Artifacts in Cardiac MRIArtifacts in Cardiac MRI

Andreas Sigfridsson, IMV/IMT/CMIV Andreas Sigfridsson, IMV/IMT/CMIV <sigge@imv.liu.se><sigge@imv.liu.se>

2007-04-272007-04-27

sigge@imv.liu.se

Dynamic imagingDynamic imaging

Images are acquired line-by-line in Images are acquired line-by-line in k-k-spacespace Motion during acquisition causes Motion during acquisition causes artifactsartifacts, not , not

blurringblurring Special care must be takenSpecial care must be taken

TriggeringTriggering

Acquisition orderingAcquisition ordering

sigge@imv.liu.se

ExampleExample

Simulation exampleSimulation example 256x256, TR=20 ms, total scan time ~5 s256x256, TR=20 ms, total scan time ~5 s

Image domain k-space

sigge@imv.liu.se

Signal saturationSignal saturation

Signal recovery - relaxationSignal recovery - relaxation

11 T

t

z eM

0t

0 5 10 15-0.5

0

0.5

1

0 5 10 15-0.5

0

0.5

1

Mz

Mx

1TTR

0 0.2 0.4 0.6 0.8 1 1.2-0.5

0

0.5

1

0 0.2 0.4 0.6 0.8 1 1.2-0.5

0

0.5

1

Mz

Mx

1TTR

sigge@imv.liu.se

Signal saturationSignal saturation

Progress to steady-stateProgress to steady-state

901 TTR

0 0.2 0.4 0.6 0.8 1 1.2-0.5

0

0.5

1

0 0.2 0.4 0.6 0.8 1 1.2-0.5

0

0.5

1

Mz

Mx

sigge@imv.liu.se

Varying signal during inflowVarying signal during inflow

The MR signal is increased during inflowThe MR signal is increased during inflow

Signal increase related to flow velocitySignal increase related to flow velocity More magnetization because of T1 More magnetization because of T1

decay/recoverydecay/recovery

Static spins, saturated signal

Flowing spins, fresh signal

sigge@imv.liu.se

Varying signal intensityVarying signal intensity

Simulated six sinusoidal heart beatsSimulated six sinusoidal heart beats

0 50 100 150 200 250 3000.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

Intensity Image (animation)

sigge@imv.liu.se

Varying signal intensity, resultsVarying signal intensity, results

Reconstructed imageReconstructed image

High contrastNormal contrast

sigge@imv.liu.se

Varying signal intensity, k-spaceVarying signal intensity, k-space

Image artifacts because of k-space Image artifacts because of k-space samplingsampling Different lines are acquired with different Different lines are acquired with different

signal intensitysignal intensity

sigge@imv.liu.se

Varying signal intensity, analysisVarying signal intensity, analysis

Analyze signalAnalyze signal

dynamicstatic ImagetIntensityImagetImage *)()(

)(*)()()( ydynamicyystaticy kKkIntensitykKkK

dynamicstatictedreconstruc ImagetIntensityImageImage ))(FT(

)*FT()FT()FT( dynamicstatic KIntensityKK

)FT()FT()FT()FT( dynamicstatic KIntensityKK

sigge@imv.liu.se

Slow sine variationSlow sine variation

0 50 100 150 200 250 3000

20

40

60

80

100

120

140

0 50 100 150 200 250 3000.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

Intensity FT(Intensity)

sigge@imv.liu.se

Varying signal intensity, resultsVarying signal intensity, results

Reconstructed imageReconstructed image

High contrastNormal contrast

sigge@imv.liu.se

Fast sine variationFast sine variation

Intensity FT(Intensity)

0 50 100 150 200 250 3000.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

0 50 100 150 200 250 3000

20

40

60

80

100

120

140

sigge@imv.liu.se

Fast sine variationFast sine variation

0 50 100 150 200 250 3000.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

0 50 100 150 200 250 3000

20

40

60

80

100

120

140

0 50 100 150 200 250 3000

20

40

60

80

100

120

140

0 50 100 150 200 250 3000.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

Intensity FT(Intensity)

sigge@imv.liu.se

““Physiological” variationPhysiological” variation

Aortic flow (handcrafted, not actual data)Aortic flow (handcrafted, not actual data)

0 50 100 150 200 250 3000

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

0 5 10 15 20 250

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

Intensity (one beat)

Intensity (whole scan)

sigge@imv.liu.se

““Physiological” variationPhysiological” variation

Intensity variation (animation)

sigge@imv.liu.se

““Physiological” variationPhysiological” variation

0 50 100 150 200 250 3000

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

0 50 100 150 200 250 3000

10

20

30

40

50

60

70

80

90

Intensity

FT(Intensity)

sigge@imv.liu.se

““Physiological” variationPhysiological” variation

Reconstructed imagesReconstructed images

High contrastNormal contrast

sigge@imv.liu.se

RemediesRemedies

TriggeringTriggering Wait for ECG R-tagWait for ECG R-tag Wait additional trigger delay timeWait additional trigger delay time Acquire one (or several) k-space linesAcquire one (or several) k-space lines

Acquisition reorderingAcquisition reordering Measure current cardiac (or respiratory phase)Measure current cardiac (or respiratory phase) Choose k-space line to acquireChoose k-space line to acquire Creates a slow varying intensity curve along k-spaceCreates a slow varying intensity curve along k-space

sigge@imv.liu.se

Triggering - random phaseTriggering - random phase

If triggering is not accurate, random If triggering is not accurate, random variations will still be leftvariations will still be left

0 50 100 150 200 250 3000

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

0 50 100 150 200 250 3000

10

20

30

40

50

60

70

80

90

Intensity FT(Intensity)

sigge@imv.liu.se

Random phase, reconstructedRandom phase, reconstructed

Normal contrast High contrast

sigge@imv.liu.se

Acquisition reorderingAcquisition reordering

0 50 100 150 200 250 3000

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

Intensity

0 50 100 150 200 250 3000

10

20

30

40

50

60

70

80

90

FT(Intensity)

sigge@imv.liu.se

Acquisition reorderingAcquisition reordering

0 50 100 150 200 250 3000

10

20

30

40

50

60

70

80

90

OrderedRandom

0 50 100 150 200 250 3000

10

20

30

40

50

60

70

80

90

sigge@imv.liu.se

Acquisition reorderingAcquisition reordering

Reconstructed imagesReconstructed images

High contrastNormal contrast

sigge@imv.liu.se

Actual examplesActual examples

ECG triggering failedECG triggering failed

sigge@imv.liu.se

Dynamic imagingDynamic imaging

Images are acquired line-by-line in Images are acquired line-by-line in k-k-spacespace Motion during acquisition causes Motion during acquisition causes artifactsartifacts, not , not

blurringblurring Special care must be takenSpecial care must be taken

TriggeringTriggering

Acquisition orderingAcquisition ordering Artifacts might be predictable through use of Artifacts might be predictable through use of

Fourier transformFourier transform