adam-scc - montreal biotech · adam-scc. scc = 4,770,000 cell/ml. dna labeling cell count: pi...
TRANSCRIPT
NanoEnTek Inc.
ADAM-SCCAuto-microscopic Somatic Cell Counter
www.nanoentek.com
• Automated fluorescent microscope + Disposable micro plastic chip• Compatible to the Direct Microscopic Somatic Cell Count•Test tool kit can be provided for user’s convenience• Composed of Device, Somatic Cell count Kit
ADAM-SCC
Somatic Cell count Kit
www.nanoentek.com
DMSCC, 1000 × ADAM-SCC
SCC = 4,770,000 cell/ml
DNA labeling Cell Count: PI staining method• Somatic cell count• Cell viability test / mammalian cell count
Real-time live image view• 4 × objectives, Fluorescent images
www.nanoentek.com
200 400 600 800 1000
200
400
600
800
1000 y = 1.045x - 14.038
R2 = 0.99
C-R
eade
r (X1
,000
/ml)
Eastern Lab. Standard solution (X1,000/ml)200 300 400 500 600 700
200
300
400
500
600
700 y = 1.04x - 12.16
R2 = 0.99
C-R
eade
r (X1
,000
/ml)
Standard Sol. (X1,000/ml)
Correlation between 12 standard somatic cell solution (Eastern Lab., USA) and ADAM-SCC
Correlation between 3 kind standard somatic cell solution (NVRQS, Korea) and ADAM-SCC
Validation & Verification
Accuracy for somatic cell standard solution
By National Veterinary Research Quarantine Service (Korea)
www.nanoentek.com
Functional Comparison of Competitors
Feature ADAM-SCC DMSCCHigh-Throughput
EquipmentD Company
Process Time Fast Very slow Very Fast Fast
Personal Error No Yes No Yes
Method Similar Standard Method Standard Method Flow cytometry Similar Standard Method
Repeatability <5% at 100,000 / mL - <5% at 100,000 / mL <12% at 100,000 / mL
Sample Prep. Simple Complex Simple Simple
Measure Cell size Yes Yes No No
View Cell Morphology Yes Yes No No
Calibration No - Need No
Cell Types Somatic Cell Most Cells Somatic Cell Somatic Cell
Throughput High Low High High
Device Price Low - High Low
Peripheral Device Chip, Stain Solution Microscope, Reagent Many Cassette, Lysis Buffer
Device Size Portable - Big Small
Maintenance Coast Low - High Low