diagnostics for kodak directview cr 950 system 26aug04
Post on 01-Jun-2018
319 Views
Preview:
TRANSCRIPT
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
1/206
EASTMAN KODAK COMPANY, 2004 HEALTH IMAGING
Confidential
RestrictedInformation
{Diagnostics}{Production}{HealthImaging}{ExternalAndInternal} Publication No. DG4826-126AUG04
Supersedes
28JUL03
DIAGNOSTICS
for the
KodakDirectViewCR 950 SYSTEM
Service Code: 4826
Important
Qualified service personnel must repair this equipment.
H196_0500GC
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
2/206
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
3/206
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
4/206
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
5/206
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
6/206
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
7/206
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
8/206
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
9/206
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
10/206
DIAGNOSTICS Using the Diagnostics
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
11/206
G OS CS Us g t e ag ost cs
26AUG04
DG4826-1
Page11 of 206
Section 1: Using the Diagnostics
System Service Diagnostics Screen
Important
For systems with Software Version 4.1, you must have a Session ID for access to theService Functions and Diagnostics. See SERVICE BULLETIN 843.
1 At the main menu, touch:
[Service Functions]
[ENTER]
[Diagnostics]
2 Does the System Service Diagnostics screen display?
3 Check that the TX and RX LEDs are blinking on the MASTER CENTRAL PROCESSINGUNIT BOARD A2 and on the NETWORK CARD in the computer.
4 At the main menu, touch:
[Service Functions]
[ENTER]
[EXIT TO DESKTOP]
5 Select My Computer>Control Panel>Network>Protocols.
6 Double-click [TCP/IP Protocol].
7 Select [ADAPTER 1].
Yes No
Advance to the test or procedure. Continue with Step 3.
DIAGNOSTICS Using the Diagnostics
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
12/206
g g
26AUG04
DG4826-1
Page12 of 206
8 Check the values for the NETWORK CARD that is connected to the MASTER CENTRALPROCESSING UNIT BOARD A2:
IP Address 192.168.0.1
Subnet Mask 255.255.255.0
Default Gateway 192.68.0.99
DIAGNOSTICS Using the Diagnostics
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
13/206
26AUG04
DG4826-1
Page13 of 206
Functions of the System Service Diagnostics Screen
Table 1 Component Tests
BUTTON Description
[Individual ComponentControl]
Allows you to actuate and check the status of: STEPPER, ALIGNMENT CAMS
MAIN BELT STEPPER
LOAD/UNLOAD BAR
ALIGNMENT CAMS
ELEVATOR, SIZE CHANGER
MAIN CAM, SCANNER
MAIN CAM MOTOR
SLOW SCAN CONTROLLER
LASER DIODE
ERASE LAMPS
[Component Control: Step] Allows you to check the status of the components byexecuting the steps:1. Advancing the CASSETTE to SLOT 1
2. Loading the CASSETTE
3. Loading the PLATE4. Scanning the PLATE
5. Erasing the PLATE
6. Ejecting the PLATE
7. Ejecting the CASSETTE
DIAGNOSTICS Using the Diagnostics
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
14/206
26AUG04
DG4826-1
Page14 of 206
Table 2 Calibration
BUTTON Description
[View Collector Profiles]
Important
For the procedure, see Calibration for the CR 950SYSTEM.
Displays each COLLECTOR PROFILE for each size ofCASSETTE, including:
number of steps from HOME to START OF SCAN
OFFSET - number of pixels advanced from START tothe diagnostic image
Amplitude - number of pixels from OFFSET to lastpixel of the diagnostic image
PMT 1 - 2: correction value of each PMT for gainadjustment
DIAGNOSTICS Using the Diagnostics
http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf -
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
15/206
26AUG04
DG4826-1
Page15 of 206
BUTTON Description
[Scan Calibration]
Caution
All COVERS must be installed when using this option. For the procedure, see Calibration for the CR 950
SYSTEM.
Completes the calibration of:
GALVO
PMT 1
PMT 2
Starts a COLLECTOR PROFILE for all sizes ofSTORAGE PHOSPHOR PLATES.
Allows calibration of the PLATES exposed at 15 - 25 mR.
Displays the date and time of the last successfulcalibration.
Provides the ability to:
make a backup of the current calibration files from theSCANNER to the computer
restore from the computer backup
remove the existing calibration
DIAGNOSTICS Using the Diagnostics
http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf -
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
16/206
26AUG04
DG4826-1
Page16 of 206
Table 3 Component Adjustments
BUTTON Description
[Set Extraction Bar Home]
Important
For the procedure, see EXTRACTION BAR HOMEPOSITION.
Allows you to set and check the home position of theEXTRACTION BAR:
[FINE NUDGE UP]
[FINE NUDGE DOWN] [HOOKS UP]
[HOOKS DOWN]
[COARSE NUDGE UP]
[COARSE NUDGE DOWN]
[SAVE] - sets the selected home position
[CANCEL] - resets the home position to the last recordedvalue
[Extraction Bar Positions] Allows you to move the EXTRACTION BAR to: [HOME]
[UPPER PLATE GUIDE ROLLER]
[LOWER PLATE GUIDE ROLLER]
[ERASE]
[RETURN TO HOME]
[NUDGE UP]
[NUDGE DOWN]
DIAGNOSTICS Using the Diagnostics
http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf -
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
17/206
26AUG04
DG4826-1
Page17 of 206 [CAM Position Actuation]
Important
For the test, see DUPLEX CAM.
Allows you to: check the status of the CASSETTE HANDLING AY by
rotating the DUPLEX CAM to:
[UNPINCH POSITION]
[HOOKS UP POSITION]
[PINCH POSITION]
[UNLATCH POSITION]
[LATCH POSITION]
[ADVANCE CASSETTE to SLOT 1]
[LOAD CASSETTE]
[EJECT CASSETTE]
[HOME SLOW SCAN]
[HOME CAM]
reset the DUPLEX CAM after jams:
[CLEAR CASSETTE] [RELEASE CASSETTE]
[Slow Scan CurrentCalibration] Important
For the procedure, see Calibration for the CR 950SYSTEM.
Completes the calibration of the SLOW SCAN MOTOR forthe correct current at all 3 phases of the MOTOR.
BUTTON Description
DIAGNOSTICS Using the Diagnostics
http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf -
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
18/206
26AUG04
DG4826-1
Page18 of 206 [Slow Scan Preload
Adjustment]Moves the SLOW SCAN AY. This allows you to:
remove the EXTRACTION BAR
make an adjustment to the SLOW SCAN AY
[Upper Transport Mechanism] Allows you to: move the MAIN BELT
[MOVE number of times]
[INDEX number of times]
[Full Revolution]
[Jog Fine]
[Jog Coarse]
move the ELEVATOR up or down
engage or disengage the LOAD/UNLOAD BAR
move the SLOW SCAN to the home position
move the MAIN CAM to the home position
BUTTON Description
DIAGNOSTICS Using the Diagnostics
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
19/206
26AUG04
DG4826-1
Page19 of 206
Table 4 Tests
BUTTON Description
[Light Test]
Important
For the test, see Light Test.
Checks for excessive light to the PMTs during the loadingand ejecting of a CASSETTE.
Allows you to:
actuate or deactuate the high voltage of the PMTs
AUTO-ZERO the PMTs
Checks the signal for noise.
[Light Test] [Advanced] Checks for excessive light to the PMTs during the loadingof a PLATE.
Allows you to energize and de-energize:
GALVO
LASER DIODE
ERASE LAMPS
DIAGNOSTICS Using the Diagnostics
26AUG04
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
20/206
26AUG04
DG4826-1
Page20 of 206 [Slow Scan Current and
Velocity Tests] Caution
An increase in current might indicate that a bind is
occurring in the SLOW SCAN AY.
For the test, see Slow Scan Current and VelocityTests.
Drives the SLOW SCAN AY to the limits in bothdirections.
Displays for the SLOW SCAN AY:
speed variation
current
Start of Scan
End of Scan
[Laser Power]Laser Warning
Prevent direct exposure to the laser beam.
For the test, see Laser Power.
Allows you to:
check the power of the LASER DIODE set the power values for a new LASER DIODE
store and restore the laser calibration file
BUTTON Description
DIAGNOSTICS Using the Diagnostics
26AUG04
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
21/206
26AUG04
DG4826-1
Page21 of 206
BUTTON Description
[Galvo Test]
Important
For the test, see Galvo Test.
Allows you to: set the Amplitude of the GALVO
check the response of the LASER DIODE
[PMT Test]
Important
For the test, see PMT Test.
Checks the functions of the 2 PMTs.
[Galvo Plot]
Important
For the test, see GALVO PLOT.
Allows you to: check:
response of the GALVO
automatic gain control - AGC
2.5 V reference set the GALVO without an OSCILLOSCOPE
DIAGNOSTICS Using the Diagnostics
26AUG04
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
22/206
26AUG04
DG4826-1
Page22 of 206
Table 5 Logs
BUTTON Description
[Error Log]
Important
For the procedure, see Checking the Error Logs.
Allows you to check: error code and description
date and time
CASSETTE ID
scan count
[Error Frequency Log]
Important
For the procedure, see Checking the Error Frequency Log.
Allows you to: check:
frequency of each error code
date and time of last error
collate error codes by:
date
error number
frequency
DIAGNOSTICS Using the Diagnostics
26AUG04
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
23/206
26AUG04
DG4826-1
Page23 of 206 [Actuation Log]
Important
For the procedure, see Checking the Actuation Logs.
Allows you to check: SCAN CYCLES - this value cannot be reset
BATCH ERASE CYCLES - the number of times that theBatch Erase function was used
TOTAL POWER-ON HOURS
LASER DIODE SCAN CYCLES
LASER HOURS (hours and minutes)
LASER HOURS LAST CLEARED
[Export Log Files]
Important
For the procedure, see Storing the Error Logs on a Disk.
Records onto a FLOPPY DISKETTE: error logs
Actuation Logs
parameters of the SLOW SCAN
BUTTON Description
DIAGNOSTICS Using the Diagnostics
26AUG04
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
24/206
DG4826-1
Page24 of 206 [Clear Error Log]
Important
For the procedure, see Clearing the Error Logs.
Clears all the errors from the error logs.
Displays the date when the error logs were last cleared.
[Import Files]
Important
If the value is 7000, the home position is not adjusted
automatically. Set the value manually.
For the procedure, see Installing Files from a Disk.
Records files from a FLOPPY DISKETTE onto the CR 950SYSTEM:
Actuation Logs
parameters of the SLOW SCAN
home position of the SLOW SCAN
BUTTON Description
DIAGNOSTICS Using the Diagnostics
26AUG04Table 6 Repetitive Cycles
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
25/206
DG4826-1
Page25 of 206
Table 6 Repetitive Cycles
BUTTON Description
[AutoLooper] Automatically processes the CASSETTE to check if it isreliable.
Completes a test cycle includes scanning and erasing.
Processes all the CASSETTES on the LOAD BELT.
Uses all sizes of the CASSETTES.
Disables the automatic checking Light Leak feature,allowing operation with the PANELS removed.
Displays status screens for the SENSORS.
Select [GIVEN NUMBER OF TIMES] to set the numberof times to scan for each CASSETTE.
[Transport] Cycles the MAIN BELT.
Uses a CASSETTE to test:
MAIN BELT operation of loading
Displays status of the SENSORS at each step.
DIAGNOSTICS Using the Diagnostics
26AUG04Table 7 Miscellaneous
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
26/206
DG4826-1
Page26 of 206
Table 7 Miscellaneous
BUTTON Description
[Display Software Versions]
Important
For the procedure, see Displaying the Software Versions.
Displays the software version number of: CR 950 SYSTEM
MASTER CENTRAL PROCESSING UNIT BOARD
MOTION SYSTEM CONTROL BOARD BOOT
MOTION SYSTEM CONTROL BOARD
SLOW SCAN CONTROLLER BOARD BOOT
SLOW SCAN CONTROLLER BOARD FIXED APP
SLOW SCAN CONTROLLER BOARD PRODUCT APP
DIGITIZER BOARD
MEDICAL IMAGE MANAGER
DIAGNOSTICS
[Download to Circuit Boards]
Important
For the procedure, see Downloading Software to the
BOARDS.
Allows you to download the necessary software to all theBOARDS from the HARD DRIVE.
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826 1 BUTTON D i ti
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
27/206
DG4826-1
Page27 of 206 [System Self Test]
Important
For the test, see System Self Tests
.Checks the operation of the MASTER CENTRALPROCESSING UNIT BOARD A2 by running the followingtests:Completes the following diagnostic tests:
GALVO BOARD COMMUNICATION - checks the
connection of the CABLE between:
GALVO BOARD A4
MASTER CENTRAL PROCESSING UNIT BOARD A2
GALVO REFERENCE - reads the GALVO AGCCHANNEL connected to the reference voltage of theGALVO
GALVO MOTOR - reads the voltage of the GALVO AGCto check for problems with the GALVO
GALVO SWEEP - checks the functions of reading andwriting of the REGISTERS on the GALVO BOARD A4
LASER OFFSET
PMT BOARD TEST - checks the connection of theCABLE between:
MASTER CENTRAL PROCESSING UNIT BOARD A2
PMT/DAS BOARD A5
MASTER CENTRAL PROCESSING UNIT BOARD
MEMORY TEST - checks for errors in the reading andwriting of data to each of the image buffers
BUTTON Description
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826 1 Diagnostic Tests
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
28/206
DG4826-1
Page28 of 206
Diagnostic Tests
Light Test
1 On the System Service Diagnostic screen, touch [Light Test].
2 Touch [TEST].
3 Wait for the test to complete.
4 Check that PASS displays in AUTO-ZERO.
5 Touch [START GRAPH].
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
29/206
DG4826-1
Page29 of 206
6 Check that the value is approximately 4 - 5. Use [ZOOM IN].
7 Touch [HV ON].
8 Check that the value increases to approximately 50 - 51.
9 Select:
[ZOOM OUT]
[CLEAR/STOP]
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1 Slow Scan Current and Velocity Tests
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
30/206
DG4826 1
Page30 of 206
Slow Scan Current and Velocity Tests
1 On the System Service Diagnostic screen, touch [Slow Scan Current and VelocityTests].
2 Insert the 35 x 43 CASSETTE into the CR 950 SYSTEM.
3 Touch:
[LOAD 35 x 43 CASSETTE]
[START CURRENT TEST]
4 Wait for the test to complete.
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
31/206
Page31 of 206
5 Check that the value is:
5 -180 between Start of Scan and End of Scan
< 660 between End of Scan and End of Travel
no spikes > 30
6 If the results are not correct, check the error logs. Use Checking the Error Logs.
7 Touch [START VELOCITY TEST].
8 Wait for the test to complete.
9 Check that the results between Start of Scan and End of Scan are below 90 andstable.
10 If the results are not correct, check the error logs. Use Checking the Error Logs.
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1 11 Touch [EJECT THE CASSETTE].
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
32/206
Page32 of 206 12 Remove the CASSETTE.
Laser Power
Laser Warning
Prevent direct exposure to the laser beam.
1 On the System Service Diagnostic screen, touch [Laser Power].
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1 2 Touch:
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
33/206
Page33 of 206 [LASER ON]
[READ]
3 Check Laser Power is 28 - 32 mW.
4 Is the value correct?
Galvo Test
1 On the System Service Diagnostic screen, touch [Galvo Test].
Yes No
Continue with the next test. 1. Check the connections to theLASER DRIVER BOARD A17.
2. Install a new OPTICAL
MODULE.
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1
P
2 Touch:
http://ar4826_1.pdf/http://ar4826_1.pdf/http://ar4826_1.pdf/http://ar4826_1.pdf/ -
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
34/206
Page34 of 206 [START GALVO]
[LASER ON]
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1
Page
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
35/206
Page35 of 206
Laser Warning
Prevent direct exposure to the laser beam.
3 Open the TOP COVER.
4 Insert one sheet of paper into the CR 950 SYSTEM between:
FOLD MIRROR
COLLECTOR
5 Observe the laser beam on the sheet of paper.
6 Touch [-100] until the Set Amplitude value is 400.
H196_1801HC
TOP COVER
H196_1801HCA
COLLECTOR
paper
laser beam FOLD MIRROR
CR 950 SYSTEM
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1
Page
7 Observe the laser beam on the sheet of paper.
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
36/206
Page36 of 206 8 Touch [+100] until the Set Amplitude value is 3000.
9 Observe the laser beam on the sheet of paper.
10 Does the laser beam change size correctly?
PMT Test
1 On the System Service Diagnostic screen, touch [PMT Test].
2 Touch [START PMT TEST].
3 Check that PASSED displays for both PMT 1 and PMT 2.
Set Amplitude Laser Beam
400 short
3000 long
Yes No
Continue with the next test. Check the Error Codes for the
cause of the malfunction.
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1
Page
4 If one of the PMTs displays FAILED, install a new PMT. See PMTs and COLLECTORAY
http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf -
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
37/206
Page37 of 206
AY.
5 If both PMTs display FAILED, check the connections between:
PMT BOARD
DAS BOARD
GALVO PLOT
Important
The Galvo Plot screen has more Cassette Size options with Software Version 4.1.
1 On the System Service Diagnostic screen, touch [Galvo Plot].
2 Touch [START GALVO PLOT].
http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf -
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
38/206
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1
Page Normal GALVO PLOT
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
39/206
g39 of 206
4 Check the values:
5 Are the values correct?
GALVO PLOT normal
AGC 10.0 V 0.1
2.5 vRef 2.5 0.1
Yes No
Continue with the next test. Adjust the GALVO BOARDuntil theGALVO PLOT is normal.
GALVO PLOT with low HOOK GALVO PLOT with excessive HOOK
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1
Page
System Self Tests
1 On the System Service Diagnostic screen touch [System Self Test]
http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf -
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
40/206
40 of 206 1 On the System Service Diagnostic screen, touch [System Self Test].
2 Touch [START MCPU SELF TEST].
3 Wait approximately 20 seconds for the test to complete.
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1
Page41 f 206
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
41/206
41 of 206
4 Observe the results for each test.
5 If any of the tests display FAILED, check the error logs. Use Checking the Error Logs.
6 See the Error Codesto determine the possible cause and action for the problem.
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1
Page42 of 206
Diagnostic Procedures
Checking the Error Logs
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
42/206
42 of 206 Checking the Error Logs
1 On the System Service Diagnostic screen, touch [Error Log].
2 Touch either:
[Sort By Date],
[Sort By Error Code], or
[Sort By Cassette ID]
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1
Page43 of 206
Checking the Error Frequency Log
1 On the System Service Diagnostic screen, touch [Error Frequency Log].
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
43/206
43 of 206 1 On the System Service Diagnostic screen, touch [Error Frequency Log].
2 Touch either: [Sort By Date],
[Sort By Error Code], or
[Sort Error Frequency]
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1
Page44 of 206
Checking the Actuation Logs
1 On the System Service Diagnostic screen, touch [Actuation Log].
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
44/206
44 of 206 y g , [ g]
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1
Page45 of 206
Storing the Error Logs on a Disk
1 On the System Service Diagnostic screen, touch [Export Log Files].
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
45/206
y g [ p g ]
2 Insert a disk into the DRIVE.
3 Touch either:
[COPY ERROR LOGS TO FLOPPY],
[COPY ACTUATION LOGS TO FLOPPY], or
[COPY SLOWSCAN PARAMETERS FILE TO FLOPPY]
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1
Page46 of 206
Clearing the Error Logs
1 On the System Service Diagnostic screen, touch [Clear Error Log].
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
46/206
2 Touch [CLEAR ERROR LOG].
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1
Page47 of 206
Installing Files from a Disk
1 On the System Service Diagnostic screen, touch [Import Files].
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
47/206
2 Insert the disk into the DRIVE.
3 To install the ACTUATION LOGS to the CR 950 SYSTEM, touch [COPY ACTUATIONLOGS FROM FLOPPY].
Important
If the value is 7000, the home position is not adjusted automatically. Set the value manually.
4 To install the parameters of the SLOW SCAN, touch:
[GET SLOWSCAN HOME POSITION FROM FLOPPY]
[UPDATE SLOWSCAN PARAMETER FILE FROM FLOPPY]
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1
Page48 of 206
Displaying the Software Versions
1 On the System Service Diagnostic screen, touch [Display Software Versions].
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
48/206
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1
Page49 of 206
2 Check that each BOARD has the correct communication:
Message Description Action
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
49/206
The version numberis in the field.
The BOARD isResponding.
Check that the newest and correctversion of the software is installed.
All fields are blank. The diagnostics is notoperating. 1. Do a shutdown of the computer.2. De-energize the CR 950
SYSTEM.
3. Energize the CR 950 SYSTEM.
Not Responding
Important
The message is normalduring initialization.
Failure of communicationsto the BOARD.
1. Check the connections to theBOARD.
2. Download the software to theBOARD.
3. If you cannot successfullydownload the software to theBOARD, install a new BOARD.
Not Reporting The BOARD has a failure of
communication becauseanother BOARD is NotResponding.
1. Check the BOARD that is Not
Responding.
2. Do the test again.
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1
Page50 of 206
Downloading Software to the BOARDS
1 On the System Service Diagnostic screen, touch [Download to Circuit Boards].
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
50/206
Caution
Downloading the SSC BOOT can cause damage to the SLOW SCAN CONTROLLERBOARD A6. Only download the SSC BOOT if necessary.
DIAGNOSTICS Using the Diagnostics
26AUG04
DG4826-1
Page51 of 206
2 Touch either:
[DOWNLOAD MCPU APPLICATION],
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
51/206
[DOWNLOAD MSC APPLICATION],
[DOWNLOAD SSC BOOT and APPLICATIONS],
[DOWNLOAD SSC APPLICATIONS], or
[DOWNLOAD SSC PARAMETERS]
DIAGNOSTICS Error Codes26AUG04
DG4826-1
Page52 of 206
Section 2: Error Codes
Overview
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
52/206
Overview
The first 2 digits of the error code identifies the subsystem or component with the error.Example: 10003
The 1 and the 0 indicate that the error occurred in the subsystem BAR CODE.
1 Use Table 8to help identify the component with the error.
2 Advance to the error code to diagnose the error.
Table 8 Subsystem Identification
ID No. Subsystem
10 BAR CODE
11 LATCH/UNLATCH CAM
12 SLOW SCAN AY
13 ERASE LAMPS
14 POWER SUPPLY15 DATA PATH
16 Calibration
18 UPPER CONTROL
19 Self Test for the MASTER CENTRAL PROCESSING UNIT
22 - 32 MOTION SYSTEM CONTROL BOARD A1
Windows2000 Subsystem
47 Shutdown Management
49 DISK MANAGER
50 DICOM
55 Installation
DIAGNOSTICS Error Codes26AUG04
DG4826-1
Page53 of 206
BAR CODE
10003 - READER did not read the CASSETTE BAR CODE
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
53/206
Cause 1: The CASSETTE has a bar code that you cannot read or it does not have a barcode.
Action: Install the correct bar code.
Cause 2: The internal BAR CODE READER malfunctioned.
Action: 1. Observe that the BAR CODE READER is not blocked.
2. Check the connections to the BAR CODE READER.
3. If the BAR CODE READER emits a beep when you insert a CASSETTE, but
the CASSETTE does not load, check CONNECTOR RJ45 between the BARCODE READER and the MASTER CENTRAL PROCESSING UNIT BOARDA1.
4. If the BAR CODE READER does not emit a beep when you load aCASSETTE, check CONNECTOR RJ45 between the BAR CODE READERand the MOTION SYSTEM CONTROL BOARD A1.
5. Open the FRONT DOOR.6. Close the FRONT DOOR to initialize the MOTION SYSTEM CONTROL
BOARD A1.
7. Touch:
[Diagnostics]
[Component Control: Step] [Read CASSETTE ID]
8. If the CASSETTE ID does not display, do again with another CASSETTE.
9. If the CASSETTE ID does not display, install a new BAR CODE READER.
DIAGNOSTICS Error Codes26AUG04
DG4826-1
Page54 of 206
LATCH/UNLATCH CAM
11001 - DUPLEX CAM did not move to the Home Position
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
54/206
11002 - DUPLEX CAM did not move to PINCH position
11003 - DUPLEX CAM did not move to CLAMP/UNLATCH position
11004 - DUPLEX CAM did not move to UNCLAMP/LATCH position
11005 - DUPLEX CAM did not move to UNPINCH position
11006 - CAM OUTER SENSOR did not detect position of DUPLEX CAM
11007 - CAM INNER SENSOR did not detect Home Position for DUPLEXCAM
Cause: The DUPLEX CAM malfunctioned.
Action: Do the checkout for the CASSETTE HANDLING AY.
Cause: The DUPLEX CAM malfunctioned.
Action: Do the checkout for the CASSETTE HANDLING AY.
Cause: The DUPLEX CAM malfunctioned.
Action: Do the checkout for the CASSETTE HANDLING AY.
Cause: The DUPLEX CAM malfunctioned.
Action: Do the checkout for the CASSETTE HANDLING AY.
Cause: The DUPLEX CAM malfunctioned.
Action: Do the checkout for the CASSETTE HANDLING AY.
Cause: The DUPLEX CAM malfunctioned.
Action: Do the checkout for the CASSETTE HANDLING AY.
Cause: The DUPLEX CAM malfunctioned.
Action: Do the checkout for the CASSETTE HANDLING AY.
DIAGNOSTICS Error Codes26AUG04
DG4826-1
Page55 of 206
11009 - DUPLEX CAM: Movement failure
Cause: The DUPLEX CAM malfunctioned.
Action: Do the checkout for the CASSETTE HANDLING AY
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
55/206
11010 - DUPLEX CAM: Software command failure on the MSC BD
11014 - DUPLEX CAM: Timeout Error
11015 - Move command failure because SLOW SCAN not at home position
11016 - MSC BOARD A1: Software Error for DUPLEX CAM
Action: Do the checkout for the CASSETTE HANDLING AY.
Description: The DUPLEX CAM received a command to move that was not correct.
Cause 1: The DUPLEX CAM malfunctioned.
Action: Do the checkout for the CASSETTE HANDLING AY.
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Description: The DUPLEX CAM did not move in the time allowed.
Cause: The DUPLEX CAM malfunctioned.
Action: Do the checkout for the CASSETTE HANDLING AY.
Cause 1: The DUPLEX CAM malfunctioned.
Action: Do the checkout for the CASSETTE HANDLING AY.
Cause 2: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Description: The MOTION SYSTEM CONTROL BOARD A1 did not provide the RTXCTIMER. The TIMER determines the timeouts for the motion of the
DUPLEX CAM.
DIAGNOSTICS Error Codes26AUG04
DG4826-1
Page56 of 206
11017 MSC BOARD A1 S f E f DUPLEX CAM
Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
56/206
11017 - MSC BOARD A1: Software Error for DUPLEX CAM
11018 - DUPLEX CAM did not move to next position
11019 - DUPLEX CAM did not move to PINCH position
11020 - DUPLEX CAM did not move to UNLATCH position
Description: The MOTION SYSTEM CONTROL BOARD A1 did not set up thecommand to move for the DUPLEX CAM.
Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Cause: The DUPLEX CAM malfunctioned.Action: Do the checkout for the CASSETTE HANDLING AY.
Cause 1: The DUPLEX CAM malfunctioned.
Action: Do the checkout for the CASSETTE HANDLING AY.
Cause 2: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Cause 1: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Cause 2: The DUPLEX CAM malfunctioned.
Action: Do the checkout for the CASSETTE HANDLING AY.
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
57/206
DIAGNOSTICS Error Codes26AUG04
DG4826-1
Page58 of 206
SLOW SCAN AY
12001 - SLOW SCAN EXTRACTION BAR did not move to home position
Cause 1: The EXTRACTION BAR malfunctioned
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
58/206
12002 - EXTRACTION BAR REFERENCE SENSOR S9 did not actuate
Cause 1: The EXTRACTION BAR malfunctioned.
Action: 1. Remove any obstructions in the path of the EXTRACTION BAR.
2. Check the SLOW SCAN MOTOR.
3. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
4. Check the EXTRACTION BAR HOME POSITION.
Cause 2: The EXTRACTION BAR REFERENCE SENSOR S9 malfunctioned.
Action: Do the checkout for the EXTRACTION BAR SENSOR S9.
Cause 1: The EXTRACTION BAR REFERENCE SENSOR S9 operates correctly, butanother malfunction occurred.
Action: 1. Remove any obstructions from the path of the EXTRACTION BAR.
2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Cause 2: The EXTRACTION BAR REFERENCE SENSOR S9 malfunctioned.
Action: Do the checkout for the EXTRACTION BAR SENSOR S9.
DIAGNOSTICS Error Codes26AUG04
DG4826-1
Page59 of 206
12003 - PLATE PRESENT SENSOR S5 did not detect PLATE
Cause 1: The PLATE PRESENT SENSOR S5 operates correctly, but another malfunctionoccurred.
http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf -
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
59/206
Action: 1. Check:
a. PLATE was removed from the CASSETTEb. SLOW SCAN MOTOR operates correctly
2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
3. Remove any obstructions from the path of the EXTRACTION BAR.
Cause 2: The PLATE PRESENT SENSOR S5 malfunctioned.Action: 1. Check the status of the PLATE PRESENT SENSOR S5 in the diagnostics.
2. Check that LED DS29 changes status on the MOTION SYSTEM CONTROLBOARD A1.
3. Check the power to the PLATE PRESENT SENSOR S5.
4. Check the alignment of the EMITTER and DETECTOR on the PLATEPRESENT SENSOR S5.
DIAGNOSTICS Error Codes26AUG04
DG4826-1
Page60 of 206
12004 - PLATE PRESENT SENSOR S5 did not actuate
Cause 1: The PLATE PRESENT SENSOR S5 operates correctly, but another malfunctionhas occurred.
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
60/206
Action: 1. Check:
a. PLATE was removed from the CASSETTEb. SLOW SCAN MOTOR operates correctly
2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
3. Remove any obstructions from the path of the EXTRACTION BAR.
Cause 2: The PLATE PRESENT SENSOR S5 malfunctioned.Action: 1. Check the status of the PLATE PRESENT SENSOR S5 in the diagnostics.
2. Check that LED DS29 changes status on the MOTION SYSTEM CONTROLBOARD A1.
3. Check the power to the PLATE PRESENT SENSOR S5.
4. Check the alignment of the EMITTER and DETECTOR of the PLATEPRESENT SENSOR S5.
Cause 3: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page61 of 206
12009 - Caution! SLOW SCAN MOTOR drawing too much power
Cause: A malfunction occurred in one or more of the following components: COIL BOARD
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
61/206
12012 - SLOW SCAN AY: Timeout Error
12013 - Software Error in SLOW SCAN Caused Error in MSC BD
EXTRACTION BAR REFERENCE SENSOR S9
SLOW SCAN AYAction: 1. Check for an obstruction.
2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Description: The PLATE did not move within the correct time.
Cause: A malfunction occurred in one or more of the following components: COIL BOARD A7
SLOW SCAN AY
Action: 1. Check for an obstruction.
2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Description: A command for the PLATE that is not correct was made by the systemsoftware.
Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: 1. Load the software for the MOTION SYSTEM CONTROL BOARD A1.
2. Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page62 of 206
12014 - MSC BOARD A1: Software Error for SLOW SCAN
Description: This error might occur during the diagnostics if the service person startsto move the PLATE before the PLATE stops moving.
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
62/206
12100 - SSC BOARD A6: Software Error
12101 - SLOW SCAN had a velocity malfunction during scan
Cause: The SLOW SCAN AY malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Description: The SLOW SCAN AY will not move.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: 1. De-energize the CR 950 SYSTEM.
2. Energize the CR 950 SYSTEM.
3. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Description: The SLOW SCAN AY exceeded the speed limit when moving the SCANPLATE.
Cause 1: The speed of the SLOW SCAN exceeded the limit. A band might appear on theimage.
Action: Do the Slow Scan Current and Velocity Tests.
Cause 2: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
63/206
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page64 of 206
12108 - SLOW SCAN stopped during scan
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
64/206
12109 - SLOW SCAN stopped during scan
12110 - SLOW SCAN did not communicate with the MSC
12111 - SLOW SCAN did not move
12112 - SLOW SCAN did not move
BOARD A7.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Description: The SLOW SCAN AY did not move because of a communication errorwith the SLOW SCAN CONTROLLER BOARD A1.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page65 of 206
12113 - SSC BOARD A6: NVRAM malfunctioned
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
65/206
12115 - SSC BOARD A6: Software Error
12116 - SSC BOARD A6 did not communicate with MSC
12120 - SLOW SCAN did not move
BOARD A7.
Cause 1: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Description: The SLOW SCAN AY did not stop moving before the next command tomove. This error indicates a communication error with the SLOW SCANCONTROLLER BOARD A6.
Cause 1: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page66 of 206
12121 - SLOW SCAN did not move
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
66/206
12122 - SLOW SCAN COIL BOARD A7 malfunctioned
12123 - SSC BOARD A6: Software Error
12124 - SSC BOARD A6: Software Error
12125 - SLOW SCAN COIL BOARD A7 malfunctioned
BOARD A7.
Description: The SLOW SCAN AY stopped or did not move at all.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Description: The MOTOR CONTROLLER for the SLOW SCAN AY malfunctioned.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Description: An error occurred during the initialization the SLOW SCAN AY.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Description: The SLOW SCAN AY stopped or did not move at all.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page67 of 206
12126 - Caution! SLOW SCAN MOTOR is overheated
Description: The SLOW SCAN MOTOR was using excessive power and is too hot.
Cause: Malfunction in one or more of:COIL BOARD A7
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
67/206
12130 - SSC BOARD A6: Software Error
12131 - SSC BOARD A6 did not communicate with MSC
12132 - SSC BOARD A6 did not communicate with MSC
COIL BOARD A7
EXTRACTION BAR REFERENCE SENSOR S9
SLOW SCAN AY
Action: 1. Check for an obstruction.
2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Cause 1: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Cause 2: The MOTION SYSTEM CONTROLLER BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page68 of 206
12201 - SSC BOARD A6: Programming Error
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
68/206
12202 - 12260 - SSC BOARD A6: Memory Error
BOARD A7.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Error
Code Description
12202 Error occurred during downloading to the SLOW SCAN CONTROLLER BOARD A6.
12203 Memory error on the SLOW SCAN CONTROLLER BOARD A6.
12204 Memory error on the SLOW SCAN CONTROLLER BOARD A6.
12216 Communication error between the SLOW SCAN CONTROLLER BOARD A6 and theMOTION SYSTEM CONTROL BOARD A1.
12217 Communication error between the SLOW SCAN CONTROLLER BOARD A6 and theMOTION SYSTEM CONTROL BOARD A1. Error EM17 - Receive Buffer Overflow.
12218 Error during downloading of code or parameters for the SLOW SCANCONTROLLER BOARD A6.
12219 Error during downloading of code or parameters for the SLOW SCANCONTROLLER BOARD A6.
12220 Error EM20 - New Bootstrap Program, Download all Boot and Program Modules.12221 Malfunction of the SLOW SCAN CONTROLLER BOARD A6.
12222 Checksum error occurred during downloading of code to the SLOW SCANCONTROLLER BOARD A6.
12223 Checksum error occurred during downloading of code to the SLOW SCANCONTROLLER BOARD A6.
12228 Error occurred during downloading of the code to the SLOW SCAN CONTROLLERBOARD A6.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page69 of 206
12230 Error EM30 - Host Flash Download Copy Boot Memory Error.
12240 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM40 Illegal Record T pe
Error
CodeDescription
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
69/206
Error EM40 - Illegal Record Type.
12241 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM41 - Invalid Extended Address.
12242 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM42 - Invalid Erase Block ID.
12243 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM43 - Receive Record Checksum Error.
12244 Error when initializing. Error EM44 - File Sum of Record Checksum Error, DownloadProgram.
12245 Program error. Error EM45 - Boot Checksum Write Error.
12246 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM46 - FIX_APP Code Block Checksum Write Error.
12247 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM47 - Code Page Checksum Write Error.
12248 Error EM48 - Parameters Save Flash Error.
12250 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM50 - Flash Write Address Out Of Range.
12251 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM51 - Flash Program Error: Erase.
12252 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM52 - Flash Program Error: Low Voltage.
12253 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM53 - Flash Program Error: Program.
12254 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM54 - Flash Program Error: Sequence.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page70 of 206
12255 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM55
12260 Error recording the parameters to the SLOW SCAN CONTROLLER BOARD A6
Error
CodeDescription
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
70/206
12896 - SSC BOARD A6: Unexpected Reset
12901 - SSC BOARD A6: Unexpected Retry
12902 - SSC BOARD A6: Unexpected Retry
12260 Error recording the parameters to the SLOW SCAN CONTROLLER BOARD A6.
Error EM60 - Parameters Save Flash Error.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Description: The PLATE was not scanned. The command is executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Description: The SLOW SCAN AY did not move to the ERASE position. Thecommand is executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page71 of 206
12903 - SSC BOARD A6: Unexpected Retry
Description: The SLOW SCAN AY did not move to the home position. The commandis executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
71/206
12904 - SSC BOARD A6: Unexpected Retry
12905 - SSC BOARD A6: Unexpected Retry
12906 - SSC BOARD A6: Unexpected Retry
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Description: The COARSE NUDGE UP command did not occur. The command isexecuted again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Description: The COARSE NUDGE DOWN command did not occur. The command is
executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Description: The FINE NUDGE UP command did not occur. The command isexecuted again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page72 of 206
12907 - SSC BOARD A6: Unexpected Retry
Description: The COARSE NUDGE DOWN command did not occur. The command isexecuted again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
72/206
12908 - SSC BOARD A6: Unexpected Retry
12911 - SSC BOARD A6: Unexpected Retry
12913 - SSC BOARD A6: Unexpected Retry
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Description: The PLATE did not move to the CASSETTE. The command is executedagain.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Description: The SOFT RESET command did not occur. The command is executed
again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Description: The EXTRACTION BAR did not move to the mechanical adjustmentposition one. The command is executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page73 of 206
12914 - SSC BOARD A6: Unexpected Retry
Description: The command MOVE TO EXTRACTION BAR REFERENCE SENSORdid not occur. The command is executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
73/206
12915 - SSC BOARD A6: Unexpected Retry
12930 - SLOW SCAN: Unexpected Retry during Current Test
12931 - SLOW SCAN: Unexpected Retry during Velocity Test
Cause e S O SC CO O O 6 a u ct o ed
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Description: The command MOVE TO THE PLATE GUIDE POSITION did not occur.The command is executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Description: The system was not able to move to the position selected in the
diagnostics. The command is executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Description: The system was not able to move to the position selected in thediagnostics. The command is executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page74 of 206
12941 - Unexpected Retry during Communication to MSC BD
Description: A message could not be sent to the SLOW SCAN CONTROLLERBOARD A6. The command is executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
74/206
ERASE LAMPS
13001 - 13008 - ERASE LAMP PAIRS
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Code Log Message13001 ERASE LAMP PAIR No. 1 malfunctioned
13002 ERASE LAMP PAIR No. 2 malfunctioned
13003 ERASE LAMP PAIR No. 3 malfunctioned
13004 ERASE LAMP PAIR No. 4 malfunctioned
13005 ERASE LAMP PAIR No. 5 malfunctioned
13006 2 ERASE LAMP PAIRS malfunctioned
13007 3 ERASE LAMP PAIRS malfunctioned
13008 4 ERASE LAMP PAIRS malfunctioned
Cause: One or more ERASE LAMP PAIRS malfunctioned.
Action: Install a new ERASE LAMP PAIR.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page75 of 206
13009 - All ERASE LAMPS malfunctioned
Cause 1: The power to the ERASE LAMPS malfunctioned.
Action: 1. De-energize the ERASE LAMPS.
2. Check for 120 V AC across the AC side of RELAY K1.
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
75/206
3. If 120 V AC is not indicated at RELAY K1, check:
ERASE LAMP AY is fully closed
FUSE F1
TERMINAL BLOCK TB1
condition of the CONNECTOR P1/J1 on the ERASE LAMP AY
condition of the INTERLOCK SWITCH S8
wires to the POWER SUPPLY PS1
4. If 120 V AC is indicated on RELAY K1, connect the MULTIMETER to the ACside of the RELAY.
5. Energize and de-energize the ERASE LAMPS. With the LAMPS energized, you
should measure < 2 V AC.6. If the 120 V AC does not change to a Lo signal, check the DC side of RELAY
K1:
connect the MULTIMETER TL-4114 negative to the violet wire
connect the MULTIMETER TL-4114 positive to the blue wire
check for 4.5 - 5.0 V DC on K1 with the ERASE LAMPS energized7. Check the CONNECTOR J10 on the MOTION SYSTEM CONTROL BOARD
A1.
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page76 of 206
13010 - ERASE LAMP PAIRS: Timeout Error
13011 - Command to Erase did not occur
Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
76/206
13011 - Command to Erase did not occur
13012 - MSC BOARD A1: Invalid Command for ERASE LAMPS
13013 - Failure to allocate RTXC timer during Erase Process
13899 - General Erase Error
POWER SUPPLY
14001 - SLOW SCAN: Malfunction of 24 V Power
Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Description: The software did not provide the RTXC TIMER during the ERASEfunction.
Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Description: The 24 V DC CIRCUIT BREAKER of the SLOW SCAN is open.
Cause: Malfunction of the 24 V DC power distribution.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
77/206
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page78 of 206
14005 - Front DOOR Open
Cause 1: The INTERLOCK SWITCH S8 malfunctioned.
Action: 1. Check the wires to the INTERLOCK SWITCH CONNECTOR P5/J5.
2. Install a new INTERLOCK SWITCH S8.
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
78/206
14006 - DUPLEX CAM: Initialization Error
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Description: The PLATE PRESENT SENSOR S5 is blocked when the SLOW SCANAY is in the home position.
Cause 1: The PLATE PRESENT SENSOR S5 operates correctly, but another malfunctionhas occurred. This error occurs when initializing, either when the CR 950 SYSTEMis energized or when a mechanical subsystem initializes.
Action: Check: PLATE was removed from the CASSETTE
SLOW SCAN MOTOR operates correctly
obstruction in the path of the EXTRACTION BAR
Cause 2: The PLATE PRESENT SENSOR S5 malfunctioned.
Action: Check: status of the PLATE PRESENT SENSOR S5 in the diagnostics
LED DS29 changes status on the MOTION SYSTEM CONTROL BOARD A1
power to the PLATE PRESENT SENSOR S5
alignment of the EMITTER and DETECTOR of the PLATE PRESENT SENSORS5
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
79/206
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page80 of 206
14012 - UPPER CONTROL INTERLOCK was pressed while the CASSETTEwas advancing
Cause: The DEAD STOP INTERLOCK S24 actuated.
Action: Check: operation of the DEAD STOP INTERLOCK S24
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
80/206
14013 - Initialization error at the UPPER CONTROL AY
14019 - Invalid Software Version detected for this Model
DATA PATH
15001 - MCPU BOARD A2: Fatal Error in Software
15002 - MCPU BOARD A2: Fatal Error in Software
operation of the DEAD STOP INTERLOCK S24
SENSOR S22 in SLOT 8 of the UPPER CONTROL AY
Cause: A CASSETTE is in SLOT 0 or SLOT 8 when the CR 950 SYSTEM is energized.
Action: Remove the CASSETTE.
Cause: The MSC BOARD has reported a model number that is not consistent with thecurrent version of MCPU BOARD software.
Action: Download the correct version of MCPU and MSC BOARD software. Do theprocedure Downloading Software to the BOARDS.
Description: A PSOS system error occurred that is not possible to repair. The EventLog includes the PSOS system error number, module, and line number.
Cause: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
Description: A system software error occurred. The Event Log will include the PSOSsystem error number, module, and line number.
Cause: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
81/206
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page82 of 206
15006 - Communications Error between MCPU BD A2 and GALVO BD A4
Description: The process of the DMA for the MASTER CENTRAL PROCESSINGUNIT BOARD A2 did not complete within the expected time.
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
82/206
15007 - Communications Error between MCPU BD A2 and GALVO BD A4
15008 - Start of Scan Timeout
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.Cause 2: The DIGITIZER BOARD A3 malfunctioned.
Action: Do the checkout for the DIGITIZER BOARD A3.
Cause 3: The GALVO BOARD A4 malfunctioned.
Action: Do the checkout for the GALVO BOARD A4.
Description: The process of the DMA for the MASTER CENTRAL PROCESSINGUNIT BOARD A2 did not complete within the expected time.
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
Cause 2: The DIGITIZER BOARD A3 malfunctioned.
Action: Do the checkout for the DIGITIZER BOARD A3.
Cause 3: The GALVO BOARD A4 malfunctioned.
Action: Do the checkout for the GALVO BOARD A4.
Description: A timeout occurred for the START OF SCAN.
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
Cause 2: The DIGITIZER BOARD A3 malfunctioned.
Action: Do the checkout for the DIGITIZER BOARD A3.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page83 of 206
15012 - Communications Error between MCPU BOARD A2 and NT
Cause 3: The GALVO BOARD A4 malfunctioned.
Action: Do the checkout for the GALVO BOARD A4.
Description:Windows2000 did not return a response to a message within a given
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
83/206
time.
Cause 1: The TCP/IP connection malfunctioned.
Action: Check the TCP/IP connection to the computer. See Step 5, Page 11.
Cause 2: Failure of the Windows2000 software.
Action: De-energize and energize the CR 950 SYSTEM.
Cause 3: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page84 of 206
15013 - Image Quality Warning
Description: The image quality might be wrong. The MASTER CENTRAL PROCESSINGUNIT BOARD A2 received a system error from the MOTION SYSTEMCONTROL BOARD A1 and was automatically repaired. If a CASSETTE was inthe CR 950 SYSTEM when the error occurred, the TOUCH SCREEN displays
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
84/206
15014 - LASER: Calibration did not occur
15015 - PLATE: Size calibration did not occur
a message about a problem with the image.Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Cause 3: A light leakage occurred.
Action: Do the Light Test.
Cause 4: The INTERLOCK SWITCH S8 malfunctioned. The DOOR might be open duringscan.
Action: Check the INTERLOCK SWITCH S8.
Cause: The LASER DIODE malfunctioned.
Action: Do the calibration for the LASER DRIVER BOARD A17- use Optics Module.
Cause: The CR 950 SYSTEM does not have the calibration data for all sizes of PLATES.Action: 1. Do the Calibration for the CR 950 SYSTEM.
2. Do the checkout MASTER CENTRAL PROCESSING UNIT BOARD A2.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page85 of 206
15016 - Laser power high
Cause: The LASER DIODE in the DETECTOR is backward.
Action: Check: LASER REFERENCE VOLTAGE in the PMT/DAS BOARD A5
LASER DIODE is installed correctly
http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf -
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
85/206
15017 - PLATE: Calibration did not occur
15018 - There was an error reading or writing to the PCMCIA card
Important
A malfunction of the PCMCIA CARD is the last possible cause. Check all othermalfunctions before installing a new PCMCIA CARD.
15040 - Communications Error Between MCPU and the PC
Cause: The customer scanned a PLATE of a size that was not recognized.
Action: 1. Do the Calibration for the CR 950 SYSTEM.
2. Use the diagnostics to complete the size calibration of the PLATE.
Cause 1: The PCMCIA CARD operates correctly, but another malfunction occurred.
Action: Check the error logs for the malfunction. Use Checking the Error Logs.
Cause 2: The PCMCIA CARD malfunctioned.
Action: Install a new PCMCIA CARD.
Description: The TCP/IP SOCKET COMMUNICATIONS were not completed or didnot occur between the MASTER CENTRAL PROCESSING UNITBOARD A2 and the host computer with Windows2000.
Cause 1: The shutdown of the CR 950 SYSTEM was not completed in the correctsequence.
Action: Use the correct shutdown sequence to de-energize the CR 950 SYSTEM.
http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf -
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
86/206
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page87 of 206
Calibration
16001 - Calibration Error: Did not find START OF SCAN position
Description: During the calibration, the CR 950 SYSTEM did not locate the STARTSCAN position within the given time.
C 1 Th f th PLATE i t t
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
87/206
16002 - Calibration Error: Did not find Offset
Cause 1: The exposure of the PLATE is not correct.Action: Use the DOSIMETER to check exposure of the PLATE.
Cause 2: Calibration of the CR 950 SYSTEM is not correct.
Action: Do the Calibration for the CR 950 SYSTEM.
Cause 3: The alignment of the optics is not correct.
Action: Check the adjustment: FOLD MIRROR
GALVO BOARD
Cause 4: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Cause 5: The PMT/DAS BOARD A5 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.
Description: The GALVO did not detect the Offset.Cause 1: The exposure of the PLATE is not correct.
Action: Use the DOSIMETER to check the exposure of the PLATE.
Cause 2: The GALVO is not aligned correctly.
Action: Do:
adjustment of the GALVO BOARD checkout for the GALVO BOARD A4
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page88 of 206
16003 - Calibration Error: Did not find Amplitude
Cause 3: The PLATE has scratches on the edges.
Action: Check the PLATE for scratches and scan another PLATE.
Check the edges and the degree of reflection.
D i ti Th GALVO did t d t t th A lit d
http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf -
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
88/206
16005 - Calibration Error: Routine did not receive image in the timeallowed
Description: The GALVO did not detect the Amplitude.
Cause 1: The exposure of the PLATE is not correct.
Action: Use the DOSIMETER to check the exposure of the PLATE.
Cause 2: The GALVO is not aligned correctly.
Action: Do:
adjustment of the GALVO BOARD
checkout for the GALVO BOARD A4.
Cause 3: The PLATE has scratches on the edges.
Action: Check: PLATE for scratches and scan another PLATE
edges and the degree of reflection
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
Cause 2: The DIGITIZER BOARD A3 malfunctioned.
Action: Do the checkout for the DIGITIZER BOARD A3.
Cause 3: The GALVO BOARD A4 malfunctioned.
Action: Do the checkout for the GALVO BOARD A4.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page89 of 206
16006 - Calibration Error: Did not obtain image
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
Cause 2: The DIGITIZER BOARD A3 malfunctioned.
Action: Do the checkout for the DIGITIZER BOARD A3
http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf -
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
89/206
16010 - Possible Light Leakage
16011 - Imaging System Error occurred during Initialization
Action: Do the checkout for the DIGITIZER BOARD A3.Cause 3: The GALVO BOARD A4 malfunctioned.
Action: Do the checkout for the GALVO BOARD A4.
Description: Large quantities of light occurred during the AUTO-ZERO.
Cause: The DOOR is not closed, a PANEL is loose, or a light leakage occurred in theCASSETTE HANDLING SYSTEM.
Action: 1. Do the Light Testto locate the source of the light leakage.
2. Repair the light leakage.
Cause: The PMT/DAS BOARD A5 malfunctioned.
Action: 1. Check the CABLE between the PMT BOARD and the DAS BOARD.
2. Do:
checkout for the GALVO BOARD A4
System Self Tests
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page90 of 206
16012 - PMT/DAS BOARD A5: Calculation Error occurred duringInitialization
Description: The value for one or more of the following parameters did not match thespecifications when the CR 950 SYSTEM initialized: Fine or CoarseOffset or Fine or Coarse Ratio.
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
90/206
16013 - Imaging System contains noise beyond the limit
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
Cause 2: The PMT/DAS BOARD A5 malfunctioned.
Action: Do the checkout for the PMT/DAS BOARD A5.
Description: The peak-to-peak signal was high when the CR 950 SYSTEM was setto AUTO-ZERO.
Cause 1: The PMT/DAS BOARD A5 malfunctioned.
Action: Do the checkout for the PMT/DAS BOARD A5.
Cause 2: Noise in the POWER SUPPLY PS1.Action: Do the checkout for the POWER SUPPLY PS1.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page91 of 206
16014 - Imaging System did not zero or calibration of the PMT did notoccur
Cause 1: The DOSIMETER malfunctioned.
Action: Check that the DOSIMETER operates correctly.
Cause 2: The exposure on the PLATE is not correct during calibration
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
91/206
16016 - Malfunctioning PMT
16019 - Calibration could not find offset, rotate GALVO counterclockwise
16020 - Calibration could not find offset, rotate GALVO clockwise
Cause 2: The exposure on the PLATE is not correct during calibration.Action: Do the calibration with another PLATE.
Cause 3: The X-RAY TUBE malfunctioned.
Action: Expose the PLATE using an X-RAY TUBE you did not use before.
Cause 4: The Imaging System did not auto zero.
Action: Do the checkout for: PMT/DAS BOARD A5.
GALVO BOARD A4.
Description: A PMT does not match the requirements of the diagnostic tests.
Cause: The PMT/DAS BOARD A5 malfunctioned.
Action: Do the checkout for the PMT/DAS BOARD A5.
Cause: The calibration could not locate the Offset.
Action: Rotate the GALVO counterclockwise.
Cause: The calibration could not locate the Offset.
Action: Rotate the GALVO clockwise.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page92 of 206
16021 - Malfunctioning PMT1: Gain out of range
Description: For a given high voltage, the low gain of PMT1 might prevent it fromreaching the correct signal level.
Cause 1: The exposure on the PLATE is not correct during calibration.
Action: Do the Calibration for the CR 950 SYSTEM.
http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf -
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
92/206
16022 - Malfunctioning PMT2: Gain out of range
Cause 2: The alignment of the optics is not correct.
Action: Do the adjustment: FOLD MIRROR
GALVO BOARD
Cause 3: The PMT malfunctioned.Action: Install a new PMT.
Cause 4: The PMT/DAS BOARD A5 malfunctioned.
Action: Do the checkout for the PMT/DAS BOARD A5.
Description: For a given high voltage, the low gain of PMT 2 might prevent it fromreaching the correct signal level.
Cause 1: The exposure on the PLATE is not correct during calibration.
Action: Do the Calibration for the CR 950 SYSTEM.
Cause 2: The alignment of the optics is not correct.
Action: Do the adjustment: FOLD MIRROR
GALVO BOARD
Cause 3: The PMT malfunctioned.
Action: Install a new PMT.
Cause 4: The PMT/DAS BOARD A5 malfunctioned.
Action: Do the checkout for the PMT/DAS BOARD A5.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page93 of 206
UPPER CONTROL
18001 - SIZE CHANGER: error
Description: The STEPPER MOTOR exceeded the time allowed to move.
Cause: The SIZE CHANGER HOME SENSOR malfunctioned.
Action: a. Check the CABLES and CONNECTORS to the SIZE CHANGER HOME
http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf -
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
93/206
18002 - UPPER CONTROL AY: timeout error
18007 - UPPER CONTROL AY: Wrong command error
18008 - UPPER CONTROL AY: Move error
Action: a. Check the CABLES and CONNECTORS to the SIZE CHANGER HOMESENSOR.
b. Replace the SIZE CHANGER HOME SENSOR.
Cause: Size .
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Description: The UPPER CONTROL AY exceeded the time allowed to move.
Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Cause: The UPPER CONTROL MOTION MECHANISM malfunctioned.
Action: Check the error logs for the cause of the malfunction. See Checking the ErrorLogs.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page94 of 206
18010 - Invalid STEPPER MOTOR error
Description: The command to move the STEPPER MOTOR is not within the limits ofthe MOTOR in the UPPER CONTROL AY.
Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
94/206
18011 - STEPPER MOTOR did not clear HOME SENSOR
18012 - Failure to do the setup of the SENSOR INTERRUPT
18014 - STEPPER MOTOR: jog timeout error
18013 - Invalid DC MOTOR INTERRUPT
Cause 1: The SLOT 8 SENSOR S22 is actuated when it should be open.
Action: Check the SLOT 8 SENSOR S22.
Cause 2: The SLOT 0 SENSOR S20 is actuated when it should be open.
Action: Check the SLOT 0 SENSOR S20.
Cause 3: The CASSETTE PRESENT SENSOR S19 is actuated when it should be open.
Action: Check the CASSETTE PRESENT SENSOR S19.
Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Description: The command to move the DC MOTOR is not within the limits of theMOTORS in the UPPER CONTROL AY.
Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page95 of 206
18015 - Failure to advance CASSETTE
Cause 1: The CASSETTE was removed.
Action: Insert the CASSETTE again.
Cause 2: The LOAD SENSORS 1 - 8 did not detect the CASSETTE.
Action: 1. Check the LOAD SENSORS 1 - 8.
-
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
95/206
18016 - No CASSETTES to advance
18017 - Cannot advance because the CASSETTE is in SLOT 8
2. Check the test points on the TOP SENSOR BOARD A11.
Description: The UPPER CONTROL AY received a command to advance aCASSETTE but the LOAD STATION SENSORS indicate that no
CASSETTES are available.
Cause: No CASSETTES are detected in SLOTS 1 - 8.
Action: 1. Insert a CASSETTE.
2. Check the SLOT SENSORS 1 - 8.
3. Check the test points for the TOP SENSOR BOARD A11.
Cause: The CASSETTE is in SLOT 8.
Action: 1. Remove the CASSETTE.
2. Check the SENSOR SLOT 8.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page96 of 206
18018 - MAIN BELT: Home position failure
Description: The MAIN BELT STEPPER MOTOR did not move to the home position.
Cause 1: The MAIN BELT SENSOR S21 malfunctioned.
Action: Check the MAIN BELT SENSOR S21.
Cause 2: The MAIN BELT is not correctly aligned.
http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf -
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
96/206
18019 - LOAD BAR: Home position failure
18020 - UNLOAD BAR: Home position failure
y gAction: Check the alignment of the COG BELTS.
Cause 3: The STEPPER MOTOR DRIVER BOARD A12 malfunctioned.
Action: Do the checkout for the STEPPER MOTOR DRIVER BOARD A12.
Cause 4: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Description: The LOAD BAR STEPPER MOTOR did not move to the home position.
Cause 1: Cannot determine the home position of the LOAD BAR.
Action: Check the adjustment of the LOAD BARS.Cause 2: The STEPPER DRIVER BOARD 13 malfunctioned.
Action: Do the checkout for the STEPPER MOTOR DRIVER BOARD A13.
Cause 3: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Description: The UNLOAD BAR did not move to the home position.
Cause 1: Cannot determine the UNLOAD BAR home position.
Action: Check the adjustment of the UNLOAD BARS.
Cause 2: The STEPPER MOTOR DRIVER BOARD A14 malfunctioned.Action: Do the checkout for the STEPPER MOTOR DRIVER BOARD A14.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page97 of 206
18024 - ALIGNMENT CAM: Clear timeout error
Description: The ALIGNMENT CAM did not move to the CLEAR POSITION in thecorrect time.
Cause 1: A mechanical bind occurred.
Action: Check the adjustment of the ALIGNMENT CAMS. See ALIGNMENT CAMS /BELTS.
http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf -
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
97/206
18025 - ALIGNMENT CAM: Engage timeout error
Cause 2: The ALIGNMENT CAM CLEAR SENSOR malfunctioned.
Action: Check the ALIGNMENT CAM CLEAR SENSOR. See ALIGNMENT CAMS /BELTS.
Cause 3: The DC ALIGNMENT CAM MOTOR M5 malfunctioned.
Action: 1. Check the CONNECTORS J1/P1.
2. Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Description: The ALIGNMENT CAM did not move to the ENGAGE POSITION in theavailable time.
Cause 1: A mechanical bind occurred.
Action: Check the adjustment of the ALIGNMENT CAMS. See ALIGNMENT CAMS /BELTS.
Cause 2: The ALIGNMENT CAM ENGAGE SENSOR malfunctioned.
Action: Check the ALIGNMENT CAM ENGAGE SENSOR. See ALIGNMENT CAMS /
BELTS.
Cause 3: The DC ALIGNMENT CAM MOTOR M5 malfunctioned.
Action: 1. Check the CONNECTORS J1/P1.
2. Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
DIAGNOSTICS Error Codes
26AUG04
DG4826-1
Page98 of 206
18026 - ELEVATOR: Up timeout error
Description: The ELEVATOR did not move to the up position in the available time.
Cause 1: The UPPER ELEVATOR SENSOR S12 malfunctioned.
Action: Check the UPPER ELEVATOR SENSOR S12.
Cause 2: The COUPLING from the DC MOTOR to the SHAFT is broken.A ti I t ll COUPLING
http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf -
8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04
98/206
18027 - ELEVATOR down timeout error
Action: Install a new COUPLING.
Cause 3: A mechanical bind occurred.
Action: Check the alignment of the ELEVATOR.
Cause 4: The DC MOTOR malfunctioned.
Action: Check the CABLE to the DC MOTOR.
Cause 5: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Description: The ELEVATOR did not move to the down position in the available time.Cause 1: The LOWER ELEVATOR SENSOR S11 malfunctioned.
Action: Check the LOWER ELEVATOR SENSOR S11.
Cause 2: The COUPLING from the DC MOTOR to the SHAFT is broken.
Action: Install a new COUPLING.
Cause 3: A mechanical bind occurred in the ELEVATOR.Action: Check the alignment of the ELEVATOR.
Cause 4: The DC MOTOR in the ELEVATOR malfunctioned.
Action: Check the connections to the DC MOTOR.
Cause 5: The