  • 7/29/2019 Bicon Brevis Technique Manual

    1/16

    Technique Manual

    Brevis Overdenture Abutment

  • 7/29/2019 Bicon Brevis Technique Manual

    2/16

  • 7/29/2019 Bicon Brevis Technique Manual

    3/16

  • 7/29/2019 Bicon Brevis Technique Manual

    4/16

    1

    Brevis Chairside Technique

    Uncover implants2

    A registration of the dentures occlusal

    relationship prior to the uncovering ofthe implants will assure that the denture

    is not inappropriately displaced by the

    abutment or its housing.

    occlUsal registration1

    2

    Use a shoulder depth gauge to facilitate the selection of the abutment height.

    Use a shoUlder depth gaUge to determine abUtment height3

    Uncover implants

    Place guide pins into the implants todetermine their axial inclinations.

    place gUide pins4

    Uncover implants2

    Uncover each implant using a small

    crestal incision and use the healing plugremoval instrument to facilitate theremoval of the black healing plug.

    Uncover implants2

  • 7/29/2019 Bicon Brevis Technique Manual

    5/16

    2

    Brevis Chairside Technique

    Rotate a combination of 0 and/or 15 angled abutments to achieve parallelism prior to their being

    seated. Brevis abutments are available in heights of 2.0, 4.0 and 6.0mm.

    abUtments5a parallel abUtments5b

    Place soft wax in the denture to act as a pressure indicator to determine the relative position of theabutments. Alternatively, the top of the abutment may be marked with a felt tip pen to indicate thelocation of the abutment on the denture.

    place blUe wax6a mark abUtments6b

    Place a Brevis housing on each abutmentintra-orally.

    brevis hoUsing7

    Liberally relieve denture to accommodatethe Brevis housings. Confirm clearancefor housings by placing the denture overthe housings.

    relieve dentUre8

    4.0 15 6.0 04.0 02.0 152.0 0 6.0 15

    2.0 0 6.0 156.0 04.0 154.0 02.0 153.0mmP

    ost

    2.0mmP

    ost

  • 7/29/2019 Bicon Brevis Technique Manual

    6/16

    3

    Brevis Chairside Technique

    Remove the Brevis housings andplace a piece of rubber dam over theabutments to act as an apron to preventacrylic from locking onto an abutment.nsure that the ruer a rtetsthe unerut the autent raryl, y lang t ave theshuler the autent.

    remove hoUsings9

    Inject Vaseline under the rubberdam aprons to serve as an additionalprecaution to prevent acrylic fromlocking under the abutments during thechairside pickup of the Brevis housings.

    inject vaseline10

    Place the denture into the mouth andinstruct patient to lenh laterally nttn rlls t assure rer seatng the enture.

    assUre proper seating12

    Inject flowable acrylic around the Brevishousings and into the relieved portionsof the existing denture.

    inject acrylic11

  • 7/29/2019 Bicon Brevis Technique Manual

    7/16

    4

    Use an occlusal registration jig to prevent inadvertent displacement of the denture during

    the chairside technique.

    Use the 15 Brevis abutment to help achieve parallelism for non-parallel implants.

    The denture should NOT rock or pivot on the abutments or the housings.

    Do not make a direct impression of the overdenture abutments.

    Use a rubber dam and Vaseline to prevent locking of the denture beneath the undercut

    of the abutments.

    Place acrylic into a syringe for ease of use and greater control.

    A too viscous mix of acrylic may inappropriately displace the orientation of the Brevis housing

    on the abutment.

    It is essential for the patient to clench bilaterally on cotton rolls to ensure proper seating of the

    housings in the denture.

    If the denture is too retentive, slightly relieve the inside of the o-ring lumen with a round bur.

    A common cause of accelerated o-ring wear is a Brevis housing whose retentive acrylic was

    polymerized while the Brevis housing was not appropriately aligned on the abutment.

    3keys t uess

    Note: If the denture is inadvertently locked onto the Brevis abutment, it is advisable to tap it off rather than

    attempting to cut it off.

    The denture may be notched to facilitate the placement of a tapping instrument. Either the denture will

    be removed from the abutment or the abutment will be removed from the implant.

    Brevis Chairside Technique

    Polish the excess acrylic around theBrevis housings after the removal of therubber dam.

    polish excess acrylic13

    Radiograph of two Bicon implants andBrevis abutments. it s araunt thatverentures e letely tssuerne an nly lant retane.i the aryl s t vsus r nlylae n the enture, t ay auseslaeent the husng resultngn a salgnent the husng an

    exessve wear the ruer -rngs.

    radiograph14

  • 7/29/2019 Bicon Brevis Technique Manual

    8/16

    5

    Indirect Transfer Technique

    Uncover implants2

    A registration of the dentures occlusal

    relationship prior to the uncovering ofthe implants will assure that the denture

    is not inappropriately displaced by the

    abutment or its housing.

    occlUsal registration1

    2

    Use a shoulder depth gauge to facilitate the selection of the abutment height.

    Use a shoUlder depth gaUge to determine abUtment height3

    Uncover implantsplace gUide pins4

    Uncover implants2

    Uncover each implant using a small

    crestal incision and use the healing plugremoval instrument to facilitate theremoval of the black healing plug.

    Uncover implants2

    Place guide pins into the implants todetermine their axial inclinations.

  • 7/29/2019 Bicon Brevis Technique Manual

    9/16

    6

    Indirect Transfer Technique

    Rotate a combination of 0 and/or 15 angled abutments to achieve parallelism prior to their being

    seated. Brevis abutments are available in heights of 2.0, 4.0 and 6.0mm.

    abUtments5a parallel abUtments5b

    4.0 15 6.0 04.0 02.0 152.0 0 6.0 15

    2.0 0 6.0 156.0 04.0 154.0 02.0 153.0mmP

    ost

    2.0mmP

    ost

    Seat black plastic impression caps ontothe Brevis abutments.

    seat impression caps6

    Make impression using Brevisimpression caps. d nt ae a retressn the ttanu autents.

    make impression Using caps7

    Block out undercuts on the model.

    block oUt UndercUts9

    Insert the aluminum transfer dies into theimpression caps prior to the pouring of amaster stone model.

    insert transfer dies8

  • 7/29/2019 Bicon Brevis Technique Manual

    10/16

    7

    Removing/Inserting Rubber O-Ringinto Brevis Housing

    Remove rubber o-rings with a scaling

    instrument or explorer.

    remove hoUsings1

    Insert the explorer in between the o-ring

    and Brevis housing and loosen theo-ring from the housing.

    insert explorer2

    Using cotton pliers, squeeze the o-ringinto a figure eight and insert it intothe housing.

    sqUeeze o-ring3

    Place the entire o-ring into the retentivegroove within the Brevis housing.

    place o-ring4

    Uncover implants

    If the o-ring is fully inserted, yet not securein the retention lip of the housing, use anexplorer to move the o-ring into place.

    move o-ring into place5

  • 7/29/2019 Bicon Brevis Technique Manual

    11/16

    8

    The design of the Bicon abutment system is such that the hemispherical base of the abutment

    does not sit ush with the neck of the implant. By design, there is a space below the abutmentpost when the abutment is fully seated. When viewing this on a radiograph, it can be seen as aradiolucency. please nte that the st any autent ay nt e fe. lterng theautent st n any anner r usng a eentng eu wll aet the rtnal ft the lng taer resultng n a tentally nn-retentve autent. The following diagramdepicts the nal seating of a Brevis abutment as well as the method for measuring each abutment.

    3Measurement Guide

    The height of the Brevis abutment is measured from thetop of the implant to the shoulder of the Brevis abutment.

    The available heights are 2.0, 4.0 or 6.0mm. The total heightfrom the top of the implant to the top of the abutment isapproximately 5.0, 7.0 or 9.0mm, respectively.

    brevs utent

    Rubber O-Ring Maintenance

    Each patient maintains and inserts his denture dierently. The o-ring should be changed every 6 to 24

    months depending upon its wear.

    The o-rings should not be soaked in a solution with high alcohol content. Some cleaning solutions

    may dry out the rubber o-rings and cause them to lose retention faster than normal.

    An inappropriately aligned housing will result in excessive wear of the rubber o-ring.

    If excessive wear of o-ring is noted, remove and re-align the housing in denture with a suciently

    owable acrylic injected onto the housing and into the relieved denture.

    3keys t uess

    Height

  • 7/29/2019 Bicon Brevis Technique Manual

    12/16

    9

    Chairside Brevis Technique

    Place soft wax to record relative position

    of abutments.

    1

    Record position of abutments in soft wax.

    Alternatively, abutments could be markedwith a wet ink prior to inserting denturefor recording an ink transfer mark.

    2

    Mark position of abutments on denture.

    3

    Liberally relieve denture to accommodateBrevis housings and confirm positionintra-orally.

    4

    Place rubber dam apron over abutments. Apply Vaseline under rubber dam apron.

    65

  • 7/29/2019 Bicon Brevis Technique Manual

    13/16

    10

    Snap Brevis housing onto abutments. Sufficiently apply fluid acrylic to Brevis

    housing to prevent displacement ofhousing upon insertion of denture.

    Place minimal acrylic into relieved denture.

    9

    Patient clenches bilaterally on cotton rollswhile acrylic sets.

    10

    Polish excess acrylic from denture andremove palatal coverage for maxillarydentures.

    11

    Chairside Brevis Technique

    7 8

  • 7/29/2019 Bicon Brevis Technique Manual

    14/16

    11

    Place plastic alignment caps on

    abutments and record impression.

    1

    Insert aluminum transfer dies into

    alignment caps in the impression andpour the master model.

    2

    Snap Brevis housing on to the aluminumtransfer die and block out undercut withstone.

    3

    Indirect Transfer Technique

    Brevis Abutment

    2.0mm

    3.0mm

    3.0mm

    4.0mm

    3.0mm

    6.0mm

    Guida alla misurazione Example

    The height of the Brevis

    abutment is measured from

    the top of the implant to

    the shoulder of the Brevisabutment. The available

    heights are 2.0, 4.0 or 6.0mm.

    The total height from the top

    of the implant to the top of the

    abutment is approximately 5.0,

    7.0 or 9.0mm, respectively.

  • 7/29/2019 Bicon Brevis Technique Manual

    15/16

    12

    Brevis Abutments*

    2.0mm PostDescription Height Angle Post Part Number 0 15

    2.0mm 0 Brevis Abutment 2.0mm Post 2.0mm 0 2.0mm 260-100-404

    2.0mm 15 Brevis Abutment 2.0mm Post 2.0mm 15 2.0mm 260-100-405

    4.0mm 0 Brevis Abutment 2.0mm Post 4.0mm 0 2.0mm 260-100-406

    4.0mm 15 Brevis Abutment 2.0mm Post 4.0mm 15 2.0mm 260-100-407

    6.0mm 0 Brevis Abutment 2.0mm Post 6.0mm 0 2.0mm 260-100-408

    6.0mm 15 Brevis Abutment 2.0mm Post 6.0mm 15 2.0mm 260-100-409

    Brevis Abutments*3.0mm Post

    Description Height Angle Post Part Number 0 15

    2.0mm 0 Brevis Abutment 3.0mm Post 2.0mm 0 3.0mm 260-300-434

    2.0mm 15 Brevis Abutment 3.0mm Post 2.0mm 15 3.0mm 260-300-435

    4.0mm 0 Brevis Abutment 3.0mm Post 4.0mm 0 3.0mm 260-300-436

    4.0mm 15 Brevis Abutment 3.0mm Post 4.0mm 15 3.0mm 260-300-437

    6.0mm 0 Brevis Abutment 3.0mm Post 6.0mm 0 3.0mm 260-300-438

    6.0mm 15 Brevis Abutment 3.0mm Post 6.0mm 15 3.0mm 260-300-439

    *Each abutment is packaged with one titanium Brevis housing.

    Restorative Components

    Description Part Number

    Brevis Abutment Chairside Kit 260-100-212 Rubber O-Ring Brevis Housing

    Brevis Impression Kit without Housing 260-100-218

    Impression Cap Aluminum Transfer Die

    Brevis Abutment Impression Kit 260-100-217

    Rubber O-Ring Brevis Housing Impression Cap Aluminum Transfer Die

    Brevis Rubber O-Rings (10) 260-100-014 Rubber O-Ring

    Brevis Abutment System

  • 7/29/2019 Bicon Brevis Technique Manual

    16/16

    Copyright 2008 Bicon LIT-255 R1208

    WORLD HEADQUARTERS

    501 rrway

    bstn, m 02130 U

    tel: 800.88.bico u 617.524.4443

    ax: 800.28.bico u 617.524.0096

    we: www.n.

    e-al: surt@n.