chapter 40 radiation protection procedures alara and occupational exposure

84
Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Upload: marlee-elvidge

Post on 14-Dec-2015

226 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Chapter 40 Radiation Protection Procedures

ALARA and Occupational Exposure

Page 2: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

ALARA

• ALARA stands for As Low As Reasonably Achievable. It is the basic principle of radiation protection procedures.

• There is much that we can do to keep exposure to the patient and the operator as low as possible.

• The chiropractor is unique in the fact that you can perform radiography and refer your patients for other types of examinations.

Page 3: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Occupational Exposures

• In radiologic technology, 95% of the occupational exposure comes from fluoroscopy and mobile radiography.

• Neither would be used in your office so the worst case scenario is that you would receive 5% of the exposure that a technologist would receive.

Page 4: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Occupational Exposures

• During radiography, the operator should be behind a protective barrier. – These barriers are usually considered as secondary

barriers so protection would be from tube leakage and scatter from the patient. The tube should never be pointed toward this type barrier.

– If the barrier can have the tube angled toward the barrier. It must be a primary barrier.

– Staying behind the barrier effectively eliminates the source of occupational exposure if the shielding is adequate.

Page 5: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Occupational Exposures

• Medical Imaging Exposures– Fluoroscopy: All personnel will wear protective apron.

If extremities get into the beam lead gloves can be worn.

• The radiologist will usually be close to the machine during fluoroscopy so their exposure will be higher than that the technologist. Aprons between the Image intensifier and Bucky Slot covers reduce radiologist exposure.

• The technologist should stand as far away from the table as possible during the exam and move closed only when necessary.

• The radiologist will use short burst of exposure and keep the exposure time as short as possible. The 5 minute clock timer will alarm when 5 minutes of fluoroscopy has been used.

Page 6: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Occupational Exposures

• Medical Imaging Exposures– Mobile radiography:

• The technologist must wear a lead apron during mobile plain film or fluoroscopy examinations.

• An apron must be assigned to each portable machine.

• The exposure cord for portable radiographic machines must be 2 meters long to maximize distance from the tube during exposures.

Page 7: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Occupational Exposures

• Radiology Ancillary Staff– Assuming the rooms are adequately shielded,

the receptionist, file room and darkroom staff should not receive any occupational exposure.

– Radiology ancillary staff should not be used to hold patients during radiography.

Page 8: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Occupational Radiation Monitoring

• Occupational Radiation Monitoring is required if there is any likelihood that an individual will receive more than 1/10 of the recommended dose.

• With just plain film radiography, monitoring may not be required as long as the operator stays in the control booth during all exposures.

• There are some exams such as stress views of the ankle where the operator would be in the room with the patient. If this is done, monitoring would be necessary.

• If the operator ever holds a patient monitoring would be necessary.

Page 9: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Occupational Radiation Monitoring

• Occupational radiation monitoring offers no protection against exposure. It merely records the exposure received.

• If needed, find a certified laboratory to process the dosimeters.

• Types of monitors– Film badges– TLD– OSL

Page 10: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Film Badges

• Film badges have been used since the 1940’s and are still used today.

• Exposures below 10 mR are not measured on the film.

• Filters along with the window in the badge allow estimation of the energy of the exposure.

• The must be worn with the proper side to the front.

• They are typically worn on the collar so they would remain outside the lead apron.

Page 11: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Film Badges

• Advantages– Inexpensive– Easy to handle and process– Reasonably accurate

• Disadvantages– Can not be reused– Sensitive to heat and humidity– Must be changes monthly

Page 12: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

TLD

• TLD has several advantages over film badges.– Not sensitive to heat or humidity– Measure exposures to 5mR More sensitive

and accurate.– Can be changed quarterly instead of monthly

• Disadvantages– Cost but changing badges less frequently

than monthly eliminates cost problem.

Page 13: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Optically Stimulated Luminescence

• All of the advantages of the TLD over film badges plus:– Can be re-read to confirm exposure– More accurate than TLD

Page 14: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Where to wear the monitor

• The whole body badge is typically worn at collar level so it can be outside the lead apron.

• Fetal monitoring badges used during pregnancy are worn at waist level under the apron.

• Hand or finger TLD’s are worn on the extremity.

Page 15: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Occupational Radiation Monitoring Reports

• State and federal regulations require that the results of the occupational radiation monitoring program be recorded in a precise fashion and maintained for review.

• Specific information is required to be on the report including current and cumulative exposure.

• Each site of monitoring must be identified separately.

• There will also be a control monitor to measure the background exposure during transport, handling and storage.

Page 16: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Occupational Radiation Monitoring Reports

• The supplier of the badges must know the type of radiation for proper calibration of the equipment.

• The badges are control are shipped back to the supplier in a timely manner.

• For lost or damaged badges, a health physicist will estimate the exposure.

• The annual exposure is discussed with each worker and receipt of the information is documented. Monthly reports may be posted but care must be taken with sensitive information.

Page 17: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

• Lead apron used for operator or patient protection must be the equivalent of 0.5mm of lead.

• They must be worn when in a room during the exposure or during fluoroscopy.

• Half aprons are effective means to provide gonad protection during radiography.

• 0.25 mm of lead aprons should be avoided as they only attenuate 66% of the beam at 76 kVp.

Protective Apparel

Page 18: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

• Aprons used in interventional radiology should be a wrap around type. Thyroid shields may also be worn.

• Lead gloves can be worn when the hands are in the beam.

• Aprons must be stored on specially designed racks or laid flat on the floor. They are never folded.

• Aprons are tested annually for cracks or holes in the lead, usually by fluoroscopy.

Protective Apparel

Page 19: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

• During fluoroscopy radiologic technologist should stand as far as possible from the machine. – Standing behind the radiologist offers added

protection.– If you must be in the room, position your body

as far away from the primary beam as possible.

Position

Page 20: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Patient holding

• Many patients will find the x-ray examination to be physically demanding. Some may not be capable of staying in position.

• This is a particular challenge for weight-bearing radiography. Mechanical supportive devices are limited for erect studies.

• If you have a radiographic table, the patient may be examined recumbent. Sponges may be used as supportive devices.

Page 21: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Patient holding

• Radiology or office staff should never hold a patient. Family or friends may be called upon to assist the patient.

• The person assisting the patient must wear a lead apron and if their hands will be in the beam lead gloves.

• Position the person as far away from the primary beam as possible.

• Since the person holding the patient may be a parent, make sure they are not pregnant.

Page 22: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Reducing Unnecessary Patient Dose

• As a doctor, you have the responsibility to determine if the radiography is necessary and justified.

• There are more practice guidelines available every year to assist in determining if the examination will yield necessary diagnostic information.

• There are many examinations that are performed knowing that they will yield little helpful information so they in no way justify the patient radiation dose.

Page 23: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Reducing Unnecessary Patient Dose

• Check to see if the patient has previous examinations that may make the new examination not necessary.

• You may be sued if you don’t take films and the treatment plan fails because you missed something the films would show.

• The yield of information must be greater than the risk of radiation exposure.

Page 24: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Reducing Unnecessary Patient Dose

• Routine x-ray examinations should not be performed.

• Used the most accurate tests to confirm or rule out your working diagnosis.

• Consider using MR instead of CT

Page 25: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Repeat Examinations

• One area of unnecessary patient exposure is repeated x-ray examinations. Past estimates of frequency has been as high as 10% but they should normally not exceed 5%.

• Most of the retakes are of the lumbar spine, abdomen and thoracic spine.

• Most retake are due to the exposure factors being incorrect resulting in an over exposed or under exposed film. Proper measurement are important.

Page 26: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Repeat Examinations

• Positioning errors account for about 25% of retakes. Proper training and practice is important to fine tune positioning skills.

• Motion causes about 11% of retakes so proper patient communication during the exam is important.

• But do not be afraid to retake a poor quality film. If you can not see a problem makes it likely you will miss it. Poor quality exams are never justified.

Page 27: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Radiographic Technique

• Use as high kVp as possible to get adequate contrast and reduce patient exposure.

• Collimate the beam to slightly smaller than film size or the area of interest, whichever is smaller.

• Use the fastest-speed screen-film combination consistent with the nature of the examination.

Page 28: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Positioning

• When taking films with the patient seated, do not allow the gonad to be in the primary beam. Position the patient lateral to the beam.

• For female patients turn the patient PA to reduce breast and gonad exposures when possible.

Page 29: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Patient shielding

• Some form of patient shielding should be used on all patients able to reproduce.

• All children should have shielding.

• Pre-menopausal women should be shielded except when the shield would interfere with the examination.

• Men should be shielded beyond 50 years.

Page 30: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Patient shielding

• Patient shielding includes contact shields and shadow shields.

• Contact shields are placed on the patient and include aprons, the heart shaped filter and the bell.

• Shadow shields are placed between the patient and the tube. Here we attach it to the tube.

Page 31: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Patient shielding

• Shielding must be used when the gonads lie in or near the useful beam and when it does not interfere with obtaining the required diagnostic information.– Accurate placement is extremely important.

Repeated examinations can result form improper placement of the shield.

• Proper patient positioning and collimation should not be relaxed when gonad shields are in use.

Page 32: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Ten Commandments of ALARA

1. Understand and apply the cardinal principles of radiation control: time, distance and shielding.

2. Do not allow familiarity to result in a false security.

3. Never stand in the primary beam.

4. Always wear protective apparel when not behind a protective barrier.

5. Always wear a radiation monitor and position it outside the protective apron at collar level.

Page 33: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Ten Commandments of ALARA

6. Never hold a patient during radiographic examinations. Use mechanical restraining devices when possible. Otherwise, use patients or friends to hold the patient.

7. The person holding the patient must wear protective apron and if possible, gloves.

8. Use gonadal protective on all people of childbearing age when it will not interfere with the examination.

Page 34: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Ten Commandments of ALARA

9. Examinations of the pelvis or lower abdomen of a pregnant patient should be avoided whenever possible, especially during the first trimester.

10.Always collimate to the smallest field size appropriate to the examination.

Page 35: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Chapter 31 Quality Control

• Two areas of activity are designed to ensure the best possible image quality with the lowest possible exposure and minimum costs. – Quality Assurance deals with people– Quality Control deals with instrumentation

and equipment.

Page 36: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Chapter 31 Quality Control

• Two areas of activity are designed to ensure the best possible image quality with the lowest possible exposure and minimum costs. – Quality Assurance deals with people– Quality Control deals with instrumentation

and equipment.

Page 37: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Ten Step Approach to Quality Assurance

1. Assign responsibility

2. Delineate scope of care

3. Identify aspects of care

4. Identify outcomes that effect the aspects of care.

5. Establish limits of the scope of assessment.

Page 38: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Ten Step Approach to Quality Assurance

6. Collect and organize data.

7. Evaluate care when outcomes are reached.

8. Take action to improve care

9. Assess and document actions

10.Communicate information to organization-wide QA Program

Page 39: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

QA Projects

• Things that QA can evaluate includes– Scheduling of patients– Instructions given to patients– Wait times in the office– Interpretation of films– Retake analysis– Record accuracy

Page 40: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

QA Program

• Quality Assurance deals with people and processes used to complete tasks.

• QA involves training and record keeping.• As the owner of the equipment, you will be

responsible for your radiology services.• The State of California Department of

Radiologic Health established the Standards of Good Practice that is the foundation of QA and QC in radiography.

Page 41: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

QA and QC Requirements

• Degree of requirements vary by state. California and New York have very tight standards for quality control of the radiographic and processing equipment.

• We are required by statue to teach QA and QC in the radiology program. It is covered in detail in 9th Quarter. My textboook covers QC in detail.

Page 42: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Quality Control

• An acceptable QC program has three steps:

1. Acceptance Testing

2. Routine performance monitoring

3. Maintenance

Page 43: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Acceptance Testing

• The x-ray machine, cassettes and film processor or digital system are the largest capital expense you may experience.

• It makes economic sense to make sure that the equipment meets the performance standards.

• It is recommended that a third party such as a health physicist do the testing.

Page 44: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Acceptance Testing

• Areas that should be tested include on the x-ray machine:– Shielding of Room– Focal spot size– Calibration of mA, timer or mAs – Calibration of kVp– Linearity of exposure– Beam alignment– Grid centering– Collimation– Filtration (HVL)

Page 45: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Acceptance Testing

• Areas that should be tested on the x-ray cassettes:

• Screen contact

• Screen speed

• Light leaks

• Light spectrum matching

Page 46: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Acceptance Testing

• Areas that should be tested on the x-ray film processor:

• Developer temperature

• Replenishment rates

• Travel time

• Water flow

• Hypo retention

Page 47: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Quality Control

• The acceptance testing ensures that the machine was installed and calibrated properly.

• The performance may drift or deteriorate over time. Consequently, periodic testing is required to monitor the performance.

• With the exception of film processing most testing is annual or semiannual.

Page 48: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Quality Control

• After a major repair, the machine should be retested to ensure that it was repaired properly.

• When the testing shows that the machine is not performing properly, service or preventive maintenance is required.

• Manufactures establish recommended preventive service schedules. When these are followed many repairs become unnecessary.

Page 49: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Radiographic Quality Control

• Areas of concern in x-ray machine– Focal Spot Size will impact spatial resolution– Filtration will impact patient exposure– Collimation will impact patient exposure– kVp calibration will impact image quality and

exposure.– Exposure timer accuracy will impact image

quality and exposure

Page 50: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Radiographic Quality Control

• Areas of concern in x-ray machine– Exposure linearity will impact exposure and

image quality– Exposure reproducibility will impact exposure

and image quality.– Alignment of tube and image receptor will

impact exposure and image quality.

Page 51: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Focal Spot Testing

• When the machine is installed or the tube is replaced, the focal spot size should be measured. Then annually thereafter.

• A pin hole camera, star test pattern or line pair test tool.

• As the tube ages, the focal spot tends to grow and spatial resolution is lost.

Page 52: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Filtration

• The filtration is measured but determining the half value layer of the beam at specific exposure levels. Minimum filtration is 2.5mm aluminum.

• As a tube ages, tungsten will plate the x-ray port and increase filtration. This can cause technique problems. Inadequate filtration will significantly increase patient exposure.

Page 53: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Collimation

• If the collimation is misaligned, intended anatomy can be missed.

• It can be tested in many ways from using pennies to using test patterns.

• Misalignment can not exceed ± 2% of the SID.

• It is tested semiannually and after the replacement of the collimator lamp.

Page 54: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

kVp Calibration

• In diagnostic radiology, any change will impact patient exposure. A variation of about 3% will impact contrast and image density.

• Can be tested with filtered ion chambers, filtered photodiodes or even a cassette with calibrated filters.

• Tested annually.

Page 55: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Exposure Timer Accuracy

• The exposure time is the responsibility of the operator. It will impact the density of the image and spatial resolution.

• Tested with an ion chamber, multi-meter internally or even a spinning top.

• Exposure time must be within ±5% for exposure times greater than 10 ms and ±20% less than 10 ms.

Page 56: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Exposure Linearity

• Many combinations of mA and time will produce the same mAs value. The ability of the machine to produce a constant level of exposure with various combinations of mA and time is called exposure linearity.

• Can be tested with a step wedge and densitometer or rate meter.

• Should be within 10% for adjacent stations.

Page 57: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Exposure Reproducibility

• Any exposure using the same factors should produce the same level of density and contrast on the image.

• Sequential exposure should be reproducible to within ± 5%

• Can be tested with a rate meter or step wedge and densitometer.

Page 58: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Performance standards for x-ray equipment

Measurement Frequency Tolerance US Tolerance Ca

Filtration Annually ≥2.5 mm Al ≥2.7mm Al

Collimation Semiannually 2% of SID 2% of SID

Focal Spot Annual ± 50% ±50%

kVp Annual ±10% ±2 kVp between 60 & 100 kVp

Timer Annual ±5% > 10 ms

±20 % ≤20 ms

3 phase ± 5%

1 phase ± 10%

Linearity Annually ± 10% ± 10%

Reproducibility Annual ± 5% ±5%

Page 59: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Darkroom and Processing

• The development of the image is dependent upon the temperature of the developer, it’s concentration and how long the film is in the developer.

• The film is sensitive to variations in the environment and processing from the time it is manufactured until it is processed.

• Darkroom and Processor QC is the key process of Quality Control.

Page 60: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Processing

• Processor densitometry is performed daily before the first patient is exposed.

• A sensitometer is used to produce a step wedge image on the film that is evaluated with a densitometer.

• The densitometer reads the optical density of the processed image.

• A digital thermometer is used to test the chemical temperatures in the processor.

Page 61: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Processing

• Key densities on the processed film are measured and then graphed.

• Base plus Fog is measured on an area of unexposed film to check the darkroom environment.

• Speed is tested at the level of exposure that produces a density of 1.25OD

• Contrast is tested at the level that produced a density of 0.40 OD and one that produced a density of 2.20.

Page 62: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Processing

• By monitoring these densities, problems with film processing can be detected before image quality deteriorates.

• In 9th Quarter we will cover how to perform processor QC and problem solve.

Page 63: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Waste Records

• Since used fixer is classified as a hazardous waste material, it is important to maintain accurate records of usage and disposal.

• The extent of records vary by city, county and state. You are responsible for the proper disposal of the waste. Some regions include developer as hazardous waste.

Page 64: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Silver Recovery

• If the silver ions are removed from the fixer, it may be disposed of in the normal waste when diluted with water.

• There are two primary types of silver recovery systems.– Metallic replacement uses steel wool and

can recover 95% of the silver in the effluent– Electrolytic recovery passes direct current

through the solution and nearly pure metallic silver is deposited on the cathode.

Page 65: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Silver Recovery

• Old radiographic films and repeated films are retained for silver recovery. X-ray images can not be disposed of in normal trash.

• They also can not be used to clean the processor rollers.

• Waste recovery companies will either burn or chemically remove the silver from the film.

Page 66: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Performance standards for film processor and darkroom equipmentMeasurement Frequency Tolerance CA Tolerance US

Sensitometry Daily B+F: ±0.05

MD:± 0.10 OD

Contrast ±0.10OD

B+F:±0.08 OD

MD± 0.15 OD

Contrast ±0.15 OD

Safelight Semiannual < 0.05 OD in 2 minutes

n/a

Darkroom temp Monthly 70ºF ± 5º n/a

Darkroom humidity

Monthly 50% ± 10% n/a

Developer temp Daily ± 0.5ºF 2-3º

Replenishment Daily ± 5%

Transport Annual ±3% ±3%

Page 67: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Accessory QC

• The cassettes and screens are the area of chief concern. Problems with either will result in artifacts on the images and increased retakes.

• The screens need to be properly cleaned frequently. Mammography screens are cleaned daily. California recommends monthly cleaning.

Page 68: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Accessory QC

• Dirty screens produce white artifacts on the image.

• Multiple white artifacts indicate the need to replace the screens.

• Screen contact is tested semiannually. A problem with screens contact will cause a loss of resolution.

• As screens age, they loose speed so this is also tested.

Page 69: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Accessory QC

• The other important accessory is the gonad protection devices and lead aprons.

• Improper care of the apron can result in cracks and holes in the lead that reduces their effectiveness.

• Aprons and shields are tested semiannually. The easiest was to test them is with video-fluoroscopy but film can be used.

Page 70: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Performance Standards for Accessories

Measurement Frequency Tolerance CA Tolerance US

Screen contact Semi annual No problems detected

No problems detected

Aprons/shields Annually No holes No holes

Screen matching

Annually ±0.05 OD for all cassettes used

Screen cleaning Monthly Semimonthly

Page 71: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Record keeping

• All electromechanical devices need periodic service.

• There are three types of maintenance.– Scheduled maintenance such as processor

monthly or weekly service. It includes observing moving parts and lubrication.

– Preventive maintenance is planned service and replacement of parts at regular intervals before they fail at inopportune times.

Page 72: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Record keeping

– Non-scheduled maintenance is the worst type of service because it impacts patient service. It may also be very expensive.

– With proper scheduled and preventive maintenance, non-scheduled service can be minimized.

– All service schedules should meet manufacture recommendations.

• All service should be documented as part of the quality assurance program.

Page 73: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Retake Analysis

• Required part of a QA program in California.

• Evaluation includes– View repeated– Cause of the repeat

• Rate of retakes should be less than 5%. • Information can be gathered from the log

that the state mandates for patients being exposed to radiation.

Page 74: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Retake Analysis

• Done every three months using a relatively large sample of data to see trends in:– Type of examination being repeated.– Reasons for the repeated films.– Determine if additional training or review is

needed.– Determine if equipment service might be

required.

Page 75: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Repeat Analysis

• Your patient log can be designed to capture both films usage and repeated films.

• Data is gathered from log for analysis

• Repeated films can be put into two main categories:– X-ray Personnel Errors– Equipment malfunctions

Page 76: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

X-ray Personnel Errors

• Failure to measure patient.

• Use of improper technical factors (mAs, kVp or distance)

• Incorrect positioning

• Improper Collimation

• Improper use of accessories such as cassettes, grids or filters

Page 77: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

X-ray Personnel Errors

• Improper handling of exposed or unexposed films.

• Failure to clearly communicate to the patient breathing instructions and to remain still.

• Failure to observe patient during exposure.

Page 78: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

X-ray Equipment and Accessory Failure or Malfunctions

• Inaccurate calibration of kVp and mA.

• Inaccurate timer calibration

• Dirty or damaged cassettes

• Improperly labeled or damaged grids

• Malfunctioning collimator

• Improper film storage or processor function.

Page 79: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Reasons for Retake Films

• Over or under exposure accounts for over 50% of retakes nationally.

• Errors in positioning (25%)

• Patient motion 11%

• Processing errors 6%

• Wrong view, beam alignment,cassette screen or grid errors and artifacts.

Page 80: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Retake Films by Region

• Cervical Spine Exams = 7%

• Thoracic Spine Exams = 17%

• Lumbar and Abdominal X-rays = 40%

• Skull, Chest,Lower Extremities = 15%

• Majority or retake results in unnecessary exposure to gonads or blood forming organs of the body.

Page 81: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Daily Log Design

• Most regulatory agencies will require a log of patients having radiation exposure.

• Columns can be added to capture film usage, repeats, views repeated and reason.

• This data can be gathered and analyzed.

Page 82: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Retake Analysis

• In this example, most of the retakes were of the T-spine. Potential reasons include:– Improper use of filters– Incorrect

measurements– Faulty technique chart

Page 83: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

Retake Analysis

• The reason of each retake is recorded and percentages are computed to determine the overall rate and rate by reason.

• Less than 5% is ideal.

Page 84: Chapter 40 Radiation Protection Procedures ALARA and Occupational Exposure

End of Lecture

Return to Lecture Index