catVTM. Controled atmosphere Vibrating magnetometer. Sample environment. Argon output. Sample Length 12mm Max Diameter 10mm Max. Motion Fosc = 8Hz Amp =15.84mm. Z applied Controlled field 0 to+-150µT. Argon input. Thermocouple. Signal. Moving Sample Holder. Coils Z axis - PowerPoint PPT PresentationTRANSCRIPT
catVTM
Controled atmosphere
Vibrating
magnetometer
Sample environment
Argon input
HeaterRoom to 700°C Cooler
Coils Z axisMagnetic momentSensor
Thermocouple
X Y CoilsField canceled
MotionFosc = 8HzAmp =15.84mm
Mumetal screens
Argon output
Z applied Controlled field0 to+-150µT
SampleLength 12mm MaxDiameter 10mm Max
Signal
MovingSampleHolder
Preparing the Experiment
A succession of profiles predifined by the user:
what is a profile ?
one ramp caracterized by:
heating or cooling rate.
Field to be applied during the ramp.
the dwelling temperature to reach.
Field to be applied at this stage
Possibility to link as many profiles as needed.
Heating or cooling rate °C/minRamp field µTDwell Temperature °CDwell field µTDwell time min
Dwelling time : 0 to hoursDwell Field
Experiment: One or Many Profiles
Profiles.txt 7 0 100 50 1010 50 200 0 5.
One File
Time
Temperature
Running the experiment
• A labview program runs automatically the profiles.txt file
• Every profile (one line in profiles.txt ) measurement is saved in one file.dat
• Possibility to modify the Profiles while the experiment is running by removing or adding line(s)