a study of linear sweep voltammetric determination … · with cyclic voltammetry, the difference...

114
A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION OF CHROMIUM (vi), MOMI[1M (III), AND SULPHATE IONS IN CHROMIUM PLATING SOLUT ION BY YEUNG CRYING KIN A Thesis submitted in partial fulfilment of the requirement for the degree of Master of Philosophy in The Chinese University of Hong Kong. 1984 . Thesis committee: Dr. P.K. Hon Dr. W.K. Li Dr. S.P. So Prof. Robert A. Oet eryoung

Upload: others

Post on 01-Jan-2020

12 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION OF

CHROMIUM (vi), MOMI[1M (III), AND SULPHATE IONS

IN CHROMIUM PLATING SOLUT ION

BY

YEUNG CRYING KIN

A Thesis submitted in partial fulfilment of the

requirement for the degree of Master of Philosophy

in The Chinese University of Hong Kong.

1984

. Thesis committee:

Dr. P.K. Hon

Dr. W.K. Li

Dr. S.P. So

Prof. Robert A. Oet eryoung

Page 2: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a
Page 3: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

1

ACKNOWLEDGEMENT

I wish to express my sincere gratitude to Dr. P.K. Hon for his

invaluable guidance and encouragement during the course of this

research work and the fulfilment of my thesis.

I am grateful to Mr. M.C. Wong for his assistance in taking care

of the instruments.

I should also thank to Dr. F.T. Wong( Hong Kong Productivity

Centre) and Mr. P. C. Tang( Hong Kong TRI -0 Co.) for supplying

the plant sample solutions.

Department of Chemistry

The Chinese University of

Hong Kong

May 1984 Yeung Chung Kin

Page 4: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

2

ABSTRACT

Linear sweep voltammetric analysis using mercury drop as the

working electrode is used to determine the Cr(VI)( as CrO3) Cr( III

and SO 4= contents in common chromium plating solutions.

A simple and sensitive determination method for the Cr(VI) is

introduced. Properly diluted sample electrolyte is detected by a distinct

reduction wave of the Cr(VI) with E P at -0.25 to -0.30 V in NH 3 -NH 4 Cl

buffer solution. Good linearity of Ipto the concentration of Cr(VI) can

be obtained and little interference is encountered.

For the Cr(III) determination, three methods have been investigated:

(a) Cathodic process to determine the Cr(III) in K.SCN- HOAc buffer

solution with Ep at -0-75 to -0.80 V

(b) Anodic process to determine the Cr(III) in KSCN- HOAc buffer solution

with EP at -0-75 to -0.7 0 V

(c) After oxidation of the Cr(III) to Cr(VI), total Cr(VI) is measured

in NH3- NH4C1 buffer solution, and the Cr(III) content is indirectly

deduced.

The results show that, anodic method (b) and oxidation method (c)

are better than cathodic method (a) for less interference and can be

satisfactorily used in analyzing the. Cr(III) content in chromium plating

baths.

For SO4= determination, the SO ions are precipitated as PbSO

by a known and excess amount of Pb2+4 The excess Pb2+ in solution is

then measured in NaOAc- HOAc buffer solution by a well defined

reduction wave with Ep at -0.40 to -0.50 V. The content of SO 4= can

be' derived from the amount of Pb2+ found.

Both the calibration curve and the standard addition method have

been used in the S0d= determination and their r.eult a are compared.

Page 5: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

3

It can be concluded that, three main components in chromium

plating bath can be simply and rapidly analyzed by Linear Sweep

Voltammetry, and the accuracy of determinations( Less than 5%

of error) can meet the requirement of industrial analysis.

Page 6: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

CONTENTS

ACKNOWLEDGEMENTS

ABSTRACT

PART I INTRODUCTION

1.1. Linear Sweep Voltammetry

1.2. Electroplating Bath Analysis by Polarographic

Method

1.3. Analysis of Chromium Plating Bath

1.4. Inst rum ents

PART II DETERMINATION OF CHROMIUM (VI)

2.1.Introduction

2.2. Experimental

2.3. Results and Discussion

2.4. Conclusion

PART III DETERMINATION OF CHROMIUM (III)

3.1. Introduction

3.2. Experimental

3.3. Results and Discussion

3.4. Conclusion

PART IV DETERMINATION OF SULPHATE IONS

4.1. Introduction

4.2. Experimental

4.3. Results and Discussion

4.4. Conclusion

SUMMARY

REFERENCES

PACE

i

ii

1

3

3

5

6

6

9

30

31

33

36

77

79

80

82

103

104

105

Page 7: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

PART I TNTR ODTT PT T ON

1-1- Linear Sweet) Vol t ammeirv

Since its discovery in 1921, polarography has been brought about

a renaissance in 1980s. Its advancement is mainly attributed to the

increasing demand of rapid and sensitive analysis techniques and the

widespread availability of high quality and comparatively low cost

Recently, a variety of polarographic techniques have been developed,

such as: fast direct current, linear sweep voltammetry, normal pulse,

differential pulse, alternating current, stripping polarographic

techniques ,etc.

In linear sweep voltammetry( L.S.V.), a rapidly linear changing

potential( scan rate is generally more than 10 mvsec.) is applied, and

the peak current is measured as the function of applied potential.

In reversible electrode process, peak potential is related to

E2 in d.c. polarography.

For reversible process, is independent of the scan rate

Peak current I for reversible process can be given by:

= a. ii n JJ¥ o

Where: k: constant

n s number of electrons transferred

As electrode area

D: diffusion coefficient

V: potential scan rate

C i bulk concentration of electroactive species.

Page 8: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

When other experimental parameters are held constant, the

I P i s lin e a r ly proportional to C and is used to measure the electroactive

species in analytical works.

With cyclic voltammetry, the difference o f the peak potential fo r

cathodic process, (Ep) c, and that fo r anodic process, (Ep) a is i

‘ < V a - 2.2nP

( V c ‘ (▼) ^ 25°Cn

Furthermore, there are several advantages in applying t.S.V.

High scan rate is applied, so the I-E polarogram recording is

faster;

2. Comparing to d.c. polarography higher resolution can be obtained.

In addition, sensitiv ity can be down to 10— — 10~ M;

3- Several components whether they are different elements or the

same element of different valences can be detected by proper techniques

and selected electrolytes;

4- The working electrode is not restricted to the mercury drop, but

platinum, gold, silver, glassy carbon, etc* can also be used;

5- When the static mercury drop is used, changing of the dropping

is easy and it is required only one to several drops of mercury in one

determination.

However, L.S.V. has some limitations :

1. The output signal I p is dependent on the scan rate and the

surface area o f the electrode, these variations would affect the accuracy

of determination and must be defined and controlled carefully;

2. The sensitiv ity is lower when compared with the d ifferen tia l

pulse polarography ( 10 - 10 8 M ) and stripping voltammetric techniques

( 1(T8 - 10~9 M ) .

- 2 -

Page 9: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

1,2. ElectroDlatin Bath Analvfiin bv PolaroflraDhic Method

Control and monitor the content of plating bath is a prime impor¬

tant factor to ensure the quality of the products and to attain the

most economic benefit in production. In addition to the main metal ion

content, other organic additive reagents and impurities, etc, should

be cheeked rrmtlnpl v.

The conventional volumetric, gravimetric analysis methods,

spectrophotometry and potentiometric titration methods have been used

5for electro dating baths analvsis.

Polarographic analysis is getting greater application nowadays,

Rothstein et al. discussed the application of applying polarographic

method in analysis of brass, copper, electroless copper, gold, nickel,

palladium, tin-lead and zinc plating baths.

Bond et al. applied DPP and stripping voltammetric methods

7to V7.P tracp pi pmpnt p in inr. d! at inr hprth-

QOkinaka and Wolowodiuk developed new procedure in determinatior

of gold and cyanide contents by polarography and cobalt content by

coulometric method in analysis of hard gold plating bath.

1.3. Analysis of Chromium Plating Baths

Chromium plating is one of the most important plating process used

extensively in both decorative and functional applications. Incorpo¬

rated with bright nickel as undercoating, chromium plating provides

an attractive appearance and durable protective coating. 1 Hard Chromiui

plated articles are coated with a high hardness and excellent anti-

corrosion and anti-abrasion coating.

Page 10: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

In common chromium plating bath, the contents of the main compo—

n on+ a or a d d f a! 1 au a•

The optimum CrO_ to HoS0.( or SO.) ratio is about 100 I 1.3 2 4 4

And the Cr(lll) content should also be kept in the range of 1- 10 gl.

The S04~ is an indispensable component in chromium plating bath. It

acts as 1 catalyst 1 in the Dlatine- nronees

The Cr(lll) concentration is critical as well, if it is too low,

the plating efficiency is low; however, if it is too high, the chromium

coating would be darken and the conductance of electrolyte tends to be

By reviewing previous works concerned with the polarographic analysis

of chromium plating bath, it can be noted that, most of them are not in

details and can only be used as a simple guide for the analysis of

5chromium plating bath using d.c. polarographic method. Some polarogra¬

phic analytical methods for the Cr(Vl) and Cr(lll) are all in general

applications but not specifically applied for the chromium plating bath

analysis. 1 The method for SO~ determination is rather compli-4

A, 12

This research work is intended to develop or modify the L.S.V.

method using the static mercury drop as working electrode to determin

the Cr0_, Cr(lll), and SO~ contents in chromium plating baths.

The main points of the research work are 2

1. To examine the behaviour of reversibility of oxidationreduction

wave by which the measurements are based on;

2. To search for the most appropriate operating conditions for

the determination of the main components;

Page 11: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

3. To establish calibration curve and determine sample solutions!

4 To evaluate interference of co-exist components and impu-

rities in determination of the main components;

In view of its rapid and easy operation, linear sweep volt amine-

try technique is preferred in industrial routine analysis. In addi¬

tion, its adequate sensitivity and accuracy can meet the requirements

A Tr o+ witn n

1. Analyzer:'• PAR 1 Model 174A Polarographic Analyzer,

2. Electrode system s 1 PAR 1 Model 303-

Working electrode: SMDE( Static Mercury Drop Electrode)

Counter electrode: Pt wire.

Reference electrode s Ag AgCl.

Electrolysis cell: Borosilicate glass cell.

3. Recorder: 1 PAR 1 Model RE 0089 XY Recorder.

4. pH Meter: Chemtrix 60A.

Peak potential vs SCE is approximately 0.05 V more negative tha

the value given.

Unless otherwise specified all E or E. values in the following

dencriut ionn are vn A c Ac CI-

Page 12: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

6

PART II DETERMINATION OF CHRQ`IIIJM (vi)

2.1. Introduction

Chromium trioxide( Cr0 3) is the main chromium source in chromium

plating bath. CrO 3 is conventionally determined by redox volumetric

titration. 3' 4 JReoently, some other instrumental analytical methods,

such as: spectrophotometry and pot entiometric titration have also been

developed. 5 Moreover, the applications of olarop gacaphis and volt smm e-

tric analysis have been growing rapidly. Bobrowski and Barchanaka 1 3

applied oscillopolarographic method to determine Cr(VI). in chromium pla-

ting bath using triethanolamine and NaOH as supporting solution.' PAR

introduced differential pulse polarographic method using NH- NH Cl as3 4

supporting solution for Cr(VI) determination, however, there was no

detail information specifically for the determination of the Cr(VI) in

chromium plating bath.

This part of research work is to extend the method proposed by

' PAR', by linear sweep volt anmietric technique for the determination of

Cr(VI)( or Cr03) in chromium plating bath. Attempts will be made to

study: the -reduction irreversibility optimum operating conditions

and interference of cationic impurities.

2.2. Experimental

2.2.1. Reagent s

Unless otherwise specified, all reagents used were of analytical

reagent grade or guarantee grade. Deionized water was used in prepa-

ration of all reagents.

(1) c r(VI) stock solution( 1000 ppm

2.8287 g of 82Cr207 was dissolved in 500 ml of water, added

Page 13: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

50 ml 0.1N HoS0. and diluted to 1000 ml.

(2) NH_- NH.C1 buffer solutions

54 g of NHCl was dissolved in 800 ml water, the pH value was

adjusted by dropwise addition of concentrated NH solution to a pH3

value of 9.55, then diluted to 1000 ml.

The following pH values of- NHCl buffer solutions were pre¬

pared as well: 7.54, 8.54, 10.44, 10.87, 11.80, 12.52, 13.02, 13.15•

(3) Standard solution of Cr(lll), 1 gl

3.8O84 g of Cr(N0:z)_.9H_0 was dissolved in 300 ml water, added

50 ml of 0»1N HN0_ and diluted to 500 ml.

(4) Standard solution of Cu+, Fe, Ni+, Zn+

These standard solutions were used for the studies of cat ionic

interferences in the Cr(Vl), Cr(lll), and SO~ determinations.

(a) Cu+ solution, 5 gl

157081 g of Cu(OAc) .HO was dissolved in water and diluted to

1000 ml.

(b) Fe~ solution, 5 gl

36.1683 g of Fe(N0) .9H.0 was dissolved in water, added 10 ml

0.1N HN0_ and diluted to 1000 ml.3

(c) Ni+ solution, 5 gl

21.1940 g of Ni(OAcJgHgO was dissolved in water, and diluted to

1000 ml.

(d) Zn+ solution, 5 gl

10.4245 g of ZnCl was dissolved in water, added 10 ml of 0.1N

HC1, then diluted to 1000 ml.

(5) Chromium plating sample solutions

Three laboratory prepared sample solutions and two plants sample

solutions were used for determinations of the Cr(Vl), Cr(lll), and

Page 14: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

POnt Pn+. q 4 ri+ Vi o f nl 1? n rr vrn-rlro

Sam r0 a

(1)

(2)

(3)

(4)

(5)

150

250

Ann

2.5 C

5.0C

q nr

2.50

3.00

S.SO

Plant sample supplied by TRI-0 Co. H.K.

Plant sample supplied by H.K. Productivity Centre

• 1. Added as

2. Added as

3. Added as

2.2.2. Calibration of CrO, contents in sample solutions

For comparing the Cr0_ results measured by linear sweep voltamme-

trie method, the sample solutions were standardized by redox volumetric

4+!+ r» n 4- -1 rsy- HPVir roonl+ o 1 r.r c?•

(1:

(2:

(3:

(4:

( •

147.3- 0.5

247.3- 0.8

393.7- 0.9

232.4- 0.7

258.0 i 0.8

h Average value of 3 replicated measurements.

9. 9. Anal van a nrnrpurp hv T. _S-1

fl} 1 ml of sample solution was pipetted into a 100 ml flask

Page 15: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

then diluted to the mark. A 1 ml aliquot of the diluted sample solution was

further diluted to 100 ml in flask( i.e. a total of 104-folds dilution).

(2) A 5 ml aliquot of the diluted solution was pipetted into the

electrolysis cell, mixed with 5.00 ml of NEL- NH CI buffer solution

( pH= 955)• Then proceeded to the L.S.V. operation.

(3) L.S.V. operating conditions s

dispersion: 1- 2 min.

Potential scan: from -0.10 V to -1.60 V, recording scan: from

-0.10 V to -0.35 V.

Full current scale: 20 .MA. Each 1 measurement is the average

value of three measurements for every sample.

Scan rate: 100 mvsec.

Hg drop size: M

Unless otherwise specified, all the polarogram recording procedure

followed the above conditions.

2.3« Results and Discussion

2.3.1. Reduction polarograms of Cr(Vl) and Cr(lII

Cathodic reduction behavior of the Cr(Vl) and Cr(lll) in NH- 4

buffer solution axe illustrated by polarograms shown in Figure 2-1 and

Table 2-1.

It can be observed that, reduction polarogram for the Cr(Vl) shows a

distinct wave peak( E= -0.28 V). This reduction wave can be used

in the Cr(Vl) determination. Furthermore, it could be expected that

less interference would be encountered, since the E is located at a

rather Dositive res-ion of Dotential scanning-.

Page 16: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

c

0

c

2 pA

i -.1 i 1 o i' c: _1 7 -1.9 —2.1

2 pA

f| 1 I i i' IT)

E (V, vs AgAgCl)E

Page 17: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

Table 2-1. Reduction wave characteristics of Cr(Vl) and Cr(lll)

Ion

Cr(Vl)

50 ppm

Cr(III]

50 ppm

Potential scan

From -0.10 V

to -3.10 V

From -0.10 V

to -4.10 V

ep v

-0.28

-1.46

-1.54

-1.72

-1.48

-1.70

Wave form

Very sharp

Defined

Defined

Sharp

Defined

Poorly defined

2.32. Irreversibility of electrode reaction

2Bond pointed out that, a totally irreversible reduction

2electrode reaction should possess the following characteristics:

(1) Anodic wave peak current (I) is about zero;P a

(2) Cathodic wave peak current (I) is proportional to the

square root of scan rate V;

(3) Cathodic E b ecome more negative with increasing scan rate V;P

(4) (1) is linearly proportional to the concentration of thep c

electroactive species.

To study the irreversibility of electrode reaction of Cr(Vl) and

search for the appropriate operation conditions( scan rate, Hg drop

size, etc.-), following examinations were executed.

(l) Cathodic and anodic waves

Polarograms representing cathodic and anodic processes for the

Cr(Vl) are shown in Figure 2-2. Due to the great- reduction tendency

of Cr(Vl), no oxidation wave in anodic process was observed.

Page 18: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

(2) Dependent of I on V

Standard 10f 20, and 30 PP® Cr(Vl) solutions were examined by

different scan rates, and the I values are concluded in Table 2-2.

Table 2-2. Effect of on I

10 20 30

50 100 200 50 100 200 50 100 200

1.30 1.72 2.60 2.50 3.46 4.80 3.30 4.86 6.60

0.19 0.17 0.18 0.37 0.35 0.34 0.47 0.49 0.47

It can be concluded that, I is directly proportional to V®,1 P

and Ip V comes nearly to a constant for each specific concentra¬

tion of Cr(Vl).

( Anodic)

••• i

-0.5 -0.2 -0.1 0

( Cathodic)

-0.1 -0.2 -0.3 -0.4 -0.8

Figure 2-2. Anodic and cathodic waves of Cr(Vl)( 10 ppm

Page 19: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

(3) Effect of the scan rate on E and IP P

In evaluating the effect of scan rate on E and I 9 and to findP P

out the proper scan rate in determinations, standard solution of 10 ppm

Cr(Yl) was examined at different scan rates. The results are shown

in A 9-7.

Table 2-3 Effect of scan rate on E and IP I

Scan rateWave form

2

5

10

20

50

100

200

Ron

0.1

0.1

0.1

0.1

0.1

0.1

r i

-0.220

-0.224

-0.226

-O.2J0

-0.245

-0.255

-0.970

Undefined

Broaden

Veil defined

Broad€

From the above investigation, it can be noted that, the sc«

rate would have a great effect to both Ep and 1.

Page 20: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

(a) For a definite concentration of Cr(Vl), I is directly

proportional to the square root of the scan rate.

(b) As far as the effect of the scan rate on the E is concerned,

increasing scan rate would lead to a more negative value of the

(c) For obtaining a well defined wave peak, reproducible I values

and a faster recording of polarogram, the scan rate should be

kept in a definite value for all determinations. Knowing the effect

of the scan rate on the E and I, the scan rate should have a certain

value( more than 20 mvsec.), however, it should not be too

fast for easy control of polarogram recording. In general, the

scan rate in the range of 50- 200 mvsec is preferable. In the

following determinations, except for specific cases, the scan rate

is fixed at 100 mvsec.

Page 21: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

(4) Effect of the mercury drop size on 1

There are three mercury drop size modes( small, medium and

large) available for the K 33 static mercury drop electrode system

used in this work. The mercury drop size for H 30 3 system has been

calibrated by weighing method and the results are as follows:

S M L

Vt. of Hg( gdrop)

Vol. of Hg( mm 'drop)

0.0014

0.103

) 1.0

0.0027

0.199

1 1.9 J

0.005'

0.419

4.1

It is worthy to note; that, the drop size of the static mercur;

electrode will have an effect to the I measurements( Table 2-4)•

Table 2-4 Effect of the mercury drop Bize on It

Cr(Vl)

( PPtn

Hg drop siz

t( n A

4A

M

1.1 4£ 2.4

Ol

s 1 1

2. A 1 cc5.2

50

S

2.9

M

4.66

]

e.i

For a specific concentration of Cr(Vl) and scan rate( e.g.

100 mvsec), Ip is varied as the mercury drop size.

In order to have reproducible 1 measurements, the mercury dro;

size must be kept in a specific mode. In the following measurements

the mercury drop size is kept in the medium( M) mode.

Page 22: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

16

2.3.3. Effect of the pH value of buffer solution

To evaluate the effect of the pH of the buffer solution, 25 ppm

Cr(VI) standard solution was measured at different pH values of buffer

solution.( Table 2-5 and Figure 2-3, Figure 2-4)

Table 2-5. Effect of pH value of buffer solution

PH ofWave form

buffer solution

-0.167.54 3.60 Sharp

8.34 -0.193.84

-0.279.55 3.90

-0.3410.443.70

-0.3810.87Broaden3.10

11.80 -0.48 2.02

-0.5612.52 1.70

-0.7413.02 2.04

13.55 -0.93 2.50

The effect of changing pH value of buffer solution can be des-

cribed as follows:

(1) When the pH value of buffer solution is changed from about

7.0 to 9.0, the Ip value is getting higher and the E P is shifted to the

more negative side.

(2) When the pH of the buffer solution is higher than 10.5, the

Ep is continuously shifted to the more negative value, but the I isP

reduced. In addition, the wave form becomes poorly defined. the reason

for variation of EP and Ip with changing pH value of buffer solution

can be proposed that, if the pH value of the supporting solution is up to

certain extent that the reduction product Cr(III) would form a hydrolyzed

product Cr201. xH90 or cr(II I)- NH, complexes, so thus the

Page 23: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

17

IP ( A )4

3

2

1

6 8 10 12 14

pH

Figure 2-3. Effect of pH of buffer solution on IP.

Ep-1.0

-0.8

-0.6

-0.4

-0.2

6 8 10 12 14

pH

Figure 2-4. Effect of pH of buffer solution on Ep .

Page 24: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

electrode reaction tenda to be more irreversible, and E will shift to

the more negative value. On the other hand due to the electrolyte is

unstirred, if the hydrolyzed Cr(lll) or complex products are accumulated

around the vacinity of electrode area and less Cr(lll) ions are diffused

back, transportation of the Cr(Vl) from the bulk solution to the diffusion

layer is delayed, thus the attainment of I will be at an earlier stage

and I will be lower.

For rather high pH value of the buffer solution( pH higher than

13.0), the variation trend of I is reversed. It is presumably ascribed

to the amphoteric nature of the hydrolyzed Cr(lll) and the accumulation

of the hydrolyzed Cr(lll) does not exist any more, so the I value will be

rrVioT

(3) In order to obtain a well define wave peak and reproducible I

and Ep measurements, the pH of the buffer solution should be kept in the

range of 8.5- 10.5. In all the following determinations, the pH value

of the buffer solution was 9• 55•

2.34 Calibration curve

It can be found that, the I value is linearly proportional to the

concentration of.Cr(Vl) in a rather wide range. As the content of Cr0_

in the chromium plating bath is generally in the range of 150- 400 gl,

so that the standard Cr(Vl) solution in the range of 5- 50 ppm( corres¬

ponding to 96.2- 962 gl of CrO in the plating solution, because the

sample solution has been diluted from the original plating solution by

10 -folds) should be used to establish the calibration curve.

In order to examine the upper and lower limits of detection, the Cr(VI

standard solutions in ranges of 5- 500 ppm and 0.5-5 Ppni are also used

Page 25: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

to set up the calibrations It should be noted that9 they are not

applicable to the determination of GrO in real samples9 the purpose

of calibration hereby is to search for the upper and lower limits of

detection. Calibrating procedure is followed the procedure described

in 2.2.(~( 1.

Blank S 10 20 V) AO RH

0.20 0.98 1.81 3-36 5.02 6.60 8.30

Pull current scale is 20JUA. The calibration curve is shown in

T34 O C

t r 1 1 M« t if V 1« C 1 U ft-

Blank 50 75 100 150 200 500 400 500

0 8.50 12.3 16.0 25.3 30.6 41.2 45.0 48.0

Pull current scale is 200 ju A. The calibration curve is shown i

Figure 2-6. It cam be shown that, if the concentration of Cr(Vl) is

higher than 200 ppm, I value is no longer linearly related to the

AAn A V O+ 4 AA A rv»f T7T 1

Blank 0.5 0.75 1 1.5 2 4 1

0.30 0.30 0.30 0.43 0.50 0.58 0.87 1.02

Pull current scale is 5 The calibration curve is shown in

Figure 2-7. It can be noted that, when the concentration of Cr(Vl) is

lower than 1 ppra, 1 is hardly detected.

By means of the least squares fitting method, linear relationship

of 1 to the concentration of Cr(Vl) can be established and correlation

coefficient (R) can be calculated as well. The results are tabulated in

Page 26: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

8

7

6

5

4

3

2

1

C ( ppm )

10 20 30 40 50

Figure 2-5. Calibration curve for Cr(VI) ( 5 - 50 ppm ).

IP ( A)

Page 27: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

21

50

40

30

20

10

C ( ppm )

100 200 300 400 500

Figure 2-6. Calibration curve for Cr(VI) ( 50 - 500 ppm ).

1.2

0.8

0.4

C (ppm)

1 2 3 4 5 6

Figure 2-7. Calibration curve for Cr(VI) ( 0.5 - 5 ppm ).

IP ( A)

IP ( A)

Page 28: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

22

Tabl 0 2-6

Table 2-6. Relationship of I (std) and concentration ofD

and correlation coefficient, R

Concn. range of

Relation o K

0.5- 50.275+0.148 C 0.995

5- 50 0.170+ 0.162 C 0.998

50- 2001.060+ 0.148 C 0.994

In the following determinations, formula

was used to calculate the Cr(VI concentratfnn.

As the sample being measured has been diluted by 104 _fn7

from the original plating solution, therefore the content of CrO3 will be s

2.3.5. Detection limit

By using 1 ppm of standard Cr(VI) solution. 12 trial m PAA,7Tam can o

for rere taken in 3puffer solution. Standard deviation

mean value ofnd the detection limit are calculated.

Detection limit

Page 29: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

2--6- Analvnla of aaiiit)] 0 noltrh-fon

Measurement of sample solutions prepared as described in 2.2,1,

are performed by calibration curve method.

Table 2—7- Analveie of naronl e «o1irfc,fnr»a

found Relative

(1

(2

(3

(4

147.

247.

393.'

252.,

255.

7.5S

13.1

20.S

12.1

•17 0

146- 1

252± 5

402 t 5

237- 5

250 t 4

-0.9

+1.9

+2.1

+2.0

-2.0

1 iv STO rf A val IIS f TATS A Q+ A1 Til SO OIITSm C

The polarograms for the samples are shown in Figure 2-

Page 30: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

24

2 A

Sample No. 5

4

3

2

1

Blank

E ( V, vs Ag/AgCl )

-0.1 -0.2 -0.3 -0.4

Figure 2-9. Polarograms for samples.

I

Page 31: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

2.3.7• Interference of cations

To assess the interference of different cations to the determi¬

nation of Cr(Vl), different cation solutions were measured under the

same conditions as the sample solution. The E values and wave forms areP

are shown in Table 2-8.

Table 2-8. E and wave forms of different cations.

Cations

50 ppm, each

Wave form

-0.10

-O.36

-0.90

-0.94

— 1• 16

-1.28

-1.0

-0.46

-1.0

-0.71

Broaden

Poorly defined

SharD

Sharp

Broaden

Undefined

Broaden

Sharp

In common chromium plating bath, the most probably present

cationic impurities are s Cu, Fe5, Hi2+ and Zn2+, hence these ca-

tionic interferences on Cr(Vl) determination are discussed as follows.

(l) Interference of fhi+

Since the E values for Ci ( -0.10 V, -O.36 V) are close to that

ofit would probably interfere the Cr(Vl)

determination.

21To estimate the upper limit of tolerable Cu content for the Cr(vi]

determination, 7.5 ppm of standard Cr(Vl) solution( corresponding to

144 gl of CrO in plating solution) was mixed with different amounts of

Cu( 1-5 ppm, correspond to 10-50 gl of Cu2+ in plat ins: bath, since

Page 32: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

4

it is a 10- fold dilution of the original sample)• The results of

CrOj determination are concluded in Table 2—9 depicted in Figure

O4.Table 2—9. Interference of Cu

CrO (gl)

O.Chi aAAaA( xrl

CrO found (gl) Error(±%)

144

0

10

20

SO

146

153

158

163

+1.4

+6«1

+9-5

+ 1 3

It can be derived from Figure 2-10 that, if the addition of Cu2+

is up to 7.5 then +50 per cent of relative error( adopted as th€

lowest acceptable accuracy) would be resulted to the determination of

2+

CrO. Therefore, when the amount of Cu present in plating bath is

about 20 times less than that of CrO( 7.5144 120), it would3

cause interference. In real plating solution, as an impurity present

2+in chromium plating bath, the Cu content is generally only 1-2 gl,

2+hence Cu should not pose a problem to the CrO_ determination.

Outline the concentration limii

2+of the Cu causing+ 50%

error on the CrO, found.

,CrOT (gl)

16'

155

15

1 AS

nr

20 40

Cu+ added (gl)

( Actual CrO,; 144 l)

2+Figure 2-10. Plot of the Cu added vs CrO, found.

Page 33: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

In special cases if it is necessary to remove the interference

2+

of Cu, it can be done by precipitation as Cu(OH)( K= 1.3 x 10~202 ap•

The procedure is as follows i

1. A 2 ml aliquot of sample solution is pipetted into a 100 ml

flask, then diluted to the mark.

2. A 5 ml aliquot of diluted solution is pipetted into a 200 ml

beaker, add 50 ml water and 5- 10 ml 1N NaOH. The solution is heated

to boiling for 2-3 minutes, then filtered. The filtrate is collected in

a 250 ml flask. Wash the ppt and filter paper 3-4 times with 150 ml

Wat PT and i 1 lit. o+. ho cnl 11+. i nn+. n +ho marlr

3. A 25 ml aliquot of resulting solution is diluted to 100 ml in

a flask. Then proceed for the determination of CrO as usual.

The availability of the above mentioned procedure for eliminating

2+

Cu interference has been verified by the measurement of sample solution

2+containing known amounts of CrO and Cu. The results of the determina-

tions of CrO before and after treatment are compared and shown in

Tahlo 9-in.

rs

Table 2-10. Elimination of Cu+ interference

Snl 11+ rm CrO found Relative error

192

0

10

in

194

1213

21Q;

+ 1.0

+ 1 1

+0.5

2+1 Before removal of Cu

2+2 After removal of Cu

Page 34: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

(2) Interference of Fe'f Hi2 and Zn2

It is derived from Table 2—8 that9 reduction waves of Fef Hi2

2+

and Zn are all shoving much more negative values of E than that ofP

the Cr(Vl) in NH- NCl buffer solution, so it is expected that these

cations should not cause interference to the determination of CrO.

To evaluate their interference quantitatively, 7.5 ppm of Cr(Vl)

solution( corresponding to 144 g1 CrO in plating bath) was mixed

seperately with different amounts of Fe5, Hi2, and Zn2( the maximum

content of cations added is 200 ppm which is corresponding to 2000 gl

of cations content in the plating bath after 10-folds dilution). The

results are shown in Table 2-11.

Table 2-11. Interference of Fe', Ni2, Zn2

The ratio of cations

added to CrO,

CrOj found

(gD

Relative error

(±90

o

1.04

14.0

147

146

136

+2.1

+1.4

-5.6

144 Hi2 0

1.04

14.0

144

146

152

0

+1.4

+5.6

Zn2 0

1.04

14.0

145

147

152

+0.7

+2.1

+5.6

Negative error is ascribed to the formation of Fe(0H) in

alkaline solution when a definite concentration of Fe is added, thus

the Cr(Yl) would be occluded into the ppt and the concentration of

the electroactive species is reduced.

Page 35: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

From the above investigation it can be known that, only when the

I. p. O.

amounts of Fe f Ni, Zn are as high as 14 times( 2000144141)

greater than that of CrOy these cations will bring about i 5.0 per

cent error to the determination of CrO, Since the cationic impurities

present in chromium plating bath axe normally only a few grams per liter,

therefore, they should not cause interference.

(3) Interference of Cr(lll)

In common chromium plating bath, a proper content of Cr(lll)

( 1- 10 gl) must be present. The value for the Cr(lll)( about

-1.48 V, referred to Table 2-1) is much more negative than that of the

Cr(Vl), so it would not cause interference to CrO determination.

Estimation for the interference of the Cr(lll) was carried out in

the same way as the interference of the cationic impurity. The results

are shown in Table 2-12.

Table 2-12. Interference of Crfllll

Thp ratio of Erf TIT

o nn f A

CrO found Error

0

1.04

14.0

on A

146

147

147

-MR

+ 1.4

+ 2.1

+ 2.1

+2.8

From the above investigation it can be noted that, even the

addition of the Cr(lll) is as high as 20 times greater than that of

CrO( 3000144 c 201), no significant error is observed, hence the

Cr(lll) should not bring about interference to the CrO determination,

Page 36: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

24 Conclusion

In NHj- NH4C1 buffer solution( pH« 9-55), the Cr(Vl) shows

a distinct irreversible reduction wave with Ep at -0.25 to -0.50 V

( vs AgAgCl). Its I values are linearly proportional to the

concentration of Cr(Vl) in a wide range( 1- 200 ppm). Detection

The reduction wave of the Cr(Vl) can be successfully and easily

applied to the determination of the CrO content in chromium plating

baths. With a proper dilution, sample solution can be directly detected

by linear sweep voltammetry. Determination error is less than 2.5 per

cent( determinations from 5 samples)• Except for rather high amounts of

Cu other components and cat ionic impurities would not cause inter¬

ference.

Page 37: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

T1TT TTTDETI8KIKATI0K OP CHRCKIIK (ill)

4

51 Introduction

When a new chromium plating bath ia prepared, it is necessary to

electrolyse the GrOj solution until an adequate amount of Cr(Hl) are

formed from the auxiliary cathode plates. The Cr(lll) content should be

measured at time intervals dnrinv ths sisct-Tv!

The Gr(lll) content should be analyzed routinely in the plating

operation especially if the operation is seemed out of control or the

ratio of the cathode to the anode surface area is not in hal anna.

By redox titration, the Gr(lll) content is normally

measured by the difference between the total amount of chromium after

oxidizing the Cr(lll) to Cr(Yl) and the actual amount of Gr(Vl) in th

_ A

Toshimura et si. developed a polarographic method for Cr(lll) deter¬

mination using EDTA as the masking reagent for Cr(Vl) in 2KP, and Li CIO4

was used as supporting electrolyte. f PAR• proposed that, Cr(lll) can

be determined by Differential Pulse Polarography in NaSCK- EOAc buffer

•rvl n+? ftn Vk-rrf r»r Hs+o41 4nfArma4. nn won tH vori.

This part of work is attempted to use the L.S.Y. method for the Cr(lll)

M m M 4 A M AM 4 Ml AA 4 M A M A A Vl

The Cr(lll) was first seperated from Cr(Vl) by the hydrolysis oi

the Cr(lll) in ammoniaeal solution. It was analyzed by three differer

(a) Cathodic process to determine the Cr(lll) in KSCN— EOAc buffer

solution with I at -0.75 to -0.80 V( refer to the reduction of theP

ttt 1 rvfTT. Ficure 3-1)•

Page 38: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

(b) Anodic process to determine the Cr(lll) in ASCK- BOAc buffer

solution with F at -0.75 to —0.70 T( refer to the re-oxidation of theP

Cr(ll) that is the reduction product of the Cr(lll) on the electrode,

-v

(c) Oxidation of the Cr(lll) to Cr(Yl) in HjCL- EaOH medium. The

total amount of Cr(Vl) and the amount of Cr(Vl) before oxidation were

eenerately measured in KH_- RH.C1 buffer solution.

The electrode reversibility of Cr(lll) in ESCK- BOAc buffer eolnticn

was examined, end it can be verified that, the reversibility is dependant

on the concentration of the Cr(lll) and the scan rate. In addition, the

effect of the initial potential on the I measurement was studied.

Proper operating conditions for hydrolysis, heating temperature and

time, pH value of the buffer solution, etc. would be examined.

i 2The interferences of cations t Cr(Vl), Cu, Fe» Ei were evaluated,

A method of eliminating the cationic interferences using gravimetric

seneration was examined as well.

I

5 MJ

( Cathodic'

V (T

-0.5 -0.7 -0.9 -1.1

(Anodic

1 1 1

—1.0 —0.8 —0.6 —O.d

Figure 3-1. Polarograms of cathodic anodic waves of the Cr(lll)

(150 ppm) in KSCN- HOAc buffer eolution.

Page 39: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

3.2. Experimental

Unless other wise specified, all reagents used were of A.R. grades.

(1) Standard Cr(lll) solution, 1 gl

1.0000 g of chromium powder (pure) was dissloved in 25 ml conc. HC1

and diluted to 1000 ml.

(2) KSCN- HOAc buffer solutions

19• 5 g of KSCN was dissolved in 800 ml water, and the pH was adjusted

to 3.20 by addition of conc. HOAc. In addition, the following pH values of

buffer solutions were prepared as well: 1.00, 1.50, 2.50, 4.20, 6.40.

(3) H2O2( 30%)

(4) 1N HC1

45 ml of conc. HC1 was dissolved and diluted to 500 ml.

(5) 1N NaOH

22 g of NaOH was dissolved in water and diluted to 500ml.

(6) 10% NaOH

10 g of NaOH was dissolved in water and diluted to 100 ml.

(7) 1.5% sodium anthranilate solution

3.5 g of anthranilic acid was dissolved in 20 ml 1N NaOH, then diluted

to 200 ml with water.

3.32. Calibration of Cr(lll) and determination of cationic impurities

in plant sample solutions

The Cr(lll) contents in unknown plant sample solutions were calibra¬

ted with conventional redox titration. The results were regarded as the

actual values in the samples determinations. In addition, the contents of

cationic impurities( Cu, Fe, Ni) were determined and their result

were used in the studies of the cationic interferences in the Cr(lll) de-

t ermination.

Page 40: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

2+

The Cu content was measured by iodometric titration after se-

2+ A

perating the Cu from other sample electrolytes

3+

The Fe content was determined by EDTA titration using the Tiron

as indicator( pH« 2.50). 4

2+

The Ni content was measured gravimetrically by dimethylglyoxime

5precipitation.

The results are summarized in the following table t

Samples Cr(lll)

(gi) (gi) (gD (gD

Sample No. 4

Orti-o Ro» x.=;nin_ 10 i.ioio.nf 1.18±0.11 2.020.12

Samule No. 5

(H.K. Productivitv Cpnt-rp 1- ssin.07 0.49-0«02 1.260.12 0.10±0.01

( Note t all listed results are average value of 3 replicated measurements'

3.2.3 Analysis procedure

(1) Determination of the Cr(lll) in KSCN- HOAc buffer solution

(a) A 10 ml aliquot of sample solution was pipetted then diluted to• •

100 ml in a flask. A 20 ml aliquot of the diluted solution was pipetted

into a 250 ml beaker, added 100 ml of water, heated to 60-70°C, then added

cone, NH_ dropwise with constant stirring to pH= 8.0, heated to boilingj

for 2-3 minutes.

(b) The Cr(lll) hydroxide ppt was filtered with filter paper No. 541

( or filter crucible No. 3)» then washed with 100- 150 ml of hot water.

The ppt was redissolved in a small amount of 1 N HC1, and the filter paper

was washed with 50 ml hot water. The filtrate was diluted to 100 ml in

a flask.

(c) A 10 ml aliquot of the solution was pipetted and mixed with

10 ml KSCN- HOAc buffer solution( pH 320) in a 80 ml beaker, heated

to 95- 100°C for 2-5 minutes. Then cooled down to room temperature.

Page 41: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

(d) A 10 ml aliquot of the resulting solution was pipetted into

the electrolysis cell and measured with L»S.V.y— w—— V. rr w A 9 W V»«

L.S.V. instrumental setting conditions t

Cathodie Dronesn: Potential scan: from -0.50 to -350 V.

Recording scan: from -0.50 to -0.90 V.

Full current scale: 50 JA.

Scan rate: 100 mvsec.

Hg drop size s M

N dispersion: 2 min.

Anodic Drocess s Potential scan: from -1.00 to +2.00 V

Recording scan from -1.00 to -0.60 V.

Full current scale: 50 JA.

scan rate: 100 mvsec.

Hg drop size: li

dispersion: 2 min.

(2) Oxidation methoi

(a) By the procedure as mentioned previously in (l) (b), the ppt

was dissolved by 1N HC1 into a 100 ml beaker, adjust the pH to 3-4 by drop-

wise addition of 1N NaOH. Transferred to a 100 ml flask, diluted to the mark.

A 25 ml aliquot was pipetted into a 200 ml beaker, added 5-10 ml

1N RaOH and 5-10 ml O. The excess HO was removed by heating gently

until no more gas evolved( if any ppt was formed, it should be filter¬

ed.), then transferred to a 100 ml flask and diluted to the mark.

(b) The measurement of Cr(Vl) was followed the procedure described i

2.2.3 The content of Cr(lll) was calculated as the difference between

the total amount of Cr(Vl) after oxidation and the amount of Cr(Vl)

r a T A A AV A i Avn

Page 42: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

3 3. Result s and discussion

331. Effect of the residual Cr(Vl) on Cr(lll) determination

It is necessary to give an aocount on the effect of the residual

Cr(Vl) after hydrolysis and seperating process to the Cr(lll) determi¬

nation In addition, the effectiveness of seperating Gr(Hl) from large

amounts of Cr(Vl) in the sample solution by hydrolysis of the Cr(lll)

should h« H+.nd'fod-

(a) Effect of Cr(Vl) on CrO_ determination

To evaluate the effect of Cr(Vl) on Cr(lll) determination, standard

solutions of Cr(lll)f Cr(Vl), and their mixture were examined catho-

dically( c) and anodically( a)• Their wave peak potentials and the

wave forms are compared( Table 3-1 a)

Table 3-1 Effect of Cr(VT) on Cr(lll) determination

Solution (E) (Y) Wave form (c)(Ep)a (V) Wave form (a)

Cr(lll) 50 pin -0.20

-0.74

—063

Broaden

Sharp

Poorly definec

-0.70 Sharj

-072 BroadenCr(Vl) 50 ppa

Cr(lll) 50 ppm

+ Cr(Vl) 50 pn -0.74 Poorly definec -0.7C Sharp

Bventhough both the cat ho die and anodic waves of the Cr(Yl) in

KSCN- HOAc buffer solution are not well defined, but they may affect

the Cr(lll) determination, because the (®p)c value of Cr(Yl) is more

positive than that of the Cr(lll) in cathodic process, and the (Rp)a Is

more negative than that of the Cr(lll) in anodic process

Page 43: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

To evaluate quantitatively the effect of Cr(Vl) on Cr(lll) deter-

mination, standard 40 ppm Cr(lll) solution( corresponding to the

amount of 2.0 gl of Cr(lll) in original plating solution after a

50-fold dilution) was mixed with different amounts of Cr(Vl) solutions

( 20— 20.0 ppm). The Cr(lll) was measured cathodically and anodi—

pal 1 V. Thp rPQIll+ Q o -no nroocnl•?« T. V

Table 5-1 b. Effect of Cr(Vl) on Cr(lll) detenninatior

Cr(lll) Det ection

( nTtn l

CYr( TTT frvnriH

( nrjm

Relative erroi

A J•

10.0

15.0

41. C

41.6

A 0

+2.5

+4.0

+ 1 3

10.C

15.C

20.C

40.6

41.6

45.0

+ 1.5

+4.C

+ 1 2

Po+ V.-V3 -J« -TV.««~„• Cf TTT(

hvrA 'i r rvn AAftPO•{ T T T}( TIT

( Pal ml a+. i nns are referred to 3.3.8.

Prom Table 3-1 b, it can be seen that, to keep relative erroi

less than +5-0 per cent, the residual Cr(Vl) should be kept below

Page 44: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

(b) Effectlveneao of aeperating Cr(lll) froo Cr(Vl) by

hydrolysis

It should be noted thatf during hydrolysis of the Cr(lll) the

Cr(Vl) is easily occluded into the gelatinous Cr(lll) hydroxide ppt.

For this reason, the Cr(VT) present in the hydrolysis solution should

be kept as low as possible and the Cr(lll) hydroxide should be washed

thoroughly to remove the Cr(Vl).

The residual Cr(Vl) occluded in the Cr(lll) hydroxide was measured

in NHj— NHCl buffer solution and the results are concluded in Table

3-2.

Table 5-2. Effectiveness of seperatine Crflll) from CrfVl}

Qotti a a a! Ar»Residual Cr(Vl) detected

(ppm

150

250

2.0C

5.oc

ft nr

5.45

11.(

iO C

( See section 2.5.4.

From the above results it can be concluded that, residual Cr(VI

content is no more than 15 ppm in the Cr(lll) hydroxide precipitatioi

Therefore, the residual Cr(Vl) should not lead to unacceptable error

in the Cr(lll) determination.

5.5.2. The dH effect on Crflll} hvdrolvsi«

In order to ensure the completeness of hydrolysis of the Cr(lll)

and to prevent the dissolution of the Cr(lll) hydroxide, the pH value

the ammonlacal solution should be controlled carefully.

Page 45: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

To test the pH effect of the hydrolysis solution on Cr(lll) deter-

mi nation, electrolyte containing known amount of Cr(lll)( 500 gl)

was hydrolyzed in different pH values of the ammoniacal solution. The

Cr(lll) was measured by cathodic and anodic processes The results are

shown in Table 3-3 and illustrated graphically in Figure 5-2

Table 3-3« The pH effect of hydrolysis

pH of hydrolysis

solut ion

Cathodic

CrflTI found Reooverv

(gi) (90

Anodic

Crflll found Recovery

(gD (56)

7.00

7.50

7.80

8.50

9.00

10.0

1.18

4.06

5.04

4.95

4.85

4.47

2.54

23.6

81.2

101

97.0

95.6

89.4

t;n ft

1.63

4.05

4.98

4.83

4.76

3.74

2.64

32.6

81.0

99.6

96.6

95.2

74.8

52.2

Note: 1. The actual concentration of Cr(lll) was 5.00 g

2. Calculation of Cr(lll):

Cathodic: Cr(lll) (l

Anodic s Crflll) (l

1 flee sent ion P. 1

From the above investigation it can be concluded that, the proper

pH range of the hydrolysis should be 7.80 to 8.50. And a pH b 8.0 valu

was used for all determinations.

Page 46: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

100

60

20

i I6 8 10 12

( Anodic)

Actual Cr(lll) content: 5.00 gl

Cr(lll) recovery

( % )

pH

100

60

20

6 8 10 12

PH

Cr(ill) recovery

( % )

( Cathodic)

Actual Cr(lll) content: 5.00 gl

Figure 5-2. Plot of pH of the hydrolysis solution

vs Cr(lll) recovery.

Page 47: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

3.3.3- Effect of acids for dissolving the Cr(lll) hydroxide

9

Different acids used for dissolving the Cr(lll) hydroxide would

affect the reproducibility of 1 measurement and the wave form. In order

to study the effect of acid used for dissloving the Cr(lll) hydroxide,

1N HC1, 1N HNO, 1N were used seperately to dissolve the Cr(lll)

hydroxide formed by hydrolysis of a sample containing 5.00 gl of Cr(lll).

The Cr(lll) contents were measured by cathodic and anodic processes and

their results were compared with that of the 100 ppm standard Cr(lll)

solution( corresponding to the amount of 5.00 gl Cr(lll) in the plating

solution after 50-folds dilution). The results are shown in Table 3-4

Table 3-4 Effect of acids for dissolving the Crflll} hvdroxidc

cr(in;

Cathodic Anodic Cathodic Anodic

WflVP f OTTO

Cathodic Anodic

St andard

solution

In 1N HC1

In 1N HN03

In 1N H SO

_o n£ _n 7 o 3.92

3.93

3.15

2.95

3.72

3.72

3.10

3.40

Sharj Sharp

Levelled

off

Prom Table 3-4 it can be known that, using 1N Hd or 1N HN0, for5

dissolving the Cr(lll) hydroxide, a sharp wave form can be obtained. Bui

for the 1N HSO, a poorly defined wave is obtained in cathodic process.

Therefore, the 1N HC1 was used in our work, so that the results will

be commensurate with the Cr(lll) standard solution.

Page 48: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

3.34. Effect of heating time and temperature

It has been observed that, when the Cr(lll) was added to the

KSCN— EOAc buffer solution, there was no apparent change at room tem¬

perature, and the detected wave form was poor, I signal was low. However,P

at higher temperatures a pale violet color appeared in the mixing solution,

and resulted in a better wave form and higher Ip signal when the Cr(lll)

is measured at room temperature. This indicates that, a more stable

Cr(lll) SCN complex is slowly formed which is a better electrode active

species. And the mixing solution of the Cr(lll) and KSCN- HOAc buffer

solution should be heated before measured at room temperature by L.S.V.

In order to study the effect of heating time and temperature for the

mixing Cr(lll) with buffer solution, standard 80 ppm Cr(lll) solutions

were measured cathodically and anodically. The results of 1 measurements

under different conditions are shown in Figure 5-5.

From the results of investigation, it can be known that, if the

mixing solution is kept in room temperature( 25°C), sharp wave form is

hardly observed even the solution is stand for a long time( up to 60

min.). When the solution is heated to 60°C before measurement, a 30

minutes time is necessary to obtain a sharp wave form for cathodic process.

If the heating temperature is at 80°C, and heating for 5-10 minutes, a

o

sharp wave form and reproducible I values will be obtained. At 95- 100 C,

2-5 minutes of heating time will be enough to get sharp wave form and

reproducible I measurements after cooling down for both cathodic andP

anodic processes.

It is therefore to use the heating temperature near the boiling

point( 95- 100°C) and 2-5 minutes as the operating conditions for

the mixing Cr(lll) and buffer solution before measured by L.S.Y. in

all other analysis works.

Page 49: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

5.5-

5.0.

2.5-

Ip( -MA) ( Cathodic process)

95-100°C

80°C

60 C

25 C

Time( min.)i i i 1 1 1—

0 10 20 30 40 50 60

Anodic process)Ip()

5-5

3.0_

2.5

i 1 1 1 1 r—

0 10 20 30 40 50 60

Time( min,)

25°C

95-ioo°cC

80 °C

Figure 3-3. Plot of the I values vs the heating times at different

t emperatures.

Page 50: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

3 35 Effect of pH value of KSCN— HOAc buffer solution

In order to evaluate the effect of pH value of KSCN- HOAc buffer

solution used in the determination of Cr(lll), 50 ppm standard Cr(lll)

solution was measured at different pH values of buffer solution by

cathodic and anodic processes. The results of determinations are

shown in Table 5—5 and illustrated srranhi r.al 1 v in mrp A.

Table 5—5 Effect of pH value of buffer solution

pH of

hn f f pr on! nh nr

Wave form

Cat h o di r

A J•

1.00

1.50

2.50

5.20

5.60

4.20

5.20

6.40

1.00

1.50

2.50

5.20

5.60

4.20

5.20

6.40

-0.84

-0.85

-0.78

n in

-0.74

-0.7 3

_n 7

1.50

1.70

2.06

2.56

2.52

2.50

1.80

Broaden

Sharp

Undefined

1.15

1.1C

1.12

1,96

1.91

1.88

1.50

Undefined

Broaden

Sharp

Page 51: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

(- Cathodic)

Ip( JUA)2.5

1.5

' i

1 3 5 7

pH

( Anodic)

( JAA)2.0

1.0

I 1 I

13 5 7

pH

Figure 5-4 Plot of pH of the buffer solution vs.

I.P

Page 52: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

The above results clearly show thatf the pH of the buffer solution

has a great effect on the C r ( l l l ) determination both by the cathodic

and anodic methods*

In low pH values, the electrode process would be more irreversible

and the value tends to be more negative*

When the pH value becomes higher, the electrode reaction is essential­

ly reversible and the I values are getting maximum. However, i f the pH

value is too high ( more than 4*0 ), the C r ( l l l ) would be partially

hydrolyzed and the 1 value would be lower* A further increase in the pH

value ( pH value beyonds 6.0 ), the C r ( l l l ) would be precipitated, thus

the wave form detected becomes undefined.

It can be concluded that, in order to have a well defined wave form

and reproducible 1 measurements in either cathodic or anodic process,

the pH of the buffer solution should be in the range of 3.0 - 4.0. The

pH sr 3*20 was used for a ll measurements.

3. 3.6. Reversib ility of electrode reaction of the C r (l l l ) in KSCN -

HOAc buffer solution

By investigating the revers ib ility of an electrode reaction, the

mechanism of the electrode process can be more accurately derived, and

the information about the operating conditions by which the cathodic

and anodic processes can be successfully employed can be provided.

For the C r ( l l l ) determination, we prefer to use the anodic process

rather than the cathodic process because the anodic process can to le ­

rate the interference o f cationic impurities in chromium plating bath.

Therefore, the revers ib ility of electrode reaction was studied in more

details.

The revers ib ility of the electrode reaction was investigated from

two aspects s

- 46 -

Page 53: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

(1) The effect of the Cr(lll) concentration and the scan rate on E:P

(2) The effect of the Cr(lll) concentration and the scan rate on I•P

(l) The effect of the Cr(lll) concentration and the scan rate on EP

It can be observed that, both the Cr(lll) concentration and the scan

rate will affect the E of the electrode reaction, so their effects on

the Ep and the reversibility were studied

Different concentrations of the Cr(lll) solutions( 50, 100, 200,

300, and 500 ppm) were measured at diffrent scan rates( 20, 50, 100, 200,

and 500 mvsec) independently by the cathodic and anodic processes.

The difference in absolute values of the peak potential of the catho¬

dic process (E) from the peak potential of the anodic process (E) mea-P c p'a

sured by different concentrations of the Cr(lll) solutions at 100 mvsec-

(AEp)c-al Were comPared with the theoretical value, i.e. O.O56V, for the

reversible one electron transfer electrode reaction. The results are

summarized in Table 3-6, and shown in Figure 3-5.

Table 3—6. Effect of the Cr(lll) concentration on E

Deviation of

from 0.05

Cr(lll)

5C

10C

150

200

300

400

Ron

0.0

0.0'

O.Ol

0.1(

0.1:

0.1,

0.11

+0.00,

+0.01

+0.02

+0.04

+0.06

+0.08

+0.09

It can also be shown that, for other scan rates, the variation of th

are similar to that for scan rate= 100 mvse

On the other hand, the effect of the scan rate or

Page 54: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

different concentrations of the Cr(lll) is shown in Table V7.

Table 5—7• Effect of the scan rate on E

Cr(lll)

I rnw»

Scan rate

onr

20

50

100

200

Ron

20

50

100

200

(;nn

2C

5C

10C

20C

-0.75

-0.75

-0.76

-0.77

-n.7Q

-0.79

-0.79

-0.80

-0.81

-0.81

-0.8C

-0.8C

-0.81

-0.82

_n q 1

-0.71

—0.71

-0.70

—0.7 0

-0.68

-0.70

-0.70

-0.70

-0.70

-n££

-0.70

-0.70

-0.69

-0.69

n CC

0.01

0.04

0.06

0.07

0.11

0.09

0.09

0.10

0.11

0.18

0.1(

0.1(

o.u

0.1:

0.16

Prom the above results, several points can be shown that

1 At- 1 i I Till vvT riT i a o v rl m n v rv, values ai

close to 0.056 V, this indicates the electrode reaction is one electro;

At hirii Cr(lll) concentration, the (E_) values are shifted to more

nocrp+'lvo vh-M a +V»o (IT. trnTnia.ei otp nhlftpri to mnrp nnnltlvp. thun th

i J 4 U A A. 1 A A.A« A A n A+ 4 n «n A o 4 Atrl ot

Prom Pi cure -8. it can be shown that, when tfr vali

reaches 0.112 V( twice amounts of the theoretical value i.e. O.056 V)

+ Vio rtnT»t»oonftnHnff fVrTTT POT! r»PTrt Tflt. 1 nn 1 R AllOut ?S0 DTTD. whprPUDon t hi

Page 55: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

Figure 3-5* Relatioship of I ( A E )I P c-a

and the C r ( l l l ) concentration

C r ( l l l ) (ppm)

Figure 3-6. The ratio of ( I ) / ( I ) VS the C r ( l l l )p c p aconcentration.

- 49 -

Page 56: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

electrode reaction is arbitrarily regarded as irreversible.

(2) Prom Table 5-7, it can be noted that, if the scan rate ia 200 mvaec

or leas both (Ep)c and (E are nearly constants( within± 0.01 V

deviation) for a given concentration of the Cr(lll), and indicate the

rpVPTfllhl'l nf+« a« 4- 2

(2) The effect of the Cr(lll) concentration and the scan rate on I.P

In order to evaluate the effects of the Cr(lll) concentration and

the scan rate on 1, different concentrations of the Cr(lll) solutions

( 40, 120, 150, 200, 500, 500 ppn) were measured at various scan rates

( 20, 50, 100, 200, 500 mvsec), both in cathodic and anodic processes

The (I) (l) ratios from different Cr(lll) concentrations at

the scan rate of 100 mvsec are listed in Table 3—8, shown in

T P ft_ Vf f pot. r f t. h D Cn( TT T onn% ont i rr rir% T

fW TTT m

40

12(

15C

20C

30C

c;nr

1.06

1.02

1.0C

0.96

0.9 3

0.87

i are also observed for other seacL

T A 4 4 A W V M A O O M A A Vs A A A A VA VI A A and

from various concentrations of the Cr(lll) arp c p- a

4 1 1 ii A« A A4 4 M m «.V1 A 7

Page 57: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

Table —Q- Effer»+. nf+ vo oo t

Cr(lll) Scan rate

40

20

50

100

200

0.14

9!

99

0.15

. 0.17

0.13

99

99

0.14

1.08

tt

tt

1.07

1.21

120

20

50.

100

200

0.41

0.42

0.42

ft

0.41

0.40

0.39

0.98

1.00

1.02

1.08

1.13

200

20

50

100

200

500

0.56

0.57

If

ft

fK SA

0.59

If

0.58

0.57

n Rq

0.95

0.97

0.98

1.00

1.06

Hereby several points can be derived:

(1) Except for high Cr(lll) concentrations( more than 300 ppra) and

high scan rates( more than 200 mvsec), the (I) (i) ratios ares p'c pa

close to unity( 1.0- 0.1) and the (I), (I) valuesp c pa

are nearly constants( within i 0.1 deviation) for a given concentration

2of the Cr(lll). Therefore, the electrode reaction is reversible.

(2) The (p)c rati°s vary with the Cr(lll) concentration. At

low Cr(lll) concentrations( less than 150 ppm), the (p)c ratio:

are larger, but when Cr(lll) concentrations increase, the (p)c (p)a

ratios decrease.

(3) Eventhough the (i) and (I) are both increaseed on high scanp c Pa

rates, but it can be observed that, the (I) is increased much fasterp c

than the (I), hence the (I) (I) ratio is getting larger onx p'a p c pa 0 00

Page 58: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

higher scan rates.

(3) Summary

After evaluating the effects of concentration and the scan rate

upon E and I, conclusion can be drawn hereby.

The electrode reaction of the Cr(lll) in KSCN- HOAc buffer solu¬

tion is reversible one electron transfer process whwnever the bulk

concentration of the Cr(lll) is less than 250ppm and the scan rate is

lower than 200 mvsec.

It may be proposed that, the reversibility of an electrode reaction

depends upon the concentration of the electroactive species, this is

a second criterion for reversibility of an electrode reaction.

3.3.7. Effect of the initial scan potential

The behavior of the Cr(lll) electrode reduction-oxidation has been

reported. Zanello and Raspi reported the electro-oxidation of the

Cr(lll) in alkaline hydroxide solution at a platinum electrode, the pro¬

cess involves the oxidation of the Cr(lll) to the chromate. Walsh and

Earley studied the reduction potentials for the Cr(lll) with SCN

and other ligands in 0.1 M NaClO. Two waves were observed when the com¬

plexes were reduced at the DME, the first wave( E1= -0.81V, vs S.C.E.)

was ascribed as the reduction of the Cr(lll) complex to the Cr(ll) species

the second wave was complicated by the reduction of hydrogen ion.

In order to investigate the stepwise reaction of the Cr(ll) in KSCN-

HOAc buffer solution, the effect of the initial scan potential on the I

measurement was studied. By using different values of the initial scan

potential, the I values for different concentrations of the Cr(lll)

were measured. The results are shown in Table 3-10, Figure 3-7 a and

Figure 3-7 b.

Page 59: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

10

8'

6

4

r.

50 100 1 50 200

Cr(ill), ppm

A

, B

'v,

Scan Potential

A: -1.10 to +1.90

B: -1.00 to +2.00

C: -0.90 to +2.10

V: -0.80 to +2.20

Figure 3-7 a. The relation of the concentration of the Cr(lll) and

the I at different scan potentials.P

Table 5-10. The effect of the initial potential

Pa+DT+1 O

Csthodi i

-0.20 to -3.2C

-0.40 to -3.4C

-0.50 to -3.5C

-0.60 to -3.6c

Anodic

-0.80 to +2.2C

-0.90 to +2.1C

-1.00 to +2.0C

-1.10 to +1.9C

Cr(lII'

50 ppn

Cr(lll)100 mm

Cr(lll)

150 ppn

Cr(lll)

200 ppm

Blank

2.05 4.4C 6.1C 7.75 0.50

1.95

2.25

2.70

3.20

5.5C

3.95

4.4

5.2

5.10

5.70

6.30

7.85

6.55

7.15

8.00

9.50

0.65

0.70

Page 60: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

, 1 1 1o _1 n —0_ R -0.6 -0.4

_i o _i' n —oTft —b.6 —0J4

_i o _i n _n fi -0.6 -0.1

-1'2 -0.8 -6.6

Scan Potential( V

Figure 3-7. Effect of the initial scan potential on the 3

in onMin nrnrp c 5( f!r TTT• 100 Tinm

Page 61: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

It can be seen that, in the cathodic process, for the reduction of

the Cr(lll)(= -0.70 V), the I values are independant on the

For the anodic process, the I values increase with the more nega-

tive values of the initial scan potential. This points out that the

formation of the Cr(ll) species should be occurred on the electrode sur¬

face prior to the bxidation reaction. The more negative value of the

initial scan potential was applied, the more favarable for the Cr(ll)

formed and the more Cr(ll) species oxidized to the Cr(lll). It can also

be noted that, with the more concentrated Cr(lll) solution, the variation

of the I values with the changing initial scan potentials is more ob-

vious. If the initial scan potentialwas kept at more negative than -1.20 V

the wave form became poorly defined or overloading should be occurred.

Furthermore, the effect of the holding time for the initial scan

potential prior to the initiation of the scanning was studied for the

anodic process. After connecting the electrolysis cell with the polaro-

graphic analyzer, the initial scan potential was held at a definite

period before starting the potential sweep. Two potential scannings were

studied and standard Cr(lll) 100 ppm solution was used. The results are

shown in Table 3-11

Page 62: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

Table 3-11. The effect of the holding time of the initial

RCfin rkntntl «1

Scan potential

( rr

Initial potential

Held time( min.}

-i.nn -tn 4- 9.no

0.2

0.5

1.0

2.0

4.0

4.40

4.40

4.40

4.42

4.45

4.60

0.:

0.5.

1.C

2.C

4.C

o r

5.1

5.1'

5.2!

5.2!

5.4

5.2!

One may expect that, the longer the holding time for the

negative initial potential, the more Cr(ll) should be formed and thus

more Cr(lll) is detected. However, experimental results did not show

such a trend. It may suggest that, the reduction of the Cr(lll) to Cr(ll

on the electrode surface is very fast and can be completed in a few

seconds or less.

Therefore, in the anodic process, the Cr(lll) would be instan¬

taneously reduced to Cr(ll) on the electrode surface and then reoxidized

by the one electron transfer reaction in the anodic process.

In our works, the initial scan potential for the anodic process

var kTrt at annroximately 10 aec. and the scanning was in the range from

-1. nn v 4. p.on l

Page 63: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

3»3.8. Calibration rurvp

In normal chromium plating bath, the Cr(lll) content is in the

range of 1.0 to 10.0 gl. After 50 folds dilution, the resulting solu¬

tions will have the Cr(lll) concentration in the range of 20- 200 ppm.

In order to get a wider application, calibrations for the Cr(lll)

in the range of 200- 2000 ppm and 0.5- 20 ppm are also set up for the

cathodic process and 200- 600 ppm for anodic process.

(l) Cathodic process

(a) Cr(lll) 20- 200 ppm

Cr(lll)

(ppm) Blank 20 50 75 100 150 200

( If A0.50 2.00 2.91 3.70 4.45 6.04 7.70

The calibration curve is shown in Figure 3-8.

(b) Crflll) 200- 2000 num

Cr(lll]

( pom B1 an k 200 300 500 800 1000 1500 2000

IP

( if A0.50 7.80 11.3 17.1 27.1 31.4 48.7 64.1

It can be observed that, when the Cr(lll) is higher than 1000 ppm,

the E value of the reduction wave is shifted to more negative, and

the peak becomes broaden.

(c) Cr(lll) 0.5- 20 ppm

Cr(lII

I ppm Blank 0.5 1.0 2.0 5.0 8.0 10 20

IP

f II A0.50 0.88 0.95 1.00 1.15 1.35 1.46 2.0(

If the Cr(lll) is less than 1.0 ppm, the peak current is hardl

H of. or»h oH.

Page 64: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

Jp (A)8

7

6

5

4

3

2

1

0 50 100 150 20 0 250

Cr(lll) (ppm)

Figure 3—8. Calibration curve( cathodic determination,

Cr(lll) 20- 200 ppm).

Page 65: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

(d) The relation of 1 and the Cr(lll) concentration, the correlation

coefficient, R, have been calculated by least squares fitting and the

results are summarized in Table 3-12.

Table 3-12. Calibration of the Cr(lll) in cathodic process

Cr(lll)

(ppm)

Calibration of Crflll) (ppm) Correlation

coefficient R

0.5- 20

20- 200

200- 2000

!p= 0.898+ 0.0557 c

I= 1.334+ 0.0316 C

I= 1.650+ 0.0312 C

0.998

0.999

0.998

The calibration curve in 20- 200 ppm range is used for the Cr(lll)

determination in actual sample solutions.

(2) Anodic process

fa) Crflll) 20- 200 com

Cr(lll)

B1 an k 20 30 50 100 120 150 200

1.50 1.90 2.60 4.50 5.11 6.22 7.85

It can be observed that, if the Crflll) is less than 20 ppm,

I measurement is not reliable.r

(rv' ttt onn_ nrm

CrClIl]

f unm B1 ank 200 500 550 600 70C

n.6( 7.85 18.6 21.1 23.1

It can be seen that, if the Crflll) is higher than 600 ppn

+. Y o onnH i r» javo no a r Vi o rrm o 0 ofM n orl-

Page 66: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

8

7

6

5'

4-

3-

2-

1-

0 50 100 150 200 250

Figure 5-9. Calibration curve( Anodic process,

Cr(lll) 20- 200 ppn).

Page 67: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

(o) Table 5-15 above the linear equation of the Cr(lll) (ppm) to

I CttA), and the corresponding correlation coefficient, E.

Table 5-15 Calibration of the Cr(lll) for anodic proceee

Cr(lll) Correlation of the Cr(lll) (ppn)

to I (jwi)

Correlation

coefficient B

20- 200

ocn- dnr

I« 0.644 0.0555 C

I« 0.297 4- 0.0578 C

0.9997

0.9990

Calibration curve for Cr(lll) in 20- 200 ppn range is ehovn in

TH ifiir®_ Q_

Detection limit was determined by a diluted Cr(lll) standard solutior

Beplicated 1 measurements( 11 valueB) were made using the 10 pp

Cr(lll) standard solution in KSCN- HOAc buffer solution. The results

ma€ T «i A e

F

OA mvk A j) 4+

Detection limi

Page 68: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

( Arr»rf 1 r rvrro.o

The 20 ppn Cr(lll) standard solution was Measured anodically

in KSCN— HOAc buffer solution and 11 replicated measurements were

made. Thp TPRnl tn DTP an fnl 1 nuc.

iveraffft of thp T vpI-iipr.

..10. Analysis of sample solutions

The Cr(lll) contents in the sample solutions were determined

seperately by cathodic process, oxidation process and anodic process

For plant sample solutions No. (4) and No. (5)» the Cr(lll) contents

were also analyzed by the redox titration, and the results were used

as the actual values in the determinations( referred to 3.32.)•

Page 69: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

Table 3-14 Analysis of samoles bv eathodio nrocenn

Sample n-r( ttt Relative error

(D

(2)

(3;

(4

f c

2.50

5.00

8.00

3-50

1 .rs

2.39- 0.26

5.14 -0.15

8.10 i 0.43

2Not detected

1- A1 1 n ir

-4.4

+2.8

+1.3

+ 17

1 Average of 4 replicated measurements which were calculated By:

( Note: the samples have been diluted by a 50-fold dilution.)

2 Cathodic method was not applicable for those sample solutions

p+containing high concentration of cat ionic impurities especially the Ni.

For detail discussion see 3«3«H

f• j i.

Table 3-15. Analysis of samples by oxidation procesj

Sample Found Crflll Relative erroi

(1:

(2:

(3:

(4:

2.50

5.00

8.00

3.50

2.57- 0.0

5.19- 0.0

7.76~ 0.2

3.40- 0.1

1.S7 i n.n

+2.1

+3.I

-5.

-2.!

+1.

Average of 4 replicated measurements and the Cr(lll) conte

Page 70: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

Where

I P1

I P2

IP

I value measured before oxidation of C r ( l l l ) .P

1 value measured after oxidation of C r ( l l l ) , with

four folds dilution.

0.170 + 0.162 C ( 2.3*4* ) was used in calculations.

1 A xp - 4 I pJ - I p,

( 3) Anodic process

Table 3-16 . Analysis o f samples by anodic process.

Sample

No.

Actual

C r (lll ) (g/1)

Pound C r (ljl )

(g/D *

Relative error

( -% )

( 1) 2.50 2.61 1 0.15 +4*4

( 2) 5*00 4.95 - 0.08 -1.0

( 5) 8.00 8.09 - 0.41 +1.1

( 4) 3*50 3.62 i 0.20 +3*4

( 5) 1*55 1.55 - 0.07 -1*3

* Average of 4 replicated measurements and the C r (ll l ) content

was calculated by :I - 0.844

C r (l l l ) (g/ l) = ----E-------------- x 0.050.0355

Equation of 1 = 0.844 + 0.0355 C ( 3.3.8. ) was used to calculate

the C r (ll l ) (ppm).

Figure 3-10 shows the polarograms for the C r (ll l ) determination

in cathodic and anodic processes.

It can be observed that, when the cathodic process was applied in

real plant samples which contained cationic impurities, the wave forms

would became undefined or poorly defined, so measurements of the C r (ll l )

would not be accurate. On the other hand, for anodic and oxidation pro­

cesses, satisfactory results were obtained with relative errors less than

-5 *0 per cent. Therefore, anodic and oxidation processes can be applied

for the determination of the C r (l l l ) in chromium plating baths.

- 64 -

Page 71: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

Sflrrml p No.

( s}

( 4)

( 3

( 2

(1)

Blank

I I I

_n R -D.7 -0

Cathodic process

Sample

No.

( 5)

7( 4)

( 3)

( 2)

( 1)

Blank

-1.0 -0.8 -0.6

Anodic process

Figure 3-10. Polarograms for determination of the Cr(Vl) in

the samples.

Page 72: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

3• 3• 11• Cat i oni n i n+. gt fot en

(1) Wave characteristics of Cu2+, Fe4, Ni2+ in KSCN- HOAc buffer

The normal cat ionic impurities in chromium plating bath are Cu2+,

5+ o|

Pe, Ni. Their interferences on the Cr(lll) determination should be

To evaluate the cat ionic interference, 50 ppm standard solutions oi

2+ 5+ 2+Cu, Fe and Ni were examined in KSCN- HOAc buffer solution se-

perately• Their wave characteristics are compared with that of 50 ppm

standard Cr(lll) solution( Table U17)

Table 3-1T Wave characteristics of cations ir

KSCN —HOAn Vm TTor cnl n+ i nn

Tnne Cathodie

A J•

EP

(V)

Wave form

T T Tc

a

—0 .76

-0.7 9

Sharp

tt

Cu2 c

a

-0.54

—0.47

_n. 9Q

Shar

It

If

c —0.1

-o.9

_n a i

Poorly define

Broaden

-0.2C

-0.66

-O.Af

Poorly defim

Sharp

flrnaH o-n

From Table 3-17 it can be seen that:

In cathodic process, most E values for Cu2+, Fe, Ni2+ ions ar

less negative than that of the Cr(lll), so they may cause interferer

for Cr(lll) determination. Especially the Ni whose E( -0.68 V) iD

Page 73: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

close to Ep for the Cr(lll)( -0.76 V), and its interference should

be more prominent.

For anodic process, most values for the interfering cations

axe less negative than that of the Cr(lll), hence their interference

would not be so serious.

To look into the interference quantitatively for different cationic

impurities, solutions containing known amount of the Cr(lll)( 2.00 gl)

to which different contents of cations were added and measured seperately

by oxidation process, cathodic process and anodic DrrmPSR.

In common plating bath analysis, a relative error within i 5.0 per

cent should be acceptable. The tolerable limit of interference is defined

as the amount of cationic impurity( gl) to cause a relative error

of- 5-0 per cent in the recovery of the Cr(lll)( gl).

9,(2) Interference of Cu

The results of determinating electrolyte containing known content

°f Cr(lll)( 2.00 gl) with addition of the different amounts of Cu+

are concluded in Table 3-18 graphically shown in Figure 3-11.

Table 3-18. Interference of Cu+ on Cr(lll) determination

Sample solution Oxidation process Cathodic process Anodic process

2.0C

0.50

1.00

2.00

5.00

10.0

2.06

2.05

2.02

1.81

1.66

2.00

2.08

2.13

3.29

1.98

2.02

2.07

2.10

Cr(lll) (gl)

Page 74: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

Where

p2 are peak currents for the Cr(Vl) detected before

and after oxidation of the Cr(lll) respectively.

Negative deviation occurs when a large amount of Cu2+ was added.

This is ascribed to the interference of the Cu2+ on Cr(Vl) determination

( 2.37) and causing the increase of I• On the other hand, afterP1

2+

the Cu was removed as Cu(0H)2 in the oxidizing alkaline solution( HgO+

NaOH), the I_ for the Cr(Vl) would no longer be affected by the Cu2+ in-

terference and the A I value became lower.

(3) Interference of Fe

The interference of Fe on the Cr(lll) determination is summarized

in Table 3-19 and in Figure 3-12

Table 3-19 Interference of Fe on Cr(lll) determination

Sample s olution Oxidation process Cathodic process Anodic process

2.0C

1.00

2.00

5.00

10.0

20.0

2.02

1-95

1.92

1.88

1.70

1.98

2.02

1.82

1.78

1.98

2.05

2.08

2.50

3.62

(4) Interference of N.

For the interference of the N; on the determination of the Cr(lll),

it is summarized in Table 3-20 and in Figure 3-13.

Page 75: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

Table 3-20. Interference of Ni2+ on C r (l l l ) determination.

Sample

C r (l l l )

(«/D

solution

Ni^added

(g/ l)

Oxidation process Cathodic process Anodic process

C r (l l l ) found

(g/ l)C r (l l l ) found

(g/D

C r (l l l ) found

(g/D *

0.02 1.95

0.04 2.01

0.10 2.20

0.16 3.20

2,00 0.50 2.02

1.00 2.05

2.00 1.96 2.08

4.00 - 2.10

5.00 1.92 2.16

10.0 1.88 not detected

15.0 1.80

20.0 1.74

* When the Ni content was getting higher, anodic wave would be

levelled o ff, and may not be measurable.

( 5) The tolerable concentrations of the interfering cations in the

C r (l l l ) determination.

From Figure 3-11 to 3-13 the highest tolerable concentrations

of cationic impurities are summarized in Table 3— 21*

- 69 -

Page 76: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

fr 4 o4 4 Ar A AO O

2.1

L—r

A A a H i A nrroo(

Cr(lll) found (f:]

t

2

r—j•

-- 2

-- 1

2+

Figure 3-11. Interference of Cu

Actual Cr(lll) content: 2.00 g1.2++

The concentration limit of the Cu causing— 5»(

_ . . » •

Page 77: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

Oxidation process

Cr(lll) found (gl)

2.5

1.5

0.5

2.1

1.9

510

15 25

Cathodic process

Cr(lll) found (gl)

Anodic process

Cr(lll) found (gl)5.5

2.5

1.5-

0

4.0

10 20

2.1

1.9

5.5

2.5

1.5

o6.2

10 20

2.1

1.9

figure 5-12. Interference of Fe

Actual Cr(lll) content: 2.00 gl.

The concentration limit of the Fe causing- 5.0%

error on Cr(lll) determination.

Page 78: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

Chr i dat 1 on Tvrn oa«a q

Cr(lll) found (gl)

2.5

1.5

0.5

2.1

1.9

5

I

o r?

15 25

Po+ V» n o vw A

Cr(lll) found (gl)

3-5

2.5

1 R

0

t

o.nft 0.1i

- 2.1

1 Q

Anodic process

-j Cr(lll) found (gl)

2.75

2.25

02 4

1 2.1

- 1.9D.

Figure -1. Interference of 1

Actual Cr(lll) content: 2.00 g,

- The limits of Ni concentration causing- 5.0 9

error on Cr(lll) determination.

Page 79: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

Table 3-21. Concentration limits of lnt.prfor)nr na+inno

na+.Vin-ir nrrtonoo Anodic process Oxidation process

1.fi

r no

R o

6.2

4.0

3.0

1 0

9.0

( Note: the nominal Cr(lll) content is 2.00 gl'

It can be seen that, the anodic method is less subjected to the

cationic interferences as compared to the cathodic method. Application

of the cathodic process will be limited, particularly when the nickel

2+plating is done prior to the chromium plating and the drag in of Ni

f%.nnt.flrrn'nfl+. inn i a sprinn.q.

In common chromium plating bath, the impurity contents of Cu,

Fe and Ni are in the range of 1-2 gl( sometimes the Fe content

could be higher in hard chromium plating bath due to the anodic disso¬

lution of the ferrous base articles.), in such case the anodic process

i o nrof orahl o in anal vc i a unrVo_

The oxidation process can tolerate larger amounts of cationic

interferences( except for Cu), but more time is needed in carrying

out the analysis. Therefore, the oxidation process should be applied onl

when the eontente of Fe or Ni are unusually hieh.

At present the anodic polarographic method is still not popular,

yet it should be complementary to the cathodic process with respect to

some aspects such as• sensitivity, tolerance of impurities, etc.

(6) Elimination of cationic interference

Generally, oxidation or anodic process can be directly appliec

to the determination of Cr(lll) in chromium plating bath. When the

Page 80: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

impurity contents are exceptionally high, they should be removed before

the analysis.

Elimination of interfering cations can be accomplished by gravi¬

metric seperation using anthranilic acid( 0—aminobenzoic acid) as

precipitating agent.

The anthranilic acid can be used to precipitate different cations

selectively by controlling the pH and temperature carefully. Table 3— 22.

lists the operating conditions for selective precipitation by anthra-

nilic acid.

Table 3-22 Operating conditions for selective precipitation

by anthranilic acid.

Cations ppt color ppt conditions

Grey green to

violet

pH 5.0, room temp,

or pH 2-3, 60°C

Blue green

Brown

Grey green

pH 2.0-30, room temp.

pH 3.0-3.5, room temp.

pH 35-40, room temp.

A procedure for seperating the interferences of Cu, Fe and

Ni Waa DPrfnrmpH fnll nuc•

(l) As the procedure described in 32 3 (a) the Cr(lll) hydro¬

xide ppt was dissolved in 1N HC1, then the pH was adjusted to 2.0 by

dropwise addition of 10% NaOH. Then 1.5% of sodium anthranilate was

added dropwise with stirring until the precipitate just appeared, and

excess 5 ®1 waa added. The pH would be 30- 40. Let the solution

stand for 20- 30 minutes.

Page 81: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

(2) The precipitate of the interfering cations was filtered and

washed with 1: 30 of 1.5% sodium anthranilate solution and 100 ml

of whter.

The filtrate was combined and heated to 60- 70°C, then added

with 1N NaOH to a pH= 8.0, and boiled the solution for 2—3 minutes.

(3) Precipitate of the Cr(lll) hydroxide was dissolved in a small

amount of 1N HCi. The color of the solution free from other cationic

impurities should be pale violet.

The solution obtained was diluted to 100ml in a flask and proceed

by usual procedure for the Crflll} determination.

To examine the effectiveness of seperating interfering cations,

Cu+, Fe, and Ni+( 10.0 gl of each) were added seperately to the

sample solutions containing 2.00 gl of Cr(lll). The Cr(lll) contents

in the sample solutions were determined after removal of the interfering

cations. The results are listed in Table 3-25.

It can be concluded that, simultaneous removal of interfering

cations Cu, Pe, Ni using anthranilic acid as the precipitating

agent could be applied to analyze the sample solutions with serious

interference. But one trouble in applying the seperating method for the

cathodic process is that, the removal of the Ni is hardly complete,

because the ppt pH for the Ni and that for the Cr(lll) are not much

different and the cathodic process is not tolerated to the presence of Ni

Therefore, in analysis works when the Ni interference is serious,

oxidation method should be applied. If the Ni is less than A.0 gl,

anodic process may also be used in the Cr(lll) determinations.

Page 82: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

Table 5-23. Determination of Cr(lll) after removal of

cat ionic interference.

O N 1Cathodic process Anodic process

Cr(lll) found Error Cr(lll) found Error

2.00

Cu2+

10.0

Fe54

10.0

XTNi

1.91

1.91

-4.5

-4.5

1.92

2.02

-4.0

+ 1.0

hi 4- «-»Vv 11.91 -4-5

Q r~ n+ t rr~ Oxidation nrnp.psR

n. 2+Ou

Fe

u. 2Ni

+0.

1 .9 -4.0

1.95 -0.?

The seperating method described is not successful for the Nj

2+removal, because even a trace amount of Ni left would caus

interference to the Cr(lll) determination in cathodic proces

Page 83: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

5 4 Conclusion

The Cr(IIl) content in chromium plating solution can be determined

either by the direct measurement in KSCN- HOAc buffer solution or

the measurement in NH- NHCl buffer solution after oxidizing the

(Yrl TTTl +N rv.1 TTT

Seperation of the Cr(III) from large amounts of the CrCVT} can

be accomplished by hydrolysis of the Cr(IIIJ in ammoniacal solution

( pH 8.0). After dissloving the CrHIJ hydroxide ppt in 1N HClf

it is mixed with KSCN- HOAc buffer solution( pH» 320) and heated

to about 95 C, then cooled down. The Cr(III) content can be determined

directly by L.S.V. either in the cathodic or anodic nrocess-

When the potential scan rate is less then 200 mvsec and the

Cr(III) concentration is less than 250 ppm, the electrode reaction is

For anodic process, the 1 measurements are dependant on the appliec

initial scan potential and it can be proposed that, the Cr(II) is formed

on the electrode surface and then re-oxidized to the Cr(III).

For cathodic process, linearity of 1 to the CrIII) concentration

has been obtained in the range of 1- 1000 ppm, and the detection limit

is around 2.1 ppm. In anodic process, good linearity of I to the Cr(III)

concentration is in the range of 20- 600 ppm, and the detection limit

arnnnfl R.O nrm-

The anodic process for determination of the CrIII) is preferabl

to cathodic process due to the former process to be more tolerable

to cationic interference.

Page 84: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

Another method for the CrIII) determination is based on oxidizing

the CrIII) to Cr(Vi; and to measure the Cr(YI) in- NH.C1 buffer

solution, The Cr(III) content is derived by the difference of the total

Cr(YI) and the residual Gr(VT) Trrior in nyi datl nn.

By calibration curve method, the anodic or oxidation process

can be used in analysis of the Cr(III) content. And the relative

errors are less than 50 percent. The cathodic process is subjected

to high cationic interference( especially the Mi), it can not

be used for the real Tlant fiamrle anal vein-

The cationic interferences of the Cu, Fe 9 and Ni can be

successfully removed by selective precipitation using anthranilic

acid as the precipitating agent.

Page 85: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

PART IV DETERMINATION OP SULPHATE IONS

41 Introduction

Traditional gravimetric determination of SO in chromium plating

bath is by BaSO precipitation, but the analysis is rather time con¬

suming. An alternative method is done by using a calibrated centrifuge

tube to estimate the volume of BaSO.

5Wild mentioned a d.c. polarographic analysis by measuring

2+the reduction of the Pb concentration after forming the insoluble

TTVvOn

12=

Bobrowski and Owoc reported that, the SO was precipitated in

2+

an excess Fb solution, the FbSO was seperated and redissolved. The2+

Pb was measured by oscillopolarographic method.

Here we attempt to determine the SO by L.S.V. method. After

reducing the Cr(Vl), the S0~ is precipitated by a known and excess

2+ 2+amount of Fb, the excess Pb can be measured by L.S.V. This will

eliminate the filtration and redissolution steps. In addition, the

L.S.V. method should be more accurate and rapid as compared to the

D.C. Polarography.

2+The measurement of the Fb by L.S.V. was performed in NaOAc

HOAc buffer solution with a well defined wave peak( E= -0.40 toP

-0.50 V).

Determination conditions such as s the time and temperature for

2+precipitation, concentration of the standard Pb solution, pH effect

of the buffer solution, etc. will be examined. In addition, the cations

and CI interferences will also be studied.

Page 86: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

4-2. Experimental

A O 1„

All reagents used were of A.R. or G.R. grades

(1) Reduction reagent

Formic acid, minimum 88%.

(2) 0.02 M Pb standard solution

3-3157 g of Pb(N0) was dissolved in 500 ml water, added 10 ml

HNO solution( 1: 4) and diluted to 500 ml.

0.1 M Pb standard solution

16.5600 g of Pb(N0) was dissolved in 200 ml water, added 20 ml

HNO solution( 1: 4) and diluted to 500 ml.

(5) NaOAc- HOAc buffer solutions

55 g of NaOAc was dissolved in 850 ml water, added glacial acetic

acid to adjust the pH to 4.58, then diluted to 1000 ml.

In addition, the following pH values of NaOAc- HOAc buffer solutions

were prepared as well: 2.05, 2.60, 5.52, 5-12, 5-80, 7-50.

(4) Standard S0 solution, 10 gl

14.7958 g of NaSO was dissolved in 200 ml water and diluted to

500 ml.

(5) Standard CI solution, 2 gl

1.6484 g of NaCl was dissolved in 200 ml water and diluted to

500 ml.

(6) AgNO soluti on, 5 gl

5 g of AgNO was dissolved in water and diluted to 100 ml.

Page 87: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

42.2. Calibration of SO, in sample solutions

The correct amounts of SO in chromium plating solutions were

standardized by gravimetric method, The results are as follows.

Samnl e snl ut i nn .9D Found SO

2.50

3.00

2.45 t 0.08

2.93t 0.05

5.68 t 0.09

3.10 i 0.07PI on+ aormVl o

Plant sample

2.50 i 0.06

Average of 3 replicated measurements.

4.2.3. Procedure of SO determination by L.S.V.

(1) A 10 ml sample solution was pipetted into a 400 ml beaker, added

30 ml water and 15- 20 ml 88% formic acid, heated to boiling for 30

minutes until the color of the solution became dark green, after cool¬

ing, diluted to 100 ml in a volumetric flask.

(2) 5 nil of the reduced solution was pipetted into a centrifuge tube,

added 500 ml standard 0.02 M Fb solution. Let stand for 25 minutes,

and then centrifuged for 1-2 minutes.

(3) 2 ml of the centrifugate was pipetted into an electrolysis cell,

added 8.00 ml NaOAc— HOAc buffer solution f pH= 4.8

(4) L.S.V. experimental conditions s

Scanning potential setting: from -O.30 V to -1.80 V

Recording potential scanning: from -0.30 V to —0.50

Scan rate 2 100 mvsec

Current full scale 2 IOuA

Hg drop size 2 M

N di sue-rni on 2 2 minutes

Page 88: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

4.3 Results and discussion

Effort of t.Vi Crr( TTT

2+

The Pb was measured in the presence of large amounts of Cr(lll) as

a result from the Cr(Vl) reduction in formic acid Polarogram( Figure

4—1) shows that, the E value for the Pb+( —0,43) is much more

positive than that for the Cr(lll)( -1.10 V), hence, the Cr(lll) should

2+not have interference on the Pb determination. To prove this respect,

following test was performed. Solution with a fixed amount of SO~4

( 2.50 gl) but different concentrations of CrO were used. After

2+

reducing the CrO and precipitating the PbSO, the excess Pb was mea-

SlITPlI. TVip rp.qnl+. Q of T valnpq crp 1 i a+. or} -in T'a'hl o A 1

Table 4—1 Effect of the Cr(lll) on determination of PI

100 150 250 300

17.5 17.3 17.5 17.4

Average value of 3 replicated measurements

From the above results, it can be known that, the I values areP

practically unchanged( variation within i 0.1 J1A). Therefore, the

presence of the Cr(lll) would not pose a serious effect on the deter-

2+mination of Pb in NaOAc- HOAc buffer solution. In order to minimize

the uncertainty of the matrix and to get more accurate results in

analysis of real sample solutions, the standard addition method should b

used( see 4- 37-)•

Page 89: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

Cr( III)— Cr(ll'

-T r I-1

9x s..

Page 90: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

452. Effect of PbSO precipitation time

The PbSO is precipitated in the excess Pb solution, the time

for completion of precipitation under specified condition should be

The sample containing a known amount of( 2.50 gl) was used

in the test. After reducing the Cr(Vl) and adding the standard Pb+

solution, let the solution stand for different times ranging from 5 to

120 monutes, then centrifuged 1-2 minutes. The Pb was measured in the

rO a -r r% a j+ -i -Pt-» 4-

The I values at different precipitation times are shown in Table

Table 4-2. Effect of PbSO. rut time

out time( min 1

5

1(

2(

3C

4C

6C

ipr

18.;

17.£

17.5

17.4

17.3

16.4

15.9

It can be shown that, I becomes essentially constant if the precipi¬

tation time is kept within 20-50 minutes.

In order to get reproducible I measurements, the ppt time should b

defined. In the following determinations, the time for ppt was 25 minute

and the rentrifuxral time wan 1— 2 minutes.

Page 91: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

4.3.3. Effect of concentration of Pb+

In order to determine the appropriate concentration of Pb+ needed

to precipitate S0=, different standard Pb+ solutions( 0,001, 0.005,

0.01, 0.02, 0.04t 0.05, 0.10 M) were used for precipitation. For each

case, the calibration was set up. Calibration curves are shown

%

in Figure 4-2.

The linear equations and the correlation coefficients R from least

squares fitting method are summarized in Table 4-3- However, for the

cases of 0.001 M, 0.005 M 0.10 M of Fb+ standard solutions, their

calibrations are rather poor, and would have no practical use.

Table 4-3 Comparison of slopes and R of calibrations

Slope b R

0.01

0.02

0.04

0.05

9.90- 0.00598 C

20.4- 0.0115 C

37.3- 0.00936 C

49.7- 0.00844 G

0.00598

0.0115

0.00936

0.00844

0.980

0.998

0.991

0.997

C: Concentration of SO solutions used were 10, 100, 200

250, 500, 800 and 1000 ppm, respectively.

2+From Table 4-3 it can be known that, the 0.02 M Fb gives the

highest slope and correlation coefficient. Therefore, the standard

2+0.02 M Pb solution was used in all actual determinations of

the SO~ in the following works.

Page 92: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

6C

5

4(

3(

2(

f I i»i

~ rr Crr onn 1 no

SO. (dtn nr

r r

f OP

Pb2+( M)

Figure 4-2. Calibration curves of I( -A) vs SO (ppn) at

different concentrations of Pb standard solutio:

Page 93: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

4..4. FITT prt r f rTtn nit oH An 4

Temperature will affect the edibility of PbSCL( K= 1.70 x 10~8).4 sp

To investigate the temperature effect, following test was carried out.

The solutions with fixed amount of( 2.50 gl) were used. After

reducing the CrO,, solutions were individually mixed with 0.02 M Pb2+f

centrifuged for 1 minute, then the centrifuge tubes were placed seperate-

ly in water baths at different temperatures, settled down for 20- 25

minutes, then immediately measured the 1 values for Fb in an

aliquot of the clear solution.

The results of 1 for different temperatures of the ppt are shown

TnhT p A—A- Pffprt nf tpmnprat.nrp nf rm+ Ph.O

Tpmnpro+nro( C

1

r

2(

y

3!

5(

16.5

17. C

17.2

17.3

17.4

17.9

1Q. 1

Experimental data demonstrate that, with increasing temperature,

I values would be increased because of more FbSO, dissolved. However.

if the temperature is in the range of 20- 35 °C, the variation of wa

minimum( i 0.1 julA). This would be an advantage because one can carry

out the analysis at room temperature without taking too much precaution

to control thp tpmnpratnrp in rrreeioit at ion.

Page 94: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

435 Effect of pH of the NaOAc— HOAc buffer solution

To study the pH effect of the buffer solution, similar expe¬

riments were done for a given sample solution at different pH values

of buffer solutions. The results of E and I values are tabulated in

Table 4-5» emd shown in Figure 4-3, 4-4.

Table 4 5» Effect of pH of the NaOAc— HOAc buffer solution.

tH of the buffeT sol lit, i nn

2.05

2.60

3.52

4.38

5.12

5.80

7.40

0.414

0.416

0.423

0.446

0.462

0.479

0.481

12.2

14.9

17.4

17.5

17.3

17.0

16.1

I

The Ep for Pb is shifted to more negative value when the pH of

the buffer solution is higher, it is presumably due to the fact that

the Pb2+ becomes hydrolyzed at high pH value and the electrode reaction

would be more irreversible and a more negative E is resulted.

On the other hand, the increasing pH of the buffer solution causes

the I measurements come up to higher values. Whereas, I values tend

to be reduced when the pH value is increased to such extent that, the

2+

Fb would be rapidly hydrolyzed( pH more than 5-0), consequently,

I values are getting lower.

From Figure 4-3 it can be known that, to obtain a better reproducibi¬

lity measurement of I, the pH of the buffer solution should be kept in

range of 35-45 ancL a pH= 438 of the buffer solution was used in all

M A A M

Page 95: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

20

15

10

0

I III

2 4 6 8 10

PH

Figure 4~3« Effect of pH on I.P

0.48

0.46

0.44

0.42

0.40

1 i' i f i

2 4 6 8 10

PH

Figure 4-4• Effect of pH on E.P

Page 96: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

4.3 .6. Calibration curve

Following procedures were used to set up the calibration curve for

the SO “ determination. A 10 ml o f standard solution containing 250 g/l

CrOj was reduced by formic acid, then transferred to a 100 ml o f flask.

Known amount o f SO ~ was added and diluted to the mark. The SO “ contents4 4

were measured seperately according to the procedure described in 4*2.5.

( l ) 10 - 1000 ppm SO ” ( corresponding to 0.1 - 10 g/l o f SO in the

orig inal p lating solution. )

sv(ppm) Blank 10 50 100 200 250 500 700 1000

IP

(^ A ) 20.0 19*8 19.7 19*2 18.2 17.5 14.7 12.4 9.40

By applying the least squares f it t in g , the I ( JWA) was related to

the concentration o f SO "" ( C, ppm ) by :

20.4 - 0.0115 C ( C : 50 - 1000 ppm )

And the correlation c o e ffic ien t, R = 0.998

The calibration curve is shown in Figure 4-5. It is adopted as the

working curve fo r determinations o f real sample solutions.

From Figure 4-5, i t can be observed that, when the SO - is less

than 50 ppm, the lin ea r ity is lo s t.

(2 ) 1000 - 3500 ppm SO ( corresponding to 10 - 35 g/l of SO in

orig inal plating solution. )

To look into the usefulness o f this method for higher SO concen­

trations, the SO - in the range o f 1000 - 3500 ppm was also calibrated.

- 90

Page 97: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

91

20

18

16

14

12

10

8

SO=4 (ppm)

0 200 400 600 800 1000

Figure 4-5. Calibration curve for SO 4-( 10- 1000 ppm)

Page 98: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

12

8

4

1000 1500 2000 2500 5000 3500

S04 (ppm)

Figure4-6. Calibration curve for SO( 1Q00- 5500 ppm).

1000 1250 1500 2000 2500 5000 5500

9.60 8.40 6.80 4.10 1.40 0.60 0.50

The calibration curve obtained is shown in Figure 4-6. It can be

seen that, if the concentration of SO is higher than about 2500 ppm

deviation of linearity for the curve is observed. The levelling off

of the curve is presumably due to insufficient Fb in the resulting

mixture and the SO= is in excess.

By least squares fitting, the I( 0U) is related to C (ppm) by:

And the correlation coefficient. R= O.Q7c

Page 99: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

437• Analysis of sample solutions

The SO contents in the sample solutions prepared( 2.2,1.) were

determined independently by calibration curve method and standard

addition method.

(1) Determination of SO, by calibration curve method

Table 4-6. Determination of SO, by calibration method.

Sample

No.

(1)

(2)

(3)

(4)

(5)

2.50•

5.00

5.50

5.10 1

2.SO 1

2.62 t 0.24

2.78± 0.25

5.04 t 0.45

5.50± 0.54

2.71- 0.52

+4.8

-7.5

-8.4

+6.5

+8.4

Relative erxor

1 Values obtained by gravimetric determination( 42.2.)•

2 Average value from 4 replicated measurements.

(2) Determination of SO, bv standard addition method

The determination procedure was according to the processes des¬

cribed in 4-2.3 In addition, the sample was measured after 1.00 gl

of standard solution was added to the mixing solution of the sample

2+with the standard 0.02 M Pb solution. The results are shown in Table

4-7.

Page 100: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

Table 4-7• Determination of SO. by standard addition method

Sample Relative error

(D

(2)

(3)

(4)

2.50

5.00

5.50

5.10

o qr

2.60- 0.06

5.06 i 0.10

5.29- 0.15

5.16 i 0.12

P.41 i 0.07

+4.0

+ 2.0'

-5.8

+ 1.9

-5.6

A ir V n A 4 Y A Tl 1 A A 4 A4 m A A nil Y» Am AAf A

In standard addition method, the calculation of SO. contenl' 4

The concentration of the Pb+ in the NaOAc- HOAc buffer

: Peak current measured for the sample with addition of

1.00 gl SO~ standard solution.

: Peak current measured for the sample without addition of

1.00 gl S0- standard solution.

: Concentration of the S0~ standard solution added( 1.00 gl)

: Volume of sample solution( 10 ml).

: Volume of SO~ standard solution added( 1.00 ml).

Since the sample solution was mixed with NaOAc- HOAc buffer solutio

bv volume in the ratio of 1: 4. therefore, the amount of residual

Page 101: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

2+

concentration of the Pb in the mixing solution is equal to 5 times

of detected. In addition, the amount of SO. was proportional to

2+the reduced amount of the Pb; and an equal volumes of sample solution

2+were mixed with 0.02 M Pb, therefore:

(3) Comparison of the calibration curve and standard addition

By standard addition method, higher accuracy could be obtained

in the determination of SO~

In preparing the calibration curve, the solutions with a fixed

amount of Cr0_( 250 gl) were used. However, the Cr0_ content in the

actual sample solutions may be from 200 to 400 gl, the matrix is certain¬

ly varied from sample to sample. The relative errors obtained were rather

high( up to i 8.5 °o).

With standard addition, the uncertainty of the matrix interference

would be minimized. The relative errors were within i 5«0 per cent. This is

satisfactory in routine analysis for chromium plating solutions.

Polarograms for two sample solutions by standard addition method

are shown in Figure 4-7•

Page 102: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

(b)

II II

-0.5 -0.4 -0.5 -0.5 -0.4 -0.

Figure 4-7. Polarograms for sample solutions b;

standard addition method,

(a) Sample No. 1, SO: 2.50 gl

( b) Samnle No. 4. SO.: 5.10 gl

Page 103: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

il.. ft. T nf a a A

(l) Cat ionic interference

To study the interferences of cationic impurities on the SO~

determination, cations were individually added to the sample solutions

and their reduction E values and wave forms in NaOAc- HOAc buffer

solution are listed in Tahl p -ft.

Table 4-8. Reduction behaviour of cations in NaOAc- HOAc

nY» o A1 n't

r at

_n 1

-1.M

-1. PR

Wave form

Sharx

Sharp

Broaden

Undefined

Shari

Among the common cationic impurities( Cu+, Fe~, Ni+) in

chromium plating solution, Cu+ would mostly interfere the measurement

of the Pb+, thus lead to an error to the determination of SO it is

because the E value of Cu+ is close to and more positive than that of

the Pb+ in NaOAc- HOAc buffer solution.

For a quantitative study of the effect of Cu on SO det erminat io;

different amounts of Cu+ were added to the sample solution containing

the same amounts of SO~~( 2.50 gl) and CrO,( 250 gl). Then the

contents of the SO~ were determined. The results are shown in Table

Page 104: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

Table 4—9. Interfprpnrp r»f

: on

Relative error

5.00

10.0

15.0

20.0

xn r

1.98

1.92

1.05

1.81

1.65

-1.0

-4.0

-7.5

-9.5

-1 7

Prom Table 4-9 it can be known that, for the relative error within

+ 2+

- 5.0 per cent, the Cu content should not be higher than 10 gl. In real

2+

situation, the Cu impurity in the chromium plating bath is generally

only about a few grams per liter, so it should not pose a problem to

the determination of SO~ in the normally operated chromium platiner bath.

It should also be noted that, the negative deviation at high Cu2+

2+may be explained by the fact that, the E value for Cu is more positive

P21

than that of Fb, and the I value for the Pb is inversely proportional

to the SO~ ront ent«

2+Whenever a high amount of Cu present, it may be removed by preci-

2+

pitating the Cu in alkaline solution before reducing the CrO in the5

2+sample solution. The procedure for Cu removal may be carried out as

f a! 1 AUo

A 10 ml of sample solution was pipetted into a 500 ml beaker,

30 ml water and 10 ml 2N NaOH were added( pH= 12- 14). Heated to boiling

for 2—3 minutes. The cupric hydroxide ppt was filtered and washed

vf+. h AH_ An ml ft f Vift+ w 0+ or

Then 5- 10 ml of 1 1 5 HOAc solution was added to the filtrate

( pH= 1- 2), and 15- 20 ml of 88% formic acid was added, boiled

for 10 minutes.

Page 105: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

Warmed and evaporated the solution to a final volume about 80 ml.

After cooling, transferred the solution into a 100 ml flask and diluted

to the mark. The SO. was determined subseauentlv.

r

To verify the effectiveness of the above mentioned method for Cu

removal, sample solution containing known amount of SO-( 2.00 gl)

2+and large amounts of Cu( 15 gl) was examined. The results of

S0~ determinations before and after removal of Cu?+ are shown in

Table 4-10. Elimination of Cu+ interference

Q am T1 orl 1? found Relative error

2.00

0

1 s

15

1.92

1.85

2+before Cu removal

2. 0C

2+after Cu removal

-4.0

-7.5

+ 2.5

Average of 2 replicated measurements by calibration curve and

standard addition method.

(5) Anionic interference

Both CI and N0_ are harmful anionic impurities in chromium plating5

bath, their presence will deteriorate the plating process and degrade

the quality of coating, so their contents must be kept in low levels( ge¬

nerally less than 0.5 gl)• The CI may cause interference in the S0~

measurement because the CI will form a sparingly soluble PbCl9( K=c. S P

- , „

To investigate the magnitude of the interfering CI, diffrent

amounts of CI were added to sample solutions with a fixed amount of

SO~( 2.00 gl). The SO~~ contents were measured as usual.

Page 106: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

The results are listed in Tahle 4~11 and shown in Figure 4'-8

Table 4-11. Interference of Cl~

CI added SO. found Relative error

2.00

0.50

1.00

2 00

5.00

5.00

1-95

2.08

2.28

2.38

2.52

-2

+4.0

+ 1 4

+ 1 9

+2 6

2.6

2.4

2.2

2.1

2.0

1.1

2 4 6

Outline the concentration limit 0f Cl~ causing

+ 5-0% error on the determination of SO~.

Figure 4-8. Interference of CI on the determination of SO.,

Page 107: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

From Figure 4—8 it can be seen that, when the CI content is more

than 1.10 gl, it would cause an error which exceeds +50 per cent limit.

Normally, the CI impurity in the chromium plating bath is below 0.5 gl,

therefore, it would not have a serious effect to the determination

r on=

When the CI content was found untolerably high, it can be removed%

by the following oronedure-

A 10 ml of sample solution was pipetted into a 300 ml beaker, 30 ml

water and 15— 20 ml of 88% formic acid were added. The solution was

heated to boiling for 2-3 minutes. To this warm solution 5 ml of 5% AgKO

and 5 ml 1N HN0_ were added.

The AgCl ppt was filtered by a filtering crucible( No. 3). The

filtrate was transferred into a 100 ml flask and diluted to the mark.

r Vl O QO woo+ Vi on m o Q ai i oo uoncl

The AgCl ppt could be dried and weighed for measuring the Cl~ content

if it is needed.

The effectiveness of the above method for CI removal has been tested

To the sample solutions containing the same amounts of S0~( 2.00 gl)

and CrO( 250 gl), 2.00 gl of Cl~ was added. The contents

were measured before and after the removal of Cl~. In addition, the Cl~

contents in real sample solutions( 2.2.1.) were also determined.

The results are summarized in Table 4-12.

Page 108: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

Table 4-12. Elimination of CI interference

SO. found Relative error AgCl found CI found

(gi! (i%) (g) (gi)

o

o no

2.00

2.00

1.95

2.25

1.92

-2.5

4-1

before CI removal

-4.0

after CI removal 0.0762 1.90

3.10( Plant

sample)

2.50( PI ant

n ott» rO r 1

3.20

2.42

+ 3.2

-3.2

0.0004

trace

0.21

trace

Average of 2 replicated measurements bv calibration curve method

and standard addition method.

From Table 4-12, it can be shown that, the interference of Cl~~

can be eliminated effectively by the above mentioned procedure. It can

also be known that, in those normally operated chromium plating baths

the CI contents are at low levels.

Page 109: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

4.4- Conclusion

An analysis method-for SO measurement in common chromium plating

bath has been studied.

After reducing the Cr(Vl) by formic acid, a known and excess amount

of 0.02 M Pb standard solution was added to precipitate the.SO After

centrifuge, an aliquot of the clear solution was mixed with NaOAc- HOAc

buffer solution( pH= 438)• The excess Pb was measured by L.S.V.

technique with the reduction E at -0.40 to -0.50 V.

The standard addition method gave a better accuracy than the cali-

bration curve memethod.

If the amount of Cu is more than 10 gl and CI is more than

1.1 gl both will cause an error exceeding S.O uer cent.

Compared with the classical gravimetric method, the procedure for

S04~ determination by L.S.V. technique is easier and more rapid. Even-

though it is not so accurate and is more subjected to interference, but

it can be satisfactorily applied in the routine analysis works for SO

in the normally operated chromium plating baths.

Page 110: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

104

SUMMARY

Linear sweep voltammetric method has been studied to determine

the Cr(VI), Cr(III), and SO= contents in chromium4 plating baths,

The Cr(VI) content can be easily and directly measured by a

distinct reduction wave in NH- NH C1 buffer solution. The error3 4

is below 2.5 per cent. Except for large amounts of Cu2+ more than

7.5 g/1), there is no other interference f found.

The Cr(III) content can either be measured directly by the anodic

process- in KSCN- HOAc buffer solution or by oxidazing of the

Cr(III) to Cr(VI) and the Cr(III) content can be deduced indirectly.

The relative error is below 5.0 per cent. In anodic method, except for

large amounts of Ni2+( more than 4.0 g/1 and in theoxidation

method, except for large amounts of Cu 2+( more than 3.0 g/l other

cat ions do not cause interference.

For S04 determination, after precipitating PbSO, the SO4 4

content can be indirectly deduced from measuring the excess Pb2+ 1 eft

in solution with a well defined reduction wave in NaOAc- HOAc buffer

solution. The accuracy by the standard addition method is better than

by the calibration curve method. E3cce for largeamounts of Cl-( more

than 1.1 g/1), a--id Cu2+( more than 10 g/l), other impurities

generally present in the chromium plating bath would not cause

int erference.

Page 111: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

T VI IJi.T! imTSTV

1. A.M. Bond, J. Electro anal. Chem., 118, 581-394, (1981)•

2. A.M. Bond, 1 Modern Polarographic Method in Analytical Chemistry 1,

Marcel Dekker Inc., New York, (1980).

3. 1 The Canning Hand Book in Electroplating•, V. Canning Ltd.,

_

次 遂 社 抄 技 嫂 说 贫 论 ’ 邊 戚 氛 分 先 劣 於 公 说 ’ 姨 乂 , 务

(1Q7 Ar-» w•%«« s

5. P.W. Wild,' Modern Analysis for Electroplating 1, Finishin,

Pl'h! -»Q+ -T f»TV«. T+ TT«rnt-r+ Hill M; T71 T J f +Cn A

6. M.L. Rothstein, W.M. Peterson and H. Siegerman, Plating and Surface

Finishing. June. 78-83. C1981

7. E.S. Pilkington, C. Weeks and A.M. Bond, Anal. Chem., 48, 1665-1 S6

( 1Q7 6

8. Y. Okinaka end C. Wolovodiuk, Plating and Surface Finishing, Sep.

cn_cc

9.• Chromium by Differential Pulse Polarography', Application Briei

C-2, Princeton Applied Research Corporation, Princeton, New Jersej

(1976).

10. C. Yoshimura, X. Miyamoto and Y. Nitta, Kinki Daigaku, Rikogakubu,

Kenk-yu Kokoku (Japan), 14» 39-42, (1979)•

11. Li Nan-qiang, Tian Xiao-ya, Geo Xiao-xia, Ko Hsueh Tung Poa,

• A Special Issue of Mathematics, Physics and Chemistry• (China),

25. 253-255, (1980).

12. A. Bobroweki and T. Owoc, Chem. Anal. (Warsaw, Poland), 26, 125-12

(1981).

Page 112: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a

106

13. A. Bobrowski and B. Barchanska, Chem. Anal. (Poland), 24,

857-860, (1979).

14. Piero Zanello and Giorgio Raspi, Anal. Chim. Acta., 88,

237-243, (1977).

15. J.H. Walsh and J.E. Earley, Inorg. Chem., 3(3), 343-347, (1964).

16. R. Belcher and L. Gordon (General Editors), International

Series of Monographs on Analytical Chemistry, Vol-7,' Gravimetric

Analysis' Part I. Pergamon Press, Oxford, (1963).

17. H.F. Willard, L.L. Merritt, Jr. and J.A. Dean,' Instrumental

Methods of Analysis' 5th edition, D. Van Nostrand Co., New York,

(1974).

Page 113: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a
Page 114: A STUDY OF LINEAR SWEEP VOLTAMMETRIC DETERMINATION … · With cyclic voltammetry, the difference of the peak potential for cathodic process, (Ep)c, and that for anodic process, (Ep)a