cah_bhme_whitepaper
TRANSCRIPT
In-Vitro (Bench) Analysis of Bypass HMEs
Introduction
Bypass Heat Moisture and Exchangers (BHMEs)
simplifies aerosol delivery to ventilated patients
utilizing passive humidification without breaking
the circuit. A BHME provides the ability to optimize
aerosol delivery by positioning the medication
delivery device in the most ideal position of 6” to
18” back from the patienti.
An ideal or perfect Bypass HME would not influence
an aerosolized particle, simply letting it pass
through the device unchanged. Quantifying the
change of aerosol through BHMEs is a complex
process dependent upon a variety of clinical and
device-related variables. Patient breathing
patterns, the choice and dilution of drug and
device selection all have the potential to affect
aerosol. In-vitro (benchmark) testing can provide
the clinician with useful information regarding the
bypass performance.
Purpose
The purpose of this technical bulletin is to illustrate
the in-vitro (bench) bypass performance
characteristics of the AirLife® BHMEs. Two
different test methods were used to demonstrate
the aerosol impact on the BHME called the
Andersen Cascade Impactor and Laser Diffraction
Test.
Andersen Cascade Impactor Analysis
Products Tested
Testing was conducted on the following products:
� 003020, AirLife® Adult Nonfiltered Bypass Heat
and Moisture Exchanger (BHME)
� 003021, AirLife® Adult Filtered Bypass Heat and
Moisture Exchanger (BHME)
� No HME in-line. Dry Circuit only. Illustrate what
would be the perfect “BHME”
Particle Size Test Method
An Andersen eight-stage cascade impactor (ACI)
with USP inlet was used and assayed with a
spectrophotometer (277 nm). Particle size
characterization was performed using an
albuterol/saline solution. Cascade impaction was
chosen for the following reasons:
� Cascade impaction measures aerodynamic
diameter directly, which accounts for the density
and irregular shape of drug particles. It is
believed that aerodynamic diameter more
accurately predicts the behavior of aerosol as it
is delivered into the patient’s lungs
� There is more historical data on particle size
measurement using cascade data than any other
method. Relative comparison with historical data
can be readily made.
� Cascade impaction is one of the USP methods for
characterization of particle distributions.
Definition of Respirable Fraction
“Respirable fraction” is the percent of aerosol
generated, by mass, which falls below an
aerodynamic diameter of 5µm. It has been
reported that particles less than 5µm will penetrate
beyond the upper airways and deposit into the
tracheobronchial and pulmonary regions of the
lung. This measurement is often used to describe
the quality of aerosol. The mass collected between
the cascade plate cutoff points of 0.4µm and
4.7µm were utilized to quantify the mass below
5µm.
Aerosol Optimization
and the AirLife® BHME
Andersen Cascade Impactor Analysis
91.2% 91.1% 90.8%
75.0%
80.0%
85.0%
90.0%
95.0%
100.0%
% M
ass
of
Alb
ute
rol
Su
lfa
te
be
twe
en
0.4
- 4
.7 μ
m
No Device 003020 003021
Conclusions
There is no statistically significant difference in the respirable fraction between delivering aerosol through the
AirLife® BHMEs (003020 & 003021) and delivering aerosol with the devices removed (p-values > 0.445, 95%
confidence).
Laser Diffraction Test for Aerosol
Products Tested
Testing was conducted on the following test
groups:
� 003020, AirLife® Adult Nonfiltered Bypass Heat
and Moisture Exchanger (BHME)
� 003021, AirLife® Adult Filtered Bypass Heat
and Moisture Exchanger (BHME)
� 19912, Gibeck Humid-Flo® Heat and Moisture
Exchanger
� 68-1000, CircuVent® HME/HCH Bypass
(Assembled with AirLife 003005, Adult Filtered
HCH)
� No HME in-line. Dry Circuit only. Illustrate
what would be the perfect “BHME”
Particle Size Test Method
Laser diffraction was performed using a Malvern
Spraytech with an inhalation cell. A constant
extraction flow rate of 14 Lpm through the Malvern
Spraytech was used. Particle size characterization
was performed using an albuterol/saline solution.
Laser diffraction was chosen because the laser
diffraction test method is less time consuming than
the cascade impactor test method, facilitating
testing of competitive products.
Conclusions
� There is a statistically significant difference in
the respirable fraction between delivering
aerosol through the any of the tested devices
and delivering aerosol with the devices
removed (p-values < 0.002, 95% confidence).
� The AirLife® BHMEs (003020 & 003021)
influence the respirable fraction less than the
DHD CircuVent® (p-values < 0.001, 95%
confidence).
� There no statistically significant difference in
the respirable fraction between delivering
aerosol through the AirLife® 003021 and
delivering aerosol through the Gibeck Humid-
Flo® (p-value = 0.638, 95% confidence).
Summary Conclusion
In either test method, the AirLife® BHME has comparable performance to using no HME device on
ventilated patients and should be considered minimal risk of aerosol knockdown to ventilated
patients.
i Patrick J Dunne MEd RRT FAARC (December 2004). Respiratory Care Protocols: Benefits for Patients, Therapists and Hospitals. www.aarc.org/protocol_course/protocols_book.pdf
James B Fink, MS RRT, Martin J Tobin MD, and Rajiv Dhand MD. Bronchodilator Therapy in Mechanically Ventilated Patients Respiratory Care. Jan 1999: 44-1, pg. 57.
Timothy B. Op’t Holt, EdD, RRT. Ventilation for Life. Aerosol Therapy during Mechanical Ventilation. AARC Times. July 2000: pg. 22.
Laser Diffraction Test for Aerosol
80.8%
85.7%
82.9%83.3%
88.2%
50
55
60
65
70
75
80
85
90
95
100
% V
olu
me
< 5
.0 μ
m
No HME
AirLife 003020
AirLife 003021
Gibeck Humid-Flo
DHD CircuVent
© Cardinal Health, Inc. or one of its subsidiaries. All rights reserved. AirLife is a registered trademark of Cardinal Health, Inc. or one of its subsidiaries. Humid-Flo is a registered trademark of Teleflex Medical or one of its subsidiaries. CircuVent is a registered trademark of Smiths Medical one of its subsidiaries. Lit. No. 3RT2925
Cardinal Health 22745 Savi Ranch Pkwy. Yorba Linda, CA 92887 www.cardinalhealth.com/respiratory