A New Era in High Speed Analysis with Dynamic MRM and UHPLC
Tim Schlabach, Marketing
Anabel Fandino, Ph.D., R&D
PeteStone, Ph.D., Marketing
Agilent 1290 Infinity LC Attributes for MS - Infinitely Better for LC/MS
Page 4
1200bar @ 2mL/min for highest resolutionReduced Ion & Matrix Suppression
HT-Solution for up to 2,000 samples/dayComplete System Integration MassHunter
Less than 100 uL delay volumesFirst peaks can elute in < 6 seconds
High Pressures
LowDelay Vol.
High Throughput
Agilent 6000 Series LC/MS: Speed for UHPLC
Page 5
Fast MRMsfor UHPLC
40 MS Scans/sec10 MS/MS scans/sec
BestData
QualityFast Scans for UHPLC
Agilent 6460
Agilent 6530
1 ms dwell for MRM30 ms Pos/Neg Switching> 200 MRMs/ sec
Fast MRMs with 6400 Series QQQ – metabolism polarity switching in 30 millisecs plus 6 transitions (2 ISTDs)
Avg W1/2 = 1.3 sec 35 points across RSD [%] = 1.8
Avg W1/2 = 0.71 sec 15 points across RSD [%] = 2.9
Avg W1/2 = 1.3 sec 35 points across RSD [%] = 2.7
Avg W1/2 = 0.52 sec 11 points across RSD [%] = 6.5
Verapamil, 0.75 min
Buspirone, 0.40 min
Dextromethorphan, 0.53 min
Diclofenac, 1.04 min
Peaks widths with UHPLC only 2 seconds or less > ten uniform measurements across peak REQUIRED
Page 8
W = 2.1 sec
Cycle time < 115 ms
Avg. W1/2 = 0.72 sec
Verapamil, peak RSD 5%
Page 10
Dynamic MRM for 6400 Series Triple Quads monitors ion transitions only when compounds elute
With DMRM the number of concurrent MRMs are fewer than with Time Segments
Cycle Time:Concurrent MRMsDwell Time:
Page 11
Agilent 1290 Infinity LC + 6460 Triple Quad
9 replicates analyzed at 200 pg levelMean area %RSD = 3.2
Analysis of 600 Pesticide DMRM Transitions in 15 min.
17 min.
1290/6230 Accurate Mass TOF Pesticide Screening250 pesticides, 50 pg on column
3 min.
1.5 min.
Most peaks 1-2 seconds wide!
The Need for Speed in UHPLC/MS/MS
Cycle Times (MRM): <150 milliseconds
Dwell Times (MRM or DMRM): <= 2 millliseconds
Polarity Switching (MRM): < 50 milliseconds
Spectral scans: at 5 Hz or faster
PPM Mass Accuracy: at 5 Hz
20 K Resolution: at 5 Hz