article 4 metodo ultrasonido.doc
TRANSCRIPT
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
1/24
yright ASME Internationalided by IHS under license with ASME
eproduction or networking permitted without license rom IHS
!icensee"Savannah #iver Nuclear Solutions$%&''())**'
Not or #esale+ *,$-,$-*'* ''.*,.%- M/0
2010 SECTION V ARTICLE 4
ARTICLE 4
ULTRASONIC EXAMINATION
METHODS FOR WELDS
(10)
T-410 SCOPE
This Article provides or references requirements for
weld examinations, which are to be used in selecting and
developing ultrasonic examination procedures when
exam- ination to any part of this Article is a
requirement of a referencing Code Section. These
procedures are to be used for the ultrasonic examination
of welds and the dimen- sioning of indications for
comparison with acceptance stan- dards when required by
the referencing Code Section the referencing Code
Section shall be consulted for specific requirements for
the following!
(a) personnel qualification " certification
requirements
(b) procedure requirements " demonstration,
qualifica- tion, acceptance
(c) examination system characteristics
(d) retention and control of calibration bloc#s
(e) extent of examination and " or volume to bescanned
(f) acceptance standards
(g) retention of records
(h) report requirements
$efinitions of terms used in this Article are contained
in %andatory Appendix &&& of Article '.
T-420 GENERAL
The requirements of this Article shall be used together
with Article (, )eneral *equirements. *efer to
(a) special provisions for coarse grain materials and
welds in T-+'(
(b) special provisions for computeried imaging tech-
niques in T-+'
(c) %andatory Appendix &&& for Time of light $iffrac-
tion /T0$1 techniques
(d) %andatory Appendix &2 for phased array manual
rastering techniques
T-421 Written Pr!e"#re Re$#ire%ent&
T-421'1 Re$#ire%ent&' 3ltrasonic examination shall
be performed in accordance with a written procedure
which shall, as a minimum, contain the requirements
listed in
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
2/24
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
3/24
yright ASME Internationalided by IHS under license with ASME
eproduction or networking permitted without license rom IHS
!icensee"Savannah #iver Nuclear Solutions$%&''())**'
Not or #esale+ *,$-,$-*'* ''.*,.%- M/0
ARTICLE 4 2010 SECTION V
TABLE T-421REQUIREMENTS OF AN ULTRASONIC EXAMINATION PROCEDURE
Requiree!"
E##e!"i$%
&$ri$'%e
N(!e##e!"i$%
&$ri$'%e
1eld conigurations to be e2amined+ including thicknessdimensions and base material product orm 3pipe+ plate+ etc45 6 4 4 4
0he suraces rom which the e2amination shall be perormed 6 4 4 4
0echni7ue3s5 3straight beam+ angle beam+ contact+ and$or immersion5 6 4 4 4
Angle3s5 and mode3s5 o wave propagation in the material 6 4 4 4
Search unit type3s5+ re7uency3ies5+ and element si8e3s5$shape3s5 6 4 4 4
Special search units+ wedges+ shoes+ or saddles+ when used 6 4 4 4
9ltrasonic instrument3s5 6 4 4 4
Calibration :calibration block3s5 and techni7ue3s5; 6 4 4 4
/irections and e2tent o scanning 6 4 4 4
Scanning 3manual vs4 automatic5 6 4 4 4
Method or discriminating geometric rom law indications 6 4 4 4
Method or si8ing indications 6 4 4 4
Computer enhanced data ac7uisition+ when used 6 4 4 4
Scan overlap 3decrease only5 6 4 4 4
Personnel perormance re7uirements+ when re7uired 6 4 4 4
Personnel 7ualiication re7uirements 4 4 4 6
Surace condition 3e2amination surace+ calibration block5 4 4 4 6Couplant. brand name or type 4 4 4 6
Automatic alarm and$or recording e7uipment+ when applicable 4 4 4 6
#ecords+ including minimum calibration data to be recorded 3e4g4+ instrument
settings5 4 4 4 6
T-4,2'2 C*)""in./Se)r! Unit& +r Te!ni$#e
One'(
$ual element search units using an angled pitch-
catch technique shall be used. The included angle between
the beam paths shall be such that the effective focal spot
of the search unit is centered in the area of interest.
T-4,, C#*)nt
T-4,,'1 Gener)*' The couplant, including additives,
shall not be detrimental to the material being examined.
T-4,,'2 Cntr* + Cnt)%in)nt&
(a) Couplants used on nic#el base alloys shall not con-
tain more than '6 ppm of sulfur.
(b) Couplants used on austenitic stainless steel or tita-
nium shall not contain more than '6 ppm of halides
/chlo- rides plus fluorides1.
T-4,4 C)*ir)tin *!3&
T-4,4'1 Gener)*
T-4,4'1'1 Re+*e!tr&' Specified reflectors /i.e.,
side-drilled holes, flat bottom holes, notches, etc.1 shall
be used to establish primary reference responses of the
equipment. An alternative reflector/s1 may be used pro-
vided that the alternative reflector/s1 produces a sensitivity
equal to or greater than the specified reflector/s1 /e.g.,
(See paragraph T-+umber grouping asone of the materials being examined. or the purposes of
this paragraph, =->os. (, ;, +, 'A through 'C, and ('A
through (' materials are considered equivalent.
(b) Dissimilar Metal Welds. The material selection
shall be based on the material on the side of the weld from
which the examination will be conducted. &f the
examination will be conducted from both sides,
calibration reflectors shall be provided in both materials.
T-4,4'1', (#)*it' =rior to fabrication, the bloc#
material shall be completely examined with a straight
beam search unit. Areas that contain an indication
exceeding the remaining bac#-wall reflection shall be
excluded from the beam paths required to reach thevarious calibra tion reflectors.
T-4,4'1'4 C*)""in.' 4hen the component material
is clad, the bloc# shall be clad by the same welding proce-
dure as the production part. &t is desirable to have compo-
nent materials which have been clad before the drop outs
or prolongations are removed. 4hen the cladding is
deposited using an automatic welding process, and, if due
to bloc# sie, the automatic welding process is
impractical, deposi- tion of clad may be by the manual
method.
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
4/24
yright ASME Internationalided by IHS under license with ASME
eproduction or networking permitted without license rom IHS
!icensee"Savannah #iver Nuclear Solutions$%&''())**'
Not or #esale+ *,$-,$-*'* ''.*,.%- M/0
--``,,,,,,`,``,``,`,`````,,,`,,-`-`,,`,,`,`,,`--- ++
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
5/24
--
,,,,,,
,
,
,
,
,,,
,,
--
,,
,,
,
,,
---
4=- 3'*(5 ,4- 3'(*5 '- 3=**5 -* 3%**5
'=4== 3===5
'* 3-%*5
( 3-**5
% 3'-%5>4( 3'-*5
-4(( 3,-5
'4,= 3>=5
'4*> 3-)5
** % 3'-%5 '* 3-%*5 '% 3=,%5 -* 3%**5
E2amination Surace /iameter+ in4 3mm5
(10)
T-4,4'1'5 He)t Tre)t%ent' The calibration bloc#
shall receive at least the minimum tempering treatment
required by the material specification for the type and
grade. &f the calibration bloc# contains welds other than
cladding, and the component weld at the time of the
exami- nation has been heat treated, the bloc# shall
receive the same heat treatment.T-4,4'1'6 S#r+)!e Fini&' The finish on the
scanning surfaces of the bloc# shall be representative of
the scanning surface finishes on the component to be
examined.
T-4,4'1'7 *!3 C#r8)t#re
T-4,4'1'7'1 M)teri)*& Wit Di)%eter& Gre)ter
T)n 20 in' 9500 %%:' or examinations in materials
where the examination surface diameter is greater than
6 in. /'66 mm1, a bloc# of essentially the same curvature,
or alternatively, a flat basic calibration bloc#, may be
used.
T-4,4'1'7'2 M)teri)*& Wit Di)%eter& 20 in'
9500 %%: )n" Le&&' or examinations in materials where
the examination surface diameter is equal to or less than
6 in. /'66 mm1, a curved bloc# shall be used. ?xcept
where otherwise stated in this Article, a single curved
basic calibration bloc# may be used for examinations in
the range of curvature from 6.@ to (.' times the basiccalibration bloc# diameter. or example, an in /66 mm1
diameter bloc# may be used to calibrate for examinations
on surfaces in the range of curvature from
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
6/24
+'
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
7/24
--
,,,,,,
,
,
,
,
,,,
,,
--
,,
,,
,
,,
---
ARTICLE 4 2010 SECTION V
examining from the convex surface by the straight beam
contact technique, Appendix ) may be used.
T-4,4'2 Nn-Piin. C)*ir)tin *!3&
T-4,4'2'1 )&i! C)*ir)tin *!3' The basic cali-
bration bloc# configuration and reflectors shall be as
shown in ig. T-+;+..(. The bloc# sie and reflectorlocations shall be adequate to perform calibrations for the
beam angles used.
T-4,4'2'2 *!3 Ti!3ne&&' The bloc# thic#ness /T1
shall be per ig. T-+;+..(.
T-4,4'2', *!3 R)n.e + U&e' 4hen the bloc#
thic#ness ( in. /' mm1 spans two weld thic#ness ranges
as shown in ig. T-+;+..(, the bloc#Ds use shall be accept-
able in those portions of each thic#ness range covered by
( in. /' mm1 of the calibration bloc#Ds thic#ness. As an
the base material shall be within ( in. /' mm1 of the
calibration bloc# thic#ness when the examination is per-
formed from the base material surface. The thic#ness of
the base material on the calibration bloc# shall be at least
twice the thic#ness of the cladding when the examination
is performed from the clad surface.
T-440 MISCELLANEOUS RE(UIREMENTS
T-441 I"enti+i!)tin + We*" E;)%in)tin
Are)&
(a) Weld Locations. 4eld locations and their identifica-
tion shall be recorded on a weld map or in an identifica-
tion plan.
(b) Marking. &f welds are to be permanently mar#ed,
low stress stamps and"or vibratooling may be used. %ar#-
example, a calibration bloc# with a thic#ness of ((E in. ings applied after final stress relief of the component shall
/; mm1 could be used for weld thic#nesses of 6.' in. not be any deeper than;EB+ in. /(. mm1.
(10)
/(; mm1 to .' in. /B+ mm1.T-4,4'2'4 A*tern)te *!3' Alternatively, the bloc#
may be constructed as shown >onmandatory Appendix F,
ig. F-+;(.
T-4,4', Piin. C)*ir)tin *!3&' The basic calibra-
tion bloc# configuration and reflectors shall be as shown
in ig. T-+;+.;. The basic calibration bloc# curvature shall
be in accordance with T-+;+.(.onmandatory Appendix A however, a dif-
ferent system may be utilied provided it meets the above
requirements.
T-450 TECHNI(UES
The techniques described in this Article are intended for
applications where either single or dual element search
units are used to produce!(a) normal incident longitudinal wave beams for what
are generally termed straight beam examinations or
(b) angle beam longitudinal waves, where both
refracted longitudinal and shear waves are present in the
material under examination. 4hen used for thic#ness
measurement or clad examination, these examinations are
generally con- sidered to be straight beam examinations.
4hen used for weld examinations, they are generally
termed angle beam examinations or
(c) angle beam shear waves, where incident angles in
wedges produce only refracted shear waves in the material
under examination are generally termed angle beam
exami- nations.Contact or immersion techniques may be used. 7ase
materials and"or welds with metallurgical structures pro-
ducing variable attenuations may require that longitudinal
angle beams are used instead of shear waves. Additionally,
computeried imaging techniques may enhance the detect-
ability and evaluation of indications.
0ther techniques or technology which can be demon-
strated to produce equivalent or better examination sensi-
tivity and detectability using search units with more than
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
8/24
+B
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
9/24
--
,,,,,,
,
,
,
,
,
2010 SECTION V ARTICLE 4
FI* T-4+4*2*1 NON-PIPIN CALIBRATION BLOCS
= T
'$-T
) in4 3'%* mm5'$>T
D
'$-T =$> T
C0
'$-T
T
'$-T
D
Cladding 3i present5
'$-T'$-T
Minimum dimensions
D " '$- in4 3'= mm5
1idth " ) in4 3'%* mm5!ength " = 2 0hickness
C$%i'r$"i(! B%./ %e
1e%2 T3i./!e##4 " T3i./!e##4 T Di$e"er
i!* () i!* () i!* ()
9p to ' 3-%5=?> 3'&5 or t
=?=- 3-4%5
@ver ' 3-%5 through - 3%*5 ''?- 3=(5 or t
'?( 3=5
@ver - 3%*5 through > 3'**5 = 3,%5 or t=?') 3%5
@ver > 3'**5 t ' 3-%5 :Note 3'5;
N(".3 Die!#i(!#4
i!* ()
Notch depthp -B 0 Notch
widthp'?> 3)5 ma24 Notch
lengthp ' 3-%5 min4
Minimum dimension4
ENE#A! N@0ES.
3a5 Holes shall be drilled and reamed '4% in4 3=( mm5 deep minimum+ essentially parallel to the e2amination surace43b5 Dor components e7ual to or less than -* in4 3%** mm5 in diameter+ calibration block diameter shall meet the re7uirements o 0>=>4'4,4-4
0wo sets o calibration relectors 3holes+ notches5 oriented &* deg rom each other shall be used4Alternatively+ two curved calibration
blocks may be used4
3c5 0he tolerance or hole diameter shall be '?=- in4 3*4( mm54 0he tolerance or hole location through the calibration block thickness
3i4e4+ distance rom the e2amination surace5 shall be '?( in4 3= mm54
3d5 Dor blocks less than=?> in4 3'& mm5 in thickness+ only the
'?-0 sidedrilled hole and surace notches are re7uired4
3e5 All holes may be located on the same ace 3side5 o the calibration block+ provided care is e2ercised to locate all the relectors 3holes+
notches5 to prevent one relector rom aecting the indication rom another relector during calibration4 Notches may also be in the sameplane as the inline holes 3See Appendi2 F+ Dig4 F>='54 As in Dig4 F>='+ a suicient number o holes shall be provided or both angle and
straight beam calibrations at the'?>0+
'?-0+ and
=?>0 depths4
35 Notch depths shall be '4)B0 minimum to -4-B0 ma2imum4 1hen cladding is present+ notch depth on the cladding side o the block shall
be increased by the cladding thickness+ C0 3i4e4+ '4)B0 G C0 minimum to -4-B0 G C0 ma2imum54
3g5 Ma2imum notch width is not critical4 Notches may be made by E/M or with end mills up to'?> in4 3)4> mm5 in diameter4
3h5 1eld thickness+ t+ is the nominal material thickness or welds without reinorcement or+ or welds with reinorcement+ the nominal material
thickness plus the estimated weld reinorcement not to e2ceed the ma2imum permitted by the reerencing Code Section4 1hen two or morebase material thicknesses are involved+ the calibration block thickness+ 0+ shall be determined by the average thickness o the weld alternatively+
a calibration block based on the greater base material thickness may be used provided the reerence relector si8e is based upon the
average weld thickness4
N@0E.
3'5 Dor each increase in weld thickness o - in4 3%* mm5 or raction thereo over > in4 3'** mm5+ the hole diameter shall increase'?') in4 3'4% mm54
+
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
10/24
--
,,,,
,,
,
,
,
,
,,,
,,
--
,,
,,
,
,,
---
ARTICLE 4 2010 SECTION V
FI* T-4+4*+ CALIBRATION BLOC FOR PIPE
L Nominal wallthickness 3T5
C0
Arc length
Cladding 3i present5
I Notches shall be located not closer than 0 or ' in4 3-% mm5+ whichever is greater+ to any block edge or to other notches4
ENE#A! [email protected] 0he minimum calibration block length 3!5 shall be ( in4 3-** mm5 or (0+ whichever is greater4
3b5 Dor @/ > in4 3'** mm5 or less+ the minimum arc length shall be -,* deg4 Dor @/ greater than > in4 3'** mm5+ the minimum arc
length shall be ( in4 3-** mm5 or =0+ whichever is greater4
3c5 Notch depths shall be (B0 minimum to ''B0 ma2imum4 1hen cladding is present+ notch depths on the cladding side o the block shall
be increased by the cladding thickness+ C0 3i4e4+ (B0 G C0 minimum to ''B0 G C0 ma2imum54 Notch widths shall be'?> in4 3) mm5
ma2imum4 Notch lengths shall be ' in4 3-% mm5 minimum4
3d5 Ma2imum notch width is not critical4 Notches may be made with E/M or with end mills up to'?> in4 3) mm5 in diameter4
3e5 Notch lengths shall be suicient to provide or calibration with a minimum = to ' signaltonoise ratio4
two transducer elements may be used. The demonstration
shall be in accordance with Article (, T-('6/a1.
T-451 C)r&e Gr)in M)teri)*&
3ltrasonic examinations of high alloy steels and high
nic#el alloy weld deposits and dissimilar metal welds
between carbon steels and high alloy steels and high
nic#el alloys are usually more difficult than ferritic weld
examina- tions. $ifficulties with ultrasonic
examinations can be caused by an inherent coarse-
grained and"or a directionally- oriented structure, which
can cause mar#ed variations in attenuation, reflection, and
refraction at grain boundaries and velocity changes within
the grains. &t usually is neces- sary to modify and"or
supplement the provisions of this Article in accordance
with T-('6/a1 when examining such welds in thesematerials. Additional items, which may be necessary, are
weld moc#ups with reference reflectors in the weld
deposit and single or dual element angle beam
longitudinal wave transducers.
T-452 C%#teri=e" I%).in. Te!ni$#e&
The ma8or attribute of Computeried &maging Tech-
niques /C&Ts1 is their effectiveness when used to
character- ie and evaluate indications however, C&Ts
may also be
used to perform the basic scanning functions required for
flaw detection. Computer-processed data analysis and dis-
play techniques are used in con8unction with automatic or
semi-automatic scanning mechanisms to produce two andthree-dimensional images of flaws, which provides an
enhanced capability for examining critical components
and structures. Computer processes may be used to
quantita- tively evaluate the type, sie, shape, location,
and orienta- tion of flaws detected by ultrasonic
examination or other >$? methods. $escriptions for
some C&Ts that may be used are provided in
>onmandatory Appendix ?.
T-460 CALIRATION
T-461 In&tr#%ent Line)rit Ce!3&
The requirements of T-+B(.( and T-+B(. shall be met
at intervals not to exceed three months for analog type
instruments and one year for digital type instruments, or
prior to first use thereafter.
T-461'1 S!reen Hei.t Line)rit' The ultrasonic
instrumentDs screen height linearity shall be evaluated in
accordance with %andatory Appendix &.
T-461'2 A%*it#"e Cntr* Line)rit' The ultrasonic
instrumentDs amplitude control linearity shall be evaluated
in accordance with %andatory Appendix &&.
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
11/24
+A
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
12/24
2010 SECTION V ARTICLE 4
FI* T-4+4*4*1 CALIBRATION BLOC FOR TECNIQUE ONE
A2is o clad beads
in4 3=( mm5n4 depth
''$-mi
'$') in4 3'4% mm5
sidedrilled holeJs
relecting surace
at clad interace4
tolerance "'$)> in4 3*4> mm5
Cladding
'$( in4 3= mm5 latbottom hole
drilled to clad interace4tolerance "
'$)> in4 3*4> mm5
T-462 Gener)* C)*ir)tin Re$#ire%ent&
T-462'1 U*tr)&ni! S&te%' Calibrations shall include
the complete ultrasonic system and shall be performed
prior to use of the system in the thic#ness range under
examination.
T-462'2 C)*i1r)tin S#r+)!e' Calibrations shall be per-
formed from the surface /clad or unclad convex or con-
cave1 corresponding to the surface of the component from
which the examination will be performed.
T-462', C#*)nt' The same couplant to be used dur-
ing the examination shall be used for calibration.
T-462'4 Cnt)!t We".e&' The same contact wedges to
be used during the examination shall be used for cali-bration.
T-462'5 In&tr#%ent Cntr*&' Any control which
affects instrument linearity /e.g., filters, re8ect, or clipping1
shall be in the same position for calibration, calibration
chec#s, instrument linearity chec#s, and examination.
T-462'6 Te%er)t#re' or contact examination, the
temperature differential between the calibration bloc# and
examination surfaces shall be within 'H /(+HC1. or
immersion examination, the couplant temperature for cali-bration shall be within 'H /(+HC1 of the couplant
temper- ature for examination.
T-46, C)*ir)tin +r Nn-Piin.
T-46,'1 S&te% C)*ir)tin +r Di&t)n!e A%*it#"e
Te!ni$#e&
T-46,'1'1 C)*ir)tin *!39&:' Calibrations shall
be performed uti liing the cal ibration bloc# shown in
ig. T-+;+..(.
T-46,'1'2 Te!ni$#e&' >onmandatory Appendices
7 and C provide general techniques for both angle beam
shear wave and straight beam calibrations. 0ther tech-
niques may be used.
The angle beam shall be directed toward the calibration
reflector that yields the maximum response in the area of
interest. The gain control shall be set so that this response
is 6G 'G of full screen height. This shall be the
primary reference level. The search unit shall then be
manipulated, without changing instrument settings, to
obtain the maxi- mum responses from the other calibration
reflectors at their beam paths to generate the distance-
amplitude correction
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
13/24
--``,,,,,,`,``,``,`,`````,,,`,,-`-`,,`,,`,`,,`--- +@
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
14/24
m5
ARTICLE 4 2010 SECTION V
FI* T-4+4*4*2*1 ALTERNATE CALIBRATION BLOC FOR TECNIQUE ONE
- in43%* mm5
=$> C0'$- C0
'$> C0
C0
- C0
3min5
' in4 3typ5 ' in4 3typ5
:-% mm 3typ5; :-% mm 3typ5;
ENE#A! N@0E. All latbottom holes are'?( in4 3= mm5 diameter4 0olerances or hole diameter and depth with respect to the clad side o the
block are '?)> in4 3*4> mm54
FI* T-4+4*4*2*2 ALTERNATE CALIBRATION BLOC FOR TECNIQUE ONE
- in43%* m
' in4 3typ5 ' in4 3typ5
:-% mm 3typ5; :-% mm 3typ5;=$> C0
'$- C0'$> C0
C0
- C03min5
ENE#A! N@0E. All sidedrilled holes are'?') in4 3'4% mm5 diameter4 Holes location tolerance is
'?)> in4 3*4> mm54 All holes drilled to a
minimum depth o '4% in4 3=( mm54
'6
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
15/24
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
16/24
'( --``,,,,,,`,``,``,`,`````,,,`,,-`-`,,`,,`,`,,`---
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
17/24
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
18/24
'
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
19/24
2010 SECTION V ARTICLE 4
the scanning surface so as to scan the entire examination
volume for each required search unit.
(a) ?ach pass of the search unit shall overlap a
minimum of (6G of the transducer /pieoelectric element1
dimension parallel to the direction of scan indexing. As an
alternative, if the sound beam dimension parallel to the
direction of scan indexing is measured in accordance with
>onmanda- tory Appendix 7, 7-+BB, 7eam Spread
measurement rules, each pass of the search unit may
provide overlap of the minimum beam dimension
determined.
(b) 0scillation of the search unit is permitted if it can
be demonstrated that overlapping coverage is provided.
T-471'2 P#*&e Reetitin R)te' The pulse repetition
rate shall be small enough to assure that a signal from a
reflector located at the maximum distance in the examina-
tion volume will arrive bac# at the search unit before the
next pulse is placed on the transducer.
T-471', R)te + Se)r! Unit M8e%ent' The rate of
search unit movement /scanning speed1 shall not exceed
B in."s /('6 mm"s1, unless!
(a) the ultrasonic instrument pulse repetition rate is suf-
ficient to pulse the search unit at least six times within the
time necessary to move one-half the transducer
/pieoelec- tric element1 dimension parallel to the
direction of the scan at maximum scanning speed or,
(b) a dynamic calibration is performed on multiple
reflectors, which are within d7 of a static calibration
and the pulse repetition rate meets the requirements of
T-+int Di&t)n!e A%*it#"e
Te!-ni$#e
4hen the referencing Code Section specifies a distance
amplitude technique, weld 8oints shall be scanned with an
angle beam search unit in both parallel and transverse
directions /+ scans1 to the weld axis. 7efore performingthe angle beam examinations, a straight beam examination
shall be performed on the volume of base material through
which the angle beams will travel to locate any reflectors
that can limit the ability of the angle beam to examine
the weld volume. >onmandatory Appendix & describes a
method of examination using multiple angle beam
search units.
T-472'1 An.*e e)% Te!-ni$#e
T-472'1'1 e)% An.*e' The search unit and beam
angle selected shall be +' deg or an angle appropriate for
the configuration being examined and shall be capable of
detecting the calibration reflectors, over the required angle
beam path.
T-472'1'2 Re+*e!tr& P)r)**e* t te We*" Se)%'
The angle beam shall be directed at approximate right
angles to the weld axis from both sides of the weld /i.e.,
from two directions1 on the same surface when possible.
The search unit shall be manipulated so that the ultrasonic
energy passes through the required volume of weld and
ad8acent base material.
T-472'1', Re+*e!tr& Tr)n&8er&e t te We*"
Se)%' The angle beam shall be directed essentially
parallel to the weld axis. The search unit shall be
manipulated so that the ultrasonic energy passes through
the required volume of weld and ad8acent base material.The search unit shall be rotated (6 deg and the
examination repeated. &f the weld cap is not machined
or ground flat, the examination shall be performed from
the base metal on
both sides of the weld cap in both weld axis directions.
T-472'2 Re&tri!te" A!!e&& We*"&' 4elds that cannot
be fully examined from two directions using the angle
beam technique /e.g., corner and tee 8oints1 shall also be
examined, if possible, with a straight beam technique.
These areas of restricted access shall be noted in the
exami- nation report.
T-472', In)!!e&&i*e We*"&' 4elds that cannot beexamined from at least one side /edge1 using the angle
beam technique shall be noted in the examination report.
or flange welds, the weld may be examined with a
straight beam or low angle longitudinal waves from the
flange face provided the examination volume can be
covered.
T-47, C*)""in. Te!ni$#e&
The techniques described in these paragraphs shall be
used when examinations of weld metal overlay cladding
(1
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
20/24
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
21/24
yright ASME Internationalided by IHS under license with ASME
eproduction or networking permitted without license rom IHS
!icensee"Savannah #iver Nuclear Solutions$%&''())**'
Not or #esale+ *,$-,$-*'* ''.*,.%- M/0
ARTICLE 4 2010 SECTION V
(10)
are required by a referencing Code Section. 4hen
examina- tion for lac# of bond and clad flaw indications is
required, Technique 0ne shall be used. 4hen
examination for lac# of bond only is required, Technique
Two may be used.
T-47,'1 Te!ni$#e One' The examination shall be per-formed from the clad surface with the plane separating the
elements of the dual element search unit positioned
parallel to the axis of the weld bead. The search unit shall
be moved perpendicular to the weld direction.
T-47,'2 Te!ni$#e T
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
22/24
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
23/24
2010 SECTION V ARTICLE 4
(h) computeried program identification and revision
when used
(i) calibration bloc# identification
(j) simulation bloc#/s1 and electronic simulator/s1 iden-
tification when used
(k) instrument reference level gain and, if used,
damping and re8ect setting/s1
(l) calibration data Iincluding reference reflector/s1,
indication amplitude/s1, and distance reading/s1J
(m) data correlating simulation bloc#/s1 and electronic
simulator/s1, when used, with initial calibration
(n) identification and location of weld or volume
scanned
(o) surface/s1 from which examination was conducted,
including surface condition
() map or record of re8ectable indications detected or
areas cleared
(!) areas of restricted access or inaccessible welds
(r) examination personnel identity and, when required
by referencing Code Section, qualification level
(s) date of examination.
&tems (b) through (m) may be included in a separate
calibration record provided the calibration record identifi-
cation is included in the examination record.
T-4@, Rert
A report of the examinations shall be made. The report
shall include those records indicated in T-+@( and T-+@.
The report shall be filed and maintained in accordance
with the referencing Code Section.
-
8/10/2019 ARTICLE 4 METODO ULTRASONIDO.doc
24/24