calibration and metrology
DESCRIPTION
Calibration and MetrologyTRANSCRIPT
-
7Calibration and Metrology
Go
ran Bringert
Independent Consultant, North Andover, Massachusetts, U.S.A.
INTRODUCTION
The regulatory authorities in the U.S.A. and Europe have
expanded the scope of regulations by reference to inter-
national standards for Quality and Risk Management.
The GMP regulations are still in effect, however the
enforcement focuses on the critical risk factors
determined by risk assessments.
In September 2004, FDA issued three documents
describing the FDAs current thinking:
&
Draft Guidance for IndustryQuality Systems
Approach to Pharmaceutical Current Good Manufac-
turing Practice Regulations (1)
&
Guidance for IndustrySterile Drug Products
Produced by Aseptic ProcessingCurrent Good
Manufacturing Practice (2)
&
Pharmaceutical CGMPs for the 21st CenturyA Risk-
Based Approach Final Report (3)
And Medical Devicespecic international
standards:
&
ANSI/AAMI/ISO 13485:2003 Medical Devices
Quality management systemsRequirements for
regulatory purposes (4)
&
ANSI/AAAMI/ISO 14971:2000/A1: 2003 Medical
devicesApplication of risk management to
medical devices (5)
All of the above documents refer either directly or
indirectly to the following documents as standard
requirements for calibration:
&
ISO 9001:2000 Quality management systems
Requirements (6)
&
ISO 10012: 2003 Measurement management
systemsRequirements for measurement processes
and measuring equipment (7)
&
ISO/IEC 17025:2005 General requirements for
the competence of testing and calibration labora-
tories (8)
&
GUM Guide to the Expression of Uncertainty in
Measurement, (corrected and reprinted 1995)
issued by BIPM, IEC, IFCC, ISO, IUPAC, IUPAP,
and OIML (9)
&
ANSI/NCSL Z540-2-1997 U.S. Guide to the
Expression of Uncertainty in Measurement (10)
The GMPs in Europe and the U.S.A. have common
objectives but differ to some degree in their approaches
and have specic regulatory requirements for human and
veterinary drugs and medical devices.
Validation veries that processes perform to speci-
cations. Specications serve to denewhat performance is
needed for a consistent quality output from the process.
Measurements of critical parameters are needed to judge
theperformance of theprocess. Themeasurements have to
be accurate and repeatable. Accurate and repeatable
measurements require adequately calibrated good
quality measurement equipment. Current regulations
and standards do not specify accuracy requirements as
processes and accuracy specications vary widely across
the industry. The regulations hold the organizations of the
regulated industries responsible to set specications and
tolerances for calibrations and to verify that calibration
laboratories/providers have the competency required for
compliance. The reader should keep in mind that the
standards and regulations cover implementation
of Quality Management Systems and that calibration is
only one of the several components of the Quality
Management System.
Validation of Pharmaceutical Processes is not
possible without reliable and repeatable measurements.
The Predicate Rules (G!Ps) require that critical
measurements be performed with adequately calibrated
measuring devices.
Heat Penetration Studies are performed to calculate
the accumulated lethality, F
0
, in the load. The accumulated
F
0
is the time integral of the lethality function:
LZ 10
TKT
b
=z
At a base temperature T
b
Z1218C and zZ108C, the
effect of 18C error in measured temperature at 1218C
results in approximately 26% error in the lethality
calculation.
This is why in section IX sub-clause C 2 Equipment
Controls and Instrument Calibration, of the 2004 Guidance
for IndustrySterile Drug Products Produced by
Aseptic ProcessingcGMPs Validation of Aseptic
Processing and Sterilization (2), FDA states:
For both validation and routine process control,
the reliability of the data generated by sterilization
cycle monitoring devices should be considered to be
of the utmost importance. Devices that measure cycle
parameters should be routinely calibrated. Written
procedures should be established to ensure that these
devices are maintained in a calibrated state.
Temperature and pressure monitoring devices for
heat sterilization should be calibrated at suitable
Abbreviations used in this chapter: CAPA, corrective action, preventive
action; CFR, Code of Federal Regulations; cGMPs, current good
manufacturing practices; FDA, Food and Drug Administration;
GMP, good manufacturing practice; GUM, Guide to the Expression
of Uncertainty in Measurement; NIST, National Institute of Stan-
dards and Technology; RTD, resistance temperature detector; SOP,
standard operating procedures; TUR, test uncertainty ratio.
-
intervals. The sensing devices used for validation
studies should be calibrated before and after validation
runs.
ANSI/ISO/DIS 17665:2004 Sterilization of health care
productsMoist heatDevelopment, validation and routine
control of a sterilization process for medical devices (11),
requires that calibration of the validation system, measuring
chain, shall be veried before and after each stage
of validation.
These statements are a clear indication that regulat-
ory authorities consider the integrity of temperature
measurements a critical part of the validation of thermal
sterilization processes. It is important that the validation
SOP reects the theoretical and practical aspects of how
to achieve and maintain high-accuracy temperature
measurements in conjunction with thermal validation.
Due to the large number of topics covered by the
current standards and regulations, this chapter will focus
on the practical issues of calibration based on the assump-
tion that the organization has a quality management
system that is compliant with applicable requirements.
The reader should refer to the complete regulatory
documents, referenced in this chapter, to fully assess
what has to be in place within the readers organization
for a competent and compliant calibration function for
the organizations specic application area. Depending
on the origin, some of these documents are available free
from the Internet while all ISO standards and regulations
are copyright protected and can be purchased as hard
copy or downloadable from any national ISO-afliated
organization.
APPLICATION SPECIFIC REGULATION
(PREDICATE RULES)
U.S. FDA GMP
&
21 CFR Part 58 Good Laboratory Practice for Noncli-
nical Laboratory Studies (12)
&
21 CFR Part 211 Current GoodManufacturing Practice
for Finished Pharmaceuticals (13)
&
21 CFR Part 606 Current Good Manufacturing Prac-
tice for Blood and Blood Components (14)
&
21 CFR Part 820 FDA, Subchapter HMedical
Devices, Quality System Regulation (15)
For full text, go to (16) and under Reference Room
select Code of Federal Regulation.
Europe GMP
&
The European GMP is published in EUDRALEX
Volume 4Medicinal Products for Human and Veter-
inary Use: GMP (17).
Medical Devices
There are three current standards specic to medical
devices recognized by the FDA and EU:
&
ANSI/AAMI/ISO 13485:2003 Quality System
Requirements for Medical Devices (4)
&
ANSI/AAMI/ISO 14971:2000/A1:2003 amendment,
Medical DevicesApplication of risk management
to medical devices (5)
&
ANSI/ISO/DIS 17665:2004 Sterilization of health care
productsMoist HeatDevelopment, validation and
routine control of a sterilization process for medical
devices (11); a draft standard expected to revise and
replace ISO 11134:1994, ISO 13683:1997, and CEN
554:1994 to keep current with technology and moist
heat sterilization practices
&
ISO 17025 has two main sections, Management
requirements and Technical requirements. Section 4,
Management requirements, has 15 sub-clauses
covering eight pages and Section 5, Technical require-
ments, has 10 sub-clauses covering 14 pages. It is
beyond the scope of this chapter to review the detailed
requirements, whichwill be left to the reader to review
as needed.
Corrective Action/Preventive Action (CAPA)
ISO 9001 (6), ISO 10012 (7), ISO 17025 (8), and GMPs have
mandatory requirements for CAPA.
GMPs (PREDICATE RULES)
Regulatory Requirements for Calibration in Europe
and the United States
In Guidance for IndustrySterile Products Produced by
Aseptic ProcessingcGMP, September 2004 (2), the
FDA has included text boxes with quotes from the CFR
and stated:
The quotes included in the text boxes are not intended
to be exhaustive. Readers of this document should
reference to complete CFR to ensure that they have
complied, in full, with all relevant sections of the
regulations.
GMP Calibration Requirements
This section will identify calibration requirements, as of
April 2004, dened by CGMP regulations, the Predicate
Rules including the 21CFR 820 Quality System
Regulations, and EU regulations.
&
21 CFR part 58Good Laboratory Practice for Noncli-
nical Laboratory Studies (12)
&
21 CFR part 211Current Good Manufacturing Prac-
tices for Finished Pharmaceuticals (13)
&
Sec. 211.68 Automatic, mechanical, and electronic
equipment.
&
Sec. 211.160 Laboratory ControlsGeneral require-
ments.
&
Sec. 211.165 Testing and release for distribution.
&
Sec. 211.194 Laboratory records.
&
21 CFR part 606Current Good Manufacturing
Practice for Blood and Blood Components (14)
&
Sec. 606.60 Equipment
&
Sec. 606.100 Standard Operating Procedures
&
Sec. 606.160 Records
&
21 CFR part 820 Quality System Regulation (15)
&
Sec. 820.72 Inspection, measuring and test equipment
ISO/DIS 17665:2004
Sterilization of health care productsMoist heat
Development, validation and routine control of a ster-
ilization process for medical devices (11).
100 II: SUPPORT AND UTILITY SYSTEMS
-
Summary, Standards, and Regulations
The standards and regulations, in the preceding
summary, require that organizations, involved in the
validation of processes, have personnel with a thorough
knowledge and understanding of what is needed to
achieve and maintain compliance. Calibration is critical,
as reliability of the data generated by sterilization cycle
monitoring devices should be considered to be of the utmost
importance. (FDA 2004 Guidance for IndustrySterile
Drug Products Produced by Aseptic ProcessingcGMP
Validation of Aseptic Processing and Sterilization) (2).
ISO 17025-2005 (8) provides general requirements
for the competence of testing and calibration laboratories
and the GUM (9) clause 1.1 states:
This Guide establishes general rules for evaluating
and expressing uncertainty in measurement that can be
followed at various levels of accuracy and in many
eldsfrom the shop oor to fundamental research.
Therefore, the principles of this Guide are intended to be
applicable to a broad spectrum of measurements,.
An organization that intends to comply has to
dene and document, in SOP, how the organization
achieves and maintains compliance.
Estimation and Expression of Uncertainty
(An Overview)
The Guide (9) is a complex document that covers
far reaching requirements and it should be the responsi-
bility of the metrology function within the organization to
dene the criteria for uncertainties in measurement that
are needed for specic compliance requirements. Cali-
bration is used to maintain measurement errors within
acceptable limits needed to ensure that product quality
consistently meets predened product quality speci-
cations. Historically this was done by the comparison of
howmuch ameasured value differed from a true value.
The difference was dened as the measurement error.
The international metrology community has agreed that
true values are by nature indeterminate and that is why
the term true value is not used in the ISO GUM (9). In
other words, there will always be a degree of uncertainty
in any measurement.
GUM 2.3.1 denes standard uncertainty as:
uncertainty of the result of a measurement expressed
as a standard deviation.
This illustrates standard deviation in a normal
distribution, based on denitions in the U.S. GUM
ANSI/NCSL Z540-2-1997 (10).
ANSI/ISO/DIS 17665:2004 (11) requires, and FDA
Guidance for IndustrySterileDrugProductsProducedby
Aseptic Processing 2004 (2) recommends that the measure-
ment systemused forvalidationshouldbe calibratedbefore
and after validation runs. Calibration requires the use of a
temperature standard and a stable thermal source. Each of
these items contributes some degree of uncertainty to the
calibration result in addition to the uncertainties that are
inherent in each of the components of the measuring chain
(the measurement system used for validation). The Guide
(10) denes how to combine these uncertainties into a
combined standard uncertainty for the measurement
system after calibration. A coverage factor is then used as
a multiplier of the combined standard uncertainty in order
to obtain an expanded uncertainty. If Figure 1 represents
the combined standard uncertainty of a measurement
system with normal distribution, the three condence
intervals shown in the graph represent coverage factors 1,
2 and 2.57 for 68%, 95% and 99% condence intervals
respectively.
European Standard EN 554:1994 required a
minimum TUR of 3:1 for the validation of thermal
sterilization processes. This means that the validation
system should have an expanded uncertainty that is
at least three times less than the process specication
limits. This is illustrated in Figure 2. The process under
validation has specication limitsG0.58C which, with an
assumed normal distribution, can be represented by the
standard uncertainty 1sZ0.16678C. The validation
0
0.1
0.2
0.3
0.4
0.5
-4 -3 -2 -1 0
-2.57 2.57
Measured
Value
Total Uncertainty 2.57 at the
99% Confidence Interval
Standard Deviation
Probability
Density
1 Uncertainty at the
68% Confidence Interval
2 Uncertainty at the
95% Confidence Interval
1 2 3 4
Figure 1 Standard deviation in a normal
distribution.
7: CALIBRATION AND METROLOGY 101
-
system has to have a standard uncertainty at least three
times smaller or 1sZ0.0558C. This is an example of an
adequate TUR for validation of a thermal sterilization
process.
A larger expanded uncertainty of the validation
standard increases the probability for an erroneous vali-
dation result that could have severe consequences. This is
illustrated by Figure 3.
TheU.S. version ofGUM,ANSI/NCSLZ540-2-1997,
U.S. GUM (10), is an adaptation of the ISO GUM (9) to
promote consistent international methods in the
expression of measurement uncertainty within the U.S.
standardization, calibration, laboratory accreditation, and
metrology services. The U.S. Guide is identical to the ISO
Guide (corrected and printed, 1995) with the exception of
minor editorial changes to facilitate its use in the
United States.
ISO 17025:2005 (8) makes reference to GUM for
further information on estimation of uncertainty in
measurement.
ISO 10012:2003 (7) Guidance to clause 7.3.1
Measurement uncertainty refers to the GUM for
methods that can be used to combine uncertainties and
present the results. It also states that other documented
and accepted methods may be used.
To fully understand and master the methods
referred to above, it is necessary to study the documen-
tation and integrate the information and requirements into
the users Quality Management and Risk Management
Systems, a topic beyond the scope of this chapter.
Practical Discussion of How to Calibrate the
Measuring Chain
The integrity of temperature measurements is a critical
part of validation of thermal sterilization processes. It is
important that the validation SOP reects the theoretical
and practical aspects of how to achieve and maintain
high-accuracy temperature measurements in conjunction
with thermal validation.
Heat penetration studies are performed to calculate
the accumulated lethality, F
0
, within the load items. The
accumulated F
0
is the time integral of the lethality
function:
LZ 10
TKT
b
=z
At a base temperature T
b
Z1218C and zZ108C, the
effect of 18C error in measured temperature at 1218C
results in approximately 25% error in the
lethality calculation.
The FDA denition of Process Validation is:
Establishing by objective evidence that a process
consistently produces a result or product meeting its
predetermined specications.
a
The required temperature uniformity in the
chamber, according to contemporary industry standards
for terminal sterilization, should be better than or equal to
18C or 0.58C depending on the application. The combined
standard uncertainty for the instrument used for vali-
dation measurements, including temperature sensors,
should be at least three times less than the specied for
the process variable. This means that the overall system
combined standard uncertainty should be better than or
equal to G0.338C or G0.178C, respectively. All com-
ponents involved with the measurement, (from the tip
of each sensor, via the connecting wires, cold junction
reference, signal interface, analog to digital conversion,
conversion from mill volts to temperature, to display and
printout of the measured values) are referred to as the
measuring chain. Figure 4 shows ameasuring chain using
thermocouples.
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
120.00 120.50 121.00 121.50 122.00
Temperature C
Pro
ba
bility
Disrib
utio
n
Standard
2
= 120.70C:
2
= 0.055C
Process
1
= 121.10C:
1
= 0.1667C
Figure 2 Adequate test uncertainty ratio
for validation.
a
(GMP/Medical Device Quality System Manual, 4. Process Vali-
dation, 1997).
102 II: SUPPORT AND UTILITY SYSTEMS
-
Error Sources
Several variable error sources can affect the temperature
measurement accuracy in validation. Control and
management of these error sources should be recognized
as the responsibility of the people who perform the
validation. Individuals responsible for validation should
have the competency to adequately perform the
validation studies.
It is important to distinguish between systematic
and random errors. Systematic errors are eliminated by
calibration, while random errors are not eliminated by
calibration, and can only be minimized through the
application of knowledge and proper procedures. The
operator has to understand how to minimize the inu-
ence of random temperature measurement errors to
consistently achieve the accuracy required for thermal
validation of steam sterilization processes. Procedures
documented in the validation SOP and individual
training of validation personnel are necessary to main-
tain competency of the validation team.
Electronic temperature measurements for valida-
tion are acquired using temperature sensors connected
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
120.00 120.50 121.00 121.50 122.00
Temperature C
Pro
ba
bility
Distrib
utio
n
Standard
2
= 120.70C;
2
= 0.11C
Process
1
= 121.1C;
1
= 0.1667C
Figure 3 Inadequate test uncertainty ratio
for validation.
121C
Autoclave
T/C
121C
Cold
Junction
Scanner A/D conv.
CPU
Feedthrough
Process Measurement
Overall System Accuracy
Figure 4 The measuring chain is
composed of all components
involved in the measurement
system.
7: CALIBRATION AND METROLOGY 103
-
to an electronic data logger or recorder. The components
of the measuring chain contribute errors, systematic or
random, that contribute to the Overall System Accuracy.
Figure 5 identies the most signicant Random Error
Sources (in red) in the measuring chain.
Signicant random errors can occur in the following
areas:
&
Sensor design
&
Sensor location
&
Sensor wire nonhomogeneity
&
Thermal scatter at the cold junction reference
Temperature Sensors
Thermocouple type T (copper/constantan) is the
most commonly used thermocouple for temperature
measurements in validation applications due to its high
accuracy and low cost. Temperature measurement is
affected by several ambient conditions, that is why the
ASTMManual on The Use of Thermocouples in Tempera-
ture Measurement, Series MNL 12, 1993 (18) has the
following statement on its rst page:
Regardless of how many facts are presented herein
and regardless of the percentage retained, all will be
for naught unless one simple important fact is kept
rmly in mind. The thermocouple reports only what it
senses. This may or may not be the temperature of
interest. Its entire environment inuences the thermo-
couple and it will tend to attain thermal equilibrium
with this environment, not merely part of it. Thus, the
environment of each thermocouple installation should
be considered unique until proven otherwise. Unless
this is done, the designer will likely overlook some
unusual, unexpected, inuence.
Calibration
The measuring chain shall be calibrated prior to and after
calibration runs. Adequate calibration equipment shall be
used when in calibration of the measuring chain. A
temperature transfer standard, traceable to a primary
standard, and a stable thermal source are required to
perform calibration of the measuring chain. The
combined expanded uncertainty of all components of
the measuring chain, temperature transfer standard and
stable thermal source shall be determined by a competent
calibration facility and documented to be adequate for the
validation of the process equipment.
Measurement standards are classied based on
metrological qualities.
&
The primary standard has the highest metrological
qualities and is accepted without reference to other
standards of the same quantity. Primary standards are
normally kept in national measurement laboratories
and designated as national standards.
&
The secondary standard has its value assigned
by comparison with a primary standard of the same
quantity.
&
The transfer standard is used for the comparison of
standards of the same quantity.
This means that the validation study has to have
documented evidence that the measuring chain was
in calibrated state before and after the validation
study was performed. The documented procedures
shall describe the criteria chosen for the calibration
procedures. This should be backed by historical docu-
mentation that gives the rationale for the procedure. The
regulations require three consecutive successful vali-
dation runs for a successful validation. Based on the
companys risk assessment and risk management it
seems realistic to calibrate before the validation study
begins and to verify that the measuring chain remains in
calibration at the end of the third successful run.
Be patient. A frequent mistake in calibrating instru-
mentation is to takemeasurements andmake adjustments
before conditions have stabilized. Itmay takemuch longer
than expected for a system to become completely stable,
because thermal equilibration takes place exponentially
and the outputmay seem to be stable even though it is still
changing slowly. Automatic detection of stability to preset
stability criteria, offered bymodern calibration equipment
and software, is the most reliable and repeatable method
for stability determination.
Sensitivity
Long Termdrift
Temperature Coefficient
A/D Conversion Error
121C
Autoclave
T/C
121C
Sensor Design
Position
Non Homogeneity
Linearization
Thermal Scatter at Cold Junction Reference
Cold
Junction
Scanner A/D Conv. CPU
Figure 5 Random and systematic error
sources in the measuring chain.
104 II: SUPPORT AND UTILITY SYSTEMS
-
Temperature Transfer Standard
The accuracy of the transfer standard must be better
than that of the instrument being calibrated. This would
seem obvious, but it is amazing how often a voltage
calibrator is used that has a greater error than the
system being calibrated. It is important to recognize
that the accuracy of the calibration can be no better than
the standard used, and it is a mistake to change the
adjustment of a measuring system if it is already more
accurate than the standard.
The characteristics of the transfer standard must
have been determined by a procedure that is traceable to
accepted primary standards. In the U.S.A., NIST is the
accepted source of primary standards. The transfer
standards need to be calibrated by NIST relative to
their primary standards or by a qualied Standards
laboratory relative to standards calibrated by NIST.
In either case, the test results and test numbers should
be known so the calibration procedure can be traced to
the primary standards.
The transfer standard must be independent of the
measuring system. Because the output of a thermo-
couple depends on the entire circuit, it is not a
desirable transfer standard. An RTD is a device that
indicates changes of temperature by a change of
resistance. Because the resistance of an RTD is only a
function of its temperature, and the resistance can
be measured independently of the system being
calibrated, RTDs are ideal temperature transfer
standards.
The transfer standard must be stable in shipment
and tolerate other handling. As its name implies, the
purpose of the transfer standard is to transfer a
measured characteristic from one laboratory to
another. The characteristics of the standard must be
the same when received from NIST as when it was
calibrated relative to their standards. Liquid in glass
thermometers may be damaged or develop small voids
in the liquid during shipment, and therefore are not
reliable temperature transfer standards. RTDs are sensi-
tive devices that maintain their characteristics only
with careful handling and shipment.
Stable Thermal Source
Signicant random calibration errors that can occur in a
dry block temperature reference:
&
Transfer calibration
&
Stem conduction
&
Uneven heat transfer
&
Immersion depth
&
Well inserts not used
&
Stability
&
Time needed for stabilization
Reference Error, Using Thermocouples
When calibrating a thermocouple T
1
, against an
RTD transfer standard T
2
, a key contribution to error is
the difference in temperature between these devices
when placed in the reference (Fig. 6). This difference
is called transfer calibration error and is potentially the
largest contribution to calibration errors in dry block
references.
Transfer calibration error contains two components:
1. Stem conduction error, which cools the thermocouple
tip (Fig. 7).
2. Uniformity of the reference wells relative to the
standard well.
A dry block with one common large diameter well
is not suitable for calibration of a measuring chain using
multiple thermocouples (Fig. 8). The stem conduction
will cause heat losses that are greater than the heat
radiated from the walls of the single well. Figure 8
illustrates this inadequacy.
A dry block with smaller diameter wells and inserts
provides closer thermal coupling between well walls and
sensors under calibration, minimizing the effect of stem
conduction and transfer calibration error. A dry block
designed for maximum transfer calibration accuracy has
small diameter wells with inserts that t the size of the
sensors under calibration (Fig. 9).
Temperature
Measurement
Standard
Thermocouple
Temperature
Reference
T
1
T
2
Figure 6 Transfer calibration error is the temperature
difference between thermocouple tip (T
1
) and the measurement
standard (T
2
).
7: CALIBRATION AND METROLOGY 105
-
Calibration Procedure
Calibration of the measuring chain (Fig. 3) should be
done with the sensors connected to the data logger/
recorder and installed in the sterilizer via the feed
through and out through the open sterilizer door.
The sensors should be inserted into the temperature
reference bath or dry block, just outside the open steri-
lizer. Calibration should be performed prior to a
validation study and a calibration check should be
performed at the conclusion of the validation.
Prestudy Calibration
A two-point calibration should be used with calibration
points bracketing the sterilization temperature for the
process under validation, e.g., 1008C and 1308C, and
calibration checkpoint should be, e.g., 1218C, between
the two calibration points to verify the calibration.
Poststudy Verification
A two-point comparison between the temperature stan-
dard and the temperature sensors should be performed to
verify that the calibration of the measuring chain is intact.
The calibration should be documented, to provide
evidence that the temperature of the reference, the
transfer standard, and the sensors were stable before
the determination of calibration correction values. The
calibration documentation should include data on the
deviation between the temperature standard and each
temperature sensor before and after calibration. To ensure
traceability, the documentation must list the calibration
parameters and equipment including serial numbers and
last calibration dates.
SUMMARY
Validation of thermal sterilization processes requires
accurate temperature measurements to provide reliable
results. In order to ensure measurement integrity it is
necessary that the validation personnel has adequate
training and well-dened processes to follow.
Risk assessment and risk management are now
mandatory for processes that are used for manufacture
or production of products with critical tolerances.
The result of the risk assessment should serve as a basis
for denition of process tolerances and the corresponding
measurement tolerances for the process control system.
Appropriate measurement tolerances vary by appli-
cation, while a moist heat sterilization process needs to
be measured toG0.58C a depyrogenation process would
be adequately measured with a tolerance of G1.08C.
Internationally accepted good metrological stan-
dards and process control engineering practice call for
the application of TUR between the equipment under
calibration/validation and the calibration equipment
itself. For most applications the minimum TUR should
be 3:1 and preferably higher (Figs. 12). The higher
the TUR the more expensive calibration equipment has
to be used. This is an area where risk assessment and
risk management is used to determine the level of
Sensor
Under Calibration
Heat Loss Through
Copper Wire
Air Gap Between
Sensor and Well
Transfer Standard
Figure 7 Stem conduction causes heat loss and generates
calibration error.
Top View
Side View
Figure 8 Dry well with large diameter wells.
106 II: SUPPORT AND UTILITY SYSTEMS
-
compromise needed to balance the level of acceptable risk
versus the cost of more sophisticated calibration equip-
ment. According to ISO 17025 and ISO 10012
the management is responsible for the determination
and justication of the balance of risk versus cost.
Calibration/verication of the sensors monitoring
and controlling the process is as critical as the cali-
bration of the sensors used for validation of the
process. Each individual measuring chain has to be
calibrated/veried against a transfer standard, i.e.,
sensor, wiring, and measurement system, to at least the
same expanded uncertainty as required for the calibration
of the validationsystem.
The pressure sensor in the autoclave must be
calibrated in place under standard operating condi-
tions. A two-point comparison between the installed
pressure sensor and a temporarily connected pressure
transfer standard, traceable to a national standard,
should be performed. Based on the comparison, zero
and span adjustments are done on the installed
pressure sensor.
REFERENCES
1. FDA Draft Guidance for IndustryQuality Systems
Approach to Pharmaceutical Current Good Manufacturing
Practice Regulation, September 2004, U.S. Department of
Health andHumanServices FoodandDrugAdministration,
Center forDrug Evaluation andResearch (CDER) Center for
BiologicsEvaluationandResearch (CBER),Center forVeteri-
nary Medicine (CVM), Ofce of Regulatory Affairs (ORA).
2. FDA Guidance for IndustrySterile Drug Products
Produced by Aseptic ProcessingCurrent Good Manufac-
turing Practice, September 2004, U.S. Department of Health
and Human Services Food and Drug Administration
Center for Drug Evaluation and Research (CDER) Center
for Biologics Evaluation and Research (CBER) Ofce of
Regulatory Affairs (ORA).
3. FDA Pharmaceutical CGMPs for the 21st CenturyA Risk-
Based ApproachFinal Report, September 2004, Depart-
ment of Health and Human Services, U.S. Food and Drug
Administration.
4. ANSI/AAMI/ISO 13485:2003 Medical DevicesQuality
management systemsRequirements for regulatory
purposes. Association for the Advancement of Medical
Instrumentation, Arlington, VA.
5. ANSI/AAMI/ISO 14971:2002/A1:2003Application of
risk management to medical devices. Association for the
Advancement of Medical Instrumentation, Arlington, VA.
6. ISO 9001:2000 Quality management systemsRequire-
ments. International Organization for Standardization,
Central Secretariat, Geneva, Switzerland.
7. ISO 10012:2003 Measurement management systems
Requirements for measurement processes and measuring
equipment. International Organization for Standardization,
Central Secretariat, Geneva, Switzerland.
8. ISO/IEC 17025:2005 General requirements for the compe-
tence of testing and calibration laboratories. International
Organization for Standardization (ISO) and International
Electrotechnical Commission, Geneva, Switzerland.
9. ISOGuide to the Expression of Uncertainty inMeasurement
(GUM), issued by BIPM, IEC, IFCC, ISO, IUPAC, IUPAP.
International Organization for Standardization, Central
Secretariat, Geneva, Switzerland, 1995.
10. ANSI/NCSL Z540-2-1997 American National Standard for
Expressing UncertaintyU.S. Guide to the expression of
Uncertainty in Measurement. National Conference of Stan-
dards Laboratories, Boulder, CO.
11. ANSI/ISO/DIS 17665:2004 Sterilization of health care
productsMoist heatDevelopment, validation and
routine control of a sterilization process for medical
devices. International Organization for Standardization,
Geneva, Switzerland, 2004.
Top View
Transfer Standard
Well with Insert and
Three Thermocouples
Well with Insert
for Three T/Cs
Well with no
Insert
and One T/C
Side View
Figure 9 View of dry block with one large well and no insert.
7: CALIBRATION AND METROLOGY 107
-
12. FDA, Good Laboratory Practice for Nonclinical Laboratory
Studies, Title 21, Part 58.
13. FDA, Current Good Manufacturing Practice for Finished
Pharmaceuticals, Title 21, Part 211, Code of Federal
Regulations (U.S.A.).
14. FDA, Current Good Manufacturing Practice for Blood and
Blood Components, Title 21, Part 606, Code of Federal
Regulations (U.S.A.).
15. FDA, Subchapter HMedical Devices, Title 21, Part 820,
Quality System Regulation, Code of Federal Regulations
(U.S.A.).
16. http://www.fda.gov, 2006
17. European Good Manufacturing Practice, EDURALEX
Volume 4Medicinal Products for Human and Veterinary
Use, August 2004 (update of 1998 edition), European Com-
mission, Enterprise DirectorateGeneral Pharmaceuticals:
Regulatory Framework andMarket Authorizations. http://
ec.europa.eu/enterprise/pharmaceuticals/eudralex/
homev4.htm, 2006
18. ASTM Manual on the Use of Thermocouples in Tempera-
ture Measurement, Manual Series MNL 12, American
Society for Testing and Materials 1916 Race Street, Phila-
delphia, PA, 1993.
108 II: SUPPORT AND UTILITY SYSTEMS