Blood Collection System
Contents
Page
Operation Sequence of d-vacì
Vacuum Blood Collection Device 4
Non-anticoagulation Tube 5
Pro-coagulation Tube 5
Oxalate Tube 6
Gel and Clot Activator Tube 7
EDTA Tube 8
Heparin Tube 9
ESR Tube 10
PT Tube 11
d-vacì Multi-Sample Needleand Operation Sequence 12-13
Product Specifications 14-16
Accessories 16
3
Colour Code:
Tube Size:
Volume:
Anticoagulant:
Anticoagulant Status:
Vertical Static Pressure:
Mixing Mode:
Allowed Operation Temperature:
Ultimate Centrifugal Speed:
Blood Storage Time:
Deviation for Collect Volume:
Corresponding Altitude Error:
Applicable Area and Altitude:
Note:
Grey
º13x100mm, º13x75mm
2ml, 3ml, 4ml, 5ml, 6ml, 7ml
Potassium Oxalate, Ammonium Oxalate,Potassium Oxalate/Sodium Fluoride,Bioxalate Sodium Fluoride
Biphase Agents or Powder
More than 16kg/cmÇ
After collecting, immediately gently invert the tube to 180Æ 3 to 5 times.
0ÆC - 3ÆC
5000rpm
72hrs in normal temperature (Blood-sugar value)
±10% (at the allowed environmenttemperature scale)
± 200m
See the package
The blood specimen is not suitable for biochemicaland cell morphologic tests.
6
d-vacì Oxalate Tubes contain a series of oxalate anticoagulant agents.By using an inhibitor and special inside treatment, the blood samplecan be preserved for a longer period without its original status beingchanged and the metabolism of blood cells almost cease.
Oxalate Tubes are widely used to collect blood for glucosedeterminations. Sodium fluoride is used in Glucose tubes as glycolysis
inhibitor to preserve glucose when combined with anticoagulant potassiumoxalate.
Oxalate
Colour Code:
Tube Size:
Volume:
Gel Weight:
Vertical Static Pressure:
Recommended Centrifugation Speed:
Ultimate Centrifugal Speed:
Mixing Mode:
Complete Clotting Time:
Blood Natural Contraction Initial Time:
Inside Natural Negative Electric Charge:
Allowed Operation Temperature:
Maximum Error Rate of Vacuum Grade:
Corresponding Altitude Error:
Applicable Area and Altitude:
Note:
Jacinth
º16x125mm, º16x100mm, º13x100mm, º13x75mm
2ml, 3ml, 4ml, 5ml, 6ml, 7.5ml, 8.5ml, 9.5ml, 11ml, 13ml
0.7g-0.8g
8kg/cmÇ
3500rpm
5000rpm
After collecting, immediately gently invert the tube to 180Æ 3 to 5 times.
≤30mins
Approx 30mins
≥30%
0ÆC -37ÆC
±10% (at the allowed environment temperature scale)
±200m
See the package
These tubes are not fit for PCR test.
7
d-vacì Gel and Clot Activator Tubes are used to collect bloodfor clinical biochemistry and immunology. They can improve serumsurfaceand prevent substance exchange between blood cell and serum.
Guaranteeing biochemical character and chemical componentsof serum unchanged evidently for a long time, which do not needto use special tool to transport serum.
Gel and Clot Activator
Colour Code:
Tube Size:
Volume:
Inner Wall Electric Charge:
Anticoagulant Status:
Vertical Static Pressure:
Mixing Mode:
Stopper Hardness:
EDTA. K2, EDTA. K3 Concentration ISO6710:
Allowed Operation Temperature:
Maximum Error Rate of Vacuum Grade:
Corresponding Altitude Error:
Applicable Area and Altitude:
Note:
8
EDTA
Æ
Lavender
º16x100mm, º13x100mm, º13x75mm
1ml, 2ml, 3ml, 4ml, 5ml, 6ml, 7ml, 10ml
0
Liquid, Powder
More than 16kg/cmÇ
After collecting, immediately gently invert the tube to 180Æ5 to 8 times.
38A
1995
0ÆC - 37ÆC
±10% (at the allowed environment temperature scale)
± 200m
See the package
In special cases, if a syringe is required to collect blood,please remove the needle header and then inject for thequantitative blood, preventing the leakage of anticoagulant.
d-vacì EDTA Tubes use EDTA. K2, EDTA. K3 as additives.
They are widely used to collect blood for clinical hematology
examinations and are suitable for blood cell analyser.
Colour Code:
Tube Size:
Volume:
Anticoagulant:
Heparin Concentration:
Vertical Static Pressure:
Mixing Mode:
Allowed Operation Temperature:
Anticoagulant Status:
Ultimate Centrifugal Speed:
Blood Storage Time:
Maximum Error Rate of Vacuum Grade:
Corresponding Altitude Error:
Applicable Area and Altitude:
Note:
Caution:
Green
º16x100mm, º13x100mm, º13x75mm
2ml, 3ml, 4ml, 5ml, 6ml, 7ml, 10ml
Heparin Lithium, Heparin Sodium
ISO6710:1995
More than 16kg/cmÇ
After collecting, immediately gently invert the tube to 180Æ3 to 5 times.
0ÆC - 37ÆC
Hyaline Velum Additive (powder), Flocculent Powder
5000rpm
Less than 6hrs in room temperature
±10% (at the allowed environment temperature scale)
± 200m
See the package
1. Most of the indexes are not effected after routine biochemical serum is changed into plasma.
2. The collection amount should be as sufficient as required to avoid the increase of relative density of heparin, which will effect the measure of calcium and ferment.
3. The test result is unlikely to be repeated when sample is kept 6 hours after collection, especially for the index of AST, ALT, TBIL, DBIL, GGT.
Different test method will bring different influence.
9
d-vacì Heparin Tubes are widely used in blood collection andanticoagulation for clinical biochemical examinations, emergencybiochemical tests and also some items of hemorrheology.
They have the advantages of fast clotting, being fit for a wide rangeof temperatures and are highly compatible with serum samplequato.
Heparin
d-vacì ESR Tubes are used for blood collection andanticoagulation in sedimentation rate testing.The measurement refers to the Westergrenmethod.
ESR
Black
º13x100mm, º13x75mm
1.6 ml, 2.4ml
0.129mol/L Sodium Citrate
Liquid
1:4
More than 16kg/cmÇ
After collecting, immediately gently invert the tube to180Æ 3 to 5 times.
Colour Code:
Tube Size:
Volume:
Anticoagulant:
Anticoagulant Status:
Volume Ratio Between Anticoagulant and Blood:
Vertical Static Pressure:
Mixing Mode:
10
Colour code:
Tube Size:
Volume:
Anticoagulant:
Anticoagulant Status:
Mixing Mode:
Vertical Static Pressure:
Volume Ratio Between Anticoagulant and Blood:
Allowed Operation Temperature:
11
d-vacì PT Tubes are blood collection tubes, using 0.129mol/LSodium Citrate as anticoagulant, whose Volume Ratio is 1 to 9(1:9) compared with blood sample volume.
PT Tubes are mainly used to collect blood for coagulation testing,with the advantage of high accuracy in matching between theproportion of blood and anticoagulant.
PT
Blue
º13x100mm, º13x75mm
1.8ml, 2.7ml, 3.6ml, 4.5ml
Sodium Citrate 0.129mol/L
Liquid
After collecting, immediately gently invert the tubeto 180Æ 3 to 5 times.
More than 16kg/cmÇ
1:9
0ÆC - 37ÆC
d-vacì Multi-Sample Needlefor More Precision and Reliability
d-vacTM Collection Needles
Kink Resistant TubingExcellent blood flow, through kink resistant tubing.
Soft wings allow for easier veinpuncture
Safe and User Friendlyd-vacì Multi-Sample Needles have beendesigned to greatly reduce risk ofcontamination and eliminate directuser-contact with blood sample.
Smooth Sharp NeedleMaximum patient awareness minimises patientdiscomfort, d-vacì coated needles are sharpand smooth minimising pain during penetration.
Luer adaptor
12
Leak Proof NeedleSimultaneously on removing the needlefrom tube, the needle is covered andclosed off, resulting in a guaranteed leakproof needle.
13
with multi sample needle and holder
1. Select an appropriate tube for intended test
and remove cover from the Multi-Sample Needle.
Attach the needle to the holder and screw until tight.
2. Carry out routine disinfection.
Select puncture site and perform veinpuncture.
3. Push tube into holder, puncturing stopper.
Center tube in holder to prevent sidewall penetration
and premature vacuum loss.
Remove tube from holder as soon as it is filled to stated volume.
Do not allow contents of tube to contact the stopper
or the end of the needle during these procedures.
4. If multiple blood sample needle is necessary, remove tube and place
new tube into the holder. Repeat procedure from step 3.
5. As soon as blood stops flowing into the last tube, remove
needle from vein, applying pressure to puncture site with
dry sterile swab until bleeding stops. Invert the tube 5-8
times to mix the blood.
6. Dispose of the needle with an appropriate device.
Operation Sequence of d-vacì
Vacuum Blood Collection Device