Background The Hemochron Configuration Manager (HCM) software program allows the customer to configure their Hemochron Signature Elite instruments to assist in meeting the most common needs of their facility’s indiidual compliance program! The configuration is performed on a personal computer with installed HCM software! The general configuration options aaila"le and their intended purposes are summari#ed in the Hemochron Signature Elite $hole Blood Microcoagulation System %perator’s Manual! The &'' eature %ne feature aaila"le in the Configuration Manager is the &'' limit! This setting is utili#ed alo ng with the Electronic uality Control (EC) and *i+uid uality Control (*C) lockouts ! ,f &'' limits are ena"led- the user wi l  "e offer ed the choice to run a patient test if the *C interal has "een e.ceeded! The allowa"le num"er of tests that can "e  perfor med i s sele cted within C onfigu ration Manage r (fro m / to 01/)! The ins trumen t disp lays the nu m"er o f remai ning tests that can "e run (the &'' oerride) "efore C must "e performed! Setting the &'' feature would also allow for patient testing to "e performed in the eent that an EC or *C test has failed! ,n the eent the facility does not want the a"ility to run patient testing when either the C interal has elapsed or a C has failed- the &'' feature should "e set to #ero (/)! 2dditional ,nformation and Contact ,nformation Each facility should identify the features aaila"le in the Configuration Manager software program that would "est comply with their +uality assessment plan! 2ny additional +uestions regarding the use of the Configuration Manager software and the function of the Hemochron Elite should "e referred to 2ccria3s 4istri"utor 5artner- ..... Support 4epartment at 666766676666!

Upload: kalletiser

Post on 09-Jan-2016

4 views

Category:

Documents


0 download

DESCRIPTION

The Hemochron Configuration Manager (HCM) software program

TRANSCRIPT

7/17/2019 Background

http://slidepdf.com/reader/full/background-5690c0510af6d 1/1

Background

The Hemochron Configuration Manager (HCM) software program allows the customer to configure their Hemochron

Signature Elite instruments to assist in meeting the most common needs of their facility’s indiidual compliance program! The

configuration is performed on a personal computer with installed HCM software! The general configuration options aaila"le

and their intended purposes are summari#ed in the Hemochron Signature Elite $hole Blood Microcoagulation System

%perator’s Manual!

The &'' eature

%ne feature aaila"le in the Configuration Manager is the &'' limit! This setting is utili#ed along with the Electronic

uality Control (EC) and *i+uid uality Control (*C) lockouts! ,f &'' limits are ena"led- the user wi l "e offered

the choice to run a patient test if the *C interal has "een e.ceeded! The allowa"le num"er of tests that can "e

 performed is selected within Configuration Manager (from / to 01/)! The instrument displays the num"er of remaining

tests that can "e run (the &'' oerride) "efore C must "e performed!

Setting the &'' feature would also allow for patient testing to "e performed in the eent that an EC or *C test has failed! ,n

the eent the facility does not want the a"ility to run patient testing when either the C interal has elapsed or a C has failed-

the &'' feature should "e set to #ero (/)!

2dditional ,nformation and Contact ,nformation

Each facility should identify the features aaila"le in the Configuration Manager software program that would "est

comply with their +uality assessment plan! 2ny additional +uestions regarding the use of the Configuration Manager 

software and the function of the Hemochron Elite should "e referred to 2ccria3s 4istri"utor 5artner- ..... Support