ac/dc power ppwamd65 series 65w power · pdf file• medical en/ul60601 ... bf rated output...

2
Medical EN/UL60601-1 3rd Edition Approval (Pending), BF Rated Output Class II IEC320-C8 AC Inlet ErP Efficiency Level V Overload, Overvoltage and Short Circuit Protection Class II Double Insulated RoHS Compliant 100% Burn-In and Hi-Pot Testing AC/DC Power Your Partners in Power……... Specifications subject to change. PPWAMD65: May 9, 2014 PPWAMD65 Series 65W Power Supply Input Input Voltage Input Frequency Input Current Safety Isolation Inrush Current 90-264VAC 47-63 Hz 2A @ 90VAC, 1A @264VAC max. 4.0kVAC Input to Output 1.5kVAC Input to Ground <50A peak @ 115VAC, cold start at 25ºC <100A peak @ 230VAC, cold start at 25ºC Total Output Output Voltage Hold Up Time Turn On Time Efficiency 40W-65W, see table See table >16 msec @ full load and 115VAC/60Hz input <3 seconds >87% min @ 115~230Vac, 50/60(Except +5V model) Output Environmental & Operating Operating Temperature Storage Temperature Humidity MTBF: Compliance 0ºC to +60ºC , 100% power to 40ºC with derating to 50% power at 60ºC -10ºC to +70ºC 0% - 90% non-condensing >100,000 hours per MIL-HDBK-217F @ full load and at 25ºC Safety Approvals USA Canada Europe EMC: Harmonic Currents Voltage Flicker Electrostatic Discharge Radiated Immunity EFT/Burst Surge Immunity Conducted Immunity Magnetic Field Dips / Interruptions FCC Class B Radiated & Conducted EN55011: Class B Radiated & Conducted EN61000-3-2:2006+A1: 2009+A2:2009 IEC 61000-3-3: 2008 IEC 61000-4-2: 2009 IEC 61000-4-3: 2006+A1: 2008+A2: 2010 IEC 61000-4-4: 2004+A1: 2010 IEC 61000-4-5: 2006 IEC 61000-4-6: 2009 IEC 61000-4-8: 2010 IEC 61000-4-11: 2004 Power Partners, Inc. | 43 Broad Street | Hudson, MA 01749 Tel: (978)567-9600 | Fax: (978)567-9601 Website: www.powerpartners-inc.com Electrical Specifications Protection Overvoltage Short Circuit Overtemperature Overcurrent 120%-150% of rated voltage. Latching mode. Cycle AC for recovery Hiccup mode. Auto recovery. Optional 120%-150% of rated current. Auto- restart when fault removed. Desktop General Dimensions AC Input DC Output Plug Weight 4.17”L x 2.36”W x 1.18”H IEC320-C8, Class II Barrel Type: 5.5mm x 2.5mm x 10mm Center positive polarity (others upon request) 0.55 pounds UL60601-1 CSA C22.2 No. 601.1-M90 NEMKO EN60601-1 CB Report CE Mark Isolation: 4000VAC input to output, 2 x MOPP 1500 VAC input to ground,1 x MOPP

Upload: dinhdien

Post on 10-Feb-2018

212 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: AC/DC Power PPWAMD65 Series 65W Power · PDF file• Medical EN/UL60601 ... BF Rated Output • Class II IEC320-C8 AC Inlet • ErP Efficiency ... • 100% Burn-In and Hi-Pot Testing

• Medical EN/UL60601-1 3rd EditionApproval (Pending), BF Rated Output

• Class II IEC320-C8 AC Inlet• ErP Efficiency Level V• Overload, Overvoltage and Short

Circuit Protection• Class II Double Insulated• RoHS Compliant• 100% Burn-In and Hi-Pot Testing

AC

/DC

Po

wer

Your Partners in Power……...

Specifications subject to change. PPWAMD65: May 9, 2014

PPWAMD65 Series 65W Power Supply

Input

Input Voltage Input Frequency Input Current Safety Isolation

Inrush Current

90-264VAC 47-63 Hz 2A @ 90VAC, 1A @264VAC max. 4.0kVAC Input to Output 1.5kVAC Input to Ground <50A peak @ 115VAC, cold start at 25ºC <100A peak @ 230VAC, cold start at 25ºC

Total Output Output Voltage Hold Up Time

Turn On Time Efficiency

40W-65W, see table See table >16 msec @ full load and 115VAC/60Hz input <3 seconds >87% min @ 115~230Vac, 50/60(Except +5V model)

Output

Environmental & Operating

Operating Temperature

Storage Temperature Humidity MTBF:

Compliance

0ºC to +60ºC , 100% power to 40ºC with derating to 50% power at 60ºC -10ºC to +70ºC 0% - 90% non-condensing >100,000 hours per MIL-HDBK-217F @ full load and at 25ºC

Safety Approvals

USA Canada Europe

EMC:

Harmonic Currents Voltage Flicker Electrostatic Discharge Radiated Immunity EFT/Burst Surge Immunity Conducted Immunity Magnetic Field Dips / Interruptions

FCC Class B Radiated & Conducted EN55011: Class B Radiated & ConductedEN61000-3-2:2006+A1: 2009+A2:2009IEC 61000-3-3: 2008 IEC 61000-4-2: 2009 IEC 61000-4-3: 2006+A1: 2008+A2: 2010 IEC 61000-4-4: 2004+A1: 2010 IEC 61000-4-5: 2006 IEC 61000-4-6: 2009 IEC 61000-4-8: 2010 IEC 61000-4-11: 2004

Power Partners, Inc. | 43 Broad Street | Hudson, MA 01749 Tel: (978)567-9600 | Fax: (978)567-9601 Website: www.powerpartners-inc.com

Electrical Specifications

Protection

Overvoltage

Short Circuit Overtemperature Overcurrent

120%-150% of rated voltage. Latching mode. Cycle AC for recovery Hiccup mode. Auto recovery. Optional 120%-150% of rated current. Auto-restart when fault removed.

Des

kto

p

General Dimensions AC Input DC Output Plug

Weight

4.17”L x 2.36”W x 1.18”H IEC320-C8, Class II Barrel Type: 5.5mm x 2.5mm x 10mm Center positive polarity (others upon

request) 0.55 pounds

UL60601-1 CSA C22.2 No. 601.1-M90 NEMKO EN60601-1 CB Report CE Mark

Isolation: 4000VAC input to output, 2 x MOPP 1500 VAC input to ground,1 x MOPP

Page 2: AC/DC Power PPWAMD65 Series 65W Power · PDF file• Medical EN/UL60601 ... BF Rated Output • Class II IEC320-C8 AC Inlet • ErP Efficiency ... • 100% Burn-In and Hi-Pot Testing

Your Partners in Power……... Power Partners, Inc. | 43 Broad Street | Hudson, MA 01749 Tel: (978)567-9600 | Fax: (978)567-9601 Website: www.powerpartners-inc.com

Specifications subject to change. PPWAMD65: May 9, 2014

Models and Ratings Chart

PPWAMD65 Series 65W Power Supply

AC

/DC

Po

wer

Mechanical Outline

Des

kto

p

Note: 1. Ripple and noise are measured at oscilloscope 20MHz bandwidth by a 10uF electrolytic capacitor and a

0.1uF ceramic capacitor in parallel at output connector.

Note: 1. AC Inlet : IEC320-C82. Plastic Case Color : Black

3. Tolerance : ± 0.5mm

4. Unit : mm

Standard Output Cable

1 AWM 1185 # 18 AWG/1C + Shielding, UL 80ºC 300V VW-1, 1050 mm.

2 AWM 2464 # 18 / 20 AWG / 4C + SHIELDING, UL 80°C 300V VW-1, 1050 mm.