Digital Multimeter Measurement Errors Series System Cabling Errors and DC Voltage Measurement Errors in Digital Multimeters Application Note AN 1389-1 When making measurements with a digital multimeter (DMM), common errors will crop up. The following discussion will help you eliminate potential measure- ment errors and achieve the greatest accuracy with a DMM. This paper covers system cabling errors and dc voltage measurement errors. For an overview of ac voltage measurement errors, see Application Note AN 1389-3. For a discussion of resistance; dc current; ac current; and frequency and period measurement errors, see Application Note AN 1389-2. (NOTE: The Agilent 34401A, a 6-1/2-digit, high-performance DMM with both benchtop and system features, will be used as an example throughout this article). Introduction

Upload: thangarajic

Post on 01-Oct-2015

215 views

Category:

Documents


0 download

DESCRIPTION

agilent

TRANSCRIPT

  • Digital Multimeter Measurement Errors Series

    System Cabling Errors and DCVoltage Measurement Errorsin Digital Multimeters

    Application Note AN 1389-1

    When making measurements with a digital multimeter(DMM), common errors will crop up. The followingdiscussion will help you eliminate potential measure-ment errors and achieve the greatest accuracy witha DMM. This paper covers system cabling errors anddc voltage measurement errors. For an overview ofac voltage measurement errors, see Application NoteAN 1389-3. For a discussion of resistance; dc current;ac current; and frequency and period measurementerrors, see Application Note AN 1389-2. (NOTE: TheAgilent 34401A, a 6-1/2-digit, high-performance DMMwith both benchtop and system features, will be usedas an example throughout this article).

    Introduction

  • Radio Frequency Interference Most voltage-measuringinstruments can generate falsereadings in the presence of large,high-frequency signal sourcessuch as nearby radio and televisiontransmitters, computer monitors,and cellular telephones. High-frequency energy can also becoupled to the multimeter onthe system cabling. To reduceinterference, try to minimize theexposure of the system cabling tohigh-frequency RF sources. If theapplication is extremely sensitiveto RFI radiating from the multi-meter, use a common mode chokein the system cabling, as shown inFigure 1, to attenuate multimeteremissions.

    Thermal EMF Errors Thermoelectric voltages, the mostcommon source of error in low-level voltage measurements, aregenerated when circuit connectionsare made with dissimilar metalsat different temperatures. Eachmetal-to-metal junction forms athermocouple, which generatesa voltage proportional to thejunction temperature. It is agood idea to take the necessaryprecautions to minimizethermocouple voltages andtemperature variations in low-level voltage measurements.The best connections are formedusing copper-to-copper crimpedconnections. Figure 2 showscommon thermoelectric voltagesfor connections between dissimilarmetals.

    Noise Caused by Magnetic Fields When you make measurementsnear magnetic fields, take pre-cautionary steps to avoid inducingvoltages in the measurementconnections. Voltage can beinduced by either movement ofthe input connection wiring ina fixed magnetic field, or bya varying magnetic field. Anunshielded, poorly dressedinput wire moving in the earthsmagnetic field can generateseveral millivolts. The varyingmagnetic field around the acpower line can also inducevoltages up to several hundredmillivolts. Be especially carefulwhen working near conductorscarrying large currents.

    Where possible, route cablingaway from magnetic fields, whichare commonly present aroundelectric motors, generators,televisions and computer monitors.In addition, when you areoperating near magnetic fields,be certain that the input wiringhas proper strain relief and is tieddown securely. Use twisted-pairconnections to the multimeter toreduce the noise pickup loop area,or dress the wires as closelytogether as possible.

    2

    System Cabling Errors

    Figure 1.

    Figure 2.

    Copper-to- Approx. V/ C

    Copper < 0.3Gold 0.5Silver 0.5Brass 3Beryllium Copper 5Aluminum 5Kovar or Alloy 42 40Silicon 500Copper-Oxide 1000Cadmium-Tin Solder 0.2Tin-Lead Solder 5

  • Noise Caused by Ground Loops When you measure voltages incircuits where the multimeter andthe device-under-test (DUT) areboth referenced to a commonearth ground, a ground loop isformed. As shown in Figure 3,any voltage difference betweenthe two ground reference points(Vground) causes a current to flowthrough the measurement leads.This causes errors, such as noiseand offset voltage (usually power-line related), which are added tothe measured voltage.

    The best way to eliminate groundloops is to maintain the multimetersisolation from earth; do notconnect the input terminals toground. If the multimeter mustbe earth-referenced, be sure toconnect it and the DUT to thesame common ground point.This will reduce or eliminate anyvoltage difference between thedevices. Also, whenever possible,make sure the multimeter andDUT are connected to the sameelectrical outlet.

    3

    Figure 3.

    RL = Lead ResistanceR i = Multimeter Isolation ResistanceVground = Voltage Drop on Ground Bus

  • Common Mode Rejection Ideally, a multimeter is completelyisolated from earth-referencedcircuits. However, there is finiteresistance between the multimetersinput LO terminal and earthground, as shown in Figure 4.This can cause errors when youmeasure low voltages that arefloating relative to earth ground.

    Noise Caused by Injected Current.Residual capacitances in themultimeters power transformercause small currents to flow fromthe LO terminal to earth ground.The frequency of the injectedcurrent is the power line frequencyor possibly harmonics of thepower line frequency. The injectedcurrent is dependent upon thepower line configuration andfrequency. A simplified circuitis shown in Figure 5.

    4

    DC Voltage Measurement Errors

    Figure 4.

    Vf = Float VoltageRs = DUT Source Resistance ImbalanceRi = Multimeter Isolation Resistance (LO-Earth)Ci = Multimeter Input Capacitance ( 200 pF LO-Earth)

    Error (v) =Vf x RsRs + Ri

    Figure 5.

    Injected Current(50/60 Hz ac line leakage current)

  • With Connection A (see Figure 6),the injected current flows fromthe earth connection provided bythe circuit to the LO terminal ofthe DMM, adding no noise to themeasurement. However, withConnection B, the injected currentflows through the resistor R,thereby adding noise to themeasurement. With Connection B,larger values of R will worsenthe problem.

    Loading Errors Due to InputResistance Measurementloading errors occur whenthe resistance of the DUT is anappreciable percentage of themultimeters own input resistance.Figure 7 shows this error source.

    To reduce the effects of loadingerrors, and to minimize noisepickup, set the Agilent 34401Asinput resistance to greater than10 G for the 100 mVdc, 1 Vdc,and 10 Vdc ranges. The inputresistance is maintained at 10 Mfor the 100 Vdc and 1000 Vdcranges.

    5

    Figure 7.

    Vs = Ideal DUT VoltageRs = DUT Source ResistanceRi = Multimeter Input Resistance (10 M or >10 G)

    Error (%) = 100 x RsRs + Ri

    Figure 6.

    Note: Themeasurementnoise caused byinjected currentcan be significantlyreduced by settingthe integrationtime of the DMMto 1 power linecycle (PLC) orgreater.

  • Loading Errors Due to Input BiasCurrent The multimeters inputcapacitance will "charge up" dueto input bias currents when theterminals are open-circuited (ifthe input resistance is 10 G).The multimeters measuringcircuitry exhibits approximately30 pA of input bias current forambient temperatures from 0 Cto 30 C. Bias current will double(x2) for every 8 C change inambient temperature above 30 C.This current generates smallvoltage offsets dependent uponthe source resistance of the DUT.This effect becomes evident for asource resistance of greater than100 k, or when the multimetersoperating temperature is signifi-cantly greater than 30 C.

    6

    Figure 8.

    ib = Multimeter Bias CurrentRs = DUT Source ResistanceCi = Multimeter Input Capacitance

    Error (v) ib x Rs=

  • A modern DMM such as the34401A has many features to helpprotect you from many commonsources of error. However, whenmaking low-level or precisemeasurements, you must bemore aware of sources thatcontribution to measurementerrors. Environmental conditions such as changes in temperature,RF signals and electromagneticfields will have a significantimpact on the quality of DMMmeasurements. Once themeasurement environmenthas been optimized, propercabling can significantly reducemeasurement errors. The remainingsources of error can be calculatedand added to the measurementuncertainty.

    For more information aboutthe Agilent 34401A, go towww.agilent.com/find/34401a

    7

    Conclusion

  • By internet, phone, or fax, get assistance with all your test & measurement needsOnline assistance: www.agilent.com/find/assist

    Phone or FaxUnited States:(tel) 1 800 452 4844

    Canada:(tel) 1 877 894 4414(fax) (905) 282 6495

    China:(tel) 800 810 0189(fax) 800 820 2816

    Europe:(tel) (31 20) 547 2323(fax) (31 20) 547 2390

    Japan:(tel) (81) 426 56 7832(fax) (81) 426 56 7840

    Korea:(tel) (82 2) 2004 5004(fax) (82 2) 2004 5115

    Latin America:(tel) (305) 269 7500(fax) (305) 269 7599

    Taiwan:(tel) 080 004 7866 (fax) 080 028 6331

    Other Asia Pacific Countries:(tel) 1 800 375 8100(fax) (65) 836 0252(e-mail) [email protected]

    Product specifications and descriptions in this document subject to change without notice. Agilent Technologies, Inc. 2002Printed in the USA January 31, 20025988-5511EN

    Agilent Technologies' Test and Measurement Support, Services, and AssistanceAgilent Technologies aims to maximize the value you receive, whileminimizing your risk and problems. We strive to ensure that you get the testand measurement capabilities you paid for and obtain the support you need.Our extensive support resources and services can help you choose the rightAgilent products for your applications and apply them successfully. Everyinstrument and system we sell has a global warranty. Support is available forat least five years beyond the production life of the product. Two conceptsunderlie Agilent's overall support policy: "Our Promise" and"Your Advantage."Our PromiseOur Promise means your Agilent test and measurement equipment will meetits advertised performance and functionality. When you are choosing newequipment, we will help you with product information, including realisticperformance specifications and practical recommendations fromexperienced test engineers. When you use Agilent equipment, we can verifythat it works properly, help with product operation, and provide basicmeasurement assistance for the use of specified capabilities, at no extracost upon request. Many self-help tools are available.Your AdvantageYour Advantage means that Agilent offers a wide range of additional experttest and measurement services, which you can purchase according to yourunique technical and business needs. Solve problems efficiently and gain acompetitive edge by contracting with us for calibration, extra-cost upgrades,out-of-warranty repairs, and on-site education and training, as well asdesign, system integration, project management, and other professionalengineering services. Experienced Agilent engineers and techniciansworldwide can help you maximize your productivity, optimize the return oninvestment of your Agilent instruments and systems, and obtain dependablemeasurement accuracy for the life of those products.