acquity in chromatography laboratory: analytical ... uplc... · analytical considerations ......
TRANSCRIPT
-
Andrew Aubin
Patricia McConville
ACQUITY UPLC in the Chromatography Lab: Analytical Considerations
-
2005 Waters Corporation
ACQUITY UPLC
ACQUITY UPLC in the Chromatography Lab: Analytical Considerations
Ultra Performance LC delivers improved speed, sensitivity and resolution that can benefit many liquid chromatography applications in today's laboratory. We will discuss analytical performance along with considerations and recommendations for successful implementation of UPLC in the chromatography laboratory.
-
2005 Waters Corporation
Why Advance the Science and Technology of LC?
Challenges in the LC Lab
Chromatography Resolution Reproducibility Ruggedness Sensitivity Speed Detector optimization
Business Complex samples Regulatory
considerations Increased # samples Time pressures Cost savings
Can you meet todays challenges?
-
2005 Waters Corporation
HPLC: 7 Analgesics in 25 minutes
AU
0.00
0.12
0.24
0.36
0.48
Minutes0.00 2.50 5.00 7.50 10.00 12.50 15.00 17.50 20.00 22.50 25.00
Pea
k1 -
5.47
0
Pea
k2 -
8.42
3
Pea
k3 -
9.94
9
Pea
k4 -
11.7
38
Pea
k5 -
16.4
10
Pea
k6 -
17.4
04
Pea
k7 -
19.0
37
ComponentsPeak #1 Acetaminophen Peak #2 - 2-AcetomidophenolPeak #3 - CaffeinePeak #4 - AcetanilidePeak #5 - Acetylsalicylic AcidPeak #6 - Salicylic AcidPeak #7 Phenacetin
Gradient Table
-
2005 Waters Corporation
UPLC:7 analgesics in 25 seconds
ComponentsPeak #1 Acetaminophen Peak #2 - 2-AcetomidophenolPeak #3 - CaffeinePeak #4 - AcetanilidePeak #5 - Acetylsalicylic AcidPeak #6 - Salicylic AcidPeak #7 Phenacetin
-
2005 Waters Corporation
Why Change from HPLC to UPLC?Time and solvent savings
Column Volume
(X5)
System Volume
(x3)
Re-equilibration time (min.)
Total cycle time
(min.)
# of injections in 24hours
Solvent consumed per injection
1.0 mL/min.
41
9931.3 mL/min
35.2mL
1.9mL
10.2.HPLC
4.6 x 100 525 min. run time
1.04
35.2
1.45
1.66mL
(8.3mL)
620uL
(1.86mL)
UPLC
2.1 x 50 1.7
0.41 min run time
0.20mL
(1.0mL)
120uL
(0.360mL)
Total Liters of solvent for 1000 injections:HPLC: 35.2L Vs. UPLC: 1.9L
-
2005 Waters Corporation
Why Change from HPLC to UPLC?More information about your sample
UPLC1.7 m
Peaks = 168Pc = 360
2.5X increase
AU
0.00
0.02
0.04
0.06
0.08
AU
0.00
0.02
0.04
0.06
0.08
Minutes
0.00 5.00 10.00 15.00 20.00 25.00 30.00 35.00 40.00 45.00 50.00 55.00 60.00
HPLC4.8 m
Peaks = 70Pc = 143
-
2005 Waters Corporation
Why Change from HPLC to UPLC?Improved sensitivity
100 pg/mL
Time1.00 2.00 3.00
%
3
Mass1 > Mass2796
1.722.08Rt = 2.08 min, Customer Compound A,
100pg/mL, 10L injection. Phenomenex column 2.0x30mm, 4umPeak Width = 6 secs
100 pg/mL
Time0.30 0.60 1.00
%
0
Mass1 > Mass21.15e4
0.83
0.54Rt = 0.83 min, Customer Compound A,100pg/mL, 10L injection. ACQUITY UPLC Column C18 2.1x50mm, 1.7umPeak Width = 1.8 secs
HPLC
UPLC
Method Comparison 100pg/mL, 10L injectionNicholas Ellor
-
2005 Waters Corporation
Speed, Sensitivity, ResolutionAU
0.00
0.10
0.20
0.30
0.40
0.50
0.60
Minutes0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50
HPLC
-
2005 Waters Corporation
HPLC TodayThe state of the science
The quality of the separation in HPLC is both driven and limited by column chemistry
Optimized chemistry can bring speed, sensitivity and resolution benefits to the chromatographic process
Particle size and selectivity are primary contributors to column performance
Todays classical HPLC instruments cannot maximize the reality of state of art smaller particles
-
2005 Waters Corporation
Smaller ParticlesThe enabler of productivity
Smaller particles provide: increased efficiency maintain efficiency over a
wider linear velocity ability for both added
resolution and increased speed of separation
Particles are central to the quality of the separation
van Deemter plot for classical HPLC
-
2005 Waters Corporation
Fundamental Resolution EquationAt constant column length
NRs
In UPLC systems, N (efficiency) is the primary driverAssuming the Selectivity and retentivity are the same as in HPLCThen, Resolution, Rs, is proportional to the square root of N
Rs = N4 ( -1 ) kk+1( )
System Selectivity RetentivityEfficiency
And, Efficiency (N), is inversely proportional to Particle Size , dp
dpN 1So:
dp 3X,Therefore if: Rs 1.7X N 3X, Then:
-
2005 Waters Corporation
Smaller ParticlesThe enabler of productivityThe promise of the van Deemter plot
-
2005 Waters Corporation
Summary of Key System Attributes(at constant column length)
Resolution Improvement
Speed Improvement
Sensitivity Improvement
Back Pressure
1.7 vs. 5 m particles 1.7X 3X 1.7X 27X1.7 vs. 3 m particles 1.3X 2X 1.3X 6X
-
2005 Waters Corporation
Influence of Particle Size on Chromatography
AU
0.000
0.010
0.020
0.030
0.040
0.050
Minutes
0.00 2.00 4.00 6.00 8.00 10.00 12.00 15.00
4.8 m, 0.2 mL/min, 354 psi
AU
0.000
0.010
0.020
0.030
0.040
0.050
Minutes0.00 1.00 2.00 3.00 4.00 5.00 6.00
Theory1.7X Resolution3X Faster1.7X Sensitivity25X Pressure
1.7 m, 0.6 mL/min, 7656 psi Actual1.5X Resolution2.6X Faster1.4X Sensitivity22X Pressure
Isocratic methods with 2.1 x 50 mm columns
-
2005 Waters Corporation
Optimal Linear Velocity on ACQUITY UPLCTM:Isocratic Caffeine Metabolites Separation
HPLCNon-Optimal Linear Velocity
UPLCTMOptimal Linear Velocity
AU
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
0.16
Minutes
0.00 0.50 1.00 1.50 2.00 2.50 3.00
ACQUITY UPLCTM BEH C182.1 x 50 mm, 1.7 m
Alliance 2695F = 0.3 mL/minPSIMAX = 4,200
T (4) = 1.63N (4) = 5,400
Rs (2,3) = 0.97
ACQUITY UPLCTM BEH C182.1 x 50 mm, 1.7 m
ACQUITY UPLCTMF = 0.6 mL/minPSIMAX = 8,400
T (4) = 1.02N (4) = 10,100Rs (2,3) = 2.25
AU
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
0.16
Minutes0.00 0.50 1.00 1.50 2.00 2.50 3.00
1
2 3
4
Same ACQUITY UPLC Column on HPLC vs UPLC instruments
-
2005 Waters Corporation
ACQUITY UPLCTM TechnologyUltra performance by design
Detectors:Optical and/or Mass SpecTunable UV or Photodiode ArrayOptimized flow cell for UPLCTMHigh speed data sampling for
Column Manager:Innovative pivot designPositions column to detectorE-CordTM connection
Sample Manager:Low dispersion XYZZ FormatFast cycle timesLow carryoverPlates and/or vialsOptional Sample Organizer
Binary Solvent Manager:High pressure blendingBinary gradientsFour solvent choicesOn-line degassingLow dispersion designUPLC pressure capabilities
Sample Organizer: (option)Expands capacity (22/15/8)Shuttles plate feedHeated/chilled
System Considerations:Small FootprintRedesigned tubing and fittingsConsolidated waste managementIntegrated system diagnosticsConnections InsightTM remote diagnostics
-
2005 Waters Corporation
UPLCUPLC Technology Challenges
How do we take advantage of smaller particles?
Pressure tolerant particle/column
High pressure fluidic modules
Holistic design, low system volume, optimized flow paths
Reduced cycle time autosampler with minimum carryover
High speed detectors; optical and mass
Controlled and coordinated system interaction Software designed for system integration Comprehensive diagnostic suite
-
2005 Waters Corporation
ACQUITY UPLCTMChemistry Proposition
Advancing the Science
Requires innovation in every aspect of column: New patented bridged hybrid particle required for pressure
tolerance and outstanding chromatographic performanceSub 2m particlesPorous for optimum mass transfer
Column hardwareNew patented frit technology to retain particles New end fittings for high pressure/low dispersion operation
Packing technologyNew column packing processes to optimize stability
eCordTMNew information chip to store column history
-
2005 Waters Corporation
Reversed-Phase Column Selectivity Chart
SunFire C18
YMC-Pack PolymerC18
Hypersil CPS Cyano
YMC-Pack CN
Waters Spherisorb S5 P
Hypersil BDS PhenylNova-Pak Phenyl
YMC-Pack Phenyl
Hypersil PhenylInertsil Ph-3
YMC-Pack Pro C4
YMCbasic
Symmetry C8YMC-Pack Pro C8
Nova-Pak C8
XTerra MS C18 Symmetry C18
YMC-Pack Pro C18
Inertsil ODS-3
YMC-Pack ODS-A
Nova-PakC18
YMC J'sphereODSL80 Nucleosil C18
Waters Spherisorb ODS2
Waters Spherisorb ODS1Resolve C18
Bondapak C18
YMC-Pack ODSAQ
YMC J'sphere ODSH80YMC J'sphere ODSM80
Inertsil CN-3
Waters Spherisorb S5CN
Nova-Pak CN HP
SymmetryShield RP8
SymmetryShield RP18
XTerra RP8
XTerra RP18
-0.6
-0.3
0
0.3
0.6
0.9
1.2
1.5
1.8
2.1
2.4
2.7
3
3.3
3.6
-1.5 -0.5 0.5 1.5 2.5 3.5
(ln [
] am
itrip
tylin
e/ac
enap
hthe
ne)
XTerra MS C8
Luna C18 (2)
ACQUITY UPLC BEH C18
XTerra Phenyl Luna
Phenyl Hexyl
ChromolithTM RP-18
Atlantis dC18
Zorbax XDB C18ACT Ace C18
Zorbax SB C18
SunFire C8
LunaC8 (2)
ACQUITY UPLC BEH C8
ACQUITY UPLC BEH Phenyl
(ln [k] acenaphthene)012005 ACQUITY UPLC Shield RP18
-
2005 Waters Corporation
ACQUITY UPLC Column Selectivity Comparison for Analgesics
C8
C18
RP18
Phenyl
1
ACQUITY UPLC BEH C18 1.7m ACQUITY UPLC BEH RP18 1.7m ACQUITY UPLC BEH C8 1.7m ACQUITY UPLC BEH Phenyl 1.7m
1. Acetaminophen2. 2-Acetomidophenol3. Caffeine4. Acetanilide5. Acetylsalicylic Acid6. Salicylic Acid7. Phenacetin
23
45 6
7
2
6
6
-
2005 Waters Corporation
Column Lifetime:>8,500 Injections
ConditionsColumn: ACQUITY UPLC BEH
C18 2.1 x 50 mmMobile Phase A: 0.1% HCOOHMobile Phase B: 0.1% HCOOH in ACNFlow Rate: 0.65 mL/minGradient: Time Profile
(min) %A %B Curve0.0 95 5 Initial 0.1 95 5 60.4 5 95 30.5 95 5 110.8 95 5
Injection Volume: 5 LTemperature: 40oCDetection: UV @ 254 nm
Pmax: 6,650 psi
-
2005 Waters Corporation
Column LifetimeCustomer Example
Isocratic aging experiment Cycle: low pressure runs/injections (data shown) 200/400/600 etc.
high pressure runs/injections (13,000 psi) low pressure runs/injections
This aging cycle was repeated
Columns lasted over 4,000 high pressure injectionsConditions (Isocratic Slow Flow/Low Pressure)Column: ACQUITY UPLC BEH C18 (2.1 x 50 mm, 1.7 m)
ACQUITY UPLC BEH Shield RP18 BEH (2.1 x 50 mm, 1.7 m) Part Number: 186002350Mobile Phase A: 0.1% HCOOH in WatersMobile Phase B: 0.1% HCOOH in ACNFlow Rate: 0.5 mL/min (~6,000 psi)Condition: 67% A / 33% BInjection Volume: 5.0 LSample Concentration: 208 g/mL ketorolac tromethamine and 13 g/mL naproxen in MeOHTemperature: 65oCDetection: UV 225 nm (Sample Rate: 10 Hz, Filter: Normal)
Conditions (Isocratic Fast Flow/High Pressure)Columns: ACQUITY UPLC BEH C18 (2.1 x 50 mm, 1.7 m)
ACQUITY UPLC BEH Shield RP18 (2.1 x 50 mm, 1.7 m) Mobile Phase A: 0.1% HCOOH in WaterMobile Phase B: 0.1% HCOOH in ACNFlow Rate: 1.250 mL/min (Pressure ~13,000 psi)Condition: 62% A / 38% BInjection Volume: 5.0 LSample Concentration: 208 g/mL ketorolac tromethamine and 13 g/mL naproxen in MeOHTemperature: 65oCDetection: UV 225 nm (Sample Rate: 10 Hz, Filter: Normal)
-
2005 Waters Corporation
Column LifetimeCustomer Example
First Injection on Column No high pressure runsN = 3483 for Peak 1 N = 6352 for Peak 2
Injection after4000 high pressure runs
N = 3573 for Peak 1N = 6466 for Peak 2
AU
- 0 .4 0
- 0 .2 0
0 .0 0
0 .2 0
0 .4 0
M in u te s0 .0 0 0 .2 0 0 .4 0 0 .6 0 0 .8 0 1 .0 0 1 .2 0 1 .4 0 1 . 6 0 1 .8 0 2 .0 0 2 .2 0 2 .4 0 2 .6 0 2 .8 0 3 .0 0
1.16
5
1.97
4
AU
-0.40
-0.20
0.00
0.20
0.40
Minutes0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00
1.16
8
1.97
8
AU
-0.40
-0.20
0.00
0.20
0.40
Minutes0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00
1.17
8
2.00
0
Ketorolac Naproxen
Chromatograms obtained under low (~6000 psi) isocratic conditions
Injection after1600 high pressure runs
N = 3659 for Peak 1N = 6489 for Peak 2
-
2005 Waters Corporation
pH Stability At The Extremes
What about chemical stability of the column at the extremes of pH, temperature and pressure?
A 2 x 2 experimental design (at UPLC pressures, of course):
30oC 60oCpH 2.0pH 11.3
-
2005 Waters Corporation
Column Stability at pH 2.0Direct Overlays
Minutes0.00 1.00 2.00 3.00 4.00
Minutes0.00 1.00 2.00 3.00 4.00
2000
1000
1
Injection
60oC8,100 psi
5.00
ACQUITY UPLC BEH C18, 2.1 x 50 mm30% ACN (v/v) with 0.1% TFA (pH 2.0) at 0.9 mL/min
Analytes: protriptyline, amitriptyline, butyrophenone
30oC11,400 psi
5.00
-
2005 Waters Corporation
Column Stability at pH=11.3
Minutes1.50 3.00 4.50
Inj. 1
Inj. 500
Inj. 1000
Minutes1.50 3.00 4.50
Acetonitrile-Pyrrolidine Buffer (pH=11.3) 45:55 (v/v) at 0.9 mL/min flow rateAnalytes: butyrophenone, protriptyline and amitriptyline
Inj. 1
Inj. 500
Inj. 750
60oC, 7,700 psi30oC, 11,500 psiOffset Overlays
-
2005 Waters Corporation
UPLCUPLC Technology Challenges
How do we take advantage of smaller particles?
Pressure tolerant particle/column
High pressure fluidic modules
Holistic design, low system volume, optimized flow paths
Reduced cycle time autosampler with minimum carryover
High speed detectors; optical and mass
Controlled and coordinated system interaction Software designed for system integration Comprehensive diagnostic suite
-
2005 Waters Corporation
ACQUITY UPLCTM TechnologyUltra performance by design
Detectors:Optical and/or Mass SpecTunable UV or Photodiode ArrayOptimized flow cell for UPLCTMHigh speed data sampling for
Column Manager:Innovative pivot designPositions column to detectorE-CordTM connection
Sample Manager:Low dispersion XYZZ FormatFast cycle timesLow carryoverPlates and/or vialsOptional Sample Organizer
Binary Solvent Manager:High pressure blendingBinary gradientsFour solvent choicesOn-line degassingLow dispersion designUPLC pressure capabilities
Sample Organizer: (option)Expands capacity (22/15/8)Shuttles plate feedHeated/chilled
System Considerations:Small FootprintRedesigned tubing and fittingsConsolidated waste managementIntegrated system diagnosticsConnections InsightTM remote diagnostics
-
2005 Waters Corporation
ACQUITY UPLCTMBinary Solvent Manager
Binary gradient Independent style piston drives Serial flow for each solvent, parallel
gradient formation High Pressure Gradient
Solvent select, 4 solvent choices On board solvent degasser 50l mixer, ~120L total system volume Ability to deliver non-linear gradients UPLC pressure capabilities
0-1mL/min to 15K psi, 1 -2mL/min to 9K psi
Dedicated transducer for each head. Enhances performance, provides diagnostic information
-
2005 Waters Corporation
BSM Components
Pump A Pump B
SSV A SSV BDegasserTee
FilterMixer
Vent Valve
PlungerWash
-
2005 Waters Corporation
From Serial to Parallel Flow
Two independently driven primary and accumulatorpiston drives Separate drive motors Two transducers Two inlet check valves
Four Solvents Capability A1 or A2 and B1 or B2 Two solvents per gradient
Six Channel Degasser 4 chromatographic eluents 2 solvents for Sample Manager
Strong & weak wash
B B A A
B2B2B1B1
A2A2
Solvent A1Solvent A1
-
2005 Waters Corporation
Pressure Envelope Fluidic components optimized to support the pressure
demands of UPLC separations Seals, head, bolts, check valves, fittings, tubing, mixer, pressure
transducers
UPLC compatible pressure with flow rate range
1 2
9000
15000
Pressure(psi)
Flow Rate(mL/min)
-
2005 Waters Corporation
Tee/ Filter/ Mixer
Glass Bead Mixing Technology Tortured Mixing Durable and will not disintegrate or affect delicate compounds Highly resistant to attack by most liquids and/or vapors.
Glass Bead Mixer Standard 4 x 5, approximate volume 50 L (P/N 700002631) Optional Mixer 6 x 6, approximate volume 100 L (P/N 205000252)
Filter Similar to Alliance, 2 micron filter (P/N 700002674)
-
2005 Waters Corporation
New Tubing & Fittings
HPLC Fittings Traditional drawn tubing
Traditional Waters Fittings
UPLCTM Fittings Center-less ground tubing
Gold Compression Screw 2 Piece Swagelock Ferrules
-
2005 Waters Corporation
ACQUITY 9/1000 SS316 Tubing
@75 Times
-
2005 Waters Corporation
Standard 5/1000 inch 316 SS Tuning
@75 Times @500 Times
-
2005 Waters Corporation
New ACQUITY 2/1000 316 SS Tubing
@75 Times @1000 Times
-
2005 Waters Corporation
ACQUITY UPLCTMBinary Solvent Manager gradient steps
N = 6
1% Steps 0% B to 10% BFlow Rate=0.50 mL/minuteBackpressure ~9300 psi
-
2005 Waters Corporation
ACQUITY UPLCTMBinary Solvent Manager Auto Blend
Injections of Caffeine with Dial-a-Mix at 1.0mL/min4.0% to 5.0% ACN in 0.1% Increments
4.0% ACN
5.0% ACN
-
2005 Waters Corporation
Original Assay Method from the USP for: Benzocaine, Butamben, and Tetracaine Hydrochloride Topical Solution
Mobile phase Prepare a mixture of methanol, water, and 0.25 M sodium 1-heptanesulfonate (500:500:20). Make adjustments if necessary (see System Suitability under Chromatography (621).
Chromatographic system (see Chromatography [621])The liquid chromatograph is equipped with a 313-nm detector, and a 3.9-mm 30-cm column that contains packing L1. The flow rate is about 2 mL per minute.
Standard preparation Transfer about 140 mg of USP Benzocaine RS, accurately weighed, to a 100-mL volumetric flask with the aid of 25 mL of methanol, and swirl. Transfer about 140 J mg of USP Butamben RS, accurately weighed, to the same volumetric flask with the aid of 25 mL of water, J being the ratio of the labeled amount, in percent, of butamben to the labeled amount, in percent, of benzocaine in the Topical Solution. Transfer about 140J mg of USP Tetracaine Hydrochloride RS, accurately weighed, to the same volumetric flask with the aid of 25 mL of water, J being the ratio of the labeled amount, in percent, of tetracaine hydrochloride to the labeled amount, in percent, of benzocaine in the Topical Solution. Sonicate for about 1 minute, dilute with Diluent to volume, and mix.
Procedure Separately inject equal volumes (about 10 L) of the Standard preparation and the Assay preparationinto the chromatograph, record the chromatograms, and measure the peak areas for the major peaks.
Chromatograph the Standard preparation, and record the peak responses as directed for Procedure: the relative retention times are about 0.3 for benzocaine, 0.8 for butamben, and 1.0 for tetracaine; the resolution, R, between the benzocaine peak and the butamben peak and between the butamben peak and the tetracaine peak is not less than 2; and the relative standard deviation for replicate injections is not more than 2.0% for each of the three analyte peaks.
http://www.uspnf.com/uspnf/pub/data/v28230/usp28nf23s0_alpha-18-140.xml#usp28nf23s0_c11rs113http://www.uspnf.com/uspnf/pub/data/v28230/usp28nf23s0_alpha-18-140.xml#usp28nf23s0_c11rs157http://www.uspnf.com/uspnf/pub/data/v28230/usp28nf23s0_alpha-18-1412.xml#usp28nf23s0_c11rs1149
-
2005 Waters Corporation
ACQUITY UPLCTMBSM Performance- RT Reproducibility
Standard Deviations of 35, 94, 125 milliseconds
-
2005 Waters Corporation
Alliance HPLCAnalytical Performance with isocratic conditions
AU
0.00
0.02
0.04
0.06
0.08
Minutes0.50 1.00 1.50 2.00 2.50 3.00 3.50
Column XTerra RP C18, 5 um, 4.6 X 250 mm Solvent A 10 mMol Ammonium Bicarbonate pH=10Solvent B AcetonitrileFlow Rate - 2.0 mL/minColumn Temperature 45 CIsocratic 40% A + 60% BDetection - PDA in 2D mode, 301 nm at 4.8 nm bandwidthData rate: 5 pps filtering constant 0.5Injection Volume - 2 uL of 0.07 mg/mL (5/100 dilution of USP Standard)
Average TR %RSD 0.133%Average Peak Area %RSD 0.467%Average Peak Width @ 4.4% 7.2 secAverage w Peak Width 3.2 sec.Minimum Resolution 2.9Total Cycle Time 351 secSolvent Used 11.7 mL
Separation of Benzocaine, Butamben, and Tetracaine
-
2005 Waters CorporationAU
0.00
0.10
0.20
0.30
0.40
0.50
0.60
Minutes0.05 0.10 0.15 0.20 0.25 0.30 0.35 0.40 0.45 0.50 0.55
Ben
zoca
ine
- 0.2
35
But
ambe
n - 0
.370
Tetra
cain
e - 0
.492
Column ACQUITY UPLCTM BEH C18 2.1 X 50 mm, 1.7 m Solvent A 10 mMol Ammonium Bicarbonate pH=10Solvent B AcetonitrileFlow Rate - 1.0 mL/minColumn Temperature 45 CIsocratic 50% A + 50% BDetection - PDA in 2D mode, 301 nm at 4.8 nm bandwidthData rate: 20 pps Filter constant: 0.1Injection Volume 2 uL of 0.07 mg/mL (5/100 dilution of USP Standard)
Average TR %RSD 0.370%
Average Peak Area %RSD 0.467%
Average 4 Sigma Peak Width 2.1 sec
Average w Peak Width 0.93 sec.
Minimum Resolution 4.2
Total Cycle Time 65 sec
Solvent Used 1.1 mL
USP Separation of Benzocaine, Butamben, and Tetracaine
ACQUITY UPLCTMAnalytical Performance with isocratic conditions
-
2005 Waters Corporation
Speed, Sensitivity, ResolutionAU
0.00
0.10
0.20
0.30
0.40
0.50
0.60
Minutes0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50
HPLC
-
2005 Waters Corporation
Changes to Original Assay Method from the USP for: Benzocaine, Butamben, and Tetracaine Hydrochloride Topical Solution
Mobile phase Prepare a mixture of acetonitrile, water containing 10 mMol ammonium bicarbonate at a pH of 10.0 (1:1) for UPLC, 4:6 for HPLC. Make adjustments if necessary (see System Suitability under Chromatography (621).
Chromatographic system (see Chromatography [621])The liquid chromatograph is equipped with a 301-nm detector, and a 4.6-mm 25-cm column that contains packing L1 (XTerra RP18, 5 um) for HPLC and a 2.1 X 50 mm 1.7 um for UPLC. The flow rate is about 1 mL per minute for UPLC and 2 mL per minute for HPLC.
Standard preparation Transfer about 140 mg of USP Benzocaine RS, accurately weighed, to a 100-mL volumetric flask with the aid of 25 mL of methanol, and swirl. Transfer about 140 J mg of USP Butamben RS, accurately weighed, to the same volumetric flask with the aid of 25 mL of water, J being the ratio of the labeled amount, in percent, of butamben to the labeled amount, in percent, of benzocaine in the Topical Solution. Transfer about 140J mg of USP Tetracaine Hydrochloride RS, accurately weighed, to the same volumetric flask with the aid of 25 mL of water, J being the ratio of the labeled amount, in percent, of tetracaine hydrochloride to the labeled amount, in percent, of benzocaine in the Topical Solution. Sonicate for about 1 minute, dilute with Diluent to volume, and mix. Further dilute this stock 5 in100 for UPLC use.
Procedure Separately inject equal volumes (about 2 L) of the Standard preparation and the Assay preparationinto the chromatograph, record the chromatograms, and measure the peak areas for the major peaks.
Chromatograph the Standard preparation, and record the peak responses as directed for Procedure: the relative retention times are about 0.3 for benzocaine, 0.8 for butamben, and 1.0 for tetracaine; the resolution, R, between the benzocaine peak and the butamben peak and between the butamben peak and the tetracaine peak is not less than 2; and the relative standard deviation for replicate injections is not more than 2.0% for each of the three analyte peaks.
http://www.uspnf.com/uspnf/pub/data/v28230/usp28nf23s0_alpha-18-140.xml#usp28nf23s0_c11rs113http://www.uspnf.com/uspnf/pub/data/v28230/usp28nf23s0_alpha-18-140.xml#usp28nf23s0_c11rs157http://www.uspnf.com/uspnf/pub/data/v28230/usp28nf23s0_alpha-18-1412.xml#usp28nf23s0_c11rs1149
-
2005 Waters Corporation
ACQUITY UPLCTMAnalytical Performance with Gradient conditions
AU
-0.10
0.00
0.10
0.20
0.30
0.40
0.50
0.60
0.70
Minutes0.00 0.05 0.10 0.15 0.20 0.25 0.30 0.35 0.40 0.45 0.50
Ace
tani
lide
- 0.
176
Ace
toph
enon
e -
0.23
0
Pro
piop
heno
ne -
0.27
3
But
yrop
heno
ne -
0.31
0B
enzo
phen
one
- 0.3
22
Val
erop
heno
ne -
0.34
6
Hex
anop
heno
ne -
0.3
83
Hep
tano
phen
one
- 0.
420
Oct
anop
heno
ne -
0.4
56
Average TR %RSD 0.109%Average Peak Area %RSD 0.368%
Average 4 Sigma Peak Width 0.369 secAverage w Peak Width 0.210 secMinimum Resolution 2.15Total Cycle Time 56 secMaximum Pressure 13,500 psi
Column: 2.1 x 50 mm @ 65CMobile Phase A: 0.1% Formic Acid in WaterMobile Phase B: 0.1% Formic Acid in ACNFlow Rate: 1.3 mL/minGradient: 30% - 90% B over 0.5minutes, Curve 5Wash Solvents: Strong 50/50 Water/ACN (200L);
Weak 95/5 Water/ACN 600L)Wavelength: 240nmData Rate: 40 pts/s, Filter Constant: 0.0
6 injections9 phenones(3L, 0.1 mg/mL)30 sec. gradient
-
2005 Waters Corporation
ACQUITY UPLCTMAnalytical Performance with Gradient conditions
AU
-0.10
-0.05
0.00
0.05
0.10
0.15
0.20
0.25
0.30
0.35
0.40
Minutes0.00 0.05 0.10 0.15 0.20 0.25 0.30 0.35 0.40 0.45 0.50
Ace
tam
inop
hen
- 0.2
08
2-A
ceto
mid
ophe
nol -
0.2
25
Caf
fein
e - 0
.264
Ace
tani
lide
- 0.3
06
AS
A -
0.32
3
Phe
nace
tin -
0.35
8
Sal
icyl
ic A
cid
- 0.3
80
10 Replicate Injections of a7 component Analgesic Mix
(5L, 0.1 0.2 mg/mL in water)
Column: 2.1x50mm @ 65CMobile Phase A: 0.1% Formic Acid in WaterMobile Phase B: 0.1% Formic Acid in ACNFlow Rate: 1.3 mL/minGradient: Time %A %B Curve
Initial 93 7 -0.10 80 20 40.18 65 35 60.41 93 7 6
Wash Solvents: Strong 50/50 Water/ACN, 100LWeak 98/2 Water/ACN, 500L
Wavelength: 240nmData Rate: 40 pts/s, Filter Constant: 0.0
Ave RT %RSD 0.115%Ave Area %RSD 0.133%Minimum Resolution 1.70Total Cycle Time 56 sec
Maximum Pressure 11,000 psi
-
2005 Waters Corporation
ACQUITY UPLCNon-Linear Gradient Capability
Gradient Profiles on ACUITY UPLC
-
2005 Waters Corporation
Utility of Non-linear gradient capability
Method Development Challenge:
Gradient elution of 9 component mix To go as fast as possible
To meet a minimum resolution requirement of 2.0
Start with generic method 30s gradient with 15s hold at final conditions. (5% to 95% ACN)
Run gradient using curves 3-9
-
2005 Waters Corporation
Chromatograms of Curves 3-9A
U
0.00
0.50
AU
0.00
0.50
AU
0.00
0.50
AU
0.00
0.50
AU
0.00
0.50
AU
0.00
0.50
AU
0.00
0.50
Minutes0.00 0.05 0.10 0.15 0.20 0.25 0.30 0.35 0.40 0.45 0.50 0.55 0.60 0.65 0.70 0.75
Curve 9
Curve 8
Curve 7
Curve 6
Curve 5
Curve 4
Curve 3
-
2005 Waters Corporation
Run Time Results and Minimum Resolution
Curve 3 4 5 6 7 8 9
Resolution 1.32 1.76 2.17 3.01 3.59 3.61 3.36
Elution Time of Last Peak
0.34 0.40 0.46 0.57 0.64 0.66 0.68
Shortest Run Time with required resolution is using Curve 5
-
2005 Waters Corporation
ACQUITY UPLCTMSample Manager
Pressure tolerant, low volume flow-through isolation & injection valves
Fast cycle times, low carry over
2 Plate capacity, multiple formats Well plates, microtube plates, vials
Thermal control (4 - 40C)
Low dispersion column management
Sample Organizer (optional)
New Technologies - all-new injection process
-
2005 Waters Corporation
Pod Style Injector
Modular design (removable Pod) permits easy access
-
2005 Waters Corporation
ACQUITY UPLC Loops available
2 L5 L10 L20 L50 L
-
2005 Waters Corporation
ACQUITY UPLCTMSample Manager Performance - Precision
Component Summary Area RT ReportProject Name: PITTCON_2005Reported by User: Andrew Aubin (Aubin)
1
2
3
4
5
6
7
8
9
10
Mean
Std. Dev.
% RSD
SampleName Date Acquired Inj Vol(ul) Inj Vial 3,4-MDA 2,3-MDA 3,4-MDMA 2,3-MDMA
2,3+3,4 MDA MDMA
2,3+3,4 MDA MDMA
2,3+3,4 MDA MDMA
2,3+3,4 MDA MDMA
2,3+3,4 MDA MDMA
2,3+3,4 MDA MDMA
2,3+3,4 MDA MDMA
2,3+3,4 MDA MDMA
2,3+3,4 MDA MDMA
2,3+3,4 MDA MDMA
2/14/2005 10:28:57 AM
2/14/2005 10:33:42 AM
2/14/2005 10:36:38 AM
2/14/2005 10:39:35 AM
2/14/2005 10:42:31 AM
2/14/2005 10:45:27 AM
2/14/2005 10:48:24 AM
2/14/2005 10:51:20 AM
2/14/2005 10:54:14 AM
2/14/2005 10:57:11 AM
5.00
5.00
5.00
5.00
5.00
5.00
5.00
5.00
5.00
5.00
1
1
1
1
1
1
1
1
1
1
2:D,1
2:D,1
2:D,1
2:D,1
2:D,1
2:D,1
2:D,1
2:D,1
2:D,1
2:D,1
252746
251676
252557
251382
251354
250666
252694
252331
252360
252659
252043
722
0.29
169926
169802
169395
169090
168659
169286
168439
169251
168626
168037
169051
604
0.36
235323
235335
235366
232696
234721
234564
235468
235365
234133
234904
234787
856
0.36
164552
163520
163371
162642
163114
163830
163120
163195
164507
162978
163483
636
0.39
Component Summary For Area
-
2005 Waters Corporation
ACQUITY UPLCTMSample Manager Performance - Linearity
Are
a
-50000
0
50000
100000
150000
200000
250000
Amount0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00
R2 = 0.9998
0.1 mg/mL Acetaminophen with 20L loop in Pressure-Assist Mode
-
2005 Waters Corporation
ACQUITY UPLCTMSample Manager Performance -Carryover
10 mg/mL Human Insulin
Carryover 0.0036%
Wash Conditions: 200L Strong (6:3:1 / 0.1% phosphoric in Water:ACN:IPA) 600L Weak (0.01M HCl)
-
2005 Waters Corporation
ACQUITY UPLCTMSample Manager Performance -Carryover
Carryover 0.0036%
Wash Conditions: 200L Strong (6:3:1 / 0.1% phosphoric in Water:ACN:IPA) 600L Weak (0.01M HCl)
-
2005 Waters Corporation
ACQUITY UPLC Minimized CarryoverAlprazolam in Plasma
Plasma Blank
314.1>210.2
260.2>116.0
309.0>205.2
1000 ng/mL
Peak Area96015
Peak Area747309
Peak Area97963
MRM transition
Plumb
-
2005 Waters Corporation
Cycle Times?
ACQUITY UPLC Injection-to-Injection Cycle TimesInjection-to-injection cycle times are calculated as the time it takes for the injector to perform all of the tasks during the injection cycle from the point of initiation by the data acquisition software until the valve turns and the injection is made. To determine this value a series of injections are monitored by a detector that is continually collecting data, even between the run times. The difference in time between corresponding peaks of each injection gives the total cycle time. Subtracting the run time from the total cycle time will yield the injection-to-injection cycle time.
Injection Interval Total Cycle Time(sec)
Run Time(sec)
Injector Cycle Time(sec)
1 64.9 36.0 28.9
2 65.1 36.0 29.1
3 65.0 36.0 29.0
4 64.1 36.0 28.1
5 63.0 36.0 27.0
6 65.5 36.0 29.5
7 67.0 36.0 31.0
8 66.9 36.0 30.9
9 65.2 36.0 29.2
Average 65.2 36.0 29.2
-
2005 Waters Corporation
ACQUITY UPLCTMColumn Manager
Column Manager
Symbiotic with Sample Manager Flow path distance optimized 65C upper limit Up to 150 mm column
Pivot positioning Stacked mode
w/ optical detector Swung out mode
w/ MS detector only Column may be accessed from
either side
-
2005 Waters Corporation
Attaching The Column to the System
High Pressure Finger Tight FittingsHolds >15000 psi
Patent Pending
-
2005 Waters Corporation
Column Outlet
- One piece finger tight fittings work well- Cut 0.0025 Peek to length- Keep as short as possible to minimize dispersion
-
2005 Waters Corporation
Column Heating
Column heating considerations Consistent temperature for RT stability Solvent viscosity is lower at higher temperatures Lower backpressures at higher temperatures
AU
0.00
0.05
0.10
0.15
0.20
0.25
0.30
Minutes1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80
60 C
50 C55 C
45 C40 C
35 C
30 C
-
2005 Waters Corporation
ACQUITY UPLCTMColumn Manager with eCord
eCordTM Technology
eCordeCordreaderreader Column Information
Certificate of Analysis
Column Usage History Total number of samples Injections Pressure/Temperature history And much more
eCordTM
-
2005 Waters Corporation
ACQUITY UPLCTMeCordTM
Columns
eCordTM
-
2005 Waters Corporation
eCordColumn QC
-
2005 Waters Corporation
ACQUITY UPLCTMDetectors
Faster eluting peaks require higher data rates High light throughput/transmission Faster digital filter time constants
UPLC separations require low dispersion flow cells Maintain chromatographic fidelity
Tunable UV, PDA and ELSD Options
MS Options
-
2005 Waters Corporation
ACQUITY UPLCTMFlow Cell Design
Light Guided UPLCTM flow cells 10 mm pathlength, 500 nL volume Flow cell channel is the inside of a low-index Teflon AF tube Total internal reflection at walls, like optical fiber cladding
Teflon AF
water
dTeflon AF
-
2005 Waters Corporation
ACQUITY UPLCTMLow dispersion UPLC optics
Time2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00 30.00
%
0
100
%
0
100 0.67
22.92
0.78
2.09
1.30
11.208.62
2.57
3.55 4.205.615.15 7.336.49 7.93
21.6621.10
20.3117.13
14.5112.71 16.32
19.4318.26
22.23
26.01
23.15
24.40
25.17
26.51
27.34
28.7728.12
0.69 23.050.81
2.58
1.37
11.62
9.11
6.135.864.883.734.36
7.06 7.678.38 9.99
21.7321.1918.3817.3214.78
13.6512.96 16.6015.39
19.5020.40
22.30
23.30
26.01
23.5124.45
25.8425.09
26.61
27.32
27.5129.18
UV cell in line75/500/75
No UV cell
Column2.1x100
400L/min
Effect of UV Cell on MS Data
-
2005 Waters Corporation
Low dispersion UPLCTM opticsExpanded Region
Time10.80 11.00 11.20 11.40 11.60 11.80 12.00 12.20
%
0
100
%
0
100
11.20
11.62
11.2810.93
No UV cell
UV cell in line
0.43mins
0.43mins
-
2005 Waters Corporation
Data Acquisition RatesImpact on UV chromatography data
1 pt/s
2 pt/s
5 pt/s
10 pt/s
20 pt/s
40 pt/s
Minutes0.50 0.52 0.54 0.56 0.58 0.60 0.62 0.64 0.66 0.68 0.70 0.72 0.74
-
2005 Waters Corporation
Effect of Filtering Constant Effect of Filtering Constant on Chromatographyon Chromatography
AU
0.000
0.010
0.020
0.030
0.040
0.050
0.060
0.070
0.080
Minutes0.50 0.52 0.54 0.56 0.58 0.60 0.62 0.64 0.66 0.68 0.70 0.72 0.74 0.76 0.78 0.80
No Filtering0.1s0.3s0.5s1.0s
5 component Analgesic Mix on 2.1x50mm ACQUITY UPLC Column, 30s gradient
-
2005 Waters Corporation
UV Spectra on ACQUITY PDA
-
2005 Waters Corporation
Peak Purity and Library Match Capabilities
-
2005 Waters Corporation
ACQUITY UPLCTMOptional Sample Organizer
Supports high throughput/high capacity
Expands capacity-mix or match 22 thin (microtitre) plates
8,448 samples max 15 medium height plates 8 full height (vial) plates
384 x 2mL vials
Temperature controlled
Programmable Access to any sample location
Supports Walk-up Operation
-
2005 Waters Corporation
ACQUITY UPLCTMOptional FlexCart
Supports inlet mobility
-
2005 Waters Corporation
UPLCUPLC Technology Challenges
How do we take advantage of smaller particles?
Pressure tolerant particle/column
High pressure fluidic modules
Holistic design, low system volume, optimized flow paths
Reduced cycle time autosampler with minimum carryover
High speed detectors; optical and mass
Controlled and coordinated system interaction Software designed for system integration Comprehensive diagnostic suite
-
2005 Waters Corporation
System Information
-
2005 Waters Corporation
Sample Syringe Diagnostic
-
2005 Waters Corporation
Needle Seal - Static vs Dynamic TEST
Static Dynamic Static Dynamic
-
2005 Waters Corporation
ACQUITY UPLCTMDesigned with Intent
UPLCTM enabled, laboratory rugged
Convenience AQT Qualification Small footprint Integrated waste management eCordTM Technology Intuitive Control Console Multiple plate formats High capacity sample organizer FlexCart
Performance and Reliability UPLC pressure capabilities Low dispersion fluidics Rapid, clean injections Fast detection Rugged chemistry Analytical performance Extensive onboard diagnostics Connections InsightTM
UPLCTM = a higher quality of (more) information faster
-
2005 Waters Corporation
QUESTIONS?
-
2005 Waters Corporation
Acknowledgements
John Morawski
Tanya Jenkins
Eric Grumbach
Robert Plumb
Jeffrey Mazzeo
ACQUITY UPLC in the Chromatography Lab: Analytical ConsiderationsACQUITY UPLCWhy Advance the Science and Technology of LC?HPLC: 7 Analgesics in 25 minutesUPLC:7 analgesics in 25 secondsWhy Change from HPLC to UPLC?Time and solvent savingsWhy Change from HPLC to UPLC?More information about your sampleWhy Change from HPLC to UPLC?Improved sensitivityUPLC method verses customers HPLC methodSpeed, Sensitivity, ResolutionHPLC TodayThe state of the scienceSmaller ParticlesThe enabler of productivityFundamental Resolution Equation At constant column lengthSmaller ParticlesThe enabler of productivitySummary of Key System Attributes(at constant column length)Influence of Particle Size on ChromatographyOptimal Linear Velocity on ACQUITY UPLCTM:Isocratic Caffeine Metabolites SeparationACQUITY UPLCTM TechnologyUltra performance by designUPLC Technology ChallengesACQUITY UPLCTMChemistry PropositionReversed-Phase Column Selectivity ChartACQUITY UPLC Column Selectivity Comparison for AnalgesicsColumn Lifetime:>8,500 InjectionsColumn LifetimeCustomer ExampleColumn LifetimeCustomer ExamplepH Stability At The ExtremesColumn Stability at pH 2.0Column Stability at pH=11.3UPLC Technology ChallengesACQUITY UPLCTM TechnologyUltra performance by designACQUITY UPLCTMBinary Solvent ManagerBSM ComponentsFrom Serial to Parallel FlowPressure EnvelopeTee/ Filter/ MixerNew Tubing & FittingsACQUITY 9/1000 SS316 TubingStandard 5/1000 inch 316 SS TuningNew ACQUITY 2/1000 316 SS TubingACQUITY UPLCTMBinary Solvent Manager Auto BlendACQUITY UPLCTM BSM Performance- RT ReproducibilityAlliance HPLCAnalytical Performance with isocratic conditions Speed, Sensitivity, ResolutionACQUITY UPLCTMAnalytical Performance with Gradient conditionsACQUITY UPLCTMAnalytical Performance with Gradient conditionsACQUITY UPLC Non-Linear Gradient CapabilityUtility of Non-linear gradient capabilityChromatograms of Curves 3-9Run Time Results and Minimum ResolutionACQUITY UPLCTMSample ManagerPod Style InjectorACQUITY UPLC Loops availableACQUITY UPLCTM Sample Manager Performance - PrecisionACQUITY UPLCTM Sample Manager Performance - LinearityACQUITY UPLCTM Sample Manager Performance -CarryoverACQUITY UPLCTM Sample Manager Performance -CarryoverACQUITY UPLC Minimized CarryoverAlprazolam in PlasmaCycle Times?ACQUITY UPLCTMColumn ManagerAttaching The Column to the SystemColumn OutletColumn HeatingACQUITY UPLCTMColumn Manager with eCordACQUITY UPLCTMeCordTMeCordColumn QCACQUITY UPLCTM DetectorsACQUITY UPLCTMFlow Cell DesignACQUITY UPLCTMLow dispersion UPLC opticsLow dispersion UPLCTM opticsExpanded RegionData Acquisition Rates Impact on UV chromatography dataEffect of Filtering Constant on ChromatographyUV Spectra on ACQUITY PDAPeak Purity and Library Match CapabilitiesACQUITY UPLCTM Optional Sample OrganizerACQUITY UPLCTMOptional FlexCartUPLC Technology ChallengesSystem InformationSample Syringe DiagnosticACQUITY UPLCTMDesigned with IntentQUESTIONS?Acknowledgements