolumn returning to bulb via surface tension forces

requires 2-3 mins to reach thermal equilibriumrequires 2-3 mins to reach thermal equilibrium

unsuitable for insertion in certain orificesunsuitable for insertion in certain orifices

Page 68: 20275003 Physics Measurement 2

T Measurement: Non-electrical

alcoholalcohol thermometers thermometers

cheaper than mercurycheaper than mercury

useful for very low T, mercuryuseful for very low T, mercury solid at solid at -39°C-39°C

unsuitable for high T, alcohol boils at unsuitable for high T, alcohol boils at 78.5°C78.5°C

expansion also tends to be less linear than expansion also tends to be less linear than

mercurymercury

Bimetallic stripsBimetallic strips

Bourdon gauge Bourdon gauge pressurepressure

Page 69: 20275003 Physics Measurement 2

T Measurement: Electrical

resistanceresistance thermometer thermometer

metalsmetals R R linearlylinearly with with TT

frequently use a platinum wire resistor, or frequently use a platinum wire resistor, or

similarsimilar

accuracy improved with a Wheatstone accuracy improved with a Wheatstone

bridgebridge

Page 70: 20275003 Physics Measurement 2

Resistance Thermometer

Platinum Wire T α R (linear) Disadv. R Not sensitive α R α T wire Battery T

Page 71: 20275003 Physics Measurement 2

T Measurement: Electrical

thermistorthermistor metal oxidesmetal oxides R R exponentiallyexponentially with with TT made exceeding smallmade exceeding small rapid thermal equilibrationrapid thermal equilibration narrow reference rangenarrow reference range different thermistors for different scalesdifferent thermistors for different scales accuracy improved with a Wheatstone bridgeaccuracy improved with a Wheatstone bridge Accuracy reduced with exposure to severe T, Accuracy reduced with exposure to severe T,

egeg. . sterilisationsterilisation

Page 72: 20275003 Physics Measurement 2

Thermistor

Thermistor oxide of R exponentially with Temp. metal Adv : small - rapid change

- accessible to remote location Disadv : Drift in calibration

R To

Page 73: 20275003 Physics Measurement 2

T Measurement: Electrical

thermocouplethermocouple based on the Seebeck effectbased on the Seebeck effect at the junction of two dissimilar metals a small at the junction of two dissimilar metals a small

voltage is produced, the magnitude of which is voltage is produced, the magnitude of which is

determined by the temperature determined by the temperature metals such as copper and constantan (Cu+Ni alloy)metals such as copper and constantan (Cu+Ni alloy) requires a constant reference temperature at the requires a constant reference temperature at the

second junction of the electrical circuitsecond junction of the electrical circuit may be made exceeding small and introduced may be made exceeding small and introduced

almost anywherealmost anywhere

Page 74: 20275003 Physics Measurement 2

Thermocouple – “Seebeck effect”

Reference Junction Copper Constantan Junction Potential

mV Measuring junction Temp

Page 75: 20275003 Physics Measurement 2

Thermocouple

Junction of 2 different metals P. Diff (α To) Seebeck effect Metal 1 eg Cu V Metal 2 eg. Constantin

Ref Measured Adv. Large linear range Temp Temp can be very small (eg.Ice) Disadv. Small output 40μv

Page 76: 20275003 Physics Measurement 2

Specific Heat Capacity

heat required to raise the temperature of 1 kg heat required to raise the temperature of 1 kg

of a substance by 1 K (J/kg/K)of a substance by 1 K (J/kg/K) water SHCwater SHC = 4.18 kJ/kg/K or, 1 kcal/kg/K= 4.18 kJ/kg/K or, 1 kcal/kg/K blood SHCblood SHC = 3.6 kJ/kg/K= 3.6 kJ/kg/K

infusion of 2000 ml of blood at 5°C, requiring infusion of 2000 ml of blood at 5°C, requiring

warming to 35°C, warming to 35°C, require require 2 kg x 3.6 kJ/kg/°C x (35-5)°C = 216 kJ2 kg x 3.6 kJ/kg/°C x (35-5)°C = 216 kJ

would result in the person's temperature falling by would result in the person's temperature falling by

~ 1°C~ 1°C

Page 77: 20275003 Physics Measurement 2

Specific Latent Heat

the heat required to convert 1 kg of a substance the heat required to convert 1 kg of a substance from one phase to another at a given temperaturefrom one phase to another at a given temperature latent heat of vaporisation latent heat of vaporisation (from liquid to vapour)(from liquid to vapour)

LHV of waterLHV of water at 100°C = 2.26 MJ/kgat 100°C = 2.26 MJ/kg at 37°Cat 37°C = 2.42 MJ/kg = 2.42 MJ/kg the lower the T the greater the latent heat requiredthe lower the T the greater the latent heat required

as T rises, the latent heat falls until ultimately it as T rises, the latent heat falls until ultimately it reaches reaches zerozero at a point which corresponds with the at a point which corresponds with the critical temperaturecritical temperature = 373°C = 373°C

Page 78: 20275003 Physics Measurement 2

Humidity

Page 79: 20275003 Physics Measurement 2

ABSOLUTE HUMIDITY

• Mass of water vapour (g) in a given

volume of air (m3)

numerically = mg / 1L

• Fully saturated air

@ 20oC contains 17mg/L water

@ 37oC contains 44mg/L water

Page 80: 20275003 Physics Measurement 2

RELATIVE HUMIDITY

• Defined as ratio of mass of water vapour in a given volume of air to the mass required

to fully saturate that volume of air at a given temperature. (%).

• By ideal gas equation, mass is proportional to number of moles present.

Relative humidity = actual vapour pressure saturated vapour pressure

Page 81: 20275003 Physics Measurement 2

HAIR HYGROMETER

Principle : Hair lengthens as humidity increases

Accuracy : - Low - Accurate between RH 15-85% - very simple & cheap

RH LHair

Page 82: 20275003 Physics Measurement 2

WET AND DRY BULB HYGROMETER

WetGauze

Water

T1 T2

Air

- - - - - - - -- - - -

Page 83: 20275003 Physics Measurement 2

WET AND DRY BULB HYGROMETER

T1 = temperature of wet bulb decreases because of evaporation in wet gauze.

Lower humidity causes more evaporation and T1 decreases more.

Humidity T1 – T2

δ - % humidity from tables

Page 84: 20275003 Physics Measurement 2

DEW POINT

• Defined as temperature at which ambient air is fully saturated

• At this point condensation occurs

Page 85: 20275003 Physics Measurement 2

REGNAULT’S HYGROMETER

Thermometer

Silver Tube

Ether

Bubble

Condensation at “ dew point”

AIR

RH = SVP at dew point SVP at ambient temperature

OR from tables

Page 86: 20275003 Physics Measurement 2

HUMIDITY TRANSDUCERS

• Principle : When a substance absorbs water, its resistance or capacitance changes.

• Substance is incorporated into electrical circuit as resistor or dielectric portion of a capacitor.

Page 87: 20275003 Physics Measurement 2

HUMIDITY TRANSDUCERS

Advantages :1. Extremely sensitive2. Rapid response - can be used as

servo-systems.

Disadvantages :1. Display hysteresis – unsuitable for critical applications where high degrees of accuracy required.

Page 88: 20275003 Physics Measurement 2

MASS SPECTROMETER for measuring humidity

- Used to measure water vapour pressure

- Rapid response - can be used to measure breath –by- breath changes.

- Disadvantage - very expensive

Page 89: 20275003 Physics Measurement 2

WEIGHING TECHNIQUES

a) Weighing quantity of water vapour that has condensed in a known volume of air

or

b) Warming air so that all water droplets areevaporated & then weighing volume of air.

c) Absorption techniquesAbsorption of water vapour in either

concentrated sulphuric acid, silica gel or anhydrous CaCl2

Page 90: 20275003 Physics Measurement 2

PRESSURE – Physics and Measurement

Page 91: 20275003 Physics Measurement 2

PRESSURE

• Defined as Force per unit area• Units : Pascal (Pa) or Newton per square meter ( N.m-2).

Newton = Force that will accelerate 1 kg (N) mass at 1ms-2.

1 Pascal = 1 N acting on area of 1m2

• Gravity = 9.81m.s-2

Page 92: 20275003 Physics Measurement 2

PRESSURE UNITS

1 kPa = 10.2 cm H2O1 kPa = 10.2 cm H2O

= mm Hg= mm Hg

[mercury = 13.6 times as dense as water][mercury = 13.6 times as dense as water]

1 bar = 100 kPa1 bar = 100 kPa

= mm Hg= mm Hg

Page 93: 20275003 Physics Measurement 2

GAUGE PRESSURE

• Pressure relative to atmospheric pressure. i.e. zero at atmospheric pressure

• Gauge Pressure may be determined by how much pressure is above or below atmospheric pressure.

Page 94: 20275003 Physics Measurement 2

ABSOLUTE PRESSURE

•Pressure relative to a true zero pressure(i.e.vacuum)

Therefore,

Zero gauge pressure = 1 atmosphere absolute

• Gauge pressure 1 atmosphere = 760 mmHg

= 101.325 kPa

= 2 atmosphere

absolute

Page 95: 20275003 Physics Measurement 2

MEASUREMENT OF BLOODPRESSURE

UNITS OF PRESSURE

Unit Value

Pascal ( Pa) N/m2 SI Unit

mmHg 133.3Pa 1 mmHg = 7.5kP

bar 105Pa

Torr almost = 1 mmHg

cmH2O ~ 1-Pa

Page 96: 20275003 Physics Measurement 2

PRESSURE

P = p x g x h

p = density of fluid

g = acceleration due to gravity

H = Height of column

Conversions 10cmH2O = 7.4 mmHg 10mmHg = 13.6 cmH2O

Mercury 13.6 x water density

Page 97: 20275003 Physics Measurement 2

INDIRECT BLOOD PRESSUREMEASUREMENT

Principlea) Utilise cuff - occlude pulse

a) Detection of return of pulse or blood flow distal to cuff.

Page 98: 20275003 Physics Measurement 2

OCCLUDING CUFF

• Cuff pressure transmitted to tissues surrounding artery.

• Cuff width = 40% limb circumference

• Bladder - at least half circumference - Centred over artery

• Cuff level with heart

Page 99: 20275003 Physics Measurement 2

ERRORS WITH CUFF

a) Too narrow or to loose cuff overestimate

SP and DP.

b) Too wide

underestimate SP and DP

(Pressure = F/A)

Page 100: 20275003 Physics Measurement 2

CUFF – AHA STANDARDS

Small Adult 10 x 24 cm 22 – 26 cm

Adult 13 x 30 cm 27 – 34 cm

Large adult 16 x 38 cm 35 x 44 cm

Adult thigh 20 x 42 cm 45 – 52 cm

Bladder sizeArm

Circumference

Page 101: 20275003 Physics Measurement 2

DEVICES MEASURING CUFFPRESSURE

Mercury manometers - Used to be standard

- Now phased out

- Column must be vertical

- Air vent on top of column

Aneroid gauges - Convenient

- Commonly under-read BP

- Calibrated 6 monthly

Page 102: 20275003 Physics Measurement 2

FLOW DETECTION DISTALTO CUFF

Palpation Finger palpation

Finger Photoplethysmograph

Auscultation Audible range – Korotkov sound

Ultrasound (5MH2) range – Arteriosounde

Subaudible (10-40mH2) range – Infrasound

Oscillometry Defection of oscillations

von Recklinghausen’s oscillations

NO2 invasive BP

Page 103: 20275003 Physics Measurement 2

KOROTKOV SOUNDS

Phase I: First appearance of tapping sounds

Phase II: Brief softening of sounds

Auscultation gap : disappearance of sounds

Phase III: Return of sounds

Phase IV: Muffing of sounds

Phase V: Disappearance of all sounds

Page 104: 20275003 Physics Measurement 2

KOROTKOV SOUNDS

Cuff deflation rate = 2-3 mm Hg per sec.

Rapid deflation = underestimate BP

Note auscultatory gap may be present

DBP = point of disappearance of sound

Difference between phase IV & V ~ 5 mmHg

Page 105: 20275003 Physics Measurement 2

OSCILLOMETRY

• Basis of NIBP

• Only one cuff acts as (i) occluding cuff

(ii) Sensor using microprocessor

• Cuff both actuator & transducer

Page 106: 20275003 Physics Measurement 2

OSCILLOMETRY

Oscillations begins at SBP

Maximal at MAP

Abruptly diminishes at DBP

Page 107: 20275003 Physics Measurement 2

OSCILLOMETRY - NATURE OFOSCILLATIONS

• Diamond shaped pattern

• Pressures at oscillation between 2 heart beats compared.

• Average is recorded

• MAP most accurate

Page 108: 20275003 Physics Measurement 2

OSCILLOMETRY

Advantage : 1. Not operator dependent

2. “hands free”

3. Automated

Disadvantage :

1. Inaccurate in shock or

arrhythmia.

2. Cuff placement important

3. Bruising + skin damage

4. Venous congestion

5. Nerve compression

Page 109: 20275003 Physics Measurement 2

FINAPRES

• Cuff placed around finger

• Changes in volume of arterial blood in

finger detected by plethysmography

• MAP - cuff inflated to maximal

- cuff pressure approximates

arterial pressure waveform

Page 110: 20275003 Physics Measurement 2

ARTERIAL TONOMETER

• Force transducer placed over artery with under lying bone.

• Electrical signal - reproduces arterial waveform

• Needs to be calibrated 5 – 10 min against oscillometric measurements

Page 111: 20275003 Physics Measurement 2

INVASIVE (DIRECT) BP MEASUREMENT

Advantages

1. Continuous monitoring

2. Trends observed

3. Accuracy over wide range

4. Enables visual analysis of pulse

pressure

Page 112: 20275003 Physics Measurement 2

VISUAL ANALYSIS OF WAVEFORM

Myocardial Contractility

Upstroke of pulse pressure ~ LV dp/dtSteep upstroke = strong LV contraction

Stroke Volume

Area under systolic ejection ~ LV stroke volume

Systemic Vascular Resistance

Low diastolic notch = Rapid run off & Steep down stroke low SVR

CIRCULATING BLOOD VOLUME

Exaggerated beat to beat variation with

ventilation = hypovolaemia

Page 113: 20275003 Physics Measurement 2

INDICATIONS FOR INVASIVE BP

1. Rapid changes in BP

2. Monitor effects of potent hypotensive or

vasopressor agents

3. During CP bypass

4. Operation with volume shifts eg. AAA or

phaeochromocytoma.

5. Shock

6. Difficult access eg. Morbid obesity

Page 114: 20275003 Physics Measurement 2

VARIATION OF BP

• Waveform distorted the further away from the heart.

• High frequency components eg. incisura damped out / disappear.

• Systolic BP increases distally.

• Hump becomes more prominent in diastolic part of waveform

Page 115: 20275003 Physics Measurement 2

VARIATION OF BP

• SBP increases towards periphery

• DBP decreases towards periphery

• MAP - only slight drop

eg. MAP radial 5% < MAP aorta

• Pulse Pressure increases towards periphery

Causes : a) Reflection of pressure wave from peripheral arterioles

b) Resonance

Page 116: 20275003 Physics Measurement 2

BP AND POSTURE

- 42 mmHg

0

+80 mmHg

Page 117: 20275003 Physics Measurement 2

COMPONENTS OF INVASIVE BP

• Mechanical coupling of blood to transducerintravascular catheterconnecting tubingstop lock

• TransducerConverts pressure changes to voltage changes

• Electronic processing

• Display + recorder

Page 118: 20275003 Physics Measurement 2

REQUIREMENTS FOR ACCURACYINVASIVE BP

1. STATIC Accuracy - Ability to measure stationary events- No baseline/sensitivity drifts

- Input & output linearity - No hysteresis

2. DYNAMIC Accuracy - Ability to accurately record over rapid changes.

3. PHYSIOLOGICAL Reactance - measuring system must have effect on event recorded / measured.

Page 119: 20275003 Physics Measurement 2

TRANSDUCER

• Coverts pressure energy movement electrical signal

• Commonly diaphragm resistance movement change Wire stretch resistance

• Wheatstone bridge used to convert resistance change to voltage signal

Page 120: 20275003 Physics Measurement 2

STATIC CALIBRATION

• Linear response (straight line) between pressure and output voltage.

OutputGain = slope of line

offset

Pressure

• Offset - Transducer output at zero pressure• Gain - Change in output for given change in pressure

must be constant.• Sensitivity (factory set) at 5V/V/mmHg

Page 121: 20275003 Physics Measurement 2

MEAN BLOOD PRESSURE

a

bMAP

• Average Pressure• Equal to pressure when a = b• Electronically averaged instantaneous measurements• Highly damped system eg. aneroid gauge gives MAP• MAP = diastolic pressure + 1/3 pulse pressure = SBP + 2DBP

3

Page 122: 20275003 Physics Measurement 2

DYNAMIC RESPONSE

• Basic or Fundamental Harmonic (1st) Heart Rate 60 beats/min = 1 Hz Heart Rate 120 beats/min = 2 Hz

• By Fourier Analysis Fundamental = f = 2Hz (HR 120) 2nd Harmonic = 2 x f = 4Hz 3rd Harmonic = 3 x f = 6Hz 10th Harmonic = 10 x f = 20Hz

• Accurate waveform without amplitude distortion achieved with 10th harmonic ~ 20Hz

Page 123: 20275003 Physics Measurement 2

PRACTICAL ASPECTS OF DYNAMIC RESPONSE

• Natural frequencies of clinical systems approx, 30Hz

• Acceptable if damping ratio is optimal

• To achieve dynamic accuracy - fundamental frequency (fo) must be maximised.

Page 124: 20275003 Physics Measurement 2

MAXIMISING Fo

Note : fo = 1 E 2 M

To minimize fo Minimize M (mass)

- Minimal volume of fluid in transducer- Short tubing

Maximise E, modulus of elasticity- Stiff tubing- Stiff diaphragm- Eliminate air bubbles

Page 125: 20275003 Physics Measurement 2

FACTORS THAT INCREASE DAMPING

• Increase fluid viscosity eg. blood clots

• Narrow tubing eg. kinked catheter

• Increased tubing length

• Decreased stiffness of tubing

Page 126: 20275003 Physics Measurement 2

LIQUID MANOMETERS(Absolute Pressure)

Mercury Barometer

••

••

• ••

••

••

••••

•• •

••

• Mercury

Torricellian Vaccum

hP

Measures Absolute Pressure

Page 127: 20275003 Physics Measurement 2

LIQUID MANOMETERS (Gauge Pressure)

P

h - Amount by which pressure exceeds atmospheric

Note tube open at both ends

Page 128: 20275003 Physics Measurement 2

LIQUID MANOMETERSMethods to increase sensitivity

• Use low density liquid

• Amplify vertical movement of

meniscus.

a) inclined plane manometer

b) differential liquid manometer

Page 129: 20275003 Physics Measurement 2

MECHANICAL PRESSURE GAUGE

Bourdon Gauge

Wheel

Pointer

Coiled tube unwinds at high pressure

Pressure

Fixed Point Low

Pressure

HighPressure

Cross Section

• Usually for measuring high pressure• Can be adapted for temperature or flow measurement

Page 130: 20275003 Physics Measurement 2

MECHANICAL PRESSURE GAUGE

Aneroid Gauge

P

BellowsExpands with pressure

Pointer

Lever System - amplifies change

Uses : BP, Airway Pressures on IPPV

Page 131: 20275003 Physics Measurement 2

DIAPHRAGM GAUGE

• Pressure measurement made by sensing movement of flexible diaphragm.

• Diaphragm movement sensed by a) Direct movement of levers etc. (not sensitive) b) Optical method pressure Diaphragm Mirror

stretched more rotated & curved

c) Electromechanical transducers

Page 132: 20275003 Physics Measurement 2

OPTICAL ELECTROMECHANICALTRANSDUCERS

Principles :

1. Increased pressure Diaphragm more convex

2. Light beam reflected off silvered surface of diaphragm on to photoelectric cell.

3. Reflected light beam more divergent.

4. Light intensity sensed by photo-electric cell decreases and electrical output falls

Page 133: 20275003 Physics Measurement 2

OPTICAL ELECTROMECHANICALTRANSDUCERS

P1 P2

Mirror Mirror

Photoelectric

Cell

Photoelectric

CellSlivered surface

Divergent reflected light

Convergent

Reflected light

Sliveredsurface

LightSource

Page 134: 20275003 Physics Measurement 2

STRAIN GAUGE ELECTROMECHANICALTRANSDUCERS

Principles :

Wire stretched or compressedChange in length and diameteratomic stricture change

Resistance change

Page 135: 20275003 Physics Measurement 2

STRAIN GAUGE TRANSDUCER

Wired compressed

Movable block

Fixed pointDiaphragm

P

Resistance wire stretch

1. Resistance wires arranged in 2 sets2. When pressure increases, one set stretches & other set compresses3. Difference in resistance is measured by wheatstone bridge system.

Page 136: 20275003 Physics Measurement 2

BONDED STAIN GAUGE

PStrain gaugeBonded todiaphragm

Single Bond Double Bond

1. Resistance wires in zig-zag patter cemented to diaphragm surfaces.

2. Robust but subject to hysteresis.3. Resistance wire – low temperature coefficient.4. Double bonded strain gauge one stretched & other

compressed.

Page 137: 20275003 Physics Measurement 2

WHEATSTONE BRIDGE ARRANGEMENT OF RESISTANCE WIRES

OUTPUT OUTPUT

StrainGaugeelement

StrainGaugeelement

Half bridge circuit Full bridge circuit

Page 138: 20275003 Physics Measurement 2

CAPACITANCE TRANSDUCER

2ND plateDiaphragmas one plate

Charge

Characteristics : 1. Very sensitive 2. High natural frequency 3. Temperature drift

4. Unstable

Page 139: 20275003 Physics Measurement 2

VARIANCE INDUCTANCE TRANSDUCER

P Iron Core

Diaphragm

Coil magnetic field

Page 140: 20275003 Physics Measurement 2

WHEATSTONE BRIDGE

Wheatstone bridge is a special arrangement of resistors designed to amplify change in resistance.

Balanced Wheatstone bridge

A

R1R2

R adjust R measure

Ammeter readszero

R measure R adjust

= R2

R1

Page 141: 20275003 Physics Measurement 2

FREQUENCY RESPONSE

1. Measurement systems respond to restricted range of frequencies.

2. Input signals of same amplitude at different frequencies will produce output over a limited range of frequencies.

3. Within this frequency range, response may be more sensitive to some frequencies than others.

4. Response of system (system gain) plotted against signal frequency is called “ Frequency Response of System”.

Page 142: 20275003 Physics Measurement 2

FREQUENCY RESPONSE OFA SYSTEM

Bandwidth

Lower cut off Upper cut off

Frequency

Systemgain

Page 143: 20275003 Physics Measurement 2

DETERMINANTS OFFREQUENCY RESPONSE

MECHANICAL SYSTEM

Inertial elements (eg. mass)Compliance elements (eg. spring)

ELECTRICAL CIRCUIT

InductanceCapacitance

Page 144: 20275003 Physics Measurement 2

NATURAL OR RESONANT FREQUENCY

1. When a constant amplitude waveform is applied at increasing amplitude occurs at resonant ornatural frequency (fo) of the system.

2. Beyond fo (higher frequencies), amplitude of oscillations increase and then fall to zero.

3. Fo depends on inertial and compliance etc.

Page 145: 20275003 Physics Measurement 2

AMPLITUDE AS FREQUENCYINCREASES

Natural or resonantFrequency (fo)(maximal oscillation)

Increasing frequency

Amplitude decreasedBeyond fo

Amplitude of oscillation

Page 146: 20275003 Physics Measurement 2

ENERGY INTERCHANGE INOSCILLATING SYSTEM

1. Continental interchange between kinetic energy of mass in motion and potential energy.

2. Kinetic energy = ½ mv2.

3. Narrow Tube

(a) More energy required to make given mass of fluid to oscillate because it has to reach higher velocity.

(b) Catheter fluid velocity > fluid velocity at diaphragm.

(c) E, Effective mass catheter > E, diaphragm.

(d) Larger effective mass lower fo.

Page 147: 20275003 Physics Measurement 2

RESONANT FREQUENCY

OUTPUT

FREQUENCY FO Resonantfrequency

Page 148: 20275003 Physics Measurement 2

UNDAMPED NATURAL FREQUENCY

Fo = 1 2

S M

Fo = undamped natural frequency

S = stiffness of transducer diaphragm

M = Effective mass

Page 149: 20275003 Physics Measurement 2

HIGH UNDAMPED NATURALFREQUENCY

Catheter - Transducer System

- Needs high fo

- Occurs when fluid velocity is minimized

- Achieved by;

(a) Stiff diaphragm

(b) Short and wide catheter.

Page 150: 20275003 Physics Measurement 2

DAMPING

• Defined as tendency of a system to resist oscillations caused by a sudden change.

• In mechanical devices, damping arises from frictional effects on mechanical moving parts.

• In fluid operated devices, damping is caused by vicious forces that oppose fluid movement.

• In electrical devices, electrical resistance oppose passage of electrical currents.

Page 151: 20275003 Physics Measurement 2

EXTENT OF DAMPING

Underdamping - Results in oscillation and over-estimation of measurement (overshoot ofoutput) .

Overdamping - Results in slow response and underestimation of measurement.

Critical damping - No overshoot of output signal but speed of response is too slow.

Page 152: 20275003 Physics Measurement 2

SIGNAL AMPLITUDE & DAMPING

2

0.50.1 0.5 1.0 1.5

1

0.2

0.5

0.64D=1.0

RelativeAmplitude

Page 153: 20275003 Physics Measurement 2

OPTIMAL DAMPING

1. State of damping in which(a) Minimal overshoot(b) Response speed only slightly reduced

2. D = 0.64 (i.e. 64% critical damping)

(a) 7% overshoot

(b) Response speed only minimally reduced

Page 154: 20275003 Physics Measurement 2

PHASE SHIFT RESPONSE

1. Waveform or signal composed of series of component frequencies.

2. Each component waveform undergoes different time delay or phase shift.

3. Phase shift is time delay expressed as an angle(radians).

4. At fo, waveform delayed by 90%

5. Other frequencies, phase lag is linearly related at D = 0.64 I.e. phase distortion minimal at D = 0.6.

Page 155: 20275003 Physics Measurement 2

SPECIFICATIONS OF TRANSDUCERS

1. To avoid waveform, amplitude and phase distortion, catheter-transducer system should have undamped natural frequency 25-40Hz .

2. Standard transducer – undamped natural frequency of 100 Hz or more.

3. Catheter - tap - cannula arrangement reduces natural frequency of the system.

Page 156: 20275003 Physics Measurement 2

DIRECT BP MEASUREMENT SPECIFICATIONS

1. Transducer : - Frequency response > 100 Hz i,.e. resonant frequency > 100Hz.

2. Tubing & cannula : - Lowers fo and adds damping.

Length increase - lower fo, more damping

Compliant tube - lower fo, more damping

Small bore tube - lower fo, more damping

air/clot in tube - lower fo, more damping

Factors that increases fo tend to lower damping.

Page 157: 20275003 Physics Measurement 2

CARDIAC OUTPUT MEASUREMENT

Page 158: 20275003 Physics Measurement 2

USES OF CARDIAC OUTPUT MEASUREMENT

1. GENERAL ICU

- Cardiac performance assessment in shocked patients.

- Management of inotropes and vasoconstrictors

- Optimisation of PEEP Therapy.

2. OPERATING THEATRES

- Major Anaesthetic eg. AAA, Liver transplant

- Anaesthesia in severe cardiac disease (eg. L V failure, Recent MI)

Page 159: 20275003 Physics Measurement 2

CARDIAC OUT MEASUREMENT USES

3. Post-cardiac Surgery Intensive Care Units4. Coronary Care Units / Laboratories

- Assessment of severity of ischaemic & valvular disease

- Management of inotropes vasoconstrictors and vasodilators.

Page 160: 20275003 Physics Measurement 2

INFORMATION FROM C.O. MEASUREMENT

Cardiac output L/min

Cardiac Index = CO Surface area L/min/m2

Systemic vascular resistance = MAP-RAP mmHg/L/min CO or PRU

dyne.sec.cm-5

Pulmonary Vascular Resistance = MPAP-LAP CO

mmHg/L/min or PRU

dyne.sec.cm-5

Page 161: 20275003 Physics Measurement 2

INFORMATION FROM C.O. MEASUREMENT

LV stroke work = MAP x SV gm.m

RVstroke work = MPAP x SV gm.m

Oxygen Consumption = CO x (CaO2-Cv-O2) ml/min

Oxygen Delivery = CO x CaO2 ml/min (D O2)

Page 162: 20275003 Physics Measurement 2

DIRECT CO MEASUREMENT

1. ELECTROMAGNETIC FLOWMETER PROBE

a) Periaorticb) Intraaortic

2. Ultrasonic Flow Probe

3. Intravascular thermal velocity transducer

Page 163: 20275003 Physics Measurement 2

INDIRECT CO MEASUREMENT

A. INVASIVE METHODS1. Fick method (1970) i) Direct (O2 Consumption) ii) Indirect (CO2 production)2. Dye dilution (Stewart, 1894,

Hamilton, 1979)3. Thermodilution (Fegler, 1954)

NON-INVASIVE METHODS1. Radioactive tracer dilution (radiocardiography)2. Bollisto cardiography3. Pneumocardiography4. Impedance Plethysmography

Page 164: 20275003 Physics Measurement 2

ELECTROMAGNETIC FLOW PROBE

Faraday’s Law states that :-

“When a conductor moves with a given velocity across theLines of force of a uniform magnetic field, an electromotiveForce will be induced at right angles to the flow, and will be Proportional to the velocity of the conductor”

Page 165: 20275003 Physics Measurement 2

MAGNETIC FIELD

Blood flow

Magnetic H

V

E

a

Magnetic field is held at right angles to blood flowElectromotive force induced at right angles to moving conductor, the blood flow.

NOTE : EMF technique measures blood velocity

Page 166: 20275003 Physics Measurement 2

ELECTROMOTIVE FORCE, E

+a E = v. H 2a 10 -8

-a

V = Velocity of bloodH = strength of magnetic field (gauss)2a = length of conductor or diameter of blood vessel.

Page 167: 20275003 Physics Measurement 2

BLOOD VELOCITY

Blood velocity = Flow rate (cm3 / sec) Cross section area of vessel (cm2)

Flow rate = Blood velocity x cross section area

Page 168: 20275003 Physics Measurement 2

TYPES OF EMF FLOW PROBES

NOTE : Peri or intra-aortic flow probes measure CO (less coronary blood flow)

(a) Periaortic Flow Probe• used in open heart surgery• rarely used clinically

(b) Intravascular Flow Probe• introduced via peripheral artery• invasive• velocity depends on exact site (maximal in centre of blood vessel)

Page 169: 20275003 Physics Measurement 2

ULTRASONIC FLOW PROBE

Measures velocity of flowing fluid

(a) Pulsed Ultrasound- Cuff transducer around artery- Pulse of 5 MHz from piece-electric crystal

in cuff.- receiver crystal downstream- Transit time between

Page 170: 20275003 Physics Measurement 2

ULTRASONIC FLOW PROBE

(B) Doppler EffectPrinciple : frequency shift of emitted wave frombarium titanate crystal when it is reflected frommoving fluid column

Disadvantage :Cannot detect difference between forward and backward flow.eg. aortic blood flow : mean velocity = 40 cm/sec

systolic velocity = 120 cm/sec

Page 171: 20275003 Physics Measurement 2

ULTRASONIC FLOW PROBE

1. Probe placed at suprasternal notch with beam directed to aortic arch or

2. Transoesophageal probe with beam directed at descending aorta

3. Velocity of blood flow and cross – sectional area of aorta determined.

Page 172: 20275003 Physics Measurement 2

ACCUCOM C.O. MONITOR

1. Continuous C.O reading

2. Oesophageal probe containing dual crystal Doppler probe transducer to measure velocity in descending aorta.

3. Second Doppler probe placed at suprasternal notch used to calibrate instrument

4. Aortic diameter determined by echocardiogram or monogram

Page 173: 20275003 Physics Measurement 2

2 D COLOUR FLOW – ECHO-DOPPLER

1. Pulse ultrasound for imaging of cardiac flow

2. L V outflow tract imaged and cross sectional area determined.

3. Velocity of blood flow in LV – outflow tractmeasured.

4. CO = Cross Section Area x Velocity.

5. Colour Doppler to demonstrate direction of flow

Page 174: 20275003 Physics Measurement 2

INTRAVASCULAR THERMAL VELOCITY TRANSDUCER

1. Heated thermistor placed in moving liquid stream dissipates heat as a function of flow velocity.

2. Probe maintained at fixed position in blood stream within vessel of fixed diameter.

3. Velocity signal translated into flow.

Page 175: 20275003 Physics Measurement 2

INDIRECT METHODS OF C.O. MEASUREMENT

I INVASIVE METHODS Direct Fick Method Indirect Fick Method Dye dilution Method Thermodilution technique

II NON INVASIVE METHODS

Radioactive tracer dilution Ballisto cardiography Thoracic impedance plethysmography

Page 176: 20275003 Physics Measurement 2

FICK PRINCIPLE(Adolph Fick – 1870)

States that : “ the flow of a liquid in a given period ofTime is equal to the amount of substance entering orLeaving the stream / or organ) in the same period of Time, divided by the difference between the concentration of the substance before and after thePoint of entry or exit “

Page 177: 20275003 Physics Measurement 2

DERIVATION OF FICK PRINCIPLE

Substance added ( M mg)

o

Amount ofIndicatorAt entry ()

+Amount of IndicatorAdded (M)

=Amount ofIndicator atExit ( o )

Concentration = Amount VolumeAmount = Volume x concentration

Page 178: 20275003 Physics Measurement 2

DERIVATION OF FICK PRINCIPLE

ConcentrationAt entry ( ) x

Volume at () +

Mass added (M)

= CE VE

Dividing at entry = Vol at exit

Conc. At entry x flow rate + M (amount added/min)

=ConcAtExit

x Flowrate

Therefore Flow Rate

=Amount indicator added per min Exit conc. – entry conc.

Page 179: 20275003 Physics Measurement 2

HYDRAULIC ANALOGUE MODEL

M mgs-1

Q ml sec –1

C mg / m

V ml

Page 180: 20275003 Physics Measurement 2

INDICATOR CONCENTRATION CHANGE AT CONSTANT INFUSION WITH NO INDICATOR INPUT

Conc.(mg ml-1)

Time

C1

C max

Page 181: 20275003 Physics Measurement 2

INDICATOR CONCENTRATION(known concentration Co at input with constant infusion)

Conc

C max

Co

Time

Page 182: 20275003 Physics Measurement 2

INDICATOR CONCENTRATION vs TIME (Bolus Injection)

yo

y1

y1 = yoe-ke Conc

Time

Page 183: 20275003 Physics Measurement 2

FICK METHOD

Used in 3 ways

1. Direct Fick - uses oxygen uptake as indicator Co = ___Vo2____

CaO2 - CVO2

2. Indirect Fick – uses CO2 production as indicator

Co = VCO2

CVCO2 – CaCO2

3. INERT GAS METHODeg. N2O xe137, K85, K7a

- Used for specific organ blood flow measurement- basis of Kety – Schmidt Method

Page 184: 20275003 Physics Measurement 2

DIRECT FICK METHOD

Assumptions ;

1. Steady State of both flow (Q) and oxygen consumption.

2. CaO2 and CVO2 constant.

3. Closed systemI.e. blood is only source of substance taken up.

Page 185: 20275003 Physics Measurement 2

DIRECT FICK METHOD

Measurement of O2 consumption

1. Breathe O2 (FIO2) via one-way valve

2. Expires into Douglas bag or Tissot Spirometer

E = Volume measured Time

.V

Page 186: 20275003 Physics Measurement 2

DIRECT FICK METHOD

Calculation of Oxygen consumption

VO2 = InspiredGas Vol

x FiO2 - ExpiredGas Vol

x FEO2

Page 187: 20275003 Physics Measurement 2

DIRECT FICK METHODOxygen consumption calculations

Nitrogen is in steady state

V inspired N2 = V expired N2

VI x FIN2 = VE x FEN2

VI = VE x FEN2

FIN2

FIN2 = I - FIO2

FEN2 = 1 – FEO2 – FECO2

VI = VE x 1 – FEO2 – FECO2

1 – FIN2

VO2 = (VI x FIO2) – VE x FEO2

Page 188: 20275003 Physics Measurement 2

DIRECT FICK METHOD

CaO2 = Hb (g/L) x SaO2 x 1.34 ( mIO2) + 0.003 x PaO2

SaO2 measured using arterial blood and calibrated oximeter

Page 189: 20275003 Physics Measurement 2

DIRECT FICK METHOD

( Mixed Venous Oxygen Content)

Need pulmonary artery blood for mixed venousSample.

CVO2 = Hb x SVO2 x 1.34 dissolved component usually ignored

Page 190: 20275003 Physics Measurement 2

DIRECT FICK METHOD

CO = 250 ml / min_____ 200ml/L - 150 ml/L

= 5 L / Min

Features : 1. Steady state of CO, VO2, CO2 production N2 balance arterial; and venous O2 concentration.

2. Accuracy + 10% used as reference.

3. Slow, cumbersome, unsuitable for rapid measurements.

4. Unsuitable during GA - Not a steady state - Uptake of volatile agents and N2 washout.

Page 191: 20275003 Physics Measurement 2

INDIRECT FICK METHOD

1. CO2 used as indicator

2. Theoretical advantage :Mixed venous CO2 estimated by rebreathingtechnique - No need for CVP.

3. Problem :Large CO2 stores - steady state not easily

achieved.

Page 192: 20275003 Physics Measurement 2

DYE DILUTION METHOD

1. Applies to bolus of indicator

2. Also known as Stewart – Hamilton principle.

3. Basis of dye dilution and thermodilution techniques.

Page 193: 20275003 Physics Measurement 2

DYE DILUTION METHOD

ExitConc.

Ct

0 T 00time

Cmax

C (t)

Ct = C maxe-kt

Where e = 2.718K = decay constant of exponential

Page 194: 20275003 Physics Measurement 2

DYE DILUTION METHOD: CALCULATIONS

At any moment c = M V M = C x V

Integrating :

Mdt = Cdt x Vdt

o o oWhere

oMdt = Original injected

o

= total volume flow

Page 195: 20275003 Physics Measurement 2

STEWART – HAMILTON FORMULA

M =

o

cdt x Q

Q = ___M___

ocdt

Page 196: 20275003 Physics Measurement 2

DYE – DILUTION INFUSION TECHNIQUE

C exit

C entry

t o

t1

Q = __M__

t = 1

t - 0

Cdt

= __M__Ct1 - Ct0

= _______M________C exit - C entry

Page 197: 20275003 Physics Measurement 2

DYE – DILUTION METHOD

Dye - Indocyanine green (ICG)

- peak absorption = 805 nm(isobestic point of HbO2)

- non toxic

- rapid removal 18-24% per min

- 2.5-5mg injected into right atrium

Page 198: 20275003 Physics Measurement 2

DYE DILUTION – GRAPH ANALYSIS

recirculation

t

C

o

Cdt = area under curve

Methods used (a) Trapezoid Method (b) Forward method

Triangle area - Peak Height x width at halve peak height (c) Computer integrator

Page 199: 20275003 Physics Measurement 2

DYE DILUTION METHOD

EQUIPMENT

1. Withdrawal Pump/syringe ~ 20ml/min

2. Optical densitometer ~ 805 nm

3. Chart Recorder

Page 200: 20275003 Physics Measurement 2

THERMODILUTION TECHNIQUE

Bolus of “Negative Heat”

Sensor = Thermistor in pulmonary artery

Cardiac Output = ___Amount of –ve Heat (M)__

Tdto

Page 201: 20275003 Physics Measurement 2

THERMODILUTION METHOD

Negative Heat M = V1 x (TB – T1) D S

DBSB

Where M = Amount of negative heat VI = Volume injectable ( 10ml 5% D) TB = Patient’s blood temperature TI = Injectable temperature DI = Injectable density DB = Blood density SI = Specific heat of injectate

SB = Specific heat of blood

For 5% Dextrose = __DS SI__ = 1.08 DB - SB

Page 202: 20275003 Physics Measurement 2

THERMODILUTION ADJUSTMENTS

(a) Correction factor

(b) AVC = Calculated to 1.22

t

0

Tdt

t where C = _30 Cmax 100

(c) Final Formula

CO = VI x (TB – TI) x 1.08 x CT x 60

1.22

o

Tdt

Page 203: 20275003 Physics Measurement 2

THERMODILUTION

1. Room temperature D 5% used

2. Inject within 1.5 sec.

3. Inject at end-expiration

Page 204: 20275003 Physics Measurement 2

ICED vs ROOM TEMPERATURE INJECTATE

1. Random error greater with injectateat room temperature

2. Iced Injectate slightly overestimates atlow cardiac output.

3. Room temperature significantly overestimatesat CO < 2-3L/min by 20-50%.

Page 205: 20275003 Physics Measurement 2

FICK METHOD

ADVANTAGES

Correlates with direct measurements referenceMethod.

DISADVANTAGES

Slow and cumbersomeNot suitable for rapid repeated measurements.

Page 206: 20275003 Physics Measurement 2

DYE DILUTION

ADVANTAGES 1. Correlates well with direct and Fick method

DISADVANTAGES 1. Arterial cannulation needed 2. Limited to 3 measurements 3. Recirculation “Noise” 4. Unsuitable for rapid, repeated measurements

Page 207: 20275003 Physics Measurement 2

THERMODILUTION METHODS

ADVANTAGES

1. Simple and convenient

2. No blood withdrawal

3. Limited recirculation

4. Unlimited number of measurements

5. Rapid repeated measurements possible

Page 208: 20275003 Physics Measurement 2

RADIOACTIVE TRACER DILUTION

Risa washout over heart

Scintillation Counter

Difficult to calibrate

Radiation hazard.

Page 209: 20275003 Physics Measurement 2

BALLISTOCARDIOGRAPHY

Patient coupled to light bed

Ultralow frequency recording

Body acceleration (recoil) measures aortic

Acceleration ( dQ/dt) and stroke volume

F

H

L

G

I

Lo

t

I and J waves = dQ/dt

Page 210: 20275003 Physics Measurement 2

IMPEDANCE PLETHYSMOGRAPHY

12

3

4

-100Yz sinusoidal current 4 mA through chest

-Wheatstone bridge to measure resistance

-Voltage change with constant current I

-R = Voltage change I

-Resistance change reflect pulmonary blood flow

Page 211: 20275003 Physics Measurement 2

LITHIUM DILUTION CARDIAC OUTPUT

- Indicator = Lithium chloride (150 mM)

- Dose ~ 0.3 mmol via any venous line

- Artyerial litium plasma concentration measured by lithium sensitive electrode aspirated at 4 ml/min.

- Co = LiCl dose x 6- Area (1 – haemocrit)

Page 212: 20275003 Physics Measurement 2

PULSE PRESSURE ANALYSIS

- Arterial pulse pressure waveform analysed

- Cardiac output ~ Area under systolic portion of arterial waveform from diastole to end-systole

- Calibrated initial using lithium dilution technique.

Page 213: 20275003 Physics Measurement 2

Measurement of pH

Page 214: 20275003 Physics Measurement 2

Measurement of pH

pH = - logpH = - log1010 of the hydrogen ion activity (~ []'n) of the hydrogen ion activity (~ []'n) at 37°C, normal blood at 37°C, normal blood pH = 7.4 ± 0.04pH = 7.4 ± 0.04 circuit consists of,circuit consists of,

capillary tube of pH sensitive glass ® dVcapillary tube of pH sensitive glass ® dV reference buffer solution the other side of the glassreference buffer solution the other side of the glass

+ a silver/silver chloride electrode + a silver/silver chloride electrode an electrolyte solution (KCl) in contact with bloodan electrolyte solution (KCl) in contact with blood

+ a silver/silver chloride electrode+ a silver/silver chloride electrode surrounding water jacket at 37°Csurrounding water jacket at 37°C voltmetervoltmeter

Page 215: 20275003 Physics Measurement 2

Measurement of pH

Page 216: 20275003 Physics Measurement 2

MEASUREMENT OF pHpH electrode

- Depends on ion selective electrode

- pH sensitive Glass Electrode

- Utilises glass membrane which is

selectively permeable to hydrogen ions.

- Glass electrode - placed in series with 2 half cells which generate a constant potential gradient

Page 217: 20275003 Physics Measurement 2

pH ELECTRODE SYSTEMS

• Electrode consists of:

metal – conducts - electrons

electrolyte – conducts ions.

• Ag:AgCl + Hydrochloric Acid

Hg:Hg2Cl2 + saturated KCl Solution

• EMF generated at interface of 2 electrodes.

Page 218: 20275003 Physics Measurement 2

SCHEMATIC ARRANGEMENT OF pH ELECTRODE

V

HCl

Ag/AgClReferenceElectrodeSAMPLE

KCl Saltbridge

Hg / Hg2Cl2

Calomel Reference Electrode

Porous Plug

pH sensitiveglass

Potential Constant Constant Variable Constant

Voltmeter

Page 219: 20275003 Physics Measurement 2

pH ELECTRODE

Saturated KCl

• Provides salt bridge

• Completes circuit between blood sample and calomel electrode.

• Porous plug prevents diffusion of KCl into blood sample.

Page 220: 20275003 Physics Measurement 2

pH ELECTRODE

• Measures activity of H+; not concentration

• Calibrated against 2 standard buffers;

(a) pH 6.841 = Zero (b) pH 7.383

Page 221: 20275003 Physics Measurement 2

pH ELECTRODE

HCl

Ag : AgClelectrode

SENDING CIRCUITAND DISPLAY

SAMPLECUVETTE

Platinum Wire

Mercurous chloride

Mercury

Saturated KCl

Porous Plug

pH sensitive glass

Page 222: 20275003 Physics Measurement 2

Measurement of Gases

Page 223: 20275003 Physics Measurement 2

GAS ANALYSIS

CHEMICAL METHODSAbsorption in chemicals using Haldane apparatusCO2 : 10 – 20% KOH or NaOHO2 : Alkaline pyrogallol or sodium anthraquinone

PHYSICAL METHODS :• Mass spectrometers• Infra-red absorption• Polarography• Galvanic fuel cell• Ultra violet absorption• Paramagnetism• Thermal conductivity

Page 224: 20275003 Physics Measurement 2

Spectrophotometry

first used to determine the [Hb] the 1930's, by first used to determine the [Hb] the 1930's, by application of the application of the Lambert-Beer LawLambert-Beer Law

IIii = the incident light= the incident light IItt = the transmitted light= the transmitted light DD = the distance through the medium= the distance through the medium CC = the concentration of the solute= the concentration of the solute = the = the extinction coefficientextinction coefficient of the of the

solutesolute

ITrans = ISource x e- DC

Page 225: 20275003 Physics Measurement 2

Spectrophotometry

the the extinction coefficientextinction coefficient is specific for is specific for a given solute at a given wavelength of a given solute at a given wavelength of lightlight

therefore, for each wavelength of light therefore, for each wavelength of light used an independent Lambert-Beer used an independent Lambert-Beer equation can be writtenequation can be written

if the number of equations = the number if the number of equations = the number of solutes, then the concentration for of solutes, then the concentration for each one can be solvedeach one can be solved

Page 226: 20275003 Physics Measurement 2

Spectrophotometry

by convention oxyhaemoglobin by convention oxyhaemoglobin concentration, HbOconcentration, HbO22 is the fractional is the fractional

concentration as measured by concentration as measured by cooximetrycooximetry

a 4 wavelength device, and includes a 4 wavelength device, and includes COHbCOHb and and MetHbMetHb in the denominator in the denominator

%HbO2 = 100 [ HbO2 ]

Hb + HbO2 + COHb + Met Hb

Page 227: 20275003 Physics Measurement 2

ULTRA-VIOLET ABSORPTION

• Halogenated vapours absorb uv light

• used for measuring halothane

• Disadvantage : Slow response time produce toxic product

Page 228: 20275003 Physics Measurement 2

THERMAL CONDUCTIVITY (KATHAROMETERS)

1. High thermal conductivity gas - more rapid heat conductioneg. Helium 600%CO2 35% compared with air

2. Gas passed over heated wire which cools.

3. Decreased wire temperature – depends on flow rate and thermal conductivity of gas.

4. Temperature leads to wire resistance

5. Advantages : Simple and inexpensive

6. Disadvantage : Slow response time ( ~ 5s)

Page 229: 20275003 Physics Measurement 2

Measurement: Methods

Mass spectrometryMass spectrometry

Raman spectrographyRaman spectrography

Photo-acoustic spectrographyPhoto-acoustic spectrography

Infra-red spectrographyInfra-red spectrography

Page 230: 20275003 Physics Measurement 2

RAMAN LIGHT SCATTERING

1. Photon of light passes thro’ gas

2. Photon energy partly given to gas molecule

3. Light is re-emitted at longer wavelengthcharacteristic to gas.

Page 231: 20275003 Physics Measurement 2

Measurement: Raman Spectrography

Raman scatteringRaman scattering occurs with illumination with high intensity occurs with illumination with high intensity argon laserargon laser light light

absorbed light energy produces unstable energy states absorbed light energy produces unstable energy states (rotational & vibration)(rotational & vibration)

emitted low energy light, Raman lightemitted low energy light, Raman light measured at 90° to the laser pathmeasured at 90° to the laser path

can be used to identify all types of molecules in the gas can be used to identify all types of molecules in the gas sample, and has been incorporated into new monitors sample, and has been incorporated into new monitors (RASCAL) which instantaneously identify & quantify CO(RASCAL) which instantaneously identify & quantify CO22 and and inhalational agentsinhalational agents

Page 232: 20275003 Physics Measurement 2

Measurement: Photo-acoustic spect.

relies on the absorbance of IR light by COrelies on the absorbance of IR light by CO22 gas expansiongas expansion

IR light is pulsed at IR light is pulsed at acoustic frequenciesacoustic frequencies and the and the energy absorbed is detected by a microphoneenergy absorbed is detected by a microphone

amount of light absorbed is measured amount of light absorbed is measured directlydirectly without the need for a reference chamberwithout the need for a reference chamber no zero point driftno zero point drift

other claimed advantages over IR spectrometry,other claimed advantages over IR spectrometry, higher accuracyhigher accuracy increased reliabilityincreased reliability reduced maintenance & reduced need for reduced maintenance & reduced need for calibrationcalibration

Page 233: 20275003 Physics Measurement 2

MASS SPECTROMETER

PRINCIPLE :

1. Gas passed into ionizing chamber

2. Electron beam ionizes gas

3. Ions diffuse thro’ slit in chambers

4. Negatively charge plate accelerate ions

5. Different particles streams separate according to mass & charge.

6. Detector plate

Page 234: 20275003 Physics Measurement 2

MASS SPECTROMETER

Detector

Low charge / mass ratio

Deflection Angle

High charge / mass ratio

GAS

AcceleratorPotentialOn screen electrode

Magnetic field

Page 235: 20275003 Physics Measurement 2

MASS SPECTROMETER

ADVANTAGES

1. Rapid response time ( < 0.1s)2. Can measure variety of gases (May be affected by water vapour)

DISADVANTAGES

1. Complex2. Expensive

Page 236: 20275003 Physics Measurement 2

Capnometry

capnometrycapnometry is the measurement and display of is the measurement and display of COCO22 concentrations on a digital or analogue concentrations on a digital or analogue displaydisplay

capnographycapnography is the graphic recording of is the graphic recording of instantaneous respired COinstantaneous respired CO22 concentrations concentrations during the respiratory cycleduring the respiratory cycle

Page 237: 20275003 Physics Measurement 2

Capnometry

first IR COfirst IR CO22 measuring and recording apparatus measuring and recording apparatus was introduced by was introduced by LuftLuft in 1943 in 1943

expensive, bulky and principally only used for expensive, bulky and principally only used for researchresearch

widespread use within the last 10-15 years with widespread use within the last 10-15 years with cost and size reductioncost and size reduction

ASA closed claims ASA closed claims 93%93% of anaesthetic of anaesthetic mishaps preventable by mishaps preventable by ETCOETCO22 / SpO / SpO22

Page 238: 20275003 Physics Measurement 2

INFRA-RED ABSORPTION

PRINCIPLE :

1. Molecule composed of 2 or more dissimilar atoms absorb infra red light.

2. Absorption of 2.5 - 25 m cause covalent bonds to bend and vibrate; increasing rotational speed.

3. Different gas molecules absorb specific of infra red light.

4. Detecting increased absorption allows their concentrations to be determined

Page 239: 20275003 Physics Measurement 2

Measurement: IR Spectroscopy

Lambert-Beer lawLambert-Beer law applies, (cf. Hb) applies, (cf. Hb) more compact and less expensivemore compact and less expensive assymetric, assymetric, polyatomicpolyatomic gases gases of two or more of two or more

molecules, absorb IR radiation (> 1.0 µm)molecules, absorb IR radiation (> 1.0 µm) HH22O, NO, N22O, COO, CO22

absorbance peak is characteristic for a gasabsorbance peak is characteristic for a gas COCO22 ~ 4.28 µm ~ 4.28 µm

Page 240: 20275003 Physics Measurement 2

Measurement: IR Spectroscopy

glass absorbs IR radiationglass absorbs IR radiation chamber windows must be made of a crystalchamber windows must be made of a crystal

sodium chloride or sodium bromidesodium chloride or sodium bromide

calibration may be achieved by filling the chamber calibration may be achieved by filling the chamber with a COwith a CO22 free gas, or by splitting the incident beam free gas, or by splitting the incident beam

and passing this through a and passing this through a reference chamberreference chamber

Page 241: 20275003 Physics Measurement 2

Measurement: IR Spectroscopy

the use of a reference beam also allows for the use of a reference beam also allows for compensation for variations in the output of the compensation for variations in the output of the IR sourceIR source

the sample chamber is made small, so that the sample chamber is made small, so that continuous analysis is possible continuous analysis is possible

the the response timeresponse time ~ 100 ms~ 100 ms enabling end-tidal COenabling end-tidal CO22 estimations and real-time estimations and real-time

graphical analysisgraphical analysis

Page 242: 20275003 Physics Measurement 2

INFRA-RED GAS ANALYSER(SPECTROPHOTOMETER)

1. LED split infra-red into different .

2. Sample chamber is transilluminated and IR absorption measured.

3. Reference chamber transilluminated & absorption allowscalibration.

4. IR absorption in sample chamber compared with reference chamber.

Page 243: 20275003 Physics Measurement 2

INFRA-RED SPECTROPHOTOMETER

REFERENCE

Known CO2

SAMPLE CELL

Light splitterChopper

Detector

ADVANTAGES : Fast response for CO2 N2O and volatile anaesthetic agentDISADVANTAGES : Rapid respiratory rates

decrease accuracy

Page 244: 20275003 Physics Measurement 2

ETCO2 : Classification 1

side-streamside-stream sensor is located within the main unit and gas is aspirated sensor is located within the main unit and gas is aspirated

from the circuitfrom the circuit sampling flow rate may besampling flow rate may be

highhigh > 400 ml/min, or> 400 ml/min, or lowlow < 400 ml/min< 400 ml/min

optimal gas flowoptimal gas flow is considered to be 50-200 ml/min, is considered to be 50-200 ml/min, ensuring reliability with both adults and childrenensuring reliability with both adults and children

exhaust gases contain anaesthetic agentsexhaust gases contain anaesthetic agents & should be & should be routed to the routed to the scavengingscavenging unit unit

Page 245: 20275003 Physics Measurement 2

ETCO2 : Classification 2

mainstreammainstream sensor is located at the patient, with a curvette placed sensor is located at the patient, with a curvette placed

within the circuitwithin the circuit

these are heated to > 39° to prevent occlusion by water these are heated to > 39° to prevent occlusion by water vapourvapour

no mixing of gases occurs during sampling and the no mixing of gases occurs during sampling and the response time is more rapidresponse time is more rapid

curvettes tend to be bulky, add dead space, are heated, curvettes tend to be bulky, add dead space, are heated, and are expensive if dropped & brokenand are expensive if dropped & broken

Page 246: 20275003 Physics Measurement 2

ETCO2 : Sources of Error

Atmospheric pressure differencesAtmospheric pressure differences NN22OO HH22OO OthersOthers

OO22

alinearityalinearity volatile agentsvolatile agents

Page 247: 20275003 Physics Measurement 2

ETCO2 : PAtm

direct effectsdirect effects gas densitygas density

for a given chamber thickness, no. of molecules for a given chamber thickness, no. of molecules increasesincreases

eliminated by calibration against a known Peliminated by calibration against a known PCO2CO2 (% (%

x Atm.)x Atm.) units calibrated against Cunits calibrated against CCO2CO2 require correction require correction

(1%:1%)(1%:1%) IR absorbanceIR absorbance

intermolecular forces ® IR absorbance for a given [COintermolecular forces ® IR absorbance for a given [CO22]]

PPAtmAtm ~ 1% ~ 1% absorbance ~ 0.5-0.8%absorbance ~ 0.5-0.8%

Page 248: 20275003 Physics Measurement 2

ETCO2 : PAtm

direct effects (continued)direct effects (continued) sampling sampling flow rateflow rate may reduce sample chamber may reduce sample chamber

pressurepressure units should be calibrated for a given sample rateunits should be calibrated for a given sample rate

PEEPPEEP maymay PPCO2CO2 reading (some unit compensate reading (some unit compensate automatically)automatically)

PEEP ~ 20 cmHPEEP ~ 20 cmH22O O PPCO2CO2 ~ 1.5 mmHg ~ 1.5 mmHg

Page 249: 20275003 Physics Measurement 2

ETCO2 : PAtm

indirect effectindirect effect : volume percent : volume percent, ,

where Pwhere PCO2CO2 = F = FCO2CO2 x Atm. x Atm.

where Pwhere PAtmAtm at calibration is different to the time of measurement at calibration is different to the time of measurement

Page 250: 20275003 Physics Measurement 2

ETCO2 : N2O

absorbs IR at absorbs IR at 4.5 µm4.5 µm (cf. CO(cf. CO22 ~ 4.28 µm) ~ 4.28 µm)

NN22OO falsely elevatedfalsely elevated CO CO22 readings readings effect minimised by a narrow bandwidth filtereffect minimised by a narrow bandwidth filter

however, presence of Nhowever, presence of N22O molecules results in O molecules results in collision broadeningcollision broadening of the absorbance peak of of the absorbance peak of COCO22

resulting in apparently resulting in apparently elevated COelevated CO22 readings readings

Page 251: 20275003 Physics Measurement 2

ETCO2 : N2O

simplest correction is to calibrate the monitor with simplest correction is to calibrate the monitor with the same background gas as is to be used during the same background gas as is to be used during anaesthesiaanaesthesia

alternatively correction factors may be applied,alternatively correction factors may be applied, 50% N50% N22OO P'P'CO2CO2 ~ P ~ PCO2CO2 x 0.9 x 0.9 70% N70% N22OO P'P'CO2CO2 ~ P ~ PCO2CO2 x 0.94 x 0.94

Page 252: 20275003 Physics Measurement 2

ETCO2 : H2O

condensed watercondensed water

result in falsely result in falsely highhigh readings readings

prevented in mainstream units by heating the sensorprevented in mainstream units by heating the sensor

side-stream units use water trapsside-stream units use water traps

some units use semipermeable Nafionsome units use semipermeable Nafion®® tubing tubing

Page 253: 20275003 Physics Measurement 2

ETCO2 : H2O

water vapourwater vapour mainstream analysers measure breathing circuit gasmainstream analysers measure breathing circuit gas

generally saturated at body T. but may be affected by the generally saturated at body T. but may be affected by the

use of humidifiers, FGF's, and the ambient T.use of humidifiers, FGF's, and the ambient T.

side-stream units, cooling of the gases results inside-stream units, cooling of the gases results in water vapour pressure, andwater vapour pressure, and

apparent apparent increaseincrease in P in PCO2CO2 ~ ~ 1.5-2%1.5-2%

Page 254: 20275003 Physics Measurement 2

ETCO2

transit timetransit time creating a creating a phase shiftphase shift, but no distortion, but no distortion gas is subject to gas is subject to mixingmixing with overdamping of a with overdamping of a

square waveformsquare waveform results in underestimation of ETCOresults in underestimation of ETCO22, especially in , especially in

childrenchildren this error increases both with,this error increases both with,

increased width and length of the sample tubingincreased width and length of the sample tubing reduced sample flow rates < 50 ml/minreduced sample flow rates < 50 ml/min higher frequency breathing patternshigher frequency breathing patterns

Page 255: 20275003 Physics Measurement 2

ETCO2

rise timerise time TT10-9010-90

time to change from 10% to 90% of the final valuetime to change from 10% to 90% of the final value

depends on size of the depends on size of the sample chambersample chamber and and flow rateflow rate

capnographs used clinically ~ 50-600 mseccapnographs used clinically ~ 50-600 msec

prolongation may decrease the slope of phase II, and prolongation may decrease the slope of phase II, and

underestimation of anatomical dead spaceunderestimation of anatomical dead space

ETCOETCO22 in adults at < 30 bpm with in adults at < 30 bpm with ± 5% ± 5% accuracyaccuracy

faster units are required in children, Tfaster units are required in children, T7070 < 80 msec < 80 msec

Page 256: 20275003 Physics Measurement 2

ETCO2

rise timerise time TT10-90 10-90 (continued)(continued)

response times have been markedly response times have been markedly reduced by,reduced by,

more powerful signal amplifiersmore powerful signal amplifiers minimising the volume of the sample minimising the volume of the sample

chamberchamber use of relatively high sample flow rates > use of relatively high sample flow rates >

150 ml/min150 ml/min

Page 257: 20275003 Physics Measurement 2

ETCO2 : Other Factors

oxygenoxygen OO22 does not directly absorb IR light does not directly absorb IR light

may affect reading by collision broadeningmay affect reading by collision broadening

results in falsely results in falsely lowlow P PCO2CO2 readings readings

not as great as with Nnot as great as with N22O (some units incorporate O (some units incorporate correction)correction)

Page 258: 20275003 Physics Measurement 2

ETCO2 : Other Factors

halogenated agentshalogenated agents absorb IR light at absorb IR light at ~ 3.3 µm~ 3.3 µm

interference is not clinically significantinterference is not clinically significant

alinearityalinearity of CO of CO22 analysis analysis

the concentration of the calibration gas should be as close the concentration of the calibration gas should be as close as possible to the measured gas sampleas possible to the measured gas sample

Page 259: 20275003 Physics Measurement 2

Severinghaus CO2 Electrode

Severinghaus developed the COSeveringhaus developed the CO22 electrode in electrode in

19581958

modern arterial blood gas analysis was bornmodern arterial blood gas analysis was born Essentially a modified pH electrodeEssentially a modified pH electrode

provides a direct measure of Pprovides a direct measure of PCO2CO2 from the from the change in pHchange in pH

Page 260: 20275003 Physics Measurement 2

Severinghaus CO2 Electrode

circuit consists of,circuit consists of, a closed cylinder of a closed cylinder of pH sensitive glasspH sensitive glass in the centre in the centre 2 electrodes, 1 inside, the other outside the cylinder2 electrodes, 1 inside, the other outside the cylinder a surrounding solution of a surrounding solution of sodium bicarbonatesodium bicarbonate

a thin film of bicarbonate impregnated nylon mesh a thin film of bicarbonate impregnated nylon mesh covering the end of the cylinder covering the end of the cylinder

a thin, a thin, COCO22 permeable membrane permeable membrane covering the end covering the end

of the electrodeof the electrode

Page 261: 20275003 Physics Measurement 2

Severinghaus CO2 Electrode

Page 262: 20275003 Physics Measurement 2

Severinghaus CO2 Electrode

COCO22 diffuses from the blood sample through the diffuses from the blood sample through the membrane into the nylon mesh and by the membrane into the nylon mesh and by the formation of formation of carbonic acidcarbonic acid lowers the pH of the lowers the pH of the bicarbonate solutionbicarbonate solution

the change in pH alters the dV across the glass, the change in pH alters the dV across the glass, such that,such that,

pH ~ pH ~ loglog1010PPCO2CO2

Page 263: 20275003 Physics Measurement 2

CO2 Electrode

output of output of voltmetervoltmeter calibrated in terms of P calibrated in terms of PCO2CO2

electrode accuracy ~ 1 mmHgelectrode accuracy ~ 1 mmHg

response time ~ 2-3 minsresponse time ~ 2-3 mins

as for the pH electrode, the COas for the pH electrode, the CO22 electrode kept at electrode kept at 37°C and regularly calibrated with known 37°C and regularly calibrated with known concentrations of COconcentrations of CO22

Page 264: 20275003 Physics Measurement 2

Measurement of OXYGEN

Page 265: 20275003 Physics Measurement 2

OXYGEN MEASUREMENT ELECTROCHEMICAL METHODS

• Based on electrochemical reaction in buffer

solution occurring between 2 electrodes,

involving gas molecules.

• 2 Devices

(a) Polarographic electrode

(b) Fuel cell.

Page 266: 20275003 Physics Measurement 2

Measurement of Oxygen

Leyland Clarke developed the Leyland Clarke developed the

polarographic oxygen electrode in 1956polarographic oxygen electrode in 1956

prior to this the POprior to this the PO22 had not been measured had not been measured

Page 267: 20275003 Physics Measurement 2

Other Methods

POPO22 may also be measured by, may also be measured by, Volumetric - van Slyke/NeillVolumetric - van Slyke/Neill Clarke electrodeClarke electrode Fuel cellFuel cell ParamagneticParamagnetic Hummel Cell - paramagneticHummel Cell - paramagnetic Optode - photoluminescence quenchingOptode - photoluminescence quenching Raman scatteringRaman scattering Mass spectrometerMass spectrometer

Page 268: 20275003 Physics Measurement 2

PARAMAGNETISM

• Paramagnetic : attracted toward magnetic field eg. oxygen

• Diamagnetic : repelled by magnetic field eg. nitrogen

• Paramagnetic molecules = 2 unpaired electrons in outer electron shell spinning in the same direction.

Page 269: 20275003 Physics Measurement 2

PAULING TYPE OF PARAMAGNETICOXYGEN ANALYSER

MAGNET POLE

MAGNET POLE

MAGNET POLE

Gas O2

Nitrogen In GlassDumb-Bell

Light beam Detector

Slow response ( 5 – 20 s)

Page 270: 20275003 Physics Measurement 2

RAPID PARAMAGNETIC O2 ANALYSERS

Sample Reference Gas

Magnetic field

Gas Mixture out

Made more compactRapid response time

DifferentialPressure transducer

Page 271: 20275003 Physics Measurement 2

POLAROGRAPHIC ELECTRODE

PRINCIPLE

• 1 pair of electrodes in electrolyte solution

• Electrodes maintained at potential difference

• Current through electrolyte solution dependent on gas concentration in solution

• Reaction driven by voltage applied to electrodes

Page 272: 20275003 Physics Measurement 2

CLARKE OXYGEN ELECTRODE

• Cathode - Platinum covered by permeable membrane

• Anode - Silver/Silver chloride covered by membrane

• Electrolyte Solution - KCl

• Electrodes connected to DC voltage 0.6V

• Electrons produced by Ag / AgCl anode migrate to cathode to reduce O2 molecules.

Page 273: 20275003 Physics Measurement 2

Clarke Electrode

the circuit consists of,the circuit consists of, DC voltage source (0.6 V)DC voltage source (0.6 V)

ammeterammeter

platinum cathodeplatinum cathode

silver/silver chloride anodesilver/silver chloride anode

electrolyte solution (KCl), andelectrolyte solution (KCl), and

OO22-permeable membrane-permeable membrane

Page 274: 20275003 Physics Measurement 2

Clarke Electrode

Page 275: 20275003 Physics Measurement 2

Clarke Electrode

Ohm’s Law: for any resistive Ohm’s Law: for any resistive circuit:circuit: I I V V

for the Clarkefor the Clarkeelectrode there is aelectrode there is aplateau voltage rangeplateau voltage range I does not change withI does not change with VV however:however: II POPO22

this occurs as the cathode this occurs as the cathode reaction requires both Oreaction requires both O22

and free electronsand free electrons

Page 276: 20275003 Physics Measurement 2

Clarke Oxygen Electrodes (Cont’d)

• Platinum Cathode - O2 + 4e 2 O

(reduction) 2 O + 2 H2O 4 OH

reaction at the platinum cathode,reaction at the platinum cathode,OO22 + 2H + 2H22O + 4eO + 4e-- 4OH4OH--

At Ag / AgCl Anode (oxidation) :

4 Ag 4 Ag+ + 4e-

Current flow between both electrodes measured

Page 277: 20275003 Physics Measurement 2

Clarke Electrode

current flow being in direct proportion to the current flow being in direct proportion to the consumption of oxygenconsumption of oxygen

the platinum electrode cannot be inserted directly the platinum electrode cannot be inserted directly into the blood stream as protein deposits form an into the blood stream as protein deposits form an affect its accuracyaffect its accuracy

Page 278: 20275003 Physics Measurement 2

CLARKE ELECTRODES

Advantages : Robust

Portable

Disadvantages : Limited life span

Silver anode eventually used

up by current

Page 279: 20275003 Physics Measurement 2

FUEL CELL

Cathode : Silver - reduces O2 molecules in solution.

Anode : Lead : 2Pb + 40H 2Pbo +2H2O + 4e-

Electrolyte : potassium bicarbonate

SAMPLE

No polarising current required

Lead Anode

M Potassium Bicarbonate Solution

Silver CathodeO2 + 4e + 2H2O 4OH-

Page 280: 20275003 Physics Measurement 2

FUEL CELL

Advantages :CompactNo power supply requiredUnaffected by N2O

Disadvantage :

Slow response timelife-span 6-12 months

Page 281: 20275003 Physics Measurement 2

OPTODES - PRINCIPLE

1. Oxygen has the property of “quenching” fluorescenceof certain dyes.

2. Dyes exposed to light – electrons excited and release photons when they return to their original state (fluorescence).

3. Oxygen absorbs energy from excited electrons electrons return to original state without releasing photon

4. Absorption of light and reduction in light emitted is proportional to PO2

Page 282: 20275003 Physics Measurement 2

OPTODE - MECHANISM

• Optical fibre with dye coated tip

• O2 permeable membrane cover

• Sequential illumination of fibre causes dye to fluorescence

• Intensity of fluorescence depends on oxygen concentration at tip

• Fluorescence measured by photo multiplication.

Page 283: 20275003 Physics Measurement 2

OPTODE : Uses

• Intravascular PO2 monitoring

• Advantages :Independent of blood flowStableRapid response times

• Disadvantage :expensiveDye deteriorate with timeFibrin deposition

Page 284: 20275003 Physics Measurement 2

Oximetry

Kramer optically measured the OKramer optically measured the O22 in animals in the early in animals in the early 1930's1930's

Karl Matthes in 1936 was the first to measure OKarl Matthes in 1936 was the first to measure O22 from from transmission of transmission of redred and and blue-greenblue-green light through the human light through the human earear

the term oximeter was coined by Millikan the term oximeter was coined by Millikan et alet al. in the 1940's. in the 1940's they developed a lightweight oximeter, a smaller version they developed a lightweight oximeter, a smaller version

of Matthes' design, which measured SaOof Matthes' design, which measured SaO22 by by transillumination of the earlobe using red & green filters transillumination of the earlobe using red & green filters covering Kramer's barrier layer photocellscovering Kramer's barrier layer photocells

Page 285: 20275003 Physics Measurement 2

Oximetry

the signal detected from the photocell under the the signal detected from the photocell under the green filter later proved to be in the green filter later proved to be in the IR rangeIR range

there were two technical problems with this approach,there were two technical problems with this approach,

there are many non-Hb light absorbers in tissuethere are many non-Hb light absorbers in tissue

the tissues contain capillary & venous blood in addition to the tissues contain capillary & venous blood in addition to arterial bloodarterial blood

Page 286: 20275003 Physics Measurement 2

TRANSMISSION OXIMETRY

Based on absorbance laws

Blood consists of a mixture of

Oxyhaemoglobin and Deoxyhaemoglobin

Page 287: 20275003 Physics Measurement 2

ABSORBANCE CURVES FOR HbO2 AND Hb

Absorbance

660 805 940

RED INFRARED

OXY Hb

DEOXY Hb

wavelength

ISOBESTIC WAVELENGTH

Page 288: 20275003 Physics Measurement 2

ABSORBANCE CURVES

Secondary Peaks of Absorbance

660 nm - Deoxyhaemoglobin

940 nm - Oxyhaemoglobin

805 nm - Isobestic point

defined as point at which absorbances of HbO2 and Hb are equal.

Depends on haemoglobin concentration

Page 289: 20275003 Physics Measurement 2

CO - OXIMETER

• Measures Oxygen saturation

• Based on absorbance curves

• Requires haemolysis of blood sample before “sats” measurement

• 2 types - Reflectance

- Transmission

Measure at 4 wavelength to enable measurement of metHb and HbCO

Page 290: 20275003 Physics Measurement 2

CO-OXIMETER

ADVANTAGES :

Light absorbance measured at several

wavelength enables fraction estimation.

DISADVANTAGES :

Cannot provide continuous monitoring

Expensive cost and maintenance

Page 291: 20275003 Physics Measurement 2

PULSE OXIMETRY

Page 292: 20275003 Physics Measurement 2

Pulse Oximetry

early 1970's, Japanese engineer Takuo Aoyagi working on a early 1970's, Japanese engineer Takuo Aoyagi working on a

dye dilution method for CO, using an earpiece densitometerdye dilution method for CO, using an earpiece densitometer noted that the noted that the pulsatile componentspulsatile components of the red & IR of the red & IR

absorbances were related to SaOabsorbances were related to SaO22

prototype, built by Nihon Khoden, was tested clinically in prototype, built by Nihon Khoden, was tested clinically in

1973 and the first commercial prototype available in 19741973 and the first commercial prototype available in 1974 further refinements were required and widespread use did further refinements were required and widespread use did

not eventuate until the early not eventuate until the early 1980's1980's

Page 293: 20275003 Physics Measurement 2

Pulse Oximetry

the signal detected from the photocell under the the signal detected from the photocell under the green filter later proved to be in the green filter later proved to be in the IR rangeIR range

there were two technical problems with this approach,there were two technical problems with this approach, there are many non-Hb light absorbers in tissuethere are many non-Hb light absorbers in tissue the tissues contain capillary & venous blood in addition to the tissues contain capillary & venous blood in addition to

arterial bloodarterial blood

Page 294: 20275003 Physics Measurement 2

Pulse Oximetry

these were overcome by first measuring the these were overcome by first measuring the absorbance of the ear while it was compressed absorbance of the ear while it was compressed to remove all bloodto remove all blood

after this after this bloodless "baseline"bloodless "baseline" measurement measurement the ear was heated to the ear was heated to "arterialise""arterialise" the blood the blood

this device was shown to accurately predict this device was shown to accurately predict intraoperative desaturations, however, due to the intraoperative desaturations, however, due to the technical difficulties was never adopted on masstechnical difficulties was never adopted on mass

Page 295: 20275003 Physics Measurement 2

Nomenclature

SaOSaO22 = 100.(O = 100.(O22 content)/(O content)/(O22 capacity) capacity) arterial blood saturation measured arterial blood saturation measured in vitroin vitro OO22 capacity the amount of O capacity the amount of O22 which can combine with which can combine with

reduced Hb, reduced Hb, withoutwithout removing COHb or MetHb removing COHb or MetHb thus, at high Pthus, at high PaO2aO2 the SaO the SaO22 = 100% = 100%

irrespective of the [COHb + MetHb]irrespective of the [COHb + MetHb] HbOHbO22= = oxyhaemoglobin concentrationoxyhaemoglobin concentration

multiwavelength spectrometers measure all speciesmultiwavelength spectrometers measure all species SaOSaO22 computed from P computed from PO2O2 and pH approximates SaO and pH approximates SaO22, not , not

HbOHbO22

SpOSpO22 = = pulse oximeter saturationpulse oximeter saturation

Page 296: 20275003 Physics Measurement 2

Methodology

2 wavelengths of light,2 wavelengths of light, redred = 660 nm= 660 nm IRIR = 910-940 nm= 910-940 nm

the signal is divided into two components,the signal is divided into two components, acac == pulsatile arterial bloodpulsatile arterial blood dcdc == non-pulsatile arterial bloodnon-pulsatile arterial blood

+ tissue + capillary blood + venous + tissue + capillary blood + venous bloodblood

NB: all pulse oximeters assume that only the NB: all pulse oximeters assume that only the pulsatile absorbancepulsatile absorbance is arterial blood is arterial blood

Page 297: 20275003 Physics Measurement 2

AC AND DC SIGNALSRECEIVED BY PULSE OXIMETER

ACVariable absorption due to pulsatile arterial blood

DC

Absorption due to arterial blood

Absorption due to venous blood

Tissue absorptionTISSUE

VENOUS BLOOD

Page 298: 20275003 Physics Measurement 2

Methodology

for each wavelength, the oximeter for each wavelength, the oximeter determines the ac/dc fractiondetermines the ac/dc fraction independent of the incident light intensityindependent of the incident light intensity

= = pulse added absorbancepulse added absorbance the the ratio (R)ratio (R) of these is calculated, of these is calculated,

R =R = (ac absorbance/dc absorbance)(ac absorbance/dc absorbance)RedRed

(ac absorbance/dc absorbance)(ac absorbance/dc absorbance)IRIR

= A= A660nm660nm / A / A940nm940nm

Page 299: 20275003 Physics Measurement 2

R and SpO2

this value varies from,this value varies from,

SaOSaO22 = 100% = 100% R = 0.4R = 0.4 (0.3)(0.3)

SaOSaO22 = 85% = 85% R = 1.0R = 1.0

SaOSaO22 = 0% = 0% R = 3.4R = 3.4

Page 300: 20275003 Physics Measurement 2

R and SpO2

Page 301: 20275003 Physics Measurement 2

Methodology

the photo-detector diodes of the sensor will also the photo-detector diodes of the sensor will also register register ambient lightambient light

interference is reduced by cycling the lightinterference is reduced by cycling the light red only red only infrared only infrared only both off both off repeated at 480-1000 Hz in an attempt to subtract the repeated at 480-1000 Hz in an attempt to subtract the

ambient light signal, even when this is oscillatingambient light signal, even when this is oscillating this allows accurate estimation of SpOthis allows accurate estimation of SpO22 at arterial at arterial

pulse frequencies ~ 0.5-4 Hz (30-240 bpm)pulse frequencies ~ 0.5-4 Hz (30-240 bpm) data is averaged over several cyclesdata is averaged over several cycles

Page 302: 20275003 Physics Measurement 2

Uses: Oxygenation

anaesthesia & recoveryanaesthesia & recovery intensive careintensive care emergency care & transportemergency care & transport labourlabour premature & newborn infantspremature & newborn infants home & hospital monitoring for SIDShome & hospital monitoring for SIDS patients in remote locations eg XRay, MRIpatients in remote locations eg XRay, MRI "office" procedures eg. dentistry, endoscopy"office" procedures eg. dentistry, endoscopy

Page 303: 20275003 Physics Measurement 2

Uses: Circulation

systolic BP & pleth waveform appearancesystolic BP & pleth waveform appearance inflation better than deflationinflation better than deflation

sympathetic blockade with central neuraxis anaesthesiasympathetic blockade with central neuraxis anaesthesia autonomic dysfunction with valsalva manoeuvreautonomic dysfunction with valsalva manoeuvre anecdotally reported usesanecdotally reported uses

patency of the ductus arteriosuspatency of the ductus arteriosus level of ischaemia in PVDlevel of ischaemia in PVD patency of arterial graftspatency of arterial grafts circulation in reimplanted digits or graftscirculation in reimplanted digits or grafts

Page 304: 20275003 Physics Measurement 2

Uses: Therapy

optimise Foptimise FIIOO22 in ventilated patients in ventilated patients

optimise CPAP or PEEPoptimise CPAP or PEEP

extubation of ventilated patientsextubation of ventilated patients

adjust Oadjust O22 therapy in preterm infants therapy in preterm infants

no consensus on optimal levelsno consensus on optimal levels

optimisation of home Ooptimisation of home O22 therapy therapy

Page 305: 20275003 Physics Measurement 2

Signal:Noise

Freund Freund et al.et al. 1.12%1.12% failure failure cumulative > 30 mins in 11,046 anaestheticscumulative > 30 mins in 11,046 anaesthetics

Gilles Gilles et al.et al. found a found a 1.1%1.1% incidence incidence 2 x 15 mins in 1,403 anaesthetics2 x 15 mins in 1,403 anaesthetics

automatic gain controlsautomatic gain controls amplification of low signal strengthsamplification of low signal strengths

low signal to noise ratio low signal to noise ratio most new meters give "low signal strength" warnings once most new meters give "low signal strength" warnings once

the ac component falls below an arbitrary fraction of the the ac component falls below an arbitrary fraction of the total transmitted light (0.2% for the Biox-Ohmeda)total transmitted light (0.2% for the Biox-Ohmeda)

Page 306: 20275003 Physics Measurement 2

Low S:N Causes

low perfusion pressurelow perfusion pressure

motion artefactmotion artefact

ambient lightambient light

skin pigments & dyesskin pigments & dyes

probe positionprobe position the "penumbra effect" the "penumbra effect"

Ventilation - a large paradox may lead to searchingVentilation - a large paradox may lead to searching

venous pressure wavesvenous pressure waves - TI, reflectance operation- TI, reflectance operation

electrocauteryelectrocautery - most unit are now immune- most unit are now immune

MRI interferenceMRI interference - rare, usually lead distorts MRI- rare, usually lead distorts MRI imageimage

Page 307: 20275003 Physics Measurement 2

Ultrasound and anaesthesia

Page 308: 20275003 Physics Measurement 2

ULTRASOUND

• Sound = disturbance propagating in material (Air, water, tissue or solid)

• Characterized by frequency and intensity.

• Frequency measured in hertz

• ULTRASOUND = sounds waves > 20 KHz Cannot be perceived by human ear.

Page 309: 20275003 Physics Measurement 2

WAVELENGTH OF SOUND

• Sound Wavelength = Velocity frequency

• Shorter wavelength higher resolution less penetration

• Compromise between penetration and resolution required.

Page 310: 20275003 Physics Measurement 2

SOUND PRODUCTION

• Ultra-sound probe = Transducer containing an array of piezo-electric crystals.

• Electrical voltage applied to crystals causes piezo- electric crystals to oscillate at resonant

frequency.

• Electrical energy - converted to sound energy.

Page 311: 20275003 Physics Measurement 2

ELECTRICAL ENERGY

Electrical Energy

Piezo – electriccrystals oscillate

Sound

Electrical energy

Page 312: 20275003 Physics Measurement 2

ULTRASOUND PROPAGATION

In homogenous tissues : -

ultrasound is absorbedAbsorption – least in fluids greatest in solid tissues

Absorbed energy converted to heat (small)

Amount of heat dissipated hence useless

Page 313: 20275003 Physics Measurement 2

ULTRASOUND PROPAGATION

In heterogenous tissues :

• Ultrasound strikes interfaces

• Wave is either a) refracted - transmitted – thro’ interface b) reflected - depends on smooth (specular) or non-smooth

surfaces.

• Bone and calcium more reflective

Page 314: 20275003 Physics Measurement 2

PULSED SOUND WAVES

• Used to prevent transmitted and reflected sound waves.

• Pulse repetition frequency = 10 – 20 Hz

• Longer path sound wave travels - lower is PRF

Page 315: 20275003 Physics Measurement 2

PULSES OF ULTRASOUND

Usually 2.5 to 7.5 MHz

Frequency - resolution - penetration

Page 316: 20275003 Physics Measurement 2

IncidentReflected Wave

Surface

Incident Wave

“Scattering” of Ultrasound

ULTRASOUND REFLECTION

Page 317: 20275003 Physics Measurement 2

REFLECTED ULTRASOUND (ECHO)

Two quantities measured :

(a) Time delay between sound transmission and reception of reflected echo.

(b) Intensity of reflected signal High echo reflection - whiteLess reflection - greyNo reflection - Black

Page 318: 20275003 Physics Measurement 2

A - MODE (AMPLITUDE)

Brief ultrasound pulses in one direction.

Reflected ultrasound amplitude plottedAgainst time

Time & distance from probe

Amplitude

Peaks = reflective interface

Time (distance)

Page 319: 20275003 Physics Measurement 2

B - MODE (BRIGHTNESS)

Brief ultrasound pulse in one direction

Reflected ultrasound measured

Amplitude = Brightness of reflected ultrasound

Page 320: 20275003 Physics Measurement 2

M – MODE (MOTION)

• Repeated B-Mode pulses graphed against time base.

• > 1000 pulses per second

• Good resolution

• Provides one-dimension image against time.

• useful for value motion

Page 321: 20275003 Physics Measurement 2

2-D ULTRASOUND

• Multiple crystals (linear or phased array) or moving crystals

• Sequential B-mode pulses across 90o

• Single image displayed

• Real time movement

Page 322: 20275003 Physics Measurement 2

DOPPLER PRINCIPLE

• Frequency of transmitted sound from a moving object

alters depending on velocity and direction of object.

• Change in frequency proportional to

a) ultrasound frequency

b) Cosine of angle between ultrasound beam

direction and moving object.

Page 323: 20275003 Physics Measurement 2

DOPPLER SIGNAL

• Maximal signal when sound moves towards probe – higher pitch (frequency).

• Lower pitch when sound moves away from probe.

• Pitch change is due to compression and rarefaction of sound waves.

Page 324: 20275003 Physics Measurement 2

USES OF DOPPLER

• Examine direction and velocity of blood flow in vessels and heart

• Estimate velocities and therefore measure pressure gradients, using Bernoulli equation

P = 4V2

• Types of Doppler used ;

a) Pulse wave b) Continuous Doppler

Page 325: 20275003 Physics Measurement 2

PULSED-WAVE DOPPLER

• Depends on Doppler shift

• Doppler shift frequency of reflected waves which depends on velocity or reflected wave.

• Used to measure velocity of red blood cells V = FDC

2fo Cos Q V = Velocity of red blood cells FD = Doppler shift

Fo = Ultrasound frequency Q = angle between flow and sound wave.

Page 326: 20275003 Physics Measurement 2

PULSED-WAVE DOPPLER- Limitations -

• Large angles - results inaccurate

• High velocity flows > 0.6m/s cannot be accurately

measured by intermittent pulses (causes “aliasing”)

Page 327: 20275003 Physics Measurement 2

CONTINUOUS WAVE DOPPLER

Separate crystals - emit & receive ultrasound continuously along 1 axis

Frequency Spectrum & velocity of interfaces

Graph of Velocity range vs time plotted

Page 328: 20275003 Physics Measurement 2

CONTINUOUS WAVE DOPPLER

Advantages :

Can measure fast flows

Calculate valve gradients

Disadvantages :

Small incident angle required

Page 329: 20275003 Physics Measurement 2

COLOURED DOPPLER

Pulsed wave used on 2 D scan

Velocity depicted as colour

Advantage : Easy visualisation

Disadvantage : high velocity – colour reversal rapid turbulent flow produce colour “jets”

Page 330: 20275003 Physics Measurement 2

TOE PROBE

Phase array 2 D probe

64 piezo-electric crystals

Mounted on gastroscope (9mm)

Can be monoplane biplane (2 array) multiplane (rotating array)

Page 331: 20275003 Physics Measurement 2

CLINICAL APPLICATIONS OF ULTRASOUND

1. Examination of structure

Brain

Neck

Chest - pleural fluid

Obstetrics

Abdominal structures

Blood vessels

2. Interventional Procedures

Guide placement of needles

Page 332: 20275003 Physics Measurement 2

CLINICAL APPLICATIONS - DOPPLER

1. Sense blood flow in blood vessels

e.g. Thrombo-embolism

Thrombosis

2. Measurement of blood pressure

a) sense onset of blood flow

b) sense movement of arterial wall

Page 333: 20275003 Physics Measurement 2

CLINICAL APPLICATIONS –DOPPLER (CONT’D)

3. Cardiac output measurement

a) mean velocity

b) cross-sectional area of aorta or left ventricular

outflow tract.

4. Fetal Heart movements and heart rate

5. Valve functional, Myocardial wall movement

6. Transcranial Doppler - Velocity of blood flow in

cerebral vessels.