accepta water analysis equipment test reagents catalogue
DESCRIPTION
water analysisTRANSCRIPT
-
www.accepta.com 1
-
www.accepta.com 2
Table of Contents
Section Details Page2.0 : General Information 2.1 - 2.4 Accepta Ltd. Details, Services & Ordering. 3
3.0 : Test Method Selector 3.0 Test Parameter / Method Selector 4
4.0 : Test Kits 4.1 Comprehensive Engineer Test Kit (Various versions)4.2 Cooling (Open & Closed) (Standard & Backpack versions)4.3 Closed Water Systems4.4 Chlorination4.5 Boiler Water4.6 Legionella Compliance4.7 Ion Exchange Resin Kit4.8 Manual Silt Density Index
823272829313334
5.0 : Pocket Test Meters 5.1 PH1 - pH/Temperature Meter5.2 EC1 - Condy/TDS/Temp/Salinity Meter5.3 PH100 - Extech pH/Temperature Meter5.4 EC400 - Extech Condy/TDS/Temp/Salinity Meter5.5 EC500 - Extech Combined pH/Condy/TDS/Temp Meter5.6 CL200 - Extech Total Chlorine/Temperature Meter5.7 DO600 - Extech Oxygen/Temperature Meter5.8 RE300 - Extech Redox (ORP) /Temperature Meter5.9 RefG300 - Glycol Refractometer
363738394041424344
6.0 : Professional Test Meters 6.1 Martini MI806 - 4 in 1 pH/Conductivity/TDS/Temp Meter6.2 Eutech pH 11 - pH Meter6.3 Eutech Con 11 - Conductivity Meter6.4 Eutech PC300 - Combined pH/ Condy/TDS/Temp Meter6.5 Jenway 350 - pH Meter6.6 Jenway 470 - Conductivity/TDS/Temp Meter6.7 Jenway 430 - Combined pH/ Conductivity/TDS/Temp Meter6.8 TRANSEC - Conductivity/TDS/Salinity/Temp Meter6.9 TRANSPH - pH/ORP Meter6.10 Borescope
46474849505152535455
7.0 : Drop Test Kits 7.1 Drop Test Kits - Acidity to Zinc 57
8.0 : MiniKits 8.1 MiniKits - Acidity to Zinc 64
9.0 : Comparator Kits 9.1 Comparators - Aluminium to Zinc 67
10.0 : Titration Test System 10.1 Titration Cartridge Tests - Alkalinity to Zinc 74
11.0 : Test Strips 11.1 Test Strips - Ammonia to Zinc with many pH ranges 77
12.0 : Bacteria Testing 12.1 Oxidising Kits (Extech pen)12.2 Non Oxidising Kits (Isothiazolone etc)12.3 ATP Meter & Swabs12.4 Dipslides & NRB/SRB Tube Tests12.5 Incubators12.6 Universal Rapid Legionella Test Kit12.7 Dip-slide Comparator App
80818283848586
13.0 : Corrosion Coupons 13.1 Corrosion Coupons & Test Racks 88
14.0 : PrimeLab Photometer 14.1 PrimeLab (Single or Multi Range) Photometer14.2 Parameters & Reagents
9092
15.0 : Temperature Meters 15.1 - 15.4 Temperature Meters 102
16.0 : Equipment 16.1 Equipment & Components 107
17.0 : Reagents 17.1 Reagents Available 110
-
www.accepta.com 3
Section 2.0
2.2 Laboratory & Support ServicesOur extensive laboratory is able to offer a wide range of services which include:
Full system water analysis Deposit analysis. Corrosion coupon weighing. Determination of product factors for tests like Phosphonate, Nitrite, Molybdate etc. Calibration services on an extensive range of meters.
2.3 Training Services
With many years experience in the water treatment and testing market we are highly trained in advis-ing clients in the use and application of test kits and reagents.
We offer specific training in the use of test kits and the various test options available.
Training can be carried out at our offices/laboratory or can be offered on site and the training canbe tailored to suit your application and requirements and a wide range of subjects can be covered.
2.4 Ordering & TermsOrders can be accepted by the following methods or via our on line portal.Email:Phone:Fax:
General1) Prices, where quoted are exclusive of VAT and delivery.2) For Terms & Conditions of Sale, please refer to our website www.accepta.com/terms-conditions3) Orders received will generally be dispatched within 2 working days.4) A standard delivery charge will apply on all orders.5) The delivery charge for dip slides and/or large test kits and reagent orders will be dependant onthe weight of the goods, please ask for a quote for delivery charges.6) The shelf life of most reagents is greater than 2 years when stored in correct conditions. Pleaserefer to the relevant reagent Safety Data Sheet (SDS) for full details.7) All reagents and kits can be branded with your own logo, please contact us for details.8) Product SDS sheets are available own branded and in many languages, please ask for details.9) All electronic equipment is warranted for 12 months from date of dispatch.10) Probes for pH, Conductivity and other meters are warranted for 6 months from date of dispatch.11) Returns must be notified in writing within 48 hours of receipt of goods.12) All payments are due within 30 days from date of invoice.13) If an invoice remains unpaid after our agreed payment terms, we reserve the right to forward theinvoice to a debt collection agency which will incur an additional 20% surcharge.14) Some images of test kits and components shown in the catalogue may differ from the actualgoods supplied due to stock availability.
2.1 Company Information
AcceptaStatham HouseTalbot RoadManchester M32 0FPUnited Kingdom
Tel : +44 (0) 161 977 2334Fax : +44 (0) 870 135 6389Email : [email protected]
[email protected]+44 (0) 161 877 2334 (e-mail confirmation is required on all telephone orders)+44 (0) 870 135 6389
-
www.accepta.com 4
Section 3.0Test Method SelectorThis useful selector can be used to help identify the test methods we have available for thoseparameters you wish to monitor. Simply select the parameter from the left column, then work acrossto identify which test methods are available. If you need help identifying the most appropriate testmethod for your application please contact Acceptas technical team.
Parameter PrimeLab Drop Test TitrationTests
Comparator Powder PillowEquivalents
TestMeters
Acidity
Active Oxygen
Alkalinity
Aluminium
Amine
Ammonia
Anionic
Boron
Bromine
Carbohydrazide
Chelant
Chloride
Chloramines
Chlorine (TS)
Chlorine Dioxide (TS)
Chlorite
Chromium
COD (LR, MR & HR) Requires heater block
Colour
Conductivity/TDS
Copper (TS)
Cyanuric Acid
DEHA
DBNPA Biocide Kit
Fluoride
Glutaraldehyde Biocide Kit
Glycol (PEG/MEG)
Hardness, Calcium
Hardness, Stop/Go
Hardness, Total (TS)
-
www.accepta.com 5
Parameter PrimeLab Drop Test TitrationTests
Comparator Powder PillowEquivalents
TestMeters
Hydrazine
Hydroquinone
Hydrogen Peroxide(TS)
Iodine
Iron (TS)
Isothiazalone Biocide Kit
Magnesium
Manganese (TS)
Molybdate
Nickel
Nitrate
Nitrite
Ozone
Peracetic Acid (TS)
pH (TS)
Phenol
PHMB
Phosphate
Phosphonate
Polyacrylate
Polyamine
Polymers
Potassium (TS)
QAC
Silver
SodiumHypochlorite
Sulphate
Sulphide
Sulphite
Suspended Solids
Silica
Tannin
Section 3.0Test Method Selector (continued)
-
www.accepta.com 6
Parameter PrimeLab Drop Test TitrationTests
Comparator Powder PillowEquivalents
TestMeters
THPS Biocide Kit
Total Oxidant
Triazole
Turbidity
Zinc
Note: Parameters above with a (TS) denotes where Test Strips are available as an alternative testingmethod.
IF THE PARAMETER YOU ARE INTERESTED IN IS NOT LISTED, PLEASE CONTACT US
Section 3.0Test Method Selector (continued)
-
www.accepta.com 7
Section 4.0Test Kits
-
www.accepta.com 8
Section 4.1Engineers Test Kit with PrimeLab - Eng3005P
Acceptas Eng3005P comprehensive combined test kit is based onphotometer and titration gun chemistries and has been developedas a complete kit, specifically for water treatment professionalsinvolved with the maintenance and management of boiler, coolingwater and closed water systems.Water systems require careful test monitoring to ensure the correctrisk management regimes and water treatment programmes aremaintained at all times. Failure to maintain the correct watersystem conditions can lead to serious problems, including reducedsystem efficiency, the build-up of scale, potentially damagingcorrosive water conditions and increased microbiological activity,which may result in the proliferation of Legionella bacteria and theincreased risks associated with Legionnaires disease.
Contents Range CodeAluminium Case complete with foam & white testing board Eq5026
Waterproof Instructions Eq5084
1 x 20ml Luer Lock Syringe Eq5075
1 x 10ml Luer Lock Syringe Eq5073
1 x 5ml Standard Syringe Eq5077
1 x 1ml Standard Syringe Eq5074
1 x Filter Holder 25mm Eq5044
1 x Filter Paper 25mm GF/C (50pk) Eq5048
1 x Filter Paper 25mm 0.45 micron (50pk) Eq5061
1 x 125ml Plastic Test Bottle Eq5008
4 x 65ml Plastic Test Bottle Eq5018
1 x 5/10ml Tube Eq5015
1 x 250ml Washer Bottle Eq5082
Jenway 430 Combined pH/Conductivity/TDS & Temp Meter Eq5119
PrimeLab Photometer Plp1001
10ml Glass Cells for PrimeLab Photometer (6) Pla1100
Alkalinity P (Titration) 0 - 2000
PA1 (65ml) Dt1001
Alkalinity cartridge (2 x 20ml for all Alkalinities) Crt91002
Alkalinity M (Titration) 0 - 2000
TA4 (65ml) Dt1004
Alkalinity OH (Titration) 0 - 2000
PA3 (65ml) Dt1007
Our comprehensive combined test kit is presented in its own robust Aluminium case and includes allthe equipment and reagents necessary to effectively test and monitor boiler, cooling and closedsystem waters to ensure optimum operating conditions are achieved and maintained.
This test kit includes the following:
-
www.accepta.com 9
Section 4.1Engineers Test Kit with PrimeLab - Eng3005P (continued)
Contents Range Code
Bromine (PrimeLab) 0.05 - 18
PL DPD 1 A (30ml) Plr1102
PL DPD 1 B (30ml) Plr1103
PL DPD Nitrite Powder (20g) Plr1104
Chloride (Titration) 0 - 2000
BC1/CC1 (65ml) Dt1023
Chloride Cartridge (2 x 20ml) Crt91003
Chlorine (PrimeLab) 0.01 - 8
PL DPD 3 C (30ml) (also requires PL DPD 1 A & B) Plr1104
Chlorine Dioxide (PrimeLab) 0.02 - 15
PL DPD Glycine (30ml) (also requires PL DPD 1 A & B) Plr1106
Hardness, Total (Titration) 0 - 2000
TH2 (65ml) Dt1046
TH5 (65ml) Dt1047
Hardness Cartridge HR (2 x 20ml for all Hardness) Crt91007
Hardness, Calcium (Titration) 0 - 2000
CH2 (65ml) Dt1043
CH1P (20g) Dt1050
Hardness, Stop/Go (Drop Test) 0 - 4
Hardness Stop/Go Solution (65ml) Dt1129
Iron Soluble & Total (PrimeLab) 0 - 30
PL Iron HR 1 (65ml) Plr1119
PL Iron HR 2 (30ml) Plr1120
Molybdate HR (PrimeLab) 5 - 200 (with dilution 50 - 2000)
PL Molybdate 1 (65ml) Plr1139
Nitrite (PrimeLab) 0 - 3000
PL Nitrite HR 2 (65ml) Dt1072
Phosphate HR (PrimeLab) 0 - 100
PL Phosphate HR 1 (65ml) Plr1145
PL Phosphate HR 2 (65ml) Plr1146
Phosphonate (PrimeLab) 0 - 20
PL Phosphonate 1 (20g) Plr1169
PL Phosphonate 2 (20g) Plr1170
PL Phosphonate 3 (65ml) Plr1171
PL Phosphonate 4 (20g) Plr1172
-
www.accepta.com 10
Section 4.1Engineers Test Kit with PrimeLab - Eng3005P (continued)
Contents Range CodeSulphite (PrimeLab) 5 - 50
PL Sulphite HR 1 Tablets (50pk) Plr1203
PL Sulphite HR 2 Tablets (50pk) Plr1204
Tannin (PrimeLab) 0 - 300
PL Tannin 1 (65ml) Plr1176
PL Tannin 2 (30ml) Plr1177
Calibration Solutions for Jenway 430 Meter
Standard Conductivity 12.88mS (65ml) Dt1154
pH 4 (65ml) Dt1119
pH 7 (65ml) Dt1118
pH 10 (65ml) Dt1161
20% Acetic Acid (65ml) Dt1103
Refractometer (PEG & MEG) 0 - 60% RefG300
1 x Digital Titrator (For Titration complete with bracket) Eq5022
6 x Titrator Accessory Pack Eq5023
-
www.accepta.com 11
Section 4.1Engineers Test Kit with PrimeLab - Eng3005P/DT
Contents Range CodeAluminium Case complete with foam & white testing board Eq5026
Waterproof Instructions Eq5084
1 x 20ml Luer Lock Syringe Eq5075
1 x 10ml Luer Lock Syringe Eq5073
1 x 5ml Standard Syringe Eq5077
1 x 1ml Standard Syringe Eq5074
1 x Filter Holder 25mm Eq5044
1 x Filter Paper 25mm GF/C (50pk) Eq5048
1 x Filter Paper 25mm 0.45 micron (50pk) Eq5061
1 x 125ml Plastic Test Bottle Eq5008
4 x 65ml Plastic Test Bottle Eq5018
1 x 5/10ml Tube Eq5015
1 x 250ml Washer Bottle Eq5082
Jenway 430 Combined pH/Conductivity/TDS & Temp Meter Eq5119
PrimeLab Photometer Plp1001
10ml Glass Cells for PrimeLab Photometer (6) Pla1100
Alkalinity P (Drop Test) 5 - 2400
PA1 (65ml) Dt1001
PA2/TA2 (65ml) Dt1005
TA3 (65ml) Dt1006
Alkalinity M (Drop Test) 5 - 2400
TA4 (65ml) (also requires PA2/TA2 & TA3) Dt1004
Alkalinity OH (Drop Test) 5 - 2400
PA3 (65ml) (also requires PA1, PA2/TA2 & TA3) Dt1007
Acceptas Eng3005P/DT comprehensive combined test kit isbased on photometer and drop test chemistries and has been de-veloped as a complete kit, specifically for water treatment profes-sionals involved with the maintenance and management of boil-er, cooling water and closed water systems.Water systems require careful test monitoring to ensure the correctrisk management regimes and water treatment programmes aremaintained at all times. Failure to maintain the correct watersystem conditions can lead to serious problems, including reducedsystem efficiency, the build-up of scale, potentially damagingcorrosive water conditions and increased microbiological activity,which may result in the proliferation of Legionella bacteria and theincreased risks associated with Legionnaires disease.
Our comprehensive combined test kit is presented in its own robust Aluminium case and includes allthe equipment and reagents necessary to effectively test and monitor boiler, cooling and closedsystem waters to ensure optimum operating conditions are achieved and maintained.
This test kit includes the following:
-
www.accepta.com 12
Section 4.1Engineers Test Kit with PrimeLab - Eng3005P/DTContents Range Code
Bromine (PrimeLab) 0.05 - 18
PL DPD 1 A (30ml) Plr1102
PL DPD 1 B (30ml) Plr1103
PL DPD Nitrite Powder (20g) Plr1104
Chloride (Drop Test) 0 - 3000
BC1/CC1 (65ml) Dt1023
BC2 (65ml) Dt1026
CC2 (65ml) Dt1024
Chlorine (PrimeLab) 0.01 - 8
PL DPD 3 C (30ml) (also requires PL DPD 1 A & B) Plr1104
Chlorine Dioxide (PrimeLab) 0.02 - 15
PL DPD Glycine (30ml) (also requires PL DPD 1 A & B) Plr1106
Hardness, Total (Drop Test) 5 - 1200
TH2 (65ml) Dt1046
TH3 (65ml) Dt1044
H/TH4/CH3 (65ml) Dt1049
TH5 (65ml) Dt1047
Hardness, Calcium (Drop Test) 5 - 1200
CH2 (65ml) Dt1043
CH1P (20g) (also require TH3 & H/TH4/CH3) Dt1050
Hardness, Stop/Go (Drop Test) 0 - 4
Hardness Stop/Go Solution (65ml) Dt1129
Iron Soluble & Total (PrimeLab) 0 - 30
PL Iron HR 1 (65ml) Plr1119
PL Iron HR 2 (30ml) Plr1120
Molybdate HR (PrimeLab) 5 - 200 (with dilution 50 - 2000)
PL Molybdate 1 (65ml) Plr1139
Nitrite (PrimeLab) 0 - 3000
PL Nitrite HR 2 (65ml) Dt1072
Phosphate HR (PrimeLab) 0 - 100
PL Phosphate HR 1 (65ml) Plr1145
PL Phosphate HR 2 (65ml) Plr1146
Phosphonate (PrimeLab) 0 - 20
PL Phosphonate 1 (20g) Plr1169
PL Phosphonate 2 (20g) Plr1170
PL Phosphonate 3 (65ml) Plr1171
PL Phosphonate 4 (20g) Plr1172
-
www.accepta.com 13
Section 4.1Engineers Test Kit with PrimeLab - Eng3005P/DT
Contents Range CodeSulphite (PrimeLab) 5 - 50
PL Sulphite HR 1 Tablets (50pk) Plr1203
PL Sulphite HR 2 Tablets (50pk) Plr1204
Tannin (PrimeLab) 0 - 300
PL Tannin 1 (65ml) Plr1176
PL Tannin 2 (30ml) Plr1177
Calibration Solutions for Jenway 430 Meter
Standard Conductivity 12.88mS (65ml) Dt1154
pH 4 (65ml) Dt1119
pH 7 (65ml) Dt1118
pH 10 (65ml) Dt1161
20% Acetic Acid (65ml) Dt1103
Refractometer (PEG & MEG) 0 - 60% RefG300
-
www.accepta.com 14
Section 4.1Comprehensive Engineers Test Kit - Eng3005
Contents Range CodeAluminium Case complete with foam & white testing board Eq5026
Waterproof Instructions Eq5084
1 x 20ml Luer Lock Syringe Eq5075
1 x 10ml Luer Lock Syringe Eq5073
1 x 5ml Standard Syringe Eq5077
1 x 1ml Standard Syringe Eq5074
1 x Filter Holder 25mm Eq5044
1 x Filter Paper 25mm GF/C (50pk) Eq5048
1 x Filter Paper 25mm 0.45 micron (50pk) Eq5061
1 x 125ml Plastic Test Bottle Eq5008
4 x 65ml Plastic Test Bottle Eq5018
1 x Plastic Comparator Eq5085
4 x 10ml Comparator Cells Eq5006
1 x 5/10ml Tube Eq5015
1 x 250ml Washer Bottle Eq5082
Jenway 430 Combined pH/Conductivity/TDS & Temp Meter Eq5119
Alkalinity P (Titration) 0 - 2000
PA1 (65ml) Dt1001
Alkalinity Cartridge (2 x 20ml for all Alkalinities) Crt91002
Alkalinity M (Titration) 0 - 2000
TA4 (65ml) Dt1004
Alkalinity OH (Titration) 0 - 2000
PA3 (65ml) (also requires PA1) Dt1007
Acceptas Eng3005 comprehensive combined test kit is based ontitration gun & comparator chemistries and has been developed asa complete kit, specifically for water treatment professionalsinvolved with the maintenance and management of boiler, coolingwater and closed water systems.Water systems require careful test monitoring to ensure the correctrisk management regimes and water treatment programmes aremaintained at all times. Failure to maintain the correct watersystem conditions can lead to serious problems, including reducedsystem efficiency, the build-up of scale, potentially damagingcorrosive water conditions and increased microbiological activity,which may result in the proliferation of Legionella bacteria and theincreased risks associated with Legionnaires disease.
Our comprehensive combined test kit is presented in its own ro-bust Aluminium case and includes all the equipment and reagents necessary to effectively test andmonitor boiler, cooling and closed system waters to ensure optimum operating conditions areachieved and maintained.
-
www.accepta.com 15
Section 4.1Comprehensive Engineers Test Kit - Eng3005 (continued)
Contents Range Code
Bromine (Comparator) 0 - 5
DPD No1 Solution A (30ml) Dt1505
DPD No1 Solution B (30ml) Dt1506
DPD Nitrite Powder (20g) Dt1295
Chloride (Titration) 0 - 2000
BC1/CC1 (65ml) Dt1023
Chloride Cartridge (2 x 20ml) Crt91003
Chlorine (Comparator) 0 - 4
DPD No3 Solution C (30ml) (also requires DPD 1 A & B) Dt1507
Iron LR (Comparator) 0 - 1
Fe5 (65ml) Dt1110
Sulphuric Acid 50% (65ml) Dt1155
Iron HR (Comparator) 0.5 - 10
Fe10 (65ml) (also requires Sulphuric Acid 50%) Dt1247
Hardness, Total (Titration) 0 - 2000
TH2 (65ml) Dt1046
TH5 (65ml) Dt1047
Hardness Cartridge HR (2 x 20ml for all Hardness) Crt91007
Hardness, Calcium (Titration) 0 - 2000
CH2 (65ml) Dt1043
CH1P (20g) Dt1050
Hardness, Stop/Go (Drop Test) 0 - 4
Hardness Stop/Go Solution (65ml) Dt1129
Molybdate VHR (Comparator) 0 - 500
Molybdate HR (65ml) Dt1127
Nitrite (Titration) 0 - 2000
N1 (65ml) (also requires Sulphuric Acid 50%) Dt1072
Nitrite Cartridge (2 x 20ml) Crt91009
Phosphonate (Titration) 0 - 20
P1/2 (65ml) Dt1514
P3 (65ml) Dt1080
P4L (65ml) Dt1291
Phosphonate Cartridge (2 x 20ml) Crt91004
Phosphate HR (Comparator) 0 - 80
PB1 (65ml) Dt1111
PB2 (65ml) Dt1112
-
www.accepta.com 16
Section 4.1Comprehensive Engineers Test Kit - Eng3005 (continued)
Contents Range Code
Sulphite (Titration) 0 - 100
S1 (40g) Dt1092
Sulphite Cartridge (2 x 20ml) Crt91005
Tannin (Drop Test) 0 - 300
TN1 (40g) Dt1094
TN2 (65ml) Dt1095
Calibration Solutions for Jenway 430 Meter
Standard Conductivity 12.88mS (65ml) Dt1154
pH 4 (65ml) Dt1119
pH 7 (65ml) Dt1118
pH 10 (65ml) Dt1161
20% Acetic Acid (65ml) Dt1103
Refractometer (PEG & MEG) 0 - 60% RefG300
1 x Digital Titrator (For Titration complete with bracket) Eq5022
6 x Titrator Accessory Pack Eq5023
-
www.accepta.com 17
Section 4.1Engineers Test Kit with Titration Syringe - Eng3005TS
Contents Range CodeAluminium Case complete with foam & white testing board Eq5026
Waterproof Instructions Eq5084
1 x 20ml Luer Lock Syringe Eq5075
1 x 10ml Luer Lock Syringe Eq5073
1 x 5ml Standard Syringe Eq5077
1 x 1ml Standard Syringe Eq5074
1 x Filter Holder 25mm Eq5044
1 x Filter Paper GF/C (50pk) Eq5048
1 x Filter Paper 25mm 0.45 micron (50pk) Eq5061
1 x 125ml Plastic Test Bottle Eq5008
4 x 65ml Plastic Test Bottle Eq5018
1 x Plastic Comparator Eq5085
4 x 10ml Comparator Cells Eq5006
1 x 5/10ml Tube Eq5015
1 x 250ml Washer Bottle Eq5082
Jenway 430 Combined pH/Conductivity/TDS & Temp Meter Eq5119
Alkalinity P (Titration) 0 - 2000
PA1 (65ml) Dt1001
Alkalinity cartridge (2 x 20ml for all Alkalinities) Crt91002
Alkalinity M (Titration) 0 - 2000
TA4 (65ml) Dt1004
Alkalinity OH (Titration) 0 - 2000
PA3 (65ml) Dt1007
Acceptas Eng3005TS comprehensive combined test kit is basedon titration syringe & comparator chemistries and has been de-veloped as a complete kit, specifically for water treatment profes-sionals involved with the maintenance and management of boiler,cooling water and closed water systems.Water systems require careful test monitoring to ensure the correctrisk management regimes and water treatment programmes aremaintained at all times. Failure to maintain the correct watersystem conditions can lead to serious problems, including reducedsystem efficiency, the build-up of scale, potentially damagingcorrosive water conditions and increased microbiological activity,which may result in the proliferation of Legionella bacteria and theincreased risks associated with Legionnaires disease.
Our comprehensive combined test kit is presented in its own ro-bust Aluminium case and includes all the equipment and reagents necessary to effectively test andmonitor boiler, cooling and closed system waters to ensure optimum operating conditions areachieved and maintained.
This test kit includes the following:
-
www.accepta.com 18
Section 4.1Engineers Test Kit with Titration Syringe - Eng3005TS(continued)Contents Range Code
Bromine (Comparator) 0 - 5
DPD No1 Solution A (30ml) Dt1505
DPD No1 Solution B (30ml) Dt1506
DPD Nitrite Powder (20g) Dt1295
Chloride (Titration) 0 - 2000
BC1/CC1 (65ml) Dt1023
Chloride Cartridge (2 x 20ml) Crt91003
Chlorine (Comparator) 0 - 4
DPD No3 Solution C (30ml) (also requires DPD 1 A & B) Dt1507
Hardness Total (Titration) 0 - 2000
TH2 (65ml) Dt1046
TH5 (65ml) Dt1047
Hardness Cartridge HR (2 x 20ml for all Hardness) Crt91007
Hardness Calcium (Titration) 0 - 2000
CH2 (65ml) Dt1043
CH1P (20g) Dt1050
Hardness Stop/Go (Drop Test) 0 - 4
Hardness Stop/Go Solution (65ml) Dt1129
Iron Soluble & Total LR (Comparator) 0 - 1
FE5 (65ml) Dt1110
Sulphuric Acid 50% (65ml) Dt1155
Molybdate VHR (Comparator) 0 - 500
Molybdate HR (65ml) Dt1127
Nitrite (Titration) 0 - 2000
N1 (65ml) (also requires Sulphuric Acid 50%) Dt1072
Nitrite Cartridge (2 x 20ml) Crt91009
Phosphonate (Titration) 0 - 20
P1/2 (65ml) Dt1514
P3 (65ml) Dt1080
P4L (65ml) Dt1291
Phosphonate cartridge (2 x 20ml) Crt91004
Phosphate HR (Comparator) 0 - 80
PB1 (65ml) Dt11111
PB2 (65ml) Dt1112
-
www.accepta.com 19
Section 4.1Engineers Test Kit with Titration Syringe - Eng3005TS(continued)Contents Range CodeSulphite (Titration) 0 - 100
S1 (40g) Dt1092
Sulphite Cartridge (2 x 20ml) Crt91005
Tannin (Drop test) 10 - 300
TN1 (40g) Dt1094
TN2 (65ml) Dt1095
Calibration Solutions for Jenway 430 Meter
Standard Conductivity 12.88mS (65ml) Dt1154
pH 4 (65ml) Dt1119
pH 7 (65ml) Dt1118
pH 10 (65ml) Dt1161
20% Acetic Acid (65ml) Dt1103
Refractometer (PEG & MEG) 0 - 60% RefG300
6 x Titrator Accessory Pack Eq5023
-
www.accepta.com 20
Section 4.1Engineers Test Kit with Drop Tests - Eng3005DT
Contents Range CodeAluminium Case complete with foam & white testing board Eq5026
Waterproof Instructions Eq5084
1 x 20ml Luer Lock Syringe Eq5075
1 x 10ml Luer Lock Syringe Eq5073
1 x 5ml Standard Syringe Eq5077
1 x 1ml Standard Syringe Eq5074
1 x Filter Holder 25mm Eq5044
1 x Filter Paper GF/C (50pk) Eq5048
1 x Filter Paper 25mm 0.45 micron (50pk) Eq5061
1 x 125ml Plastic Test Bottle Eq5008
4 x 65ml Plastic Test Bottle Eq5018
1 x Plastic Comparator Eq5085
4 x 10ml Comparator Cells Eq5006
1 x 5/10ml Tube Eq5015
1 x 250ml Washer Bottle Eq5082
EC500 Combined pH/Conductivity/TDS/Temp/Salinity Meter EC500
Alkalinity P (Drop Test) 5 - 2400
PA1 (65ml) Dt1001
PA2/TA2 (65ml) Dt1005
TA3 (65ml) Dt1006
Alkalinity M (Drop Test) 5 - 2400
TA4 (65ml) (also requires PA2/TA2 & TA3) Dt1004
Acceptas Eng3005DT comprehensive combined test kit is basedon photometer and titration gun chemistries and has been devel-oped as a complete kit, specifically for water treatment profession-als involved with the maintenance and management of boiler, cool-ing water and closed water systems.Water systems require careful test monitoring to ensure the correctrisk management regimes and water treatment programmes aremaintained at all times. Failure to maintain the correct watersystem conditions can lead to serious problems, including reducedsystem efficiency, the build-up of scale, potentially damagingcorrosive water conditions and increased microbiological activity,which may result in the proliferation of Legionella bacteria and theincreased risks associated with Legionnaires disease.
Our comprehensive combined test kit is presented in its own robust Aluminium case and includes allthe equipment and reagents necessary to effectively test and monitor boiler, cooling and closedsystem waters to ensure optimum operating conditions are achieved and maintained.
This test kit includes the following:
-
www.accepta.com 21
Section 4.1Engineers Test Kit with Drop Tests - Eng3005DTContents Range Code
Alkalinity OH (Drop Test) 5 - 2400
PA3 (65ml) (also requires PA1, PA2/TA2 & TA3) Dt1006
Bromine (Comparator) 0 - 5
DPD No1 Solution A (30ml) Dt1505
DPD No1 Solution B (30ml) Dt1506
DPD Nitrite Powder (20g) Dt1295
Chloride (Drop Test) 0 - 3000
BC1/CC1 (65ml) Dt1023
BC2 (65ml) Dt1026
CC2 (65ml) Dt1024
Chlorine (Comparator) 0 - 4
DPD No3 Solution C (30ml) (also requires DPD 1 A & B) Dt1507
Hardness, Total (Drop Test) 5 - 1200
TH2 (65ml) Dt1046
H/TH4/CH3 (65ml) Dt1049
TH5 (65ml) Dt1047
TH3 (65ml) Dt1044
Hardness, Calcium (Drop Test) 5 - 1200
CH2 (65ml) Dt1043
CH1P (20g) (also requires H/TH4/CH3 & TH3) Dt1050
Hardness, Stop/Go (Drop Test) 0 - 4
Hardness Stop/Go Solution (65ml) Dt1129
Iron LR (Comparator) 0 - 1
FE5 (65ml) Dt1110
Sulphuric Acid 50% (65ml) Dt1155
Iron HR (Comparator) 0.5 - 10
FE10 (65ml) (also requires Sulphuric Acid 50%) Dt1247
Molybdate VHR (Comparator) 0 - 500
Molybdate HR (65ml) Dt1127
Nitrite (Drop Test) 0-2000
N1 (65ml) Dt1072
N2 (65ml) Dt1073
Phosphonate (Drop Test) 0 - 20
P1/2 (65ml) Dt1514
P3 (65ml) Dt1080
P4L (65ml) Dt1291
P5 (65ml) Dt1075
-
www.accepta.com 22
Contents Range CodePhosphate (Comparator) 0-80
PB1 (65ml) Dt1111
PB2 (65ml) Dt1112
Sulphite (Drop Test) 5 - 300
S1 (40g) Dt1092
S2 (65ml) Dt1093
Tannin (Drop Test) 10-300
TN1 (40g) Dt1094
TN2 (65ml) Dt1095
Calibration Solutions for Extech EC500 Meter
Standard Conductivity 1413S (65ml) Dt1109
pH 4 (65ml) Dt1119
pH 7 (65ml) Dt1118
pH 10 (65ml) Dt1161
20% Acetic Acid (65ml) Dt1103
Refractometer (PEG & MEG) 0-60% RefG300
Section 4.1Engineers Test Kit with Drop Tests - Eng3005DT
-
www.accepta.com 23
Section 4.2Cooling (Open & Closed) Water Test Kit - Cool3002
Acceptas Cool3002 comprehensive Cooling (Open & Closed)Water Test Kit is based on drop test and colour comparatorchemistries and has been developed specifically for engineersand water treatment professionals involved with the mainte-nance and management of both open and closed cooling sys-tems to ensure optimum operating conditions.Cooling water systems require careful test monitoring to ensurethe correct water conditions are maintained at all times. Failureto maintain correct treatment programs can lead to seriousproblems including reduced efficiency, the build-up of scale,potentially damaging corrosive water conditions and increasedmicrobiological activity, which may include the growth ofLegionella bacteria. Replacements and components areavailable, please refer to replacement reagent section.
Contents Range Code2019 Maxi Case with Foams Eq5027
Waterproof instructions Eq5084
1 x 20ml Luer Lock Syringe Eq5075
1 x 10ml Luer Lock Syringe Eq5073
1 x Filter Holder 25mm Eq5044
1 x Filter Papers GF/C (50pk) Eq5048
1 x Filter Paper 25mm 0.45 micron (50pk) Eq5061
2 x 65ml Plastic Test Bottle Eq5018
1 x Plastic Comparator Eq5085
2 x 10ml Comparator Cells Eq5006
1 x 5/10ml Tube Eq5015
PH1 - pH/Temperature Meter 0 - 14 pH1
EC1 - Conductivity/TDS/Temperature Meter 0 - 20.00 mS EC1
Alkalinity M (Drop Test) 5 - 2400
TA4 (65ml) Dt1004
PA2/TA2 (65ml) Dt1005
TA3 (65ml) Dt1006
Bromine (Comparator) 0 - 5
DPD No1 Solution A (30ml) Dt1505
DPD No1 Solution B (30ml) Dt1506
DPD Nitrite Powder (20g) Dt1295
Chloride (Drop Test) 0 - 3000
BC1/CC1 (65ml) Dt1023
BC2 (65ml) Dt1026
CC2 (65ml) Dt1024
The Cooling (Open & Closed) Water test kit is presented in itsown robust case and includes all the equipment and reagents necessary to effectively test and moni-tor system waters.
-
www.accepta.com 24
Section 4.2Cooling (Open & Closed) Test Kit - Cool3002 (continued)
Contents Range CodeHardness, Calcium (Drop Test) 5 - 1200
CH2 (65ml) Dt1043
CH1P (20g) (also requires TH3 & H/TH4/CH3) Dt1050
Hardness, Total (Drop Test) 5 - 1200
TH2 (65ml) Dt1046
H/TH4/CH3 (65ml) Dt1049
TH5 (65ml) Dt1047
TH3 (65ml) Dt1044
Iron LR (Comparator) 0 - 1
FE5 (65ml) Dt1110
Sulphuric Acid 50% (65ml) Dt1155
Iron HR (Comparator) 0 - 10
FE10 (65ml) (also requires Sulphuric Acid 50%) Dt1247
Molybdate VHR (Comparator) 0 - 500
Molybdate HR (65ml) Dt1127
Nitrite (Drop Test) 0 2000
N1 (65ml) Dt1072
N2 (65ml) Dt1073
Phosphonate (Drop Test) 0 - 20
P1/4L (65ml) Dt1514
P5 (65ml) Dt1075
Calibration Solutions for EC1 & pH1 Meter
Standard Conductivity 1413S (65ml) Dt1109
pH 4 (65ml) Dt1119
pH 7 (65ml) Dt1118
pH 10 (65ml) Dt1161
Hardness, Stop/Go (Drop Test) 0-4
Hardness Stop/Go Solution (65ml) Dt1129
-
www.accepta.com 25
Section 4.2Cooling (Open & Closed) Test Kit (Backpack) - Cool3002BP
Contents CodeWaterproof Back Pack Case with Foam Inserts Eq5026E
Waterproof Instructions Eq5084
1 x Plastic Comparator Eq5085
2 x 10ml Comparator Cells Eq5006
3 x Rigid Bottle 250ml Wide Neck Eq5064
2 x 65ml Plastic Test Bottle Eq5018
1 x Filter Holder 25mm Eq5044
1 x Filter Papers GF/C (50pk) Eq5048
1 x 10ml Luer Lock Syringe Eq5073
1 x 20ml Luer Lock Syringe Eq5075
1 x 5/10ml Tube Eq5015
PH1 - pH/Temperature Meter 0 - 14 pH1
EC1 - Conductivity/TDS/Temperature Meter 0 - 20.00 mS EC1
Alkalinity M (Drop Test) 5 - 2400
TA4 (65ml Dt1004
PA2/TA2 (65ml) Dt1005
TA3 (65ml) Dt1006
Hardness, Total (Drop Test) 5 - 1200
TH2 (65ml) Dt1046
H/TH4/CH3 (65ml) Dt1049
TH3 (65ml) Dt1044
TH5 (65ml) Dt1047
The Cooling (Open & Closed) Backpack Test Kit is presented in its own robust case and harness andincludes all the equipment and reagents necessary to effectively test and monitor system waters.
This test kit includes the following:
Acceptas cooling (Open & Closed) Test Kit (Backpack Version)is designed for use on Cooling Towers and many types ofClosed Water Systems such as Chilled Water Circuits and Heat-ing systems.
The Backpack Test Kit has most of the reagents andcomponents that are normally contained in a standard Open &Closed Test Kit but formatted into a backpack design for use bythe commuter in busy commercial areas.
It includes all the equipment and reagents necessary to performthe below noted tests. Replacement reagents and componentsare available which make this kit a lasting addition to any watertreatment professional. Simple to use and fully portable.
-
www.accepta.com 26
* For true Soluble and Total Iron determination, Sulphuric Acid 50% and 0.45 micron filter papersare required and can be ordered separately.
Section 4.2Cooling (Open & Closed) Test Kit (Backpack) - Cool3002BP(continued)Contents Range: CodeHardness, Calcium (Drop Test) 5 - 1200
CH2 (65ml) Dt1043
CH1P (20g) (also requires TH3 & H/TH4/CH3) Dt1050
Hardness, Stop/Go (Drop Test) 0-4
Hardness Stop/Go Solution (65ml) Dt1129
Chloride (Drop Test) 0-3000
BC1/CC1 (65ml) Dt1023
BC2 (65ml) Dt1026
CC2 (65ml) Dt1024
Iron LR (Comparator) 0-1
FE5 (65ml) See below note* Dt1110
Iron HR (Comparator) 0.5 -10
FE10 (65ml) See below note* Dt1247
Bromine (Comparator) 0-5
DPD No1 Solution A (30ml) Dt1505
DPD No1 Solution B (30ml) Dt1506
DPD Nitrite Powder (20g) Dt1295
Molybdate VHR (Comparator) 0-500
Molybdate HR (65ml) Dt1127
Nitrite (Drop Test) 0-2000
N1 (65ml) Dt1072
N2 (65ml) Dt1073
Phosphonate (Drop Test) 0-20
P1/4L (65ml) Dt1514
P5 (65ml) Dt1075
Calibration Solutions for EC1 & pH1 Meter
Standard Conductivity 1413S (65ml) Dt1109
pH 4 (65ml) Dt1119
pH 7 (65ml) Dt1118
pH 10 (65ml) Dt1161
-
www.accepta.com 27
Section 4.3Closed Water Systems Test Kit - Clo3003/Clo3003F
Contents Range Code2011 Medium Case with Foams Eq5028/2
Waterproof instructions Eq5084
1 x 20ml Standard Syringe Eq5168
1 x Plastic Comparator Eq5085
2 x 10ml Comparator Cells Eq5006
1 x 5/10ml Tube Eq5015
PH1 - pH/Temperature Meter 0 - 14 pH1
EC1 - Conductivity/TDS/Temperature Meter 0 - 20.00 mS EC1
Iron HR (Comparator) 0.5 - 10
FE10 (65ml) Dt1247
Molybdate VHR (Comparator) 0 - 500
Molybdate HR (65ml) Dt1127
Nitrite (Drop Test) 0 - 2000
N1 (65ml) Dt1072
N2 (65ml) Dt1073
Calibration Solutions for EC1 & pH1 Meter
Conductivity Standard 1413s (65ml) Dt1109
pH 4 (65ml) Dt1119
pH 7 (65ml) Dt1118
pH 10 (65ml) Dt1161
This kit is presented in its own robust case and includes all the equipment and reagents necessary toeffectively test and monitor Closed Water System to ensure optimum operating conditions areachieved and maintained.
This test kit includes the following:
Please Note: CLK003F includes Filtration & Sulphuric Acid 50% for True Soluble and Total Iron.
Acceptas comprehensive Closed Water System test kit is basedon colour comparator and drop test chemistries and has beendeveloped specifically for engineers and water treatmentprofessionals involved with the maintenance and managementof closed systems to ensure optimum operating conditions.
Closed water systems require careful test monitoring to ensurethe correct water conditions are achieved during the course of atreatment programme.
Failure to maintain the correct dosage levels can lead to scaling,corrosion and increased biological activity which can lead topotentially unstable water conditions.
Liquid equivalent reagents are available on request.
-
www.accepta.com 28
Section 4.4Chlorination Test Kit - Chlor3004
Contents Range Code2004 Mini Case with Foams Eq5030
Waterproof instructions Eq5084
1 x Plastic Comparator Eq5085
2 x 10ml Comparator Cells Eq5006
1 x Tablet Crusher Eq5079
PH1 - pH/Temperature Meter 0 - 14 pH1
Chlorine LR (Comparator) 0 - 4
DPD No1 Solution A (30ml) Dt1505
DPD No1 Solution B (30ml) Dt1506
Chlorine HR (Comparator) 10 - 300
Chlorine HR (65ml) Dt1179
Acidifying Solution (65ml) Dt1180
Calibration Solutions pH1 Meter
pH 4 (65ml) Dt1119
pH 7 (65ml) Dt1118
pH 10 (65ml) Dt1161
DPD No3 Solution C (30ml) Dt1507
This kit includes all the reagents and equipment required to allow the monitoring of low rangechlorine (0 to 4ppm) and high range chlorine (0 to 300ppm). These are required to ensure the correctsystem conditions are achieved.
This test kit includes the following:
Acceptas chlorination test kit is based on colour comparatorchemistry and has been developed specifically for engineersand water treatment professionals involved with the chlorinationand disinfection of water systems.
Water system chlorinations require careful test monitoring toensure the correct water conditions are achieved during thecourse of the disinfection procedure. Failure to achieve thecorrect dosage levels can lead to unsuccessful disinfections,whilst elevated levels can lead to potentially corrosive waterconditions.
This kit is presented in its own robust case and includes all theequipment and reagents necessary to effectively test and mon-itor the chlorination procedure.
-
www.accepta.com 29
Section 4.5Boiler Water Test Kit - Blr3001
Contents Range Code2019 Maxi Case with Foams Eq5027
Waterproof Instructions Eq5084
1 x 20ml Luer Lock Syringe Eq5075
1 x 10ml Luer Lock Syringe Eq5073
1 x Filter Holder 25mm Eq5044
1 x Filter Papers GF/C (50pk) Eq5048
2 x 65ml Plastic Test Bottle Eq5018
1 x Plastic Comparator Eq5085
2 x 10ml Plastic Comparator Cells Eq5006
PH1 - pH/Temperature Meter 0 - 14 pH1
EC1 - Conductivity/TDS/Temperature Meter 0 - 20.00 mS EC1
Alkalinity P (Drop Test) 5 - 2400
PA1 (65ml) Dt1001
PA2/TA2 (65ml) Dt1005
TA3 (65ml) Dt1006
Alkalinity M (Drop Test) 5 - 2400
TA4 (65ml) (also uses PA2/TA2 & TA3) Dt1004
BC2 (65ml) Dt1026
CC2 (65ml) Dt1024
BC1/CC1 (65ml) Dt1023
Chloride (Drop Test) 0 - 3000
PA3 (65ml) (also uses PA1, PA2/TA2 & TA3) Dt1007
Alkalinity OH (Drop Test) 5 - 2400
1 x 5/10ml Tube Eq5015
This professional Boiler Water Test kit is presented in its own robust case and includes all theequipment and reagents necessary to effectively test and monitor system waters to ensure optimumoperating conditions are achieved and maintained.
This test kit includes the following:
Acceptas comprehensive Boiler Water Test Kit is based on acolour comparator and drop test chemistries and has been de-veloped specifically for engineers and water treatment profes-sionals who are involved with the maintenance and manage-ment of boilers and boiler systems.
Boiler Water and feed-water systems require careful testmonitoring to ensure the correct water conditions are maintainedat all times. Failure to maintain the correct water treatmentprograms can lead to serious problems including reduced boilerefficiency, scale, corrosive water conditions and reduced steampurity.
-
www.accepta.com 30
Section 4.5Boiler Water Test Kit - Blr3001 (continued)
Contents Range CodeHardness, Stop/Go (Drop Test) 0 - 4
Hardness Stop/Go Solution (65ml) Dt1129
Hardness, Total LR (Drop Test) 5 - 120
TH2 (65ml) Dt1046
TH5 (65ml) Dt1047
TH3 (65ml) Dt1043
Phosphate (Comparator) 0 - 80
PB1 (65ml) Dt1111
PB2 (65ml) Dt1112
Sulphite (Drop Test) 5 - 300
S1 (40g) Dt1092
S2 (65ml) Dt1093
Calibration Solutions for EC1 & pH1 Meter
Standard conductivity 1413S (65ml) Dt1109
pH 4 (65ml) Dt1119
pH 7 (65ml) Dt1118
pH 10 (65ml) Dt1161
20% Acetic Acid (65ml) Dt1103
Liquid equivalent reagents are available on request.
-
www.accepta.com 31
Section 4.6Legionella Compliance Test Kit - Leg3006
Contents Range Code
2019 Maxi Case with Foams Eq5027
Waterproof Instructions Eq5084
1 x 20ml Standard Syringe Eq5168
1 x 10ml Standard Syringe Eq5177
2 x 65ml Plastic Test Bottle Eq5018
1 x 5/10ml Tube Eq5015
1 x Plastic Comparator Eq5085
2 x 10ml Plastic Comparator Cells Eq5006
PH1 - pH/Temperature Meter 0 - 14 pH1
EC1 - Conductivity/TDS/Temperature Meter 0 - 20.00 mS EC1
Alkalinity M (Drop Test) 5 - 2400
TA4 (65ml) Dt1004
PA2/TA2 (65ml) Dt1005
TA3 (65ml) Dt1006
Bromine (Comparator) 0 - 5
DPD No1 Solution A (30ml) Dt1505
DPD No1 Solution B (30ml) Dt1506
DPD Nitrite Tablets (50pk) Dt1131
Chloride (Drop Test) 0 - 3000
BC1/CC1 (65ml) Dt1023
BC2 (65ml) Dt1026
CC2 (65ml) Dt1024
Replacement reagents and other components are available, please refer to the relevant section.
This test kit includes the following:
Acceptas Legionella compliance kit is based on colour compar-ator and drop test chemistries and has been developed specifi-cally for engineers and water treatment professionals involved inthe weekly and monthly analysis of water systems.
Cooling Water systems require careful test monitoring to ensurethe correct water conditions are achieved to reduce the risk ofLegionella. Failure to keep the system under control can lead toscale, corrosion and the growth of the potentially fatal Legionellabacteria.
This kit is presented in its own robust case and includes all theequipment and reagents necessary to effectively monitor thewater conditions.
-
www.accepta.com 32
Section 4.6Legionella Compliance Test Kit - Leg3006 (continued)
Contents Range CodeHardness, Total (Drop Test) 5 - 1200
TH2 (65ml) Dt1046
TH5 (65ml) Dt1047
TH3 (65ml) Dt1044
H/TH4/CH3 (65ml) Dt1049
Hardness, Calcium (Drop Test) 5 - 1200
CH2 (65ml) Dt1043
CH1P (20g) (also requires TH3 & H/TH4/CH3) Dt1050
Hardness, Stop/Go (Drop Test) 0 - 4
Hardness Stop/Go Solution (65ml) Dt1129
Iron LR (Comparator) 0 - 1
FE5 (65ml) Dt1110
Calibration Solutions for EC1 & pH1 Meter
Standard Conductivity 1413S (65ml) Dt1109
pH 4 (65ml) Dt1119
pH 7 (65ml) Dt1118
pH 10 (65ml) Dt1161
-
www.accepta.com 33
Section 4.7Resin Test Kit - Res3008
Contents Range Code2004 Mini Case with Foam Eq5030
Waterproof Instructions Eq5084
1 x Optic Magnifier Eq5034
1 x 65ml Plastic Test Bottle Eq5018
1 x White Spoon, 0.5ml Eq5382
1 x RA1 Iron Test Reagent (65ml) Dt1338
1 x RA2 Iron test Reagent (65ml) Dt1336
1 x RA3 Organic Test Reagent (65ml) Dt1339
1 x De-ionised Water (65ml) Dt1210
The Resin Test kit is presented in its own robust case and includes all the equipment and reagentsnecessary to effectively test and monitor the qualitative condition of ion exchange resins.
This test kit includes the following:
Ion exchange Resins should be checked when the plant isserviced to determine if the ion exchange resin is sufferingfrom fouling or damage which will reduce the plant efficiency.
Acceptas portable and easy to use kit allows the qualitativeanalysis of ion exchange resins for the three main problemsencountered when water softeners or ion exchange plantefficiency is reduced.
The ion exchange resin can be checked for Iron fouling,organic fouling and resin bead optical assessment. Failure tomaintain good quality resins will lead to serious problemsincluding reduced efficiency and the build-up of scale andincreased microbiological activity, which may include thegrowth of Legionella bacteria.
-
www.accepta.com 34
Section 4.8Manual Silt Density Kit SDI001
Contents Range CodeWaterproof Instructions Eq5084
47mm Filter Holder Eq5086
0.45m Micron Filter Papers (47mm) (50pk) Eq5181
500ml Measuring Cylinder Eq5183
Stopwatch/Clock Eq5239
Dull Blunt Tweezers Dt1240
SDI Assembly Dt1182
These fine un-dissolved suspended particles tend to settle on the surface and in the process, foul themembrane. The net result is an increase in feed pressure, reduced flow and high conductivity ofpermeate. Spiral wound and hollow fine fibre membranes are the most vulnerable and once fouled,cleaning of the element is often difficult.
Use of the KitSilt Density Index (SDI) is an empirical test developed for measuring the rate of fouling of nanofiltration (NF) or RO membranes.
Suspended solids and colloidal materials in feed water are one of the biggest problems in reverseosmosis systems. Although most systems have some pre-treatment, including 5-micron pre-filters,these fine particles are responsible for the fouling of reverse osmosis membranes.
In order to measure the degree of this fouling problem, a concept called Silt Density Index is used.Here, filtration rates are calculated by exposing a 0.45-micron filter to the feed water under pressure.
An SDI of less than 5 is typically considered acceptable for reverse osmosis systems. This meansmembranes should foul at a very low rate at SDI values of less than 5. Although this concept worksmost of the time, there are exceptions when a lower SDI (less than 3) is desirable due to the natureof the suspended solids in feed water.
Depleting supplies of fresh water, ground water contaminationand the rising cost of water have compelled industries toincreasingly turn to the desalination of brackish/sea water aswell as water recovered from treated industrial effluent andsecondary treated municipal sewage.
All these recycled water systems use reverse osmosis (RO)system for removal of dissolved solids. The presence of finelydivided particles in feed water to an RO system can causeserious problems as they are too small to be retained byconventional pre-treatment systems and they find their way tothe RO membrane.
The function of an RO membrane is to concentrate dissolvedsolids on one side of the membrane and allow water moleculesto permeate through the other side. Suspended, colloidalparticles are larger than dissolved solids and are rejected by themembrane.
-
www.accepta.com 35
Section 5.0Pocket Test Meters
-
www.accepta.com 36
Section 5.1PH1 - pH/Temperature Meter
Acceptas waterproof pH1 Meter has a built in smart chip which offers manyadvanced functions and features for a stick style pH Meter.
Features:
Waterproof and rugged enough to withstand harsh environments.
LCD display.
Displays Max & Min values with data hold.
Easy to replace pH electrode.
Exchangeable with an ORP Probe for Redox.
Auto recognition of buffer solutions.
Automatic temperature compensation.
Simultaneous display of pH and temperature.
Auto shut off after 10 minutes.
Icons on display for easy recognition of function.
Specifications:
pH:Temperature:Resolution:Accuracy:Power:Dimensions:Weight:
Optional Extra:Redox Probe
Replacement Probe:pH Probe
0.00 to 14.00pH0 to 90C0.01 pH0.01pH, 1.8F/1C2 x CR2032 Button Batteries195 x 40 x 36mm150g
Code:RE1P
Code:pH1P
Unique flat electrode which is very robust and isalmost unbreakable.
-
www.accepta.com 37
Section 5.2EC1 - Conductivity/TDS/Temperature/Salinity Meter
Acceptas waterproof EC1 Meter has a built in smart chip which offers manyadvanced functions and features for a stick style EC Meter.
Features
Unit is waterproof and rugged enough to withstand many harshenvironments.
LCD Display.
Displays Max & Min values with data hold.
Easy to replace conductivity electrode.
Auto recognition of conductivity solutions.
Automatic temperature compensation.
Simultaneous display of conductivity and temperature.
Auto shut off after 10 minutes.
Icons on display for easy recognition of function.
Specifications
Conductivity:
TDS:
Salinity
Temperature:Power:
Dimensions:
Weight:
Replacement Probe:EC/TDS Probe
Ranges
0 to 199.9S/cm200 to 1999S/cm2.00 to 19.99mS/cm0 to 99.9ppm (mg/L)100 to 999ppm (mg/L)1.00 to 9.99ppt (g/L)0 to 1000ppm100 to 999ppm (mg/L)1.00 to 12.00ppt0 to 90F2 x CR2032 Batteries
36 x 173 x 41mm
110g
EC1P
Max Resolution
0.1S/cm1S/cm0.01mS/cm0.1ppm (mg/L)1ppm (mg/L)0.01ppt (g/L)
1ppm (mg/L)
0.1C
Basic Accuracy
2% FS
2% FS
1ppm / 0.01ppt
0.2C
-
www.accepta.com 38
Section 5.3PH100 - Extech pH/Temperature Meter
The waterproof ExStik pH100 Meter has a built in smart chip which offersadvanced functions and features to a stick style pH Meter not found anywhereelse and has a robust flat surface electrode.
Features: Flat surface pH electrode, rugged for harsh environments.
Internal memory stores up to 15 labelled readings for easy recall.
CAL feature, tells you when to recalibrate your meter.
Analogue bar graph to conveniently view trends.
Renew indicator tells you its time to replace your electrode.
1, 2 or 3 point calibration automatically recognises buffer solutions.
Automatic temperature compensation & waterproof to IP 57.
Simultaneous display of pH and temperature & indicates reading stability.
Data Hold, Auto Power Off and Low Battery Indication.
Waterproof design floats in water.
Interchangeable probes available for Chlorine & ORP
Optional weighted base (EX006) holds plastic cup and ExStik Meter.
Complete with flat surface pH electrode, protective sensor cap, samplecup with four 3V CR-2032 button batteries, and 1.2m neck strap.
Specifications:
pH:Temperature:Resolution:Accuracy:Power:Dimensions:Weight:
Optional Extra:
Weighted Base and 5 Solution Cups
Replacement Probe:Probe (PH105)
0.00 to 14.00pH23 to 194F (-5 to 90C)0.01pH, 0.10.01pH, 1.8F/1C4 x CR2032 Button Batteries35.6 x 172.7 x 40.6mm110g
Code:
EX006
Code:Eq5039
-
www.accepta.com 39
Section 5.4EC400 - Extech Conductivity/TDS/Temp/Salinity Meter
The waterproof ExStik Conductivity/TDS/Salinity Meter offers one high accuracymulti-range sensor which measures four parameters (Conductivity, TDS, Salinityand Temperature).
Features Auto ranging meter offers 3 ranges to measure, pure water, salinity solu-
tion, waste, boiler and cooling water using the same conductivity cell.
6 units of measure: S/cm, mS/cm, ppm, ppt, mg/L and g/L.
Internal memory stores up to 15 labelled readings for easy recall.
Waterproof design to withstand wet environments meets IP57 Standard.
Adjustable Conductivity to TDS ratio factor from 0.4 to 10 convenientlycalculates the TDS value with easy to replace conductivity cell module.
Large digit, digital display with analog bar graph to indicate sample trends.
Self calibration on power up, data hold, Auto Power Off & low batteryIndication.
Automatic Temperature Compensation of 2% per C.
Simultaneous display of Conductivity, TDS, or Salinity and temperature.
EC400 includes meter and Conductivity cell, protective sensor cap,sample cup with cap, batteries and 1.2m neck strap.
Specifications
Conductivity:
TDS:
Salinity:
Temperature:Power:
Dimensions:
Weight:
Optional Extras:Weighted Base and 5Solution Caps
Replacement Probe:EC405 Probe
Ranges
0 to 199.9S/cm200 to 1999S/cm2.00 to 19.99mS/cm0 to 99.9ppm (mg/L)100 to 999ppm (mg/L)1.00 to 9.99ppt (g/L)0 to 99.9ppm (mg/L)100 to 999ppm (mg/L)1.00 to 9.99ppt (g/L)32 to 149F (0 to 65C)4 x CR44W Batteries
36 x 173 x 41mm
110g
Code:EX006
Code:Eq5032
Max Resolution
0.1S/cm1S/cm0.01mS/cm0.1ppm (mg/L)1ppm (mg/L)0.01ppt (g/L)0.1ppm (mg/L)1ppm (mg/L)0.01ppt (g/L)0.1F/C
Basic Accuracy
2% FS
2% FS
2% FS
1.8F/1C
-
www.accepta.com 40
Section 5.5EC500 - Extech Combined pH/Conductivity/TDS/Temperature Meter
The waterproof ExStik II pH/Conductivity Meter Combination comes with arugged flat surface pH electrode and an Auto-ranging high accuracy conductivitycell.
Features
Measures 5 parameters including Conductivity, TDS, Salinity, pH, andTemperature using one electrode.
9 units of measure: pH, S/cm, mS/cm, ppm, ppt, mg/L, g/L, C, F
Analog bargraph indicates trends.
Memory stores up to 25 labelled readings
Adjustable Conductivity to TDS ratio from 0.4 to 1.0; 0.5 fixed Salinity ratio.
RENEW feature alerts user when electrode needs replacement.
Auto Power Off and Low Battery Indicator & waterproof to IP57.
EC500 meter includes electrode, protective sensor cap, sample cup withcap, four 3V CR-2032 button batteries and 1.2m neck strap.
EC510 Kit includes EC500, 84S/cm, 1413S/cm and 12.88mS/cmcalibration standards, pH buffer pouches (1 each of 4, 7,10 pH plus rinsesolution), weighted base, 3 sample cups with caps and case.
Specification
Conductivity:
TDS/Salinity:
pH:
Temperature:Waterproof:Memory:Dimensions:Weight:
Optional Extras:Weighted Base and 5Solution Cups
Replacement Probe:Probe (EC505)
Range
0 to 199S/cm200 to 1999S/cm2.00 to 19.99mS/cm0 to 99.9ppm (mg/L)100 to 999ppm (mg/L)1.00 to 9.99ppt0.00 to 14.00pH
32 to 149F (0 to 65C)iP6725 datasets36 x 186 x 41mm110g
Code: EX006
Code: Eq5041
Max Resolution
0.1S/cm
0.1ppm (mg/L)
0.01pH
0.1F/C
Basic Accuracy
2% FS
2% FS
0.01pH
1.8F/1C
-
www.accepta.com 41
Section 5.6CL200 - Extech Total Chlorine/Temperature Meter
The waterproof ExStik CL200 Chlorine Meter gives a direct reading of TotalChlorine down to 0.01ppm and is unaffected by colour/turbidity of the sample.
Features
Direct reading of Total Chlorine provides fast and easy measurements (lessthan 2 minutes).
Low detection limit down to 0.01ppm.
Wide dynamic range of measurement up to 10ppm (use dilution method for10 to 50ppm).
Unaffected by sample colour or turbidity.
Automatic electronic calibration.
Memory stores, tags and recalls up to 15 readings.
Unique flat surface chlorine electrode.
Waterproof to IP67 standard.
The state of the art LCD simultaneously displays ppm of TotalChlorine and Temperature plus analog bargraph indicator.
Complete with meter and flat surface Chlorine electrode, ExTab reagenttablets (50 tests), protective sensor cap, sample cup with cap, four 3VCR-2032 Button Batteries, and a 1.2m neck strap.
Specifications
Range in ppm: 0.01 to 10.00ppm (10 to 50ppm using dilution method)Temperature: 23 to 194F (-5 to 90C)Resolution: 0.01ppm ; 0.1F/CAccuracy: 10% of reading 0.01ppm; 1.8F/1CPower: 4 x CR-2032 Button BatteriesDimensions: 35.6 x 172.7 x 40.6mmWeight: 110g
Optional Extras
Weighted Base and 5 Solution Cups: Code: EX006Replacement Probe (CL205) Code: Eq5040
-
www.accepta.com 42
Section 5.7DO600 - Extech Dissolved Oxygen/Temperature Meter
The waterproof ExStik II Dissolved Oxygen Meter incorporates an easy toreplace screw on membrane cap with optional extension cables.
Features
Oxygen level displayed as % Saturation from 0 to 200.0% orConcentration from 0 to 17.00ppm (mg/L).
Adjustable Altitude Compensation (0-20,000ft in 1,000ft increments).
Adjustable Salinity Compensation from 0 to 50ppt.
Analog bargraph indicates trends.
Memory stores up to 25 data sets with DO and Temperature reading.
Self calibration on power up.
Data Hold, Auto Power Off, Low Battery Indicator.
Waterproof to IP57.
Optional 1m or 5m extension cable with probe weight.
Complete with DO electrode, protective sensor cap, spare membrane cap,electrolyte, four 3V CR-2032 Button Batteries and 1.2m neck strap.
Specification Range Max Resolution Basic Accuracy
DO (saturation mode): 0 to 200.0% 0.1% 2% FSDO (concentration mode): 0 to 20.00ppm (mg/L) 0.01ppm (mg/L) 0.4ppm (mg/L)Temperature: 32 to 122F (0 to 50C) 0.1F/C 1.8F (1C)
Dimensions 36 x 176 x 41mmWeight: 110gPower: 4 x CR2032 Button Batteries
Optional Extras
Weighted Base and 5 Solution Cups: Code: EX006
Replacement Probe: Code: D0605
-
www.accepta.com 43
Section 5.8RE300 - Extech ORP Meter
The Extech ORP Meter has flat surface electrodes which provides On-the-Spotmeasurements and also has a large display with bar-graph.
Features:
ORP from 999 to 999mV, resolution of 1mV and 4mV accuracy.
Automatic electronic self calibration.
Simulated analogue bar graph displays change in ORP reading.
Memory stores and recalls 15 readings and saves last calibrated value.
Data validation indicators show user the reading is stabilised.
Large 3-1/2 digit (2000 count) digital display with bar graph.
Data Hold, Auto Shut-off and Low Battery Indicator.
Easy to replace electrode module is interchangeable with pH and Chlorinemodules; (built in chip automatically recognises the type of probe pluggedin for specific measurements).
Waterproof IP57 floats in water & protects the meter in wet environments.
Complete with meter and electrode, protective sensor cap, sample cup withcap, four 3V CR-2032 Button Batteries and neck strap
Specifications
ORP (mV): 999 to 999mVResolution: 1mVAccuracy: 4mVPower: 4 x CR2032 Button BatteriesDimensions: 35.6 x 172.7 x 40.6mmWeight: 110g
Optional Extras
Weighted Base and 5 Solution Cups: Code: EX006
Replacement Probe: (RE305) Code: Eq5031
-
www.accepta.com 44
Section 5.9Refractometer - Glycol Tester (RefG300)
Acceptas refractometer is a precision optical instrument designedto measure the concentration of Glycol as PEG or MEG displayedas a % and freeze point
It is easy to use, you simply place a few drops of the treated solu-tion with the pipette included on to the prism glass and hold up tothe light. You can then read the concentration of Glycol in % orfreeze point on to the lens.
Our Refractometer has a built-in automatic temperaturecompensation system.
Measuring Range
Ethylene Glycol: 60C - 0C with a resolution of 5C
Polypropylene: 50C - 0C with a resolution of 5C
Battery Acid: 1.100-1.400sg with a resolution of 0.01sg
Other Refractometers for Coolants, Brix, Sugars etc. are available on request.Please contact us for details.
-
www.accepta.com 45
Section 6.0Professional Test Meters
-
www.accepta.com 46
Section 6.1Martini MI806 - 4 in 1 pH/Conductivity/TDS/Temp
The Martini Mi806 is a multi-function portable meter measuringpH, Conductivity, TDS and temperature using a single probe.
Measures 4 parameters with 1 single probe.
You can select from a range of calibration buffers andalso the temperature scale (C or F).
The multi-parameter probe Mi806P includes pH/EC/TDSand temperature sensors all in one rugged unit.
TDS factors from 0.45 to 1.00, range of temperaturecoefficients () from 0.0 to 2.4% for greater reproducibility.
Stability Indicator prompts when the reading stabilizes.Auto-Hold Function freezes reading for later viewing.
Large and Easy-to-Read display provides simultaneousreadings of pH and Temperature or EC/TDS andtemperature
Specifications
Range: pHECTDSTemp
Resolution: pHECTDSTemp
Accuracy (25C): pHEC/TDSTemp
Typical EMC Deviation: pHEC/TDSTemp
Typical Compensation:
pH Calibration:EC Calibration:EC/TDS Conversion Factor:Probe (Included):
Environment:Battery Type (Included):Battery Life:Auto-off:Dimensions:Weight (with battery)
0.00 to 14.00 pH0.00 to 20.00 mS/cm0.00 to 10.00 ppt0.0 to 60.0C / 32.0 to 140.0F0.01pH0.1 mS/cm0.01 ppt0.1C / 0.1F0.01 pH2% FS0.5C / 1F0.02 pH2% FS0.5C / 1FAutomatic from 0 to 60C;6 adj. from 0.0 to 2.4%/C1 or 2-point with automatic buffer recognitionAutomatic, 1 pointAdj. from 0.45 to 1.00Mi806P (MA851D/1) amplified pH/EC/TDS/Temperatureprobe with DIN connector and 1m cable
0 to 50C / 32 to 122F; 100% RH max1 x 9ValkalineApprox. 300 hoursAfter 8 minutes200 x 85 x 50mm260g
Replacment Probe: Code: MI806PMI806 Probe
-
www.accepta.com 47
Section 6.2Eutech pH11 - pH/Temperature Meter
The Eutech pH11 is a simple to use, accurate, reliablehand held meter which has been designed to handle thetough environment of field work.
Large, Easy to Read display with simultaneousreadings with user-friendly icons.
5-point Push-Button Calibration with Auto-Bufferrecognition for quick calibration with no mistakes.
Automatic temperature compensation.
Stores/Recalls up to 100 Data sets with temperaturereadings.
Supplied with probe & batteries.
Stability indicator prompts stable reading.
Supplied with pH buffers solution 4.0 & 7.0
Self-Diagnostic with message codes for easytroubleshooting.
Specifications Ranges Max Resolution Basic Accuracy
pH:mV:
-2 to 16.000 to 1000 mV
0.01pH0.01pH
0.02 pH 2mV
Temperature: -5 to 105C 0.1C 0.5C
Power: 4 x AAA Batteries
Dimensions: 85 x 185 x 40 mm
Weight: 224g
-
www.accepta.com 48
Section 6.3Eutech CON 11 - Conductivity Meter
The Eutech CON 11 is a simple to use, accurate, reliablehand held meter which has been designed to handle thetough environment of field work.
Large, Easy-to-Read LCD Display with simultaneousreadings and user-friendly icons.
Stability indicator prompts when reading stabilizes.
Auto - ranging Conductivity to detent the appropriatemeasurement range - ensures the best resolution.
User Customisable Functions for options:Normalisation Temperature. Temperature coefficient,TDS factor & cell constant.
Automatic temperature compensation.
Supplied with calibration solution.
Auto-off feature after 5 minutes to save battery life.
Single & Multi-point Automatic or Manual Calibration.
Specifications Ranges Max Resolution Basic Accuracy
Conductivity: 0 to 19.99 S/cm20.0 to 199.9 mS/cm2.00 to 19.99 mS/cm20.0 to 199.9 mS/cm
0.1S/cm0.01mS/cm
0.1C
1%
1.8F/1C
Temperature: -9 to 120C
Power: 4 x AAA Batteries
Dimensions: 85 x 185 x 40 mm
Weight: 224g
-
www.accepta.com 49
Section 6.4Eutech PC300 - Combined pH/Condy/TDS/Temp Meter
The Eutech PC 300 is designed to readily withstand therigours of portable analysis, while reliably measuring pH,Conductivity and Temperature in a simple professionalmanner.
Waterproof and Dust proof Housing - IP67 rated.
Multi-parameter- toggle between pH/temperature andConductivity/temperature at touch of buttons.
Multi-point Push-button Calibration (up to 3/4 points forpH/Conductivity) - for easy, quick calibration.
Auto-Ranging for Conductivity measurements, fastresponse and best resolution over range.
User-Customization for enhanced flexibility, viewing ofConductivity/TDS cell constants per range, electrodediagnosis, pH slope/offset/mV base value.
Stability indicator prompts reading stability.
Auto-Hold Function automatically freezes readingwhen stabilized for easy viewing and recording.
Auto-Power Off saves battery power after non-use.
Self-Diagnostic message codes for troubleshooting.
Large Easy-to-Read Custom LCD with simultaneousdisplay for optimum viewing with user-friendly icons
Specifications Ranges Max Resolution Basic Accuracy
pH: -2 to 16.00 0.01pH 0.02 pH
mV: 0 to 1000 mV 1 Mv 2mV
Temperature: -5 to 105C 0.1C 0.5C
Power: 4 x AAA Batteries
Dimensions: 85 x 185 x 40 mm
Weight: 224g
-
www.accepta.com 50
Section 6.5Jenway 350 - pH Meter
The Jenway 350 pH Meter offers simple yet professionalperformance for the operator who requires reliable resultsfrom trusted technology. With its single pH input andmanual temperature compensation the 350 is easy to use,but also offers such advanced features as the 32 readingnon-volatile memory.
Features Power on self-check routine with 32 reading
data storage.
Environmentally protected case to IP65.
Automatic or Manual calibration with 500hours battery life.
Simultaneous pH and temperature reading.
Full range of application specific electrodesand consumables.
Specifications
pH Range:Resolution (Accuracy):Manual Temperature Compensation:Resolution:Auto Buffer Recognition:Calibration:Power:Battery Life:Size:Weight:
-2.00 to 16.00pH0.01pH ( 0.02pH)0 to 100C / 32 to 212F0.1C / 1F4.00, 7.00, 9.22 & 10.00 (manual override)User selectable 1 or 2 point2 x AA batteries500 hours typical (@25C with alkaline cells)175 x 75 x 35 mm250g
The custom back-lit liquid crystal display simultaneously shows the pH reading and the manually settemperature compensation value that is simply adjusted with the up and down arrow keys. A one ortwo point calibration can be selected and, for ease of operation, automatic buffer recognition can beused. In special applications this can be overridden, enabling calibration to be set at any two pointsacross the full pH range.
An indication of battery life is also permanently shown on the display. An automatic switch off facilityhelps to conserve battery life. This product is ideal for use in field or any portable situation.
-
www.accepta.com 51
Section 6.6Jenway 470 - Conductivity/TDS/Temp Meter
The Jenway 470 Conductivity/TDS/Temp Meter is a generalpurpose hand held meter offering a wide measuring range, whichis further complimented by the auto-ranging function that ensuresresults are displayed with the highest resolution that can beachieved.
Features Automatic range selection for optimum resolution.
Automatic or manual calibration.
32 reading data storage.
Conductivity and TDS modes.
Power on self-test routine.
Specifications:
Conductivity:Ranges:Resolution (Accuracy):TDS Ranges:
Resolution (Accuracy):Temperature Range:Resolution Accuracy:ATC & Manual ranges:Cell constant:Reference Temperature:Temperature Coefficient:Auto Standard Recognition:Probe input:Power:Battery Life:Size:Weight:
6 switched ranges0 to 19.99S - 0 to 1999mS*0.1S to 1mS* (0.5% 2 digits)6 switched ranges0 to 19.99mg/l - 0 to 1999g/l*0.01mg/l to 1g/l* (0.5% 2 digits)-10 to 105C / 14 to 221F0.1C/1F (0.5C/1F)0 to 100C / 32 to 212FDigitally settable 0.01 to 19.9918, 20 or 25C settable0.00 to 4.00%/C linear10S, 84S, 1413S or 12.88mS (manual override)Mini-DIN conductivity/temperature2 x AA batteries500 hours typical (@25C with alkaline cells)175 x 75 x 35 mm250g
* only with cell constant > 5
Calibration can be carried out automatically against standard solutions at 10S, 84S, 1413S and12.88mS or a manual calibration can be made against any other standard value.
Standard conductivity measurements can be made quickly and easily using the default settings butfor more complex applications a full range of adjustments can be made. The reference temperaturecan be set to 18, 20 or 25C and the temperature coefficient set from 0 to 4%/C. When using thealternative TDS range the E/C ratio can be set from 0.5 to 0.8 to ensure conformance with alloperating procedures.
A simple dual function memory bar enables up to 32 readings to be stored and recalled from the non-volatile internal memory.
An indication of battery life is also permanently shown on the display. An automatic switch off facilityhelps to conserve battery life. This product is ideal for field or any portable situation.
-
www.accepta.com 52
Section 6.7Jenway 430 - Combined pH/Conductivity/TDS/Temp Meter
The 430 is designed to readily withstand the rigours of portable anal-ysis, while reliably measuring pH and Conductivity in a simple andprofessional manner.
Features
Simultaneous readout of pH, Conductivity or TDS andtemperature
Fully featured pH and Conductivity meters in a single,hand held unit
Automatic or manual calibration of both functions Power on self-check routine Calibration diagnostics Environmentally protected case to IP65
Specifications:
pH Range (1 or 2 point call):Resolution:Accuracy:Conductivity Ranges:Resolution:Accuracy:TDS Ranges:Resolution:Accuracy:Temperature Ranges:Resolution:Accuracy:ATC Range:Manual Temp. Comp.:Auto Buffer Recognition:Auto Standard Recognition:Cell Constant:EC Ration (TDS):Reference Temperature:Temperature Coefficient:Power:Battery Life:Size:Weight:
0 to 14pH0.01pH0.02pH5 switched ranges: 0 to 19.99S - 0 to 199.9mS0.01S, 0.S, 1S, 0.01mS, 0.1mS0.5% 2 digits5 switched ranges: 0 to 19.99mg/l - 0 to 199.9g/l0.01mg/l, 0.1mg/l, 1mg/l, 0.01g/l, 0.1g/l0.5% 2 digits-9.9 to +99.9C / 32 to 212F0.1C/ 1F0.5C / 1F0 to 99.9C / 32 to 212F0 to 99.9C / 32 to 212F4, 7, 9.22 and 10.00 (manual override facility)10S, 84S, 1413S & 12.88mS0.01 to 19.990.50 to 0.8018, 20 or 25C0.00 to 4%/C2 x AA batteries100 hours typical (@25C with alkaline cells)175 x 75 x 35 mm250g
This product allows the simultaneous readout of pH, Conductivity or TDS and Temperature using thecustom backlit liquid crystal display. Calibrations are clearly indicated, together with the parameter inerror.
It offers 1 or 2 point pH calibration and automatic buffer recognition with manual override.Conductivity calibration can be performed on standard solutions or by direct cell constant entry. Theconductivity/TDS range is automatically selected for optimum resolution and temperaturecompensated using an adjustable temperature coefficient.
The 430 is supplied with a combined conductivity and pH probe assembly (a temperature sensor isalso incorporated to provide automatic temperature compensation). The probe connects to theinstrument via a single mini-DIN connector for operator convenience.
An indication of battery life is also permanently shown on the display. An automatic switch off facilityhelps to conserve battery life. This product is ideal for field or any portable situation.
-
www.accepta.com 53
Section 6.8TRANSEC - Conductivity/TDS/Salinity/Temp Meter
The Conductivity/TDS Meter HC3010 is a highly sophisticated meter withability to automatically display and convert in Siemen (S), mSiemen (Ms),TDS in ppm and Salinity in ppt.
The meter is versatile and able to preset at different conversion factors.User can select to display in Conductivity or TDS. Its 99 data memorycoupled with real time clock enable you to record reading with date andtime.
For QC control, data can be analysed to display maximum and minimumreadings over all saved data. The meter comes complete in a hard casewith a set of calibration fluids for on the spot calibration.
Auto-range measurement
Multi range display of conductivity and TDS
Dual conductivity or TDS and temperature reading display
Save up to 99 data in memory with real time clock
Able to show maximum and minimum reading over saved data
Stability indicator prompts whenever reading stabilizes
Automatic temperature compensation (ATC)
Supplied with probe & batteries
Auto-off feature to save battery life.
Specifications Ranges Max Resolution Basic Accuracy
Conductivity: 0 to 1999 S/cm 20 to 199.9 mS/cm2.00 to 19.99 ms/cm20.0 to 199.9 ms/cm
0.1S/cm 0.01mS/cm
Temperature: -5 to 105C 0.1C 1.8F/1C
Power: 4 x AAA Batteries
Dimensions: 40 x 195 x 36 mm
Weight: 150g
-
www.accepta.com 54
Section 6.9TRANSPH - pH (also ORP Meter with additional probe)
The pH Meter HP3040 is a portable yet highly accurate pH meter.
Used in all industries, the meter uses the regular BNC connector and ac-cepts a variety of electrodes for special measurement needs.
The meter comes complete in a hard case with calibration buffer solutionsand offers ATC (automatic temperature compensation).
Large LCD with multiple display, displays pH, temperature and timesimultaneously
Multi point calibration - Automatic buffer recognition with built instandards pH 4, 7 & 10
Custom set available - Manual set to any buffer standard
Up to 5 point calibrations
Capable of reading mV readings with additional probe
Stores/recalls up to 99 data sets with date & time
Supplied with probe & batteries
Stability indicator prompts whenever reading stabilises
Self- Diagnostic with message codes for easy troubleshooting
Auto-off feature to save battery life
NOTE: Additional probe required for ORP (mV) measurements - sold separately
Specifications Ranges Max Resolution Basic Accuracy
pH 0-14 0.01pH 0.02pH
Temperature: -5 to 105C 0.1C 0.5C
Power: 4 x AAA Batteries
Dimensions: 40 x 195 x 36 mm
Weight: 180g
MV 0.1 to 199.9 mV / 0 to 1999 mV 0.2 mV / 2mV 0.2mV/2 mV
-
www.accepta.com 55
Section 6.10Borescope/Wireless Inspection Camera - BR200/BR250
Cooling tower fill Pack, Ductwork & Calorifier Monitoring
Each Kit comes complete with: 4 x AA Batteries Rechargeable Display Battery Micro SD Card Memory card with SD Adapter USB Cable Video Interconnect Cable AC Adapter Magnetic Base stand and hard case
We offer two affordable Borescopes
The BR200 has a 17mm diameter camera head with a length of 1 meter which can be extended byuse of up to 2 extension cables each of 0.96m.
The BR250 has a 9mm diameter camera head with a length of 0.9 meter which cannot be extended.
Both cameras have a detachable wireless colour display which can be viewed at 10m from thesource and all images/video can be transferred to your PC via the microSD card (with included SDadaptor) or the USB cable (included) and viewed using Windows Media Player.
The 9mm (BR250) and 17mm (BR200) cameras are waterproof and have 4 LED lamps with dimmerto illuminate the viewed object with a glare free close up field of view. The gooseneck cable retainsits configured shape.
A memory card is includedfor capturing JPEGimages and video (AVI)with day time stamp.
9mm or 17mm diametercamera head with flexiblegooseneck cable with 4 built inbright LED lights which easilyilluminate dark areas.
The detachable wireless3.5 colour TFT LCD screencan be viewed form remotelocation up to 10m from thesource.
Order Options:BR200 - 17mm diameter BorescopeBR250 - 9mm diameter BorescopeBR200-EXT - BR200 Borescope Extension Cable
-
www.accepta.com 56
Section 7.0Drop Test Kits
-
www.accepta.com 57
Section 7.1Drop Test Kits
Acceptas high quality drop test kits are available in a range of parame-ter specific options to suit most analysis requirements. Each self con-tained kit is housed in its own rugged case and comes complete withtitration bottle, 65ml reagents, tablets or powders; a crushing rod (if re-quired) and an easy to follow waterproof instruction manual.
Each drop test kit or replacement reagent pack includes reagentsnecessary to perform between 100 and 200 tests.
Replacement reagents are available, refer to back of catalogue.
Parameter Range Order Code
Acid Strength 0-25% Drp19079
Cleaning Acid Strength Tablets (100pk) Dt1025
pH Test Strips 0-14 (100pk) Str6022
Acidity 50-20,000 Drp19001
PA1 (65ml) Dt1001
ACD2 (65ml) Dt1002
ACD3 (65ml) Dt1003
Alkalinity P 5-2400 Drp19002
PA1 (65ml) Dt1001
PA2/TA2 (65ml) Dt1005
TA3 (65ml) Dt1006
Alkalinity M 5-2400 Drp19003
PA2/TA2 (65ml) Dt1005
TA3 (65ml) Dt1006
TA4 (65ml) Dt1004
Alkalinity OH 5-2400 Drp19004
PA1 (65ml) Dt1001
PA2/TA2 (65ml) Dt1005
PA3 (65ml) Dt1007
TA3 (65ml) Dt1006
Alkalinity P, M & OH 5-2400 Drp19005
PA1 (65ml) Dt1001
PA2/TA2 (65ml) Dt1005
PA3 (65ml) Dt1007
TA3 (65ml) Dt1006
TA4 (65ml) Dt1004
Individual Replacement Components & complete Replacement Packs are available for each ofthe above parameters.
-
www.accepta.com 58
Section 7.1Drop Test Kits (continued)
Parameter Range Order Code
Anionics (As Sodium Lauryl Sulphate) 5-2000 Drp19006
P1/M (50ml) Dt1012
P2/3 (65ml) Dt1013
P9 (65ml) Dt1014
P10 (65ml) Dt1015
Bromine Total 0-20 Drp19007
TBR1 (65ml) Dt1016
TBR2 (65ml) Dt1017
Carbohydarzide 5-50 Drp19074
CZ1 (65ml) Dt1652
CZ2 (65ml) Dt1653
Carbon Dioxide 10-300 Drp19073
PA1 (65ml) Dt1001
CDX2 (65ml) Dt1646
Chelant (Free) (as EDTA) 10-240 Drp19053
FC1 (65ml) Dt1018
FC2 (65ml) Dt1019
FC3 (65ml) Dt1020
Chelant LR (Total) (as EDTA) 10-240 Drp19008
CT2 (65ml) Dt1022
P1/4 Tablets (50pk) Dt1074
Chloride LR & HR 0-3000 Drp19009
BC1/CC1 (65ml) Dt1023
BC2 (65ml) Dt1024
CC2 (65ml) Dt1024
Chlorine LR & HR (Free) 0-300 Drp19010
FCL1 (65ml) Dt1027
FCL2 (65ml) Dt1028
CD4 (40g) Dt1077
FCL3 (65ml) Dt1029
Chlorine Dioxide ULR 0.05-1.0 Drp19072
CD5 (40g) Dt1292
CD7 (65ml) Dt1114
CD6 (20g) Dt1293
Individual Replacement Components & complete Replacement Packs are available for each ofthe above parameters.
-
www.accepta.com 59
Section 7.1Drop Test Kits (continued)Parameter Range Order Code
Chlorine Dioxide VLR 0-12 Drp19011
CD01 (65ml) Dt1036
CD02A (65ml) Dt1122
Chlorine Dioxide LR 0-24 Drp19012
CD01 (65ml) Dt1036
CD02 (65ml) Dt1037
CD02A (65ml) Dt1122
Chlorine Dioxide VLR-HR 0-600 Drp19013
CD01 (65ml) Dt1036
CD02 (65ml) Dt1037
CD02A (65ml) Dt1122
CD03 (65ml) Dt1038
Hardness Stop/Go 0-4 Drp19014S
Hardness Stop/Go Solution (65ml) Dt1129
Hardness Yes/No 0-20 Drp19014T
Yes/No Hardness Tablets (50pk) Dt1121
Hardness Calcium 5-1200 Drp19015
CH2 (65ml) Dt1043
CH1P (20g) Dt1050
H/TH4/CH3 (65ml) Dt1049
TH3 (65ml) Dt1044
Hardness Total 5-1200 Drp19016
TH3 (65ml) Dt1044
TH2 (65ml Dt1046
TH5 (65ml) Dt1047
H/TH4/CH4 Dt1049
Hardness Calcium & Total 5-1200 Drp19043
CH2 (65ml) Dt1043
TH2 (65ml Dt1046
CH1P (20g) Dt1050
TH5 (65ml) Dt1047
H/TH4/CH3 (65ml) Dt1049
TH3 (65ml) Dt1044
Individual Replacement Components & complete Replacement Packs are available for each ofthe above parameters.
-
www.accepta.com 60
Section 7.1Drop Test Kits (continued)Parameter Range Order Code
Hydrogen Peroxide LR 1-100 Drp19051
HP1 (65ml) Dt1051
HP3 (65ml) Dt1053
Hydrogen Peroxide HR 10-5000 Drp19017
HP1 (65ml) Dt1051
HP2 (65ml) Dt1052
Hydrogen Peroxide LR & HR 5-5000 Drp19054
HP1 (65ml) Dt1051
HP2 (65ml) Dt1052
HP3 (65ml) Dt1053
Hypochlorite (Sodium) 0-15% Drp19060
CL1P (20g) Dt1030
CL2A (50pk) Dt1031
CL5 (65ml) Dt1035
CL2B (10g) Dt1032
Iron VHR 50-6000 Drp19033
HFE1 (65ml) Dt1055
HFE2 (65ml) Dt1056
HFE3 (20g) Dt1057
HFE4 (65ml) Dt1058
Molybdate LR 0-50 Drp19018
MB1 (65ml) Dt1063
MB2 (5g) Dt1064
MB3 (65ml) Dt1065
Nitrite 10-2000 Drp19020
N1 (65ml) Dt1072
N2 (65ml) Dt1073
Peracetic Acid 10-6000 Drp19045
CL2A (50pk) Dt1031
CL2B (10g) Dt1032
HP1 (65ml) Dt1051
CL7 (65ml) Dt1054
pH 4-12 Drp19022
Universal pH Indicator Dt1164Individual Replacement Components & complete Replacement Packs are available for each ofthe above parameters.
-
www.accepta.com 61
Section 7.1Drop Test Kits (continued)Parameter Range Order Code
Phosphonate (Short) (HEDP & AMP or PBSAM) 0-20 Drp19023
P1/4 Tablets (50pk) or Dt1074
P1/4L (65ml) Dt1260
P5 (65ml) Dt1075
Phosphonate (S/Long) (HEDP & AMP or PBSAM)(Use in areas where the Alkalinity is greater than500ppm)
0-20 Drp19059
P1/2 (65ml) Dt1514
P3 (65ml) Dt1080
P4 Tablets (100pk) or Dt1080
P4L (65ml Dt1291
P5 (65ml) Dt1075
Phosphonate (PBTC) 0-50 Drp19032
PO1 (65ml) Dt1351
PO2 (65ml) Dt1352
PO3 (65ml) Dt1353
Polyacrylate & Total Polymer 0-20 Drp19025
A1 (65ml) Dt1082
A2 (65ml) Dt1083
A3 (100ml) Dt1087
A4 (65ml) Dt1084
A5 (65ml) Dt1085
A6 (65ml) Dt1086
25mm Filter Holder Eq5044
0.45 filter papers (50pk) Eq5061
Polyamine (as CTAB) 0-20 Drp19026
P1/M (50ml) Dt1012
P2/3 (65ml) Dt1013
P4/2 (65ml) Dt1088
QAC (cationic) 0-1000 Drp19027
QA1 (65ml) Dt1089
QA2 (65ml) Dt1090
QA3 (65ml Dt1091
Individual Replacement Components & complete Replacement Packs are available for each ofthe above parameters.
-
www.accepta.com 62
Section 7.1Drop Test Kits (continued)Parameter Range Order Code
Sulphite 5-300 Drp19028
S1 (40g) Dt1092
S2 (65ml) Dt1093
Silver 0-60 Drp19080
Silver Reagent (40g) Dt1727
Silver Indicator (65ml) Dt1728
Tannin 5-200 Drp19029
TN1 (40g) Dt1094
TN2 (65ml) Dt1095
Total Oxidant 0.02-6.0 Drp19030
TO1 (65ml) Dt1213
TO2 (40g) Dt1214
TO3 (40g) Dt1215
TO4 (65ml) Dt1216
Zinc 0-5 Drp19031
Z1 (65ml) Dt1096
Z4T (50pk) Dt1101
Z5 (65ml) Dt1097
Individual Replacement Components & complete Replacement Packs are available for each ofthe above parameters.
-
www.accepta.com 63
Section 8.0MiniKits
-
www.accepta.com 64
Section 8.1MiniKits
Acceptas high quality MiniKits are available in a range of parameterspecific options to suit most analysis requirements.
Each self contained kit is housed in its own rugged case and comescomplete with 30ml reagents, tablets or powders; a test bottle and easyto follow waterproof instructions. Each kit will perform between 30 and50 tests.
Additional parameters and replacement reagents are also available,please refer to Replacement Reagents section.
Parameter Range Order Code
Acidity 5-6000 Min41001
PA1 (30ml) Dt1001
ACD3 (30ml) Dt1003
Alkalinity P 5-2000 Min41002
PA1 (30ml) Dt1001
TA3 (30ml) Dt1006
Alkalinity M 5-2000 Min41003
TA3 (30ml) Dt1006
TA4 (30ml) Dt1004
Alkalinity OH 5-2000 Min41004
PA1 (30ml) Dt1001
PA3 (30ml) Dt1007
TA3 (30ml) Dt1006
Bromine Total 0-20 Min41005
TBR1 (30ml) Dt1016
TBR2 (30ml) Dt1017
Chelant Total 0-240 Min41006
P1/4L (30ml) Dt1260
CT2 (30ml) Dt1022
Chloride LR 5-200 Min41007
BC1/CC1 (30ml) Dt1023
CC2 (30ml) Dt1024
Chloride HR 100-3000 Min41008
BC1/CC1 (30ml) Dt1023
BC2 (30ml) Dt1026
Chlorine LR 2-30 Min41009
FCL1 (30ml) Dt1027
FCL2 (30ml) Dt1028
Individual Replacement Components & complete Replacement Packs are available for each ofthe above parameters.
-
www.accepta.com 65
Section 8.1MiniKits (continued)Parameter Range Order Code
Hardness Calcium 10-1200 Min41012
CH2 (30ml) Dt1043
H/TH4/CH3 (30ml) Dt1049
CH1P (10g) Dt1050
Hardness Total 10-1200 Min41013
TH2 (30ml) Dt1046
H/TH4/CH3 (30ml) Dt1049
TH5 (30ml) Dt1047
Hydrogen Peroxide 10-5000 Min41022
HP1 (30ml) Dt1051
HP2 (30ml) Dt1052
Molybdate 0-500 Min41027
MLB1 (30ml) Dt1440
MLB2 (30ml) Dt1441
Nitrite 0-2000 Min41014
N1 (30ml) Dt1072
N2 (30ml) Dt1073
pH 0-14 Min41016
Universal pH Solution (30ml) Dt1164
Phosphonate 0-20 Min41017
P1/4L (30ml) Dt1260
P5 (30ml) Dt1075
Sulphite 5-200 Min41018
S1 (20g) Dt1092
S2 (30ml) Dt1093
Tannin 10-300 Min41019
TN1 (20g) Dt1094
TN2 (30ml) Dt1095
QAC Cationics 0-1000 Min41020
QA1 (30ml) Dt1089
QA2 (30ml) Dt1090
QA3 (30ml) Dt1091
Zinc 0-2 Min41021
Z1 (30ml) Dt1096
Z4T (30pk) Dt1101
Z5 (30ml) Dt1097Individual Replacement Components & complete Replacement Packs are available for eachof the above parameters.
-
www.accepta.com 66
Section 9.0Comparator Test Kits
-
www.accepta.com 67
Section 9.1Comparator Test Kits
Acceptas high quality comparator test kits are available in a rangeof parameter specific options to suit most analysis requirements.
Each self contained kit is housed in its own rugged case and comescomplete with a high quality plastic comparator, comparator disk,reagents, liquids, tablets or powders; 2 x 10ml plastic cells, a tabletcrusher (if required) and an easy to follow waterproof instructions.
Each comparator test kit or replacement reagent pack includesreagents necessary to perform between 50 and 100 parameterspecific tests.
Parameter Range Order Code
Aluminium 0-0.3 Com81001
Aluminium Disk 256200
Aluminium No1 Tablets (50pk) Dt1147
Aluminium No2 Tablets (50pk) Dt1148
Amine 0-15 Com81060
Amine Disk 255005
Amine Indicator Solution (65ml) Dt1643
Amine Buffer Solution (65ml) Dt1644
Amine Solvent (100ml) Dt1645
Ammonia 0-1 Com81002
Ammonia Disk 256210
Ammonia No1 Tablets (50pk) Dt1245
Ammonia No2 Tablets (50pk) Dt1244
Bromine 0-5 Com81003
Bromine Disk 256280
DPD No1 Tablets (50pk) Dt1106
DPD Nitrite Powder (20g) Dt1295
Bromine (Liquid) 0-5 Com81003L
Bromine Disk 256280
DPD No1 Solution A (30ml) Dt1505
DPD No1 Solution B (30ml) Dt1506
DPD Nitrite Powder (20g) Dt1295
Carbohydrazide LR 0-0.65 Com81049
Carbohydrazide LR Disc 255011
Carbohydrazide LR Solution (65ml) Dt1654
Carbohydrazide LR Indicator (65ml) Dt1741
Individual Replacement Components & complete Replacement Packs are available for eachof the above parameters.
-
www.accepta.com 68
Section 9.1Comparator Test Kits (continued)Parameter Range Order CodeChlorine VLR 0-1 Com81004
Chlorine VLR Disc 256010
DPD No1 Tablets (50pk) Dt1106
DPD No3 Tablets (50pk) Dt1107
Chlorine VLR (Liquid) 0-1 Com81004L
Chlorine VLR Disc 256010
DPD No1 Solution A (30ml) Dt1505
DPD No1 Solution B (30ml) Dt1506
DPD No3 Solution C (30ml) Dt1507
Chlorine LR (Free & Total) 0-4 Com81005
Chlorine LR Disc 256020
DPD No1 Tablets (50pk) Dt1106
DPD No3 Tablets (50pk) Dt1107
Chlorine LR (Liquid) 0-4 Com81005L
Chlorine LR Disc 256020
DPD No1 Solution A (30ml) Dt1505
DPD No1 Solution B (30ml) Dt1506
DPD No3 Solution C (30ml) Dt1507
Chlorine HR (Total) 10-300 Com81006
Chlorine HR Disc 256030