aespire 7100 - ge healthcare worldwide/media/downloads/us/product/... · aespire 7100 exceptional...

Post on 16-Apr-2018

298 Views

Category:

Documents

9 Downloads

Preview:

Click to see full reader

TRANSCRIPT

Features• Enhancedmonitorintegrationcapabilitieswithour

Cardiocap*/5andDash*seriesofmonitors

• Largecolorventilatordisplaywithcolorwaveformandalarmmessageindicator

• Lightweightandcompactforeasymaneuverability

• OptionalintegratedauxiliaryO2flowmeterandsuctioncontrol

Advanced Breathing System (ABS)• Onestepbag/ventswitchturnstheventilatoron/off

• Minimalnumberofpartsandtubeconnectionsmayhelptoreducethepotentialforleaksandmisconnects

• Easeofdisassembly(notools)

• Autoclavableandlatex-free

7100 Ventilator• VolumeandPressuremodeswithelectronicPEEP

• Exhaledvolume,airwaypressureandinspiredoxygenmonitoringcapabilities

• Directaccesstoventilatorparametersettings

• Smartalarmsdirectusertospecificproblemsandaffectedparameters

• Pressurebargraphforvisualreferenceonabreath-by-breathbasis(optionalpressurewaveformavailable)

Aespire*shownwithCardiocap*/5monitorandTec*7Vaporizers

Aespire7100ExceptionalperformanceCompactdesign

Improved low flow/reduced life cycle costs• Onlyonescheduledmaintenancecheckperyear

• Freshgasflowcompensation–automatically(availablewithtidalvolumecompensationoption)

• MinimumO2flowof50mL

• Dualairflowtubesstandardforhigherresolutionoflowflows

GEHealthcare

Physical specifications

Dimensions

Height: 134.5cm/52.9in

Width: 72cm/28.3in

Depth: 73cm/28.7in

Weight: Approximately108kg/238lbs

Top shelf

Weightlimit: 34kg/75lbs

Width: 66cm/26in

Depth: 40cm/15.75in

Work surface

Height: 81.7cm/32.2in

Size: 2160cm2/334in2

DIN rail

Sideofmachine: 34.5cm/13.6in

Drawers (internal dimensions)

Height: 17.5cm/6.9in

Width: 33cm/13in

Depth: 26.5cm/10.4in

Absorber bag arm (optional)

Armlength: 30.5cm/12in

Bagarmheight(adjustable): 87cm/34.3in 104cm/40.9in

Casters

Diameter: 12.5cm/5in

Brakes: Individuallocking frontcasters

Ventilator operating specifications

Positive End Expiratory Pressure (PEEP)

Type: Integrated,electronicallycontrolled

Range: OFF,4to30cmH2O (incrementsof1cmH2O)

Ventilator monitored values

Tidalvolume: 5to1500mL,1mLresolution

Minutevolume: 0to99.9L/min,0.1L/minresolution

Breathingrate: 0to65breathsperminute, 1breathperminuteresolution

Oxygenpercentage: 5%to110%,1%resolution

Airwaypressure:–9to99cmH2O,1cmH2Oresolution

Alarm settings

Tidalvolume(VTE): Low:OFF,5to1500mL High:20to1600mL,OFF

Minutevolume(VE): Low:OFF,0.1to10L/min High:0.5to30L/min,OFF

Inspiredoxygen (FiO2): Low:18to100% High:21to100%,OFF

Apneaalarm: MechanicalventilationON:<5mL breathmeasuredin30seconds

MechanicalventilationOFF:<25mL breathmeasuredin30seconds

Lowairwaypressure: Changeof<4cmH2OabovePEEP

Pressure(Plimit)range: 12to99cmH2O (incrementsof1cmH2O)

Sustainedairwaypressure:6to30cmH2O+PEEP (adjustedbasedonventilatorsettings)

Subatmosphericpressure: Paw<–10cmH2O

Alarmsilencecountdowntimer: 120to0seconds

Modes of ventilation

VolumeControlmode Withtidalvolumecompensation(optional)

Pressuremode(optional)

Ventilation parameters

Tidalvolumerange: 45to1500mL (VolumeControlmode)

Incrementalsettings: 45to100mL (incrementsof5mL)

100to300mL (incrementsof10mL)

300to1000mL (incrementsof25mL)

1000to1500mL (incrementsof50mL)

Pressure (PInspired)range: 5to50cmH2O (incrementsof1 cmH2O)

5to1500mL volumedelivery

Rate: 4to65breathsper minute(increments of1breathperminute)

Inspiratory/expiratoryratio: 2:1to1:6 (incrementsof0.5)

Inspiratorypauseadjust: 5%to60%of inspiratorytime (incrementsof5%)

Colorventilatorydisplay

Ventilator accuracy

Delivery/monitoring accuracy

Volumedelivery: >200mL=betterthan±10% SetTV

75to200mL=betterthan±20mL

<75mL=betterthan±15mL

Pressure(PInspired)deliveryrepeatability: ±2cmH2O

PEEPdeliveryrepeatability: ±2cmH2O

Volumemonitoring: >200mL=betterthan±10%

75to200mL=betterthan±20mL

<75mL=betterthan±15mL

Pressuremonitoring: Betterthan±2cmH2Oand±5% ofreading(whicheverisgreater)

Ventilator components

Flow transducer

Type: Variableorificeflowsensor

Dimensions: 22mmODand15mmID

Location: Inspiratoryoutletand expiratoryinlet

(Optionalautoclavablesensoravailable)

Oxygen sensor

Type: Galvanicfuelcell

Ventilator pneumatics

Pressurerangeatinlet: 240kPato700kPa/35psigto100psig

Peakgasflow: 70L/min+freshgasflow

Flowrange: 2to70L/min

Flowcompensationrange: 200mL/minto15L/min

Ventilator screen

Displaysize: 120mmx92mm

Displaydensity: 1/4colorVGA

Battery back-up

Backuppower: Demonstratedbatterytime undertypicaloperating conditionsis90+minutes whenfullycharged.Battery timeunderextremeconditions is30minutes.

Batterytype: Internalrechargeablesealed leadacid

Communication port

Serialinterface: IsolatedRS-232C compatibleport

Anesthetic agent delivery

Delivery

Vaporizers: Tec*5,Tec6Plus,Tec7

Numberofpositions: 2

Mounting: Tool-freeinstallation Selectatec*manifold interlocksandisolates vaporizers

Tec6Plusvaporizer Tec7vaporizer

Electrical specifications

Current leakage

100/120V: <300µA

Power

Powerinput: 100-120Vac,50/60Hz

Powercord: Length:5m/16.4ft Rating:15A@120Vac

Inlet/outlet modules

Systemcircuit breakers: 15A

Outlets 4outletsonback,3-2A,1-3Aindividual(optional): breakers,optionalisolationtransformer

Pneumatic specifications

Auxiliary common gas outlet

Connector: ISO22mmODand15mmID

Gas supply

Pipelineinputrange: 240kPato600kPa/35psigto88psig

Cylinderinput: Pinindexedinaccordancewith CGA-V-1orDIN(nutandgland); containsinputfilterandcheckvalve

Note:Maximum3cylinders; twoinboardmounted,one outboardmounted.

Primaryregulatordiaphragmminimumburstpressure: 2758kPa/400psig

Primaryregulatornominaloutput: <338kPa/49psig Pinindexedcylinder connections

O2 controls

Method: Proportionatedecreaseof N2OwithreductioninO2pressure

Supplyfailurealarm: Range:193kPato221kPa/ 28psigto32psig

Soundsatmaximumvolume every10seconds

O2flush: Range:25to75L/min

Flowmeters

O2ranges: 0.05to0.95L/min and1.0to15.0L/min; MinimumO2flow: 50mL/min±25mL

N2Oranges: 0to0.95L/minand 1.0to10.0L/min

Airrange: 0to0.95and1to15L/min

Pneumatic specifications, continuedCalibration: Percentof Accuracy fullscaleflow (%offlowrate)

100 ±2.5% 90 ±2.5% 80 ±2.6% 70 ±2.7% 60 ±2.9% 50 ±3.1% 40 ±3.4% 30 ±4.0% 20 ±5.0% 10 ±8.1%

Calibrationconditions:* 20°C/68°F,101.3kPa/760mmHg

* Differentbreathingcircuitpressures,barometricpressures ortemperatureschangeflowtubeaccuracy.

Hypoxic guard system

Type: MechanicalLink-25*

Range: Providesanominalminimum25% concentrationofoxygenin O2/N2Omixture

Materials

Allmaterialsincontactwithpatientbreathinggasesarefreeofnaturalrubberlatex.

Environmental specifications

System operation

Temperature: 10°to40°C/50°to104°F

Humidity: 15to95%relativehumidity (non-condensing)perIEC68-2-3

Altitude: –440to3565m/500to800mmHg

System storage

Temperature: –15°to50°C/–5°to122°F

Humidity: 10to95%relativehumidity (includingcondensing)perIEC68-2-3

Altitude: –440to5860m/375to800mmHg

Oxygencellstorage: –15°to50°C/5°to122°F 1 0to95%relativehumidity 500to800mmHg

Electromagnetic compatibility

Immunity: Complieswithallrequirements ofEN60601-1-2

Emissions: CISPR11group1classB

Approvals: UL2601-1, CSAC22.2#601.1 EN/IEC60601-1 CE0197

Breathing circuit specifications

Operational modes

Breathingcircuitiscirclemodeonly

Carbon dioxide absorbent canister

Absorbentcapacity: 800g

Integratedexpiratorylimbwaterreservoír

Ports and connectors

Exhalation: 22mmODISO15mm IDtaper

Inhalation: 22mmODISO15mm IDtaper

Bagport: 22mmOD

Pressure gauge

Scalerange: 0to10kPa/ –20to100cmH20

Bag-to-Ventilator switch

Type: Bi-stable

Control: Controlsventilatorand directionofbreathinggas withinthecircuit

Integrated Adjustable Pressure Limiting (APL) valve

Range: 0.8to70cmH2O

Tactileknob indicationat: 30cmH2Oandabove

Adjustmentrangeofrotation: 0.8to30cmH2O(0to230°) 30to70cmH2O(230to330°)

Materials

Allmaterialsincontactwithexhaledpatientgasesareautoclavable,exceptdisposableflowsensorsandO2cell.(Autoclavableflowsensorsoptional).

Allmaterialsincontactwithpatientgasarefreeofnaturalrubberlatex.

Breathing circuit parameters

Compliance: Bagmode: 1.82mL/cmH2O

Mechanical mode: Automaticallycompensatesforcompression losseswithintheabsorberandbellows assembly

Circuitvolume: 2.7LVentMode 1.2LBagMode

Expiratoryresistance: PexpBagMode PexpVentMode Flowrate Pressuredrop Pressuredrop

10L/min 0.78cmH2O 0.77cmH2O

30L/min 1.59cmH2O 1.71cmH2O

60L/min 3.48cmH2O 3.88cmH2O

Note:Withpatientcircuitandwyepieceadd+0.89cmH2O

Anesthetic gas scavenging

Type Hospitalsystemrequired Machineconnection

Activelowflow: Highvacuum36L/min DISSevac (300mmHg)@12inHg

Passive: Passiveorexternally 30mm/0.5in attachedactivesystem MISOtaper

Activeadjustableflow: >30L/mi

©2011GeneralElectricCompany–AllRightsReserved.

GeneralElectricCompanyreservestherighttomakechangesinspecificationandfeaturesshownherein,ordiscontinuetheproductdescribedatanytimewithoutnoticeorobligation.ContactyourGErepresentativeforthemostcurrentinformation.

GEandGEMonogramaretrademarksofGeneralElectricCompany.

*Aespire,Cardiocap,Dash,SelectatecandTecaretrademarksofGeneralElectricCompany.

Datex-Ohmeda,Inc.,aGeneralElectriccompany,doingbusinessasGEHealthcare.

AN-0237-08.11-EN-US DOC0000000

About GE Healthcare GEHealthcareprovidestransformationalmedicaltechnologiesandservicesthatareshapinganewageofpatientcare.Ourbroadexpertiseinmedicalimagingandinformationtechnologies,medicaldiagnostics,patientmonitoringsystems,drugdiscovery,biopharmaceuticalmanufacturingtechnologies,performanceimprovementandperformancesolutionsserviceshelpourcustomerstodeliverbettercaretomorepeoplearoundtheworldatalowercost.Inaddition,wepartnerwithhealthcareleaders,strivingtoleveragetheglobalpolicychangenecessarytoimplementasuccessfulshifttosustainablehealthcaresystems.

Our“healthymagination”visionforthefutureinvitestheworldtojoinusonourjourneyaswecontinuouslydevelopinnovationsfocusedonreducingcosts,increasingaccessandimprovingqualityaroundtheworld.HeadquarteredintheUnitedKingdom,GEHealthcareisaunitofGeneralElectricCompany(NYSE:GE).Worldwide,GEHealthcareemployeesarecommittedtoservinghealthcareprofessionalsandtheirpatientsinmorethan100countries.FormoreinformationaboutGEHealthcare,visitourwebsiteatwww.gehealthcare.com.

GEHealthcareP.O.Box7550Madison,WI53707-7550U.S.A.

www.gehealthcare.com

top related